Workflow Set-up
in myExam Cockpit
Job Aid
2
Effective Date: 28 APR 2022| HOOD05162003269520
Unrestricted © Siemens Healthineers, 2022
Workflow Set-up in myExam Cockpit
3
Effective Date: 28 APR 2022| HOOD05162003269520
Unrestricted © Siemens Healthineers, 2022
Enable Workflow in myExam Cockpit
4
Effective Date: 28 APR 2022| HOOD05162003269520
Unrestricted © Siemens Healthineers, 2022
Characteristics in a Nutshell
1• AutoAlign
myExam Autopilot requires AutoAlign to define the region to be scanned.
− AutoAlign scout step must have an assigned AutoAlign region.
− Measurement step must have an assigned AutoAlign reference, and Copy Reference
cannot be used.
2• Automized Workflow
myExam Autopilot does not allow pauses for user confirmation of Slice Alignment
or Spine Labeling Verification (e.g., working man, or pauses).
3• Minimized User Interactions
In normal cases, the user interactions during workflows are minimized to:
− Image quality check after localizer and before exam finalization (optional step).
− Planning the transverse images.
4• Workflow Assistance
myExam Autopilot assists the user by automatically resolving situations that
typically require user interaction (e.g., SAR warnings).
− SAR Assistant, Charge Balance Model (CBM) Assistance and Boost Assistant must always
be active. 5
Effective Date: 28 APR 2022| HOOD05162003269520
Unrestricted © Siemens Healthineers, 2022
Add-ins and Interaction Steps
6
Effective Date: 28 APR 2022| HOOD05162003269520
Unrestricted © Siemens Healthineers, 2022
Decisions
1 Spine planning
Interaction step is supported by the Autopilot Spine
Planning add-in. This enables the user to visualize the
step in the Autopilot workline and to select the scan
region.
9
Effective Date: 28 APR 2022| HOOD05162003269520
Unrestricted © Siemens Healthineers, 2022
Set-up Rules Summary (1)
Below you can find the list of step-related features that are not compatible with myExam Autopilot workflows:
Feature Detail
Copy Reference The myExam Autopilot relies on AutoAlign and a fixed scanning area.
Copy References are not needed and are not supported in the workflow.
Working Man Working Man is not supported by myExam Autopilot.
Ensure the workflow doesn’t include a Working Man.
Pause step Pauses are not supported by myExam Autopilot.
You cannot include Pause steps in the workflow.
Interaction steps Interaction steps are not supported by myExam Autopilot.
They are available only to enable Image QA and Spine Planning steps.
Decision steps Decision steps for workflow decisions are not supported (except for Spine Planning).
Within myExam Autopilot, use only Patient View decisions.
MPR Planning MPR Planning & Assignment are currently not supported in myExam Autopilot. When
needed, these actions shall be performed in MR View&GO.
Manual Voice Manual voice commands are not supported. Use automatic voice commands.
commands
10
Effective Date: 28 APR 2022| HOOD05162003269520
Unrestricted © Siemens Healthineers, 2022
Set-up Rules Summary (2)
Below you can find a list of the protocol-related features that are not compatible with myExam Autopilot workflows:
Feature Detail
Inline Composing Inline Composing is not supported and cannot be used in myExam
Autopilot.
Set-n-Go Set-n-Go is not supported within myExam Autopilot.
Below you can find a list of the protocol assistants that are mandatorily required by myExam Autopilot workflows:
Feature Detail
SAR Assistant SAR Assistant must always be enabled in myExam Autopilot, for
automatic resolution.
Boost Assistant Boost Assistant must always be enabled in myExam Autopilot, for
automatic resolution (if Gradient Mode Boost is activated).
11
Effective Date: 28 APR 2022| HOOD05162003269520
Unrestricted © Siemens Healthineers, 2022
Set-up Rules Summary (3)
In the table you can find the list of additional rules required by myExam Autopilot workflows:
Feature Detail
AutoAlign (AA) A scout or localizer step must have an AA region assigned.
A measurement step must have an AA reference assigned.
Spine Planning Spine Planning interaction is required in myExam Autopilot, to enable a spine transverse
scanning decision and planning. This is supported by the Autopilot Spine Planning add-in.
Spine Planning step must follow at least one sagittal spine measurement step and the
spine sagittal step must be configured with a spine AutoAlign region.
In a workflow only one Spine Planning step is allowed and in the event of multiple
strategies, it shall be valid across strategies.
Spine scan mode This is the only workflow decision allowed in myExam Autopilot.
decision The decision name Spine scan mode decision is required and cannot be changed.
The Spine scan mode decision must directly follow the Spine planning interaction step in
the workflow.
12
Effective Date: 28 APR 2022| HOOD05162003269520
Unrestricted © Siemens Healthineers, 2022
Please note that the learning material is for training purposes only.
For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator
Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator
Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software
and hardware available at the time of the training.
The Operator Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions.
Please note: Some functions shown in this material are optional and might not be part of your system.
Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your
country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your
local Siemens Healthineers sales representative for the most current information.
The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be
liable for damages.
All names and data of patients, parameters and configuration dependent designations are fictional and examples only.
All rights, including rights created by patent grant or registration of a utility model or design, are reserved.