Mammo Digital
Installation Manual
Headquarters
Installation Manual
Table of Contents
1. INSTALLATION ................................................................................................................................................ 3
2. UNPACKING ........................................................................................................................................................................................ 5
3. DISASSEMBLE THE REAR COVER OF CONSOLE. ............................................................................................................................. 7
9. ELECTRICITY.......................................................................................................................................................................................18
BM-IME-MD 2 / 44
BEMEMS Mammo Digital
Installation Manual
1. Installation
BM-IME-MD 3 / 44
BEMEMS Mammo Digital
Installation Manual
- Plus Drivers
- Wrench Set (Six angels)
- Scissors
- Knife
- Cable Tie
- Six Angle Wrench Driver (7.5)
- Molding etc.
BM-IME-MD 4 / 44
BEMEMS Mammo Digital
Installation Manual
2. Unpacking
Console
Accessory Box
PC 1EA
Monitor 1EA
Detector 1EA
Calibration Phantom 1EA
Apron 1EA
Gonad Shield 1EA
Manual
BM-IME-MD 5 / 44
BEMEMS Mammo Digital
Installation Manual
2.1 Others (Cover for Console side and front, cover for monitor cable).
BM-IME-MD 6 / 44
BEMEMS Mammo Digital
Installation Manual
BM-IME-MD 7 / 44
BEMEMS Mammo Digital
Installation Manual
4. Main OP Setting
4.1 Please refer to the below (to move the backside cover)
BM-IME-MD 8 / 44
BEMEMS Mammo Digital
Installation Manual
BM-IME-MD 9 / 44
BEMEMS Mammo Digital
Installation Manual
5. PC Monitor connection.
BM-IME-MD 10 / 44
BEMEMS Mammo Digital
Installation Manual
BM-IME-MD 11 / 44
BEMEMS Mammo Digital
Installation Manual
6. PC Connection.
BM-IME-MD 12 / 44
BEMEMS Mammo Digital
Installation Manual
Detector Lan Cable should be connected into Graphic Card Lan Port at PC.
BM-IME-MD 13 / 44
BEMEMS Mammo Digital
Installation Manual
6.5 Internet Lan Cable Connection. (PC Main B/D Internet Port)
BM-IME-MD 14 / 44
BEMEMS Mammo Digital
Installation Manual
BM-IME-MD 15 / 44
BEMEMS Mammo Digital
Installation Manual
BM-IME-MD 16 / 44
BEMEMS Mammo Digital
Installation Manual
BM-IME-MD 17 / 44
BEMEMS Mammo Digital
Installation Manual
9. Electricity.
9.2 Electricity.
Be sure to check the rated voltage before connecting the power supply.
When installing the vehicle, check the charging status and voltage
BM-IME-MD 18 / 44
BEMEMS Mammo Digital
Installation Manual
After hearing the beep sound and lamp is on, then Main OP Power On.
BM-IME-MD 19 / 44
BEMEMS Mammo Digital
Installation Manual
BM-IME-MD 20 / 44
BEMEMS Mammo Digital
Installation Manual
BM-IME-MD 21 / 44
BEMEMS Mammo Digital
Installation Manual
11.1 Please refer to the below pictures. (Ex. Pinkview-DR Prime Detector)
11.2 Remove 4 pars as the below picture (Allen wrench) and Cable Connection.
BM-IME-MD 22 / 44
BEMEMS Mammo Digital
Installation Manual
Pinkview-DR Detector
BM-IME-MD 23 / 44
BEMEMS Mammo Digital
Installation Manual
BM-IME-MD 24 / 44
BEMEMS Mammo Digital
Installation Manual
Type: (a) PV-DR Pro (b) PV-DR Prime, Smart (c) PV-DR Plus
BM-IME-MD 25 / 44
BEMEMS Mammo Digital
Installation Manual
BM-IME-MD 26 / 44
BEMEMS Mammo Digital
Installation Manual
If X-ray images are taken in this state, grid lines may occur.
BM-IME-MD 27 / 44
BEMEMS Mammo Digital
Installation Manual
15. Choose the installation place. And make the last finish.
BM-IME-MD 28 / 44
BEMEMS Mammo Digital
Installation Manual
BM-IME-MD 29 / 44
BEMEMS Mammo Digital
Installation Manual
18. Test the movement and X-ray Exposure with ACE Phantom.
If No ACR Phantom, you can do with Calibration Phantom from the manufacturer.
BM-IME-MD 30 / 44
BEMEMS Mammo Digital
Installation Manual
19.1 Pinkview-DR software Execute (Operator: SUPERVISOR, pw: BMS) – Capital letters.
Operator as SUPERVISOR and login, CONFIG tab activated. Login as Guest, CONFIG tab NOT
activated.
CONFIG – Software – Default Setting – Operator List (put user name) – Add – PW input –
Memory.
BM-IME-MD 31 / 44
BEMEMS Mammo Digital
Installation Manual
CONFIG – Software – Default Setting – Hospital (put name address, phone no).
CONFIG – DICOM – DICOM Default Setting – AE Title, IP Address, Port Number Input.
Client AE: PinkView
BM-IME-MD 32 / 44
BEMEMS Mammo Digital
Installation Manual
20. IP Setting
BM-IME-MD 33 / 44
BEMEMS Mammo Digital
Installation Manual
21.1 Putting patient information (ID, NAME, others) - Std. Exam Click.
(Std. Exam: RCC LCC RMLO LMLO, Exam: Set the position to for image)
BM-IME-MD 34 / 44
BEMEMS Mammo Digital
Installation Manual
21.4 (Same process 21.1 to 21.2) BROWSER-> Select patient -> Click Send.
BM-IME-MD 35 / 44
BEMEMS Mammo Digital
Installation Manual
22.1 Pinkview-DR software Execute (Operator: system, PW: BMS) – Capital letters.
* Operator: Set as guest login – PW: N/A
File – Configuration – Operator Login – Setup – Operator List – Enter User Name – Add – Enter
BM-IME-MD 36 / 44
BEMEMS Mammo Digital
Installation Manual
BM-IME-MD 37 / 44
BEMEMS Mammo Digital
Installation Manual
BM-IME-MD 38 / 44
BEMEMS Mammo Digital
Installation Manual
23. IP Setting
BM-IME-MD 39 / 44
BEMEMS Mammo Digital
Installation Manual
24.1 New Patient – Putting patient information (ID, NAME, others) – Start Exam Click
or
BM-IME-MD 40 / 44
BEMEMS Mammo Digital
Installation Manual
24.3 After completion of exposure, select Browser → Select image and Click send
BM-IME-MD 41 / 44
BEMEMS Mammo Digital
Installation Manual
25. Cautions
25.1 At Manual mode, you can take x-ray image. But you must compress the target to get the x-ray
25.3 Maximum Compression force is 200N. On display, you find 210, 220. But more than
compression force shall display “----". No proper X-ray will be exposure. Please make the
compression force till 200N (Not more compression force)
25.4 Ex) Compression force is lower than the setting compression power. Move up and down
button on Sub OP. You set the maximum limit force level.
25.5 Safety function will work if you make more than 200N by manual. It is normal. Move the
compression paddle to the maximum lower position and move to the maximum high position.
25.6 For Pinkview-DR Plus, change the collimator exposure size to 18x24 for magnification image
exposure. (Normal Exposure Size: 24x30)
25.7 Test exposure ONLY with phantom. S09-01 Error may happen with metal material test with AEC
mode.
BM-IME-MD 42 / 44
BEMEMS Mammo Digital
Installation Manual
25.8 When Biopsy Mode is ON, automatic compression release will NOT work after exposure.
Please check the Biopsy mode is off when you do not use Biopsy mode.
25.9 When you rebooting the main system, please reboot Pinkview-Software also.
(Software should be executed ONLY AFTER the main system is completely ON)
BM-IME-MD 43 / 44
BEMEMS Mammo Digital
Installation Manual
BM-IME-MD 44 / 44