Professional Documents
Culture Documents
Service Manual
Contents
1 Important
2 Introduction
3 Description of functions
4 Disassembling and assembling
5 Service procedures
6 Troubleshooting
7 Preventive maintenance
8 Index
9 Revision history
10 Diagrams
1 Important
Service documentation for the SERVO-i Ventilator The System version number can be found in the Status
System consists of: window on the User interface. Make sure that the
version of the User’s Manual corresponds to the
• User's Manual including Cleaning and Maintenance
bullet System version.
manual. The User's Manual is an indispensable
complement to the Service Manual for proper The System SW version number can be found in the
servicing. Status window on the User interface. For functionality
enhancement, the latest released System SW version
• Service Manual
bullet is always recommended.
• Installation Instructions
bullet
• Spare parts information
bullet 1.1.2 Serial number
Polymeric material - total 9% Item 5: Changed approximately two and a half years
from manufacturing date.
• PA (Polyamide)
bullet
Item 6: Changed approximately every five years.
• POM (Polyoxymethylene)
bullet
Items 7 – 10: Changed when needed.
• SI (Silicone)
bullet
• TPE (Thermoplastic elastomer)
bullet Articles related to clinical applications, e.g. patient
tubing, Y sensor, nebulizer medication cups and NAVA
• PUR (Polyurethane)
bullet
catheters, not included in the list above.
• ABS (Acrylicnitrilebutadienstyrene)
bullet
• EPDM (Ethylenepropylenedienemonomer)
bullet
• PTFE (Polytetrafluoroethylene)
bullet
• FPM (Fluororubber)
bullet
• NBR (Nitrilerubber)
bullet
• PP (Polypropylene)
bullet
• PVC (Polyvinyl chloride)
bullet
SERVO-i Ventilator System, Service Manual 1-5
| 1 | Important |
Trickle
Mode Fast charging
charging
In operation 70 W 38 W
Standby 65 W 33 W
Off 35 W 6W
• Corrugated cardboard
bullet
• Stretch film of polyethylene, PE
bullet
• Shock-absorbing material of expanded
bullet
polyethylene, EPE, or expanded polypropylene,
EPP
2 Introduction
Table of Contents
2.1 Main units | 2-2
2.2 Optional equipment | 2-2
2.3 User interface | 2-5
2.4 Patient unit | 2-7
2.5 SERVO-i software structure | 2 -10
Gas trolley
• MR environment declaration
bullet
• MR environment kit – Installation Instructions
bullet
• SERVO-i – User's Manual
bullet
• Gas trolley – Installation Instructions
bullet MSync
Heliox
• MSync – User's Manual
bullet
• MSync – Installation Instructions
bullet
• SERVO-i – User's Manual
bullet • MSync – IT and security white paper
bullet
• Heliox adapter kit – Installation Instructions
bullet • Interface specification, MSync – Reference Manual
bullet
Holder
• Mirth Connect configuration in MSync –
bullet
Configuration Manual
Shelf base
The User interface can be rotated and tilted into a For further information regarding operation of the User
suitable position. Locking levers, mounting devices interface, refer to the User's Manual.
and some other items are shown in the illustration
below.
S30003
1. Display with touch screen. 12. PC Card slot with slot cover.
2. Fixed keys for immediate access to special 13. Control cable between User interface and Patient
windows. unit.
3. Main rotary dial. 14. Service connector.
4. Special function keys. 15. On/Off switch and switch cover, version
5. Direct access knobs. discontinued in production Q2 2007.
6. Mains power indicator (green). 16. On/Off switch and switch cover, version introduced
in production Q2 2007.
7. Standby indicator (yellow).
17. Panel holder
8. Start/Stop (Standby) ventilation key.
18. Locking screw, alternative mounting
9. Luminescence detector, adjusts display brightness
automatically. On User interface Type 1, the 19. Locking lever, rotation
detector is placed in the upper left corner. On User 20. Locking lever, tilting.
interface Type 2 and 3, the detector is placed 21. Serial number label. The serial number stated on
above the Fixed keys in the upper right corner. this label is the ID number of the User interface.
10. Loudspeaker grid. This serial number must be referred to when
11. Cable reel. ordering service, spare parts, etc for the User
interface.
a j f k g
S30004
(01) XXXXXXXXXXXXXX
(21) XXXXXX (11) XXXXXX
(01) XXXXXXXXXXXXXX
(21) XXXXXX (11) XXXXXX
S30006
1. Handle. 15. Module unit with six slots for connecting optional
2. Gas inlet for Air. The gas inlet for Air can also be modules, i.e. Battery modules, CO2 analyzer
used for the optional Heliox adapter. module, Edi module and Y sensor module.
3. Gas inlet for O2. 16. Nebulizer connector.
5. Mains power supply connector incl. fuses F11 and 18. Alarm output connector (previously optional,
F12. factory mounted as from Q2 2012).
6. Internal fan with filter. 19. Serial port for data communication (RS232)
(previously optional, factory mounted as from
7. Connector for external +12V DC power supply.
Q2 2012).
8. Fuse F1 for external +12V DC power supply.
20. Serial number label. The serial number stated on
9. Optional connector. this label is the ID number of the Patient unit. The
10. Control cable connector (User interface). serial number is also stored in the software
11. Serial port for data communication (RS232). memory as the System ID. This serial number
12. Expiratory outlet. must be referred to when ordering service, spare
parts, software options, etc.
13. Inspiratory section cover.
21. Battery lock.
14. Expiratory inlet.
SERVO-i Ventilator System, Service Manual 2-7
| 2 | Introduction |
1. PC 1772 Monitoring.
2. PC 1771 Control.
3. PC 1784/PC 2024 Expiratory channel with the two
connected PC 1781 Inspiratory and Expiratory
Pressure Transducers. PC 2024 replaces
PC 1784. Introduced in production Q2 2016.
4. Expiratory valve coil.
5. Module unit including PC 1775 Plug-and-play
back-plane.
6. AC/DC converter.
7. Internal fan.
8. Mains supply inlet.
9. PC 1778 DC/DC & Standard connectors.
10. PC 1785 Expiratory channel connector.
11. PC 1957 RS232/Alarm output connector.
g h j k A
S30007
S30008
14. Two gas modules, Air and O2, for regulation of the
inspiratory gas.
15. Connector muff.
16. Inspiratory pipe with housings for the O2 sensor/
cell and for the safety valve.
17. O2 cell incl. bacteria filter.
18. O2 sensor. Alternative to the O2 cell for oxygen
concentration measurement.
J K L H
19. Safety valve.
20. Temperature sensor (inside the connector).
21. Inspiratory pressure transducer tube incl. bacteria
filter, to connect the inspiratory pressure
transducer.
S30009
• Breathing BRE
bullet
• Monitoring MON
bullet
• Panel PAN
bullet
• System ID
bullet
• Edi
bullet
The Monitoring software controls all monitoring and The Edi software processes and filters the signals
alarm functions in the system, including trends of from the Edi Catheter, and transmits Edi and leads
measured values. Events, such as alarms and change data to the system.
of settings will also be logged.
The Edi software is stored in the Edi module. New
The Monitoring software is stored on PC 1772 software can be installed via a System software
Monitoring. The software must be re-installed if installation. If an Edi module is part of the system,
PC 1772 is replaced. Software related to Monitoring make sure that this module is connected during
is also stored in the O2 sensor. New software can be software installation.
installed via a System software installation.
The Edi software is executed by the microprocessor
The Monitoring software is executed by the in the Edi module.
microprocessor on PC 1772.
2.5.5 System ID
3 Description of functions
Table of Contents
3.1 General | 3-2
3.2 User interface functions | 3-2
3.3 Main back-plane | 3-4
3.4 Inspiratory section | 3-4
3.5 Expiratory section | 3-8
3.6 Pressure transducers | 3 -10
3.7 Control | 3 -10
3.8 Monitoring | 3 -11
3.9 Expiratory channel | 3 -12
3.10 Power supply | 3 -12
3.11 RS232/Alarm output connector | 3 -15
3.12 Optional equipment | 3 -16
Some features included on PC 1777 Panel are: For generation of sound, e.g. alarm. Connected to
P72 on PC 1777 Panel.
• SIMM (Single In-line Memory Module) mounted on
bullet
its connector P77. Memory type: SDRAM The loudspeaker generates different tones with
individual sound volumes. At startup and during
• PC Card Slot intended for connection/insert of a
bullet
Pre-use check the function of the loudspeaker is
PC Card. PC Cards are used to:
monitored by the microphone on PC 1777. During
- Download software into the different flash operation it is continuously monitored through current
memories situated on PC boards marked μP and sensing.
into the EEPROM on PC 1770 Main back-plane.
- Transfer patient and system data for further With the optional accessory Loudspeaker booster kit,
transfer to a computer. the alarm sound is amplified. Refer to section Optional
equipment.
- Service purpose.
For Type 3 it is estimated that the LED backlight will • PC 1780 Pneumatic back-plane.
bullet
maintain an acceptable brightness level during the
system's lifetime. 3.4.1 Gas modules – Air and O2
5. Flow transducer (Delta pressure transducer and Inspiratory valve temperature sensor
net)
The temperature of the supplied gas is measured by
6. Nozzle unit with valve diaphragm
the Inspiratory valve temperature sensor. This sensor
7. Inspiratory solenoid is situated in the gas flow.
8. Gas module key
The output signal from this sensor is used to
Each gas module is provided with an ID-PROM. The compensate for the gas density variations due to
ID information can be read by the system. temperature.
Note: The gas modules are factory calibrated. Each Supply pressure transducer
gas module must not be disassembled further than
The pressure of the supplied gas is measured by the
described in chapter Preventive maintenance.
Supply pressure transducer.
The gas flow through the gas module is regulated by The O2 cell is mounted in a housing on the Inspiratory
the Inspiratory solenoid via the Nozzle unit. pipe and is protected by a bacteria filter.
The key consists of a plastic guide mounted The O2 cell gives an output voltage proportional to the
underneath the module and a corresponding guide partial pressure of oxygen inside the Inspiratory pipe.
mounted in the Patient unit. At constant ambient pressure this output is
proportional to the O2 concentration in percent.
3.4.2 Connector muff
The life time of the cell is affected by the O2
The Connector muff connects the gas module outlets concentration. With a concentration (at the cell) in %
to the Inspiratory pipe inlet. and expected cell life time in hours the following
o o
applies at 25 C (77 F):
3.4.3 Inspiratory pipe
Expected cell life time = O2 Conc (%) x Time (h)
The Inspiratory pipe leads the gas from the Connector = 500 000 (%hours)
muff to the Inspiratory outlet.
The O2 cell is automatically calibrated each time a
The Inspiratory pipe comprises: Pre-use check is performed (if O2 is connected to the
system).
• Housing for the O2 sensor as well as housing and
bullet
locking lever for the O2 cell with its bacteria filter. If the system has continually been in use for a long
time, the measured O2 concentration may drop due
• Housing for the Safety valve.
bullet
to normal degradation of the O2 cell. This will activate
• Connection for measurement of inspiratory
bullet a nuisance alarm. For further information, refer to the
pressure.
User's Manual.
The pipe is provided with internal flanges with the
purpose to improve mixing of O2 and Air. An ID PROM is integrated into each O2 cell. Its ID
information and remaining lifetime can be read by the
The O2 sensor requires a changed design of the system.
Inspiratory pipe. The O2 sensor and the Inspiratory
pipe are equipped with a mechanical key to prevent Note: Pre-use check is recommended to use to
that the O2 sensor is mounted on wrong type of calibrate the O2 cell.
Inspiratory pipe. The O2 cell can be used on both
versions of the Inspiratory pipe.
The O2 sensor is mounted in a housing on the A Temperature sensor is integrated into the connector
Inspiratory pipe as an alternative to the O2 cell. on top of the O2 sensor/cell. This Temperature sensor
measures the temperature inside the Inspiratory
section.
The opening conditions for the safety valve are: 3.5.1 Expiratory cassette
OUTLET
INLET
The Ultrasonic flowmeter is a measuring device for Via a Bacteria filter inside the cassette, the Expiratory
the expiratory gas flow, using ultrasound technique pressure tube connects the cassette to the Expiratory
with two ultrasonic transducers/receivers. The pressure transducer. The filter and the connector are
measuring process is controlled from the main block integrated parts of the cassette. The filter protects the
PC 1784/PC 2024 Expiratory Channel. transducer on PC 1781 Pressure transducer from
contamination.
Expiratory valve
3.6.2 PC 1781 Expiratory pressure transducer This desired constant Inspiratory flow value is used
to generate the flow reference signals Insp Flow Ref 1
Function identical to PC 1781 Inspiratory pressure and Insp Flow Ref 2 with the same relation and same
transducer. handling as described above under 'Regulation of
Inspiratory flow…' except this occurs during expiration
time.
• MR environment considerations.
bullet Notes:
• The System software must be re-installed if
bullet
PC 1784/PC 2024 is replaced.
3.9 Expiratory channel
• MR environment considerations.
bullet
System overview, functional main block F.
If the internal DC supply voltage +12 V_Unreg drops The fan will start with half effect at approx. 33 °C
below 10 V, due to power supply failure, the power (91 °F) and with full effect at approx. 43 °C (109 °F).
supply source will automatically switch. The following When the temperature drops below approx. 37 °C
power supply source priority is used: (99 °F), the fan turns to half effect and when the
temperature drops below approx. 27 °C (81 °F), the
1. Mains power fan stops.
2. External +12 V DC supply (if connected)
The air inlet is protected by a filter that must be
3. Backup Battery modules
cleaned or replaced during the Preventive
Power supply selection is managed by: maintenance.
3.10.5 PC 1775 Plug-and-play back-plane The life span for the Battery module is calculated to
two and a half year from manufacturing date. Normal
Connects the Optional modules that are inserted in time for logistics and storage are included in this
the Module unit. calculation. The calculation corresponds thus to an
estimated operational time of two years. Manufacturing
PC 1775 also controls:
date (YearMonthDay) is printed on the battery label.
• Charging / discharging of the Battery modules.
bullet
• Switching between Mains power/External 12 V DC
bullet
and Battery module power supply.
Connection slots for 6 optional modules, i.e. Battery System software version V2.01.00 (or higher) includes
modules, CO2 analyzer module, Edi module and Y an improved monitoring of the battery status. This
sensor module. System software will, among others, monitor:
3.10.8 Control cable Preferably the NC contact should be used, as this will
detect any interruption in the alarm system or
This Control cable connects the Patient unit and the unintentional disconnection of the cable between the
User interface. The cable can be partly wound on the system and the ventilator. Furthermore, if both the NO
cable reel on the rear side of the User interface. and the NC-contacts are utilized, an even higher safety
level for unforeseen failures can be achieved. Note
Notes: that, in both cases, even an intentional disconnection
• The Control cable must only be connected or
bullet of the ventilator will trigger the external alarm.
disconnected when the system is switched Off.
Pin configuration and signal names in N67 – Alarm
• MR environment considerations (two ferrite
bullet output connector can be found in chapter Diagrams.
blocks mounted on the Control cable).
For further information, refer to the 'Alarm output
connector – Reference Manual'.
3.11 RS232/Alarm output connector To enable the Alarm output function, the following
System overview, functional main block A. parts are required:
Aerogen Pro and Aerogen Solo are devices intended The Compressor Mini is well insulated against noise
to aerosolize physician-prescribed medications, for and therefore does not cause disturbance when used
inhalation, that are approved for use with a general during operations.
purpose nebulizer.
The Compressor Mini is equipped with a standby
• Aerogen Pro: Reusable nebulizer unit.
bullet function. In the standby mode, the compressor will
start to deliver compressed air if the hospital central
• Aerogen Solo: Single use nebulizer unit.
bullet
gas supply fails.
The Aerogen Pro/Aerogen Solo are designed to
operate in-line with the ventilator circuit in acute and For further information, refer to separate Compressor
sub acute care environments. It operates without Mini documentation.
changing patient ventilator parameters.
3.12.4 CO2 analyzer module
To enable the nebulizer function, the following parts
are required: The CO2 analyzer module is an optional accessory
that is connected to the Module unit. There are two
• Aerogen Module.
bullet versions of the CO2 analyzer module:
• Nebulizer software. The nebulizer software is
bullet
individually created for each specific ventilator and • CAPNOSTAT III.
bullet
can only be installed on this ventilator. The
nebulizer function must be enabled in the
• CAPNOSTAT 5. Requires System version 7.1
bullet
3.12.2 Battery module The CO2 analyzer module is, via a cable, connected
to a Capnostat sensor mounted on an airway adapter
The Module unit allows up to six Battery modules. To
at the Y-piece. The sensor uses a solid state and IR
guarantee safe battery backup, always use at least
based optical system with no moveable parts. It
two batteries.
measures the difference between a reference light
For further information, refer to section Battery beam and one filtered for CO2 wavelength.
modules above in this chapter.
To enable the CO2 analyzer function, the following The Edi Catheter is placed into the esophagus to the
parts are required: gastric ventricle. The catheter receives the signals
from the diaphragm excitation. The Edi module filters
• CO2 analyzer module. The module occupies two
bullet the signals and passes them to the system which uses
plug-in slots. the signals for monitoring and regulation. It also
displays the Edi signal as a waveform on the User
• CO2 analyzer software. The CO2 analyzer software
bullet
is individually created for each specific ventilator interface.
and can only be installed on this ventilator. The CO2
The Edi module is available in a 50 Hz and a 60 Hz
analyzer function must be enabled in the version (to be used in 50 Hz alt. 60 Hz environment).
configuration software. A built-in filter removes interference from surrounding
• For CAPNOSTAT III: CO2 CAPNOSTAT III sensor
bullet equipment.
kit
If an Edi module is part of the system, make sure that
• For CAPNOSTAT 5: CAPNOSTAT 5 mainstream
bullet this module is connected during software installation.
CO2 sensor Software stored in the module will be checked/
• CO2 airway adapters. The CO2 airway adapter is
bullet updated during the software installation.
available as disposable or reusable.
If the Edi module is exposed to rough handling, e.g.
For electrical safety test of the module, refer to chapter dropped on the floor, an electrical safety test of the
Preventive maintenance section Safety inspection. module must be performed. For electrical safety test
of the module, refer to chapter Preventive maintenance
section Safety inspection.
3.12.5 Edi module (NAVA)
The Edi module is an optional accessory used during 3.12.6 Gas cylinder restrainer
NAVA ventilation. The Edi module requires one free
slot in the Module unit. The Gas cylinder restrainer is mounted on the rear
side of the optional Mobile cart. It is intended for two
To enable NAVA, the following parts are required: gas cylinders, maximum volume five litres each.
3.12.10 Humidifier and Humidifier holders The Mobile cart is designed for carrying the User
interface, the Patient unit and all required optional
Recommended active (heated) humidifier is Fisher & equipment.
Paykel MR 850. For further information, refer to
manufacturer's documentation.
• MR stickers
bullet The Servo Ultra Nebulizer is intended for administering
The complete procedure to classify a SERVO-i as MR nebulizing drugs to patients requiring mechanical
conditional is described in the MR environment ventilation or positive pressure breathing assistance
declaration. via an endotracheal tube or face mask/prongs.
Table of Contents
4.1 General | 4-2
4.2 Preparations | 4-2
4.3 PC boards and electrical components | 4-2
4.4 Information stored in the system | 4-4
4.5 Assembling guidelines | 4-5
4.6 User interface overview | 4-6
4.7 User interface - common parts | 4-8
4.8 User interface - Type 1 | 4 -15
4.9 User interface - Type 2 | 4 -19
4.10 User interface - Type 3 | 4 -23
4.11 Inspiratory cover | 4 -25
4.12 Inspiratory channel | 4 -25
4.13 Safety valve membrane | 4 -27
4.14 Gas modules | 4 -28
4.15 PC 1780 Pneumatic back-plane | 4 -29
4.16 Patient unit front cover | 4 -30
4.17 Expiratory cassette | 4 -31
4.18 PC 1785 Expiratory channel connector | 4 -34
4.19 Expiratory valve coil | 4 -34
4.20 PC 1771, PC 1772 and PC 1784/PC 2024 | 4 -35
4.21 AC/DC converter | 4 -36
4.22 Module unit including PC 1775 Plug-and-play | 4 -36
back-plane
4.23 PC 1778 DC/DC & Standard connectors | 4 -38
4.24 Internal fan | 4 -39
4.25 PC 1957 RS232/Alarm output connector | 4 -40
4.26 PC 1770 Main back-plane | 4 -41
4.27 Fixed Battery module | 4 -42
4.28 Control cable | 4 -43
Remove and insert the PC boards very carefully to Other technical errors may occur after a software
avoid damage to the connectors. installation. These errors will disappear after the first
restart.
Note: ESD sensitive components.The
PC boards contain components that
are highly sensitive to static electricity.
When working with ESD sensitive
components, always use a grounded
wrist band and a grounded work
surface. Adequate service tools must
always be used.
• PC 1771 Control
bullet
• PC 1772 Monitoring
bullet
• PC 1784/PC 2024 Expiratory Channel
bullet
• PC 1777 Panel
bullet
• O2 Sensor (PC 1873)
bullet
• Edi module (PC 1874)
bullet
Stored information PC 1770 PC 1771 PC 1772 PC 1777 PC 1781 PC 1784/ Expiratory O2 cell Edi
Main back- Control Monitoring Panel Pressure PC 2024 cassette O2 sensor module
plane transducer Expiratory (PC 1786)
channel
System software X X X X X X
(O2 sensor)
System info: X
§ Serial number
§ System version
§ Installed software
options
§ Operating time (total
and running hours)
§ Last preventive
maintenance
Configurations X
Date and time setting X
Logs X
Trends X
Recordings X
Pre-use check results X X X X X
including transducer result calibration calibration calibration calibration
calibrations
Barometer calibration X
Performed preventive X1
maintenance
Replaced expiratory X1 X
membrane counter
Replaced backlight X1 X
counter
Backlight operating time X
Language X
Expiratory cassette: X
§ Date of first use
§ Total operating time
§ Remaining membrane
capacity
O2 cell/sensor: X
§ Date of first use.
§ Estimated remaining
capacity (O2 cell)
1
Included in Service report as part of the Service log.
S30018
The system specifications allow unit operation also Lubricants should normally not be used when servicing
during patient transportation. Almost all parts of the the unit. If lubricants must be used, use only very small
User interface and the Patient unit assembled with amounts of grease with P/N 95 73 700. This special
screws and nuts are therefore tightened with a grease is O2 compatible.
specified torque and secured with threadlocking
adhesives. WARNING! Due to the increased O2
concentration in valves and gas
In order to maintain these specifications over time, it
channels, only special grease intended
must be ensured that after any service intervention
for use in high O2 concentrations, is
removed parts are re-assembled and secured
permitted.
according to instructions. Make sure to follow the
guidelines stated below.
PC 1777 Panel Replacement kit for PC1777A-D PC 1777, Type 2 PC 1777, Type 3
TFT Display TFT Display, Type 1 TFT Display, Type 2 TFT Display, Type 3
Backlight Backlight inverter, Type 1 Backlight inverter, Type 2 Not required in Type 3
inverter
Backlight lamp Backlight kit, Type 1 Backlight lamp, Type 2 Not required in Type 3
(white/gray cable)
Backlight Cable
Note: MR environment. On MR
environment approved units, an EMC
shield is mounted between the Touch
screen and the TFT display. Refer to the
'MR Environment kit – Installation
Instructions'.
• Lift off the rear cover from the front panel section.
bullet
All parts inside the front panel section are now
accessible.
Label strips
S30038
S30039
S30040
l
S30042
S30043
a S30160
s d S30159
S30161
S30162
S30163
g S30164
S30165
S30166
S30167
S30168
S30169
S30170
h j S30171
a l s f ;
S30024
Disassembling
Assembling
d S30030
f S30142
f
S30025
Note: MR environment. On MR s a s
S30022
environment approved units, EMC
absorbers are mounted on some Type 2 = PC 1777F – M
components on PC 1777.
f S30031
g h j S30032
k S30029
s
S30026
Note: MR environment. On MR
environment approved units, EMC Type 3 = PC 1777P or higher
absorbers are mounted on some
components on PC 1777.
f f
S30141
g
S30143
S30147
S30148
If an O2 cell is installed:
S30150
S30151
If an O2 sensor is installed:
S30152
S30153
S30149
• The membrane (3) and the valve seat (4) are now
bullet
accessible, e.g. for inspection and cleaning.
a f d s
S30063
a s S30065
S30066
Notes: s a s
• When assembling, make sure that the connector
bullet
muff properly seals around the inspiratory pipe
(1). The connector muff must not be pushed too
far onto the nozzle units (2).
S30067
• Lift off the mounting plate (2). The safety valve pull
bullet
magnet is mounted on this plate.
S30044
f d d f S30045
S30072
S30073
S30072
S30075
S30071
S30068
S30070
Notes:
• When replacing PC 1771 Control, PC 1772
bullet
Monitoring or PC 1784/PC 2024 Expiratory
Channel, it can be necessary to re-install the
system software. For further information refer to
section Replacing PC boards in this chapter.
Note: MR environment. On MR
environment approved units, EMC
absorbers are mounted on some
components on PC 1771, PC 1772 and
PC 1784/PC 2024.
d d s a
S30048
d h
Note: When assembling the unit, make sure that
the spacer (6) is in position with its flat side facing
down. The spacer, designed as a wedge, should be
pushed inwards before the screws (3) are tightened.
S30049
j
S30050
a s S30051
d f S30052
S30053
g S30054
a
s
S30055
S30056
S30057
Notes:
• When assembling units with metal air duct (4),
bullet
g
make sure that the rubber seal (5) is mounted as
shown in the illustration.
S30058
Notes:
• Units with PC 1957 RS232/Alarm output
bullet
connector must be equipped with spill protection
strips. These strips are included in the optional
RS232/Alarm output connector kit.
s
• A cable adapter must be used to connect
bullet
PC 1789 to PC 1772 of new design. This cable
adapter is delivered with spare part PC 1772 and
PC 1789.
S30060
S30079
S30080
5 Service procedures
Table of Contents
5.1 Biomed menu | 5-2
5.2 Service key | 5-4
5.3 Field Service System | 5-5
5.4 Software installation | 5-6
5.5 Leakage detection | 5 -10
5.6 Battery modules | 5 -14
5.7 Memory backup batteries | 5 -16
Functions
• Latest logs
• Selected time interval
Test results
- All other Biomed submenus are available only Display Service logs.
Preventive
during Standby and require an access code. The Maintenance Possible selections:
• 21 latest logs
• Selected time interval
default code is 1973, but the code can be Service report
history
changed in the Biomed menu. Installation log
Set date
and clock Date and time setup.
S30086
the operating capacity meter (breath counter) in the • Respiratory Rate – Lower and Upper alarm limit
bullet
Expiratory cassette. The Replaced Exp membrane • End Exp. Pressure – Lower and Upper alarm limit
bullet
button is by default disabled in the Biomed menu, but
• CPAP – Lower and Upper alarm limit (infant alarm)
bullet
can be enabled in the Field Service System
(Configuration > General). • End Tidal CO2 – Lower and Upper alarm limit
bullet
• Apnea time – Alarm limit
bullet
Edit configuration • NAVA Apnea off - Disable/Enable (infant alarm).
bullet
Software option.
Available only in Standby mode.
• Leakage fraction - Disable/Enable (NIV infant
bullet
alarm).
Alarm sound level
The alarms listed above, except Pressure alarm and
Default alarm sound level can be changed in this Apnea time alarm, can be permanently silenced in this
window. window.
Adult alarm limits and Infant alarm limits Default configuration regarding General and Units
cannot be changed in this window. Access to the Field
Default invasive alarm limits can be changed in this Service System (FSS) is required.
window:
Startup configuration
• Pressure – Upper alarm limit
bullet
Startup configuration can be changed in this window:
• Minute Volume – Lower and Upper alarm limit
bullet
• Respiratory Rate – Lower and Upper alarm limit
bullet • Patient range
bullet
• End Exp. Pressure – Lower and Upper alarm limit
bullet • Type of ventilation
bullet
• End Tidal CO2 – Lower and Upper alarm limit
bullet • Volume setting
bullet
• Apnea time – Alarm limit
bullet • Breath Cycle setting
bullet
• NAVA Apnea off - Disable/Enable (infant alarm).
bullet • Pre/post oxygenation conc. above set O2 conc
bullet
Software option.
• Option to deactivate backup ventilation
bullet
• Low Minute Volume audio off – Disable/Enable
bullet
(infant alarm). Software option. • NIV disconnect function
bullet
SERVO-i Ventilator System, Service Manual 5-3
| 5 | Service procedures |
Available only in Standby mode. With the Service key it is possible to access different
user levels of the built-in Field Service System (FSS)
Default user configuration can be copied to/from a software. There are two user levels intended for
PC Card. service:
Access code to the Biomed menu can be changed in • Make sure to log out (select Exit) from Field
bullet
Service System when completed. If the Service
this window.
key is removed while Field Service System is
open, access remains until manual exit.
Start MCare Remote Services
O2 cell adaptation
This chapter describes installation of System software A System software installation will install a new System
and Software options. SW version. System software installation is not
dependent on the serial number of the system and
Only personnel trained and authorized by Maquet shall will not alter the installed Software options.
be permitted to install System software and Software
options.
Software option
It is not recommended to install System software with
A Software option installation will install Software
a lower version number than already installed in the
options ordered for the concerned system. For
system.
functionality enhancement, the latest released System
Before starting any software installation, check: software version will also be installed.
• User's Manual.
bullet
• Gas supply required for the Pre-use check.
bullet
5.4.4 System software installation • In the Report PM window, check Time to Next
bullet
Preventive maintenance. Adjust if required.
• Software stored in the Edi module will be
bullet
checked/installed during the software installation. • Perform a Pre-use check according to instructions
bullet
in the User's Manual.
If an Edi module is part of the SERVO-i system, the
Edi module shall be connected during the software Notes:
installation.
• If the software installation is performed after a
bullet
• Set the On/Off switch to Off.
bullet PC board replacement, the installation may fail
due to a conflict between the installation
• Insert the PC Card.
bullet
programs. If this occurs, repeat the installation
• Set the On/Off switch to On. The installation window
bullet procedure once more.
will now appear on the screen and the
semi-automatic installation procedure will start. • Following a software installation, technical alarms
bullet
Follow instructions given on the screen. may be activated, e.g. Settings lost: Restart
system! due to reset of internal memory on the
• Press the Main rotary dial to start the installation
bullet PC boards. Restart the system once more and
(or switch the system Off to cancel). On systems
check that no technical alarms are activated.
with language other than English selected, a dialog
concerning language installation may appear. Refer • Check that the version of the User's Manual used
bullet
to section Change language. together with the updated system corresponds
to the installed System software version. User's
• The installation program will now check the software
bullet Manual is not enclosed with the System software,
and hardware versions installed in the system. This
must be ordered separately if required.
check is done to detect if the installation must be
terminated due to hardware incompatibility. If • Make the customer aware of changes in the
bullet
detected that the software cannot be installed due function of the system related to the software
to hardware incompatibility, the installation will be installation.
terminated. A message will appear on the screen.
Remove the PC Card and restart the system.
5.4.6 Software option installation • Remove the PC Card and close the slot cover.
bullet
5.5.1 General Start the Leakage detection when the Pre-use check
has stopped and displayed either Leakage or
When a leakage is detected in the internal leakage Excessive leakage. Do not cancel the Pre-use check,
test, the Pre-use check will stop and display either the Redo test function will be used during the Leakage
Leakage or Excessive leakage. To locate the leakage, detection.
the three leakage tests described below can be used.
Test equipment:
• Test tube
bullet
• Leakage detector tool, P/N 68 81 146
bullet
• Manometer, local purchase ( e.g. a Cuff filling
bullet
measureing device with manometer).
• Test tube
bullet S30087
• Inspiratory pipe
bullet
• O2 cell bacteria filter seal
bullet
• Safety valve membrane
bullet
• Inspiratory pressure silicone tube
bullet S30088
Expiratory pressure
PC 1785 transducer tube Expiratory
valve coil
Expiratory
pressure
transducer
PC 1781
Part of leakage
P40 N40
P50
detector tool MOD1
S30089
Expiratory channel
Expiratory valve coil
1. Connect leakage detector and manometer at
1. Remove the expiratory cassette.
expiratory inlet.
2. Select Redo test.
2. Select Redo test.
3. Make sure the expiratory valve coil axis is
3. When the expiratory valve closes, pressurize to
activated (pushed up).
approx. 80 cmH2O. This must be done quickly
because the expiratory valve coil is closed only If the axis is activated, continue with the Expiratory
for 2-3 seconds. A pressure drop indicates a cassette.
leakage.
If not, check/replace as follows (one at a time):
If a leakage is detected in the Expiratory channel,
continue with Expiratory valve coil. If not, go to Test 1. Check the expiratory valve coil including cable
case 3. and connector.
2. Replace PC 1784/PC 2024 Expiratory Channel.
3. Replace PC 1771 Control.
Expiratory cassette
• Expiratory cassette.
bullet
Press
S30090
Inspiratory
pressure
PC 1781 transducer
P50
MOD2
PC 1784/ PC 2024
S30091
Manufacturing
date = YYMMDD
(YearMonthDay)
(01) XXXXXXXXXXXXXX
(21) XXXXXX (11) XXXXXX
Serial number
S30108
• Slot number.
bullet
• Battery module S/N.
bullet
• Charge indicator for each battery, where:
bullet
- 0 boxes filled = < 10% relative charge
S30083
- 1 box filled = 10-25% relative charge
- 2 boxes filled = 26-50% relative charge
- 3 boxes filled = 51-75% relative charge
- 4 boxes filled = 76-100% relative charge
5.7 Memory backup batteries 5.7.3 Replacing the Memory backup battery
• Cut the cable ties holding the old battery (4) to the
bullet
PC board.
5.7.2 Preparations
• Disconnect and remove the old battery.
bullet
Before disassembling or assembling the system, make
• Mount the new battery onto the PC board using
bullet
sure that: new cable ties as shown in the illustration.
a f S30085
6 Troubleshooting
Table of Contents
6.1 General | 6-2
6.2 Pre-use check | 6-2
6.3 Technical error codes and messages | 6 -10
Internal tests If possible, check the Test results log in the More
detailed mode (FSS).
(MON + PAN) 1. Make sure that the Patient unit main cover
and the User interface rear cover are
correctly mounted. Otherwise the audio
tests may fail.
2. Check in the Test results-log (FSS) if it was
the Panel test or the Monitoring test that
failed:
• If the Monitoring test failed:
bullet
Replace PC 1772 Monitoring.
• If the Panel test failed:
bullet
Replace the loudspeaker or PC 1777
Panel.
Alarm output connector test. If the Alarm output connector test failed:
Barometer test Checks that the barometric pressure measured Check the Barometric pressure value in the
by the internal barometer is within Status window.
650–1070 hPa.
If that value is within 650–1070 hPa:
Checks that the measured barometric pressure
values differ less than 8 hPa between BRE and 1. Replace PC 1771 Control
MON. 2. Replace PC 1772 Monitoring
(BRE + MON) 3. Replace PC 1784/PC 2024 Expiratory
channel
4. Replace the gas modules. Replace one gas
module at a time.
The new zero value for the pressure transducers If Pressure transducer test failed:
may not differ more than ±6 cmH2O from factory
calibration. 1. Cancel the Pressure transducer test and
continue with the remaining tests (this will
With the inspiratory pressure transducers used calibrate the safety valve). Restart the
as a reference, a new gain factor is set for the Pre-use check.
expiratory pressure transducer. The new gain 2. Check the expiratory cassette:
factor may not differ more than ±5% from factory
calibration. • If possible, replace the expiratory
bullet
cassette and check if the new cassette
During this test, the different subsystems is accepted by the Pre-use check.
concerned are compared. The difference
between the sub-systems must not be more than
• If the new cassette was accepted by the
bullet
Pre-use check, the fault was located to
±1 cmH2O at 60 cmH2O. the cassette. The fail with the old
cassette may in this case be due to
Expiratory valve coil test. Measures offset and water collected at the pressure
gain in the valve coil. transducer filter inside the cassette. Dry
the cassette properly.
(BRE + MON) 3. Check/replace PC 1781 Pressure
transducer (Insp. and Exp.). To locate the
faulty pressure transducer, replace one
transducer at a time.
4. Replace PC 1771 Control.
5. Replace PC 1772 Monitoring.
Safety valve test Checks and if necessary adjusts the opening 1. Check the inspiratory section:
pressure for the safety valve to 117 ±3 cmH2O.
• Check that the inspiratory pipe is
bullet
correctly mounted in inspiratory section.
Checks the hardware signals related to the
safety valve functions. • Check that the safety valve membrane
bullet
is clean and correctly seated in the
(BRE + MON) inspiratory pipe.
2. Replace the safety valve pull magnet.
3. Replace PC 1784/PC 2024 Expiratory
channel.
4. Replace PC 1772 Monitoring.
(BRE + MON)
Flow transducer test Checks the inspiratory flow transducer. 1. Check that the connected gas supply
Calibrates and checks the expiratory flow pressure (Air and O2) is within the specified
transducer. range.
2. Check the expiratory cassette:
Calibrates at 60% O2 and checks at 100% and
21% O2. As different gas mixtures are used • Check that the cassette is correctly
bullet
seated in the cassette compartment.
during this test, calibration of the expiratory flow
transducer will only be performed if both gases
are connected. The check using the connected
• If possible, replace the expiratory
bullet
cassette and check if the new cassette
gas, (100% alt. 21% O2) will however be is accepted by the Pre-use check.
performed. The Flow transducer test will pass if
the result of this check corresponds to the old • If the new cassette was accepted by the
bullet
Pre-use check, the fault was located to
calibration factor from a previous Pre-use check.
The same expiratory cassette must be used. the cassette. The fail with the old
cassette may in this case be due to
The new calibration factor for the expiratory flow water collected inside the cassette. Dry
transducer may not differ more than -10% to the cassette properly.
+15% from factory calibration. 3. Replace the gas modules. Replace one gas
module at a time.
During this test, the different subsystems
4. Replace PC 1785 Expiratory channel
concerned are compared. The difference
connector.
between the subsystems must not be more than
±0.3 l/min. 5. Replace PC 1771 Control.
6. Replace PC 1772 Monitoring.
(BRE + MON) 7. Replace PC 1784/PC 2024 Expiratory
channel.
Battery switch test Checks that the power supply switches to battery 1. Check that the total remaining time for the
when mains power is disconnected. Battery module(s) is >10 min. If not, allow
the battery to charge and repeat the test.
Checks that the power supply switches back to 2. Replace PC 1775 Plug-and-play
mains power when reconnected. back-plane.
This test will not be performed if: 3. Replace the Battery module(s).
(MON)
(BRE)
Y sensor test Checks the pressure and flow measurement of 1. If the patient circuit test (see above) has
the Y sensor. passed, check that there is no leakage in
the Y sensor and that there are no
This test is only performed if a Y sensor and a obstructions in the Y sensor tubings.
Y sensor module is connected to the system. 2. Replace the Y sensor.
3. Replace the Y sensor module.
Alarm state test Checks that no Technical error alarms are active Refer to section regarding Technical error codes
during the Pre-use check. and messages for further information.
(MON)
External alarm system If the option Alarm output connector is enabled, 1. Check the external alarm system.
test the user can test the external alarm system. 2. Replace PC 1957 RS232/Alarm output
connector (alt. PC 1789 Alarm output
The external alarm output signal is activated and connector).
the user must verify the external alarm.
The result of the test does not affect the outcome
of the Pre-use check.
Separate Patient circuit This test separately performs the Patient circuit See above.
test test as described above.
6.3.1 Monitoring
Error
Error message / Possible cause Recommended action
code
1 Power error. 1. Replace PC 1778 DC/DC & Standard connectors.
Error
Error message / Possible cause Recommended action
code
10 Ventilation stopped. 1. Replace PC 1784/PC 2024 Expiratory channel.
2. Replace PC 1771 Control.
The VALVES_DISABLED signal has stopped power
supply to gas modules and safety valve. 3. Replace PC 1778 DC/DC & Standard connectors.
4. Replace PC 1772 Monitoring.
24 Backup audible alarm or Remote alarm error. 1. Make sure that System SW version 3.02.02 (or
higher) is installed.
Error indicated in the backup alarm or remote alarm 2. Replace PC 1772 Monitoring
capacitor on PC 1772 Monitoring.
Error
Error message / Possible cause Recommended action
code
25 Communication error. Depending on ID # stated in the Technical alarm log.
Backup alarm buzzer error. Note: If the Patient unit front cover is removed, this
error may be activated.
Error
Error message / Possible cause Recommended action
code
34 User interface communication error. 1. Check the control cable.
2. Replace PC 1778 DC/DC & Standard connectors.
Communication error between PC 1777 Panel and
PC 1772 Monitoring during startup. 3. Replace PC 1777 Panel.
Error
Error message / Possible cause Recommended action
code
44 Alarm limit error. 1. Restart the system.
2. Replace Memory backup battery on PC 1772 if
Alarm limit settings (with checksum) are stored. If the together with error code 29.
checksum is incorrect, this Technical error alarm is
activated. 3. Replace PC 1772 Monitoring.
52 O2 sensor cannot be used with Heliox. 1. Replace the O2 sensor with an O2 cell.
53 PC 1784 Expiratory channel version conflict. 1. Replace PC 1784 Expiratory channel. PC 1784
must be of Revision 16 (or higher) when used with
Heliox.
6.3.2 Breathing
Error
Error message / Possible cause Recommended action
code
10001 Lithium battery depleted. 1. Replace battery on PC 1771 Control.
6.3.3 Panel
Error
Error message / Possible cause Recommended action
code
20001 Communication error. 1. Replace PC 1777 Panel.
20002 Backlight error. If the display is lit (and the error code is shown on the
display):
The backlight power consumption is measured. If the
power consumption is outside specified range during 1. Replace PC 1777 Panel.
>20 consecutive readings, this technical error is
activated. If the display is not lit (the display is dark):
This error message may also be logged immediately 1. Replace backlight lamps.
before the No battery capacity alarm when the system
is running in battery mode. 2. Replace PC board Backlight Inverter.
Notes:
• If one of the lamps is broken, the other lamp will
bullet
automatically be switched off. Thus, with a failure
on a backlight lamp, or likely also on PC board
Backlight inverter, the User interface display will
become dark. This error code will in such case
not be possible to see (but will be logged).
• If the User interface is placed too close to the
bullet
MR scanner, the PC board Backlight inverter
may become defective.
20003 Membrane button error. 1. Restart the system. Do not touch the User
interface during system startup. Interfering with
Membrane buttons are checked during startup. If any the knobs, keys, touch screen, loudspeaker grid,
button is detected as pressed, this technical error is etc, may affect the internal technical tests.
activated. 2. Check the User interface membrane buttons
(FSS).
3. Replace the Touch screen.
4. Replace PC 1777 Panel.
20004 Alarm sound level too low. 1. Restart the system. Do not touch the User
interface during system startup. Interfering with
The loudspeaker is checked using a microphone during the knobs, keys, touch screen, loudspeaker grid,
startup: etc, may affect the internal technical tests.
2. Replace the loudspeaker.
• On PC 1777A-C; one beep (silent before and after). 3.
bullet Replace PC 1777 Panel.
• On PC 1777D (or higher); three beeps (silent
bullet
between).
Error
Error message / Possible cause Recommended action
code
For systems with PC 1784 Expiratory channel
40001 Exp. flow meter error. Technical alarm Expiration flow meter PC 1784 and
Exp. cassette EEPROM read/write error:
See information logged together with error code 40001
in the Service log / Technical alarms. 1. Replace PC 1784 Expiratory channel.
Error
Error message / Possible cause Recommended action
code
40006 Power error. 1. Replace PC 2024 Expiratory channel.
40010 Internal temperature too high, > 77 ºC. 1. Ensure proper cooling:
- Check fan operation
The temperature is too high or temperature sensor
error on PC 2024. - Check the operating temperature
- Check the fan filter in the Patient unit.
2. Replace PC 2024 Expiratory channel.
7 Preventive maintenance
Table of Contents
7.1 General | 7-2
7.2 Preventive maintenance intervals | 7-2
7.3 Preventive maintenance calculation and | 7-2
notification
7.4 Required equipment | 7-2
7.5 Maintenance kit 5000 hours | 7-3
7.6 Performing the Preventive maintenance | 7-3
7.5 Maintenance kit 5000 hours • Make sure that gas supply hoses and connectors
bullet
are correct.
Only original parts from the manufacturer must be
used. Spare parts and maintenance kits can be • Check if there are any unexpected Technical alarms
bullet
ordered from your local Maquet representative. in the Service log available via the Biomed menu.
Technical alarms are available also via the Field
When performing this maintenance, a Maintenance Service System (FSS).
kit 5000 hours should be used.
• Perform a Pre-use check to make sure that the
bullet
system works properly before the maintenance.
The following parts shall be replaced and they are
included in the maintenance kit: • Turn the system Off.
bullet
A. Filters for the gas modules • Disconnect the mains power cable.
bullet
B. Nozzle units for the gas modules • Disconnect the gas supplies (wall and/or cylinder).
bullet
C. Bacteria filter for the inspiratory pressure • Remove patient tubing.
bullet
transducer • If fitted, remove bacteria filter from the expiratory
bullet
inlet.
S30094
7.6.1 Preparations
S30095
S30096
S30097
S30098
8 Index
Table of Contents
8.1 Alphabetic index | 8-2
T
Technical error codes and messages 6-10
Technical training 1-2
Temperature sensor 2-9, 3-7
Test case 1 5-10
Test case 2 5-10
Test case 3 5-10
TFT Display 2-6, 3-4
Threadlocking adhesives 4-5
Tightening torque 4-5
Touch screen 3-2
Touch screen incl. frame 2-6, 3-3
U
Ultrasonic flowmeter 2-9, 3-9
User interface 2-2, 2-5, 3-2
User interface controls 3-2
User interface functions 3-2
User interface panel cover 2-4, 3-20
User interface Type 1 3-2
User interface Type 2 3-2
User interface Type 3 3-2
User's Manual 1-1
Y
Y sensor measuring 2-4, 3-20
9 Revision history
Table of Contents
9.1 SERVO-i revision history | 9-2
9.2 Service Manual revision history | 9-4
PC 1777 Panel
Type 2 Type 3 PC 1784 PC 2024
PC 1771 PC 1772 Type 1 PC 1777 PC 1777 PC 1777 Expiratory Expiratory
Control Monitoring PC 1777A – D F–L M P– channel channel
SERVO-i System SW version
System SW requirement for 64 67 596 64 67 620 64 67 778 65 70 241 65 70 829 68 81 356 68 83 322 64 67 984 68 81 702
version options revision # revision # revision # revision # revision # revision # revision # revision # revision #
1.02.04 – ≥ 07 ≥ 10 ≥ 08 ≥ 00 – – – 08 – 16 –
1.03.00 – Alarm output ≥ 07 ≥ 10 ≥ 08 ≥ 00 – – – 08 – 16 –
connector
Note 1
2.00.04 – ≥ 07 ≥ 10 ≥ 08 ≥ 00 ≥ 01 – – 08 – 16 –
3.00.02 – Y sensor ≥ 07 ≥ 10 ≥ 08 ≥ 00 ≥ 01 – – 08 – 16 –
measuring
3.01.00 – O2 sensor ≥ 07 ≥ 10 ≥ 08 ≥ 00 ≥ 01 – – 08 – 16 –
3.02.01 – Edi module - ≥ 07 ≥ 10 ≥ 08 ≥ 00 ≥ 01 – – 08 – 16 –
(3.00.12) NAVA
4.00.00 – Heliox ≥ 07 ≥ 10 ≥ 08 ≥ 00 ≥ 01 – – 08 – 16 –
(3.03.00, 3.04.00, Note 2
3.05.00)
4.00.04 – ≥ 07 ≥ 10 ≥ 08 ≥ 00 ≥ 01 – – ≥ 08 –
6.00.00 – RS232/Alarm ≥ 07 ≥ 10 ≥ 08 ≥ 00 ≥ 01 – – ≥ 08 –
(3.05.01, 3.05.02, output connector
3.05.03) Note 3
Aerogen function
Note 4
7.00.04 – ≥ 07 ≥ 10 ≥ 08 ≥ 00 ≥ 01 ≥ 01 ≥ 03 ≥ 11 –
7.01.00 – CO2 analyzer ≥ 07 ≥ 10 ≥ 08 ≥ 00 ≥ 01 ≥ 01 ≥ 03 ≥ 11 –
(7.00.50) module
Capnostat 5
8.00.01 – ≥ 07 ≥ 10 ≥ 08 ≥ 00 ≥ 01 ≥ 01 ≥ 03 ≥ 11 ≥ 02
(7.00.05, 7.00.51,
7.01.02)
S30146
Part and revision numbers for the PC boards can be Note 1: PC 1789 Alarm output connector requires:
found on:
• PC 1772 Rev ≥ 15.
bullet
• Labels on the PC board (part number/revision
bullet
number). Note 2: Heliox option requires:
9.2.2 Revision 04
9.2.3 Revision 05
9.2.4 Revision 06
9.2.5 Revision 07
9.2.6 Revision 08
10 Diagrams
Table of Contents
10.1 Standard and optional connectors |10 - 2
10.2 EMC emission |10 - 3
10.3 PC 1770 Main back-plane |10 - 4
10.4 PC 1780 Pneumatic back-plane |10 - 7
10.5 Field Service System menus |10 - 9
10.6 System overview |10 -10
5
1
S30102
1. GND
2. DISABLE_VALVES_EXT.L
S30100
3. CAN_EXPANSION.H
1. +12V_UNREG_EXT_DC 4. CAN_EXPANSION.L
2. +12V_UNREG_EXT_DC 5. +12V_UNREG_EXPANSION
3. –
4. GND 10.1.4 N29 and N68 – RS232
5. GND
Recommended cable connector (local purchase):
Amphenol C 091 A, P/N T 3360 005
1. –
2. RXD_ISO
3. TXD_ISO
4. DTR_ISO
5. GND_ISO
S30101
6. –
1. +12V_UNREG_PANEL
7. RTS_ISO
2. GND
8. –
3. LED_CONTROL
9. –
4. ON_OFF_CONTROL
5. CAN_PANEL.H
6. CAN_PANEL.L
S30104
1. NO – Normally Open
2. NC – Normally Closed
3. Common
4. Common
Recommended cable connector (local purchase):
Modular plug 4/4 (4P4C)
S30105
S30154
S30155
S30156
S30157
S30158
P74
H
slot
Back-
light
B
P92
N46 A
Inspiratory pipe N70
P46 P75 TFT
ID
P18 N18 PC 1785
P76
P87
Expiratory SW1
Expiratory Expiratory
P77
channel
connector valve one-way µP
Heating foil valve P80 SW2 ID Touch
screen
P79
P81
P82
P84
P83
P85
Expiratory
N78
Ultrasonic flowmeter
outlet
N86
Bacteria filter
et
inl
tory PC1785 Type 3 - View from rear Service
ira Expiratory Expiratory connector
p
Ex P45 pressure valve coil
N45 tube
U
X1 PC 1777 Panel
I Inspiratory P72 PC Card
P74
Temp. sensor slot
section
N19 P73 N70
P19 P75 TFT
Inspiratory
ID
P76
pressure
P67
Filter
Inspiratory Gas module SW1
Back- light
tube O2 cell
P77
outlet O2
P80 µP SW2 ID Touch
N15 P1 ID screen
Inspiratory pipe P79
P81
P82
P84
P83
P85
P78
N86
Safety
Emergency
N18 P18
valve
air intake
Gas module
Safety valve Connector Air Type 2 - View from rear Service
C M connector
pull magnet muff ID Filter
N16 P1
PC 1780
N17
Pneumatic
P17 back-plane C M C
Control cable
N10
P10
User interface
B PC 1770 Main back-plane
Air O2
T Pressure transducers gas inlet
Inspiratory Expiratory
ID pressure pressure ID
PC 1781 transducer transducer PC 1781
P PC 1778
P50 P50 DC/DC &
P40 N40
PC 1785
RS232
Temperature
µP C M
C M sensor ID N29
PC 1784/PC 2024 Expiratory channel PC 1780
CAN bus
N28
User interface
N1 Patient unit
P1 ID
C Battery N21 PC 1771 Control
PC 1770
P23
PC 1789/
BP
PC 1957
C µP
PS1 ID
X1 X2 N22
P20
N20
A RS232/Alarm
output connector
PC 1775
N3
P3 P66
X N68 RS232 AC/DC
PC 1957 Alarm converter
RS232/Alarm N67
output
output connector Functional main blocks
PC 1789 U User interface functions
Nebulizer N11 N8 P8 Alarm
P66 Alarm P67 output B Main back-plane
output I Inspiratory section
P Module unit ID connector
E Expiratory section
µP ID T Pressure transducers
Battery module P31
C Control
P5 N5
M Monitoring
Battery module P32 P7 N7 F Expiratory channel
P Power supply
Optional module P33 Internal A RS232/Alarm output connector
fan
P14 N14
P34
Symbols
Optional module PC 1775
Plug- P N/PXX Connector
and- Mains inlet N/PXX Connector including CAN bus
Optional module P35 play P91 N91
AC/DC
back- converter ID ID PROM communicating
plane P90 N90 with Monitoring
S30001_16-06-23