You are on page 1of 12

Carestream X-ray Generators

(a.k.a. “Everest” Generators)


in DRX-Ascend and Q-Rad Systems

Overview

P/N 7K7752 2016-09-29


Table of Contents

1. Introduction ..................................................................................................................................................... 3
Generator Models: .......................................................................................................................................... 3
2. Labels ............................................................................................................................................................... 4
3. Generator Options ........................................................................................................................................... 5
4. Generator Specifications ................................................................................................................................. 7
5. Generator Power On / Off ............................................................................................................................... 7
6. Service and Installation ................................................................................................................................... 8
Generator Service Manuals ............................................................................................................................. 8
Service Information ......................................................................................................................................... 9
System Installation with a Carestream Generator ........................................................................................ 10
System Interconnect Diagrams for Carestream Generator Systems ............................................................ 12
Service Training.............................................................................................................................................. 12

CARESTREAM Generator in DRX-Ascend and Q-Rad Systems Page 2


1. Introduction
CARESTREAM 3-phase generators have replaced QUANTUM 3-phase generators in new DRX-Ascend and
Q-Rad Systems.
 CARESTREAM 3-phase generator cabinets are smaller, have modern control electronics, and superior
reliability.
 The generator installation steps are the same. See Section 6 for more details.
 The Generator User Interface (Quest, Odyssey, Q-Vision, DirectView, DRive) is the same.
Note: Recent updates to the Odyssey and Quest designs (Dealer Support Bulletin #1308) are not related
to the Carestream Generator.
 Quantum Single Phase and Stored Energy generator models have not changed.
 Carestream has spare parts to support the installed base of QUANTUM 3-phase generators.

Generator Models:

Quantum Generator Carestream Generator


20 kW STORED ENERGY QG-20G-SE ** Discontinued ** N/A
25 kW 208-260 VAC Single-Phase QG-25G ** Discontinued ** N/A
200-260 VAC Single-Phase QG-32G --
208-240 VAC Three-Phase → QG-32G-2
32 kW
380-480 VAC Three Phase → QG-32G-3
STORED ENERGY QG-32G-SE --
200-260 VAC Single-Phase QG-40G --
208-240 VAC Three-Phase → QG-40G-2
40 kW
380-480 VAC Three Phase → QG-40G-3
STORED ENERGY QG-40G-SE --
208-240 VAC Single Phase N/A N/A
208-240 VAC Three-Phase → QG-50G-2
50 kW
380-480 VAC Three Phase → QG-50G
STORED ENERGY QG-50G-SE --
208-240 VAC Single Phase N/A N/A
208 VAC Three Phase N/A N/A
65 kW
380-480 VAC Three Phase → QG-65G
STORED ENERGY N/A N/A
208-240 VAC Single Phase N/A N/A
208 VAC Three Phase N/A N/A
80 kW
380-480 VAC Three Phase → QG-80G
STORED ENERGY N/A N/A

CARESTREAM Generator in DRX-Ascend and Q-Rad Systems Page 3


2. Labels
Carestream Generators can be identified by the CARESTREAM X-ray Generator label.

The REF and MODEL designation for the Carestream Generator is the same as the Quantum Generator.
Quantum Generator CARESTREAM Generator

CARESTREAM Generator in DRX-Ascend and Q-Rad Systems Page 4


3. Generator Options
Q-HSS High Speed Starter: Rotor controller accelerates tube anode (10,000 RPM)
Q-HSS Quantum Generator Carestream Generator
208-240 VAC Three-Phase N/A Optional
32 kW
380-480 VAC Three Phase N/A Always Included
208-240 VAC Three-Phase N/A Optional
40 kW
380-480 VAC Three Phase N/A Always Included
208-240 VAC Three-Phase Optional Optional
50 kW
380-480 VAC Three Phase Optional Always Included
65 kW 380-480 VAC Three Phase Always Included Always Included
80 kW 380-480 VAC Three Phase Always Included Always Included

QG-150 150 kVp Output: Increases generator output to 150 kVp


Currently this option is included with all Carestream generators.
QG-150 Quantum Generator Carestream Generator
208-240 VAC Three-Phase Optional when tube ≥300 kHU Always included
32 kW
380-480 VAC Three Phase Optional when tube ≥300 kHU Always Included
208-240 VAC Three-Phase Optional when tube ≥300 kHU Always included
40 kW
380-480 VAC Three Phase Optional when tube ≥300 kHU Always Included
208-240 VAC Three-Phase Optional when tube ≥300 kHU Always included
50 kW
380-480 VAC Three Phase Optional when tube ≥300 kHU Always Included
65 kW 380-480 VAC Three Phase Always Included Always Included
80 kW 380-480 VAC Three Phase Always Included Always Included

CARESTREAM Generator in DRX-Ascend and Q-Rad Systems Page 5


 These options are all still available with the Carestream Generator.

 These options are not needed with the Carestream Generator, and are no longer available.

CARESTREAM Generator in DRX-Ascend and Q-Rad Systems Page 6


4. Generator Specifications
Carestream Generator Specifications are in these manuals:
 7K7738 Installation Planning Guide and Installation Instructions for Carestream Generator
 DC30-145 Everest Quest Service Manual
 DC30-144 Everest Odyssey_QVision Service Manual

Quantum Generator Specifications are in these manuals:


 DC30-139 - ODYSSEY II Operator Manual
 DC30-140 - QUEST II Operator Manual
 DC30-101 - Q-VISION Operator Manual
 DC30-138 Odyssey_Q-Vision HF Generator Service Manual
 DC30-141 Quest HF Generator Service Manual

5. Generator Power On / Off


Carestream Generator Power On/Off switch is on the Control Box. There is no circuit breaker on the
Carestream Generator cabinet as there was on the Quantum generator cabinet.

CARESTREAM Generator in DRX-Ascend and Q-Rad Systems Page 7


6. Service and Installation
Generator Service Manuals
Service Manuals are delivered with each system on a USB memory device. Manuals are also posted on the
Quantum FTP site.

Carestream Generator Service Manuals


 7K7738 Installation Planning Guide and Installation Instructions for Carestream Generator
 DC30-145 Everest Quest Service Manual_Rev A.pdf
 DC30-144 Everest Odyssey_QVision Service Manual_Rev A.pdf
APPENDIX A for Q-Vision and Integrated Consoles (DRX-Ascend, Q-Rad Digital, DRive)

Quantum Generator Service Manuals


 DC30-022_Rev H Generator Installation Planning Guide.pdf
 DC30-141 Quest HF Generator Service Manual_Rev B.pdf
 DC30-138 Odyssey_Q-Vision HF Generator Service Manual_Rev B.pdf
APPENDIX A for Q-Vision and Integrated Consoles (DRX-Ascend, Q-Rad Digital, DRive)

CARESTREAM Generator in DRX-Ascend and Q-Rad Systems Page 8


Service Information
Information for installation and setup of the Carestream generator is found in these documents:

Quest Odyssey Q-Vision, DRX-Ascend, DRive


CARESTREAM Generator: DC30-145 DC30-144 DC30-144
Unpack & Installation Chapter 2 Chapter 2 Chapter 2
Configuration & Calibration Chapter 3 Chapter 3 Appendix A
Troubleshooting Chapter 5 Chapter 5 Chapter 5
Adjustments & Replacements Chapter 5 Chapter 5 Chapter 5
Spare Parts List ** Chapter 5 Chapter 5 Chapter 5
** For the current list of all Carestream Generator Spare Parts see manual 7K7738 Installation Planning Guide
and Installation Instructions for Carestream Generator, Section 5.

Note: Generator Service Manual (7K3254) is not complete for the DRX-Ascend and Q-Rad Systems
and is not included with the other service manuals. This manual is not needed because all the required service
information from this manual is included in the above referenced manuals (DC30-144, DC30-145, and
7K7738).

CARESTREAM Generator in DRX-Ascend and Q-Rad Systems Page 9


System Installation with a Carestream Generator
The steps to install a Q-Rad system that has a Carestream Generator are the same as with a Quantum
Generator.

System installation steps are in Q-Rad System Service Manual DC30-109 Figures 7-15 ~ 7-20.

CARESTREAM Generator in DRX-Ascend and Q-Rad Systems Page 10


With a Carestream Generator in Step 9(D) additionally verify the DIP switch settings.

CARESTREAM Generator in DRX-Ascend and Q-Rad Systems Page 11


System Interconnect Diagrams for Carestream Generator Systems
For cable connections to the Carestream Generator refer to the appropriate interconnect diagram.

Interconnect diagrams for all system configurations are in manual 7K7738 Installation Planning Guide and
Installation Instructions for Carestream Generator.

Quantum Generator Interconnect Diagrams


Interconnect diagrams for Quantum Generator systems are found in Publication DC30-109, figures 7-4, 7-7,
and 7-10 and Publication DC30-136_ QG-DIG-DRV Service Manual_Rev C Figures 3-2 and 3-2.

Service Training
 CARESTREAM X-ray Generator Training. See Carestream X-ray Generator Training for DRX-Ascend and Q-
Rad System Rev A

CARESTREAM Generator in DRX-Ascend and Q-Rad Systems Page 12

You might also like