Thank you very much for purchasing DAIHAN LABTECH. Co., LTD products
In order to use the product correctly and most appropriately, Please carefully read and fully understand
instructions in this manual and keep it in a convenient location for further reference.
In order to protect the human life and property, please pay attention to the following sings.
Warning
WARNING
This mark indicates precautions that must be taken if there is danger of electric shock, fire, etc.
Caution
CAUTION
This mark indicates that if these precautions and operation procedure are not taken, damage to
the equipment may result.
Note
This mark indicates that all precautions should be taken for safe usage,
Hot Surface
This mark indicates that burning injury by hot surface may result.
HOT
Explosive
WARNING
‐ This equipment must be used in accordance with the specifications to prevent fire or damage to equipment.
‐ This equipment is not intended for use in locations subject to flammable or explosive gases.
‐ Do not touch high-voltage connections such as power supply terminals, etc. To avoid electric shock.
‐ DAIHAN LABTECH is not responsible if this equipment is repaired, modified of disassembled by other than
factory-approval personnel.
‐ Clogging lines, Equipment malfunction of Failure in the steam supply may cause the autoclave chamber to fill
with scalding water. If the water leaks from any part of the autoclave, do not open the chamber door burns
from scalding water may otherwise result.
‐ Potential exposure to infectious agents through either airborne, droplet or contact transmission.
‐ Do not open the chamber door at any time until the pressure gauge indicating ‘0’
-2-
CAUTION
● All precaution described in this manual should be taken to avoid damage to the equipment.
● All wiring must be in accordance with local codes and regulations.
● Prevent metal fragments or lead wire scraps from falling inside equipment case to avoid electric shock, fire of
malfunction.
● For proper operation of this equipment, provide adequate ventilation for heat dispensation.
● Turn off the power supply before cleaning the equipment.
● Do not use a volatile solvent such as paint thinner to clean the equipment. Deformation of discoloration will
occur. Use a soft, dry cloth to remove stains from the equipment.
● To avoid damage to equipment display, do not rub with an abrasive material or Press front with a hard object.
● Do not refill the water in the reservoir in operation.
● When processing the Liquids, use only flasks and self-venting automatic sealing stoppers recommended by
the manufacturer. This will help to prevent liquid in the glass containers from rupturing.
● Waster that contains bleach may harm an autoclave.
● Carefully prepare items for autoclaving loosely cover or cap containers to avoid over-pressurization.
● Keep the loads appropriately- overloading hinder the steam penetration.
● Bags should be opened, contained within the stainless steel tray or basket.
● Please an autoclave on the plat and leveled surface.
● Be sure that connect the drain and steam ventilation valve to the drain sink.
● Use distilled water only to protect the heater and sensor.
● Change the water in the chamber frequently to prevent contamination and odor
HOT
-3-
● The figures, diagrams and numeric values used in this manual are only of the purpose of illustration.
● DAIHAN LABTECH is not responsible for any damage or injury that is caused as a result of using this
equipment, equipment failure of indirect damage.
● Not portion of this document may be reprinted, copied, stored, and retrieved through any electronic, optical or
other means with prior written approval from DAIHAN LABTECH.
● Do not remove the ground prong(third pin)from the power cord, or use an ungrounded adapter. The autoclave
requires a 3-wire receptacle. If one is not available, consult with an electrician for installation.
● To reduce the risk of electric shock, do not open the body, no user serviceable parts inside.
-4-
1. Table of contents
2. Beginning
Autoclaves are such a familiar feature in many laboratories that it is easy to forget what hazards they can pose. The
autoclave’s job is to render its contents sterile, or free of any living organisms. If it fails to do so, serious health hazards
can result. The hot, pressurized steam (135℃, 2.65 kg/㎠ in gauge) that autoclaves generate to do this job makes them
serious burn hazards as well. And, because conditions created inside steam autoclaves are so extreme, autoclaves can
easily malfunction if they are not carefully.
Before using DAIHAN LabTech LAC-Series autoclave at first, read and understand the operating manual thoroughly.
If you have not kept the manual, contact a local distributor or us directly to have a copy.
DAIHAN LABTECH CO., LTD. IS NOT RESPONSIBLE FOR ANY DAMAGE OF PROPERTY OR
PERSONAL INJURY FOR ABUSE OR USAGE FOR WRONG PURPOSE OF AUTOCLAVE.
-5-
3. Characteristic
¾ Convenient use
For your convenience, automatic irrigation system and drainage system is installed and after sterilisation, in a row
without waiting time can be used
As the UP & DOWN sliding door automatically opens and closes with the touch of button.
Method to gasket levitation- sealed door, there are less concerns about burn, because users do not have to control
it by themselves. You can check the condition of door at the LCD display.
As the automatic water supply system is built in, an amount of water provides into chamber to supply constant
water. It is economic, because of water saving system inside.
- (Thermostat)
- (Electronic Limiter)
All automatic control method by Dual 8Bit Processor, there is no mechanical set point, as there are self-check and
self-control, we increased the durability.
If you press start button after closing the door, Preheating > pre- vacuum > heating > sterilization > ventilation >
dry > completion , full automatic process.
-6-
¾ Self-check
There is self-check and self-control function, if equipments damage or parts defects is happened, with alarm sound,
the machine indicate the error no.(refer to 21, 22 Page)
¾ Safety equipment
If pressure insider the chamber is found, then door is not opened.
- when the pressure over 0.03Mpa (0.3Bar)
If the pressure inside the chamber increases then, safety valve will be opened automatically
- When the steam pressure over 2.5bar
Because of maintaining a constant pressure sterilisation, sterilization system is of excellent quality.
- While sterilizing cycle is in progress, steam pressure deviation over ±0.3BAR, then error will be shown.
Prevent from over current by ELCB(Earth leakage circuit breaker)
Self-diagnostic feature tells users the status of a sterilizer.
- Sterilization of the power-on self-diagnosis system has been working from the moment a problem occurs with
the LCD screen displays an error alarm will sound.
During mechanical error or power failure, automatically, pressure will be exhausted from the product.
Heater safety mode
- After finishing cycle or error, unless you press the STOP button, heater is not working.
-7-
4. Specification and diagram of product
-8-
z SPECIFICATIONS
MODEL LAC-6205SP/ LAC-6206ES
Printer Standard
Weight 750kg
* This specification could be changed to improve the product without any notice to customer
-9-
5. preparation
- 10 -
5.2. Installation and other cautions
※ Please read the instruction before use for user safety and correct use.
z Caution during installation
1. Do not install in an inclined or bumpy area.
2. Install the sterilizer horizontally and approximately 30cm away from the wall.
3. Establish minimum spacing required for inspection and repairs.
4. Do not leave flammable or volatile materials near the sterilizer such as gasoline.
5. Do not put anything on top of the sterilizer.
6. Power supply should be wired with 3Phase 4wiring and above50A circuit breaker.
z Use proper power voltage z Do not dismantle / remodel
Check for a power line that is at least 220V/2.0KW Do not dismantle, repair. Or remodel unless you are
(single Phase) and prepare an exclusive line. a repair technician.
There is danger of fire, electric shock, and injury
Connect the ground wire for safety. due to abnormal operation.
z When not used for an extended amount of time. Underground water is can not used.
Unplug the power cord from the outlet when not you used underground water, the underground water is
used for an extended amount of contains a lot of minerals.
Time and safe keep it from moisture. A cause of equipment failure is contains a lot of
minerals.
- 11 -
5.3. Name and function of each parts
5.3.1. Front view
LCD display
printer
Chamber
pressure gauge Power switch
Jacket pressure
gauge Chamber
Water pump
pressure gauge
Tray1, 2
door
- 12 -
5.3.2. Drain view (Rear view)
Supply of
Water Line
Drain Line
- 13 -
5.4. Name and function of indicator
① ⑥
②
⑤
③
④
8 LCD monitor indication of situation, temperature, pressure and time of the machine
- 14 -
6. use
Environment for operation
▪ Transport & Storage condition
- The direction has to be as arrows. Avoid the moisture. Fragile. Do not use hook.
- Temperature of shipping: 0℃ ~ +70℃
- Moisture of shipping : relative humidity 0%~ 95%
- Pressure of atmosphere: 760mmHg
- 15 -
z When closing the door
* In case of opening the door while cycle is progressing, please press the STOP button.
As the FAIL message is shown at the LCD monitor, door is opened automatically.
- 16 -
6.1. sterilization and drying time according to the type of sterilizing water and status
Not wrapped 20 10
MOIST
Individually wrapped with cloth or paper 30 15
※ Times above do not include preheating and preparation times and indicate the sustain time at the temperature.
Reference: Perkins JJ: Principles and methods of sterilization in Health Sciences, 2nded Springfield,IL, Thomas1982P
- 17 -
6.2. Progress and installation of sterilization
z Installation
①Use the Enter button to position the CURSOR to the corresponding part and change the values by pressing the
Down/Up KEY.
*If you want to record the sterilization process, activate the PRINT button before starting to print the full process of
sterilization.
- If the STOP button was pressed or the equipment is re-operated after power failure, FAIL is displayed.
(Remove remaining pressure)
z Re-operation
- After sterilization, push the STOP button.
And then, heater can be re-operation (prevention of overheating)
- 18 -
7. Operation
■ LAC-6205SP / LAC-6206ES STEAM STERILIZER
IF the sterilizing water comes into contact with the left/right side, there is a possibility of the material becoming
wet, so be careful for the water not to come into contact.
5. Operation begins closing the door when pressing the START button.
When sterilization, as follow process by built in Program
- 19 -
If you want to stop the sterilization, Press the STOP button.
When sterilization is stopped by ERROR or STOP button, please wait until the door open after exhausting
pressure.
6. When melody rings after cycle is finished, push the STOP KEY button.
9 The melody will be stopped after 1 minute.
9 When the cycle is ended by the STOP KEY button, it can be used continuously.
9 When the cycle is ended without the STOP KEY button,
9 In case of end after do not push STOP KEY,
- Æ It needs some times for preheating and STOP KEY should be pushed to set the equipment.
7. When door is opened automatically, please take out the samples.
WARNING
The inside of the equipment and sterilizing water is hot after sterilization.
There is a possibility of burning, so wear protection gloves.
- 20 -
8. Trouble shooting
▶ Is the door open? ● Turn the power OFF and then turn it ON
If the equipment stops
▶ Is the power turned off? again. Contact the A/S center if it
during operation
▶ Is an ERROR displayed? doesn’t return to normal.
▶ Are there any impurities on the gasket? ● Check if the lock wheel system is
If vapor leaks through
▶ Is the door completely closed and displayed as CLOSE. If the situation
the door
locked? continues, contact the A/S center.
- 21 -
z TROUBLE SHOOTING
¾ ERROR NO.
5 Insufficient in boiler O O O O O O
7 NONE
Temperature doesn’t rise to its set value within
8 O
60minutes
9 Temperature not retained with the standard temp O
¾ If you couldn’t fix the equipment after doing all above mentioned countermeasures, Please apply for a service
to a technical expert.
¾ During apply for a service department, Please let him know of the Error Code & Serial Number.
¾ Above all details will be change for better equipment.
- 22 -
Overseas 2nd Factory – PT.DAIHAN LABTECH Headquarters – 247-16 Gagok-Ri, Hwado-Eup
JI. Beringin Lot 322, Batamlndo Indusstrial Park, Namyangju-City, Kyonggi-Do, Korea
Mukakuning, Batam 29433, Indonesia TEL : 82-31-593-9561 / FAX : 82-31-593-9563
India Overseas Factory & Incorporative
Tel : +62 770 612277 / Fax : +62 770 612278
DAIHAN LABTECH INDIA PVT.LTD
TEL : +91 11 2811 4662 / FAX : +91 11 2811 4664
E-mail : info@labtech.co.kr Web : www.labtech.co.kr
Quality Warranty
Descriptions Vertical Steam Sterilizer
Model LAC-6205SP
Date of Purchase
Purchase From
WARRANTY COVERAGE
LabTech’s warranty obligations for the products are limited to the terms set forth below:
Daihan LabTech Co., Ltd. ("LabTech") warrants the product against defects in materials and workmanship for a period
of one (1) year from the date of original purchase ("Warranty Period"), providing that the unit is operated according to
the instruction in the operating manual.
The guarantee comprises removal of all damages that arises during the guarantee period and that are proven to be due
to faulty material or poor workmanship.
If a defect arises and a valid claim is received by LabTech within the Warranty Period, at its option, LabTech will (1)
repair the product at no charge, using new or refurbished replacement parts, (2) exchange the product with a product
that is new or which has been manufactured from new or serviceable used parts and is at least functionally equivalent to
the original product.
This Limited Warranty applies only to the product manufactured by or for LabTech that can be identified by Name Plate.
LabTech is not liable for any damage to or loss of any products or material stored or tested in the instruments or programs, data, or other information
stored on any media contained within the product, or any non-LabTech product or part not covered by this warranty. Recovery or reinstallation of
programs, data or other information is not covered under this Limited Warranty.
This warranty does not apply: (a) to damage caused by accident, abuse, misuse, misapplication, or non-LabTech products; (b) to damage caused by
service performed by anyone other than LabTech; (c) to a product or a part that has been modified without the written permission of LabTech; or (d)
if any LabTech serial number has been removed or defaced; or (e) if the unit is not used according to its purpose; or (f) no original spare parts are
used.
LabTech warrants replacement products or parts provided under this warranty against defects in materials and
workmanship from the date of the replacement or repair for ninety (90) days or for the remaining portion of the original
product's warranty, whichever provides longer coverage for you. When a product part is exchanged, any replacement
item becomes your property and the replaced item becomes LabTech’s property. When a refund is given, your product
becomes LabTech's property.
- 24 -
APPENDIX Ⅰ
P&I schematic
- 25 -
APPENDIX Ⅱ
Electric diagram
- 26 -
APPENDIX Ⅲ
Environmental Health and Safety Internal
Weekly Inspection Log
Check for the following ;
◈ Chemical indicators supply ◈ Biological indicators supply
◈ Make sure the heater cover at the bottom of the chamber or water in the chamber clean
◈ Leakage of water or steam in pipeline
- 27 -
APPENDIX Ⅳ
Autoclave Use List
CHEMICAL
NUMBER OF PRINT YOUR
PHONE BIOWASTE INDICATOR
DATE TIME CYCLE BAGS NAME AND
EXT (YES/NO) (If biowaste accept
(If Biowaste) LABORATORY
or rejected)