You are on page 1of 7

Shenzhen HugeMed Medical Technical Development Co.

,LTD
Add.:8/F,Tempus Building A, Qingshui River Road, Luohu District ,Shenzhen City,Guangdong Province,P.R.China.
Tel:0086-755-22275866 Fax:0086-755-82267835 Web: www.hugemed.net

HugeMed VL3R Video Laryngoscope


Cleaning Instructions

HugeMed, Co., LTD


8F, Block A, Tempus Building
1st Qingshuihe Road, Luohu District
Shenzhen City, Guangdong, China
www.HugeMed.net
Shenzhen HugeMed Medical Technical Development Co.,LTD
Add.:8/F,Tempus Building A, Qingshui River Road, Luohu District ,Shenzhen City,Guangdong Province,P.R.China.
Tel:0086-755-22275866 Fax:0086-755-82267835 Web: www.hugemed.net

Cleaning

Cleaning Process:

After operation

Use wet gauze to wipe

To Cleaning chamber

Staff with protective wear

Clean by water and wipe it

Soak

Clean and wipe it dry

Sterilized by cleaner

Clean and dry

Stored in a sealed box


Shenzhen HugeMed Medical Technical Development Co.,LTD
Add.:8/F,Tempus Building A, Qingshui River Road, Luohu District ,Shenzhen City,Guangdong Province,P.R.China.
Tel:0086-755-22275866 Fax:0086-755-82267835 Web: www.hugemed.net
The following is recommended disinfectant for cleaning:

Disinfectant Material g/100g Manufactuer


TERRALIN Benzene methylene 20g Schülke
(high concentration) chloride
Phenoxy propionic 35g
aldehyde
QUATOHEX Dimethyldecyladipate B. Braun Medical
(high concentration) Ammonium chloride 14 g
Benzylidenechloride 10g
Bilateral guanidine acetate
7.5g
0.5g
Incidin Plus Glucoprotein 26.0g Ecolab Inc.
(high concentration) non-ionic surface active
agent
Pursept-A Alcohol 38.9 g Merz& Co.,
(Cooperate with Glyoxal 0.1 g
sterilized sponge) QAV 0.05 g

Warning :
Laryngoscopy should be one with a disinfectant cleaning, please read the cleaning requirements before
sterilized.
Please turn off the power before sterilized.
Handled and Monitor is not waterproof design, please avoid putting the main unit into the cleaner.

Washing
Please use the water to clean the blade and wipe dry with wet gauze until there is no stain. Use the brush to
clean the blade detailed location and use the gauze to make it dry. Please use the wet gauze to clean the
handle and monitor. And please avoid the liquid into unsealed internal.
Gauze should be used for one-time use, and brush should be one disinfected.
Please make it dry after washing.
Shenzhen HugeMed Medical Technical Development Co.,LTD
Add.:8/F,Tempus Building A, Qingshui River Road, Luohu District ,Shenzhen City,Guangdong Province,P.R.China.
Tel:0086-755-22275866 Fax:0086-755-82267835 Web: www.hugemed.net

1.1.1

Enzyme wash
For removing body fluid secretions and other organic matter in order to prevent the organic matter on
lessening disinfectants, Enzyme washing in time to avoid protein dried up, difficult to remove.
Multienzyme configuration and soaking time should be according to product specification

2, Please put the device to the Enzyme wash tank, and wipe with multiple enzyme lotion .

3, Accessories should be soak in the enzyme Lotion, and also need to clean by Ultrasonic cleaner for 5-10
minutes.

4,Multienzyme should be replaced each time after washing the blade.

1.1.2

Cleaning
Pleases use water cannons or wet gauze for cleaning and wipe the device surface until dry after sterilized
by the Multienzyme cleaner(cleaning time should be according to the user manual).

1.1.3

Sterilization

1.1.3.1

Soak by Antibacterial liquid

1, Please put the blade into the Sterilization bath or disinfection tub, and make it immersion in disinfectant.
Time: The sterilization time should not less than 20 minutes. The sterilization time to Special patients
infected with Mycobacterium tuberculosis, other mycobacteria should not less than 45 minutes.
2. Please use the wet gauze with 75% alcohol to wipe the monitor and handle after first water wiping.
3. The time for sterilization should be not less than 30 minutes.
4. Each time before treatment begins, the reusable blade used that day must be repeated sterilized .
Using the recommended soaking disinfectants with sterilization time of less than 20 minutes rinsing and
drying before for treating a patient.
5. If necessary sterilized laryngoscope using a 2% alkaline glutaraldehyde sterilization, must be soaked
for 10 hours.
Shenzhen HugeMed Medical Technical Development Co.,LTD
Add.:8/F,Tempus Building A, Qingshui River Road, Luohu District ,Shenzhen City,Guangdong Province,P.R.China.
Tel:0086-755-22275866 Fax:0086-755-82267835 Web: www.hugemed.net

1.1.3.2

Low-temperature plasma sterilizer, is not suitable for heat sterilization without moisture metals,
non-metallic precision medical instruments and supplies.

NOTICE
Peep blade containing electrical components should not be too low or too high
temperature during cleaning.
Sterilizing active ingredient is hydrogen peroxide. It is easy to penetrate the skin, when
contacted with a solution, must be equipped with latex or vinyl gloves.

1.1.4 Clean and dry


Remove the tank from the laryngoscope before disinfection, cleaning and disinfection of personnel gloves
should be replaced with sterile gauze or clean compressed air and other methods, and dried.

Warning

Peep chemical disinfectants immersed and sterilized

blade, prior to use must be thoroughly rinsed with sterile

water to remove residual disinfectant.

1.1.5 Storage

The disinfection or sterilization of the laryngoscope should be stored in a sealed box instrument after
sterilization.

Notice
The above operation is for reference only, through appropriate disinfection methods for
verification.
Shenzhen HugeMed Medical Technical Development Co.,LTD
Add.:8/F,Tempus Building A, Qingshui River Road, Luohu District ,Shenzhen City,Guangdong Province,P.R.China.
Tel:0086-755-22275866 Fax:0086-755-82267835 Web: www.hugemed.net

The reusable blade also could be Sterilization by EO, the process is as following:
Shenzhen HugeMed Medical Technical Development Co.,LTD
Add.:8/F,Tempus Building A, Qingshui River Road, Luohu District ,Shenzhen City,Guangdong Province,P.R.China.
Tel:0086-755-22275866 Fax:0086-755-82267835 Web: www.hugemed.net

Note: Please follow the recommendations of cleaning and sterilization declared

by the manufacturer and by the local health authority.

Shenzhen HugeMed Medical Technical Development Co., LTD


8F, Block A, Tempus Building, 1st Qingshuihe Road, Luohu District, Shenzhen City ,Guangdong, China
Tel : +86-755-22275866,22275899 Fax : +86-755-22275833 E-mail:server@hugemed.net