Professional Documents
Culture Documents
This section will provide you with a list of the order in which the equip-
ment should be: installed, setup, calibrated etc. After you have com-
pleted a step refer back to this page to see what the next step is.
There will be times when the manuals with need to be used in con-
junction with one another. This list is here to help you use the equip-
ment more efficiently, you should still read all of the manuals thor-
oughly before using this list. By reading all of the manuals and then
referring to this list you will have a better understanding of how the
equipment works.
1. First, adapt the generic chamber. In the “Adapting a Generic Chamber” manual, refer to the:
“Installation Overview” section, for an overview of how to install the generic chamber. Also use
the “Adapting a Chamber for a ProCO2” section for information and directions on how to
adapt the chamber for a ProCO2.
2. Now, installing the ProCO2. In the “ProCO2 Model 120” manual refer to the: “BioSpherix
Supplied Parts” and “User Supplied Parts” sections for a parts list. Also refer to the:
“Setup of Gas Supply”, “ProCO2 Front Panel Installation” and “ProCO2 Back Panel Instal-
lation” for how to make all the connections.
3. Next, familiarize yourself with the push button interface. In the “ProCO2 Model 120”
manual refer to the: “Operation”, “Program Functions Menu” and “Configuration” sections
for information regarding the push button interface.
4. Now, calibration. In the “ProCO2 Model 120” manual refer to the: “Calibration” section
for information on how to calibrate the sensor.
5. Next, tuning. In the “ProCO2 Model 120” manual refer to the: “Tuning” section for spe-
cific information on fine tuning. You shouldn’t have to use this section much, because most
tuning is done at the factory.
6. Now, operating. In the “ProCO2 Model 120” manual refer to the: “Single Setpoint Con-
trol” section for information on how to operate with the ProCO2. Also refer to the: “Setting
the Alarm Setpoint” section for directions for how to set a specific alarm setpoint.
7. Finally, maintenance issues. In all of the manuals refer to; “Maintenance” sections on
how to maintain the respective equipment.
The nominal range of carbon dioxide concetration is 0.1-20.0%. Advanced feedback algo-
rithms handle practically any carbon dioxide gas dynamic.
Infusion rates are adjustable up to 35 SCFH, sufficient for
most lab chambers. Installation is easy. Operation is simple.
The unit moves easily from one host chamber to another.
Please read and follow the safety and operations instructions
PO Box 87 on the following pages. Be careful. Any pressurized gas can
19 Demott St. be dangerous. Know what you are doing and do it safely.
Lacona, New York 13083
PHONE 315-387-3414
FAX 315-387-3415
TOLL FREE US/CAN 800-441-3414
www.biospherix.com
Safety Instructions........................................................................3
Operation........................................................................................9
Configuration...............................................................................11
Calibration...............................................................................12-13
Tuning......................................................................................14-17
Maintenance.................................................................................20
Warranty........................................................................................21
3
ProCO2 model 120
version 1.0
SAFETY INSTRUCTIONS
ELECTRICAL POWER SOURCE unit should be connected to a power supply only of the
type described in the operating instructions or as marked on unit.
POWER CORD PROTECTION power supply cords should be routed so that they are not
likely to be walked on or pinched by items placed upon or against them.
ELECTRIC SHOCK do not remove cover of ProCO2 due to presence of uninsulated "dan-
gerous voltage" within product's enclosure.
ELECTRIC SHOCK unit should never be used where it can fall or be pushed into water.
ELECTRIC SHOCK if modifying host chamber, be careful not to drill or cut into electrical
wires hidden behind chamber wall. Never drill or cut blindly.
NONUSE PERIODS power cord of unit should be unplugged from electrical outlet when left
unused for long period of time.
FALLING OBJECTS AND LIQUID SPLASH care should be taken so that objects do not fall
on equipment and liquids do not spill, splash, or drip onto or into unit enclosure or
power cord.
PRESSURIZED GAS secure all connections with hose clamps. Never exceed pressure
limits. Bleed all lines before disconnecting. Wear safety glasses at all times.
LOW OXYGEN ATMOSPHERES never enter a chamber which has a low oxygen atmo-
sphere because of severe danger of suffocation. Host chamber should be in well venti-
lated room. Control gas (nitrogen or other low oxygen gas) continuously leaks out of
chamber and should never be allowed to build up in room outside of chamber.
HIGH OXYGEN ATMOSPHERES never enter a chamber which has a high oxygen atmo-
sphere due to danger of oxygen toxicity. Never smoke or allow any source of fire or
spark in or around a chamber with high oxygen atmosphere. Oxygen radically promotes
combustion and can be explosive. Host chamber should always be in well ventilated
room. Oxygen continuously leaks out of chamber and should never be allowed to build
up in room outside chamber.
VENTILATION unit should be situated so that its' location or position does not interfere
with proper ventilation. Neither ProCO2 nor host chamber should be in poorly ventilated
areas.
HEAT unit should be situated away from heat sources such as radiators, heat registers,
stoves, or other appliances or processes that produce heat.
CLEANING do not immerse unit in water. Do not wipe unit with wet cloth or sponge or
paper. Clean only with a dry cloth.
4
BioSpherix Supplied Parts ProCO2 model 120
version 1.0
This section will familiarize you with the components that come with
the ProCO2.
ProCO2 unit
This section will list the parts that the user will need to supply in order to
operate the equipment.
1. Qty. 2 regulators, either a one stage regulator or a two staged regulator. Either way you
will need two gauges, one for the amount of gas in the tank and one for the gas coming out
of the tank. (2500 PSIG input 0-60 PSIG output, recommended). PSIG stands for Pounds
per Square Inch Gauge as opposed to PSIA which stands for Pounds per Square Inch
Absolute.
2. Qty. 1 compressed gas source (tank) of carbon dioxide.
3. Qty. 1 compressed gas source (tank) of carbon dioxide/oxygen mix (ratio should be 10-
90).
4. Qty. 1 compressed gas source (tank) of nitrogen (optional, see “Calibration” section for
information).
6
Setup of Gas Supply ProCO2 model 120
version 1.0
This section will describe how to setup the gas supply, for information
on how to connect the gas supply to the machine, see the calibration
sections and the “Single Setpoint Control” section.
Control gas must be supplied through a 1/8 in. ID tube to the back panel of the
ProCO2. If the tubing coming from the compressed gas source (tank) isn’t 1/8 in. then use
one of the provided gas fittings to adapt the tubing to 1/8 in. The pressure must be regu-
lated to 0-25 PSIG. For best results use approximately 2 PSIG in aCulture Chamber and
use 15 PSIG in an Animal Chamber. (refer to “Tuning” section under the “Power” heading
for further information). Do not connect any tubing yet, wait until told to do so in the “Cali-
bration” section and “Single Setpoint Control” section.
The amount of gas used is determined by how the chamber is used, not the ProCO2.
The ProCO2 uses the least amount of gas possible, which is only what the chamber needs.
The amount of gas used is dependent on: (1) The size and leakiness of the chamber, (2)
The amount of times and how long the chamber door(s) are opened, (3) The carbon diox-
ide level being controlled.
Use a one or a two stage, 2500 PSIG input, 0-60 PSIG output regulator at the source
of the compressed gas. It is best to have two gauges so that you can monitor the amount
of gas in the compressed source and also monitor the amount passing through the output.
Never allow the pressure coming out of the compressed source to exceed 25 PSIG or
damage will occur to the ProCO2. If the regulator is not near the ProCO2 then there
should be a shutoff valve placed upstream of the unit. When taking off gas supply tube
always make sure to shut off compressed gas at the source first, then bleed the pressure
out of the line and then take off the tube from back of ProCO2.
Step-by-Step Procedure for Connecting Gas to Machine
1. Before connecting the regulator to the tank, close the regulator and make sure the tank
is also closed off.
2. Connect the regulator to the compressed nitrogen (if applicable), compressed carbon
dioxide and to the compressed carbon dioxide/oxygen mix gas sources.
3. Connect the tubing from each compressed gas source to the ProCO2 when told to do so
in the calibration procedures and the operation procedures.
4. At this point you don’t have to open the regulator or turn on the gas. You will be turning
the gas on at the beginning of the calibration/single setpoint control procedures, for now
just connect the regulator.
7
ProCO2 Front Panel Installation ProCO2 model 120
version 1.0
The following steps will explain how to set up the ProCO2 front panel
and prepare it for installation to a chamber.
1. Placing the Unit
Place the unit on a level secure surface either on the
chamber or within seven feet of the chamber. Make sure
you can easily access the front panel.
This section will describe how to setup the back panel of the ProCO2.
1. Power Supply
Plug the jack end into the port labeled “12VDC” on
back of ProCO2. Plug the 12VDC power supply into a
wall outlet. The display on the front of ProCO2 should
turn on.
2. Connecting Gas
Attach 1/8 in. ID tubing to hose barb labeled “IN” on
the back of the ProCO2. Making sure that the gas com-
ing in Never exceeds 25 PSIG or damage will occur to
the ProCO2. Depending on if you are calibrating or
operating will determine what type of gas you will be 1
using. (see “Calibration” and “Single Setpoint Control” 2
sections).
3. Connecting Sensor
Plug the sensor cable into the port labeled
“SENSOR” on back of ProCO2 by twisting the sensor
cable into the port until it pushes in, then push in until
“click is heard. When removing pull the ring back and
pull cable off. Plug the other end of the sensor cable into
the sensor itself by lining up the key on the terminal with
the key on the connector of the sensor cable. Push in
and twist the ring on. The display on the front of the
ProCO2 should respond.
This section will give an overview of some of the different modes and
functions within the ProCO2
CHANGE LEVELS
Multi-level menu of functions requires
navigating from level to level.
Index down key to
level function.
Ú
Ú
Ø
Read control SP1 Read SP1 control
Select Autotune or Park output device
output power %
×
Proportional Band / Gain SP1 manual power Read SP2 alarm
or Dead band width control % output device
Offset / Manual reset Select display resolution Sensor zero adjust Derivative sensitivity
(decimal point)
Lock main setpoint SP1 Set scale maximum Select control Display sensitivity
accuracy monitor
there's one way in. Menu items are loosely grouped together according to
bottom of level 1. Exit menu from any point. Level 4 is hidden for security;
PROGRAM MODE FUNCTIONS MENU is arranged on 4 levels. Entry into menu is
Functions reset
version 1.0
10
ProCO2 model 120
CONFIGURATION 11
ProCO2 model 120
version 1.0
DERIVATIVE SENSITIVITY
This parameter compensates
when power/load balance is such
that derivative suppresses overshoot
but does not sufficiently speed up
response to upsets.
Derivative sensitivity is frac-
tion multiplier of derivative time.
Range is 0.1 - 1.0. By fractioning
derivative time it speeds action up to
10 times faster.
STEPS:
1. With power, proportional, and
integral parameters tuned, and
derivative time tuned for over-
shoot suppression, watch
control process in work mode,
introduce upset by opening
chamber door(s), and note
speed of response to upset.
18
Single Setpoint Control ProCO2 model 120
version 1.0
This section will describe how to use single setpoint control with the
ProCO2. Single setpoint control is a way to control the level of gas in
the chamber. For more information, see the “Operation” section of this
manual. The ProCO2 can only use 0-20 for a set point. Other wise the
sensor will overload.To set a setpoint on the controller, follow these
steps:
1. Make sure the compressed carbon dioxide gas supply is connected to the back panel of
the ProCo2 at the “IN” barb.
2. Open the regulator on the compressed carbon dioxide source to 0-25 PSIG, Never
exceed 25 PSIG or damage will occur to the unit. For the best results use approxi-
mately 2 PSIG in a Culture Chamber and 15 PSIG in an Animal Chamber when operating.
3. Check to make sure that the gas is connected properly by opening the bleed valve on
the front panel. If gas is heard expelling from the “BLEED” barb then it is connected cor-
rectly. Once you have confirmed that the gas is connected properly, close the bleed valve.
4. To set a setpoint, push and hold the “*” button. While holding the “*” button use the up
and down buttons to adjust the “set” to read what you want the setpoint to be. Once you
have adjusted the number to the desired reading, release the “*” button and the controller
will begin to go towards the new setpoint. Remember you can only use 0-20 for a setpoint
anything higher will overload the sensor.
5. Flip the switch on the front panel labeled “GAS” upward. This will allow the gas to enter
the chamber.
19
Setting the Alarm Setpoint ProCO2 model 120
version 1.0
This section will explain how to set the alarm setpoint. Remember,
when you want the alarm function to be active you must make sure the
alarm switch on the front of the controller is on.
1. Push the up and down buttons at the same time for 3 seconds.
2. Now, go to “LEVL 2”. In ‘LEVL 2” navigate to “SP2.A”. Change this number to either “DV.hi”
(Deviation High) or “DV.Lo” (Deviation Low). This will determine whether the alarm activates
below or above the “SEt.2” (alarm setpoint).
3. In “LEVL 1” navigate to “SEt.2”. Adjust this number by holding the “*” button and using the up
or down buttons. Change this number to the desired alarm setpoint.
4. In “LEVL 1” navigate to “bnd.2”. This number will determine the range of the alarm setpoint.
The “bnd.2” number is a coarse adjustment, the number that you change this to will
not be the exact size of the band width. To set the alarm to activate as quickly as possible,
set the band to “0.1”. Increase the number to increase the range. Remember it is a coarse
adjustment.
5.Remember, if you want the alarm to have the ability to come on you must flip the “Alarm”
switch on the front panel.
20
Maintenance ProCO2 model 120
version 1.0
This section will describe how to check calibrate, recalibrate and how to
replace the sensor.
New Sensor
In the event that the sensor quits functioning call the number on the side of the sensor
to purchase a new one. To replace the sensor follow step “6” of “ProCO2 Installation”.
21
WARRANTY ProCO2 model 120
version 1.0
BIOSPHERIX, LTD.
ONE-YEAR WARRANTY
Except to the extent specifically stipulated, BioSpherix, Ltd. shall have no
liability or obligation under warranty, express or implied, including the implied
warranty of merchantability and any implied warranty of fitness for a particular
purpose; statutory or otherwise except as stated.
BioSpherix, Ltd. (BioSpherix) warrants that all apparatus of its manufacture has
been factory inspected and is free from defects in material and workmanship. And that
when such apparatus receives normal use and service, BioSpherix will correct defects
in material and workmanship that might occur within a twelve (12) month period from
the date of delivery to customer. This warranty is limited to the repair, replacement or
exchange of parts, subject to the exceptions listed below, which prove defective on
examination by BioSpherix. Costs assumed by BioSpherix under this warranty cover
only the cost of material and workmanship. This warranty does not cover sensors.
Manual
version 0.2 May 2006
neric Chamber
version 0.2
This manual will describe how to install and adapt a generic chamber.
Read this manual thoroughly before installing and adapting.
PO Box 87
19 Demott St.
Lacona, New York 13083
PHONE 315-387-3414
FAX 315-387-3415
TOLL FREE US/CAN 800-441-3414
www.biospherix.com
neric Chamber
version 0.2
Safety Instructions.........................................................................................3
Installation Overview.....................................................................................4
POWER CORD PROTECTION power supply cords should be routed so that they are not
likely to be walked on or pinched by items placed upon or against them.
ELECTRIC SHOCK if modifying host chamber, be careful not to drill or cut into electrical
wires hidden behind chamber wall. Never drill or cut blindly.
LOW OXYGEN ATMOSPHERES never enter a chamber which has a low oxygen atmo-
sphere because of severe danger of suffocation. Host chamber should be in well ventilated
room. Control gas (nitrogen or other low oxygen gas) continuously leaks out of chamber
and should never be allowed to build up in room outside of chamber.
HIGH OXYGEN ATMOSPHERES never enter a chamber which has a high oxygen at-
mosphere due to danger of oxygen toxicity. Never smoke or allow any source of fire or
spark in or around a chamber with high oxygen atmosphere. Oxygen radically promotes
combustion and can be explosive. Host chamber should always be in well ventilated room.
Oxygen continuously leaks out of chamber and should never be allowed to build up in room
outside chamber.
VENTILATION unit should be situated so that its’ location or position does not interfere with
proper ventilation. Host chamber should never be in a poorly ventilated area.
HEAT host chamber should be situated away from heat sources such as radiators, heat
registers, stoves, or other appliances or processes that produce heat.
HIGH CONCETRATION OF GAS never adapt host chamber when gas is evident inside of
chamber.
DRILLING PRECAUTIONS keep a firm grasp on the drill at all times, drill slowly and cau-
tiously. Beware of sharp drill bit.
MAKING CUTOUT FOR ADAPTER PLATE the cutout for the adapter plate should only be
made by an experience fabricator.
Installation Overview
Adapting a Ge-
neric Chamber
version 0.2
If the chamber being used is not one from BioSpherix then follow these
guidelines to adapt the chamber for a BioSpherix unit.
The chamber must be in a well ventilated room. Control gas will continually leak out of
chamber and should never be allowed to build up outside of chamber.
The chamber must be semi-sealed. If the chamber is sealed too tight then it may de-
velop positive pressure which sensors can’t tolerate. Positive pressure is when there is no
small opening for the gases inside the chamber to vent out or for the outside air to vent in.
The chamber can’t be too leaky or you won’t be able to hold an inner atmosphere inside of
the chamber.
It is important to have leaks when you are doing an experiment that deals with both
low levels of oxygen and high levels of oxygen. With a low level oxygen experiment you
will be infusing nitrogen (or an oxygen poor gas), which will bring the level of oxygen down,
and through the leak(s) the outside air will be bringing up the level of oxygen. This is im-
portant because without the outside air coming in, the oxygen level would drop to 0% and
stay there. With a high level of oxygen experiment, you will be infusing oxygen (or an
oxygen rich gas), which will bring the level of oxygen in the chamber up, and through the
leak(s) the outside air will be bringing down the level of oxygen, because the outside air is
made up of only 20.8% oxygen. This is important because without the outside air coming
inside the chamber the oxygen level would rise to 100% and stay there.
Plan the best access point(s) into the chamber for the sensor and the infusion tube.
Any passageway will work as long as the sensor can monitor the inside of the chamber and
gas can be infused into the chamber.
There are three ways to install the sensor and infusion tubing: either on the
chamber, in the chamber or by mounting a BioSpherix adapter plate to the cham-
ber (see the appropriate unit’s section for information on how to adapt the chamber three
ways). Along with the instrument you will also receive a muffler, which can be attached to
the end of the infusion tube if you wish.
Muffler
Adapting a Chamber for a ProCO2 Adapting a Ge-
neric Chamber
version 0.2
There are three ways to adapt the generic chamber for a ProCO2. One
way is having the sensor inside of the chamber and routing the cables
through an access port. The second way is to drill a hole in the cham-
ber to accomodate a port that is supplied by BioSpherix. The third
way is to make a cutout on the chamber that will accomodate a Bio-
Spherix adapter plate.
Inside of Chamber:
The first way to install the sensor is inside of the chamber. You will have to drill a hole
(specific size, below) so that the sensor cable and the infusion tube can both fit through.
Once the hole is drilled you can slide the sensor cable then the infusion tube through the
opening. Then, you can connect the sensor to the sensor cable inside of the chamber and
attach the muffler to the infusion tube. By installing the sensor in this manner you have cre-
ated a small leak for the gas inside to vent out and the outside air to vent in. You may need
more of a leak depending on the size of the chamber.
IMPORTANT: If you place the sensor inside the chamber it must be placed in such a
way so that it can be pulled out of the chamber for calibration purposes. In order to cali-
brate the sensor it has to be outside the chamber in the outside air.
Use the following drawing to determine the correct size of the holes and
the cutout size you will need to fit the BioSpherix adapter plate to the
chamber. You may not need to use the supplied muffler if you are
adapting the chamber in this manner.
Adapting a Chamber for a Proox 110
Adapting a Ge-
neric Chamber
version 0.2
There are three ways to adapt the generic chamber for a Proox 110. One
way is having the sensor inside of the chamber and routing the cables
through an access port. The second way is to drill a hole in the cham-
ber to accomodate a port that is supplied by BioSpherix. The third
way is to make a cutout on the chamber that will accomodate a Bio-
Spherix adapter plate.
Inside of Chamber:
The first way to install the sensor is inside of the chamber. You will have to drill a hole
(specific size, below) so that the sensor cable and the infusion tube can both fit through.
Once the hole is drilled you can slide the sensor cable then the infusion tube through the
opening. Then, you can connect the sensor to the sensor cable inside of the chamber and
attach the muffler to the infusion tube. By installing the sensor in this manner you have cre-
ated a small leak for the gas inside to vent out and the outside air to vent in. You may need
more of a leak depending on the size of the chamber.
IMPORTANT: If you place the sensor inside the chamber it must be placed in such a
way so that it can be pulled out of the chamber for calibration purposes. In order to cali-
brate the sensor it has to be outside the chamber in the outside air.
Use the following drawing to determine the correct size of the holes and
the cutout size you will need to fit the BioSpherix adapter plate to the
chamber. You may not need to use the supplied muffler if you are
adapting the chamber in this manner.
Adapting a Chamber for a Proox 360 Adapting a Ge-
neric Chamber
version 0.2
There are three ways to adapt the generic chamber for a Proox 360. One
way is having the sensor inside of the chamber and routing the cables
through an access port. The second way is to drill a hole in the cham-
ber to accomodate a port that is supplied by BioSpherix. The third
way is to make a cutout on the chamber that will accomodate a Bio-
Spherix adapter plate.
Inside of Chamber:
The first way to install the sensor is inside of the chamber. You will have to drill a hole
(specific size, below) so that the sensor cable and the infusion tube can both fit through.
Once the hole is drilled you can slide the sensor cable then the infusion tube through the
opening. Then, you can connect the sensor to the sensor cable inside of the chamber and
attach the muffler to the infusion tube. By installing the sensor in this manner you have cre-
ated a small leak for the gas inside to vent out and the outside air to vent in. You may need
more of a leak depending on the size of the chamber.
IMPORTANT: If you place the sensor inside the chamber it must be placed in such a
way so that it can be pulled out of the chamber for calibration purposes. In order to cali-
brate the sensor it has to be outside the chamber in the outside air.
Use the following drawing to determine the correct size of the holes and
the cutout size you will need to fit the BioSpherix adapter plate to the
chamber. You may not need to use the supplied muffler if you are
adapting the chamber in this manner.
Mounting a Pod 11
Adapting a Ge-
to the Chamber for a Proox C21 neric Chamber
version 0.2
Use the following drawing to determine the correct size of the holes and
the cutout size you will need to fit the BioSpherix Proox Model C21’s
Pod to the chamber.