You are on page 1of 76

®

Maintenance manual medinCNO

Maintenance manual medinCNO®

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 1 of 76
®
Maintenance manual medinCNO

®
The medinCNO contains batteries and electrical components. Consequently it cannot be
disposed of in domestic waste but must be collected separately and recycled in
accordance with local regulations.

Classification (according to COUNCIL DIRECTIVE 93/42/EEC Annex IX): IIb

Medin Medical Innovations GmbH


Adam-Geisler-Str.1 – 82140 Olching – Germany
+49 8142-44846-0 – www.medingmbh.com – info@medingmbh.com
Made in Germany

®
Maintenance manual of medinCNO Rev 06: 24.04.2015

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 2 of 76
®
Maintenance manual medinCNO

Contents

Contents .................................................................................................................................................. 3
1 General liability conditions ............................................................................................................... 6
2 Description of the device and conditions for use ............................................................................. 7
2.1 Intended use ............................................................................................................................ 7
2.2 Operation by staff .................................................................................................................... 7
2.3 Alternative respiration system ................................................................................................. 7
2.4 Combination with other devices and ancillary equipment ....................................................... 7
2.5 Technical description of function ............................................................................................. 7
2.5.1 Principles of function........................................................................................................ 7
2.5.2 Technical data ................................................................................................................. 8
2.5.3 Overview of subsections of the general technical description ......................................... 9
®
3 Cleaning, Maintenance and repairs of the medinCNO ................................................................ 10
3.1 Cleaning ................................................................................................................................. 10
3.2 Regular maintenance (replacement of the internal batteries and filters)............................... 10
®
3.3 Repairs of the medinCNO .................................................................................................... 11
®
3.4 Modifications of the medinCNO ........................................................................................... 11
®
4 Installing the medinCNO .............................................................................................................. 12
4.1 Mounting ................................................................................................................................ 12
4.2 Power supply ......................................................................................................................... 12
4.2.1 Mains operation ............................................................................................................. 12
4.2.2 Battery operation ........................................................................................................... 13
4.3 Gas supply ............................................................................................................................. 14
4.4 External alarm interface and data interface ........................................................................... 15
4.4.1 External alarm ................................................................................................................ 15
4.4.2 USB data interface......................................................................................................... 15
4.4.3 RS232 data interface ..................................................................................................... 20
4.5 Environmental conditions (usage, transport and storage) ..................................................... 20
5 Service Menu ................................................................................................................................. 22
5.1 Start Service Menu ................................................................................................................ 22
5.2 Service Menu overview ......................................................................................................... 22
5.3 Device name .......................................................................................................................... 23
5.4 Date and time ........................................................................................................................ 23
5.5 Language ............................................................................................................................... 23
5.6 Pressure unit.......................................................................................................................... 23
5.7 External Alarm ....................................................................................................................... 23

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 3 of 76
®
Maintenance manual medinCNO

5.8 FiO2 Alarm ............................................................................................................................. 24


5.9 Oxygen calibration ................................................................................................................. 24
5.10 Device check ......................................................................................................................... 24
5.11 Last maintenance .................................................................................................................. 24
5.12 Self check pneumatic unit ...................................................................................................... 25
5.13 Internal Data .......................................................................................................................... 25
6 Device check ................................................................................................................................. 27
7 Last Maintenance .......................................................................................................................... 32
®
8 Software update of medinCNO .................................................................................................... 33
8.1 Update of main CPU software ............................................................................................... 33
8.1.1 Necessary material ........................................................................................................ 33
8.1.2 Set up ............................................................................................................................ 33
8.1.3 Execution ....................................................................................................................... 33
8.2 Display software update ........................................................................................................ 34
8.2.1 Necessary material ........................................................................................................ 35
8.2.2 Set up ............................................................................................................................ 35
8.2.3 Execution ....................................................................................................................... 35
8.3 Update of the pneumatic software ......................................................................................... 36
8.3.1 Necessary material ........................................................................................................ 36
8.3.2 Set up ............................................................................................................................ 36
8.3.3 Execution ....................................................................................................................... 36
9 Alarms and error messages .......................................................................................................... 39
®
9.1 Failures during the medinCNO start up ............................................................................... 39
®
9.2 Alarms during the use of the medinCNO ............................................................................. 40
9.3 Emergency programme ......................................................................................................... 44
9.4 Signal lights ........................................................................................................................... 45
®
10 Steps during repairs and maintenance of medinCNO ................................................................. 46
10.1 Exchanging the oxygen cell ................................................................................................... 46
10.2 Opening of the case .............................................................................................................. 46
10.3 Exchanging the internal rechargeable battery ....................................................................... 47
10.4 Exchanging the internal 3V battery........................................................................................ 47
10.5 Exchanging the alarm LED .................................................................................................... 48
10.6 Exchanging the knob ............................................................................................................. 49
10.7 Exchanging the fan ................................................................................................................ 50
10.8 Exchanging the connection board ......................................................................................... 50
10.9 Exchanging the pneumatic unit ............................................................................................. 51
10.10 Exchanging the main CPU .................................................................................................... 53
10.11 Mounting the additional power supply holder ........................................................................ 55

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 4 of 76
®
Maintenance manual medinCNO

10.12 Mounting the optional RS232 interface ................................................................................. 55


10.13 Closing of the case ................................................................................................................ 56
10.14 external filter .......................................................................................................................... 57
10.14.1 Exchange of the external filter ................................................................................... 57
10.14.2 Upgrade of the external filter ..................................................................................... 57
11 Spare Parts and accessories ......................................................................................................... 59
11.1 List of spare parts .................................................................................................................. 59
11.2 Fuses ..................................................................................................................................... 60
12 Electromagnetic compatibility ........................................................................................................ 61
12.1 Electromagnetic transmission ................................................................................................ 61
12.2 Electromagnetic immunity ..................................................................................................... 61
13 Symbols ......................................................................................................................................... 65
14 Translations of the error messages into different languages ........................................................ 67
15 Device Passport............................................................................................................................. 73
16 History of changes ......................................................................................................................... 76

Attention:
® ®
Whenever the case of medinCNO is opened or whenever anything on medinCNO is modified, a
®
device check of medinCNO has to be performed and its results have to be documented on the
® ®
medinCNO device passport. Otherwise the medinCNO cannot be used and it must not be
connected to a patient under any circumstances.

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 5 of 76
®
Maintenance manual medinCNO

1 General liability conditions


The general terms of trade of medin Medical Innovations GmbH (hereafter referred as medin) are
binding.
®
Medin accepts no responsibility or liability for the safe use of the medinCNO if the device is serviced
or repaired by individuals who have not received appropriate specialist training or if an activity that
does not comply with correct use is performed. Medin also accepts no responsibility or liability for the
®
safe use of the medinCNO if the specified function tests and maintenance are not carried out. The
user is responsible for observing maintenance and function test intervals.
Medin accepts no liability for damage caused by failure to comply with these instructions.
®
If maintenance or repair of the medinCNO is performed not by medin, the organisation which
performs it is responsible for its correct execution and all actions which are performed thereby.
®
Modifications of the medinCNO
Warning:
®
The medinCNO must not be changed or modified without the permission of the manufacturer medin
and without an adequate and full checkup and function test afterwards to ensure the further safe
usage of the device.

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 6 of 76
®
Maintenance manual medinCNO

2 Description of the device and conditions for use


2.1 Intended use
® ®
The medinCNO CPAP driver is used in combination with the Medijet nCPAP generator to administer
®
CPAP therapy to premature infants and newborns. The medinCNO must be used under the
supervision of expert, specially trained staff in a clinical setting, and the patient’s oxygen saturation
must be monitored at the same time.
Warning:
®
- The medinCNO is intended for clinical use only.
®
- The medinCNO may only be used while the patient’s oxygen saturation is being monitored at
the same time.
®
- The medinCNO may only be used by and under the supervision of expert staff who have
been trained in the device.
®
- For use only in combination with the Medijet nCPAP generator.
®
- The medinCNO is a CPAP device. It is suitable only for patients who are breathing
spontaneously and not for patients who are dependent on artificial respiration.
®
- The medinCNO must not come into contact with flammable anaesthetics or flammable
agents during use.
- According to U.S. law, may this device only be sold on the prescription of a doctor (for USA:
Rx only).

2.2 Operation by staff


®
The medinCNO must be operated by a doctor or by an individual with the necessary technical skills
acting on a doctor’s instructions. Consequently, all staff operating the device must have a detailed
knowledge of these instructions for use and have participated in training in the device.
2.3 Alternative respiration system
An alternative CPAP system or respiration system (e.g. bag valve unit) must always be available while
®
the medinCNO is in use.
2.4 Combination with other devices and ancillary equipment
®
The medinCNO may only be combined with devices or consumables (e.g. CPAP ancillary equipment)
not listed in these instructions for use after consulting the manufacturer.
Use of ancillary equipment that is not approved puts the patient at risk and may impair the proper
functioning of the device.
2.5 Technical description of function
2.5.1 Principles of function
® ®
The medinCNO is a CPAP driver that can be used in combination with the Medijet nCPAP generator
®
to administer CPAP therapy. The role of the medinCNO in this CPAP system is to provide the
necessary oxygen-enriched driving flow which is fed to the nCPAP generator via the tubes connected
to the patient and is converted into CPAP pressure within the generator.
An internal electronic blender combines the air and oxygen coming from an external source and
administers the total volume so that the driving flow reaching the patient can be enriched with oxygen
as necessary to any oxygen level between 21% and 100%.
The flow volume of the gas mix within a range of 4 l/min to 15 l/min can be set depending on the mode
®
selected. According to the flow volume set, the medinCNO provides a constant drive flow. This
® ®
generates a constant CPAP pressure in the Medijet CPAP generator, or the medinCNO assesses
®
the CPAP pressure signal measured in the Medijet in order to react to the patient's breathing and
trigger a higher flow impulse in the event of apnoea or synchronously with the patient's inhalation. This
®
triggers a pressure impulse with elevated CPAP pressure in the Medijet CPAP generator. The

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 7 of 76
®
Maintenance manual medinCNO

®
medinCNO can also trigger oscillation between 5 Hz and 20 Hz in the CPAP pressure applied by
®
Medijet by means of an oscillating drive flow.
®
The monitor which is part of the medinCNO displays the parameters selected by the user, the
®
sensors’ readings, the length of time that CPAP pressure has been applied by Medijet in the form of a
graph, and any alarms.
®
The medinCNO can operate for up to two hours without mains power thanks to its rechargeable
batteries.
Warning:
®
Changes may not be made to the medinCNO without the authorization of the manufacturer medin
and without a suitable complete analysis and functional check following the change which ensure
further safe usage.
2.5.2 Technical data

Dimensions (L x W x H) 29 x 23.5 x 18.5 cm


Weight: 4.75 kg
Display 7.0“ – colour, 800 x 480 pixel
Power supply: 100 to 240 V AC / 50 to 60 Hz
Internal rechargeable battery 14.4 V DC, 2 hours run time
Gas connections: Air supply: 300 to 600 kPa (= 3.0 to 6.0 bar)
Oxygen: 300 to 600 kPa (= 3.0 to 6.0 bar)
Connector standard: DISS or NIST (as preferred)
Patient flow outlet Dimensions M22 (OD) or F15 (ID)
CPAP pressure meter Luer – 4.3 mm ID
connection
Flow: Setting range: 0 L/min to 15 L/min
Working range: 4 L/min to 10 L/min
Accuracy: ±1 L/min (in working range) ±2 L/min (outside of working
range)
Basis of measurement: DIN 1343 at 0°C, 1013.25 mbar and 0%
humidity (0/1013)
Oxygen concentration: Setting range: 21% to 100% (in flow operating range)
Measurement range: 21% to 100%
Accuracy ± 3% (Vol.)
Measurement sensor OOM 102 or MLF-16
Pressure measurement: Measurement range 0 mbar to 15 mbar - pressure graph
Accuracy ±1.3 mbar
Verification: Redundant measurement by two sensors
Push (Inspiration support Setting range
®
through the medinCNO ) Additional flow during inspiration push:
min: 0 L/min
max: 17.5 L/min minus adjusted basic flow
Duration 200 ms to 2 seconds
Accuracy
± 1 L/min (if the complete flow = basic flow + additional
inspiration flow - is within the operating range)
± 2 L/min (if the complete flow = basic flow + additional
inspiration flow - is outside of operating range)
Basis of measurement: DIN 1343 at 0°C, 1013.25 mbar and 0%
humidity (0/1013)

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 8 of 76
®
Maintenance manual medinCNO

Electronic shut-off valve In the event of an error, this valve interrupts the flow supply to the
patient and opens the tubing system to the atmosphere
Mechanical overpressure valve Opening pressure 4 kPa (= 40 mbar)
®
Operation time The medinCNO is suitable for continuous operation for up to 4
weeks without a restart during this time.

2.5.3 Overview of subsections of the general technical description


Information about these topics can be found in the following chapters:
 Environmental conditions for usage, transport and storage – see chapter 4.5

®
Characteristics and accuracies of the medinCNO – see chapter 2.5.2
 Information about the set-up of the device - see chapter 4
 Definition of the necessary gas supply – see chapter 4.3
 Description of the electric supply – see chapter 4.2
 The device is disconnected from the mains by pulling the power supply unit cable of the
® ®
medinCNO out of the electrical outlet. Thus when installing the medinCNO , it should be
ensured that this plug is always freely accessible and can be disconnected from the mains
®
without difficulty. The medinCNO does not contain any switch which disconnects it from
the mains.

®
Changes and modifications of the medinCNO are not permitted without the explicit
authorisation of the manufacturer medin – see chapter 3.4

®
Maintenance, repairs and replacement of parts of the medinCNO may only be performed
by trained and competent service technicians and only according to the instructions in this
maintenance manual and following all the warnings here and inside the operation manual
®
of the medinCNO .

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 9 of 76
®
Maintenance manual medinCNO

3 Cleaning, Maintenance and repairs of the medinCNO®


Attention:
® ®
Whenever the case of the medinCNO is opened or whenever anything on the medinCNO is
®
modified, a device check of the medinCNO has to be performed and its results have to be
® ®
documented on the medinCNO device passport. Otherwise the medinCNO cannot be used and it
must not be connected to a patient under any circumstances.
®
Additionally the medinCNO works with oxygen and for this reason, it must be ensured that no part of
®
the medinCNO which can come in contact with oxygen becomes contaminated with oil, grease or any
other substance which is not qualified to be used in contact with oxygen.
3.1 Cleaning
®
The surface of the medinCNO should be disinfected before first use and every time after it has been
used on a patient. 70% isopropyl alcohol solution can be used for this purpose.
Attention:
- The device must never be sterilised or immersed in a liquid solution.
- No abrasive substances may be applied to the surface.
- No liquid may penetrate the device.
3.2 Regular maintenance (replacement of the internal batteries and filters)
®
The medinCNO must undergo maintenance and a function check in accordance with the
manufacturer's instructions every 12 months. This must be carried out by trained specialists using
appropriate measuring and testing equipment.
The following actions must be performed during this maintenance:
- Replacement of the oxygen cell
o Type OOM-102 (REF OOM-102) oxygen cells must be replaced annually. (see
chapter 10.1)
o Type MLF-16 (REF MLF-16) oxygen cells must be replaced if they have less than one
year of service life left or no later than after three years. (see chapter 10.1)

- Test of the device’s functionality and sensor calibration (see chapter 6) – every year and after
every exchange of a component.
- The internal 3-volt battery must be replaced as required or every three years at the latest. The
replacement battery must be the type CR123A powerone of manufacturer VARTA
Microbattery GmbH (REF 39-102) (see chapter 10.4).
- The internal rechargeable battery must be replaced as required or every three years at the
latest. The replacement battery must be a rechargeable battery of the same type (can be
ordered from medin, REF 39-101) (see chapter 10.3).
®
- If the medinCNO contains a water trap, the filter contained within it (REF 39-819) should be
changed as needed (if soiled) or no later than after three years.

If an error is detected during this function test and maintenance, the device must not be used or
connected to a patient under any circumstances.
Warning:
The two internal batteries must be replaced by the batteries mentioned above and by no other
batteries. The battery replacement must be performed by trained and skilled service technicians and
according to the instructions in this maintenance manual.
®
After each maintenance or repair of the medinCNO , a complete function test has to be performed and
®
passed before the medinCNO can be used for a patient again.
Attention:
If an error is detected during this function test and maintenance, the device must not be used or
connected to a patient under any circumstances.

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 10 of 76
®
Maintenance manual medinCNO

3.3 Repairs of the medinCNO®


®
This manual provides information about the repair and maintenance of the medinCNO . On request
and for repairs which can be made by trained service personnel in the field, medin will provide you
with further information such as circuit diagrams, further descriptions or descriptions of calibrations
and will support you during the repair.
Warning:
®
Repairs of the medinCNO must be performed by trained and skilled personnel only. During these
repairs, all instructions and warnings in this maintenance manual have to be followed. After each
®
maintenance of or repair to the medinCNO , a complete function test has to be performed and passed
®
before the medinCNO can be used for a patient again. If an error is detected during this function test
and maintenance, the device must not be used or connected to a patient under any circumstances.

Medin accepts no responsibility for any actions performed during these repairs. If the maintenance or
®
repair of the medinCNO is not performed by medin, the organisation which performs it is responsible
for its correct execution and all steps which are performed thereby.
Repairs which are performed by organisations outside medin are limited to replacements of whole
parts as described in this manual (see chapter 10). It is not possible to repair defects of the individual
®
parts of the medinCNO . For this reason, it is possible to only replace the following groups completely
and it is not possible to repair defects within them:
- Pneumatic unit
- Connection board
- Main CPU
- Alarm LED
- Knob
- Rechargeable battery
- Fan
® ®
After each repair, a medinCNO device check has to be performed. Otherwise the medinCNO cannot
be used for a patient.
Attention:
® ®
Whenever the case of the medinCNO is opened or whenever anything on the medinCNO is
®
modified, a device check of the medinCNO has to be performed and its results have to be
® ®
documented on the medinCNO device passport. Otherwise the medinCNO cannot be used and it
must not be connected to a patient under any circumstances.
3.4 Modifications of the medinCNO®
Warning:
®
The medinCNO is not allowed to be modified without the authorisation of the manufacturer medin and
without a suitable complete analysis and functional check following the modification which ensures
further safe usage.

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 11 of 76
®
Maintenance manual medinCNO

4 Installing the medinCNO®


®
The medinCNO must be securely mounted and the following conditions must be met before the
device is started up:
4.1 Mounting

®
Figure 4-1: a) A bracket holding the medinCNO in position (the second cannot be seen); b) Ventilation slits; c)
Overpressure valve; d) Knurled screw
®
The medinCNO must always be securely mounted in an upright position before the device is started
up. The two mounting brackets on the back of the device (Figure 4-1 a) are inserted into a standard
track (cross-section: 10 x 25 mm) and the brackets are then fixed in place manually using the knurled
screws (Figure 4-1– d). It is important to ensure that the rear and lateral ventilation slits (Figure 4-1– b)
are not covered and that nothing is blocking the overpressure valve (Figure 4-1– c) on the base of the
device.
®
The medinCNO can be transported when it is not in use. When it is in use, it must be securely
®
mounted, however. However, if the medinCNO becomes damaged during transport or due to rough
handling, its function must be checked by a service technician. It is not allowed to be used until then.
Warning:
®
- The medinCNO must always be securely mounted in an upright position throughout the
entire time that it is in use.
®
- The medinCNO must not be covered while in use, and none of the openings or ventilation
slits must be blocked.
- The overpressure valve must be kept clear whenever the device is in use; make sure that
nothing touches the overpressure valve.
®
- A damaged medinCNO may not be used.

4.2 Power supply


4.2.1 Mains operation
®
The power supply unit (REF 39-113) of the medinCNO may only be connected to a power supply with
the following properties:
 Voltage: 100 – 240V~ (alternating current)

 Frequency: 50 – 60Hz

 Strength of current: at least 1.1A

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 12 of 76
®
Maintenance manual medinCNO

®
The power supply unit of the medinCNO has a C8 connector for connection to the mains grid. The
®
power line between the power supply unit of the medinCNO and the mains grid must therefore meet
these conditions:
 Power supply side: plug appropriate for the country

 Device side: C7 plug in accordance with EN60320

 Minimum rated voltage: the minimum rated voltage of the power line must be at least as high
as the voltage of the local mains supply. (Recommended: 240V)

 Minimum rated current: 1.1A

 Maximum length 2 metres (in order to achieve EMC in accordance with DIN EN 60601-1-
2:2007)

Spare mains cables with EU plugs can be ordered under REF 39-115 (supplied as standard as part of
the power supply unit). Alternative mains cables for other countries are available under these
numbers:
 with UK plug REF 39-116

 with plug for USA/Japan REF 39-117

 with Australian plug REF 39-118

Attention:
®
The medinCNO may only be run on the local mains grid using the power supply unit provided by the
manufacturer (REF 39-113) and suitable mains cables, otherwise the functionality and EMC of the
device cannot be guaranteed.
4.2.2 Battery operation
®
The medinCNO contains an internal battery and can operate on a fully charged battery for up to two
hours.

Figure 4-2: Battery charge status displayed in the top left-hand area of the screen

®
The charge status of the internal battery is displayed in the top left-hand area of the medinCNO
screen (Figure 4-2).
A full battery is indicated by a symbol with five lines. If 2 or fewer are left, the symbol is displayed in
®
red and the medinCNO should be connected to the mains soon. If only one line is left, the battery
alarm is triggered.
Attention:
®
When the medinCNO is running on battery power, an alarm sounds when the internal battery is low.
If this happens it is essential that the device be connected to an external power source without delay.

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 13 of 76
®
Maintenance manual medinCNO

4.3 Gas supply


®
In order to operate the medinCNO , it must be supplied with air and oxygen from an external source
that meets these conditions:
Medical oxygen:
- 300 to 600 kPa (= 3.0 to 6.0 bar)

- Oxygen content: ≥ 99.5%

- H2O ≤ 67 ppm (V/V) (free from condensate)

Medical air (Aer medicalis):


- 300 to 600 kPa (= 3.0 to 6.0 bar)
3
- Oil ≤ 0.1 mg/m , ideally oil-free

- H2O ≤ 870 ppm (V/V) and free from condensate. If, due to the operating temperature, it is
possible for condensate to form despite a relatively low concentration of H2O in the air, then
the concentration of H2O must be reduced accordingly.

®
Figure 4-3: Water trap in the medinCNO air inlet; a) Bolt for removing the condensate from the water trap

®
The medinCNO can be fitted with a water trap in the air inlet as an option (Figure 4-3) (REF 39-800).
This allows small amounts of condensate to be removed from the air supply provided to the
®
medinCNO if the water trap is checked for condensate and emptied on a regular basis. To empty the
water trap, undo the black bolt (Figure 4-3 – a) on the underside of the water trap and keep it open
until the condensate has run off (this can also be done under pressure).
Attention:
®
The functionality of the medinCNO can only be guaranteed if both the oxygen and medicinal air
supplied to it are free from condensate water, oil and contaminants. If this is not guaranteed, the
®
medinCNO could be damaged. Even the use of a water trap can only remove small amounts of
condensate from the air.
In order to avoid the risk of fire, oxygen must not come into contact with oil or grease. It is therefore
®
essential to ensure that all parts of the medinCNO and the CPAP system which come into contact
®
with oxygen (e.g. gas feed tubes, O2 inlet to the medinCNO , patient outlet, patient tubes and
®
Medijet ) remain free from oil and grease.

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 14 of 76
®
Maintenance manual medinCNO

Gas supply connections:


®
The medinCNO can be fitted standard with NIST or DISS gas supply connections as the customer
prefers. The tube kits to connect the device to the external gas source must also meet the selected
standard.
4.4 External alarm interface and data interface
Attention:
Only devices and systems which are approved for medical use and which meet the relevant electrical
®
safety and EMC standards may be connected to the medinCNO data interface and alarm interface.
4.4.1 External alarm
®
The medinCNO has an interface for connection to an external alarm system. The interface can open
or close an external current circuit in the event of an alarm. The interface thus acts only as a switch,
that is, the power supply must be ensured by the external alarm system.

Time delay:
The interface to the external alarm system is updated every 100msec. This means that the alarm at
the external system is with a maximum delay of 100 msec in comparison to the alarm in the device.

The alarm is likewise deactivated with a maximum delay of 100 msec in comparison to the alarm in the
device. The silencing of the alarm sound does not count here, but rather the actual deactivation of the
alarm (that is, correction of the alarm condition and acknowledgement of the alarm by the user).
®
The cable supplied by Medin must be used to connect the medinCNO to an external alarm system in
order to achieve EMC according to DIN EN 60601-1-2:2007 (REF 39-114) and the connection must be
set up initially by an appropriately trained technician.
Attention:
®
The medinCNO may only be connected to an external alarm system using the connection cable
provided by the manufacturer (REF 39-114), otherwise the functionality and EMC of the device cannot
be guaranteed.
4.4.2 USB data interface
The USB data interface can be used for brief access to the device’s internal data or for software
updates.
During normal operation it delivers automatically live data. Additionally there is the possibility to export
the stored trend data by pressing the “Export via USB” button in the trend view. The trend data also
includes an alarm list and a start list. Both data types will be send through the USB port and have to
be received and stored with a computer and a serial terminal software.
Attention:
The export of stored trend data interrupts the normal live data for the time necessary to export all trend
data. At the beginning of the export of trend data the export of live data is stopped and afterwards
started again, but the life data from the time interval in between will be missing.

Live Data:
®
The live data of the medinCNO contain the patient pressure, the set flow, the measured oxygen
concentration, whether a push is currently being emitted, all alarms, the set additional push flow and
the current mode. In addition, date, time, device name and software version are regularly indicated.
The live data are automatically and constantly indicated in normal operation. However, they are not
®
saved in the medinCNO . If the live data are of interest, they must be recorded and saved by a
computer or data system or the data will be lost.
The exact structure of the live data is described beyond.

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 15 of 76
®
Maintenance manual medinCNO

Trend Data:

The trend data are divided into two parts: The first part contains logfiles of saved alarms and device
starts. This recording contains the data of the last device uses and will continue to be saved even if
®
the medinCNO is switched off. The second part of the trend data is deleted each time the
®
medinCNO is switched off. It saves the data since the last device start for a maximum of 28 days. It
contains the CPAP pressure, the set flow, the measured oxygen concentration, the set additional push
flow, the mode, respiratory rate, whether the backup was active, the pushrate Rinsp, the minimum and
maximum possible push rate, the maximum pressure during the pushes, the oscillation frequency, the
set level of the oscillation amplitude, the resultant pressure of the oscillation amplitude, the set
inspiration time and the set apnoea time.
The structure of the trend data you can see beyond.

4.4.2.1 Data type:


Both live and trend data are serial data, which are transformed internal to a USB signal. But at the
computer the data handling is similar to standard serial signals (like RS232). To receive the data these
setting have to be set:
- Data Type: serial

- Baud rate: 115200

- Data bits: 8

- Parity: none

- Stop bits: 1

- Handshaking: none
®
To connect the medinCNO to a system for archiving and processing patient data, a standard USB
cable (USB 2.0 type A to USB 2.0 type B) with double winding through a ferrite of the type: 742-711-32
from the manufacturer Würth Elektronik can be used (see Fehler! Verweisquelle konnte nicht
®
gefunden werden.). Nonetheless, the function of the medinCNO must be observed to avoid
disruptions by the entire system with regard to electromagnetic influences.
Attention:
®
The data interface of the medinCNO may only be connected with the cable specified above, including
ferrite core of the type: 742-711-32 from the manufacturer Würth Elektronik with double winding and a
computer system that meets the medically relevant requirements in terms of electrical safety and
®
EMC, otherwise the functionality and EMC of the medinCNO cannot be guaranteed.
In doing so, only the following ferrite core may be used: Würth Elektronik - Type: 742-711-32.
The cable used in each case (USB cable or optional RS232 cable) must thus contain a ferrite (REF
39-125; Würth Elektronik - Type: 742-711-32, gray or black) with a double wound cable in accordance
with figure 4-4.

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 16 of 76
®
Maintenance manual medinCNO

a) b) c) d)

Figure 4-4: USB and RS232 cabel with ferrite core


a) RS232 cable without ferrite b) RS232 cable with a ferrite that is still open whereby the cable must be carried through
the core twice, that is, wrap the ferrite core once. c) RS232 cable with ferrite d) USB cable with ferrite (the cable
must also be carried through the ferrite core twice as in the case of RS232 cable)

4.4.2.2 Structure of the live data:


Example:
***;2014-04-04;14:09:11;medin-cno1004;V1.3;
04.1;06.0;039;0;0000;0008;0000;03.7;O
04.1;06.0;039;0;0000;0008;0000;03.7;O

04.2;06.0;039;0;0000;0008;0000;03.7;O
04.2;06.0;039;0;0000;0008;0000;03.7;O

07.0;08.7;021;0;0000;0000;0000;03.8;C
07.9;08.7;021;0;0000;0000;0000;03.8;C
07.9;08.7;021;0;0000;0000;0000;03.8;C
07.7;08.7;021;0;0000;0000;0000;03.8;C
07.5;08.7;021;0;0000;0000;0000;03.8;C
06.7;08.7;021;0;0000;0000;0000;03.8;C

Every 10 minutes it provides these data:

- ***;

- Date: YYYY-MM-DD; (adjustable in service menu)

- Time: HH:MM:SS; (adjustable in service menu)

- Device name (adjustable in the service menu)

- VXXX (CPU software version)

In the interim, the following data are provided every 0.1 sec (separated by ; ) via the USB
interface:

- The currently measured CPAP pressure (in the CPAP, APNOEA CPAP and SNIPPV mode) or
the CPAP average value (in oscillation mode) in XX.X mbar

- Set target flow in XX.X L/min for the basic CPAP level; in the case of additional activation of the
leak assist, the target flow + correction flow is emitted.

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 17 of 76
®
Maintenance manual medinCNO

- Measured oxygen concentration FiO2 XXX%;

- Current push activity X: 2 = Backup actively running (independent of whether there is currently
a push or not), 1 = push active (without backup), 0 = no push active;

- Alarms of high priority – shown by an active bit on the corresponding position:

o 0x0001: Disconnection

o 0x0002: high CPAP fixed

o 0x0004: high feed pressure

o 0x0008: gas feed O2

o 0x0010: gas feed air

o 0x0020: device error patient pressure measurement

o 0x0040: flow stopped – reactivation via alarm reset

o 0x0100: device error; restart necessary

- Alarms of medium priority:

o 0x0001: battery almost empty – connect the power supply

o 0x0002: CPAP pressure high (adjustable)

o 0x0004: CPAP pressure low (adjustable)

o 0x0008: oxygen concentration high

o 0x0010: oxygen concentration low

o 0x0020: internal temperature too high

- Alarms of low priority:

o 0x0001: fan blocked

o 0x0004: Leak-Assist

- Adjusted additional flow during push XX.X L/min

- Mode X:

o C = CPAP

o A = Apnoea CPAP

o S = SNIPPV

o O = Oscillation

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 18 of 76
®
Maintenance manual medinCNO

4.4.2.3 Structure of the trend data:


Example:
###;2014-04-07;14:20:14;medin-cno1004;V1.3;alarm record;
start;2014-04-04;10:59;
alarm;04;11:07;0000;0008;0000;for 14:48
alarm;04;13:50;0000;0004;0000;for 00:30
alarm;04;14:14;0002;0002;0000;for 00:36
###;2014-04-07;14:20:15;medin-cno1004;V1.3;trend record;
CPAP;Flow;FiO2;Finsp;Mode;RR;BU;Rinsp;min;max;Pinsp;Freq;AmpS;AmpP;Tinsp;Tapn;
05.6;07.0;021;00.0;C;018; ; ; ; ; ; ; ; ; ; ;
05.6;07.0;021;00.0;C;018; ; ; ; ; ; ; ; ; ; ;

5.5;07.0;024;01.0;A;024;-;000; ; ;00.0; ; ; ;0.3;10.0;
5.9;07.5;024;01.0;A;018;-;000; ; ;00.0; ; ; ;0.3;10.0;

05.8;08.0;021;02.6;S; ;B;070;25;120;10.6; ; ; ;0.2;02.2;
07.3;08.0;021;02.6;S; ;B;070;25;120;11.2; ; ; ;0.2;02.2;

02.5;06.0;026;02.5;O; ; ; ; ; ; ;07;10;05.4; ; ;
04.1;06.0;036;03.7;O; ; ; ; ; ; ;14;06;04.1; ; ;
04.1;06.0;039;03.7;O; ; ; ; ; ; ;19;06;04.9; ; ;
***;2014-04-07;14:22:51;medin-cno1004;record complete;

Start line of alarm record in trend data:

- ###;

- Date: YYYY-MM-DD; (current date, set in service menu)

- Time: HH:MM:SS; (current time, set in service menu,


the switch to local time or summer time must be
performed manually)

- Device name: medin-cnoZZZZZ(set in service menu)

- SW version: VXXX (installed main software version)

- Alarm record;

Start and alarm records:

- For every device start it displays date and time like this:

start; YYYY-MM-DD;HH:MM;

- For every new alarm it displays day, time, all active high, middle and low alarms and the
duration of this alarm:

alarm; DD; HH:MM;HHHH;MMMM;LLLL;for MM:SS

The alarm coding of all alarms is analogue to the coding in the live data (for detailed information
see above). The date is only shown as day, the month has to be calculated using the last start
entry before this alarm entry. The alarm duration displays the minutes and seconds from the
beginning of the alarm until the alarm was finished.

Records of adjusted and measured values:

The trend data contain all values saved in the trend and additional values since the last time the
device was switched on. Every 20 seconds, a line is saved. If a value is not available in the current
mode, it is replaced with blank spaces.

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 19 of 76
®
Maintenance manual medinCNO

The general structure is like this:

- Average value of the CPAP pressure in XX.X mbar

- Set flow in XX.X L/min + if applicable, correction flow with activated leak assist

- Measured FiO2 in XXX%

- Set, additional push flow in XX.X L/min

- Current mode: C = CPAP, A = APNOEA CPAP, S = SNIPPV, O = Oscillation, - = Standby

- Measured respiration rate in XXX / min

- Status of backup (B = actively running; -= currently not actively running or generally off )

- Push frequency Rinsp in XXX / min

- Minimum reachable push frequency Rinsp in XX / min caused by apnoea

- Maximum reachable push frequency Rinsp in XX / min caused by triggering synchronised with
respiration

- Maximum CPAP value during the push in XX.X mbar

- Set oscillation frequency in XX Hz

- Set amplitude level of oscillation in level XX

- Measured amplitude during the oscillation in XX.X mbar

- Set push length Tinsp in X.X sec

- Set apnoea time in XX.X sec

- Adjusted apnoea time in XX.X sec

4.4.3 RS232 data interface


This interface is optional.

The function, data and settings are identical equal with the USB interface (see chapter 4.4.2).

Attention:
®
The medinCNO data interface must be connected using the cable, including ferrite core (REF 39-
125) specified above (see Fehler! Verweisquelle konnte nicht gefunden werden.) and a computer
system that meets the medically relevant requirements in terms of electrical safety and EMC,
®
otherwise the functionality and EMC of the medinCNO cannot be guaranteed.

4.5 Environmental conditions (usage, transport and storage)


®
The medinCNO must not be used in rooms where there is a risk of explosion or near flammable
substances or locations exposed to splash water.
Environmental conditions during operation:
Temperature: 15°C to 35°C
Relative humidity: Air humidity 20% to 80% (not producing condensate)

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 20 of 76
®
Maintenance manual medinCNO

Moisture: The device must be kept dry during use.


Ambient pressure: 700hPa to 1100hPa
Altitude: ≤2000 m height above sea level
Cleanness: All hygiene regulations valid for the hospital have to be followed. Furthermore,
®
the medinCNO must be used under standard clinical room conditions.
Environmental conditions during transport or short-term storage:
Temperature: -20℃ to 50℃ (long-term storage may only take place at normal room
temperature)
Relative humidity 20% to 80% (not producing condensate)
Moisture: The device must be kept dry during transport and storage.
®
Cleanness: The medinCNO must be protected from contamination during transport and
storage.
Attention:
®
If the medinCNO is exposed to temperatures of ≤ 15°C during transport, it must be acclimatised at
room temperature for at least 24 hours at the location where it is to be used before being switched on
and used for the first time. If this is not done, the condensate which has formed could damage the
®
medinCNO .
®
The medinCNO may only be transported in the original packaging, since it contains internal lithium
®
batteries which may only be transported under special conditions. If the medinCNO contains
defective batteries, these may not be transported and must be removed in advance and disposed of in
®
accordance with local regulations. (See separate instructions on transporting the medinCNO and the
batteries it contains).

Long-term storage:
Temperature: Room temperature (about 20°C)
Relative humidity: 20% to 80% (not condensing)
Moisture: The device must be kept dry during storage.
®
Cleanness: The medinCNO must be protected from contamination during storage.
Advice:
Long-term storage has to be carried out under room temperature conditions. Otherwise the internal
batteries may become damaged and must be exchanged earlier.

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 21 of 76
®
Maintenance manual medinCNO

5 Service Menu
5.1 Start Service Menu
® ®
To start the service menu of the medinCNO , switch on the medinCNO and perform the start test
until you reach the overpressure valve check. Instead of pressing “ok” or “not ok”, you have to press
the two buttons shown in Picture 5-1 at the same time when the service menu opens.

Picture 5-1: Necessary keys to open the service menu.

If your pneumatic unit has an internal problem or if no gas supply is connected, you have to start the
®
medinCNO and wait until you get one of the following failure messages: “Gas feed pressure low,
check air”; “Gas feed pressure low, check O2” or “Device Error – device shutdown necessary” (or the
equivalent texts in your language). Then you can also open the service menu by pressing the two keys
shown in Picture 5-1.
5.2 Service Menu overview
The service menu includes several points. All of them are described in the following chapters. To
®
choose a point, use the knob of the medinCNO and scroll up and down within the service menu list.
Press the knob to choose one of the points highlighted in blue.

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 22 of 76
®
Maintenance manual medinCNO

Picture 5-2: Service menu of medinCNO®

5.3 Device name


® ®
The item “device name” in medinCNO ’s service menu allows your particular medinCNO to be
identified by an individual number.
®
If the data export of more than one medinCNO is used via the USB connector (or optional RS232
connector), this device name is exported together with the actual values and differentiates the data.
5.4 Date and time
®
medinCNO saves the current date and time which is used to save internal values such as the last
maintenance date.
If the displayed time and date is not correct, you can change it by selecting the line and pressing the
knob. Afterwards you can change any part of the date and time by turning the knob. Pressing the knob
activates the next part: year – month – day – hour – minute - second.
5.5 Language
By choosing language and pressing the knob, you can scroll through all possible languages by turning
the knob. Choose your desired language and press the knob to change the language which the
®
medinCNO displays.
This language affects all the areas which are used during normal usage. The service menu and all
sections inside the service menu are always in English.
5.6 Pressure unit
®
medinCNO is able to display pressures in mbar or in cmH2O. The displayed values are identical in
both versions, but the unit displayed next to it will be changed to the version which your users prefer.
5.7 External Alarm
®
The medinCNO includes a connector to connect it to an external alarm system.
This interface can be activated or deactivated here.
The interface has no internal power source. It can only switch your external system. It is designed as
an electric switch, which offers two possible connections:
- Version 1: The switch will close during an alarm and open again afterwards. In this situation,
there is a conductive connection during an alarm and a nonconductive connection if no alarm
is active.

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 23 of 76
®
Maintenance manual medinCNO

- Version 2: The switch will open during an alarm and close again afterwards. In this situation
there is a nonconductive connection during an alarm and a conductive connection if no alarm
is active.
Attention:
If you choose “off” at this point in the service menu, the switch is deactivated the entire time. To use
the external alarm system interface, activate this connection here (it has to display “on”).
5.8 FiO2 Alarm
®
Through this point you adjust medinCNO ’s behaviour during changes of the adjusted FiO2:
- Automatic: In automatic mode the alarm boarders of the FiO2 alarm are automatically updated
if the FiO2 is changed. E.g. you have FiO2 of 27% and the alarm boarders 22% / 32%. If you
now change the FiO2 to 40% the alarm boarders will be automatically changed to 35% / 45%,
without any further information.
- Manual: In manual mode changes in FiO2 will have no influence to the FiO2 alarm boarders.
E.g. you have FiO2 of 27% and the alarm boarders 22% / 32%. If you now change the FiO2 to
40% the alarm boarders will stay at 22% / 32%. To update the alarm boarders to the new FiO2
value the user has to open the alarm setting menu, choose FiO2 alarm refresh and confirm it.
Then the FiO2 alarm boarders will be changed to 35% / 45%.

5.9 Oxygen calibration


If you choose this item in the service menu and press the knob, oxygen calibration starts.
Attention:
®
To get correct results, make sure that the medinCNO is connected correctly to 100% oxygen and
normal air during this time.
You can use this calibration e.g. if you have exchanged the oxygen cell and before you start the
®
device test. Alternatively, you can start the medinCNO and perform the system start until you have
reached the oxygen calibration step.
5.10 Device check
The item “device check” displays the date of the last device check which was performed for the
®
medinCNO and the result of this test.
The item “device check” is also used to start a new device check. Therefore you have to highlight this
item in blue and then press the knob. See chapter 6.
5.11 Last maintenance
Next to the item “last maintenance”, the date of the last maintenance which was performed on this
®
medinCNO is displayed.
This item is also used to start the confirmation after a new maintenance. Therefore you have to
highlight this point in blue and then press the knob. See chapter 7.

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 24 of 76
®
Maintenance manual medinCNO

5.12 Self check pneumatic unit


The item “self check pneumatic unit” displays the result of the last pneumatic self test, which is
normally performed during the system start before entering the service menu.
If the value displays a “1”, something was wrong with your pneumatic unit. The position of the “1”
indicates the failure. In this case, contact medin for further information.
X X X X X X X X X X X X

Inlet pressure O2 not


is

sensor

Inlet pressure air not


pressure device outlet

within 2.8 and 6.0 bar

within 2.8 and 6.0 bar


Failure ∆P sensor O2

Failure ∆P sensor air


Leakage OUT valve
Leakage OSZ valve
Medijet

Leakage O2 valve

Leakage air valve


Failure flow O2

Failure flow air


Failure P out
to
(Connection
necessary)

Failure
Attention:
® ®
To test the P out sensor it is necessary to connect a tube set and a Medijet to the medinCNO .
Otherwise the test can be incorrect. Also variations in the pressure of your gas supply can influence
the result of the pneumatic self check – in this case, restart the test. But if there is a permanent failure
in your pneumatic unit, the pneumatic unit has to be sent to medin. It is not possible to repair a failure
in the pneumatic unit in the field.
5.13 Internal Data
The internal data field opens the internal data overview.
Using the internal data menu, you can change the flow value and the FiO2 value by using the knob.
®
medinCNO will use these parameters for manual debugging purposes.

Picture 5-3: internal data overview of medinCNO®

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 25 of 76
®
Maintenance manual medinCNO

Inside the internal data, these values are displayed, among others:
- Pressure Air and Pressure O2: Here you can see the pressure of the connected gas supply
®
system. If this value is <2.8 bar or >6.0 bar, medinCNO will not work correctly. If the values
®
are borderline, medinCNO may also trigger regular alarms, because small variations in the
pressure of a gas supply are normal.
- Pat. Pressure 1 and Pat. pressure 2: These are the values of the two internal CPAP
®
pressure sensors. If these sensors differ by 2 mbar or more, medinCNO will trigger an alarm
and cannot be used further.
- Bat. charge status: indicates the actual charging status of the internal rechargeable battery.
- Max. capacity of bat.: displays the maximum capacity which your internal battery reached
during the last charging cycle. If this value is ≥4300 mAh, you have a new rechargeable
battery. If this value is ≤3500mAh, your battery is old and has less capacity left. In this case,
®
the internal rechargeable battery has to be exchanged, because usage of the medinCNO in
battery-powered mode can no longer be guaranteed.
- Number of bat. cycles: the number displayed here counts how often the rechargeable
®
battery inside the medinCNO was discharged and charged. The battery inside the
®
medinCNO has to be exchanged if this value is 300 or higher.
- Voltage 3 V Bat. shows the actual current of the internal 3-volt battery. If this value is 2V or
less, the 3-volt battery has to be exchanged.
®
- Operating hours shows the number of hours during which this medinCNO was used.
- Last pneumatic status see chapter 5.12
- SW-Version: indicates the installed software version.
- Displaymakro version: indicates the installed display and language version.
- Last maintenance, last oxygen calibration and last device check: indicate the dates when
these points were last performed.

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 26 of 76
®
Maintenance manual medinCNO

6 Device check
If you choose the item “device check” in the service menu and press the knob, you will enter the
®
device check test. This test checks all parts and sensors in medinCNO . Through this test you can
®
check if the medinCNO is working correctly.
Attention:
®
Once a year this test has to be performed to verify that the medinCNO is still fit for service. Otherwise
®
the medinCNO functionality cannot be guaranteed and the device cannot be used for patients.
The device check has to be performed AFTER the oxygen cell has been replaced and AFTER the
oxygen calibration has been renewed.
®
The device check consists of the following tests. Its results have to be documented on the medinCNO
device passport (see chapter 15):
Test: Description: Passed if:
LED test In this test, the alarm LED and knob LED OK = if the colours red, green, and
are tested by visual examination. blue of the alarm LED and the
white light under the knob are
visible.
®
Alarm buzzer Here the alarm tone of medinCNO is OK = if the sound is audible
®
test 1 tested and the medinCNO will beep.
Alarm buzzer This test checks the alarm monitor. If OK = if a continuous tone is
test 2 everything works correctly, the test triggers triggered and audible
a 2-min continuous tone.
Advice:
After the tone is triggered, it can be stopped
before the end of the 2 minutes by pressing the
on/off button for a second. It is not necessary to
wait the entire two minutes for the test.
Key test Here all keys are tested by pressing them OK = if by pressing the individual
one after the other. keys only “ok” and no “NOT – Ok”
Advice: is generated.
The key on/off is not tested, because its
functionality is necessary to reach this test.
After a short time, if the keys are not pressed, a
not-ok is automatically generated.
Display test This test checks whether all areas of the OK = if the complete display is
display are working. black.
Knob test All three functions of the knob are tested by OK = if turning in both directions
turning it in both directions and pressing. and pressing generate an “ok”.
Power supply Here the power supply and the internal OK = if the connection and the
test connection of the power supply inside the disconnection of the power supply
®
medinCNO is tested. is identified correctly and both
Advice: generate an “ok”.
®
If the medinCNO is not accompanied by its OK = if the green light inside the
power supply, the test has to be performed power supply is glowing.
using another power supply, the field “at
customer” has to be marked, and the field LOT
stays empty.

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 27 of 76
®
Maintenance manual medinCNO

Test: Description: Passed if:


USB connection This test checks the connection of the USB OK = if the computer receives
test interface and the data output. You must “Test”.
therefore connect the USB port of the
®
medinCNO with a computer and record
the data which is communicated through
the serial port.
Advice:
Use a terminal program with these adjustments:
Baud rate – 115200
Data bits – 8
Parity – none
Stop bits – 1
Handshaking – none
RS232 This test checks the connection of the OK = if the computer receives
connection RS232 interface and the data output. You “Test”.
(optional) must therefore connect the RS232 port of
®
the medinCNO with a computer and
record the data which is communicated
through the serial port.
Advice:
Use a terminal program with these adjustments:
Baud rate – 115200
Data bits – 8
Parity – none
Stop bits – 1
Handshaking – none
External alarm During this test the interface for the
test external alarm is tested. This involves OK = if the resistance between the
testing whether the internal switch changes pins fulfils the following:
between conductive and non-conductive.
Advice:
during without
This test will work whether or not the external
Pin
alarm alarm
alarm in the service menu is activated.
non
1-2 conductive
conductive
non-
3-2 conductive
conductive

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 28 of 76
®
Maintenance manual medinCNO

Test: Description: Passed if:


Pressure sensor During this test, the three internal pressure OK = if the maximum deviation of
test sensors are compared to each other. the three pressure values is
≤ 2 mbar.
Therefore the patient pressure port and the
®
flow outlet port of medinCNO have to be
connected with a syringe like this:

Additionally, you have to block the flow


bypass for the oxygen measurement by
blocking its outlet (next to the air gas
supply inlet) like this:

Attention:
Use a soft tool (e.g. nail with a piece of
silicone tube on the end)
In this configuration you have to press the
syringe down gently and create a pressure
of 15 to 20 mbar on all three pressure
sensors.
Gas supply During this step the internal measurement OK = if the internal measured
pressure test for the gas supply pressure is compared values of air and O2 pressure
oxygen and air with an external measurement. deviate from the external
Advice: measured values ≤ 0.5 bar
®
At this point, the medinCNO must be
connected to air and oxygen. Otherwise the
device check will not be possible.

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 29 of 76
®
Maintenance manual medinCNO

Test: Description: Passed if:


®
Overpressure During this test, the medinCNO generates OK = if the overpressure valve
valve test flow at the flow outlet. In addition, the tester opens and the pressure at the
has to block the flow outlet like this: outlet port is ≤ 45 mbar while
open.

Now the pressure at the flow outlet is


checked.
Pneumatic valve During this test, the pneumatic unit OK = if the result of the valve test
test performs a self-test which checks all the is “ok”.
valves.
If the result of this test is not ok, the
®
medinCNO will display the failure values
which are described in chapter 5.12.
Attention:
For this test it is necessary, that
®
medinCNO is connected to a tube set and
®
a Medijet , otherwise the test cannot be
done correct.
Flow and During this test, the accuracy of the OK = if the external measured
concentration electronic blender is tested. Therefore the values are within these tolerances:
®
test medinCNO emits several flows and
concentrations.
Flow 4 L/min ± 1 L/min
Attention
Flow 15
These flows and concentrations have to be ± 2 L/min
L/min
measured by external, calibrated
measurement equipment at the flow outlet FiO2 ± 3 Vol%
®
of the medinCNO .
Advice:
Flow values have to be measured according to
DIN1343 at:
- 0°C
- 1013.25 mbar ambient pressure and
- 0% humidity
Oxygen sensor In this step, the oxygen measurement OK = if the external and internal
®
test: sensor inside the medinCNO is tested. measured values are within this
®
The medinCNO emits several oxygen tolerance limit:
concentrations which have to be checked
by an external, calibrated measurement
device. The value of the external and the FiO2 ± 2 Vol%
internal measurements have to be within
the tolerance range.

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 30 of 76
®
Maintenance manual medinCNO

Test: Description: Passed if:


Rechargeable In this test, the internal rechargeable OK = if the measured values are
battery test: battery is tested. within these tolerances:
Therefore the power supply must be
disconnected; otherwise the displayed
12V ≤ voltage ≤ 17V
voltage of the battery will not be correct.
number of charging cycles
The displayed values must be checked
<300
according to the given tolerances. If these
are not met, the internal battery has to be maximum charging capacity
exchanged. >3500 mAh
Attention:
The internal rechargeable battery has to be
replaced after 3 years, even if it is still ok.
Advice:
If the number of charging cycles increases by
®
300 whenever the medinCNO is started, a
warning is displayed. Thus if the value is near
300, an exchange of the internal battery is
recommended.
3V battery test: In this step of the device check the internal OK = if this tolerance is met:
3V battery is checked. Here the display
voltage is compared to the tolerances.
2V < value < 3.5V
Attention:
The 3V battery has to be replaced after 3
years, even if it is still ok.
Advice:
If the displayed voltage is > 3.5V, the
®
medinCNO does not contain any 3V battery.
To correct this, insert a battery.
® ®
medinCNO During the last step of the medinCNO OK = If all tests are passed, no
®
system start test device test, the normal medinCNO system warnings are displayed and the
®
start test is performed. Switch the medinCNO starts the CPAP
®
medinCNO on and follow all the displayed therapy.
®
steps until the medinCNO starts normal
CPAP therapy.

®
If all of these tests are passed, the medinCNO displays “everything ok” and saves the current date
and the result. This information can be seen in the service menu and in the internal data.
Attention:
®
If the device check was not passed, the medinCNO cannot be used for a patient.

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 31 of 76
®
Maintenance manual medinCNO

7 Last Maintenance
The item “last maintenance” is started within the service menu by selecting it and pressing the knob.
®
This item checks whether the maintenance of the medinCNO was performed completely and
correctly. If yes, the saved maintenance date will be refreshed.
In the last maintenance menu you have to confirm:
- that the oxygen cell has been replaced. If the date of the last oxygen calibration is too remote
(≥ 7 days) it will not be accepted.
- that the device test has been performed. If the date of the last device test is too remote (≥ 7
days) or if the result was “failed”, it will not be accepted.
®
If the date of the last maintenance is ≥11 months old, medinCNO will display a warning during the
start up which informs the user about the date of the last maintenance and that new maintenance is
(soon) necessary.

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 32 of 76
®
Maintenance manual medinCNO

8 Software update of medinCNO®


8.1 Update of main CPU software
Attention: Software updates are only available for regular Software versions x.y. Software Updates are
not supported in the version without Oscillation xCy.
8.1.1 Necessary material
®
To update the medinCNO CPU software, you need:
- Computer (at least Windows XP or newer)
- USB cable (type A to type B)
®
- Software package medinCNO CPU Update
- A correct installed driver for the USB Port

Attention: The software update only works with Windows XP or newer.

8.1.2 Set up
®
To update the software the USB port of the medinCNO has to be connected to the computer with the
USB cable.

8.1.3 Execution
1. Start the CPU Boot loader:

Press the 3 keys on the right beneath the display and additionally the on/off button until the
text ”Bootloader Vx.xx – CPU -“ appears in the lower left-hand side of the display.
®
2. Double-click on file “Update medinCNO main CPU.cmd” and wait while the window appears
slowly.

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 33 of 76
®
Maintenance manual medinCNO

3.

If the update has finished successfully, the following text appears at the bottom of the screen
identified by the red arrow:
Comparing Flash Data…
Equal!
Leaving programming mode…
Otherwise you have to repeat the process (beginning from point 2).
4. If the software update has finished successfully, you can switch off the boot loader by pressing
the hardware shut-off (second key from the left under the display and on/off). Afterwards you
®
can restart the medinCNO .

8.2 Display software update


®
There are four different software packages existing for medinCNO ’s display. Each package includes
different languages. The right package has to be chosen, depending on the language of the user and
the CPU Software of the device.

Attention: Package ZH is only to be used with CPU Software xCy without Oscillation!
Package A: Package B: Package C: Package ZH:
- English - English - English - English
- German - Czech - Chinese - Chinese
- French - Polish - Turkish
- Italian - Norwegian - Japanese
- Netherlands - Danish - Romanian
- Spanish - Swedish
- Greek - Russian
- Lithuanian

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 34 of 76
®
Maintenance manual medinCNO

8.2.1 Necessary material


®
To update the medinCNO display software, you need:
- Computer (at least Windows XP or newer)
- USB cable (type A to type B)
®
- Software medinCNO _display_package_VXXX-Y in the desired language set stored in a
location with write access
- A correct installed driver for the USB Port
8.2.2 Set up
®
To update the display software, the USB port of the medinCNO has to be connected to the computer
via the USB cable.
8.2.3 Execution
1. Start the LCD boot loader:

To do this, press the 3 keys on the upper right-hand side of the display and additionally the
on/off button until the text “Bootloader Vx.xx – LCD ” is displayed in the lower left-hand side of
the display.
®
2. Afterwards start the program “update medinCNO display.cmd” inside the software package.
3. Now a window like this should appear:

First, if necessary, the Firmware will be updated as see on the entry

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 35 of 76
®
Maintenance manual medinCNO

“Programming firmware …: …kB <..%>”.


If it is finished it writes – Ok – Rebooting – Ok

Afterwards you have to start the data update, by pressing any key.

The update you can see on the entry


“Programming firmware …: …kB <..%>”.
If it is successfully finished it writes – Ok – Verifying – Ok

4. The software update has been finished now and the bootloader can be switched off by
pressing the hardware shut-off (the second key from the left under the display together with
®
the on/off button). Now the medinCNO can be restarted.

8.3 Update of the pneumatic software


8.3.1 Necessary material
®
To update the medinCNO pneumatic software, you need:
- Computer (at least Windows XP or newer)
- USB cable (type A to type B)
®
- Software medinCNO pneumatic update (stored in a location with write access)
- A correct installed driver for the USB Port
8.3.2 Set up
®
To update the pneumatic software, the USB port of the medinCNO has to be connected to the
computer via the USB cable but it must not be switched on.
8.3.3 Execution
®
1. Double-click on the file “update medinCNO pneumatic.cmd“
2.

When prompted you have to start the boot loader (see red arrows).

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 36 of 76
®
Maintenance manual medinCNO

To do this, press the lower 3 keys on the right-hand side of the display and additionally the
on/off button until the text „Bootloader Vx.xx-AIR-” is displayed in the lower left-hand
side of the display.
Wait during the update is sent.
3. Restart the boot loader if you see the following demand:

®
Switch off the medinCNO by pressing the hardware shut-off (second key from the left under
®
the display and on/off). Afterwards you can restart the medinCNO by pressing the lower three
keys on the right side of the display and the on/off button.

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 37 of 76
®
Maintenance manual medinCNO

4. The following window appears:

If the software update has finished successfully you see the following text in the middle of the
window identified by the red arrow
Application started
Medin3090 V: x.xx.x
Otherwise you have to repeat the process (beginning from point 1).
5. If the software update has finished successfully, you can switch off the boot loader by pressing
the hardware shut-off (second key from the left under the display and on/off). Afterwards you
®
can restart the medinCNO .

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 38 of 76
®
Maintenance manual medinCNO

9 Alarms and error messages


9.1 Failures during the medinCNO® start up
® ®
medinCNO performs a system test during its start up. If failures are detected, medinCNO informs
the user about them. In the following table, the failures and their possible reasons are described:
Error message Possible reason and reaction
Device error – device shutdown necessary The internal rechargeable battery cannot be
(immediately after the confirmation that no contacted. Check the battery.
patient is connected)
Connect power supply unit – battery low The internal rechargeable battery is not loaded.
The power supply is necessary to start the
®
medinCNO .
3V battery empty – battery has to be exchanged The internal 3V battery has to be exchanged,
because its capacity is low.
Battery needs replacing – device function in If the internal rechargeable battery has been
battery mode cannot be guaranteed. Device recharged ≥300 times, it has to be exchanged,
must be connected to mains for use! because its capacity no longer can be
guaranteed. Exchange the rechargeable battery
to eliminate this error message.
Gas feed pressure low, check air This error message is displayed if the pressure
of the air gas supply is lower than 3 bar.
®
Connect the medinCNO to the gas supply or
check its pressure if it is already connected.
Gas feed pressure low, check O2 This error message is displayed if the pressure
of the O2 gas supply is lower than 3 bar.
®
Connect the medinCNO to the gas supply or
check its pressure if it is already connected.
Gas feed pressure high, check air This error message is displayed if the pressure
of the air gas supply is higher than 6 bar. Check
the pressure of the air supply.
Gas feed pressure high, check O2 This error message is displayed if the pressure
of the O2 gas supply is higher than 6 bar. Check
the pressure of the O2 supply.
Device error – device shutdown necessary If this error message appears, the pneumatic
(displayed after the pneumatic self check has self test has reported a failure. This can be
been performed, which is audible because of the caused by pressure variations during the self
click sound) test. By pressing “ok”, the self test can be
repeated. But if this failure occurs repeatedly,
the pneumatic unit is faulty and you should
contact medin.
®
Error during tube test During the tube test, medinCNO generates a
characteristic flow-CPAP curve for the used tube
®
set and Medijet . If this curve deviates too much
®
from characteristic Medijet curves, this error
message is generated. In this case, check if the
tube set is connected correctly and if all
openings are closed the entire time (e.g. both
®
Medijet openings or the holes for the
temperature sensors of the humidifiers) or if
there are leakages. In addition, CPAP
®
generators other than Medijet will not be

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 39 of 76
®
Maintenance manual medinCNO

accepted.
®
Device error - patient pressure. Device must be medinCNO includes two internal CPAP
replaced. pressure sensors. If these two sensors measure
values which differ too much, this error message
is displayed. In this case, the connection of the
pressure sensors or the main CPU which
includes the pressure sensors is possibly
defective and you should contact medin.
Oxygen calibration failed, check O2 If this error message is displayed in this section
®
of the medinCNO start up test, the oxygen
calibration could not be performed correctly.
This can be caused by an old or defective
oxygen cell or if the gas supply delivers air
instead of oxygen.
®
Fan blocked If the medinCNO cannot detect fan rotation
which is fast enough, this error message is
triggered. In this case, the fan has to be
checked.
Maintenance necessary! Next maintenance If the last maintenance was performed 11 or
necessary: YY-MM-DD more months ago, the user is reminded that the
next maintenance is due soon.

Translations of these error messages into your language can be found in chapter 14.

9.2 Alarms during the use of the medinCNO®


(1) ®
Alarm text Colour ∆t Meaning and reaction of the medinCNO
(
Disconnection (Mode CPAP, red <6s The CPAP pressure is constantly < 2.0 mbar for >
2)
Apnoea CPAP, SNIPPV) 5 seconds - the flow tube or pressure meter tube
®
are no longer connected to the medinCNO .
Please reconnect the tube.
This alarm starts the emergency programme.
Disconnection (Oszillation) red <12 The CPAP average value is constantly < 2.0 mbar
s for > 5 seconds - the flow tube or pressure meter
(2) (3) ®
tube are no longer connected to the medinCNO .
Please reconnect the tube.
The average value interprets the CPAP readings
from the past five seconds which delays the start
and end of the alarm up to 12 seconds.
This alarm starts the emergency programme.
®
CPAP pressure high red <1s CPAP pressure > 18 mbar - the medinCNO briefly
interrupts the flow.
This alarm starts the emergency programme.
Feed pressure high red <1s Feed pressure > 40 mbar, the flow supply tube is
blocked or is generating resistance that is too high.
®
Please check. - The medinCNO briefly interrupts
the flow.
This alarm starts the emergency programme.
Device Error – patient red <1s The patient pressure sensor is causing an error.
pressure. Device must be The two CPAP sensors are measuring a difference
replaced! of > 2 mbar for 500 milliseconds. The device can
no longer be used and must be replaced. Please

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 40 of 76
®
Maintenance manual medinCNO

contact a service technician.


This alarm starts the emergency programme.
®
Device error – device shut red <1s A serious error occurred within the medinCNO .
®
down necessary! The medinCNO can no longer be used and must
be replaced. Contact a service technician.
This alarm starts the emergency programme.
Check gas feed pressure O2 red 3s The O2 feed pressure is too low (<3 bar) or too
®
high (>6.0 bar) - the medinCNO is only providing
air.
This alarm starts the emergency programme.
Check gas feed pressure air red 3s The air feed pressure is too low (<3 bar) or too
®
high (>6.0 bar) - the medinCNO is only providing
oxygen.
This alarm starts the emergency programme.
CPAP pressure high yellow <87 The CPAP average pressure was continuously
(3)
s greater for > 1 second than the set upper CPAP
pressure alarm threshold.
- In a normal case, the average value interprets the
CPAP readings from the past five seconds which
delays the start of the alarm up to <8 seconds.
- If the backup is activated, there may be a
maximum delay of 85 seconds until the average
value has been fully updated, which can result in a
total alarm delay of <87 seconds.
- In SNIPPV mode without active backup, the
average value is updated every 15 seconds, and
for this reason, there is a total alarm delay of <17
seconds.
®
In this case, the medinCNO continues to work
normally.
CPAP pressure low yellow <101 For 15 seconds, the CPAP average pressure was
s constantly less than the set lower CPAP pressure
(2) (3)
alarm threshold.
- In a normal case, the average value interprets the
CPAP readings from the past five seconds which
delays the start of the alarm up to <22 seconds.
- If the backup is activated, there may be a
maximum delay of 85 seconds until the average
value has been fully updated, which can result in a
total alarm delay of <101 seconds.
- In SNIPPV mode without active backup, the
average value is updated every 15 seconds, and
for this reason, there is a total alarm delay of <31
seconds.
®
In this case, the medinCNO continues to work
normally.
Oxygen concentration high yellow 100 The oxygen concentration measured is higher than
s the alarm threshold set for ~1,5 min - the
®
medinCNO continues to operate normally.
Oxygen concentration low yellow 100 The oxygen concentration measured is lower than
s the alarm threshold set for ~1,5 min - the
®
medinCNO continues to operate normally.
®
Temperature yellow 90s The temperature inside medinCNO is too high.

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 41 of 76
®
Maintenance manual medinCNO

Check the environmental conditions and contact a


service technician.
Connect power supply unit - yellow 6s The internal battery is almost empty. The power
battery low supply unit must be connected to the mains
immediately.
Fan blocked yellow 15s The fan is blocked, please contact a service
technician.
Leak-Assist yellow <36 This alarm is triggered if the average CPAP is
s different from the target CPAP for 30 seconds and
if the leak flow of ±2 L/min is not sufficient to
compensate.
®
In this case, the medinCNO continues to work
normally.
®
pure alarm sound, - <1s This alarm is triggered if the medinCNO power
®
medinCNO switched off supply fails completely, for example because the
power supply unit was not connected to the mains
although the battery was discharged.
®
The medinCNO was switched off.
Press Alarm reset to restart red - No alarm – informational message only
flow Triggered by a prolonged high feed pressure (e.g.
if a flow tube is blocked), by several high CPAP
pressures successively or by a combination of high
feed pressures and high CPAP pressures, the flow
supply is interrupted for the time being and is only
reactivated after the Alarm Reset button is
pressed. Reactivation of the flow takes
approximately 10 seconds starting from this
moment.
Push activated Not actually an alarm, just notification that a push
turquoise - was triggered. Users can decide whether or not to
receive this information visually and aurally.
Apnoea Push activated Not actually an alarm, just a notification that a push
was triggered due to apnoea. Users can decide
turquoise -
whether or not to receive this information visually
and aurally.
Backup Push gray - Not actually an alarm, just a notification that a
backup push was triggered due to apnoea. Users
can decide whether or not to receive this
information aurally. However, this information is
always indicated visually.
(1) Delay ∆t: This corresponds to the maximum time which passes from the time the alarm condition occurs until the alarm is
triggered.
(2) After start-up, the disconnection and CPAP pressure low alarm is suppressed for the first 5 minutes of operation. However,
the alarms are activated no later than after 5 min or if a CPAP pressure > 1 mbar for at least 30 sec has already been
measured.
(3) Maximum time delay, including maximum required time to completely re-calculate the average value.

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 42 of 76
®
Maintenance manual medinCNO

The sound level of the alarm is:


In the case of A-weighted
Type of alarm Volume in (dB) background noise volume
(dB)
High priority (red) 69 53
Medium priority (blinking yellow) 67 52
Low priority (steady yellow) 67 52
Information signal (push) 68 53

Alterable alarms:

The user can set the CPAP pressure low, CPAP pressure high (yellow) alarms and the permitted
deviation interval for the oxygen concentration high / low alarms. This can be done using the setting in
the Alarm Settings menu. After the numerical values are adjusted, the modified settings become active
and are saved when this change is confirmed by pressing the dial. No other alarms can be modified.

Alarm settings upon start-up:

Upon start-up, the CPAP pressure high (yellow) alarm is set to 12 mbar and the CPAP pressure low
alarm is set to 3 mbar. The alarm interval for monitoring the oxygen levels set is always set to ±5 % at
start-up.
®
If the medinCNO is restarted, the alarms are reset to these levels. This cannot be changed by the
user.
The alarms - CPAP pressure low (yellow) and CPAP pressure high (yellow) analyse whether the
average value of the CPAP readings is outside of the alarm thresholds (can be set under Alarm
Settings – see chapter Fehler! Verweisquelle konnte nicht gefunden werden.). Thus the alarm
CPAP pressure high is triggered if the CPAP average value was above the alarm threshold for more
than one second. The alarm CPAP pressure low is triggered if the CPAP average value was below the
alarm threshold for more than 15 seconds.
In oscillation mode, the Disconnection alarm is triggered if the average CPAP value was less than
2 mbar for five seconds.
Since the Disconnection (only in oscillation mode), CPAP pressure low and CPAP pressure high
alarms refer to the average CPAP pressure, the start and end of the alarm is thus further delayed (see
table at the start of the chapter). This likewise causes the CPAP pressure low or disconnection alarm
during oscillation to not be reset until the CPAP average value is again above the alarm threshold and
not directly after adjusting the alarm threshold or correcting the problem.
The alarm oxygen concentration high/low is triggered when the measured oxygen concentration is
®
outside of the set limits. These thresholds are always automatically set by the medinCNO in
automatic FiO2 alarm thresholds mode at the level of the set interval (can be set under alarm settings;
see chapter Fehler! Verweisquelle konnte nicht gefunden werden.) around the target oxygen
concentration. In FiO2 alarm thresholds manual mode, they are set via manual updating at the level of
the set intervals around the oxygen target value active at this time. The thresholds that are active due
to the current setting are displayed in the target oxygen concentration field as superscript and
subscript numerals. Thus, for example, at a set target oxygen concentration of 30% and an alarm
interval setting of ±3%, the oxygen concentration low alarm is triggered if the measured oxygen
concentration is less than 27%, and the oxygen concentration high alarm is triggered if the measured
oxygen concentration is greater than 33%.
Warning:
When setting the alarm thresholds for CPAP pressure high (yellow), CPAP pressure low and oxygen
concentration, it must be ensured that the alarm thresholds are set so as to be as limited as necessary
for the patient since otherwise these alarms are unusable.

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 43 of 76
®
Maintenance manual medinCNO

Advice:
When using pushes which generate a CPAP pressure of 18 mbar or more and which are supplied by
means of high push flows or a long patient tube containing a constriction, the feed pressure high and
CPAP pressure high (red) alarms may be generated during pushes. This combination generally also
®
activates the text message Press alarm reset to restart flow, whereby the medinCNO stops the flow
emission until the Alarm Reset button is pushed. After pressing the button, the reactivation of the flow
takes approximately 10 seconds. To avoid this, the push flow must be decreased and constrictions in
the tube system must be eliminated.

Alarm deactivation:
All high-priority alarms (all red alarms) are latching. This means that these alarms are only deactivated
if the cause of the alarm has been corrected and the user has confirmed the alarm. Alarms with
medium and low priority (all yellow alarms) are non-latching and are deactivated automatically if the
reason for the alarm no longer exists. In these cases, the alarm text is displayed for another minute if
no other text message is displayed in the interim.
9.3 Emergency programme
®
If high-priority alarms occur during operation of the medinCNO due to errors or supply breakdowns,
®
the medinCNO tries to maintain operation as long as possible. In doing so, the normal function of the
®
medinCNO is limited as follows:
®
- Instead of oscillation, leak-assist or pushes, the medinCNO limits its operation to normal basic
®
CPAP. That is, the oscillation is stopped and the medinCNO does not emit any more manual
pushes, apnoea pushes, backup pushes and SNIPPV pushes. Likewise, the leak assist
function pauses. After the emergency programme has ended, the oscillation is not
automatically restarted but instead the user must reset this unless only the following alarms
were present: CPAP pressure high, Feed pressure high, Press Alarm Reset to restart flow or
Disconnection. In the case of these alarms, the amplitude and flow values originally set do not
need to be reset, but instead the values are gradually reset automatically.
®
- The mode change of the medinCNO is limited: The mode can be changed only if the
emergency programme has been triggered due to a disconnection alarm. Otherwise the error
must be corrected prior to changing the mode.
®
- The flow emitted by the medinCNO is limited to a maximum of 8 L/min. If a flow of ≤8 L/min
was set previously, this setting will be maintained. Otherwise, only a flow of 8 L/min will be
emitted instead of the flow set. After the emergency programme, the set flow will again be
automatically emitted in CPAP, APNOEA CPAP and SNIPPV mode. In the oscillation mode,
the user must manually perform this together with reactivation of the oscillation. (Exception:
Only these alarms were active: Disconnection, CPAP pressure high, Feed pressure high or
Press Alarm Reset to restart flow)
- If the emergency programme is triggered by a CPAP pressure high or feed pressure high
®
alarm, the medinCNO stops its flow. If the triggering high pressure decreases immediately
®
(< 200 msec), the medinCNO restarts its flow after approximately 4 seconds. If it takes longer
®
than this amount of time for the pressure to decrease, the medinCNO restarts its flow
®
gradually, whereby it takes approximately 10 seconds for the medinCNO to once again emit
its set flow. If the alarm was triggered by a blocked flow supply tube or a combination of several
alarms of the CPAP pressure high or feed pressure high type, the flow is stopped for the time
being and in addition to the feed pressure high / CPAP pressure high alarm, a text message
appears, informing the user that the flow will be restarted by pressing the Alarm Reset button.
- If the emergency programme is caused by the breakdown of one of the supply gases, the
®
medinCNO continues its basic CPAP operation with the remaining gas so that an oxygen
concentration of 21% or 100% develops.
- If there is a serious error in the device, the flow is deactivated and the device must be
restarted.

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 44 of 76
®
Maintenance manual medinCNO

9.4 Signal lights


®
The medinCNO includes two LEDs which indicate the following:
Alarm-LED:
Colour Meaning
Red An alarm with high priority is active – immediate action by the user is necessary.
Yellow An alarm with medium or low priority is active – prompt reaction by the user is
needed.
White Information signal: two short blinks indicates that a push was triggered. This signal
can be deactivated in many situations by the operator as needed

LED inside the rotation knob:


This LED is activated if the operator can change or confirm settings by turning or pressing the dial. If
this LED is deactivated, the dial does not have any function at the present moment.

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 45 of 76
®
Maintenance manual medinCNO

10 Steps during repairs and maintenance of medinCNO®


Attention:
® ®
Whenever the case of medinCNO is opened or whenever something on medinCNO is manipulated,
®
the medinCNO device check has to be performed and its results have to be documented on the
® ®
medinCNO device passport. Otherwise the medinCNO cannot be used and it must not be
connected to a patient under any circumstances.
®
Never lay the medinCNO down on its knob on the front, since doing so will damage the sensor inside
®
the knob. Lay the medinCNO down on a base (e.g. made out of polystyrene) to ensure that no
pressure is applied to the knob.
10.1 Exchanging the oxygen cell
®
To exchange the oxygen cell (OOM-102 or MLF-16) in the medinCNO , open the cover on the lower
®
side of the medinCNO .

Next, disconnect the cable and unscrew the old oxygen cell.
®
Afterwards screw the new oxygen cell into the medinCNO , connect the cable again and close the
cover.
To finish the maintenance, perform a device check, but the new oxygen cell must be calibrated first.
10.2 Opening of the case
Attention:
®
Whenever the back panel of the medinCNO is opened, it should be ensured that adequate ESD
®
safety measures are taken. Otherwise defects can occur inside the medinCNO .
®
Never lay the medinCNO down on its knob on the front, since doing so will damage the sensor inside
®
the knob. Lay the medinCNO down on a base (e.g. made out of polystyrene) to ensure that no
pressure is applied to the knob.
®
Ensure ESD safety measures before opening the medinCNO . Then loosen the 10 screws (M3x14mm
®
TX10) on the back of the medinCNO .

®
Now carefully open the back panel of the medinCNO but use caution when handling the fan cable.
Disconnect it carefully (blue arrow).

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 46 of 76
®
Maintenance manual medinCNO

Next, disconnect the rechargeable battery by removing its plug (yellow arrow). The plug is marked with
the word “AKKU”.
10.3 Exchanging the internal rechargeable battery
Attention:
During this step, ESD safety measures are necessary.
Steps which have to be performed beforehand:
- opening of the case – chapter 10.2
To exchange the internal rechargeable battery, loosen the four screw nuts shown in the following
picture.

Then lift the rechargeable battery and remove it from the case. The plug of the battery is labelled
“AKKU”.
Replace it with the new battery and screw in the 4 screw nuts.
Warning:
®
The internal rechargeable battery of the medinCNO may only be replaced by a battery of the same
type. This can only be ordered at from medin by REF 39-101. Its plug has to be connected with the
socket labelled “AKKU”.
Steps which have to be performed afterwards:
- closing of the case – chapter 10.12

10.4 Exchanging the internal 3V battery


Attention:
During this step, ESD safety measures are necessary.
Steps which have to be performed beforehand:
- opening of the case – chapter 10.2
To exchange the 3V battery, lift the holder shown in the following picture (yellow arrow).

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 47 of 76
®
Maintenance manual medinCNO

Afterwards remove the 3V battery, replace it with a new one (ensure that it is inserted in the right
direction – marked in the socket) and reinsert the holder.
Warning:
®
The internal 3V battery of the medinCNO may only be replaced by a battery of the same type: Type
CR123A manufactured by Varta Microbattery GmbH. This can be ordered from medin by REF 39-102.
It has to be inserted in the plug labelled “BAT1” in the direction marked inside the socket.

Steps which have to be performed afterwards:


- closing of the case – chapter 10.12

10.5 Exchanging the alarm LED


Attention:
During this step, ESD safety measures are necessary.
Steps which have to be performed beforehand:
- opening of the case – chapter 10.2
To exchange the alarm LED board, disconnect its cable from the main CPU. Afterwards, gently open
the clamp and remove the LED board carefully so as not to break the clamp. Once the LED board is
dismounted, the cable can be removed from the LED board.

Connect the cable to the LED board and afterwards insert it into the holder inside the case. Make sure
it is secured through the clamp.
Finally, connect the LED board with the main CPU. Use the plug holder labelled “LED Links”.
Steps which have to be performed afterwards:
- closing of the case – chapter 10.12

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 48 of 76
®
Maintenance manual medinCNO

10.6 Exchanging the knob


Attention:
During this step, ESD safety measures are necessary.
Steps which have to be performed beforehand:
- opening of the case – chapter 10.2
To exchange the knob, disconnect its cable from the main CPU. Afterwards, gently loosen the two
®
screws which secure it. Now remove the black housing of the knob at the back of the medinCNO and
the other half of the knob on the front. The knob cable has to be guided through the black part of the
knob and is removed together with the front part of the knob. After dismounting the knob, the cable
can be pulled out.

To insert the knob again, the cable has to be connected to the knob board. In the next step, the front
part of the knob has to be inserted into the case. During this step, ensure that the grey part on the
front of the knob (yellow arrow) is not pressed. A special tool must be used to push the knob into the
case (green tool and blue arrow), otherwise the sensor in the knob will be damaged.

Now put the knob cable through the big hole in the black part of the knob (yellow arrow in the following
picture) and put the black part from the back over the knob.

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 49 of 76
®
Maintenance manual medinCNO

Finally, secure the knob using the two screws (red arrow) and connect it to plug socket labelled
“Starknop” in the main CPU (blue arrow).
Steps which have to be performed afterwards:
- closing of the case – chapter 10.12

10.7 Exchanging the fan


Attention:
During this step, ESD safety measures are necessary.
Steps which have to be performed beforehand:
- opening of the case – chapter 10.2
To exchange the fan, remove the four screws which secure the fan.

To remount the fan, position the fan according to its blowing direction. The arrow which is marked on
®
the side of the fan has to be pointing at the back wall of the medinCNO (middle picture). Afterwards
secure the fan with four screws (M3 x 22mm), spring washers and plain washers.
Steps which have to be performed afterwards:
- closing of the case – chapter 10.12

10.8 Exchanging the connection board


Attention:
During this step, ESD safety measures are necessary.
Steps which have to be performed before:
- opening of the case – chapter 10.2
To exchange the connection board, open the cable plug and disconnect the cable (yellow arrow).
Then remove the four screws which secure it (red arrows). Now remove the connection board.

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 50 of 76
®
Maintenance manual medinCNO

To insert a new connection board, first ensure that the edge or flat side of the screw nuts inside the
case point toward the back of the case. Otherwise it will not be possible to insert a connection board.
After inserting of the connection board, secure it using the four screws and reconnect it to the cable.
Steps which have to be performed afterwards:
- closing of the case – chapter 10.12

10.9 Exchanging the pneumatic unit


Attention:
During this step, ESD safety measures are necessary.
Steps which have to be performed beforehand:
- opening of the case – chapter 10.2
Removal of the pneumatic unit:
®
If you have a medinCNO which includes a water trap, the water trap has to be removed. To do this,
loosen the screw marked in the following picture and remove the water trap. Normal NIST or DISS
connectors without a water trap must not be removed.

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 51 of 76
®
Maintenance manual medinCNO

The oxygen cell must also be removed– see chapter 10.1 - and the holder for the oxygen cell must be
dismounted (see right-hand picture above).

®
Afterwards, the flow outlet port of the medinCNO (left-hand picture above) and the cable which
connects it with the main CPU have to be removed. Finally, the three screws which secure the
pneumatic unit have to be unscrewed.
Now the pneumatic unit can be removed. To do this, grab the valves of the pneumatic unit with your
left hand and the gas supply connectors with your right hand. Ensure that the board of the pneumatic
unit is not touched during the entire process.

Pull the pneumatic unit with your left hand out of the device and control the direction with the fingers of
your right hand. During this removal, take care when handling the pneumatic board (especially the
area marked by the red arrow). Ensure that the case does not touch it.
Insertion of the pneumatic unit:
Insert the pneumatic unit into the case by grabbing it with your left hand while holding on the valves.
Do not touch the board of the pneumatic unit. Direct the pneumatic unit into the case and its gas
supply connectors through the holes in the bottom of the case. Make sure the board of the pneumatic
unit does not touch the case, especially the area near the red arrow in the picture above.
Then secure the pneumatic unit using the 3 screws. Secure the screws with locking varnish and insert
the holder for the oxygen cell.

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 52 of 76
®
Maintenance manual medinCNO

Now connect the pneumatic unit via the cable to the main CPU and insert the flow outlet connector.

Additionally the oxygen cell must be inserted – see chapter 10.1.


If necessary, remount the water trap; it is recommended that it be remounted after the case has been
closed, which will make it easier to insert the water trap into its holder.
Steps which have to be performed afterwards:
- closing of the case – chapter 10.12

10.10 Exchanging the main CPU


Attention:
During this step, ESD safety measures are necessary.
Steps which have to be performed beforehand:
- opening of the case – chapter 10.2
- removal of the pneumatic unit – chapter 10.9
To exchange the main CPU, first disconnect all cables which are connected to it (yellow arrows):
- the knob cable
- the LED cable
- the cable to the connection board
- the oxygen sensor board
- the cable of the display keys
Then remove the six screws which secure it (red arrows).

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 53 of 76
®
Maintenance manual medinCNO

®
Now lift the main CPU out of the medinCNO and disconnect the pressure measurement tubes.
®
Before inserting a new main CPU into a medinCNO , it must be connected to the pressure
measurement tubes. Make sure you use the correct upper nipple shown in the following picture.

Now you can insert the new main CPU into the case and secure it with 6 screws (M3 x 14mm),
containing a spring washer and a plain washer.
Finally, reconnect the cables (according to their previous disconnection). The oxygen sensor cable
may have to be inserted through the case:
- the knob cable to the plug socked labelled “Starknob”
- the LED cable to the plug socket labelled “LED Links”
- the cable to the connection board to the plug socket labelled “Anschlussplatine”
- the oxygen sensor cable to the plug socket labelled “Sauerstoff”
- the cable of the display keys to the plug socked labelled “Tastfolie”

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 54 of 76
®
Maintenance manual medinCNO

Steps which have to be performed afterwards:


- inserting the pneumatic unit – chapter 10.9
- closing of the case – chapter 10.12

10.11 Mounting the additional power supply holder


Attention:
During this step, ESD safety measures are necessary.
Steps which have to be performed before:
- opening of the case – chapter 10.2
To mount a power supply holder, open the case and puncture the case inside the two bushings at the
right side of the case (red arrows). Carefully enlarge these two holes until they have a diameter of 3
mm, but make sure that you do not damage the screw thread inside the bushings.

Now position the power supply holder to the side of the case and screw in the two M3x14 screws
(yellow arrows).

Steps which have to be performed afterwards:


- closing of the case – chapter 10.12

10.12 Mounting the optional RS232 interface

Furthermore to the USB interface an optional RS232 interface can be mounted:


- This RS232 interface is already pre-cut in the housing.
- When mounting the interface this pre-cut part must be break out carefully at the edge with a
tool e.g. screwdriver. The tool is to be set on the lower edge on the inside of the housing and
the pre-cut part can be pushed with a little force from the inside to the outside.

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 55 of 76
®
Maintenance manual medinCNO

- It is recommended to round the edge with a file to allow easy insertion of the RS232 adaptor
and avoid injuries.
- The same is to perform with the holes on the right and left side of the rectangle.

Mounting of the optional RS232 interface

10.13 Closing of the case


Attention:
During this step, ESD safety measures are necessary.

First, reconnect the rechargeable battery to the main CPU by inserting its plug into the plug socket
®
labelled “AKKU” (yellow arrow), then reconnect the fan of medinCNO . To do this, put the cable into
the plug socket next to the 3V battery which is labelled “Lüfter” (blue arrow). Then insert the back wall
®
of medinCNO but make sure that no cable is pinched between the parts of the case.

Now screw in the 10 screws (M3x14mm TX10) and tighten them gently. As a last step, seal the back
®
plate of medinCNO with a safety seal showing the sign of your company.

Attention:
®
Before using the medinCNO again to treat a patient, a device check has to be performed and passed.
Its results have to be documented on the device passport which can be found in chapter 15.

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 56 of 76
®
Maintenance manual medinCNO

10.14 external filter


10.14.1 Exchange of the external filter
The grid has to be moved towards the back side, in order to remove and change the filter. Afterwards
the grid including the new filter has to be slid into the u-shaped holder until you hear a clear sound.

Tip:
It is only allowed to use a filter of class PPI 30

10.14.2 Upgrade of the external filter


You need the mounting kid (REF 39-320).

1) put the u-shaped part into the mounting aid.

2) Put some glue (Polystyrolkleber) on the contact area at the back side of the grey u-part.
Wait around 1 minute.

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 57 of 76
®
Maintenance manual medinCNO

3) Turn the mounting aid and put it onto the case. Attention, the edges of the mounting aid have
to fit to the edges of the case! (the slanting part of the mounting aid has to be on the top).
Weigh down it with around 1kg and wait at least for 24h.

4) After that disassemble the mounting aid by lifting a bit and push it to the right side.

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 58 of 76
®
Maintenance manual medinCNO

11 Spare Parts and accessories


11.1 List of spare parts
®
There are several spare parts and accessories available for the medinCNO :
Gas supply tubes: Medin offers gas supply tubes according to several standards.
Contact medin for more information.
®
Water trap: When you order a medinCNO from medin, you can additionally
order a water trap. This water trap will then be mounted inside the
®
air supply of your medinCNO .
Patient flow outlet adapter: Connector aluminium M22/F15 REF 51091
White connector REF 39-902
Spare power supply unit: REF 39-113
Power line EU (without power REF 39-115
supply):
Power line UK (without power REF 39-116
supply):
Power line USA/Japan (without REF 39-117
power supply):
Power line Australia (without REF 39-118
power supply):
Spare internal rechargeable REF 39-101
battery:
spare 3V battery: REF 39-102
Spare oxygen cell : REF OOM-102 or REF MLF-16
Exchange filter for water trap REF 39-819
Ferrit for RS232 or USB cabel REF 39-125
Filter REF 39-317
Filter grid REF 39-318
Retrofit kit REF 39-320

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 59 of 76
®
Maintenance manual medinCNO

11.2 Fuses
®
The medinCNO includes 3 internal fuses, 2 of which can be exchanged and one of which is a
polyfuse which cannot be exchanged.
Location (according to Marking Fuse type Exchange possible?
parts in chapter 10)
Main CPU DO-L1901-1A03 Litelfuse Series 452
F1
Rating 0.375A
slow blowing Exchange through
service technician
Connection board DO-L1910-1A01 Litelfuse Series 452 possible
F1 Rating 2A
slow blowing
Pneumatic unit Tyco UF 135 Exchange only
possible through medin
Polyfuse 30V, 2.7A,
slow blowing

Warning:
If the polyfuse on the PCB of the pneumatic unit is permanently broken, it cannot be exchanged in the
field. In the event of this failure, it is very likely that the complete PCB of the pneumatic unit has
permanent damage and because of this, the complete pneumatic unit has to be sent to medin for a
complete check-up and the exchange of this polyfuse.

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 60 of 76
®
Maintenance manual medinCNO

12 Electromagnetic compatibility
®
The information in this section is provided in order to enable the operator of the medinCNO to decide
®
whether the medinCNO is suitable for its electromagnetic environment.
Attention:
®
- The medinCNO is a medical electrical device. Consequently, in order to guarantee the
®
function of the medinCNO , precautions with regard to electromagnetic compatibility must be
®
taken and the medinCNO must be set up and used in accordance with the conditions set out
below.
- Portable or mobile RF communications equipment (e.g. mobile phones) can affect the
® ®
medinCNO and must therefore not be used near the medinCNO . The minimum separation
distance required must be calculated in accordance with chapter 10.2 on the basis of the
transmission frequency.

12.1 Electromagnetic transmission

Guidance and manufacturer’s declaration - electromagnetic emissions


®
The medinCNO is intended for use in the electromagnetic environment specified below. The
®
customer or user of the medinCNO , or the organisation responsible, should assure that it is used in
such an environment.
Emissions test Compliance Electromagnetic environment - guidance
®
The medinCNO uses RF energy only for its
internal function. Therefore, its RF emissions
RF emissions, CISPR11 Group 1
are very low and not likely to cause any
interference in nearby electronic equipment.
RF emissions, CISPR11 Class B ®
The medinCNO is suitable for use in all
Harmonic emissions, establishments, including domestic
Class A
IEC 61000-3-2 establishments and those directly connected to
Voltage fluctuations/flicker the public low-voltage supply network that
Complies supplies buildings used for domestic purposes.
emissions, IEC 61000-3-3

12.2 Electromagnetic immunity

Guidance and manufacturer’s declaration - electromagnetic immunity


®
The medinCNO is intended for use in the electromagnetic environment specified below. The customer or
®
user of the medinCNO , or the organisation responsible, should assure that it is used in such an
environment.
IEC 60601 test Electromagnetic environment -
Immunity test Compliance level
level guidance
± 6 kV contact ± 6 kV contact Floors should be wood, concrete or
Electrostatic
ceramic tile. If floors are covered with
discharge (ESD)
synthetic material, the relative humidity
IEC 61000-4-2 ± 8 kV air ± 8 kV air should be at least 30%.
± 2 kV for power ± 2 kV for power
Electrical fast supply lines supply lines Mains power quality should be that of a
transient/burst typical commercial or hospital
IEC 61000-4-4 ± 1 kV for ± 1 kV for environment.
input/output lines input/output lines

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 61 of 76
®
Maintenance manual medinCNO

± 1 kV differential ± 1 kV differential
mode mode Mains power quality should be that of a
Surge
typical commercial or hospital
IEC 61000-4-5
± 2 kV common ± 2 kV common environment.
mode mode
< 5% UT for 1/2 < 5% UT for 1/2
cycle (> 95% dip) cycle (> 95% dip)
Mains power quality should be that of a
typical commercial or hospital
Voltage dips, short 40% UT for 5 40% UT for 5 environment.
interruptions and cycles (60% dip) cycles (60% dip)
voltage variations Thanks to its built-in battery, the
®
on power supply medinCNO continues to operate during
input lines 70% UT for 25 70% UT for 25 power mains interruptions. It therefore
IEC 61000-4-11 cycles (30% dip) cycles (30% dip) does not need to be powered from an
uninterruptable power supply or external
battery.
< 5% UT for 5 s < 5% UT for 5 s
(> 95% dip) (> 95% dip)
Power frequency The strength of power-frequency
(50/60 Hz) magnetic fields should correspond to that
3 A/m 3 A/m
magnetic field of a typical commercial or hospital
IEC 61000-4-8 environment.
Note: UT is the a.c. supply voltage prior to application of the test level

IEC 60601 test Electromagnetic environment -


Immunity test Compliance level
level guidance
Portable and mobile RF communications
equipment should be used no closer to
®
any part of the medinCNO , including
cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of
the transmitter.
Recommended separation distance:
3 Vrms 150 kHz to
Conducted RF IEC 80 MHz
61000-4-6
1 Vrms 𝑑 = 3.5√𝑃
outside of the ISB
a
bands
10 Vrms
150 kHz to 80
MHz
1 Vrms 𝑑 = 12√𝑃
a
in the ISM bands
10 V/m 𝑑 = 1.2√𝑃 80 MHz to 800 MHz
Radiated RF
80 MHz to 2.5 10 V/m
IEC 61000-4-3
GHz 𝑑 = 2.3√𝑃 800 MHz to 2.5 GHz

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 62 of 76
®
Maintenance manual medinCNO

Where P is the maximum output power


rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
b
distance in metres (m) .
The field strength from fixed RF
transmitters, as determined by an
c
electromagnetic site survey , is less than
the compliance level in each frequency
d
range.
Interference may occur in the vicinity of
equipment marked with the following
symbol:

Note 1: At 80 MHz and 800 MHz, the higher value applies.


These guidelines may not apply in all situations. Electromagnetic propagation is
Note 2:
affected by absorption and reflection from structures, objects and people.
The ISM bands between 150 kHz and 80 MHz are 6.765 MHz to 6.795 MHz;
a
13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz and 40.66 MHz to 40.70
MHz.
The compliance level in frequency ranges of 150 kHz to 80 MHz and 80 MHz to 2.5
GHz is intended to reduce the probability that a mobile/portable communication
b device inadvertently brought into the patient area would cause a disruption. For this
reason, an additional factor of 10/3 was incorporated into the formula for the
calculation of the recommended separation distance and taken into account for
transmitters in this frequency range.
Field strengths from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM
radio broadcast and TV broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, a site
c
survey is to be recommended. If the field strength determined at the site of the
® ®
medinCNO exceeds the compliance level indicated above, the medinCNO must
be observed with regard to its normal operation at each application site. If abnormal
performance is observed, additional measures may be necessary, such as
®
reporienting or relocating the medinCNO .
d Over the frequency range 150 kHz - 80 MHz, the field strength should be less than
1 V/m.

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 63 of 76
®
Maintenance manual medinCNO

Recommended separation distances between portable and mobile communications equipment


®
and the medinCNO
®
The medinCNO is intended for use in an electromagnetic environment in which radiated RF disturbances
®
are controlled. The customer or the user of the medinCNO , or the organisation responsible, can help to
prevent electromagnetic interference by maintaining a minimum distance between portable and mobile
®
RF communications equipment (transmitters) and the medinCNO , as recommended below, according to
the maximum output power of the communications equipment.

Separation distance acc. to transmission frequency


Rated m
maximum
output power 150 kHz to 80 MHz
150 kHz to 80 MHz 800 MHz to
of transmitter outside of the ISM 80 MHz to 800 MHz
in the ISM bands 2.5 GHz
bands 𝒅 = 𝟏. 𝟐√𝑷
W 𝒅 = 𝟏𝟐√𝑷 𝒅 = 𝟐. 𝟑√𝑷
𝒅 = 𝟑. 𝟓√𝑷
0.01 0,35 1.2 0.12 0.23
0.1 1,1 3.8 0.38 0.73
1 3,5 12 1.2 2.3
10 11 38 3.8 7.3
100 35 120 12 23
For transmitters rated at an output power not listed above, the recommended separation distance d in
metres (m) can be estimated using the equation applicable to the relevant column, where P is the rated
maximum output power of the transmitter in watts (W) according to the transmitter manufacturer.
The ISM bands between 150 kHz and 80 MHz are 6.765 MHz to 6.795 MHz;
Note 1:
13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz and 40.66 MHz to 40.70 MHz.
To calculate the recommended separation distance of senders in the ISM frequency
bands between 150 kHz and 80 MHz and in the frequency range from 80 MHz to
Note 2: 2.5 GHz, an additional factor of 10/3 was incorporated and used in the formula to reduce
the probability that a mobile/portable communication device inadvertently brought into
the patient area would cause a disruption.
These guidelines may not apply in all situations. Electromagnetic propagation is affected
Note 3:
by absorption and reflection from structures, objects and people.

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 64 of 76
®
Maintenance manual medinCNO

13 Symbols
®
The symbols with the meanings indicated below are used as labels on the medinCNO or in these
instructions for use:
Symbol Meaning Symbol Meaning

CPAP mode symbol CPAP apnoea mode symbol

SNIPPV mode symbol Oscillation mode symbol

Indicates devices that


Standby mode symbol contain RF transmitters or
deliberately emit RF energy

Charge status of the internal


Manual push trigger
rechargeable battery

Alarm sound paused for the


time indicated. Alarm LED
Conformity mark
and alarm text continue to be
displayed
The device contains batteries
and electrical components.
Consequently it cannot be
Direct current
disposed of in domestic
waste but must be collected
separately.

Manufacturer Device serial number

Do not use device if


Keep dry
packaging is damaged

Date of manufacture - year REF number, order number

Instruction sign “Follow the


Patient flow outlet
instructions for use”

Patient connection pressure


USB interface
meter line

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 65 of 76
®
Maintenance manual medinCNO

Electrical connection to
®
connect the medinCNO Air gas inlet
power supply unit

Interface for external alarm


Oxygen gas inlet
system

Symbols of the power supply:

Symbol Explanation Symbol Explanation

Declaration of conformity UL approval mark. Valid in the USA


according to 93/42/EEC and Canada

WEEE directive Class II equipment

Manufacturer's trademark IP code numbers

Measurement of
ambient temperature/converter Attention – Follow instructions
Temperature class

Read user manual

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 66 of 76
®
Maintenance manual medinCNO

14 Translations of the error messages into different languages


English German French Spanish Italian Netherlands Russian Lithuanian

Apnoea - push activated Apnoe - Push aktiviert Apnée - insufflation Apnea - push activated Apnea - push attivato Apneu - push geactiveerd Поддержка-апное Apnėjos įpūtimo
assistée активирована aktyvavimas
3V Battery empty - 3V Batterie leer - Batterie Batterie 3V – Hors Batería 3V agotada - la Batteria 3V scarica - la 3V batterij leeg - vervang 3 В Батарея разряжена 3V baterija išsikrovė -
battery has to be muss ausgetauscht Service – Remplacez la batería se debe cambiar batteria deve essere batterij - необходимо зарядить baterija turi būti pakeista
exchanged werden Batterie. sostituita батарею
Battery needs replacing - Batterie Austausch nötig - Remplacement de la Cambiar la batería - no Sostituzione della batteria Vervanging batterij Необходимо заменить Būtina pakeisti bateriją -
device function in battery Gerätefunktion kann im batterie nécessaire – se garantiza el necessaria - il noodzakelijk - werking батарею - работа от įrenginio darbas iš
mode cannot be Batteriebetrieb nicht Fonctionnement du funcionamiento en funzionamento del van het toestel op batterij батареи не baterijos
guaranteed. Device must garantiert werden. Gerät dispositif non garanti – batería. La unidad debe dispositivo in modalità is niet gegarandeerd. гарантирована. negarantuojamas.
be connected to mains nur mit angeschlossenem Utilisez le système conectarse a la red batteria non può essere Toestel enkel te Используйте прибор Naudoti įrenginį,
for use. Netzteil verwenden! uniquement sur secteur. eléctrica para su uso garantito. Utilizzare solo gebruiken indien только для работы от maitinamą iš elektros
con dispositivo collegato aangeschakeld aan het сети! tinklo!
all'alimentazione di rete! netsnoer!
Connect power supply Netzteil anstecken - Connectez sur secteur – Conectar la fuente de Collegare l'alimentazione Netsnoer aansluiten - Низкий заряд батареи - Prijungti maitinimo bloką -
unit - battery low Batterie leer Batterie faible. alimentación a la unidad - di rete - batteria scarica batterij leeg подключить шнур išsikrovusi baterija
si la batería esta baja питания
check gas feed pressure Gasversorgungsdruck Vérifiez la pression Comprobar presión Controllare pressione di Controleer проверьте давление tikrinti dujų tiekimo slėgį
air Luft prüfen d'alim. Air alimentación aire alimentazione Aria gastoevoerdruk Lucht подвода воздуха oras
check gas feed pressure Gasversorgungsdruck O2 Vérifiez la pression Comprobar presión Controllare pressione di Controleer проверьте давление tikrinti dujų tiekimo slėgį
O2 prüfen d'alim. O2 alimentación O2 alimentazione O2 gastoevoerdruk O2 подвода О2 O2
CPAP pressure high Hoher CPAP Druck Pression CPAP Haute Presión CPAP alta Pressione CPAP alta Hoge CPAP druk Высокое давление Per didelis CPAP slėgis
СРАР
CPAP pressure low niedriger CPAP Druck Pression CPAP Basse Presión CPAP baja Pressione CPAP bassa Lage CPAP druk Низкое давление СРАР Per mažas CPAP slėgis

Device error - device shut Gerätefehler - Erreur Système – Arrêt Error en la Unidad - Errore Dispositivo - Apparaat storing - Ошибка прибора - Įrenginio klaida - būtinas
down necessary Ausschalten nötig du dispositif obligatoire. apagar el equipo si es Arresto dispositivo uitschakelen noodzakelijk отключить прибор aparato išjungimas
necesario necessario
Device error - patient Gerätefehler - Erreur de contrôle de la Error en la unidad - Errore dispositivo Apparaat storing Ошибка прибора Įrenginio klaida ligonio
pressure. Device must be Patientendruck pression Patient – presión de paciente - la pressione paziente - patiëntdruk - vervanging давление пациента - slėgis - būtina pakeisti
replaced Geräteaustausch nötig Veuillez contacter le unidad debe ser cambio dispositivo apparaat noodzakelijk заменить прибор aparatą
Service Technique. cambiada necessario Or
Or BŪTINA PAKEISTI
Erreur Système – APARATĄ - Įrenginio
Pression Patient – klaida - paciento slėgis
Remplacement du
dispositif obligatoire
Disconnection Diskonnektion déconnection Desconectar Sconnessione Losgekoppeld Отсоединение Atjungimas
Error during tube test – Fehler während Erreur pendant le test du Error durante la prueba Errore durante il test del Storing tijdens Ошибка теста контура - Kontūro testavimo klaida
restart Schlauchtest - Neustart Circuit Patient - de tubos - reiniciar circuito - Riavviare patiëntscircuit test – перезагрузить – Restartas
Redémarrage herstart
Fan blocked Lüfter blockiert Erreur Ventilateur Ventilador bloqueado Ventola bloccata Ventilator geblokkeerd Вентилятор блокирован Ventiliacija užblokuota

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 67 von 76
®
Maintenance manual medinCNO

English German French Spanish Italian Netherlands Russian Lithuanian

Feed pressure high Hoher Versorgungsdruck Pression d'alimentation Presión de alimentación Pressione di Hoge toevoerdruk Высокое давление Per didelis palaikymo
élevée elevada alimentazione alta газов slėgis
Gas feed pressure high, hoher Druck Pression élevée d'alim. Comprobar air, presión Pressione di Gastoevoer druk hoog, Высокое давление per didelis dujų tiekimo
check air Gasversorgung Luft en gaz, vérifiez l'Air de gas alta alimentazione alta, controleer Lucht подводки газа, slėgis, tikrinti orą
prüfen controllare Aria проверьте воздух
Gas feed pressure high, hoher Druck Pression élevée d'alim. Comprobar O2, presión Pressione di Gastoevoer druk hoog, Высокое давление per didelis dujų tiekimo
check O2 Gasversorgung O2 en gaz, vérifiez O2 de gas alta alimentazione alta, controleer O2 подводки газа, slėgis, tikrinti O2
prüfen controllare O2 проверьте О2
Gas feed pressure low, niedriger Druck Pression Basse – Vérifiez Baja presión de Pressione di Lage druk controleer Низкое давление, Oro padavimo žemo
check air Gasversorgung Luft Alimentation AIR alimentación de Aire, alimentazione Aria bassa toevoerdruk Lucht проверить воздух slėgio patikrinimas
prüfen verificar
Gas feed pressure low, niedriger Druck Pression Basse – Vérifiez Baja presión de Pressione di Lage druk controleer Низкое давление, O2 padavimo žemo
check O2 Gasversorgung O2 Alimentation O2 alimentación de Oxigeno, alimentazione O2 bassa toevoerdruk Zuurstof проверить О2 slėgio patikrinimas
prüfen verificar
Maintenance necessary! Wartung nötig! Entretien Obligatoire – Mantenimiento necesario! Manutenzione Onderhoud vereist! Необходима сервисная Aptarnavimas būtinas!
Next maintenance Nächste Wartung fällig: Prochaine révision Proximo mantenimiento necessaria! Volgend onderhoud: проверка! Kitas aptarnavimas
necessary: YY-MM-DD YY-MM-DD necesario Prossima manutenzione YY-MM-DD следующая проверка: būtinas:
YY-MM-DD YY-MM-DD necessaria: YY-MM-DD YY-MM-DD
YY-MM-DD
Oxygen calibration failed, Sauerstoffkalibration Défaut Etalonnage Comprobar O2, fallo en Errore calibrazione Storing O2 calibratie, Сбой калибровки О2, Deguonies kalibravimas
check O2 fehlgeschlagen, O2 Oxygène, Vérifiez O2 calibración de Oxigeno ossigeno, controllare O2 controleer O2 проверьте О2 nepavyko, tikrinti O2
überprüfen
Oxygen concentration Hohe Concentration oxygène Alta concentración de Concentrazione di Hoge Высокая концентрация Per didelė deguonies
high Sauerstoffkonzentration élevée Oxigeno ossigeno alta zuurstofconcentratie О2 koncentracija
Oxygen concentration niedrige Concentration oxygène Baja concentración de Concentrazione di Lage Низкая концентрация Per maža deguonies
low Sauerstoffkonzentration basse Oxigeno ossigeno bassa zuurstofconcentratie О2 koncentracija
Press alarm reset to Drücken Sie Alarm reset Appuyez sur alarme reset Presionar para quitar la “Premere Reset per Druk alarm activeren om Нажать отключение Nuspausti įspėjimo
restart flow um den Flow zu starten – pour redémarrer le alarma - para restaurar el riattivare il flusso” flow te herstarten» тревоги для atstatymą - atnaujinti
débit flujo возобновления потока tėkmę
Push activated Push aktiviert insufflation assistée Push activated push attivato push geactiveerd Вдох активирован įpūtimo aktyvavimas
Safety mode Sicherheitsmodus Mode Sécurité Modo de Seguridad Safety Mode Veiligheidsmodus Безопасный режим Saugus režimas
max. 8 L/min Flow max. 8 l/min Flow Débit max. 8l/min max. 8l/min Flujo Flusso max. 8l/min Max flow 8l/min макс. 8л/мин Поток max. 8l/min.
Temperature Temperatur Température Temperatura Temperatura Temperatuur температура Temperatūra
Leak-Assist Leak-Assist Comp-Fuite Leak-Assist Leak-Assist Lekcomp. Комп. Утечек Nuosr.kompens.

Backup push Backup push Insufflation Backup Backup push Push backup Backup push Вдох вент апное Backup įp.

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 68 von 76
®
Maintenance manual medinCNO

English Czech Polish Norwegian Danish Swedish Greek Chinese

Apnoea – Push activated Apnoe - tlačítko Apnoea - aktywowano Apné - push aktivert Apnø- push aktiveret Apné- push aktiverad Ενεργοποίηση άπνοιας- 发生窒息 - 已启动增压功
aktivováno push ώθησης 能

3V Battery empty - 3V baterie vybitá - baterie Bateria 3V pusta - 3V Batteri tomt - batteri 3V batteri tomt- udskift 3V batteri tomt- Byt Άδεια η μπαταρία 3V - 3V电池耗尽-
battery has to be se musí vyměnit wymień baterie må byttes batteri batteri πρέπει να αλλαχθεί η
请更换电池
exchanged μπαταρία
Battery needs replacing - Baterii je třeba vyměnit - Bateria wymaga wymiany Batteri bør byttes - Batteriet bør udskiftes, Batteriet måste bytas, Η μπαταρία χρειάζεται 请更换后备电池 -
device function in battery samotná nezaručuje - nie ma gwarancji enhetens funksjon i enhedens funktion på enhetens funktion i αντικατάσταση- Η 后备电池已不能保证设备
mode cannot be správnou funkci přístroje. prawidłowej pracy. batterimodus kan ikke batteri kan ikke batterival kan inte λειτουργία μπαταρίας δεν
garanteres. Enheden skal garanteras. Enheten 正常工作。请连接主电源
guaranteed. Device must Ten musí být zapojen do Podłącz urządzenie do garanteres. Enheten må μπορεί να εγγυηθεί την
tilsluttes til strømudtag for måste kopplas till eluttag 使用。
be connected to mains sítě. zasilania sieciowego. kobles til nettstrøm for λειτουργικότητα του
for use. bruk. at anvendes. för att användas. μηχανήματος. Για χρήση
του μηχανήματος
συνδέστε στο δίκτυο .
Connect power supply Připojte ke zdroji - baterie Słaba bateria - podłącz Koble til strømkilde - Tilslut til stikkontakt-lavt Anslut eladapter-lågt Συνδέστε το τροφοδοτικό 后备电池电量低 - 请连接
unit - battery low je slabá zasilanie batteri lavt batteri. batteri ρεύματος -χαμηλή 主电源
μπαταρία
check gas feed pressure Zkontrolujte tlak v přívodu Sprawdź ciśnienie Sjekk gasstrykk Luft kontroller gas Kontrollera ελέγξτε την πίεση 请检查压缩空气压力
air vzduchu zasilające - powietrze forsyningstryk luft gasmatningstryck Luft τροφοδοσίας αέρα
check gas feed pressure Zkontrolujte tlak v přívodu Sprawdź ciśnienie Sjekk gasstrykk O2 kontroller gas Kontrollera ελέγξτε την πίεση 请检查氧气压力
O2 O2 zasilające O2 forsyningstryk O2 gasmatningstryck O2 τροφοδοσίας Ο2
CPAP pressure high CPAP tlak vysoký Zbyt wysokie ciśnienie CPAP trykk høyt CPAP tryk høj CPAP tryck högt Πίεση CPAP υψηλή CPAP压力高
CPAP
CPAP pressure low CPAP tlak nízký Zbyt niskie ciśnienie CPAP trykk lavt CPAP tryk lav CPAP tryck lågt Πίεση CPAP χαμηλή CPAP压力低
CPAP
Device error - device shut Chyba přístroje - přístroj Błąd aparatu - wyłącz Enhetsfeil - enhet må Enhedsfejl- enheden Enhetsfel- enheten Σφάλμα μηχανήματος- 设备故障 - 请关机
down necessary třeba vypnout urządzenie slås av lukkes ned stängs av πρέπει να
απενεργοποιηθεί
απαραιτήτως το
μηχάνημα
Device error - patient Porucha přístroje - Błąd aparatu - ciśnienie Enhetsfeil - pasient trykk Enhedsfejl- Patienttryk. Enhetsfel- Patienttryck. Σφάλμα μηχανήματος- 设备故障 – 压力检测。
pressure. Device must be pacientský tlak. Přístroj pacjenta. Wymień Enhet må byttes Enheden skal udskiftes Enheten måste bytas ut. πίεση ασθενή. Το 请更换设备
replaced se musí vyměnit urządzenie μηχάνημα πρέπει να
αντικατασταθεί. Or
Or Or
Chyba přístroje - tlak Błąd urządzenia - Or 设备故障 – 压力检测。
pacienta. Přístroj se musí ciśnienie pacjenta. Σφάλμα μηχανήματος - 请更换主机
vyměnit Konieczna wymiana πίεση του ασθενή. Το
urządzenia μηχάνημα πρέπει να
αντικατασταθεί.
Disconnection Přerušení spojení Brak połączenia Frakobling Frakoblet Bortkopplad Αποσύνδεση 管路脱落

Error during tube test – Chyba v hadicovém Błąd podczas testu Feil ved test av slanger - Fejl ved Fel vid slangsystemtest - Σφάλμα κατά την 病人回路测试失败 –
restart systému - restartovat układu pacjenta restart slangesystemtest - återstarta διάρκεια ελέγχου του 请重新测试
Restart genstart σωλήνα - επανεκκίνηση

Fan blocked Ventilátor blokován Wentylator zablokowany Vifte blokkert Blæser blokeret Fläkt blockerad Μπλοκαρισμένος 散热风扇堵塞
ανεμιστήρας

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 69 von 76
®
Maintenance manual medinCNO

English Czech Polish Norwegian Danish Swedish Greek Chinese

Feed pressure high Přívodný tlak vysoký zbyt wysokie ciśnienie w Arbeidstrykk høyt Tilførselstryk høj Matningstrycket högt Πίεση παροχής υψηλή 送气压力高
układzie
Gas feed pressure high, Vysoký tlak Zkontrolujte Zbyt wysokie ciśnienie Gasstrykk høy, sjekk Luft Gas forsyningstryk høj, Gasmatningstryck högt, Υψηλή πίεση 供气压力高,请检查压缩
check air přívod vzduchu zasilające, sprawdź kontroller luft kontrollera Luft τροφοδοσίας αερίου, 空气
powietrze ελέγξτε τον αέρα
Gas feed pressure high, Vysoký tlak Zkontrolujte Zbyt wysokie ciśnienie Gasstrykk høy, sjekk O2 Gas forsyningstryk høj, Gasmatningstryck högt, Υψηλή πίεση 供气压力高,请检查氧气
check O2 přívod O2 zasilające, sprawdź O2 kontroller O2 kontrollera O2 τροφοδοσίας αερίου,
ελέγξτε το Ο2
Gas feed pressure low, Nízký tlak plynu, kontrola Niskie ciśnienie, sprawdź Gasstrykk lavt, sjekk luft Gas tilførselstryk lav, Gastrycket för lågt, Πίεση παροχής αέρα 供气压力低,请检查压缩
check air vzduchu powietrze kontroller lufttilslutning kontrollera luftanslutning χαμηλή, ελέγξτε τον αέρα 空气

Gas feed pressure low, Nízký tlak plynu, kontrola Niskie ciśnienie, sprawdź Gasstrykk lavt, sjekk O2 Gas tilførselstryk lav, Gastrycket för lågt, Πίεση παροχής αέρα 供气压力低,请检查氧气
check O2 kyslíku O2 kontroller ilttilslutning kontrollera χαμηλή, ελέγξτε το Ο2
syrgasanslutning
Maintenance necessary! Nutná údržba! Konieczna konserwacja! Vedlikehold nødvendig! Service nødvendigt! Service nödvändig! Nästa Απαραίτητη η 需进行维护保养!
Next maintenance Příští kontrola třeba: Kolejna konserwacja: Neste vedlikehold Næste service service nödvänding: συντήρηση!
下一次保养日期:
necessary: YY-MM-DD YY-MM-DD nødvendig: nødvendig: YY-MM-DD Επόμενη συντήρηση
YY-MM-DD YY-MM-DD YY-MM-DD
YY-MM-DD απαραίτητη:
YY-MM-DD
Oxygen calibration failed, Kalibrace kyslíku se Błędna kalibracja tlenu, Oksygenkalibrering feilet, Ilt kalibrering Oxygen kalibrering Βαθμονόμηση οξυγόνου 氧浓度校准失败,请检查
check O2 nezdařila, zkontrolujte O2 sprawdź O2 sjekk O2 mislykkedes, kontroller misslyckad, kontrollera απέτυχε, ελέγξτε το O2 氧气
O2 O2
Oxygen concentration Koncentrace kyslíku Zbyt wysokie stężenie Oksygen konsentrasjon Iltkoncentration høj Syrgaskoncentrationen Συγκέντρωση οξυγόνου 氧浓度高
high vysoká tlenu høy hög υψηλή
Oxygen concentration Koncentrace kyslíku Zbyt niskie stężenie tlenu Oksygen konsentrasjon Iltkoncentration lav Syrgaskoncentrationen Συγκέντρωση οξυγόνου 氧浓度低
low nízká lav låg χαμηλή
Press alarm reset to Stisknout alarm reset pro Zresetuj alarm aby Trykk alarm reset for Tryk på alarm nulstil for Tryck på larmåterställning Επιλέξτε επαναφορά του 按报警复位键恢复送气
restart flow restartování průtoku wznowić przepływ restarte flow at genstarte flow för att återstarta flödet συναγερμού για την
επανεκκίνηση της ροής
Push activated Stlačení aktivováno Aktywowano push Push aktivert Push aktiveret Push aktiverad Ενεργοποίηση ώθησης 已启动增压功能

Safety mode Bezpečnostní režim Tryb bezpieczny Sikkerhetsmodus Sikkerhedsindstilling Säkerhetsinställning Λειτουργία ασφαλείας 安全模式
max. 8 L/min Flow Tok max. 8 l/min max. 8 l/min Flow maks. 8 l/min Flow maks flow 8L/min. max flöde 8L/min. max. 8 l/min Ροή 最大流量 8 l/min
Temperature Teplota Temperatura Temperatur Temperatur Temperatur Θερμοκρασία 温度过高

Leak-Assist Komp. Úniku Komp. Przecieku Leak-Assist Laekbistand Läck-Assist Leak-Assist 泄漏补偿

Backup push Backup Push Backup push Backup push Backup push Backup push Εφεδρική ώθηση 后备通气加压

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 70 von 76
®
Maintenance manual medinCNO

English Turkish Japanese Rumanian

Apnoea - push activated Apne - Zorunlu soluk アプネア - push 作動 Apnee - apăsare activată
verildi

3V Battery empty - 3V Batarya boş - バッテリー電圧低下- 3V baterie descărcată –


battery has to be bataryayı değiştiriniz bateria trebuie înlocuită
バッテリーを交換して下
exchanged さい

Battery needs replacing - Bataryanın değişmesi バッテリー交換が必要で Bateria trebuie înlocuită –


device function in battery gerekiyor. funcționarea dispozitivului
す。(バッテリー機能が
mode cannot be numai prin alimentarea
-Batarya modundayken 低下しています。コンセ cu baterie nu poate fi
guaranteed. Device must cihazın çalışırlığı garanti ントに接続して使用して garantată. Dispozitivul
be connected to mains edilmez. Cihazı
for use. ください) trebuie conectat la
kullanmak için güç rețeaua de alimentare
bağlantısını yapın!
pentru a funcționa.

Connect power supply Güç kaynağına bağlantı コンセントを接続して下 Conectați unitatea de


unit - battery low yapınız -batarya az alimentare cu energie
さい。-バッテリー低下
electrică – baterie
descărcată

check gas feed pressure Hava kaynağı basıncını 供給空気ガスを確認して Verificați alimentarea cu
air kontrol edin gaz Presiunea aerului
下さい

check gas feed pressure O2 kaynağı basıncını 供給酸素ガスを確認して Verificați alimentarea cu


O2 kontrol edin gaz Presiunea O2
下さい
CPAP pressure high CPAP basıncı yüksek 高CPAP 圧警報 CPAP presiune mare

CPAP pressure low CPAP basıncı düşük 低CPAP 圧警報 CPAP presiune mică

Device error - device shut Cihaz hatası - cihazın 本体異常 シャットダウ Eroare de dispozitiv –
down necessary kapatılması gerekiyor este necesară oprirea
ンして下さい。
dispozitivului

Device error - patient Cihaz hatası - hasta 本体異常 圧力異常。 Eroare de dispozitiv –
pressure. Device must be basıncı. Cihazın presiune pacient.
değişmesi gerekli. 器械を交換して下さい。 Dispozitivul trebuie
replaced
înlocuit.

Disconnection Bağlantı hatası 回路開放・はずれ Deconectare

Error during tube test – Devre testinde hata - 回路チェック エラー Eroare în timpul pornirii
restart Tekrar et tubului - reluare start
restart
Fan blocked Fan bloke edildi ファンが目詰まりしてい Ventilator blocat
ます。

Feed pressure high Besleme basıncı yüksek 高供給圧警報 Presiune înaltă de


alimentare

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 71 von 76
®
Maintenance manual medinCNO

English Turkish Japanese Rumanian

Gas feed pressure high, Gaz besleme basıncı 供給ガス圧上昇 Presiune înaltă a gazului
check air yüksek, havayı kontrol de alimentare, verificare
供給空気ガスを確認して aer
edin
下さい
Gas feed pressure high, Gaz besleme basıncı 供給ガス圧上昇 Presiune înaltă a gazului
check O2 yüksek, O2'yi kontrol edin de alimentare, verificare
供給酸素ガスを確認して O2
下さい

Gas feed pressure low, Gaz besleme basıncı az, 供給ガス圧低下 Presiunea gazului de
check air havayı kontrol edin alimentare este mică,
供給空気ガスをチェック
verificați aerul

Gas feed pressure low, Gaz besleme basıncı az, 供給ガス圧低下 Presiunea gazului de
check O2 oksijeni kontrol edin alimentare este mică,
供給酸素ガスをチェック
verificați O2

Maintenance necessary! Bakım gerekli! メンテナンスが必要です Necesită întreținere!


Next maintenance Următoarea întreținere
Bir sonraki bakım ihtiyacı: !
necessary: necesară: YY-MM-YY
YY-MM-DD YY-MM-DD
YY-MM-DD
Oxygen calibration failed, O2 kalibrasyonu 酸素校正失敗 酸素配管 Reglarea oxigenului a
check O2 başarısız oldu, O2'yi eșuat, verificați O2
を確認して下さい
kontrol edin
Oxygen concentration Oksijen konsantrasyonu 高酸素濃度警報 Concentrația de oxigen
high yüksek mare

Oxygen concentration Oksijen konsantrasyonu 低酸素濃度警報 Concentrația de oxigen


low düşük mică

Press alarm reset to Akışı tekrar başlatmak 流量を復帰させるにはア Apăsați resetarea alarmei
restart flow için Alarm Sıfırlamaya pentru repornirea fluxului
ラームリセットボタンを
basınız 押して下さい。

Push activated Zorunlu soluk verildi Push 作動 Apăsare activată

Safety mode Güvenli mod セーフティーモード Mod de siguranță


max. 8 L/min Flow Flux max. 8 l/min
maks. 8 l/dk akış 最大流量 8 l/min

Temperature Sıcaklık 温度異常 Temperatură

Leak-Assist Kaçak-Asist リークアシスト Leak-Assist

Backup push Back-Up Zorunlu Soluk バックアップ push Impingere datorata backup

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 72 von 76
®
Maintenance manual medinCNO

15 Device Passport
Gerätepass/Device Passport

medinCNO® REF 3090

medin-cno SN: Power supply Lot:

CPU-Monitor SN: Front panel: CNO

Accumulator SN: Gas connection system: DISS NIST

Pneumatic Unit SN: incl. water trap

Oxygen cell SN: Power Supply Plug: X EU UK

Software Version C/P: USA/JP AUS

Displaymakro Version: Accessory: power supply holder

Test equipment: Flow analyser PF-302 Flow analyser PF-300 Other


LOW (IMT) (IMT) _______________

System Test:

LEDs Alarm LED Red Alarm LED Green Alarm LED blue
ok / nok ok / nok ok / nok

Knob LED (white ring) ok / nok

Alarm Buzzer test 1 – several tones ok / nok

Buzzer test 2 – continuous tone ok / nok

Keys Key 01 ok / nok Key 02 ok / nok Key 03 ok / nok

Key 04 ok / nok Key 05 ok / nok Key 06 ok / nok

Key 07 ok / nok Key 08 ok / nok Key 09 ok / nok

Key 10 ok / nok

Display everything black ok / nok

Knob turn clockwise turn anticlockwise press


ok / nok ok / nok ok / nok

Power supply connected ok / nok

disconnected ok / nok

green light of power ok / nok /


supply at customer

Port USB receiving “Test” ok / nok

RS232 (optional) receiving “Test” ok / nok / na/

External alarm Pin 1-2 active non-conducting ok / nok

Pin 3-2 active conducting ok / nok

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 73 of 76
®
Maintenance manual medinCNO

Pressure sensors Sensor 1 _______mbar Sensor 2 _______mbar Sensor 3 _______mbar

max derivation __________mbar ≤ 2 mbar ok / nok

Gas supply pressure O2 Air


external measured: ____________bar external measured: ____________bar
internal measured: ____________bar internal measured: ____________bar
difference ≤ 0,5bar ok / nok difference ≤ 0,5bar ok / nok

Overpressure valve Does it open? pressure in opened situation:


(Close flow outlet port
ok / nok ___________ mbar ≤ 45mbar ok / nok
manually)

Pneumatic valve test Test result: Displayed failure - if not ok:


(performed ok / nok
automatically)

Flow and oxygen given values and Tol.: external measured:


concentration test measured values inside
(Flow values have to be Flow in the tolerances
FiO2 in % FiO2 in % Flow in l/min
measured according to l/min
DIN1343 for:
- 0°C, 25±3Vol% 4 ±1l/min ok / nok
- 1013,25mbar ambient
pressure and 96±3Vol %; 4 ±1l/min ok / nok
- 0% humidity)
25±3Vol %; 15 ±2l/min ok / nok

96±3Vol %; 15 ±2l/min ok / nok

Oxygen sensor test given values and internal measured external measured both values inside
Tol.: in %: in %: the tolerances:

21% ± 2 Vol % ok / nok

40%±2 Vol % ok / nok

60% ±2 Vol % ok / nok

100%±2 Vol % ok / nok

Rechargeable battery Voltage in V: 12V ≤ Value ≤ 17V


check ______________________ ok / nok
(disconnect power
supply) Number of charging Value < 300
cycles: ______________________ ok / nok

Max. charging capacity in Value >3500mAh


mAh: ______________________ ok / nok

3 V Battery Voltage in V: 2V < Value < 3,5V


______________________ ok / nok

medin-cno system Perform the medin-cno system start test – did you pass
start test all tests and did medin-cno start?:
ok / nok

Result: every test passed: ok / nok

if not failures:

Additional accessories:

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 74 of 76
®
Maintenance manual medinCNO

Outlet converter 51091 __________________ ____________________

Production:

Date Sign Name

Final check:

Date Sign Name

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 75 of 76
®
Maintenance manual medinCNO

16 History of changes
Revision Validity date Changes
01 29.03.2012 New document
02 19.02.2013 - Complement of new languages in
survey of error messages
- Complement / change of error
messages “supply gas pressure
high”, “check supply gas pressure”
and “O2 calibration failed”
- USB data output the trend
- adjustment FiO2 alarm
- Update of alarms and error
messages
- Complement of filter and water trap
in maintenance
- reworking of display software
update
03 16.09.2013 Chapter 8: reworking of software
update:
- Update of main CPU software
- Display software update
- Pneumatic software update
04 15.07.2014 - Software change to RS232
interface
- Tables for immunity and separation
distances for life-sustaining device
revised
- Leak assist, backup function
- Optional RS232 interface, ferrites
on USB or RS232 cable
- Change in scale
- New housing design
- Alarm conditions
- Trend data, live data
- Trend
05 08.12.2014 - SW-Update without OSC
06 24.04.2015 - External filter

medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching


+49 8142-44846-0  www.medingmbh.com  info@medingmbh.com
MM_3090_E_Rev06.docx Rev 06 Stand: 24.04.2015 76 of 76

You might also like