Professional Documents
Culture Documents
Experiment No: 4
Title: Determination of Ca, Mg, Fe and Na in mineral waters by flame
atomic spectrophotometry
Name: Sharvani A/P Karthigesu
Matric number: U2103660/1
Name of Partners: 1. Low Ru Yi
2. Nur Eillia Shuhada Binti Ab Rahman
Date: 8/11/2022
Objective
Introduction
Residential water quality from different source varies due to the presence of various
trace elements that are present in the water. Though these trace elements often seem
to be very insignificant, they can lead to many interesting health problems. The trace
elements in our drinking water have to meet the Malaysian Drinking Water Standards.
If it crosses the limit, it may develop severe diseases for human, animals, and plants
and sometimes may cause death (Dkhar, E. N., Dkhar, P. S., & Anal, J. M. H., 2014).
The exposure of trace elements such as Fe, Cu, Zn, Mn, Ni, As and Pb to humans in
higher amount or the bio-accumulation of these elements in the human organ can
cause lethal health effects with various diseases, depending on the nature and
quantity of metal, through ingestion, dermal contact, and inhalation pathways
(Mirzabeygi et al., 2017). So, identification and qualification of these elements become
necessary.
A., J. (2022, September 16). Atomic absorption vs. atomic emission spectroscopy.
PSIBERG. Retrieved November 14, 2022, from https://psiberg.com/atomic-
absorption-vs-atomic-emission-spectroscopy/#google_vignette
Atomic emission spectroscopy measures the intensity of light emitted by the atoms in
excited state. When an excited atom returns to the ground level, it emits radiation
which is detected by detector. The instrumentation and working of the Atomic Emission
Spectrometry is shown in figure 2.
A., J. (2022, September 16). Atomic absorption vs. atomic emission spectroscopy.
PSIBERG. Retrieved November 14, 2022, from https://psiberg.com/atomic-
absorption-vs-atomic-emission-spectroscopy/#google_vignette
A linear calibration curve is plotted for absorbance vs concentration from measuring
the absorbance of a series of standard solutions of a specific range. The concentration
of the sample obtained will then be compared with the calibration plot to determine Fe,
Ca and Mg. For Na, a graph of emission intensity vs absorbance is plotted and the
similar steps are done.
Experimental Procedure
0.04
y = 0.0888x - 0.0065
0.03
R² = 0.9963
Absorbance
0.02
0.01
0
0.00 0.10 0.20 0.30 0.40 0.50 0.60 0.70
-0.01
Concentration of Ca (ppm)
0.1200
Absorbance
0.1000
y = 0.5785x + 0.0078
0.0800 R² = 0.9866
0.0600
0.0400
0.0200
0.0000
-0.05 0.00 0.05 0.10 0.15 0.20 0.25
Concentration of Mg(ppm)
0.1500
0.1000
0.0500
0.0000
0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5
Concentration of Fe(ppm)
Slope or m = 0.0888
The y-intercept = 0.0065
For Ca, the diluted sample of mineral water with dilution factor 100/3 has an
absorbance of 0.064 which falls outside of the calibration range (0.0014 – 0.038). The
concentration of Ca in mineral water could not be determined as the sample
absorbance is higher than the standards. Therefore the calibration equation y =
0.0888x - 0.0065 is not applicable to measure concentration of Ca.
To improve the result, we should reduce the dilution factor by 100/2 or 100/1 and
repeat the experiment till the value falls within the calibration range. It is said because,
as the volume of dilution increases, the number of molecules per unit volume
decreases.
Slope or m = 0.05785
The y-intercept = 0.0078
Slope or m = 0.0648
The y-intercept = 0.0103
For the pure water sample,
The measured absorbance, y is 0.002.
For Fe, the diluted sample of mineral water with dilution factor 100/3 has an
absorbance of 0.002 which falls outside of the calibration range (0.0399 – 0.1632).
The concentration of Fe in mineral water could not be determined as the sample
absorbance is very much lower than the standards. Therefore the calibration equation
y = 0.0888x - 0.0065 is not applicable to measure concentration of Ca.
To improve the results, we should heat the sample to concentrate it. It is because, at
100°C, only water molecules will evaporate while the metal ions don’t. As a
consequence, the number of mol per unit volume of metal ions increases and this
concentrated sample solution can be used to measure the absorbance of Fe.
0.6000
0.5000
0.4000
0.3000
0.2000
0.1000
0.0000
-0.05 0.00 0.05 0.10 0.15 0.20 0.25 0.30 0.35
Concentration of Na(ppm)
Slope or m = 1.8424
The y-intercept = 0.1182
i) For the diluted mineral water,
Dilution Factor: 100/3
The measured absorbance, y is 0.3829.
By substituting y = 0.3829 into the equation y = 1.8424x + 0.1182,
0.3829 = 1.8424x + 0.1182
x = Na concentration = 0.14 ppm
ii) For pure mineral water,
By multiplying x with dilution factor,
Concentration of Mg = 0.14 ppm x 100/3
=4.67 ppm
For Na, the diluted sample of mineral water with dilution factor 100/3 has an emission
intensity of 0.3829 which falls within the calibration range (0.1960 – 0.5492). The
concentration of Na in mineral water can be determined.
The term LOD stands for limit of detection. It is the lowest amount of a
substance that can be distinguished from the absence of that substance (zero
point) with a stated confidence level. In our experiment, the confidence level
for LOD is 95%. We can estimate the detection limit from the standard
deviation of the blank and the slope of the calibration plot along with a defined
confidence factor.
The term loq stands for limit of quantitation. It gives the smallest concentration
of an analyte in a test sample that we can determine with acceptable
repeatability and accuracy. In other words, it is the concentration at which the
entire analytical system must give a recognizable signal and acceptable
calibration point.
If the results lie between the range of LOD and LOQ, it means that a particular
analyte is present in a sample in a value that can be recorded. However, if the
value is lesser than LOD, it implies that the analyte is not detected which
means that the analyte is present in very small concentration. On the other
hand, if the value is more than LOQ, it indicates that the analyte is present in
a amount that can be detected (Madhu, 2020).
Sy
LOD = 3.3 ( )
S
Sy
LOQ = 10 ( )
S
Whereby,
Sy = SE × √N , where SE is the standard error of intercept and N is the number of tests
and
S = Slope of the graph y=mx+c
The LOD and LOQ for Ca, Fe and Na in water samples are calculated based on the
formulas above using the data analysis function in excel and the result obtained is
shown in table 10.
Table 12: The LOD and LOQ values for Ca, Mg, Fe and Na in water sample
Ca Mg Fe Na
SE of intercept 0.00174 0.00516 0.00928 0.02941
SD of
intercept,Sy 0.00388 0.01153 0.02075 0.06576
LOD, ppm 0.144 0.066 1.057 0.118
LOQ, ppm 0.437 0.199 3.202 0.357
√N 2.23607 2.23607 2.23607 2.23607
Slope, S 0.0888 0.5785 0.0648
1.8424
➢ The concentration of Ca ions can be detected as the Ca ions are
present in a high concentration(absorbance higher than calibration
ramge)
➢ The Mg concentration of sample is 5.67 ppm which is higher than LOQ.
Therefore, it can be detected.
➢ The concentration of Fe ions can’t be detected as the Fe ions are
present in a very low concentration(absorbance lower than calibration
range) and the limitation of the instruments used causes it to be unable
to be detected.
➢ The concentration of Na is 4.67 ppm. It falls within the LOD and LOQ
value which shows that Na can be detected in the sample mineral
water.
Table 12: Maximum acceptable value and concentration of element in mineral water
for Ca, Mg, Fe and Na.
There are a few other precautions that should be taken while conducting this
experiment. First, we must use a new micropipette tip to draw out different solutions
to prevent contamination. We must also ensure that there are no air bubbles at the
tip of micropipette. It is also important to ensure that there are no solution adhering
to the wall of the micropipette to obtain a very precise reading. Next, the micropipette
should always be in a upright position to prevent its dysfunctions. Then, all the
volumetric flasks must be calibrated with distilled water prior to use. When
measuring the absorbance of samples in AAS and AES, we must take the
measurement of blank twice.
Conclusions
1. Why is flame emission a more sensitive technique for some cations, mainly
the alkaline and earth alkali cations, while atomic absorption has greater
sensitivity for other cations, such as the transition metal ions?
2. Explain why AAS is so selective, i.e. why do other elements not usually
interfere in the analysis?
The atoms will absorb electromagnetic radiation at a specific wavelength. The
wavelength of light source absorbed by atoms is the same as element that is
being evaluated. Since other elements in the sample won't absorb the
wavelength of interest, this technique becomes highly selective towards the
element of interest only.
References
Madhu, M. (2020, August 9). Difference between LoD and LoQ. Compare the
Difference Between Similar Terms. Retrieved November 13, 2022, from
https://www.differencebetween.com/difference-between-lod-and-loq/
Armbruster, D. A., & Pry, T. (2008). Limit of blank, limit of detection and limit of
quantitation. The Clinical biochemist. Reviews, 29 Suppl 1(Suppl 1), S49–S52.
Mirzabeygi, M., Abbasnia, A., Yunesian, M., Nodehi, R. N., Yousefi, N., Hadi, M., &
Mahvi, A. H. (2017). Heavy metal contamination and health risk assessment
in drinking water of Sistan and Baluchistan, Southeastern Iran. Human and
Ecological Risk Assessment: An International Journal, 23(8), 1893-1905.
https://doi.org/10.1080/10807039.2017.1322895
Dkhar, E. N., Dkhar, P. S., & Anal, J. M. H. (2014). Trace Elements Analysis in
Drinking Water of Meghalaya by Using Graphite Furnace-Atomic Absorption
Spectroscopy and in relation to Environmental and Health Issues. Journal of
Chemistry, 2014, 975810. https://doi.org/10.1155/2014/975810