You are on page 1of 12

Clin Chem Lab Med 2018; 56(8): 1277–1288

Simon Degandta, Bart Peetersa, Stijn Jughmans, Nancy Boeckxb and Xavier Bossuytb,*

Analytical performance of an automated


volumetric flow cytometer for quantitation
of T, B and natural killer lymphocytes
https://doi.org/10.1515/cclm-2017-0638 Conclusions: AQUIOS CL allows for reliable fully auto-
Received July 20, 2017; accepted January 19, 2018; previously mated immunophenotyping of lymphocyte subset quanti-
­published online February 21, 2018
tation. Gating algorithms could be further improved.
Abstract Keywords: AQUIOS; flow cytometry; lymphocyte subsets.

Background: Quantitation of lymphocyte subsets (B cells,


T cells, CD4 and CD8 T cells and NK cells) classically relies
on quantitation of lymphocytes and immunophenotyping Introduction
by flow cytometry. AQUIOS CL (Beckman Coulter) is a fully
automated system that performs an onboard volumetric In clinical laboratories, quantitation of lymphocyte
cell count, automatically processes the sample (staining, subsets (helper and cytotoxic T cells, B cells and NK
lysing and fixation) and analyzes the results. We com- cells) is performed (i) to monitor CD4 T-cell counts in
pared AQUIOS CL to a dual-platform analysis and evalu- HIV-positive patients, (ii) to monitor immunosuppressive
ated analytical performance. or immunomodulatory therapy (e.g. in transplantation
Methods: We evaluated precision, sample stability, inter- and autoimmunity), (iii) to assess immune reconstitu-
sample carryover, linearity and interpanel consistency. tion post-hematopoietic cell transplantation, (iv) to assess
AQUIOS CL was compared to a dual-platform method the immune status in immune-mediated diseases such as
(Sysmex XE-5000 and BD FACSCanto-II). A total of 113 primary immunodeficiencies and (v) to classify tumors of
patient samples were included: 45 from posttransplant hematopoietic and lymphoid tissues [1–4].
patients, 44 from children and 24 random routine samples. Quantitation of lymphocyte subsets classically relies
The degree of automation was scored through the need of on (i) quantitation of the total number of lymphocytes
manual revisions triggered by AQUIOS CL run notifica- (e.g. by a hematology analyzer) and (ii) immunophenotyp-
tions and run flags. ing of lymphocyte subsets by flow cytometry (FCM). For
Results: Intrarun and interrun variability was <9.1% immunophenotyping, labeled antibodies to a set of (cell
with dedicated control material and <32.1% with patient surface) markers is used, such as CD3 (T cells), CD4 (helper
samples. Relative values of lymphocyte subsets could T cells), CD8 (cytotoxic T cells), CD19 (B cells) and CD56/
be determined up to 48 h after venipuncture when the CD16 (NK cells). Immunophenotyping is labor-intensive
sample was kept at room temperature. There was no carry- involving several preanalytical preparation steps such as
over and good linearity. Interpanel consistency was 3.3% sample lysis, cell staining and washing, which are prone
for relative values and 9.4% for absolute values. Method to human mistakes. Moreover, a dual-platform approach,
comparison showed good analytical correlation between in which data obtained by a hematology analyzer are com-
AQUIOS CL and a dual-platform method. Thirty-five bined with FCM data, suffers from high intra- and inter-
percent of the samples triggered a run notification. In 74% laboratory variation [5, 6].
of these samples, the results could be accepted without During the past decade, manufacturers made efforts
intervention, so in 26% of all samples, an unnecessary to automate these processes and developed washing sta-
notification was generated. tions and loading racks for cytometers. Besides, tubes
precoated with monoclonal antibodies were introduced
a
Simon Degandt and Bart Peeters share first authorship. as well as bead-based methods that use counting micro-
b
Nancy Boeckx and Xavier Bossuyt share senior authorship. beads as counting reference allowing quantitation of cells
*Corresponding author: Prof. Dr. Xavier Bossuyt, PhD, MD,
on the flow cytometer. Such bead-based methods elimi-
Department of Laboratory Medicine, UZ Leuven, Herestraat 49,
3000 Leuven, Belgium, E-mail: xavier.bossuyt@uzleuven.be
nate the use of a routine hematology analyzer for leuko-
Simon Degandt, Bart Peeters, Stijn Jughmans and Nancy Boeckx: cyte cell counting. In an attempt to maximally automate
Department of Laboratory Medicine, UZ Leuven, Leuven, Belgium the process, Beckman Coulter released a fully automated

Brought to you by | De Gruyter / TCS


Authenticated
Download Date | 8/6/18 9:10 AM
1278 Degandt et al.: Evaluation AQUIOS CL

volumetric flow cytometer, AQUIOS CL (Beckman Coulter and never surpassed 30 min. No more than 48 h delay was permit-
Life Sciences, Miami, FL, USA). This device is able to ted between sample collection and both measurements, except for
stability studies.
perform an onboard volumetric cell count, to automati-
The study was approved by the Local Ethics Committee. A signed
cally process the sample (staining, lysing and fixation), to informed consent was not needed as leftover samples were used.
analyze the sample including acquisition of cells, gating
of leukocyte/lymphocyte populations, and to prepare a
report. AQUIOS CL is a closed system and runs tests pre- Dual-platform approach: Sysmex XE-5000 and BD
configured by the manufacturer, or user-defined assays FACSCanto-II
through a more open configuration software. For now, two
panels are available, allowing quantitation of T cells, CD4 For the dual-platform approach, the leukocyte count was obtained
and CD8 T cells, B cells and NK cells. from a Sysmex XE-5000 device (Sysmex, Kobe, Japan) using light
So far, two studies that evaluated AQUIOS CL have scattering technology. Leukocyte differentiation was done automati-
cally by the instrument whenever possible. In case a sample was
been reported. Gossez et al. [7] compared CD4+ T lym-
flagged by the instrument and differentiation was not possible, the
phocyte quantitation by volumetric AQUIOS CL to CD4 T
sample was differentiated manually by an experienced technologist.
lymphocyte quantitation by bead-based single platform The system was calibrated and quality controlled according to the
analysis, whereas Grossi et al. [8] compared quantifica- manufacturer’s instructions.
tion of B cells, T cells, T cell subsets and NK cells by volu- A stain/lyse/wash procedure was used for immunophenotypic
metric AQUIOS CL to quantification by bead-based FCM. analysis. Based on the physicians request, the following multicolor
reagent combinations were used: CD3-FITC/CD8-PE/CD45-PerCP/
Both studies showed good agreement between the two
CD4-APC and/or CD3-FITC/CD16.56-PE/CD45-PerCP/CD19-APC (BD
methods, and Grossi et al. [8] reported a reduction in turn- Biosciences). Samples were measured on a BD FACSCanto-II TM flow
around time for the AQUIOS CL system. cytometer (BD Biosciences, San Jose, CA, USA) by analyzing 10,000
In the present study, we compared immunophenotyp- events within the lymphocyte gate. Acquisition and analysis was
ing results obtained with AQUIOS CL to results obtained done by using the BD FACSDiva Software version 6 (BD Biosciences,
San Jose, CA, USA). Gating was done manually by experienced lab
with a dual-platform analysis and performed a more com-
technicians and verified by a medical supervisor. The following lym-
prehensive analysis of the analytical performance, includ-
phocyte populations were enumerated: CD3+, CD3+/CD4+/CD8−,
ing precision, sample stability, intersample carryover and CD3+/CD4−/CD8+, CD3−/CD19+ and CD3−/CD56CD16+.
linearity of the analytical measurement range. For the
correlation study, specific attention was given to samples
with low levels of lymphocytes or lymphocyte subsets by
Single platform volumetric FCM: AQUIOS CL
including samples from a posttransplant and pediatric
population. We also documented the type and number of
The AQUIOS CL system is based on a “Load & Go” principle, which
system errors.
means that the system combines sample preparation and fluores-
cence-based cellular analysis into one system. Briefly, closed tubes
are loaded on the system. After a mixing step, the samples are cap-
Materials and methods pierced and 43 μL is pipetted into a 96-well microplate. The blood is
stained with 13 μL of a monoclonal antibody reagent, which is a com-
bination of four or five murine monoclonal antibodies. After 15 min
Samples of incubation, the blood is lysed using 435 μL of lysing reagents A
and B. Lysing reagent A is a cyanide-free lytic reagent that lyses red
For precision studies, two commercial controls were used: the blood cells in preparation for white blood cell measurement. Lysing
AQUIOS IMMUNO-TROL Cells (Beckman Coulter) and the AQUIOS reagent B slows the reaction caused by reagent A and preserves the
IMMUNO-TROL Low Cells (Beckman Coulter). Both positive cell con- white blood cells for measurement in the flow cell. Finally, 100 μL of
trols were processed in the same manner as the whole blood samples. prepared sample is aspirated for analysis. Absolute leukocyte count
A total of 113 peripheral blood samples were used for the evalu- is based on an electronic-volume measurement. For cell staining, two
ation, including 45 samples from posttransplant patients, 44 pedi- ready-to-use mixes of antibodies were used: AQUIOS Tetra-1 Panel
atric samples (age <16 years) and 24 other routine samples. In all (Beckman Coulter) (CD45 FITC/CD4 RD1/CD8 ECD/CD3 PC5) and
these patients, a routine FCM analysis of T-, B- and/or NK-subsets AQUIOS Tetra-2 Panel (Beckman Coulter) (CD45 FITC/[CD56+CD16]
was requested, and analysis on the AQUIOS CL was done on leftover RD1/CD19 ECD/CD3 PC5). A combination of both panels is referred
sample. as AQUIOS Tetra Combo. These reagents provide identification and
Peripheral blood was collected in K2-EDTA tubes and kept at enumeration of total CD3+, CD3+/CD4+/CD8−, CD3+/CD4−/CD8+,
room temperature until analysis. The blood samples used for method CD4/CD8 ratio, CD3−/CD19+/CD56CD16−, CD3−/CD19−/CD56CD16+
comparison were first analyzed with the current dual-platform lymphocytes in peripheral whole blood, together with CD45+ and
method and kept at room temperature until analysis on the AQUIOS CD45+ Low SS (lymphocytes) count. Dual-positive CD4+/CD8+ cells,
CL system. Time between the two analyses was kept to a minimum however, are not taken into account by this gating strategy. Sample

Brought to you by | De Gruyter / TCS


Authenticated
Download Date | 8/6/18 9:10 AM
Degandt et al.: Evaluation AQUIOS CL 1279

acquisition runs for 45 s. Relative (rel) values are measured as per- seven mixtures on the same day. For each parameter correlation coef-
centage of total lymphocytes and absolute (abs) counts are derived ficients (R2) were calculated after linear regression.
from the absolute leukocyte count. Gating is done automatically by
the system’s software and should only be revised when run notifi- Interpanel consistency: Both Tetra-1 and Tetra-2 Panels were ana-
cations and flags are shown on the result screen. Run notifications lyzed whenever a Tetra Combo analysis was requested (n = 71). In
are generated to convey an abnormal cell distribution or population, order to evaluate consistency between these panels in the same
which may be specimen specific. Run flags indicate an issue that sample, the difference of the common parameter, i.e. CD3+ cells, was
affects the results. Both notifications and flags need manual revision, calculated, both rel and abs. Parallel to our calculation, AQUIOS CL
but run flags are considered more critical. During the evaluation, all generated an error flag (“Inter panel count error”) whenever the dif-
automatic gatings were checked visually, but no active interfering ference of CD3+ cells (rel) of Tetra-1 and Tetra-2 in a Tetra Combo anal-
regarding gating was allowed. In samples with a run notification or ysis was greater than 15%. However, AQUIOS Tetra Software System
flag, the reason was checked, and it was investigated whether these Guide states that the difference of CD3+ cells (rel) in a Tetra Combo
samples should be excluded. Within the evaluation period, internal analysis should not be greater than ±3.5%. Moreover, an interna-
quality control was assessed by running both AQUIOS IMMUNO- tional guideline for performing single-platform absolute CD4+ T-cell
TROL Cells and IMMUNO-TROL Low Cells daily. determinations states that the difference between replicate lineage
AQUIOS CL uses a specific gating strategy for lymphocyte gating markers, such as CD3, should be ≤2% [10].
(Figure 1, see Supplemental Material).

Method comparison between AQUIOS CL


Analytical performance AQUIOS CL and ­dual-­platform approach

Precision: To analyze precision, CLSI document EP15-A3 was In total, 113 patient samples were analyzed for comparison between
­followed [9]. the current approach and AQUIOS CL. Only those tests that were
Intrarun variability was measured by analyzing (Tetra Combo requested by the physician were performed with our current method
analysis) the IMMUNO-TROL Cells (lot 6170007) 13 times in a row and and repeated on the AQUIOS CL system. Therefore, only a part of the
IMMUNO-TROL Low Cells (lot 6180010) 8 times in a row. Precision samples (n = 71) were analyzed with the Tetra Combo panel, whereas
was also measured using patient samples in order to exclude matrix other samples were analyzed with either Tetra-1 (n = 36) or Tetra-2
effects. The following samples were studied: (a) a sample with nor- Panel (n = 6) alone.
mal values was measured with Tetra Combo analysis, (b) a sample For each parameter (CD3+, CD3+/CD4+, CD3+/CD8+, CD3−/
with low CD3+/CD4+ cells was measured with Tetra-1 Panel and (c) a CD19+ and CD3−/CD56CD16+, rel and abs values), the correlation
sample with high CD3−/CD19+ cells and two samples with low CD3−/ between both methods was evaluated by a Passing-Bablok regres-
CD19+ cells were measured with Tetra-2 Panel. The normal sample sion. Confidence interval values of the slope and intercept were
was measured 20 times in a row, the other patient samples 10 times. calculated. All calculations were performed on the total cohort of
Interrun variability was measured by analyzing the IMMUNO- patient samples as well as on the pediatric and the posttransplant
TROL Cells and IMMUNO-TROL Low Cells with Tetra Combo analysis subpopulations, both for Tetra-1 and Tetra-2 Panels.
on 20 different days. To evaluate clinical agreement, all measured parameters of
Imprecision was calculated using the coefficient of variation (CV). each sample were categorized as low, normal or high for both meth-
ods based on published reference values [11]. A Cohen’s kappa coef-
ficient of agreement was calculated based on the classifications of
Sample stability: Sample stability was tested with a sample from a
healthy male volunteer. Between analyses, the sample was stored at both methods.
room temperature for five consecutive days (day 0 until day 4). Each
day, Tetra Combo analysis was performed in duplicate on the sample
Flagging and notifications of AQUIOS CL
with an interval of 24 h. For each parameter, the % difference with the
value measured on day 0 was calculated.
All run flags and notifications were documented and listed in order
Intersample carryover: Carryover was evaluated with two patient to give an idea in how many cases revision of the results was neces-
samples: a sample with low CD3−/CD19+ cells and a sample with sary. All flags and notifications due to the evaluation setting, such as
high CD3−/CD19+ cells. The Tetra-2 Panel was measured on each “Missing QC” and “Sample ID re-used”, were omitted.
sample separated by a system flush, to obtain reference values. Next,
the Tetra-2 Panel was measured 10 times alternately on both samples
without intermittent flushing of the tube system. Afterwards the 95% Statistics
confidence interval of each parameter in the 10-fold tandem run was
calculated and compared with the reference value. Statistical analysis was performed using Microsoft Excel® 2013 (Micro-
soft Corp., Redmond, WA, USA) and Analyse-it® for Microsoft Excel
Linearity analytical measurement range: Two pathological samples version 3.90 (Analyse-it Software Ltd., Leeds, UK). Imprecision was cal-
were selected for linearity assessment: one sample with low CD3−/ culated using the CV. Linearity was measured with linear regression and
CD19+ cells and one sample with high CD3−/CD19+ cells. Both sam- R2. Correlation between methods was evaluated with Passing-Bablok
ples were mixed in seven different ratios (10/0, 9/1, 8/2, 6/4, 4/6, 2/8, regression and Bland-Altman regression. Clinical agreement was evalu-
0/10). The Tetra Combo analysis was performed in duplicate on all ated by Cohen’s kappa coefficient (observed unweighted kappa).

Brought to you by | De Gruyter / TCS


Authenticated
Download Date | 8/6/18 9:10 AM
1280 Degandt et al.: Evaluation AQUIOS CL

A
All data points
1000

800

600

FS-Lin
400

200

0
0 200 400 600 800 1000
SS-Lin
B C
Cells Cells
1000 1000

800 800

600 600
EV-Lin
SS-Lin

400 400

200
200

0 Lymphs (45)
0
100 101 102 103 104 Lymphs EV
CD45 0 200 400 600 800 1000
SS-Lin
OR AND

D Lymphs all E
50 Lymphs
50
CD3– All
CD3+ All CD3–
CD3+

25 25

0 0
–101100 101 102 103 104 –101 100101 102 103 104
CD3 CD3

F G
CD3+ All CD3–
104 104

103 103
CD56/16
CD4

102 102

101 101
100 100
–101
–101
–101 100 101 102 103 104
–101 100 101 102 103 104
CD8
CD19

Figure 1: Gating strategy on AQUIOS CL.


Leukocytes are selected on a SS/FS plot in order to exclude debris (A). Lymphocytes are gated on a CD45/SS plot (B) and on a SS/EV plot (C).
All events of both gates are combined in one gate, “Lymphs all” (D), and only joint events are combined in another gate, “Lymphs” (E). From
“Lymphs all”, CD3+ events are selected on a CD3 histogram (D), and from this gate, CD4+/CD8− and CD4−/CD8+ events are measured on
a CD4/CD8 plot (F). From “Lymphs”, CD3− events are selected on a CD3 histogram (E), and from this gate, CD19+/CD56CD16− and CD19−/
CD56CD16+ events are measured on a CD19/CD56CD16 plot (G).

Brought to you by | De Gruyter / TCS


Authenticated
Download Date | 8/6/18 9:10 AM
Degandt et al.: Evaluation AQUIOS CL 1281

Results Interpanel consistency

Tetra Combo Analysis was performed on 71 samples. As


Analytical performance AQUIOS CL
eight of these samples were excluded for method compari-
son, interpanel consistency was evaluated on the remain-
Precision
ing 63 samples. The mean difference between CD3+ cells
(rel) on Tetra-1 Panel and Tetra-2 Panel was 3.3% and
For IMMUNO-TROL Cells and IMMUNO-TROL Low Cells,
between CD3+ cells (abs) 9.4%. None of the differences
all CVs were ≤5% for high counts (≥300 cells/μL or >20%)
between CD3+ cells (rel) exceeded 15%, however, for CD3+
and ≤10% for low counts (<300 cells/μL or ≤20%) both for
cells (abs), three samples exceeded the limit of 15% dif-
intrarun and interrun precision (Table 1).
ference. These were all samples with low absolute CD3+
These criteria were not always met for the patient
counts (114, 151 and 285 counts/μL). Applying the limit
samples (Table 1). For the normal sample, CD3+/CD8+ (rel
of 2% difference as stated in international guidelines, as
and abs) and CD3−/CD56CD16+ (abs) CVs were between
many as 16 samples (or 25.4%) of the relative values and
5% and 10%. The first CD3−/CD19+ low sample showed
44 samples (or 69.8%) of the absolute values should not
a high CV for CD3−/CD19+ (abs) (10.76%), but the mean
be accepted.
value was situated around the manufacturer’s defined
lower limit of quantification (25 cells/μL). For the CD3−/
CD19+ high sample, all were below 10%. The second
Method comparison between AQUIOS CL
CD3−/CD19+ low sample and the CD3+/CD4+ low sample
showed for most parameters high imprecision; however, and dual-platform approach
all values with a CV >15% were situated around the manu-
A total of 113 blood samples were initially included in the
facturers defined limits of quantification.
method comparison study. However, 10 patient samples
were excluded from the statistical analysis for different
Sample stability reasons. Two samples were highlighted by AQUIOS CL
with the run flag “Insufficient lymphocytes”. The first
Relative values for lymphocyte subsets remained stable sample had 132 lymphocytes/μL on AQUIOS CL and 442/μL
during the first 2 days but showed significant instability on XE-5000. The other sample had 85 lymphocytes/μL on
on day 3 (−17.85%) for CD3−/CD19+ cells and on day 4 for AQUIOS CL and 44/μL on XE-5000 combined with manual
the other populations (Table 2). A difference was consid- differentiation. In a third sample, there was an incorrect
ered significant if it was higher than 10% (for values ≤20% CD3+ gating in the Tetra-2 Panel, noticed through the error
or <300 cells/μL) or higher than 5% (for values >20% or flag “Inter Panel Count Error”. The fourth sample was a
≥300 cells/μL), according to the manufacturer’s impreci- sample mainly consisting of atypical blasts. The remain-
sion criteria. ing six samples showed suboptimal gating necessitat-
All absolute values were unacceptable on day 3. Abso- ing manual gating or a rerun. These were highlighted by
lute values for CD19+ cells were unacceptable after 1 day AQUIOS CL through the run notification “Potential sample
and CD3+ cells of the Tetra-2 Panel after 2 days (Table 2). or gating issue”.
The absolute lymphocyte count degraded rapidly, After exclusion, 103 samples were used for the com-
whereas the relative values were stable for at least 3 days. parison of which 63 were analyzed with Tetra Combo
panel, 34 with Tetra-1 Panel alone and six with Tetra-2
Panel alone. For the pediatric population, 42 samples
Intersample carryover
were compared, 40 with Tetra Combo panel, two with
Tetra-1 Panel alone and none with Tetra-2 Panel alone. For
During the 10-fold tandem run, all obtained values
the posttransplant population 42 samples were compared,
remained within the 95% confidence interval, so no carry­
17 with Tetra Combo panel, 25 with Tetra-1 Panel alone
over was observed (see Supplemental Material, Table 1).
and none with Tetra-2 Panel alone.

Linearity analytical measurement range


Statistical comparison
All parameters showed excellent linearity with R2 ≥0.99.
For CD3−/CD56CD16+, however, no R2 was calculated as For each parameter, correlation was measured for relative
the evaluated range was too narrow (Table 3, Figure 2). values (percentage of lymphocytes) and for absolute values
Brought to you by | De Gruyter / TCS
Authenticated
Download Date | 8/6/18 9:10 AM
1282

Table 1: Intra- and interrun precision of the AQUIOS CL Tetra-1 Panel (upper section) and the Tetra-2 Panel (lower section).

Precision Sample type Counts CD3+ T1 (rel) CD3+ T1 (abs) CD3+/CD4+ (rel) CD3+/CD4+ (abs) CD3+/CD8+ (rel) CD3+/CD8+ (abs)

Intrarun (CV%) Patient sample: normal 20 1.14 3.14 1.95 2.93 6.95 7.61
Patient sample: low CD3−/CD19+ (1) 10 n/a n/a n/a n/a n/a n/a
Patient sample: low CD3+/CD4+ 10 6.10 14.26 26.34 32.10 8.13 12.97
Degandt et al.: Evaluation AQUIOS CL

Patient sample: high CD3−/CD19+ 10 n/a n/a n/a n/a n/a n/a
Patient sample: low CD3−/CD19+ (2) 10 n/a n/a n/a n/a n/a n/a
IMMUNO-TROL 13 1.43 2.28 1.83 2.77 1.93 2.20
IMMUNO-TROL Low 8 2.33 2.63 4.10 4.85 2.58 2.65
Interrun (CV%) IMMUNO-TROL 20 2.54 3.16 1.52 3.68 3.88 3.93
IMMUNO-TROL Low 20 2.46 3.97 3.60 5.14 4.91 4.30
CD3+ T2 (rel) CD3+ T2 (abs) CD3−/CD19+ (rel) CD3−/CD19+ (abs) CD3−/CD56CD16+ (rel) CD3−/CD56CD16+ (abs)

Intrarun (CV%) Patient sample: normal 20 1.21 2.94 4.43 7.64 3.12 5.25
Patient sample: low CD3−/CD19+ (1) 10 2.35 1.98 9.59 10.76 6.44 8.79
Patient sample: low CD3+/CD4+ 10 n/a n/a n/a n/a n/a n/a
Patient sample: high CD3−/CD19+ 10 1.72 4.76 0.25 4.53 5.99 6.85
Patient sample: low CD3−/CD19+ (2) 10 1.18 3.51 13.82 13.87 16.38 18.86
IMMUNO-TROL 13 1.14 2.25 2.26 3.60 3.61 4.38
IMMUNO-TROL Low 8 2.17 3.75 4.97 6.02 4.82 9.09
Interrun (CV%) IMMUNO-TROL 20 1.76 4.88 4.08 4.86 4.69 7.60
IMMUNO-TROL Low 20 2.29 4.98 5.18 7.28 6.64 7.96

CV, coefficient of variation; n/a, not applicable.

Brought to you by | De Gruyter / TCS

Download Date | 8/6/18 9:10 AM


Authenticated
Degandt et al.: Evaluation AQUIOS CL 1283

Table 2: AQUIOS CL Tetra-1 (upper section) and Tetra-2 (lower section) sample stability at room temperature for 4 days.

Day CD3+ T1 (rel) CD3+ T1 (abs) CD3+/CD4+ (rel) CD3+/CD4+ (abs) CD3+/CD8+ (rel) CD3+/CD8+ (abs)

1 1.31 −0.33 1.38 −0.26 2.71 1.12


2 −1.84 −2.51 −1.35 −0.78 −0.95 −1.36
3 2.13 −9.53 2.84 −8.94 2.87 −8.68
4 7.57 −10.89 7.29 −11.16 8.18 −10.3

CD3+ T2 (rel) CD3+ T2 (abs) CD3−/CD19+ (rel) CD3−/CD19+ (abs) CD3−/CD56CD16+ (rel) CD3−/CD56CD16+ (abs)

1 1.85 −2.48 −8.67 −12.50 −0.34 −4.67


2 −0.33 −7.75 −5.47 −12.50 2.19 −5.69
3 2.11 −12.63 −17.85 −29.50 −1.23 −15.45
4 7.81 −17.70 −36.28 −51.50 −20.34 −39.43

Relative differences (in %) compared to reference value on day 0. Bold: difference higher than 10% (for values ≤20% or <300 cells/μL) or
higher than 5% (for values >20% or ≥300 cells/μL).

Table 3: Linearity of AQUIOS CL analytical measuring range.

CD3+ T1 CD3+ T2 CD3+/CD4+ CD3+/CD8+ CD3−/CD19+ CD3−/CD56CD16+ Lymphocytosis


(abs) (abs) (abs) (abs) (abs) (abs)

Analytical measuring 55–4700 55–4700 35–3000 45–1600 25–1000 20–1000 20–4700


range (manufacturer)
Experimentally 27.5–2285 24–2632.5 12–1104 11–1206.5 3–692 308.5–444 495.5–5885.5
checked range
R2 0.99 1.00 0.99 0.99 1.00 n/a 1.00

n/a, not applicable.

Figure 2: Linearity of AQUIOS CL analytical measuring range.

Brought to you by | De Gruyter / TCS


Authenticated
Download Date | 8/6/18 9:10 AM
1284 Degandt et al.: Evaluation AQUIOS CL

A (cell count/μL) (Figure 3). Tables with correlation values


4000 between AQUIOS CL and the dual-platform approach,
together with correlation plots are provided in the supple-
3500
mentary data (see Supplemental Material, Tables 2 and 3,
3000
Figures 1–24).
CD3+/CD4+ (count/µL): AQUIOS CL

The parameters in the Tetra-1 Panel of AQUIOS CL


2500 showed small but statistically significant proportional
negative biases for CD3+ (abs), CD3+/CD4+ (rel and abs)
2000
and CD3+/CD8+ (rel and abs) and a constant positive bias
for CD3+/CD4+ (rel) when compared to the dual-platform
1500
Passing-Bablok fit approach in 97 samples. The pediatric subpopulation
(y = 3.723 + 0.9271x)
1000 (n = 42) showed only a slight statistically significant pro-
portional negative bias for CD3+ (rel), CD3+/CD4+ (rel)
500 and CD3+/CD8+ (rel). The posttransplant subpopulation
(n = 42) showed proportional negative bias for CD3+ (abs)
0
0 500 1000 1500 2000 2500 3000 3500 4000 and CD3+/CD8+ (rel and abs), and a constant positive bias
CD3+/CD4+ (count/µL): dual-platform approach
for CD3+/CD4+ (rel).
B
The parameters in the Tetra-2 Panel of AQUIOS
400
CL showed significant proportional negative biases
200
for CD3+ (abs) and CD3−/CD56CD16+ (rel) and a con-
AQUIOS CL – dual-platform approach

0 stant positive bias for CD3−/CD56CD16+ (rel and abs)


CD3+/CD4+ (count/µL):

–200 in 69 samples. In the pediatric subpopulation (n = 40),


a proportional negative bias was observed for CD3+
–400
Mean
(rel and abs) and a constant positive bias for CD3−/
–600 (–62.5) CD56CD16+ (rel). In the posttransplant population
–800 95% LoA (n = 17), a proportional negative bias was observed for
(–358.7–233.7)
–1000 CD3−/CD56CD16+ (abs) and a constant positive bias for
CD3−/CD19+ (rel).
–1200
0 500 1000 1500 2000 2500 3000 3500 4000
CD3+/CD4+ (count/µL):
dual-platform approach + AQUIOS CL/2

Figure 3: Regression between dual-platform approach and AQUIOS CL. Clinical agreement
(A) Passing-Bablok regression (red line) between dual-platform
approach (x-axis) and AQUIOS CL (y-axis). Green lines indicate Kappa coefficients for clinical agreement between the two
reference values. This example shows Passing-Bablok regression
methods, together with the 95% confidence interval, are
for CD3+/CD4+ (abs); (B) Bland-Altman regression (full blue line)
between dual-platform approach and AQUIOS CL. Dotted blue lines
shown in Table 4.
indicate 95% limit of agreement. This example shows Bland-Altman All kappa coefficients were higher than 0.71 showing
regression for CD3+/CD4+ (abs). substantial agreement between both methods. For some

Table 4: Clinical comparison (kappa coefficient) between AQUIOS CL and the dual-platform approach.

CD3+ T1 (rel) CD3+ T1 (abs) CD3+/CD4+ (rel) CD3+/CD4+ CD3+/CD8+ (rel) CD3+/CD8+ (abs)
(abs)

Kappa coefficient 0.75 [0.62–0.88] 0.90 [0.82–0.99] 0.91 [0.83–0.99] 0.98 [0.94–1.00] 0.80 [0.67–0.92] 0.84 [0.73–0.95]
[95% CI]
CD3+ T2 (rel) CD3+ T2 (abs) CD3−/CD19+ CD3−/CD19+ CD3−/CD56CD16+ CD3−/CD56CD16+
(rel) (abs) (rel) (abs)

Kappa coefficient 0.79 [0.64–0.94] 0.97 [0.90–1.00] 0.94 [0.85–1.00] 1.00 [1.00–1.00] 0.81 [0.67–0.94] 0.71 [0.52–0.91]
[95% CI]

CI, confidence interval.

Brought to you by | De Gruyter / TCS


Authenticated
Download Date | 8/6/18 9:10 AM
Degandt et al.: Evaluation AQUIOS CL 1285

parameters, kappa coefficient was higher than 0.90, indi- the current method. Therefore, results were not omitted
cating an almost perfect clinical agreement. from method comparison. Nevertheless, when subtract-
ing these samples in counting of notifications, 35 (35%)
of 100 samples showed a notification which was not due
Flagging and notifications of AQUIOS CL to mechanical problems of the device. After revision of
these 35 samples with notifications, 26 could be accepted
Eight samples out of 113 (7.1%) had one or more run flag, without any intervention, 2 should be rerun and in 7
and 48 samples (42.5%) had one or more run notification. samples the lymphocyte-gate on EV/SS plot and on SS/
Run flags were “Inter Panel Count Error” (n = 3), CD45 plot should be optimized manually. As mentioned
“Insufficient Lymphs” (n = 2), “Over Lyse Incub” (n = 1), above, these samples were excluded from analytical
“Clog or Bubble” (n = 1), “ReRun” (n = 1) and “High Count method comparison.
Rate” (n = 1). The flag “Insufficient Lymphs” occurs when
less than 500 lymphocytes are analyzed. Samples with
this flag showed low lymphocytic counts with the current
method (442 lymphocytes/μL and 44 lymphocytes/μL). Discussion
These samples were excluded from method compari-
son as mentioned above. The sample with “High Count This study evaluated the analytical features of AQUIOS
Rate” had a leukocyte count of 213.36 × 109/L with the CL and compared AQUIOS CL to a dual-platform method
current method and mainly consisted of lymphoblasts. (Sysmex XE-5000 and BD FACSCanto-II).
This sample was also excluded from method compari- Precision study showed an acceptable intrarun and
son. Three samples flagged an “Inter Panel Count Error”. interrun variability when dedicated control material was
This flagging is specific for the Tetra Combo analysis used. Comparable results were obtained by Gossez et al.
and occurs when there is a difference greater than 15% [7] for CD3+/CD4+ (abs). However, our precision study
between the corresponding populations of the Tetra-1 and showed clearly higher CVs for patient samples (ranging
Tetra-2 Panels. In two samples, no abnormalities could be from 0.25% to 26.34% for relative values and from 1.98%
seen, but in the third sample, there was an incorrect CD3+ to 32.10% for absolute values). Few to no precision crite-
gating in the Tetra-2 tube. Therefore, this sample was also ria are available for FCM values. Ricos et al. defined a CV
excluded from method comparison. The sample with run of 12.5% as a desirable imprecision for only CD3+/CD4+
flag “Clog or Bubble” also showed the flag “ReRun”, but (abs), a criterion that was met for a patient sample with
no abnormalities were noticed. The flag “Clog or Bubble” normal values, but not for patient samples with low CD3+/
is shown whenever there is an abrupt interruption of the CD4+ cells. Moreover, high CVs were especially noticed for
events over time. The sample was rerun and the same low values, around the lower limit of the analytical meas-
values were obtained as before, but without run flags. The uring range. Focusing on the concentrations of absolute
flag “Over Lyse Incub” means that for some reason (e.g. CD4+ T-cell count with the highest impact on clinical deci-
unforeseen shutdown of the system) the sample was too sions (100–300 counts/μL), a good precision was noticed:
long in the lysis solution before measurement. The sample dedicated control material, AQUIOS IMMUNO-TROL Low
with this flag was rerun and the values corresponded to Cells, showed a mean concentration of 147.50 counts/μL
those obtained before. with a CV of 4.85%.
In all but one sample with run notifications (n = 47), Relative values of lymphocyte subset counts seemed
the notification was “Potential sample or gating issue”, to be stable at least up to 48 h when the K2-EDTA whole
meaning that on one plot the gating could be possibly blood sample is stored at room temperature, which is
incorrect. The other sample showed the notification “Low longer than demonstrated by other authors [12, 13] and
CD8”. All notifications were reviewed and apparently a longer than the manufacturer’s recommendation of 24 h.
minor problem with the device was noticed: a problem in Next to CD19+ staining, especially leukocyte count was
the flow cell rendered incorrect electronic volume (EV)/ most prone to deterioration over time, causing more insta-
side scatter (SS) plots, causing notifications in 13 consecu- ble absolute values [14]. These results imply that clinical
tive samples. AQUIOS CL uses this EV/SS plot for lympho- samples can be stored overnight before analysis without
cyte gating as an addition to lymphocyte gating on the causing significant mistakes. When only relative values
SS/CD45 plot. However, for these 13 samples, lymphocyte are considered, samples can even be stored for 2 days
counting was correct and all results correlated well with before analysis.

Brought to you by | De Gruyter / TCS


Authenticated
Download Date | 8/6/18 9:10 AM
1286 Degandt et al.: Evaluation AQUIOS CL

Interpanel consistency was acceptable according to notification “Potential sample or gating issue” which also
the applied limit of 15%. However, when using the 2% should trigger manual revision. This incorrect CD3-gating
limit as described in MMWR guideline [10], the rejection is a consequence of the gating algorithm the AQUIOS CL
rate rose to 25.4% for CD3 (rel) and 69.8% for CD3 (abs). software uses. In order to distinguish between the nega-
Method comparison showed a good analytical corre- tive and positive population, the software uses a histo-
lation between AQUIOS CL and our current dual-platform gram CD3-plot and searches for a valley between the two
method. Some small proportional and constant biases populations to place a cutoff between the positive and
were noticed, but these were too small to be of clinical the negative populations. In case only one population is
significance. present (either positive or negative), the software cannot
AQUIOS CL is presented as a “Load & Go” device. find a valley and places its cutoff before or after the only
During this study, there was indeed a clear reduction population present, together with a notification “Poten-
in hands-on time. Lab technicians only had to load rea- tial sample or gating issue”. A suggestion to optimize this
gents and racks with (closed) patient samples, while all gating strategy is to use a cutoff based on absolute fluores-
further preanalytical and analytical steps were done cence intensity values when no valley is detected.
by the AQUIOS CL system. Even gating of all popula- Another drawback of the gating strategy is neglecting
tions of interest was done automatically through gating-­ the possible presence of a dual-positive CD4+/CD8+ T-cell
algorithms by the software system. However, this last step population. The presence of less than 5% of this popula-
was the most cumbersome step, and manual review was tion in peripheral blood is considered as premature release
necessary for some samples. In order not to undermine of CD4+ CD8+ T-cells from the thymus into the periphery
the “Load & Go” principle, these cases should be kept to [15]. However, more than 5% CD4+/CD8+ T cells should be
a minimum and the users should be properly warned if considered as abnormal and should be further explored.
manual revision is needed. AQUIOS CL warns for revision Another important remark on the use of AQUIOS CL
using run flags and notifications. During this study 7.1% is its incompatibility with the stabilized blood samples
of the samples showed a run flag. In most of the cases, a universally used for certified external quality assessment
rerun could resolve the problem (i.e. for flags “Inter Panel (EQA) schemes such as UNKEQAS Immune Monitoring.
Count Error”, “Over Lyse Incub” and “Clog or Bubble”). In On AQUIOS CL, manual gating was necessary in order to
case of flag “Insufficient Lymphs”, there are insufficient prevent a slight negative bias as the AQUIOS algorithm
lymphocytes detected to obtain a trustworthy, representa- would exclude lymphocytes for not meeting gating crite-
tive lymphocyte subsets. Other flags, such as “High Count ria (even though they are genuine cells). In order to prop-
Rate” are suspected for other hematological disorders and erly evaluate the unique walkaway character of AQUIOS
need further, more elaborated immunophenotyping. In CL, EQA providers are challenged to produce a material
this study, 35% of the samples showed a run notification that is both compatible with this system but also has
(35/100). In 26 out of these 35 samples (74.3%), the results clinical relevance in terms of the cell counts used. To meet
could be accepted without intervention, whereas in nine these requirements the ISO 17043 accredited UK NEQAS
samples out of these 35 samples (25.7%) with notifications Immune Monitoring Alternative Technologies program (31
an action was needed: this included a rerun in two cases international participants as of June 2017) uses material
(5.7%) and an optimization of the gating in seven cases that is sourced from Beckman Coulter and manipulated to
(20%). In 26% of all samples an unnecessary notifica- ensure that the levels of CD4+ T lymphocytes (and other
tion was shown, so most notifications did not necessitate populations) are unique to the EQA exercise but differ-
manual revision. The relative high number of notifications ent from the supplied material. The utilization of a com-
could lead to “revision fatigue” of the users, meaning loss mercial source for the raw material from which the EQA
of concentration of the lab technician during revision samples are produced complies with ISO 17043 and the
since most notification have minimal or no impact. underlying principles of EQA.
One sample showing an “Inter Panel Count Error” This study has limitations. First, there is a selec-
flag, had an incorrect CD3+ gating in the Tetra-2 Panel tion bias as we enriched for samples from pediatric
while there was a correct CD3+ gating in the Tetra-1 patients and posttransplant patients. Second, we com-
Panel. This sample was repeated, but instead of request- pared AQUIOS CL with a dual-platform methodology
ing a Tetra Combo panel, Tetra-1 and Tetra-2 Panels were because in Belgium 82% of the clinical laboratories use
performed separately. Again one CD3+ gating was done this method. This methodology, however, is known to be
incorrectly, but this sample was not flagged as there was less precise than single-platform methods [6] and more
no interpanel comparison. However, there was a run and more abandoned. Nowadays, almost 90% of the UK

Brought to you by | De Gruyter / TCS


Authenticated
Download Date | 8/6/18 9:10 AM
Degandt et al.: Evaluation AQUIOS CL 1287

NEQAS participants of the immune monitoring scheme Author contributions: BP, NB, XB: study design. SD:
use a single platform method. Third, AQUIOS CL was only drafted the manuscript. BP, SJ: patient recruitment. SD,
evaluated for a period of 3 months. Therefore, robustness BP, SJ: data acquisition. SD, BP: statistical analysis. NB,XB
of the device over a longer period of time was not evalu- checked and improved content and English grammar
ated. We noticed a minor mechanical issue at the end of and style of the manuscript. NB, XB commented on and
the evaluation period, which caused run notifications. edited the manuscript, which was read and approved by
The error was caused by a problem in the flow cell which all authors. All the authors have accepted responsibility
resulted in incorrect EV/SS plots. This issue had only a for the entire content of this submitted manuscript and
small, insignificant impact on the results. approved submission.
To the best of our knowledge, this is the first study Research funding: None declared.
that demonstrates an acceptable analytical performance Employment or leadership: None declared.
of AQUIOS CL and good correlation with a dual-platform Honorarium: None declared.
method for the quantification of T (both CD4 and CD8), Competing interests: The funding organization(s) played
B and NK lymphocyte subsets. Gossez et al. [7] focused no role in the study design; in the collection, analysis, and
on absolute CD4 T cell enumeration and its population interpretation of data; in the writing of the report; or in the
mainly consisted of patients with HIV. A recent study decision to submit the report for publication.
of Grossi et al. [8] compared AQUIOS CL to a bead-
based method using 224 routine samples. They focused
mainly on a comparison with a bead-based method and
the influence on turnaround times. Here we evaluated
References
both Tetra-1 and Tetra-2 Panels for enumeration of all
1. Picard C, Al-Herz W, Bousfiha A, Casanova JL, Chatila T, Conley
T, B and NK lymphocytes, partly by comparing it to a ME, et al. Primary immunodeficiency diseases: an update on
dual-platform method, in a pediatric population, in the classification from the International Union of Immunological
a ­posttransplant population and in samples from our Societies Expert Committee for Primary Immunodeficiency 2015. J
routine work load. Clin Immunol 2015;35:696–726.
2. Fleisher TA, Madkaikar M, Rosenzweig SD. Application of flow
In conclusion, AQUIOS CL allows automated immu-
cytometry in the evaluation of primary immunodeficiencies.
nophenotyping of lymphocyte subset quantitation. Ana- Indian J Pediatr 2016;83:444–9.
lytical performance was acceptable and quantitative 3. Maguire O, Tario JD, Jr, Shanahan TC, Wallace PK, Minderman H.
results of the different lymphocyte subsets were compa- Flow cytometry and solid organ transplantation: a perfect match.
rable to results obtained with a dual-platform method; Immunol Invest 2014;43:756–74.
however, interpanel consistency is troublesome. Auto- 4. WHO. Consolidated guidelines on the use of antiretroviral drugs
for treating and preventing HIV infection: recommendations for
matic gating sometimes needed to be manually reviewed.
a public health approach, second edition. 2016. Available at:
AQUIOS CL warns – through run flags and notifications – http://www.who.int/hiv/pub/arv/arv-2016/en/. Accessed: 18 Jul
whenever such a manual revision is needed. As the 2017.
number of notifications was relatively high (about one- 5. Brando B, Barnett D, Janossy G, Mandy F, Autran B, Rothe G, et al.
third of all samples), gating algorithms should be further Cytofluorometric methods for assessing absolute numbers of
cell subsets in blood. European Working Group on Clinical Cell
improved to reduce this number. Overall, the system looks
Analysis. Cytometry 2000;42:327–46.
promising for quantification of lymphocyte subsets in 6. Barnett D, Granger V, Whitby L, Storie I, Reilly JT. Absolute CD4+
routine clinical practice. T-lymphocyte and CD34+ stem cell counts by single-platform flow
cytometry: the way forward. Br J Haematol 1999;106:1059–62.
Acknowledgments: This study was supported by Beck- 7. Gossez M, Malcus C, Demaret J, Frater J, Poitevin-Later F, Mon-
man Coulter through donations of laboratory equipment neret G. Evaluation of a novel automated volumetric flow cytom-
eter for absolute CD4+ T lymphocyte quantitation. Cytometry B
and supplies. This private company had no role in study
Clin Cytom 2017;92:456–64.
design and the collection of data. They did help in inter- 8. Grossi V, Infantino M, Meacci F, Bellio E, Bellio V, Ciotta G,
preting gating results with run notifications. Beckman et al. Comparison of methods and TAT assessment: volumetric
Coulter had no role in the preparation of the manuscript AQUIOS CL and bead-based FACS CANTO II cytometers. Cytom-
or the decision to submit it for publication. We thank etry B Clin Cytom 2017. doi: 10.1002/cyto.b.21513 [Epub ahead
of print].
Professor David Barnett and Liam Whitby (UK National
9. CLSI. User verification of precision and estimation of
External Quality Assessment Scheme [UK NEQAS], Shef- bias; approved guideline – third edition. CLSI document
field) for providing information and data regarding the UK EP15-A3. Carey, RN: Clinical and Laboratory Standards Institute;
NEQAS immune monitoring scheme. 2014.

Brought to you by | De Gruyter / TCS


Authenticated
Download Date | 8/6/18 9:10 AM
1288 Degandt et al.: Evaluation AQUIOS CL

10. Centers for Disease Control and Prevention (CDC). Guidelines for of storage before and after staining in a developing country set-
performing single-platform absolute CD4+ T-cell determinations ting. Indian J Clin Biochem 2004;19:95–9.
with CD45 gating for persons infected with human immunodefi- 14. Zini G. International council for standardization in ­haematology
ciency virus. Morb Mortal Wkly Rep MMWR 2003;52:1–13. (ICSH). Stability of complete blood count parameters with
11. Comans-Bitter WM, de Groot R, van den Beemd R, Neijens HJ, storage: toward defined specifications for different diagnostic
Hop WC, Groeneveld K, et al. Immunophenotyping of blood applications. Int J Lab Hematol 2014;36:111–3.
lymphocytes in childhood. Reference values for lymphocyte 15. Nascimbeni M, Shin EC, Chiriboga L, Kleiner DE, Rehermann
subpopulations. J Pediatr 1997;130:388–93. B. Peripheral CD4(+)CD8(+) T cells are differentiated effector
12. Ekong T, Hill AM, Gompels M, Brown A, Pinching AJ. The effect of memory cells with antiviral functions. Blood 2004;104:478–86.
the temperature and duration of sample storage on the meas-
urement of lymphocyte subpopulations from HIV-1-positive and
control subjects. J Immunol Methods 1992;151:217–25.
13. Jalla S, Sazawal S, Deb S, Black RE, Das SN, Sarkar A, et al. Supplemental Material: The online version of this article offers
Enumeration of lymphocyte subsets using flow cytometry: effect supplementary material (https://doi.org/10.1515/cclm-2017-0638).

Brought to you by | De Gruyter / TCS


Authenticated
Download Date | 8/6/18 9:10 AM

You might also like