You are on page 1of 6

MC Toolkit 34-ST-03-80

10/06
(Rugged & Intrinsically Safe models)
Specification and
Model MCT202 Model Selection Guide
Features
Figure 1 MCT202 Selectable Versions (Ruggedized, Zone 2 and Zone 1
- Configures any Honeywell DE device: Intrinsically-safe)
ST3000 pressure and STT3000
temperature transmitters.
- Configures any HART® device with a
published HART Device Descriptions
(DD), regardless of device
manufacturer.
- Supports easy downloading of any
new vendor’s DD files via a PC
connection.
- Intuitive graphical user interface
provides simple navigation and device
configuration.
- Available in three versions handling
almost any environmental condition:

• Ruggedized version-
Environmentally hardened with no
approvals. Ruggedized
• Zone 2 approvals: Intrinsically-
safe version available with FM
Class I, Div 2 and ATEX Zone 2
approvals.

• Zone 1 approvals: Intrinsically-


safe version available with FM
Class I, Div 1 and ATEX Zone 1
approvals.
- Fully supports all HART 5.X
commands and Universal commands
for HART 6.X and utilizes DD-IDE/SDC
625 technology and “Open” Tools
standard.

Benefits Zone 2
- Supports transmitter diagnostic and
on-line monitoring as well as transmitter
calibration.
- Standardizes both HART and DE field
instrument configuration with one hand-
held device. Configuration of common,
universal and specific device
commands is supported.
- User interface provides detailed
explanations and sub-menus during the
configuration process for ease of
operation for even the casual user.
Zone 1

Honeywell Industrial Measurement and Control, 2500 W. Union Hills Drive, Phoenix, AZ 85027
34-ST-03-80
Page 2

- Supports communication with HART The MC Toolkit Configurator is loaded with two software applications:
devices when the operator is
performing the following functions: 1. The MC Toolkit software configures Honeywell DE and HART devices

• Initial device configuration 2. The SDC 625 software configures non-Honeywell HART devices
The combination of these two applications enables true “Universal” configuration
• Starting a method
of HART devices.
• Changing a configuration
The MCT202 is rated intrinsically safe for Zones 1& 2 for both FM C and ATEX.
parameter

• Determining the actual “state”


or process variable of a
particular device

• Activating a device’s internal


diagnostic functions.
- Avoid obsolescence issues when
vendors publish new DD files

• Download new device


descriptions from the HART
website.

• Acquire the latest DD releases


from the manufacture’s
website including device
descriptions not available from
HART.
- Incorporates data reasonability
checks. Under given conditions, the
MC Toolkit configurator will not accept
configurations beyond device ranges
reducing the likelihood of costly
mistakes.

Description
The Honeywell MC Toolkit Configurator
is a Pocket PC-based engineering and
maintenance tool used to configure,
monitor, diagnose and manage smart
field instruments/devices. Designed
specifically for ease-of-use and
increased productivity, this
multipurpose tool supports both HART
and Honeywell’s Digitally Enhanced
(DE) communication protocols. The
MC Toolkit allows easy access to
instrument configuration database
parameters without the need for prior
training, user manuals or previous
experience. Each MC Toolkit comes
loaded with all registered HART Device
Description (DD) files and may be
supplemented with non-published DD
files at any time – so any HART new
device can be configured as long as a
DD can be downloaded. New DD files
are easily downloaded and installed
using ActiveSync software and the USB
cable provided.
34-ST-03-80
Page 3

Specifications

MCT-202
Pocket PC i.roc 627 (HP iPAQ hx2490 or equivalent)

Processor Intel PXA270 520 MHz

Memory 128 MB ROM and 64 MB SDRAM


1 GM San Disk Flash Memory card

Operating System Microsoft Windows Mobile 5.0

Display 3.5” Transflective TFT color with LED backlight


240 x 320 pixels, 64K color, touch screen

Power Supply Rechargeable Lithium-Ion battery plus new charger options

All AC power options accept voltage inputs (Vac) of 90 - 264. Select the version required based
on the input connection types available. Select using the pictures below.

US Type EU Type AU Type UK Type

Dimensions (6 x 2.875 x 1.5 in) (15.24 x 7.3025 x 3.81 cm)

Weight 200 g (.4409 lbs)

Operating Temperature -10 to +50°C (+14 to +122°F)

Storage Temperature -20 to +60°C (-4 to +140°F)

Operating Humidity 10 to 90% RH

Approximately 51MB (includes all HART registered device descriptions). MC


Toolkit software, including SDC 625 and HART DD files, are loaded on flash
Memory Requirements memory card. The flash memory card in the MCT202 is not removable.
Each MC Toolkit is loaded with a complete set of HART DD files available at the
time of manufacture. If the device vendor has qualified their DD with the HART
HART Device Description Foundation SDC 625 tool, the device description is assured of being compliant with
the MC Toolkit. Additional DDs can be downloaded as new releases or new
products become available through the ActiveSync connection.
European EMC Standards:
EEC Directive 89/336 (EMC)
CE Mark
FM C-US:
Intrinsically Safe – Class I, Div 1 Group A,B,C,D
Approvals Non-incendive – Class I, Div 2 Group A,B,C,D
ATEX:
Intrinsically Safe – Ex II 2G EEx ia IIC T4
Type n – Ex II 3G EEx nL IIC T4
Dust ignition proof – Ex II 2D T99oC IP65; Ex II 3D T55oC IP65

Microsoft® and Windows® are registered trademarks of Microsoft Corporation.


All other product and brand names shown above are trademarks of their respective owners.
34-ST-03-80
Page 4

MCT202 Parts

i.roc 627 (HP iPAQ Pocket PC)


Integrated HART/DE modem
6’ transmitter connector cable with one set of
alligator clips and one set of EZ Clip connectors
CD ROM (MC Toolkit + SDC 625 software)
MC Toolkit user manual
i.roc user manual
i.roc charger LG x 10
2 stylus pens
1 battery installed
1 GB SD card installed

Spare Parts and Accessories

Item Number Description MCT202


50003165 MC Toolkit Upgrade Kit: 256 MB San Disk card loaded with MC Toolkit and
SDC 625 software plus HART DD files
50018046-001 Transmitter connector cable with one set of Alligator clips and one set of EZ Available
clips
50017823-001 Stylus pens, set of two (2) Available
50017828-001 Spare charger LG x 10 (US adaptor) Available
50017828-002 Spare charger LG x 10 (EU adaptor) Available
50017828-003 Spare charger LG x 10 (UK adaptor) Available
50017828-004 Spare charger LG x 10 (AU adaptor) Available
50017822-001 Leather carrying case Available
50017824-001 Hand Strap Available
50018626-001 Docking station with charger adapter and USB cable for computer interface Available
50017827-001 Docking station with USB cable Available
50017825-001 Charger adapter for docking station Available
50018045-001 USB interface kit with charger cable Available
51453286-501 Honeywell CD-ROM software (SDC 625 and MC Toolkit software) Available
34-ST-03-80
Page 5

Model Selection Guide 34-ST-16U-61


34-ST-16U-61
Issue 1
Page 1 of 2

MC Toolkit (Rugged and I.S.) Model Selection Guide


Multiple Communication Configurator
Instructions
Select the key number and the desired Safety Approval
?
code.
Order Accessories/Spare parts as separate line items on
?
purchase order.
Key Number I II III
______ - __ - __ - __
Ruggedized Zone 2 Zone 1
KEY NUMBER
Description Selection
MC Toolkit Configurator MCT202
This MC Toolkit version comes complete with 2 stylus pens, LGx10 charger, integrated HART/DE modem, USB
interface kit w/ charger cable, 4' transmitter connector cable & clips and manuals on a CD-rom.

TABLE I - Safety Approvals


Location Approval
Approval Type Location or Classification Selection
Type Body
None No approval body certifications included ZZ
Class I, Div. 1, Groups A,B,C,D
FM C-US Intrinsically Safe
Zone 1 AEx ia IIC Zone 1
XX
Div 1 Intrinsically Safe Ex II 2G EEx ia IIC T4
ATEX o o
Dust Ignition Proof Ex II 2D T99 C IP65; Ex II 3D T55 C IP65
Class I, Div. 2, Groups A,B,C,D
Zone 2 FM C-US Non-Incendive
AEx nA [L] IIC Zone 2 YY
Div 2
ATEX Type n Ex II 3G EEx nL IIC T4

TABLE II - AC Power Options


All AC power options accept voltage inputs (Vac) of 90 - 264. Select the version required

US Type EU Type UK Type AU Type


US connector type 01
EU connector type 02
UK connector type 03
AU connector type 04

TABLE III - Options


No Options 00
Hard copy of User's Manual UM

Ordering Example: MCT202-YY-01-UM


34-ST-03-80
Page 6

Honeywell Industrial Measurement and Control


2500 W. Union Hills Drive
Phoenix, Arizona 85027 ©Honeywell International Inc.

You might also like