Edwards FloTrac Sensor Setup

1 Open FloTrac sensor 6 Pressurize the Pressure Bag until it reaches


packaging and inspect 300mmHg.
contents. Replace all
caps with non-vented 7 Fast-flush the FloTrac sensor and tap on tubing
caps and ensure that and stopcocks to remove any residual bubbles.
all connections are
tight. 8 Connect the green FloTrac connecting cable to
the green capped connector on the FloTrac
sensor. Then connect the opposite end of the
cable to the connection labeled “FloTrac” on the
back of the Vigileo monitor.

2 Remove the FloTrac


sensor from packaging 9 Connect the bedside
and insert into an monitor’s arterial
Edwards Lifesciences pressure cable to
mounting back-plate the white cable
that is secured on an connector on the
I.V. pole. FloTrac sensor.

3 To de-air and prime I.V.


bag and FloTrac system: 10 Connect tubing to arterial catheter, then aspirate
Invert NaCl I.V. bag and flush system to assure no residual bubbles
(anticoagulation per institution remain.
policy). Spike I.V. bag with fluid
administration set, keeping 11 Level the FloTrac sensor to the phlebostatic axis.
drip chamber upright. While Note: It is important to keep the FloTrac sensor
keeping I.V. bag inverted, level to the phlebostatic axis at all times to ensure
gently squeeze air out of bag accuracy of cardiac output.
with one hand while pulling a. Open the stopcock to atmospheric air.
flush tab with the other hand b. Highlight the upper left Cardiac Output frame
until air is emptied from I.V. using the navigation knob on the Vigileo
bag and drip chamber is monitor. Select Zero Arterial Pressure, then
filled half-way. select and press Zero.
c. Zero the arterial channel on the bedside monitor.
4 Insert I.V. bag into the Pressure Bag and
hang on I.V. pole (do not inflate). 12 Cardiac output will display within 40 seconds and
will update every 20 seconds thereafter.

5 With gravity only (no 13 Inspect arterial pressure trace on bedside


pressure in Pressure Bag), monitoring screen or the waveform confirmation
flush FloTrac sensor screen on the Vigileo monitor.
holding pressure tubing
in upright position as
the column of fluid raises
through the tubing,
pushing air out of the
pressure tubing until the
fluid reaches the end of
the tubing.
Edwards Vigileo Monitor Setup & Zeroing

1 Press the button 7 From the CO


on the front panel to Menu, rotate the
turn the Vigileo navigation knob
monitor ON. The until Zero Arterial
screen will display an Pressure is
opening message, highlighted and
indicating that a then press the
Power-On Self-Test knob. The Zero
(POST) is being Arterial Pressure
performed. screen will appear.

8 Open the FloTrac


sensor to
2 When the POST is atmospheric
complete, patient air. Rotate the
information (gender, navigation knob
age, height, and on the Vigileo
weight) must be monitor to Zero
entered before cardiac and press the
output monitoring knob. Select
can occur. Return to exit
screen. Close the
3 Use the navigation knob to select and enter values. FloTrac sensor to
Press Continue to confirm selection and open the atmospheric air.
Home screen.
9 Cardiac output
will be displayed
4 Connect the FloTrac
within 40
connecting cable to the
seconds after
FloTrac cable connector at
arterial pressure
the back of the Vigileo
is registered by
monitor. Align the arrows
the FloTrac sensor.
at the top of the cable
connector on the monitor
to the arrow on the
FloTrac connecting cable.

5 Connect the other end of the FloTrac cable to the


green capped FloTrac sensor.

6 Rotate navigation Rx only. See instructions for use for full prescribing information.
knob until the CO Edwards Lifesciences devices placed on the European market
meeting the essential requirements referred to in Article 3 of the
frame is outlined Medical Device Directive 93/42/EEC bear the CE marking of
in yellow and then conformity.
press the knob to Edwards, the stylized E logo, FloTrac and Vigileo are trademarks
of Edwards Lifesciences Corporation. Edwards Lifesciences is a
open CO menu. trademark of Edwards Lifesciences Corporation and is registered
in the U.S. Patent and Trademark Office.

© 2005 Edwards Lifesciences LLC


All rights reserved. AR00883

Edwards Lifesciences LLC · One Edwards Way · Irvine, CA 92614 USA · 949.250.2500 · 800.424.3278 · www.edwards.com
Edwards Lifesciences (Canada) Inc. · 1290 Central Pkwy West, Suite 300 · Mississauga, Ontario · Canada L5C 4R3 · Phone 905.566.4220 · 800.268.3993
Edwards Lifesciences S.A. · Ch. du Glapin 6 · 1162 Saint-Prex · Switzerland · Phone 41.21.823.4300
Edwards Lifesciences Japan · 2-8 Rokubancho · Chiyoda-ku, Tokyo 102-0085 · Japan · Phone 81.3.5213.5700

You might also like