Professional Documents
Culture Documents
Cryostat
Instruction Manual
Leica CM3050 S Cryostat
English V1.1 02/2000
Always keep this manual near the instrument.
Read carefully prior to operating the instrument.
Important note
Serial No.: .............. ......................................................
The information contained in this document rep-
Year of Manufacture: ............................................... resents state-of-the-art technology as well as
the current state of knowledge.
Country of Origin: Federal Republic of Germany
Leica will not assume liability for errors that
might be contained in this manual, nor for acci-
dental damage or damage arising from the de-
livery, performance or use of this manual.
Therefore, no claims can be made based on the
text and illustrations in this instruction manual.
Leica Microsystems Nussloch GmbH reserves
the right to change technical specifications
without notice since each of our products is
subject to a policy of continuous improvement.
This document is protected under copyright
laws. Leica Microsystems Nussloch GmbH
retains all rights related to this documentation.
Any reproduction of text and illustrations - or
any parts thereof - in form of printing, photo-
copying, microfiches, or other methods, includ-
ing electronic systems, requires the prior writ-
ten permission of Leica Microsystems Nussloch
GmbH.
Issued by:
The serial number and the year of manufacture
Leica Microsystems Nussloch GmbH are specified on the nameplate at the back of
Heidelberger Str. 17 - 19 the instrument.
D-69226 Nussloch
Germany
Phone: +49 6224 143-0
Fax: +49 6224 143-200
Internet: http://www.hbu.de © Leica Microsystems Nussloch GmbH
3. Installation................................................................................................................................... 16
3.1 Site requirements .................................................................................................................................... 16
3.1.1 General site requirements .................................................................................................................... 16
3.1.2 Electrical connections ........................................................................................................................... 16
3.2 Unpacking and installation .................................................................................................................... 16
3.3 Standard delivery .................................................................................................................................... 17
5. Operation ..................................................................................................................................... 24
5.1 Operating the instrument for the first time .......................................................................................... 24
5.2 Switching on/fuses ................................................................................................................................. 25
5.3 Control panel 1 ......................................................................................................................................... 26
5.3.1 Key functions in control panel 1 ........................................................................................................... 26
5.3.2 Display functions in control panel 1 .................................................................................................... 26
5.3.3 Display indications when switching on .............................................................................................. 27
5.3.4 Menu functions: ...................................................................................................................................... 28
5.3.5 Status display .......................................................................................................................................... 31
5.3.6 Indication of refrigeration state, section thickness setting and counters .................................... 32
5.4 Control panel 2 ......................................................................................................................................... 33
5.4.1 Selecting a sectioning mode / .............................................................................................................. 34
5.4.2 Setting the sectioning window ............................................................................................................. 35
5.4.3 Selecting sectioning speed .................................................................................................................. 36
5.4.4 Emergency stop ...................................................................................................................................... 36
5.4.5 Coarse feed keys .................................................................................................................................... 37
5.4.6 Trimming and sectioning functions ..................................................................................................... 38
5.4.7 Setting the section thickness ............................................................................................................... 38
5.5 Daily operation of the instrument ......................................................................................................... 39
5.5.1 Freezing specimens onto specimen .................................................................................................... 39
5.5.2 Freezing specimens onto specimen .................................................................................................... 40
5.5.3 Optional accessories for freezing ........................................................................................................ 40
5.5.4 Inserting specimen discs into the ........................................................................................................ 41
Appendix............................................................................................................................................... 59
The chapters of the Leica CM3050 S cryostat 1.1 Symbols used in this manual and
instruction manual: their meaning
Chapter 1 Structure of the manual:
Table of contents Warnings
Important information on this appear in a grey box and are
manual marked by a warning triangle .
Chapter 2 Safety
Read this chapter before
operating the instrument!
Notes
Chapter 3 Installation i.e. important user information ap-
Unpacking and installation pears in a grey box and is marked
Standard delivery, information by an information symbol.
on accessories
Chapter 4 Instrument properties
Overview Figures in brackets refer to item
Technical data (5)
numbers in drawings or to the
Chapter 5 Operation (Fig.5) drawings themselves.
Controls
Setup, daily use and shut
down
Chapter 6 Trouble shooting Instrument type:
and applications tips All information given in this instruction manual
Operating errors applies only to the instrument type indicated on
Display error messages the title page.
Temperature reference chart A name plate, indicating the instrument serial
number, is attached to the back of the instru-
Chapter 7 Disinfection, cleaning ment.
and maintenance
Required information for all inquiries:
Chapter 8 Warranty and service
For any inquiries please specify:
Appendix Electrohydraulic height adjustment Instrument type
Ordering information accessories Serial number
EC Declaration of Conformity
Alphabetical index
This instrument has been built and tested in ac- The instrument is equipped with the following
cordance with the following safety regulations safety devices:
on electrical measuring, control, regulating and
laboratory devices:
Handwheel lock
DIN EN 292 Handwheel grip centering
DIN EN 61010-1 Emergency stop function
(instruments with sectioning motor only)
EN 50082-1
Knife holder equipped with knife guard
EN 55011
IEC 1000-4
The safety devices installed by the manufactu-
rer of the instrument only constitute the basis of
as well as according to the international quality
accident prevention.
standard
Mainly responsible for accident-free operation
is above all the institution which owns the in-
DIN ISO 9001
strument and, in addition, the designated per-
sonnel who operate, service or repair the
instrument.
In order to maintain this condition and to ensure
safe operation, the operator must observe the
instructions and warnings contained in this in-
struction manual.
C T - 3 0 ° C O T - 3 5 ° C
The handwheel can be locked in 2 positions: L O C K E D
with the grip in the uppermost position (left), Unlocking:
with the grip in the lowest position (right). To unlock, push locking pin (2) to the left into
position (2a).
Display indication LOCKED disappears.
2
Instruments with sectioning motor:
The sectioning motor can now be activated
again.
Footswitch
1
Function
The footswitch performs the same task as the
RUN/STOP and RUN/ENABLE keys (activating/ 3.5 Prior to switching on the instrument
deactivating motorized sectioning / trimming).
In addition, the footswitch can be used to acti- After transport, observe a waiting pe-
vate the emergency stop function. riod of at least 4 hours before turning
the instrument on! - See also safety
instructions in chapter 2.3.1 Trans-
Models with footswitch: port.
All instruments with sectioning motor.
4.1 Overview
1 Cryostat CM3050 S
2 Control panel 1
3 Control panel 2
4 4 Storage shelf, left
2 5 5 Storage shelf, right
3
6 Rubber mat
7 Mains switch /
Automatic cutout
6 for instrument,
Automatic cutout
for sectioning motor,
Footswitch port
8 Quick-freeze shelf
9 Stationary
heat extractor
10 Mobile heat extractor
11 Specimen disc
9 12 Thermoblock (optional)
10 13 Section waste tray
14 Brush shelf
15 Specimen head w/o
specimen cooling
16 Specimen head with
specimen cooling
(Option)
17 90° Prism
1
(instruments with
11
8 specimen cooling
7 only)
18 Knife holder base
19 Knife holder CE
20 Knife holder CN
12 21 Knife holder CS
22 22 Footswitch with
protective guard
15
16
11
11
13
17
14
21
20
19
18
Operating temperature range (ambient temperature): + 18°C to + 40°C. All specifications related to tempe-
rature are valid only up to an ambient temperature of + 22°C and relative air humidity of less than 60%!
Refrigeration system 50 Hz 60 Hz
Cryochamber
Temperature selection range 0° C to -40°C (+ 2 K / - 0 K) 0° C to -40°C (+2 K / - 0 K)
Defrosting Automatic hot-gas defrost cycle, Automatic hot-gas defrost cycle,
programmable in 15-min increments; programmable in 15-min. increments;
1 automatic defrost cycle per 24 hours. 1 automatic defrost cycle per 24 hours.
Duration: 6 - 12 minutes Duration: 6 - 12 minutes
Manual defrost cycle Manual defrost cycle
Refrigeration capacity 690 W 690 W
Safety factor 3 3
Refrigerant 280 g ±5g refrigerant R 404A * 280 g ±5g refrigerant R 404A*
Compressor oil 0.6 l EMKARATE RL-22S, ICI * 0.6 l EMKARATE RL-22S, ICI *
Quick-freeze shelf
Maximum temperature - 43°C (+ 0 K / - 2 K) - 43°C (+ 0 K / - 2 K)
Number of quick-freeze stations 10 10
*) Refrigerant and compressor oil to be replaced only by trained and authorized service
personnel!
Cryocabinet
Dimensions 1 according to IEC-1010, UL 3101
Width (w/o handwheel) 800 mm
2 according to CECOMAF:
Width (including handwheel) 882 mm
Depth 766 mm Liquid temperature 45°C
Height (arm rest level) 840 mm Evaporation temperature -25°C
Overall height 1040 mm
Weight
(incl. microtome, w/o specimen cooling) ca. 180 kg
Prior to setting up the instrument, make sure to read the instruction manual
carefully and in full.
When setting up the instrument, please follow the individual chapters in the
following order:
Switching on
To switch on the instrument, place switch
(1) in upper lock-in position.
5.3.1 Key functions in control panel 1 5.3.2 Display functions in control panel 1
ON/OFF key for cryochamber illumination. The display in control panel 1 combines several
different functions:
1. When switching the instrument on:
To select the menu items for setting /
A series of standard indications is displayed,
changing instrument parameters. during which two basic parameters can be se-
lected (with the arrow keys):
Language (of display indications) and speed
In the status indication, pressing the ar- level of sectioning motor.
row keys displays the individual current See chapter 5.3.3 Display indications when
parameters. switching on.
In the menu, pressing the arrow keys
2. Menu function (activated via menu key):
changes set values.
Via the menu key a number of parameters can
As soon as a value is displayed, it is au- be selected, then be activated /deactivated via
tomatically stored. the lock/unlock key and/or changed via the ar-
row keys.
Activates the manual defrost cycle (cry- See chapter 5.3.4 Menu functions: Setting param-
ochamber). This key is functional only eters of refrigeration, time and preset counter.
when one of the menu items SET TEMP
3. Status indication function:
CT ... °C or SET TEMP OT ... °C (last
one only applies to instruments with The actual status of selected parameters is in-
dicated in the display screen.
specimen cooling) has been selected Some of those parameters are displayed auto-
first. matically (e.g. sectioning stroke phase, coarse
feed position, handwheel locked).
Lock / unlock key: Others can be displayed by pressing the arrow
- Locks the display to protect against keys; e.g. selected section thickness, preset
accidental changes of set parameters. counter status etc.
- Activates / deactivates a number of See chapter 5.3.5 Status indication and chap-
functions. ter 5.3.6 Displaying state of refrigeration, sec-
tion thickness and counters.
C M3 0 5 0 S V s . X X . X X
Software version ---
E NG L I S H
Selected display language Select language via
keys
C U T T I N G MO T O R
Sectioning motor speed S L OW
level slow / (fast) Select speed level via
keys
yes Temperature prior to
no
switching off < 0 °C
Temperature upon
switching on ³ 0 °C no
C T . . . .°C O T . . . . ° C *)
P OW E R F A I L U R E
yes
Acknowledge indication by
pressing once briefly.
C T . . . .°C O T . . . . ° C *)
D R Y M I C R O T OME C T . . . .°C O T . . . . ° C *)
1 2 : 0 0
Standard indication locked is displayed:
If this indication is displayed: Do not switch on (Display background dark)
the instrument! To unlock: press until the display back-
Prior to switching on, both microtome and ac- ground lights up:
cessories must be completely dry!
Remove microtome/accessories, disinfect if
applicable, then dry thoroughly and reinstall. C T . . . .°C O T . . . . ° C
(--> See chapters 7.1 - 7.5). n n H OM E dd
*) Indication in instruments with specimen cooling. Instruments without specimen cooling: only CT (centered) is displayed.
4. Quick-freezing at maximum specimen The automatic defrost cycle takes place once
temperature* (instruments with specimen every 24 hours at the selected defrost time.
cooling only)
7. Duration of defrost cycle:
Press key until menu item Fast object Press key until menu item Defrost period
freezing is displayed: is displayed:
F A S T O B J F R E E Z I N
O F F
D E F R O S T P E R I O D
or C HAMB E R : 0 8 M I N
F A S T O B J F R E E Z I N Use defrost keys ( or ) to adjust de-
O N frost time. (Only if necessary - see below).
Press lock/unlock key to activate or de-
activate (ON/OFF). Important information for setting the
defrost parameters
Upper line 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
Lower line 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
Section thickness
○
Sectioning window [
○
○
○ ○
○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○
A
Functional area A:
Sectioning modes
Sectioning window
Sectioning speed
Emergency stop function
B B
Functional area B: Functional area B:
Coarse feed Coarse feed
Trimming Trimming
Section thickness selection Section thickness selection
C
Functional area C:
Start / Stop sectioning
5.4.6 Trimming and sectioning functions 5.4.7 Setting the section thickness
Switching between trimming
Activate sectioning mode ( off).
and sectioning
C T . . . .°C O T . . . . ° C
Press this key to alternate between the µ 0 1 0 . 0
two functions.
Press or to select section thickness.
illuminated: Pressing one of these keys once makes the
Trimming mode is active display in control panel 1 switch to section
thickness indication (thickness value not yet
off: being altered). Pressing or further
Sectioning mode is active. starts to change the actual section thick-
ness setting.
Selecting trimming thickness
Activate trimming mode
Selectable section thickness settings:
Select trimming thickness (= sectioning
thickness at which trimming is carried out) 0 - 2 µm in 0.5-µm steps,
by pressing or . 2 - 10 µm in 1-µm steps,
10 - 20 µm in 2-µm steps,
The selected thickness is displayed in the cor-
responding LED : 20 - 60 µm in 5-µm steps,
60 - 100 µm in 10-µm steps,
Motorized trimming 100 - 300 µm in 50-µm steps.
5.5 Daily operation of the instrument 5.5.1 Freezing specimens onto specimen
Select chamber temperature and, in instru- discs with quick-freeze shelf and
ments with specimen cooling, specimen stationary heat extractor
temperature according to the type of speci-
men to be sectioned (see also chapter 6.3 Size the specimen.
Reference chart for temperature settings).
Apply sufficient amount of embedding medi-
Do not continue with step 5.5.1 before the
actual chamber / (specimen) temperature um onto ambient temperature specimen
corresponds to the set value. This may take disc.
several hours (e.g. when operating the in- Inside the cryochamber, place the specimen
strument for the first time or when the instru- onto the disc and orient.
ment had been switched off).
Place specimen disc with specimen into one
Prior to operating the instrument , of the 10 openings in the quick-freeze shelf
please read and observe the fol- (see below) and freeze.
lowing safety instructions:
Knife guard /
3
Handwheel lock
(Chapter 2.3.5)
Motorized sectioning
(Chapter 2.3.6) The freezing process can be sped up further by
Defrosting using the stationary heat extractor (1). Next to
/ handling frozen tissue the low temperature stabilizer (2) (Parking
(Chapter 2.3.7) station), 3 specimen discs (3) can be placed
within its pivoting range.
Frozen instrument parts
and accessories
(Chapter 2.3.8) Place the heat extractor (1) directly onto the
specimen surface. After approx. 30 seconds
Infectious / radioactive material the specimen is completely frozen.
(Chapter 2.3.9)
5.5.2 Freezing specimens onto specimen The specimen head is now automatically cooled
discs with 90 ° prism down to the lowest temperature (-50 °C).
(Instruments w/specimen cooling only) Once the specimen is frozen, remove the
In instruments with specimen cooling, the 90 ° disc from the prism and store it in the quick-
prism, which is inserted into the specimen head, freeze shelf until inserting it into the speci-
is an additional option to freeze specimens at
maximum temperature . men head.
Press again to activate menu item Fast
Object Freezing and deactivate the function
4
by pressing once again (see chapter
5.3.4 Menu functions item 4).
The specimen head temperature now automati-
cally returns to the previously selected set val-
ue.
Loosen screw (3) at specimen head and re-
move prism (1).
Installing the 90 ° prism
Store prism inside the cryochamber.
Loosen screw (3) at specimen head.
Insert pin (1) of prism completely into recep-
tacle (2) in the specimen head. 5.5.3 Optional accessories for freezing
specimens onto specimen discs
Important: the entire rear surface of the
prism must fit tightly against the specimen and for removing frozen specimens
head surface!
Retighten screw (3) at specimen head. Mobile heat extractor
The prism now forms a horizontal surface for In addition to the stationary heat extractor (part
quick specimen freezing. of standard delivery), a mobile heat extractor is
available as an optional accessory - see Ap-
Freezing specimens with the 90 ° prism pendix Ordering information - accessories ):
Apply sufficient amount of embedding medi-
um onto ambient temperature specimen disc,
add specimen and orient.
Insert specimen disc with specimen into the
receptacle (4) of the prism.
Press to activate menu item Fast Object Store heat extractor in chamber to precool.
Freezing and activate function by pressing Place onto specimen in quick-freeze shelf.
Remove as soon as specimen is frozen.
(see chapter 5.3.4 Menu functions item 4).
Do not store the thermoblock inside If a knife has already been inserted
the cryochamber - store at room into the knife holder, cover knife
temperature and only place into the edge with knife guard!
chamber for removing specimens
from discs.
Place cap (9) on one of the two ends so that Loosen clamping screw (3).
the correct size opening (7) for the disc to be
inserted is exposed. Insert specimen disc in opening (2) (rear
Insert pin (6) of specimen disc (5) into the surface of specimen disc must fit tightly
opening (7). against surface of specimen head).
After about 20 seconds remove the frozen Tighten clamping screw (3).
specimen from the specimen disc (use
tweezers).
If cap (9) fits too loosely, readjust with 5.5.5 Specimen orientation
screw (10). Do not tighten screw too much!
Once the specimen has been detached from Open clamping lever (4) --> (4a).
the disc, remove the thermoblock from the Orient specimen via orienting screws (1)
cryochamber. and (5).
Retighten clamping lever (4).
C T . . . .°C O T . . . . ° C
1 2 : 0 0
5.7.2 Manual chamber defrost cycle 5.7.3 Defrosting the specimen head
(instruments with specimen cooling
In addition to the automatic defrost cycle, a
manual chamber defrost cycle can be activat-
only)
ed when necessary. To activate specimen head defrosting:
Press until menu item Set chamber Press until menu item Set specimen
temperature is displayed:
temperature is displayed:
S E T T EMP C T - 3 0 ° C S E T T EMP O T - 3 5 ° C
d - dd
Press . A brief sound signals that the Press . A brief sound signals that the
manual chamber defrost cycle has started:
manual chamber defrost cycle has started:
S E T T EMP O T - 3 5 ° C
S E T T EMP C T - 3 0 ° C
d - dd
We recommend a maintenance
To acknowledge the message, press for contract once the warranty period
about 5 seconds.
has expired!
2. Coarse feed Contact your local Leica sales
Coarse feed malfunction has occurred. office for details!
C T - 3 0 ° C O T - 3 5 ° C
If one of the error messages listed
? ? ? ?
under 2, 3 and 4 is displayed, please
This message is also displayed, if the coarse contact Leica Technical Service.
feed plug has not been reconnected properly
after reinstallation of the microtome.
6.1.1 Other error indications
3. Temperature indications
(red ) in control panel 2 is illuminated:
C T - # # ° C O T - 3 5 ° C Emergency stop button accidentally activat-
or ed.
--> Unlock.
C T - 3 0 ° C O T - # # ° C
Footswitch not properly connected.
--> Check connection - connect properly,
If the # symbol is displayed instead of chamber if applicable.
or specimen temperature indication, the corre-
sponding temperature sensor is broken or a
short circuit has occurred.
Sections not properly flattened - Knife and/or anti-roll plate dirty. - Clean with dry cloth or brush.
despite correct temperature - Edge of anti-roll plate dam-
and correctly aligned anti-roll aged. - Replace anti-roll plate.
plate - Blunt knife.
- Use different part of knife or
replace knife.
Scraping noise during section- - Anti-roll plate protrudes too far - Readjust correctly.
ing stroke and specimen return beyond the knife edge and is
stroke. scraping against the specimen.
Chatter marks form during - Specimen insufficiently frozen - Refreeze specimen onto disc.
sectioning. onto specimen disc.
- Specimen disc not clamped - Check disc clamping.
tightly enough.
- Specimen holder ball joint not - Check ball joint clamping.
clamped.
- Knife not clamped tightly - Check knife clamping.
enough.
- Selected section thickness too - Refreeze specimen onto disc.
thick - specimen has come off
the specimen disc.
- Specimen very hard and inho- - Increase section thickness; if
mogeneous. possible, reduce specimen
surface.
- Blunt knife. - Use different part of knife or
replace knife.
- Knife profile inappropriate for - Use knife with a different pro-
the type of specimen to be file.
sectioned.
- Wrong clearance angle. - Readjust clearance angle set-
ting.
Condensation forms on anti-roll - Brush, forceps, cloth or other - Accessories and cleaning
plate and knife during cleaning. cleaning item too warm. material to be used pre-cooled
only. - Store on storage shelves
inside the cryochamber.
Tissue sticks to the anti-roll - Anti-roll plate too warm or - Cool or readjust anti-roll plate.
plate. incorrectly adjusted.
- Static electricity. - Eliminate cause.
- Fat on corner or edge of anti- - Remove fat (alcohol).
roll plate.
- Knife rusty. - Remove rust.
Inconsistent or insufficient - Microtome was not completely - Remove microtome from cham-
specimen feed. dry when refrigeration was ber and dry - possibly with hair-
switched on; ice build-up inside dryer (hairdryer only to be used
the micrometer mechanism. outside the cryochamber!) or at
ambient temperature.
- Microtome defective. - Call Technical Service.
Specimen disc cannot be - Specimen disc stuck to quick- - Apply strong alcohol to contact
removed. freeze shelf or specimen head point or heat the specimen
due to moisture on underside head.
of specimen disc.
Cryostat inoperational. - Mains plug not properly con- - Check mains plug connection.
nected.
- Automatic cut-out has been - Reconnect automatic cut-out
triggered. (switch instrument back on).
No or insufficient refrigerating - Stopper not placed properly - Close drain with stopper.
performance. into drain in bottom of cryo-
chamber.
- Compressor defective. - Call Technical Service.
- Leak in cooling system. - Call Technical Service.
Tissue 10 - 15 15 - 25 25 - 50
Adrenal d
Bone marrow d
Brain d
Bladder d
Breast - fatty d
Breast - little fat d
Cartilage d d
Cervical d
Fatty d
Heart and vascular d
Intestinal d
Kidney d
Laryngeal d
Lip d d
Liver d
Lung d
Lymphoid d
Muscular d
Nose d
Pancreatic d
Prostate d
Ovarian d
Rectal d
Skin with fat d
Skin without fat d
Spleenal or bloody tissue d d
Testicular d d
Thyroid d
Tongue d
Uterus curettage d
Please note: The recommended temperature settings are based on long-term experience. However, they can only be
guidelines for general orientation, as each individual specimen may require particular adjustments.
7.5 Removing / reinstalling the microtome 7.5.2 Disassembling the specimen head
(instruments w/specimen cooling only)
7.5.1 Removing the microtome
Use Allen key size 5 to remove screw (2)
When removing the microtome from the center of the specimen head (1).
make sure to observe the following
safety instructions:
Removing / reinstalling
the microtome
(Chapter 2.3.11)
8.1 Warranty
Leica Microsystems Nussloch GmbH guarantees that the delivered product
was subjected to stringent quality control to ensure that the instrument com-
plies with its technical specification.
The warranty conditions of your local Leica sales company or dealer apply.
Only authorized Leica technical service engineers may carry out repairs on
the instrument.
Any intervention on / modification of the instrument and any use of accesso-
ries other than those delivered by Leica is subject to prior approval by Leica,
unless explicitly specified otherwise in this manual.
Please note that any spare parts and accessories, which are not supplied by
Leica, have neither been inspected nor have they been approved by Leica.
Therefore, installation or use of such parts and/or accessories may impair the
design features of the instrument.
Warranty claims can be made only if the instrument has been operated ac-
cording to its designated use and according to the instructions given in this
manual.
Damage resulting from improper use of the product and/or faulty operation
invalidates the warranty.
Leica will not assume liability for any such damage.
Electrohydraulic height adjustment device for Never place your feet underneath
cryostat series CM3050 S, complete outfit, con- the U-frame when lowering the
sisting of a U-shaped frame on four height-ad- height adjustment device! Risk of
justable feet and a two-key control unit. injury!
The height adjustment device is available in 2 Only authorized Leica technical ser-
voltages: vice engineers may carry out re-
pairs on both the Leica CM3050 S
1. 230 240 V / 50 Hz order no. 0443 26147 and the height adjustment device.
This applies especially to all work
2. 230 240 V / 50 Hz order no. 0443 26147 carried out on the hydraulic system
of the height adjustment device.
1.4 Installation
Open the shipping box and remove all packing material including the Sty-
rofoam pieces.
Take the height adjustment device out of the box.
At the place of installation, place the U-frame on the floor with the open
side facing yourself. Adjust the U-frame parallel to the wall, maintaining a
distance of approx. 10 cm between wall and U-frame.
Push the (ab) button to lower the height adjustment device to the lowest
possible position.
When placing the instrument onto Plug the height adjustment device into the
the U-frame, make sure the power mains socket. The unit is now ready for use.
cord does not get caught and thus
crushed between U-frame and cry-
ostat housing.
2.2 Accessories
90° prism ........................................................................................................................................................... 0443 25949
Mobile heat extractor ..................................................................................................................................... 0443 26836
Thermoblock .................................................................................................................................................... 0398 18542
Glass anti-roll guide, complete assy., for knife holder CE ....................................................................... 0419 30402
Glass anti-roll guide upgrade kit,
for retrofitting knife holders CN, CS (for C-profile knives) ...................................................................... 0419 30403
2.4 Consumables
OCT embedding compound for freezing tissue specimens onto specimen discs,
bottle of 125 ml ................................................................................................................................................. 0201 08926
Cryostat oil No. 407, bottle of 100 ml ............................................................................................................ 0336 06099
EC Declaration of Conformity
We herewith declare, in exclusive responsibility, that the
A E
Accessories 17, 40, 41, 60 EC Declaration of Conformity 59
Anti-roll guide (glass) 60 Electrical connections 12, 16, 19
Automatic chamber defrost cycle 44 Emergency stop function 9, 11, 36
Automatic cutout for sectioning motor 20 Exchanging the fluorscent light lamp 15, 56
Auxiliary clamping device for knife holder CS 60
Available models of the CM3050 S 17
F
Finishing work 43
B Fluorescent light lamp 15, 56
Brush shelf 20 Foot switch - use for emergency stop 11
Footswitch 11, 19, 20, 34, 36, 46, 50
C Footswitch port 20
Freezing specimens onto specimen discs 39, 40
CE Mark (declaration of conformity) 59 Frozen instrument parts / accessories 13
Chamber defrost cycle (automatic) 44 Frozen specimens 45
Chamber defrost cycle (manual) 45 Fuses 25
Cleaning 53
CM3050 S - overview 20 G
Coarse feed functions 37 Glass anti-roll guide 60
Connecting the footswitch 19
Consumables 60 H
Continuous stroke 34
Handling microtome knives / blades 13
Control panel 1 20, 26, 27, 28, 29, 30, 31, 32
Handwheel grip centering 9, 10
Control panel 2 20, 33, 34, 35, 36, 37, 38
Handwheel grip centering (motorized sectioning) 13
cryochamber illumination - ON/OFF key 26
Handwheel lock 9, 10, 13, 31
Cryochamber temperature 29
cryohamber illumination (exchange) 56 I
cryohamber illumination (safety instructions) 15
Cryostat CM3050 S (overview) 20 Infectious / radioactive material 14
Cryostat oil No. 407 60 Inserting specimen discs into the specimen head 41
Installing / inserting accessories 18
D Installing the footswitch 19
Installing the handwheel 18
Daily operation of the instrument 39 Instrument nameplate 3, 7
Daily routine steps when finishing work 43 Integrated safety devices 9, 10, 11
Defrost key 26 Intermittent stroke 34
Defrosting 13, 44, 45
Defrosting / Handling frozen tissue 13 K
Defrosting the specimen head 45
Knife guard 9, 11, 13, 60
Designated use 8
Knife guard / handwheel lock 13
Disassembling the specimen head 54
Knife holder CN 20
Disinfection and cleaning 14, 52, 53
Knife holder CS 20
Display message Dry microtome 15
Knife holder base for knife holders CN, CE and CS 20, 60
Displayed error messages 46
Knife holder CE 20
Displaying parameters with arrow keys 32
Dry microtome 15, 27 Knife holders CN, CE, CS 60
Duration of defrost cycle (setting) 30 Knife rest with section collecting tray for knife 60
Knife support bar for knife holder CS 60
Knives / blades. Siehe Handling microtome knives / blades