SERVICE DEPARTMENT
Institute
Part Description
1. Test instrument for Earth Continuity and Insulation Resistance. If instrument fails, do not
proceed with service until the cause of the failure has been established and the fault
corrected.
Keyboard
1. Remove and check specimen clamp and knife-holder(s). Clean, adjust and lubricate (use
P13880 Low Temperature Grease) where necessary.
1. Check front and rear rubber membranes for damage. Replace if necessary.
4. Check the lead-screw and trunnion nut for wear and excessive play. Clean, adjust and
lubricate using microtome oil as required.
9. Check crank arm coupling for excessive wear. Adjust or replace as necessary.
Microtome (Electronic)
1. Check front and rear rubber membranes for damage. Replace if necessary.
4. Check the lead-screw and trunnion nut for wear and excessive play. Clean, adjust and
lubricate using microtome oil as required.
5. Check correct operation of Advance Alarm. Check advance jam detection. Adjust as
necessary.
6. Check crank arm coupling for excessive wear. Adjust or replace as necessary.
12. Check all microtome fixing points for tightness for screws.
Refrigeration
General