Professional Documents
Culture Documents
Aesculap Orthopaedics
Aesculap® Columbus® Revision
The innovative Revision knee system
2
Preoperative planning 4
Tibial resection 12
3
Aesculap® Columbus® Revision
Preoperative planning
4
▲ ▲ 2
▲
▲
5
▲ 1
3 2
4
▲ 3
1
4
▲
▲
▲
For implantation of a Columbus® CR femur with CR/DD/UC PE Evaluation of tibial Evaluation of stem
gliding surface and Columbus® Revision tibia plateau, the requested rotation (intra operatively) diameter/-length (5)
primary trial femur respects the femur for the operated side Evaluation of femoral
and the drill with stop for the femur pegs have to be ordered rotation (intra operatively)
additionally.
In order to be able to perform proper ligament balancing with
the distractor, the soft tissue management set NQ414 has to be
ordered.
Please ensure that the amount of received implants and instruments
sent with an individual shipment via Aesculap Loan Service are
complete and the ones your patient needs for surgery, based on your
pre-operative planning.
5
Aesculap® Columbus® Revision
Brief instructions
1 4
Set ant. reference mark and remove implants / Ant. Referenzmarkierung Gap balancing by distractor / Messung der Bandspannung mit Distraktor
anbringen und Implantate entfernen
2 5
Resection of proximal tibia / Proximale Tibiaresektion Tibia box preparation by drill / Tibiabox aufbohren
Tibia box broach preparation. Define femur size / Tibiabox mit Kompressions-
meissel präparieren. Femurgrösse bestimmen
3
Gap balancing by spacer. Tibia and trial stem selection / Bandspannungsmessung
mit Distanzblöcken. Auswahl von Tibiaplateau und Probeschaft
option
6
6 9
Resection of distal femur / Distale Femurresektion Femoral box preparation by rasp. Trial box impaction / Femurbox aufraspeln und
Probebox einführen
ML
7
Resection of femoral APC geometry / APC Femurresektion
10
Fix femoral box and stem / Femurbox und -schaft fixieren
8
Resection of femoral box geometry / Femurbox Resektion
11
Adapt trial values to definitive implants / Werte der Probeimplantate auf die
def. Implantate übertragen
7
Operative steps – Tibia First
OPTION: Applying reference marks
1 2
Assemble distal femur contact plate NQ799R and joint Remove all primary implants.
line positioner NQ798R. Set anterior mark where primary
▲
Please note:
femoral implant ends proximal. Keep value of joint line
Ensure with cross check of implant availability matrix
positioner by tightening fixation screw.
(see annex) that you will have the implants you need from
preoperative planning.
8
Preparation of the tibia
2/3
1/3
9
Operative steps – Tibia First
Tibial resection variants
4 5
1. Version:
Use stylus NE425R to determine cut depth. To see where
saw blade ends, use check device NM350R.
2. Version:
Ensure that posterior nose of alignment system (N) is at
N the same level as the reamer mark. Connect tibia cutting
guide with intramedullary connection rod. Press button
Attach alignment system on reamer until posterior on opposite side (see small image) of intramedullary nut
nose (N) is at same level like reamer mark. Tighten fixation to select cut depth. See scale on the intramedullary
screw. Connect tibia cutting guide with intramedullary connection rod device.
connection rod.
3. Version:
Attatch complete alignment system inclusive cutting
Option:
guide on reamer. Use check device in cutting slot and
Check axis with control rods NE331R and NP471R. go down until contact with tibia plateau. Tighten the
fixation screw. Select cut depth by intramedullary
connection rod device.
B
A
D
C
A: NQ648R Intramed. sliding rail, B: NQ649R Intramed. nut, C: NQ650R/NQ651R Asymm. cutting guide (Left/Right),
D: NQ647R Intramedullary connection rod
10
6 7
Fix tibia cutting guide with two parallel headless and one Release rod after turning
convergent head pin in proper rotational position. toggle counter-clockwise
to 11 o`clock position
Remove alignment system and reamer.
‘Unlocked’.
11
Operative steps – Tibia First
Tibial resection
9 10 11
t = 8.5 mm
12
Determining the size of the tibial components
12 13
1. Option: Option:
Insert reamer again until required depth mark. Select trial Select trial tibia hemi spacer of correct size below corre-
tibia plateau with best fit in ML and AP. Define position sponding side of trial tibia plateau.
with chimney NQ677R guided by the reamer. Fix chimney
by using anterior excentric screw – note value. Determine
best rotational position.
Option:
Check axis with rod inserted in the connected handle.
The tibial offset is:
T0/T0+ = ± 4 mm
A
T1-T5 = ± 6 mm B
A
14
▲
Please note:
The reamer mark defines the length of stem and the Fix trial tibia plateau in correct ML and AP position with
height of tibia box together. headed pin. Remove reamer.
See anterior marks where your hemi spacer ends in ML
direction (A) for cuts. Central mark (B) defines the middle
of trial tibia plateau.
13
Operative steps – Tibia First
Determining the size of the tibial components
15 16
2. Option: Option:
Select trial tibia plateau with best fit in ML and AP. Connect trial hemi spacer in correct size below corre-
Use ML positioner NQ676R with required trial stem inter- sponding side of trial tibia plateau.
connected. Define best rotational position.
Place trial tibia plateau with screwed trial hemi spacer
▲
17
14
Option: Gap measurements
18 19
Option: Option:
Use distractor to measure the extension gap medial and If you want, use distractor to measure the flexion gap
lateral with trial tibia plateau and if needed hemi spacer medial and lateral with trial tibia plateau and if needed
below. If necessary perform release and measure again. hemi spacer below. If necessary perform release and
Note the values. measure again. Note the values.
The metal thickness of the trial tibia plateau is 5 mm.
20
Fix trial tibia plateau in correct rotational position with
two headed pins.
15
Operative steps – Tibia First
Preparation of the tibial box
21 22
▲
16
23 25 26
24 27 28
Remove the two headed pins from trial tibia plateau.
A B C
A: Trial stem, B: NQ833R Tibia stem connector, C: Tibia broach 17
Operative steps – Tibia First
Preparation of the tibial box
29 30
Option: Option:
If tension of ligaments was not already measured before. If you want, use distractor to measure the flexion gap
medial and lateral with trial tibia plateau and if required
hemi spacer below. If necessary perform release and
Option:
measure again. Note the values.
Measure the extension gap medial and lateral with trial
tibia plateau and if required hemi spacer below.
If necessary perform release and measure again. Note
the values.
The metal thickness of the trial tibia plateau is 5 mm.
18
Operative steps – Femur First
Preparation of the femur
ML
31 32
33
Ream until reamer connected with handle NE198R has
stable fit and until required depth mark for femur
extension stems on reamer is reached.
Because of distal resection bone release, the reamer has
to be inserted deeper to ensure correct depth for stem.
19
Operative steps – Femur First
Distal femur resection
Details:
B C
A D
34 35
Option:
Check cuts with cutting check device NM350R.
Fix cutting guide with two headless and one headed pin
on anterior femur.
A B C D
A: Distal femur cutting guide NQ701R, B: Guide support NQ703R, C: Alignment system NQ702R, D: Handle NQ474R
20
B
36 38
Remove check device, joint line positioner, and Perform distal femoral cut. If needed cut additionally for
alignment system (by opening screw A and B) from distal femoral spacers in the correct slot.
femur.
▲
37
21
Operative steps – Femur First
Completing the femur resection
39 40
If cut for distal femur spacer was performed: Screw trial Place cutting guide with slided in requested femur
femoral spacer(s) behind 4-in-1 cutting guide to ensure adjustment sleeve (5° NQ705R and 7° NQ707R for
the correct distal position of the femoral cutting guide. cementless and 6° NQ706R for cemented extension
stems) inside on the reamer. Use check device in ant.
parallel cutting slot as reference to avoid anterior
notching.
Measure flexion gap. Dorsal geometry of femoral cutting
guide is corresponding with definitive femoral implant
measurements.
Option:
Connect handles to the cutting guide to adjust rotational
position.
41
Femoral rotation can also be adjusted by using trial
gliding surface in requested height with 4-in-1 cutting
guide.
22
42 43
Fix cutting guide in defined anterior/posterior position by Fix cutting guide in defined rotational position by tight-
tightening the distal screw. The now visible offset value is ening the horizontal screw.
just a rough information. The definitive offset value can
be read later on the assembly of trial femur and trial
Option:
stem.
Use check device in anterior parallel cutting slot as
reference to avoid anterior notching.
Option:
Use handle just on the opposite side of everted patella.
Use check device in anterior parallel cutting slot as
reference to avoid anterior notching.
44
Fix cutting guide in defined rotational position by using
two long headless pin through the holes of the handles.
Fix with one additional headless pin through distal hole
in cutting guide marked with L (left leg)/R (right leg).
Select your operated leg side.
23
Operative steps – Femur First
Completing the femur resection
45 46
3
4
2
47
Perform all femoral cuts in sequence below:
1. Anterior cut
2. Posterior cut (incl, spacer cut(s) if needed)
3. Posterior chamfer cut
4. Anterior chamfer cut
24
Preparation of the femoral box
Femur Femur
medialized lateralized
48 50
Attach trial femoral prosthesis in correct size connected Fix trial femur with two head pin in this defined position.
with screwed on ML adjustment adapter with correct Remove adjustment device and reamer.
angle (5°/7° for cementless extension stems; 6° for
cemented) – if needed in neutral or with medial/lateral
offset – guided by the reamer on the bone until full bone
contact of inner femoral geometry.
49
Example:
If the trial femur switched to medial, the ML-offset-
adapter with correct angle (5°/7°) need to be attached on
the trial femur in offset version medial.
This femur-assembly has to be placed on the reamer until
complete bone contact of the inner femur geometry.
25
Operative steps – Femur First
Preparation of the femoral box
51 52
Perform medial and lateral inner box cuts with oscillating To ensure optimal implant fit of the femoral extension
saw with small saw blade. stem perform medial and lateral rasp procedure to remove
all bone particles, by using rasp NQ832R with intercon-
If necessary use chisel to remove all bone particles.
nected trial stem in requested dimension.
Option:
Attach on trial femur the guide for proximal chisel box cut
NQ796R. Slide sizing device NQ795R in correct slot on the
chisel. Impact chisel NQ794R until stop.
26
Assembly of the femoral trial box
53
Example:
If the femur medialized like in the example before: the
result is that the stem is now in lateral position. So the
54 adapter (e. g. NS016R) with correct angle (5°/7°) has to
be selected in requested offset version medialized like
shown in the pink ring. This stem-assembly is now placed
on the proximal flat femur box surface.
▲
Please note:
The definitive stem implant has to correspond to the
stem-assembly. The version: medialized, neutral or later-
alized is on the adapter displayed. In the example left side
it shows in the green ring: The definitive Implant has to
be a lateralized stem.
55
Impact femoral-box-assembly until stop.
27
Operative steps – Femur First
Assembly of the femoral trial box
56 58
57
Tight trial femur extension stem AP-offset fixation screw
in this self centered position.
28
59 60
29
Operative steps – Femur First
Preparation of the patella
61 63
The thickness of the patella is measured using the patella The drilling clamp NS841R is brought around the patella
foreceps NS840R. This thickness should not be exceeded and fixed. The peg holes are drilled with the diameter
after implantation of the patella implant (see table of 6 mm drill NE480R with stop.
measurements). There is also a caliper AA847R available.
The size of the patella is established with the four color
The aim should be to achieve same level or a reduction in
coded trial patellae implants.
patella thickness following implantation.
64
The implantation is performed with the same foreceps
using the fix inside curved adapter NS842 to press patella
and cemented patella implant together.
Bone cement has also to be in the three drilled peg holes.
62
The foreceps is set to chosen resection height. The
resection is performed through the cutting slot.
30
Assembling the final implants
Please note: To apply the requested torque, all implants which are tightened with defined torque
(Tibia-/Femur stem, PE fixation screw), tight three times with torque wrench!
65 67
27 Nm
66
Select definitive implants.
Option: 68
Screw tibia hemi spacer below the tibia plateau.
Use torque wrench and stem holder NQ834R to screw
extension stem with 27 Nm with correct nut in the femur.
Use torque wrench NE184RM and tibia plateau holder
NQ835R to screw extension stem with 20 Nm below the
tibia in requested offset position.
31
Operative steps – Femur First
Assembling the final implants
70
Click femur insert NQ561
inside femur holder NQ560R
and attach handle. Slide both
retaining fingers of the holder
inside the recesses of the femur
and tighten fixation screw.
69
Sequence of implantation
steps: 71
- Tibia plateau
- Femur
- Gliding surface
- Patella
Insert tibia plateau holder
NQ399RM into the tibia
plateau and spread until
stable fit. Tighten fixation
screw and connect handle.
72
Click defined poly into the tibia plateau by inserting in
first the two posterior noses diagonal. Then the two
anterior ones top down. After hardening of bone cement,
tight poly fixation screw, which is in same package with –
adapter NP450R.
32
73
Explosion view of the Columbus® Revision knee system trial instruments.
33
Implants
Columbus® Revision implant dimensions and design
7 sizes left/right
Bone cuts similar Columbus® primary
Femur box width always 25.5 mm
Hyperextension 4°
Flexion angle 130°
Measurements in [mm]
Cemented: 6°
Length: 77, 157 mm
ML offset: Neutral
Diameter: 12, 15, 18 mm
Cylindrical and polished
4 longitudinal grooves to avoid risk of embolism
Len
gth
Cementless: 5°/7°
Length: 117, 177 mm
ML offset: Neutral, Medialized, Lateralized
Diameter: 12 – 20 mm (steps of 1 mm)
Slightly tapered
10 longitudinal grooves (Wagner profile)
Len
gth
34
Nut for extension stems in neutral and offset version.
Patellae Dimensions
D
Patella P0 27 mm x 7 mm
Patella P1 30 mm x 8 mm
Patella P2 33 mm x 9 mm
Patella P3 36 mm x 10 mm
H
35
Implants
Columbus® Revision implant dimensions and design
11 sizes
Seating for tibial extension stems
Offset: T0/T0+ = ± 4 mm, T1-T5 = ± 6 mm
Flexion angle 130°
Symmetric plateau design
Cemented
Cemented
Length: 52, 92 mm
Diameter: 12, 15, 18 mm
Cylindrical and polished
With asymmetric ‘collar’ for increased stability
3 longitudinal grooves to avoid risk of embolism
Cementless
Length: 92, 132 mm
Diameter: 11 – 20 mm (steps of 1 mm)
Slightly tapered
With asymmetric ‘collar’ for increased stability
10 longitudinal grooves (Wagner profile)
36
Spacer in 5, 10 and 15 mm
Screwed in from underneath
Anatomical medial or lateral design
Cement pockets 1 mm deep
Size T0 T1 T2 T3 T4 T5
AP 40.3 42.3 44.4 47.4 50.5 55.5
ML 24.8 26.3 28.8 31.1 33.8 36.3
ML
5 mm AP 5 mm ML 15 mm AP 15 mm ML
10 mm AP 10 mm ML
Columbus® Revision tibia spacer Tibia 0/0+ Tibia 1/1+ Tibia 2/2+ Tibia 3/3+ Tibia 4/4+ Tibia 5
Original ML dim. 62 65 70 75 80 85
With 2 tibia spacer 5 mm ML 58 61 66 71 76 81
With 2 tibia spacer 10 mm ML 54 57 62 67 72 77
With 2 tibia spacer 15 mm ML 50 53 58 63 68 73
37
Implants
Columbus® Revision implant dimensions and design
Combination options
Torque information:
27 Nm for femur extension stems
20 Nm for tibia extension stems
10 Nm for PE fixation screw
38
Material overview
2 2
2
3
3
2
4
5
2
2 2
1
2
3 UHMWPE 2
4 PEEK Optima (LT1)
5 Ti6Al4V
39
Implants
Compact matrix of implants – Femur components
Ø 12 mm Ø 13 mm Ø 14 mm Ø 15 mm Ø 16 mm
Femur 117 mm 177 mm 117 mm 177 mm 117 mm 177 mm 117 mm 177 mm 117 mm 177 mm
F1-F7 5˚ Neut NR402K NR432K NR403K NR433K NR404K NR434K NR405K NR435K NR406K NR436K
F1-F7 5˚ Med NR412K NR442K NR413K NR443K NR414K NR444K NR415K NR445K NR416K NR446K
F1-F7 5˚ Lat NR422K NR452K NR423K NR453K NR424K NR454K NR425K NR455K NR426K NR456K
F1-F7 7˚ Neut NR502K NR532K NR503K NR533K NR504K NR534K NR505K NR535K NR506K NR536K
F1-F7 7˚ Med NR512K NR542K NR513K NR543K NR514K NR544K NR515K NR545K NR516K NR546K
F1-F7 7˚ Lat NR522K NR552K NR523K NR553K NR524K NR554K NR525K NR555K NR526K NR556K
Ø 17 mm Ø 18 mm Ø 19 mm Ø 20 mm
Femur 117 mm 177 mm 117 mm 177 mm 117 mm 177 mm 117 mm 177 mm
F1-F7 5˚ Neut NR407K NR437K NR408K NR438K NR409K NR439K NR410K NR440K
F1-F7 5˚ Med NR417K NR447K NR418K NR448K NR419K NR449K NR420K NR450K
F1-F7 5˚ Lat NR427K NR457K NR428K NR458K NR429K NR459K NR430K NR460K
F1-F7 7˚ Neut NR507K NR537K NR508K NR538K NR509K NR539K NR510K NR540K
F1-F7 7˚ Med NR517K NR547K NR518K NR548K NR519K NR549K NR520K NR550K
F1-F7 7˚ Lat NR527K NR557K NR528K NR558K NR529K NR559K NR530K NR560K
Please note: If e. g. a offset stem adapter type 7° ‘medial stem’ was selected, the definitive implant must be selected in version 7° ‘medialized’.
40
Compact matrix of implants – Tibia components
Ø 11 mm Ø 12 mm Ø 13 mm Ø 14 mm Ø 15 mm
Tibia 92 mm 132 mm 92 mm 132 mm 92 mm 132 mm 92 mm 132 mm 92 mm 132 mm
T0-T5 NR171K NR181K NR172K NR182K NR173K NR183K NR174K NR184K NR175K NR185K
Ø 16 mm Ø 17 mm Ø 18 mm Ø 19 mm Ø 20 mm
Tibia 92 mm 132 mm 92 mm 132 mm 92 mm 132 mm 92 mm 132 mm 92 mm 132 mm
T0-T5 NR176K NR186K NR177K NR187K NR178K NR188K NR179K NR189K NR180K NR190K
Femur
Standard AS
NR001K NR001Z Femur F1, left
NR002K NR002Z Femur F2, left
NR003K NR003Z Femur F3, left
NR004K NR004Z Femur F4, left
NR005K NR005Z Femur F5, left
NR006K NR006Z Femur F6, left
NR007K NR007Z Femur F7, left
NR011K NR011Z Femur F1, right
NR012K NR012Z Femur F2, right
NR013K NR013Z Femur F3, right
NR014K NR014Z Femur F4, right
NR015K NR015Z Femur F5, right
NR016K NR016Z Femur F6, right
NR017K NR017Z Femur F7, right
Patellae
44
Femur extension stems, cementless, 7° neutral
Standard AS
NR502K NR502Z Femur stem, 7°, 12 x 117 mm, cementless
NR532K NR532Z Femur stem, 7°, 12 x 177 mm, cementless
NR503K NR503Z Femur stem, 7°, 13 x 117 mm, cementless
NR533K NR533Z Femur stem, 7°, 13 x 177 mm, cementless
NR504K NR504Z Femur stem, 7°, 14 x 117 mm, cementless
NR534K NR534Z Femur stem, 7°, 14 x 177 mm, cementless
NR505K NR505Z Femur stem, 7°, 15 x 117 mm, cementless
NR535K NR535Z Femur stem, 7°, 15 x 177 mm, cementless
NR506K NR506Z Femur stem, 7°, 16 x 117 mm, cementless
NR536K NR536Z Femur stem, 7°, 16 x 177 mm, cementless
NR507K NR507Z Femur stem, 7°, 17 x 117 mm, cementless
NR537K NR537Z Femur stem, 7°, 17 x 177 mm, cementless
NR508K NR508Z Femur stem, 7°, 18 x 117 mm, cementless
NR538K NR538Z Femur stem, 7°, 18 x 177 mm, cementless
NR509K NR509Z Femur stem, 7°, 19 x 117 mm, cementless
NR539K NR539Z Femur stem, 7°, 19 x 177 mm, cementless
NR510K NR510Z Femur stem, 7°, 20 x 117 mm, cementless
NR540K NR540Z Femur stem, 7°, 20 x 177 mm, cementless
45
Implants
Implant ordering details
46
Tibia spacer, cemented with screws
Standard AS
NR040K NR040Z Tibia spacer, RM/LL, T0/T0+, 5 mm
NR041K NR041Z Tibia spacer, RM/LL, T0/T0+, 10 mm
NR042K NR042Z Tibia spacer, RM/LL, T0/T0+, 15 mm
NR044K NR044Z Tibia spacer, RM/LL, T1/T1+, 5 mm
NR045K NR045Z Tibia spacer, RM/LL, T1/T1+, 10 mm
NR046K NR046Z Tibia spacer, RM/LL, T1/T1+, 15 mm
NR048K NR048Z Tibia spacer, RM/LL, T2/T2+, 5 mm
NR049K NR049Z Tibia spacer, RM/LL, T2/T2+, 10 mm
NR050K NR050Z Tibia spacer, RM/LL, T2/T2+, 15 mm NR057K NR057Z Tibia spacer, RM/LL, T4/T4+, 10 mm
NR052K NR052Z Tibia spacer, RM/LL, T3/T3+, 5 mm NR058K NR058Z Tibia spacer, RM/LL, T4/T4+, 15 mm
NR053K NR053Z Tibia spacer, RM/LL, T3/T3+, 10 mm NR060K NR060Z Tibia spacer, RM/LL, T5, 5 mm
NR054K NR054Z Tibia spacer, RM/LL, T3/T3+, 15 mm NR061K NR061Z Tibia spacer, RM/LL, T5, 10 mm
NR056K NR056Z Tibia spacer, RM/LL, T4/T4+, 5 mm NR062K NR062Z Tibia spacer, RM/LL, T5, 15 mm
48
Tibia offset stems, cemented
Standard AS
NR191K NR191Z Tibia stem, 12 x 52 mm, cemented
NR194K NR194Z Tibia stem, 12 x 92 mm, cemented
NR192K NR192Z Tibia stem, 15 x 52 mm, cemented
NR195K NR195Z Tibia stem, 15 x 92 mm, cemented
NR193K NR193Z Tibia stem, 18 x 52 mm, cemented
NR196K NR196Z Tibia stem, 18 x 92 mm, cemented
49
Instruments
Overview of instruments, X-ray templates and saw blades
NQ596 Active
NQ594 Passive
NQ29
2
Femu
X-ray templates Scale
1,1:1
r stem
s cem
e nted
Colum
Scale
1,1:1
CENTIM
bus R
evis ion
on 6° −
evisi
ETE
bus R
RS (Sca
um
r Col
157
2
Femu
2
NQ29 7
le is enla
6
3
Y
NQ292 Set of X-ray templates, scale 1.10:1 2:1 5
rged to
0
4
CENTIM
1
templat
5
ETERS
2
e magnific
F7L
6
(Sca
3
2
Y
atio
le is enla
4
n)
2
rged to
5
8
M L
templat
6
77
9
e magnific
7
10
9
12
X
10 Ø12
2:1
Ø15
13
11 Ø18
3 3
14
12 0
1mm
increment
15
13
X 0
14 5
15 10
16
N
15 P
17
15
10
17
5
NQ29
2214
18
B BR AUNTISE
EXPER B BRAU
SHARING
SHARING N
EXPER
TISE
50
Saw blades overview
9 mm 13 mm 13 mm 19 mm 23 mm
■ Thickness:
1.27 mm
10 10
8 mm 10
90
80
9 mm/ 75 mm GE231SU - - - - -
13 mm/ 90 mm - GE233SU GE236SU - - GE266SU
19 mm/ 90 mm - - - GE241SU - GE271SU
23 mm/ 90 mm - - - - GE246SU -
23 mm/100 mm - - - - GE249SU -
25 mm/ 90 mm - - - - - GE276SU
13 mm 19 mm 25 mm
■ Thickness:
1.27 mm
9 mm/ 75 mm - - -
13 mm/ 90 mm GE320SU GE327SU GE323SU
19 mm/ 90 mm - - -
23 mm/ 90 mm - - -
23 mm/100 mm - - -
25 mm/ 90 mm GE322SU GE329SU GE326SU
51
Instruments
Compact overview of instruments
NQ601 General Instruments NQ602 Manual Femur Instruments NQ603 Manual Tibia Instruments
NQ601 General Instruments NQ602 Manual Femur Instruments NQ603 Manual Tibia Instruments
NQ604 Femoral Preparation NQ605 Trial Tibial Components NQ606 Trial Femoral
Components Left
NQ607 Trial Femoral NQ608 Trial Tibia Spacer NQ609 Stem Preparation
Components Right Cemented
52
NQ610 Stem Preparation NQ611 Patella Preparation NQ612 Tibial Instruments
Cementless Set 1/2 Size 0
53
Instruments
Overview of instruments
54
NQ602 Columbus® Revision F Manual Femur Instruments
55
Instruments
Overview of instruments
56
NQ604 Columbus® Revision F Femoral Preparation
57
Instruments
Overview of instruments
58
NQ608 Columbus® Revision F Trial Tibia Spacer
59
Instruments
Overview of instruments
60
NQ611 Patella Preparation
61
Instruments
Overview of instruments
NP615R
Drill, Ø 3.2 mm
FS633
Transmitter, yellow
NE292R
Screw-tap for
Tibia-RB
FS634
NE199R Transmitter, blue
Adapter for
T-handle
NE163R FS635
RB C-hook, small Transmitter, red
NE191R NQ595R
RB C-hook Storage
NE192R
Tibia Rev.-RB
Sliding block NE171R NE324T NM769R NE441RM
NE190R 0° for tibia cutting guide Rev. adapter, Orienting Base plate Post. footplate, left
NE445R 5° for femur cutting guide link cutting block NE442RM
NE446R 6° for femur cutting guide guides Post. footplate, right
NE447R 7° for femur cutting guide
62
63
✂
Columbus ® Revision Implant Matrix – Femoral Components
Femur, cemented
27 Nm for femur stems Possible Combination Femur/Tibia
Types: F1 F2 F3 F4 F5 F6 F7 Types: F1 F2 F3 F4 F5 F6 F7
Left NR001K NR002K NR003K NR004K NR005K NR006K NR007K T0/T0+ OK OK OK + + + +
Right NR011K NR012K NR013K NR014K NR015K NR016K NR017K T1/T1+ OK OK OK OK + + +
T2/T2+ – OK OK OK OK + +
T3/T3+ – – OK OK OK OK +
T4/T4+ – – + OK OK OK OK
T5 – – + + OK OK OK
Femur stems, cemented, 6°
– not compatible
Ø 12 mm Ø 15 mm Ø 18 mm + possible, but not recommended by Aesculap
Types: 77 mm 157 mm 77 mm 157 mm 77 mm 157 mm
F1-F7 NR291K NR294K NR292K NR295K NR293K NR296K