You are on page 1of 2

Connecting Peptide (C-Peptide) CLIA 2 x 50 Test

52025078

Intended Use This product contains chemical ingredients. Contacting with skin or mucosa should
C-peptide (CLIA) is intended for in-vitro quantitative measurement of C-peptide be avoided. If the product is spilled into eyes, mouth or skin accidentally, rinse
in human serum or plasma by chemiluminescent immunoassay. with running water and seek for doctor advice if necessary.
Clinical Significance This product does not involve human source substances and has no biosafety hazard,
but it should still be considered as a potential infectious substance. The treatment,
C-peptide (C-P) is also called connecting peptide and secreted by the pancreatic  use, storage of each component and the disposal of solid and liquid wastes
cells. Proinsulin is the common precursor of C-P and insulin. After enzymatic generated in the analysis process should be hand led in accordance with the
digestion, a molecule of proinsulin is cleaved into a molecule of insulin and a corresponding measures of local biosafety guidelines or regulations.
molecule of C-P. C-P is not affected by exogenous insulin. For diabetics who have
been treated with insulin or produced insulin antibodies, the determination of C- All test equipment and related glass and plastic wares should be clean and free of
P can evaluate endogenous insulin secretion ability, which is of great significance contaminants.
for the diagnosis and treatment of diabetes. In addition, C-P is not inactivated by
liver enzymes and has a longer half-life than insulin, instead it is excreted directly Packaging materials that are non-degradation should be collected and delivered
in the urine by the kidneys, so the concentration of C-peptide in the urine within to the local process center.
24 hours can better reflect the function of pancreatic â cells. They are the main
Do not use reagent kits beyond the expiration date.
clinical significance of C-peptide detection that the diagnosis of pancreatic
dysfunction case, the analysis of etiology and disease type, and the guidance for Waste Management
the treatment plan. At present, methods commonly used for C-P detection are Reagents must be disposed off in accordance with local regulations
chemiluminescent immunoassay and enzyme-linked immunoassay.
Sample
Principle
Serum/Plasma
This C-peptide assay is a “sandwich” chemiluminescent magnetic microparticle Specimen Col lection&Storage
immunoassay. Immunomagnetic microparticles (IMMP), specimen and the Follow this package insert as well as the specimen collection tube manufacturer’s
acridinium ester (AE) labelled conjugate are mixed and incubated to form “IMMP- instructions for specimen collection and preparation. Raw materials and addictive
C-P-conjugate” complexes. After washing, the excess conjugate and other may vary among different manufactures’ collection tubes, which may result in
substances are removed, and then the Chemiluminescence Analyzer Substrate different results. This kit has not been tested with all collection tube manufactures.
Buffer is added, the resulting chemiluminescent reaction is quantified in relative Each laboratory needs to determine the feasibility of the collection tubes or serum
light unit (RLU). The value of RLU is directly proportional to the concentrations of separation products. The specimens should be free of microbial contamination,
C-P in the specimen. fibrin, cells or other particulate matter, severe lipemia, severe hemolysis. If the
sample is cloudy or has visible flocculent, the sample shall be centrifuged, and the
Kit Components
supernatant can be used for detection.

Reagent Product Code Description If testing cannot be done before 24 hours after blood collection, the specimens
52025078 shall be separated to serum or plasma and then be stored. Specimens may be stored
C-peptide R1 3.5 mL Magnetic microparticles coated with anti- for up to 24 hours at 2°C~8°C or 3 months at -20°C. Multiple freeze-thaw cycles of
human C-peptide monoclonal antibody specimens should be avoided. The sample should be transported at low
temperature.
C-peptide R2 6 mL Acridinium labelled anti-human C-peptide
monoclonal antibody Ensure the patients’ samples, calibrators and controls are at ambient temperature
(15°C~25°C) before measurement.
C-peptide R3 3.5 mL Phosphate buffer
Do not use mixed specimens.
Calibrator-1 1 mL C-peptide antigen
Volume required: 200 µL for the first c-peptide test plus 50 µL for each additional
Calibrator-2 1 mL C-peptide antigen c-peptide test from the same sample tube.

Plastic Opener 2 nos Materials provided

C-peptide Casette,Calibrator & Plastic Opener


* Note: Do not mix or interchange d ifferent batches of kit. Procedure
Follow the procedures described in Maccura i1000 Automatic Chemiluminescence
Risk & Safety Analyzer User Manual.
Material Safety data sheets (MSDS) will be provided on request Verify that test parameters are loaded before reagents are loaded, for specific
operations, please refer to the analyzer user manual.
Reagent Storage and Stability
The unopened reagents and calibrators are stable until the expiration dates stated Prior to running the kit, ensure that the magnetic microparticles are mixed to keep
on the labels, when stored upright at 2-8OC and protected from direct sunlight. them suspended and avoid formation of foam.
Avoid freezing. After opening, the reagents are stable for 35 days when used on
Place the reagents in the reagent compartment. When the reagent compartment is
board, and the calibrators are stable for 35 days at 2-8OC.
closed, the analyzer will automatically scan and read the reagent RFID information.
Open Vial Stability
Cal ibration
Once opened, the reagent casette is stable up to 35 days when kept at 2-8 OC. Ensure the calibrator to ambient temperature, mix and place it on the calibrator
rack.
On-board Calibration Stability
Calibration is stable for 7 days. Make sure that the reagent and calibrator lot are same before loading.

Precaution Ensure the related calibrator information is obtained by the analyzer, for specific
Place the reagent kit in upright position always. operations, please refer to the analyzer user manual.

For in vitro diagnostic use only. Recommended calibration frequency:


For every new reagent lot.
Package insert must be followed accordingly. Reliability of assay results cannot be
guaranteed if there are any deviations from the instructions in this package insert. The controls are out of range.

This product requires the handling of human specimens. It is recommended that After 35 days when using the same reagent lot.
all human sourced materials be considered potentially infectious and handled in
After 7 days when using the same reagent kit.
accordance with applicable laws. Al l materials contaminated with patient
specimens should be inactivated by validated procedures (autoclaving or chemical After major repair or replacing spare parts
treatment) in accordance with applicable regulations.

 N

ISO 9001:2015
www.agappe.com EN ISO 13485:2016
Rev.0.280622
2 x 50 Test
Connecting Peptide (C-Peptide) CLIA 52025078

Quality control Performance


Ensure the control to ambient temperature, mix and place it on the required rack. 1.Accuracy:
Ensure the related control information is obtained by the analyzer for specific The recovery rate is within the range of 85.0%~115.0%.
operations, please refer to the analyzer user manual.
2.Sensitivity:
Ensure that control values are within the concentration ranges.
Limit of blank is  0.3 ng/mL The limit of blank is the 95th percentile value of
Recommended quality control frequency: the zero-concentration calibrator for repeated determination.
For every new reagent.
Limit of detection is 0.04 ng/mL.
After every calibration.
The limit of detection is the lowest analytical concentration which is distinguished
Once every 24 hours when the test is in use. from the limit of blank (Miscalculation probability  = 0.05).

If the quality control procedures in user’s laboratory require more frequent use of Limit of quantification is 0.05 ng/mL.
controls to verify test results, follow the laboratory specific procedures.
3.Linearity:
Limitation
In the range of 0.10ng/mL~30.00ng/mL, the coefficient of correlation r is not less
This assay has only been validated for the specimen type of human serum and
than 0.9900.
plasma (heparin and EDTA). Other types of specimens have not been validated.
4.Precision:
Specimens of severe icterus, hemolysis, lipemia and contamination may result in
error results. The intra-lot coefficient of variation (CV) is not higher than 8.0%.
Interference due to extremely high titers of antibodies to streptavidin and intake The inter-lot coefficient of variation (CV) is not higher than 15.0%.
of high dose of biotin can occur.
5. Interference
Heterophilic antibodies in specimen can react with reagent immunoglobulins,
interfering with in vitro immunoassays. Interfering Substance Concentration (Max)

Patients routinely exposed to animals or animal serum products can be prone to Hemoglobin  5.0 g/L
this interference and anomalous values may be observed. Triglycerides  9.0 mmol/L
The results of different detection systems cannot be used interactively. Bilirubin  400 µmol/L
Reference Range Biotin  20 ng/mL
0.77 ng/mL ~ 3.08 ng/mL RF factor  1200 IU/mL
The above reference interval is the midd le 95% interval of the measurement results
6.HOOK TEST:
of 120 specimens from fasting normal population. It is recommended that each
laboratory establish its own normal range which may be unique to the population There is no high-dose hook effect at C-P concentrations up to 200 ng/mL.
it serves depending upon ethnics, geographical, patient, dietary, or environmental
factors. 7.SPECIFICITY
When the concentration of proinsulin is 10 ng/mL, or the concentration of insulin
Interpretation of Result is 500 µIU/mL, the measured results of C-P are al l not higher than 0.25 ng/mL.
The test results shall be only considered as a clinical reference rather than the BIBLIOGRAPHY
unique basis for confirming or excluding a case. For diagnostic purposes, results 1. Clark PM. Assays for insulin, proinsulin(s) and C-peptide. Ann Clin Biochem,
should always be used in combination with clinical examination, medical history 1999, 36 (5): 541-564.
and other result of inspection.
2. Wahren J, jornval l H.C-peptide makes a comeback. Diabetes Metab Res
Rev,2003,19(5): 345-347.


±

 N

ISO 9001:2015
www.agappe.com EN ISO 13485:2016
Rev.0.280622

You might also like