Professional Documents
Culture Documents
C-REACTIVE PROTEIN REAGENT KIT (LATEX ENHANCED IMMUNOTURBIDIMETRIC ASSAY) 3. Testing procedure
Sample volume: 2μL Temperature: 37℃
Reagent 1 volume: 320μL Reagent 2 volume: 80μL Main wavelength: 546nm
◆Product Name Absorbance A1
C-REACTIVE PROTEIN REAGENT KIT(LATEX ENHANCED IMMUNOTURBIDIMETRIC ASSAY) Absorbance A2
Main Wavelength 546 nm Optical Path 1.0cm Sample 1 13.53 0.66 4.88 15.28 0.29 1.90
Sample Volume 2 μL Reagent 2 Volume 80μL Sample 2 43.62 1.02 2.34 64.87 1.44 2.22
Absorbance Range 0~3.2A Reaction Time 300s 5. Method comparison: Conduct tests upon 200 samples with reagent y of the company and the on the market reagent x, the
Test Mode 2 point end assay Delay Time 24s relevance between our reagent (y) and approved market reagent (x) is: y=0.8336x+1.1386 r=0.9918.
◆Standardization Traceability
The constant value of calibrator can be traced to international reference ERM-DA472/IFCC.
Positive Positive Positive Positive Positive Positive Positive Positive Positive Positive
◆Matters Need Attention Reaction type
reaction reaction reaction reaction reaction reaction reaction reaction reaction reaction
1. Cautions for operation -3.3 lower -3.3 lower -3.3 lower -3.3 lower -3.3 lower -3.3 lower -3.3 lower -3.3 lower -3.3 lower -3.3 lower
Prozone check
1.1 The product is only for in vitro diagnosis. limit limit limit limit limit limit limit limit limit limit
1.2 Do not add reagent during the test. Calibration method Logit-log5p Logit-log5p Logit-log5p Logit-log5p Logit-log5p Logit-log5p Logit-log5p Logit-log5p Logit-log5p Logit-log5p
1.3 Volume of reagent and sample can be changed proportionally in accordance with the requirements of instrument.
2. Cautions for safety Deflection check 0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.1
2.1 Consider the product as dangerous materials that may cause HIV, HBV, HCV and other infections. To avoid the risk, use Discreteness check 0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.1
single-use gloves.
2.2 Avoid contact with skin, clothes, and eyes. Once in contact with skin or clothes, rinse the contact part with plenty of water, Sensitivity check 0.05 0.05 0.05 0.05 0.05 0.05 0.05 0.05 0.05 0.05
safety operation rule, local laws and regulations. Linearity range 1~80 1~80 1~80 1~80 1~80 1~80 1~80 1~80 1~80 1~80
◆Reference
1. Lars-Olof Hanson et al. Current Opinion in Infect. Diseases 1997; 10:196~201.
2. Chetana Vaishnavi. Immunology and Infetious Diseases 1996; 6:139~144.
3. NCCLS.Interference Testing in Clinical Chemistry;Approved Guideline,2005.
◆Date of Approval and Revision: 11/2018
◆Packaging Specification
No. Specifications Type
232021602030 R1:2×60mL R2:2×15mL Calibration sample: 1mL Dirui CS-T series package
232021602031 R1:1×40mL R2:1×10mL Calibration sample: 1mL Dirui CS-T series package
232021602032 R1:1×40mL R2:1×10mL Calibration sample: 1mL Dirui CS-T series package
Item CRP CRP CRP CRP CRP CRP CRP CRP CRP CRP
Unit mg/L mg/L mg/L mg/L mg/L mg/L mg/L mg/L mg/L mg/L
Method 2 point end 2 point end 2 point end 2 point end 2 point end 2 point end 2 point end 2 point end 2 point end 2 point end
Time 20 20 10 10 10 13 10 9 12 9
Photometric point 18~31 18~31 18~31 17~41 17~31 22~41 22~47 15~31 15~31 17~33
Main wavelength 546 546 546 546 546 546 546 546 546 546
Sub wavelength 0 0 0 0 0 0 0 0 0 0
Reagent R1/T1 320 320 320 320 320 320 320 280 280 320
R2/T2 80 80 0 80 0 80 80 70 70 0
R3/T3 —— —— 80 —— 80 —— —— —— —— 80
R4/T4 —— —— 0 —— 0 —— —— —— —— 0
Normal volume of
2 2 2 2 2 3 3 1.8 1.8 2
serum sample
Absorbance limit 3.3 3.3 3.3 3.3 3.3 3.3 3.3 3.3 3.3 3.3