Latex turbidimetry
REAGENTS
Tris buffer 20 mmol/L, pH 8.2. Sodium azide 0.95 CALIBRATION
Diluent (R1)
g/L.
Latex particles coated with goat IgG anti-human
Latex (R2)
CRP, pH 7.3. Sodium azide 0.95 g/L. Linear two points Water 0
Calibrator. C-Reactive protein concentration is stated Calibrator *
CRP-CAL
on the vial label.
Ref.:1102114 Control serum ASO/CRP/RF Level L
Optional
Ref.:1102115 Control serum ASO/CRP/RF Level H.
The Calibration is stable until 31 days. After this period the Calibration
PRECAUTIONS must be performed again in order to obtain good results.
Components from human origin have been tested and found to be
negative for the presence of HBsAg, HCV, and antibody to HIV (1/2). QUALITY CONTROL
However handle cautiously as potentially infectious. Control Sera are recommended to monitor the performance of manual
and automated assay procedures. It should be used SPINREACT
CALIBRATION Controls ASO/CRP/RF Level L (Ref.:1102114) and Level H (Ref.:
Use CRP Calibrator Reference 1107002. 1102115).
The sensitivity of the assay and the target value of the calibrator have Each laboratory should establish its own Quality Control scheme and
been standardized against the Reference Material CRM 470/RPPHS corrective actions if controls do not meet the acceptable tolerances.
(Institute for Reference Materials and Measurements, IRMM).
Recalibrate when control results are out of specified tolerances, when REFERENCE VALUES
using different lot of reagent and when the instrument is adjusted. Normal values up to 6 mg/L.
Each laboratory should establish its own reference range.
PREPARATION
Ready for use. INTERFERENCES
CRP Calibrator: Reconstitute () with 1.0 mL of distilled water. Mix Bilirrubin (20 mg/dL), lipemia (10 g/L) and rheumatoid factors (300
gently and incubate 10 minutes at room temperature before use. IU/mL) do not interfere. Hemoglobin (≥ 5 g/L), interferes. Other
substances may interfere7.
STORAGE AND STABILITY
All the components of the kit are stable until the expiration date on the NOTES
label when stored tightly closed at 2-8ºC and contaminations are Clinical diagnosis should not be made on findings of a single test
prevented during their use. Do not use reagents over the expiration date. result, but should integrate both clinical and laboratory data.
Reagent deterioration: Presence of particles and turbidity. BIBLIOGRAPHY
CRP Calibrator: Stable for 1 month at 2-8ºC or 3 months at –20ºC. 1. Lars-Olof Hanson et al. Current Opinion in Infect Diseases 1997; 10:
196-201.
Do not freeze; frozen Latex or Diluent could change the functionality of 2. Chetana Vaishnavi. Immunology and Infectious Diseases 1996; 6:
the test. 139 – 144.
3. Yoshitsugy Hokama et al. Journal of Clinical Lab. Status 1987; 1: 15
ADDITIONAL EQUIPMENT – 27.
- SPIN 800 autoanalyzer. 4. Kari Pulki et al. Sacand J Clin Lab Invest 1986; 46: 606 – 607.
- Laboratory equipment. 5. Werner Müller et al. Journal of Immunological Methods 1985; 80: 77
– 90.
SAMPLES 6. Shogo Otsuji et al. Clin Chem 1982; 28/10: 2121 – 2124.
Fresh serum. Stable 7 days at 2-8ºC or 3 months at –20ºC. 7. Young DS. Effects of drugs on clinical laboratory test, 4th ed. AACC
The samples with presence of fibrin should be centrifuged before testing. Press, 1995.
Do not use highly hemolized or lipemic samples.
PACKAGING
Ref.: MX1107101 R1. Diluent: 2 x 40 mL
BARCODED REAGENTS LOAD MUST BE PRECEDED OF A Cont.
SPINREACT “DATABASE” COPY INTO THE ANALYZER R2. Latex :1 x 20 mL
SOFTWARE. IT IS AVAILABLE UNDER REQUEST TO SPINREACT. CRP-CAL :1 x 1 mL
MXTLIS40-I 29/09/16 SPINREACT,S.A./S.A.U. Ctra.Santa Coloma, 7 E-17176 SANT ESTEVE DE BAS (GI) SPAIN
Tel. +34 972 69 08 00 Fax +34 972 69 00 99 e-mail: spinreact@spinreact.com
CRP-TURBI
PCR-turbilátex
Turbidimetría Látex
MXTLIS40-E 29/09/16 SPINREACT,S.A./S.A.U. Ctra.Santa Coloma, 7 E-17176 SANT ESTEVE DE BAS (GI) SPAIN
Tel. +34 972 69 08 00 Fax +34 972 69 00 99 e-mail: spinreact@spinreact.com
CRP-TURBI
PCR-turbilatex
Latex turbidimétrique
PRÉPARATION REMARQUES
Tous les réactifs sont prêts à l’emploi. Les diagnostics cliniques doivent être effectués sur les découvertes
d’un seul résultat de l’essai, mais doivent intégrer à la fois les données
Calibrateur CRP: Reconstituer () avec 1,0 mL d’eau distillée. Mélanger
cliniques et celles du laboratoire.
doucement et incuber pendant 10 minutes sous la température ambiante
avant l’utilisation.
BIBLIOGRAPHIE
1. Lars-Olof Hanson et al. Current Opinion in Infectious diseases 1997;
CONSERVATION ET STABILITÉ
10: 196-201.
Toutes les composantes du kit sont stables jusqu’à l’expiration de la date
2. Chetana Vaishnavi. Immunology and Infectious Diseases 1996; 6:
qui figure sur l’étiquette lorsqu’elles sont conservées hermétiquement à 2-
139 – 144.
8ºC et les contaminations sont empêchées lors de leur utilisation. Ne pas
3. Yoshitsugy Hokama et al. Journal of Clinical Laboratory Status 1987;
utiliser les réactifs au-delà de leur date d’expiration.
1: 15 – 27.
Détérioration du réactif : Présence de particules et turbidité.
4. Kari Pulki et al. Sacand J Clin Lab Invest 1986; 46: 606 – 607.
Calibrateur CRP: Stable pendant 1 mois à 2-8ºC ou 3 mois à -20ºC.
5. Werner Müller et al. Journal of Immunological Methods 1985; 80: 77
Ne pas geler ; le Latex gelé ou le Diluant peuvent modifier la fonctionnalité
– 90.
de l’essai.
6. Shogo Otsuji et al. Clin Chem 1982; 28/10: 2121 – 2124.
7. Young DS. Effects of drugs on clinical laboratory test, 4th ed. AACC
ÉQUIPEMENTS SUPPLÉMENTAIRES
Press, 1995.
- Auto-analyseur SPIN 800.
- Equipament classique de laboratoire.
PRÉSENTATION
ÉCHANTILLONS Ref.: MX1107101 R1. Diluyente: 2 x 40 mL
Cont.
Sérum frais. Stable pendant 7 jours à 2-8ºC ou 3 mois à –20ºC. R2. Látex:1 x 20 mL
Les échantillons qui contiennent la fibrine doivent être centrifugés avant CRP-CAL:1 x 1 mL
l’essai.
Ne pas utiliser des échantillons hémolysés ou lipidémiques.
MXTLIS40-F 29/09/16 SPINREACT,S.A./S.A.U. Ctra.Santa Coloma, 7 E-17176 SANT ESTEVE DE BAS (GI) SPAIN
Tel. +34 972 69 08 00 Fax +34 972 69 00 99 e-mail: spinreact@spinreact.com