CAG SIGNAL TIP SET INSTRUCTION

NOTICE

D This set should be used and will be effective only for a signal communication problem occurred in China,
like a DAS Trigger Less Error.

D The configuration of this tip set (CAGx6) has a better electrical contact reliability, but also shows a higher
wear rate than a normal configuration (CAGx1 and CA40x5).

D Once this set was installed to CT System, a check of the tip remaining length should be taken at least per
half a year. In case of its worn–out, there is a possibility of a serous damage on a slipring surface.

D When this set is installed to CT System, a Notice Label included in this set should be put on the signal
brush board in place. Do not use the configuration of this set with a normal configuration on the same slip-
way. After installation of this set, continue to use it.

Illustration 1–1 Location of the Notice Label

Notice Label

2319208, Rev0, October 15, 2001 Copyright 2001 by General Electric co.