P9015EK
Revision 4

CT
InSite Kit Installation
for system without ISC Assy

Copyrighte 1993, 1994, 1997, 2001 by General Electric Company

Operating Documentation
CT INSITE KIT INSTALLATION
GE MEDICAL SYSTEMS FOR SYSTEM WITHOUT ISC ASSY
REV 4 P9015EK

D THIS SERVICE MANUAL IS AVAILABLE IN ENGLISH ONLY.


WARNING D IF A CUSTOMER’S SERVICE PROVIDER REQUIRES A LANGUAGE OTHER
THAN ENGLISH, IT IS THE CUSTOMER’S RESPONSIBILITY TO PROVIDE
TRANSLATION SERVICES.
D DO NOT ATTEMPT TO SERVICE THE EQUIPMENT UNLESS THIS SERVICE
MANUAL HAS BEEN CONSULTED AND IS UNDERSTOOD.
D FAILURE TO HEED THIS WARNING MAY RESULT IN INJURY TO THE SERVICE
PROVIDER, OPERATOR OR PATIENT FROM ELECTRIC SHOCK,
MECHANICAL OR OTHER HAZARDS.

D CE MANUEL DE MAINTENANCE N’EST DISPONIBLE QU’EN ANGLAIS.


AVERTISSEMENT D SI LE TECHNICIEN DU CLIENT A BESOIN DE CE MANUEL DANS UNE AUTRE
LANGUE QUE L’ANGLAIS, C’EST AU CLIENT QU’IL INCOMBE DE LE FAIRE
TRADUIRE.
D NE PAS TENTER D’INTERVENTION SUR LES ÉQUIPEMENTS TANT QUE LE
MANUEL SERVICE N’A PAS ÉTÉ CONSULTÉ ET COMPRIS.
D LE NON-RESPECT DE CET AVERTISSEMENT PEUT ENTRAÎNER CHEZ LE
TECHNICIEN, L’OPÉRATEUR OU LE PATIENT DES BLESSURES DUES À DES
DANGERS ÉLECTRIQUES, MÉCANIQUES OU AUTRES.

D DIESES KUNDENDIENST–HANDBUCH EXISTIERT NUR IN


ENGLISCHER SPRACHE.
WARNUNG
D FALLS EIN FREMDER KUNDENDIENST EINE ANDERE SPRACHE BENÖTIGT,
IST ES AUFGABE DES KUNDEN FÜR EINE ENTSPRECHENDE ÜBERSETZUNG
ZU SORGEN.
D VERSUCHEN SIE NICHT, DAS GERÄT ZU REPARIEREN, BEVOR DIESES
KUNDENDIENST–HANDBUCH NICHT ZU RATE GEZOGEN UND VERSTANDEN
WURDE.
D WIRD DIESE WARNUNG NICHT BEACHTET, SO KANN ES ZU VERLETZUNGEN
DES KUNDENDIENSTTECHNIKERS, DES BEDIENERS ODER DES PATIENTEN
DURCH ELEKTRISCHE SCHLÄGE, MECHANISCHE ODER SONSTIGE
GEFAHREN KOMMEN.

D ESTE MANUAL DE SERVICIO SÓLO EXISTE EN INGLÉS.


AVISO D SI ALGÚN PROVEEDOR DE SERVICIOS AJENO A GEMS SOLICITA UN IDIOMA
QUE NO SEA EL INGLÉS, ES RESPONSABILIDAD DEL CLIENTE OFRECER UN
SERVICIO DE TRADUCCIÓN.
D NO SE DEBERÁ DAR SERVICIO TÉCNICO AL EQUIPO, SIN HABER
CONSULTADO Y COMPRENDIDO ESTE MANUAL DE SERVICIO.
D LA NO OBSERVANCIA DEL PRESENTE AVISO PUEDE DAR LUGAR A QUE EL
PROVEEDOR DE SERVICIOS, EL OPERADOR O EL PACIENTE SUFRAN
LESIONES PROVOCADAS POR CAUSAS ELÉCTRICAS, MECÁNICAS O DE
OTRA NATURALEZA.

a
CT INSITE KIT INSTALLATION
GE MEDICAL SYSTEMS FOR SYSTEM WITHOUT ISC ASSY
REV 4 P9015EK

D ESTE MANUAL DE ASSISTÊNCIA TÉCNICA SÓ SE ENCONTRA


DISPONÍVEL EM INGLÊS.
ATENÇÃO
D SE QUALQUER OUTRO SERVIÇO DE ASSISTÊNCIA TÉCNICA, QUE NÃO A
GEMS, SOLICITAR ESTES MANUAIS NOUTRO IDIOMA, É DA
RESPONSABILIDADE DO CLIENTE FORNECER OS SERVIÇOS DE TRADUÇÃO.
D NÃO TENTE REPARAR O EQUIPAMENTO SEM TER CONSULTADO E
COMPREENDIDO ESTE MANUAL DE ASSISTÊNCIA TÉCNICA.
D O NÃO CUMPRIMENTO DESTE AVISO PODE POR EM PERIGO A SEGURANÇA
DO TÉCNICO, OPERADOR OU PACIENTE DEVIDO A‘ CHOQUES ELÉTRICOS,
MECÂNICOS OU OUTROS.

D IL PRESENTE MANUALE DI MANUTENZIONE È DISPONIBILE


SOLTANTO IN INGLESE.
AVVERTENZA
D SE UN ADDETTO ALLA MANUTENZIONE ESTERNO ALLA GEMS RICHIEDE IL
MANUALE IN UNA LINGUA DIVERSA, IL CLIENTE È TENUTO A PROVVEDERE
DIRETTAMENTE ALLA TRADUZIONE.
D SI PROCEDA ALLA MANUTENZIONE DELL’APPARECCHIATURA SOLO DOPO
AVER CONSULTATO IL PRESENTE MANUALE ED AVERNE COMPRESO IL
CONTENUTO.
D NON TENERE CONTO DELLA PRESENTE AVVERTENZA POTREBBE FAR
COMPIERE OPERAZIONI DA CUI DERIVINO LESIONI ALL’ADDETTO ALLA
MANUTENZIONE, ALL’UTILIZZATORE ED AL PAZIENTE PER
FOLGORAZIONE ELETTRICA, PER URTI MECCANICI OD ALTRI RISCHI.

b
CT INSITE KIT INSTALLATION
GE MEDICAL SYSTEMS FOR SYSTEM WITHOUT ISC ASSY
REV 4 P9015EK

IMPORTANT! . . . X-RAY PROTECTION


X-ray equipment if not properly used may cause injury. Accordingly, the instructions herein contained should
be thoroughly read and understood by everyone who will use the equipment before you attempt to place this
equipment in operation. The General Electric Company, Medical Systems Group, will be glad to assist and
cooperate in placing this equipment in use.
Although this apparatus incorporates a high degree of protection against x-radiation other than the useful beam, no
practical design of equipment can provide complete protection. Nor can any practical design compel the operator to
take adequate precautions to prevent the possibility of any persons carelessly exposing themselves or others to
radiation.
It is important that everyone having anything to do with x-radiation be properly trained and fully acquainted with the
recommendations of the National Council on Radiation Protection and Measurements as published in NCRP Reports
available from NCRP Publications, 7910 Woodmont Avenue, Room 1016, Bethesda, Maryland 20814, and of the
International Commission on Radiation Protection, and take adequate steps to protect against injury.
The equipment is sold with the understanding that the General Electric Company, Medical Systems Group, its agents,
and representatives have no responsibility for injury or damage which may result from improper use of the equipment.
Various protective material and devices are available. It is urged that such materials or devices be used.

CERTIFIED ELECTRICAL CONTRACTOR STATEMENT


All electrical installations that are preliminary to positioning of the equipment at the site prepared for the equipment shall be
performed by licensed electrical contractors. In addition, electrical feeds into the Power Distribution Unit shall be performed
by licensed electrical contractors. Other connections between pieces of electrical equipment, calibrations, and testing shall
be performed by qualified GE Medical personnel. The products involved (and the accompanying electrical installations) are
highly sophisticated, and special engineering competence is required.

In performing all electrical work on these products, GE will use its own specially trained field engineers. All of GE’s electrical
work on these products will comply with the requirements of the applicable electrical codes.

The purchaser of GE equipment shall only utilize qualified personnel (i.e., GE’s field engineers, personnel of third-party
service companies with equivalent training, or licensed electricians) to perform electrical servicing on the equipment.

c
CT INSITE KIT INSTALLATION
GE MEDICAL SYSTEMS FOR SYSTEM WITHOUT ISC ASSY
REV 4 P9015EK

DAMAGE IN TRANSPORTATION

All packages should be closely examined at time of delivery. If damage is apparent, have notation “damage in
shipment” written on all copies of the freight or express bill before delivery is accepted or “signed for” by a General
Electric representative or a hospital receiving agent. Whether noted or concealed, damage MUST be reported to the
carrier immediately upon discovery, or in any event, within 14 days after receipt, and the contents and containers held
for inspection by the carrier. A transportation company will not pay a claim for damage if an inspection is not requested
within this 14 day period.

Call Traffic and Transportation, Milwaukee, WI (414) 827–3449 / 8*285–3449 immediately after damage is found. At
this time be ready to supply name of carrier, delivery date, consignee name, freight or express bill number, item
damaged and extent of damage.

Complete instructions regarding claim procedure are found in Section “S” of the Policy & Procedure Bulletins.

OMISSIONS & ERRORS

GE personnel, please use the GEMS CQA Process to report all omissions, errors, and defects in this documentation.
Customers, please contact your GE Sales or Service represenatives.

CAUTION

Do not use the following devices near this equipment. Use of these devices near this equipment could cause
this equipment to malfunction.

Devices not to be used near this equipment:

Devices which intrinsically transmit radio waves such as; cellular phone, radio transceiver, mobile radio transmitter,
radio–controlled toy, etc.

Keep power to these devices turned off when near this equipment.

Medical staff in charge of this equipment is required to instruct technicians, patients and other people who may be
around this equipment to fully comply with the above regulation.

d
CT INSITE KIT INSTALLATION
GE MEDICAL SYSTEMS FOR SYSTEM WITHOUT ISC ASSY
REV 4 P9015EK

REVISION HISTORY

REV DATE PRIMARY REASON FOR CHANGE


0 ..... Apr. 15, 1993 ... Initial release.
1 ..... Sep. 30, 1994 ... Software information (Part No.) added.
2 ..... Feb. 26. 1997 ... Included Em/Sytec 1600i/Sytec 1800i and manual title changed.
3 ..... Nov. 06, 1997 ... M/L changed
4 ..... Nov. 14, 2001 ... Modem notice added (sec2)

LIST OF EFFECTIVE PAGES

PAGE REV PAGE REV PAGE REV PAGE REV PAGE REV
Title page . . . . . . . . . . 4 a to d . . . . . . . . . . . . . . 4 i................... 2 2–1 to 2–2 . . . . . . . . . . 4 Rear page
Title page rear . . blank A .................. 4 1–1 . . . . . . . . . . . . . . . . 2 3–1 to 3–5 . . . . . . . . . . 2

A
CT INSITE KIT INSTALLATION
GE MEDICAL SYSTEMS FOR SYSTEM WITHOUT ISC ASSY
blank P9015EK

B
CT INSITE KIT INSTALLATION
GE MEDICAL SYSTEMS FOR SYSTEM WITHOUT ISC ASSY
REV 2 P9015EK

TABLE OF CONTENTS

SECTION PAGE

TABLE OF CONTENTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . i
LIST OF ILLUSTRATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ii

SECTION 1 – INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–1

SECTION 2 – MATERIAL LIST . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–1


2–1 INSITE KIT (46–320293G1) for GEMS–AM Pole . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–1
2–2 INSITE KIT (46–320293G2) for GEMS–E Pole and GEMS–A Pole( except GEYMS) 2–1
2–3 INSITE KIT (2127805 – for Em) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–2
2–4 INSITE KIT (2162605 – for Sytec 1600i/Sytec 1800i) . . . . . . . . . . . . . . . . . . . . . . . . . . 2–2

SECTION 3 – INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–1


3–1 REQUIRED TOOLS, SUPPLIES AND TEST EQUIPMENT . . . . . . . . . . . . . . . . . . . . . 3–1
3–2 INSTALLATION PROCEDURE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–1

i
CT INSITE KIT INSTALLATION
GE MEDICAL SYSTEMS FOR SYSTEM WITHOUT ISC ASSY
blank P9015EK

ii
CT INSITE KIT INSTALLATION
GE MEDICAL SYSTEMS FOR SYSTEM WITHOUT ISC ASSY
REV 2 P9015EK

SECTION 1 - INTRODUCTION

The InSite installation provides a communication network between Support Engineer at the OLC, ESC, ASC and the
CT scanner system for the purpose of servicing and supporting a CT scanner system.

An external 19.2 k baud modem is connected to the system computer through a RS–232C serial cable. The network is
completed by connecting the modem to a voice grade DID, Direct Inward Dial, phone line.

This kit is for the systems that are released without ISC Assy, that are:

D CT Pace/Pace Plus

D CT Sytec 3000, Sytec 2000/3000s, Sytec 3000, Sytec 3000 plus/4000

D CT Em/Sytec1600i/Sytec 1800i

CAUTION

PREREQUISITE:
To use InSite, the following system version is required. The related FMI must be performed
before this installation if the system is not at the appropriate version as listed below.

D CT Pace/Pace Plus Version 9.00 (FMI29105/29106) or above

D CT Syteclp 3000 Version 5.00 (FMI26043) or above

D CT Sytec 2000/3000s Version 3.00 (FMI26059) or above

D CT Sytec 3000plus Version 5.00 (FMI26044) or above

D CT Sytec 4000 Version 5.00 (FMI26045) or above

D CT Em/Sytec1600i/Sytec1800i Any Version

1–1
CT INSITE KIT INSTALLATION
GE MEDICAL SYSTEMS FOR SYSTEM WITHOUT ISC ASSY
blank P9015EK

1–2
CT INSITE KIT INSTALLATION
GE MEDICAL SYSTEMS FOR SYSTEM WITHOUT ISC ASSY
REV 4 P9015EK

SECTION 2 – MATERIAL LIST

2–1 INSITE KIT (46–320293G1) for GEMS–AM (except Sytec 1600i/Sytec 1800i)
Item Part Number Description Quantity
1 46–3278761P2 HIGH SPEED, DIAL–UP MODEM WITH POWER SUPPLY 1
2 46–278763P1 PHONE LINE CORD – 25 FEET 1
3 46–251182P363 MR/CT INLINK RATING PLATE 1
4 46–301402P1 RESTRICTED TO GEMS LABEL 1
5 46–320357P1 MODEM CABLE, 10 FEET 1
6 46–136246P16 1” WIDE VELCRO – 5 FEET 1
7 46–136246P18 1” WIDE VELCRO – 5 FEET 1
8 2144760 ISC CIRCUIT BOARD 1
9 46–320095P9 INSITE CABINET ID MAGNET – SC 1
10 46–320095P10 INSITE CABINET ID MAGNET – GPS 1
11 46–320095P11 INSITE CABINET ID MAGNET – SXFMR 1
12 46–320095P12 INSITE CABINET ID MAGNET – TME 1
13 46–320551P1 LOCKING POST KIT 1
14 46–312343P17 M3 (0.5) x 6 SCREW 2
15 P9020EE INSITE INSTALLATION INSTRUCTION (MR) 1
16 P9015EK INSITE INSTALLATION INSTRUCTION (CT) 1

2–2 INSITE KIT (46–320293G2) for GEMS–A (except GEYMS) and GEMS–E (except Sytec 1600i/Sytec
1800i)
Item Part Number Description Quantity
1 46–278763P1 PHONE LINE CORD – 25 FEET 1
2 46–251182P364 MR/CT INLINK RATING PLATE 1
3 46–301402P1 RESTRICTED TO GEMS LABEL 1
4 46–320357P1 MODEM CABLE, 10 FEET 1
5 46–136246P16 1” WIDE VELCRO – 5 FEET 1
6 46–136246P18 1” WIDE VELCRO – 5 FEET 1
7 2144760 ISC CIRCUIT BOARD 1
8 46–320095P9 INSITE CABINET ID MAGNET – SC 1
9 46–320095P10 INSITE CABINET ID MAGNET – GPS 1
10 46–320095P11 INSITE CABINET ID MAGNET – SXFMR 1
11 46–320095P12 INSITE CABINET ID MAGNET – TME 1
12 46–320551P1 LOCKING POST KIT 1
13 46–312343P17 M3 (0.5) x 6 SCREW 2
14 P9020EE INSITE INSTALLATION INSTRUCTION (MR) 1
15 P9015EK INSITE INSTALLATION INSTRUCTION (CT) 1
Note
In addition to 46–320293G2, Modem must be prepared locally. The recommended type of the Mo-
dem is: T2000 (made by Telebit company)

2–1
CT INSITE KIT INSTALLATION
GE MEDICAL SYSTEMS FOR SYSTEM WITHOUT ISC ASSY
REV 4 P9015EK

SECTION 2 – MATERIAL LIST (continued)

2–3 INSITE KIT (2127805 – for Em)


ITEM PART NO. DESCRIPTION QTY.
1 2144760 ISC ASSY 1
2 P9309US CABLE ASSY 1
3 2195668 MODEM 1
4 U0032BA VELCRO TAPE 1
5 U0033BA VELCRO TAPE 1
7 P9015EK INSTRUCTION 1
8 2196935 INSITE POWER CABLE 1

2–4 INSITE KIT (2162605 – for Sytec 1600i/Sytec 1800i)


Note
The Modem must be prepared locally.

ITEM PART NO. DESCRIPTION QTY.


1 2144760 ISC ASSY 1
2 P9309US CABLE ASSY 1
3 U0032BA VELCRO TAPE 1
4 U0033BA VELCRO TAPE 1
5 P9015EK INSTRUCTION 1

2–2
CT INSITE KIT INSTALLATION
GE MEDICAL SYSTEMS FOR SYSTEM WITHOUT ISC ASSY
REV 2 P9015EK

SECTION 3 – INSTALLATION

3–1 REQUIRED TOOLS, SUPPLIES AND TEST EQUIPMENT


Standard hand tools are used for the hardware installation. No special tools or test equipment are required.

3–2 INSTALLATION PROCEDURE


1. Confirm that a telephone line for INSITE use is available.

2. Turn the power switch of the Operator Console OFF.

3. Remove the OC Front Cover.

4. For CT Pace/Pace Plus and Eu/Sytec1600i/Sytec1800:


Remove the OC Rear Cover.

- Modifying the ISC Board


5. Check if the DIP switch setting of SW10B1 on the ISC Board is as shown in Illustration 3–1. Use proper ESD
handling procedures.

ON ND 8
ON
DS2 OFF
J3
1 J1 1 2 3 4 5 6 7 8
4
SW10B1
DS1
1 8
SW NO. ON OFF
J5 1, 5 RS–232C RS–422A
1 J2
4
2, 6 DA IN USE DB IN USE
DS3
3, 7 TXC/RXC NO USE TXC/RXC IN USE

4, 8 DTE MODE DCE MODE

Sw 1 to 4: for J5, sw 5 to 8: for J3

DIP SWITCH ON ISC ASSY


ILLUSTRATION 3–1

3–1
CT INSITE KIT INSTALLATION
GE MEDICAL SYSTEMS FOR SYSTEM WITHOUT ISC ASSY
REV 2 P9015EK

3–2 INSTALLATION PROCEDURE (continued)


Note
Steps 6. and 7. may be already completed. If so continue at the step 8. If your system is Eu/Sy-
tec1600i/Sytec1800i, also skip steps 6. and 7.

6. Remove the two screws securing connector J3 (25 pin Sub–D) to the ISC Board. Replace them with the new
ones (46–312343P17) supplied with the Inlink kit. See Illustraion 3–2.

7. Remove the two cable–connect–standoffs from connector J3. Replace them with the new ones (46–320551P1).
See Illustration 3–2.

Replace these
two screws.

Replace these
two cable–con- J3
nect–standoffs. CONNECTOR

ISC Board
(Soldering
side)

J3 CONNECTOR MODIFICATION ON ISC BOARD


ILLUSTRATION 3–2

8. Install the ISC Board in the OC (Pace series; Slot 21, Sytec series; Slot 20, Eu/Sytec1600i/Sytec1800i: Slot 7).

- Modifying the FILE–BB Board


9. For CT Pase/Pace Plus and Eu/Sytec1600i/Sytec1800: Set all the J133 (J071) Bus priority Jumpers, located on
the right side of the FILE–BB Board, to the Open Position.

AUX4 AUX
A5 B C MISC2 ISC
0

J133 J071

Pase/Pace Plus Eu/Sytec1600i/Sytec1800


BACK BOARD JUMPERS SETTING
ILLUSTRATION 3–3

3–2
CT INSITE KIT INSTALLATION
GE MEDICAL SYSTEMS FOR SYSTEM WITHOUT ISC ASSY
REV 2 P9015EK

3–2 INSTALLATION PROCEDURE (continued)

- Routing the modem cable


10. Route the modem cable as shown in Illustration 3–4 and connect the cable to the modem.

PACE Series and Sytec 2000/3000s/3000/3000plus/4000 Series

Operator Console

OC REAR DUCT

J3
ISC CONNECTOR
ASSY
MODEM (ATTACHED
WITH SWITCHES/LEDS
FACING FORWARD)
MODEM CABLE
RATING PLATE

MODEM POWER SUPPLY


TO AC OUTLET

ÀÀ
ÀÀ
REAR DUCT

Operator Console
ÀÀ
ÀÀ
MODEM POW-
ER
SUPPLY

ÀÀ
ÀÀ
ÀÀ TO AC OUTLET

ÀÀ
J3
CONNECTOR ISC

ÀÀ
ASSY

RATING PLATE

Eµ/Sytec1600i/Sytec1800i

MODEM (ATTACHED
WITH SWITCHES/LEDS
MODEM CABLE FACING FORWARD)

CABLE WIRING AND POSITION OF MODEM AND RATING PLATE


ILLUSTRATION 3–4

3–3
CT INSITE KIT INSTALLATION
GE MEDICAL SYSTEMS FOR SYSTEM WITHOUT ISC ASSY
REV 2 P9015EK

3–2 INSTALLATION PROCEDURE (continued)


11. Connect the modem cable to the ISC Board as shown in Illustration 3–5. Be sure to lock the modem cable slide
latch to the ISC board connector.

12. Connect the modem power supply cable to the modem.

13. Connect the phone line cord to the modem “TO LINE” port.

MODEM (REAR VIEW)

I O

TO PHONE LINE

MODEM POWER MODEM POWER SW


SUPPLY
TO AC OUTLET

ISC ASSY IN OC
PACE SERIES:SLOT 21
SYTEC SERIES:SLOT 20
Eµ/Sytec1600i/Sytec1800i: Slot 7

J3 CONNECTOR

INSITE HARDWARE INTER–CONNECT


ILLUSTRATION 3–5

- Attaching the modem


Note
For GEMS–AM pole (except Japan), skip steps 14. to 16. since the modem supplied in the kit already
has the Velcro attached to the modem and power supply.

14. Cut two 12 inch pieces and two 3 inch pieces of each of 46–136246P16 (U0032BA for Eu/Sytec1600i/Sy-
tec1800i) and 46–136246P18(U0033BA for Eu/Sytec1600i/Sytec1800i), at the total of 8 pieces.

Note
The lengths of the Velcro are based on the size of the Telebit 2000 modemand its power supply unit.
For all other modems, cut pieces according to physical size of the modem and the modem power
supply unit.

15. Mate the corresponding Velcro pieces together, two 12 inch 46–136246P16 (U0032BA for Eu/Sytec1600i/Sy-
tec1800i) pieces to 12 inch 46–136246P18 (U0033BA) pieces and two 3 inch 46–136246P16 (U0032BA) pieces
to 3 inch pieces 46–136246P18 (U0033BA).

Note
Extra Velcro is supplied in the kit to be used if necessary; discard any extra Velcro after installation.

3–4
CT INSITE KIT INSTALLATION
GE MEDICAL SYSTEMS FOR SYSTEM WITHOUT ISC ASSY
REV 2 P9015EK

3–2 INSTALLATION PROCEDURE (continued)


16. Remove the tape backing from one side of each piece of Velcro and stick them to the top of the modem and to the
top of the modem power supply unit as shown in Illustration 3–6.

12 INCH PIECES OF VELCRO


TOP OF MODEM

3 INCH PIECES OF VELCRO

TOP OF MODEM POWER SUPPLY UNIT

VELCRO POSITION
ILLUSTRATION 3–6

17. Remove the tape backing from the other side of Velcro strips and attach the modem (switches and LEDs facing
forward) and modem power supply unit to the side of the OC in the positions shown in Illustration 3–4.

18. Place the InSite cabinet ID magnets, supplied with the kit, on their respective cabinets .

19. Attach the InSite kit rating plate on the side of OC as shown in Illustration 3–4.

20. Reinstall the OC Front Cover.

21. For CT Pase/Pace Plus and Eu/Sytec1600i/Sytec1800i:


Reinstall the OC Rear Cover.

22. Connect the modem power supply cable to the AC outlet and the modem phone line cable to the phone jack.

23. Turn the modem power switch located on rear of modem ON as shown in Illustration 3–5.

24. Turn the power switch of the OC ON.

25. After proper modem initialization, verify that the CTS, DTR and MR LEDs of the modem light up.

3–5
CT INSITE KIT INSTALLATION
GE MEDICAL SYSTEMS FOR SYSTEM WITHOUT ISC ASSY
blank P9015EK

3–6
GE Medical Systems: Telex 3797371
P.O. Box 414, Milwaukee, Wisconsin 53201 U.S.A.
(Asia, Pacific, Latin America, North America)

GE Medical Systems – Europe: Telex 698626


283, rue de la Miniére, B.P. 34, 78533 Buc Cedex, France