P9015EK
Revision 4
CT
InSite Kit Installation
for system without ISC Assy
Operating Documentation
CT INSITE KIT INSTALLATION
GE MEDICAL SYSTEMS FOR SYSTEM WITHOUT ISC ASSY
REV 4 P9015EK
a
CT INSITE KIT INSTALLATION
GE MEDICAL SYSTEMS FOR SYSTEM WITHOUT ISC ASSY
REV 4 P9015EK
b
CT INSITE KIT INSTALLATION
GE MEDICAL SYSTEMS FOR SYSTEM WITHOUT ISC ASSY
REV 4 P9015EK
In performing all electrical work on these products, GE will use its own specially trained field engineers. All of GE’s electrical
work on these products will comply with the requirements of the applicable electrical codes.
The purchaser of GE equipment shall only utilize qualified personnel (i.e., GE’s field engineers, personnel of third-party
service companies with equivalent training, or licensed electricians) to perform electrical servicing on the equipment.
c
CT INSITE KIT INSTALLATION
GE MEDICAL SYSTEMS FOR SYSTEM WITHOUT ISC ASSY
REV 4 P9015EK
DAMAGE IN TRANSPORTATION
All packages should be closely examined at time of delivery. If damage is apparent, have notation “damage in
shipment” written on all copies of the freight or express bill before delivery is accepted or “signed for” by a General
Electric representative or a hospital receiving agent. Whether noted or concealed, damage MUST be reported to the
carrier immediately upon discovery, or in any event, within 14 days after receipt, and the contents and containers held
for inspection by the carrier. A transportation company will not pay a claim for damage if an inspection is not requested
within this 14 day period.
Call Traffic and Transportation, Milwaukee, WI (414) 827–3449 / 8*285–3449 immediately after damage is found. At
this time be ready to supply name of carrier, delivery date, consignee name, freight or express bill number, item
damaged and extent of damage.
Complete instructions regarding claim procedure are found in Section “S” of the Policy & Procedure Bulletins.
GE personnel, please use the GEMS CQA Process to report all omissions, errors, and defects in this documentation.
Customers, please contact your GE Sales or Service represenatives.
CAUTION
Do not use the following devices near this equipment. Use of these devices near this equipment could cause
this equipment to malfunction.
Devices which intrinsically transmit radio waves such as; cellular phone, radio transceiver, mobile radio transmitter,
radio–controlled toy, etc.
Keep power to these devices turned off when near this equipment.
Medical staff in charge of this equipment is required to instruct technicians, patients and other people who may be
around this equipment to fully comply with the above regulation.
d
CT INSITE KIT INSTALLATION
GE MEDICAL SYSTEMS FOR SYSTEM WITHOUT ISC ASSY
REV 4 P9015EK
REVISION HISTORY
PAGE REV PAGE REV PAGE REV PAGE REV PAGE REV
Title page . . . . . . . . . . 4 a to d . . . . . . . . . . . . . . 4 i................... 2 2–1 to 2–2 . . . . . . . . . . 4 Rear page
Title page rear . . blank A .................. 4 1–1 . . . . . . . . . . . . . . . . 2 3–1 to 3–5 . . . . . . . . . . 2
A
CT INSITE KIT INSTALLATION
GE MEDICAL SYSTEMS FOR SYSTEM WITHOUT ISC ASSY
blank P9015EK
B
CT INSITE KIT INSTALLATION
GE MEDICAL SYSTEMS FOR SYSTEM WITHOUT ISC ASSY
REV 2 P9015EK
TABLE OF CONTENTS
SECTION PAGE
TABLE OF CONTENTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . i
LIST OF ILLUSTRATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ii
i
CT INSITE KIT INSTALLATION
GE MEDICAL SYSTEMS FOR SYSTEM WITHOUT ISC ASSY
blank P9015EK
ii
CT INSITE KIT INSTALLATION
GE MEDICAL SYSTEMS FOR SYSTEM WITHOUT ISC ASSY
REV 2 P9015EK
SECTION 1 - INTRODUCTION
The InSite installation provides a communication network between Support Engineer at the OLC, ESC, ASC and the
CT scanner system for the purpose of servicing and supporting a CT scanner system.
An external 19.2 k baud modem is connected to the system computer through a RS–232C serial cable. The network is
completed by connecting the modem to a voice grade DID, Direct Inward Dial, phone line.
This kit is for the systems that are released without ISC Assy, that are:
D CT Pace/Pace Plus
D CT Em/Sytec1600i/Sytec 1800i
CAUTION
PREREQUISITE:
To use InSite, the following system version is required. The related FMI must be performed
before this installation if the system is not at the appropriate version as listed below.
1–1
CT INSITE KIT INSTALLATION
GE MEDICAL SYSTEMS FOR SYSTEM WITHOUT ISC ASSY
blank P9015EK
1–2
CT INSITE KIT INSTALLATION
GE MEDICAL SYSTEMS FOR SYSTEM WITHOUT ISC ASSY
REV 4 P9015EK
2–1 INSITE KIT (46–320293G1) for GEMS–AM (except Sytec 1600i/Sytec 1800i)
Item Part Number Description Quantity
1 46–3278761P2 HIGH SPEED, DIAL–UP MODEM WITH POWER SUPPLY 1
2 46–278763P1 PHONE LINE CORD – 25 FEET 1
3 46–251182P363 MR/CT INLINK RATING PLATE 1
4 46–301402P1 RESTRICTED TO GEMS LABEL 1
5 46–320357P1 MODEM CABLE, 10 FEET 1
6 46–136246P16 1” WIDE VELCRO – 5 FEET 1
7 46–136246P18 1” WIDE VELCRO – 5 FEET 1
8 2144760 ISC CIRCUIT BOARD 1
9 46–320095P9 INSITE CABINET ID MAGNET – SC 1
10 46–320095P10 INSITE CABINET ID MAGNET – GPS 1
11 46–320095P11 INSITE CABINET ID MAGNET – SXFMR 1
12 46–320095P12 INSITE CABINET ID MAGNET – TME 1
13 46–320551P1 LOCKING POST KIT 1
14 46–312343P17 M3 (0.5) x 6 SCREW 2
15 P9020EE INSITE INSTALLATION INSTRUCTION (MR) 1
16 P9015EK INSITE INSTALLATION INSTRUCTION (CT) 1
2–2 INSITE KIT (46–320293G2) for GEMS–A (except GEYMS) and GEMS–E (except Sytec 1600i/Sytec
1800i)
Item Part Number Description Quantity
1 46–278763P1 PHONE LINE CORD – 25 FEET 1
2 46–251182P364 MR/CT INLINK RATING PLATE 1
3 46–301402P1 RESTRICTED TO GEMS LABEL 1
4 46–320357P1 MODEM CABLE, 10 FEET 1
5 46–136246P16 1” WIDE VELCRO – 5 FEET 1
6 46–136246P18 1” WIDE VELCRO – 5 FEET 1
7 2144760 ISC CIRCUIT BOARD 1
8 46–320095P9 INSITE CABINET ID MAGNET – SC 1
9 46–320095P10 INSITE CABINET ID MAGNET – GPS 1
10 46–320095P11 INSITE CABINET ID MAGNET – SXFMR 1
11 46–320095P12 INSITE CABINET ID MAGNET – TME 1
12 46–320551P1 LOCKING POST KIT 1
13 46–312343P17 M3 (0.5) x 6 SCREW 2
14 P9020EE INSITE INSTALLATION INSTRUCTION (MR) 1
15 P9015EK INSITE INSTALLATION INSTRUCTION (CT) 1
Note
In addition to 46–320293G2, Modem must be prepared locally. The recommended type of the Mo-
dem is: T2000 (made by Telebit company)
2–1
CT INSITE KIT INSTALLATION
GE MEDICAL SYSTEMS FOR SYSTEM WITHOUT ISC ASSY
REV 4 P9015EK
2–2
CT INSITE KIT INSTALLATION
GE MEDICAL SYSTEMS FOR SYSTEM WITHOUT ISC ASSY
REV 2 P9015EK
SECTION 3 – INSTALLATION
ON ND 8
ON
DS2 OFF
J3
1 J1 1 2 3 4 5 6 7 8
4
SW10B1
DS1
1 8
SW NO. ON OFF
J5 1, 5 RS–232C RS–422A
1 J2
4
2, 6 DA IN USE DB IN USE
DS3
3, 7 TXC/RXC NO USE TXC/RXC IN USE
3–1
CT INSITE KIT INSTALLATION
GE MEDICAL SYSTEMS FOR SYSTEM WITHOUT ISC ASSY
REV 2 P9015EK
6. Remove the two screws securing connector J3 (25 pin Sub–D) to the ISC Board. Replace them with the new
ones (46–312343P17) supplied with the Inlink kit. See Illustraion 3–2.
7. Remove the two cable–connect–standoffs from connector J3. Replace them with the new ones (46–320551P1).
See Illustration 3–2.
Replace these
two screws.
Replace these
two cable–con- J3
nect–standoffs. CONNECTOR
ISC Board
(Soldering
side)
8. Install the ISC Board in the OC (Pace series; Slot 21, Sytec series; Slot 20, Eu/Sytec1600i/Sytec1800i: Slot 7).
AUX4 AUX
A5 B C MISC2 ISC
0
J133 J071
3–2
CT INSITE KIT INSTALLATION
GE MEDICAL SYSTEMS FOR SYSTEM WITHOUT ISC ASSY
REV 2 P9015EK
Operator Console
OC REAR DUCT
J3
ISC CONNECTOR
ASSY
MODEM (ATTACHED
WITH SWITCHES/LEDS
FACING FORWARD)
MODEM CABLE
RATING PLATE
ÀÀ
ÀÀ
REAR DUCT
Operator Console
ÀÀ
ÀÀ
MODEM POW-
ER
SUPPLY
ÀÀ
ÀÀ
ÀÀ TO AC OUTLET
ÀÀ
J3
CONNECTOR ISC
ÀÀ
ASSY
RATING PLATE
Eµ/Sytec1600i/Sytec1800i
MODEM (ATTACHED
WITH SWITCHES/LEDS
MODEM CABLE FACING FORWARD)
3–3
CT INSITE KIT INSTALLATION
GE MEDICAL SYSTEMS FOR SYSTEM WITHOUT ISC ASSY
REV 2 P9015EK
13. Connect the phone line cord to the modem “TO LINE” port.
I O
TO PHONE LINE
ISC ASSY IN OC
PACE SERIES:SLOT 21
SYTEC SERIES:SLOT 20
Eµ/Sytec1600i/Sytec1800i: Slot 7
J3 CONNECTOR
14. Cut two 12 inch pieces and two 3 inch pieces of each of 46–136246P16 (U0032BA for Eu/Sytec1600i/Sy-
tec1800i) and 46–136246P18(U0033BA for Eu/Sytec1600i/Sytec1800i), at the total of 8 pieces.
Note
The lengths of the Velcro are based on the size of the Telebit 2000 modemand its power supply unit.
For all other modems, cut pieces according to physical size of the modem and the modem power
supply unit.
15. Mate the corresponding Velcro pieces together, two 12 inch 46–136246P16 (U0032BA for Eu/Sytec1600i/Sy-
tec1800i) pieces to 12 inch 46–136246P18 (U0033BA) pieces and two 3 inch 46–136246P16 (U0032BA) pieces
to 3 inch pieces 46–136246P18 (U0033BA).
Note
Extra Velcro is supplied in the kit to be used if necessary; discard any extra Velcro after installation.
3–4
CT INSITE KIT INSTALLATION
GE MEDICAL SYSTEMS FOR SYSTEM WITHOUT ISC ASSY
REV 2 P9015EK
VELCRO POSITION
ILLUSTRATION 3–6
17. Remove the tape backing from the other side of Velcro strips and attach the modem (switches and LEDs facing
forward) and modem power supply unit to the side of the OC in the positions shown in Illustration 3–4.
18. Place the InSite cabinet ID magnets, supplied with the kit, on their respective cabinets .
19. Attach the InSite kit rating plate on the side of OC as shown in Illustration 3–4.
22. Connect the modem power supply cable to the AC outlet and the modem phone line cable to the phone jack.
23. Turn the modem power switch located on rear of modem ON as shown in Illustration 3–5.
25. After proper modem initialization, verify that the CTS, DTR and MR LEDs of the modem light up.
3–5
CT INSITE KIT INSTALLATION
GE MEDICAL SYSTEMS FOR SYSTEM WITHOUT ISC ASSY
blank P9015EK
3–6
GE Medical Systems: Telex 3797371
P.O. Box 414, Milwaukee, Wisconsin 53201 U.S.A.
(Asia, Pacific, Latin America, North America)