GE Healthcare

Technical
Publications

Direction 2370115-100
Revision 1

GEHC 0.2T Profile Magnet


Installation Manual

For The M20552PY, M20082PM


& M20092PY Permanent Magnet Installation Only

Copyright© 2003 - 2004 by GE HealthCare

Operating Documentation
GE Healthcare GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

DAMAGE IN TRANSPORTATION

All packages should be closely examined at time of delivery. If damage is


apparent, have notation “damage in shipment” written on all copies of the
freight or express bill before delivery is accepted or “signed for” by a General
Electric representative or a hospital receiving agent. Whether noted or
concealed, damage MUST be reported to the carrier immediately upon
discovery, or in any event, within 14 days after receipt, and the contents and
containers held for inspection by the carrier. A transportation company will not
pay a claim for damage if an inspection is not requested within this 14-day
period.

If shipment was handled by moving van service uncrate, call Traffic Milwaukee
immediately when any damage is found. At this time be ready to describe type
of damage, type of equipment, serial numbers and if possible the order number.

The above paragraph is in regard to equipment requiring installation only, and


does not apply to supply items.

20-September-2003

GE Healthcare
GE Healthcare: Telex 3797371
P.O. Box 414, Milwaukee, Wisconsin 53201 U.S.A.
(Asia, Pacific, Latin America, North America)

GE HEALTHCARE — Europe: Telex 698626


283, rue de la Minere, B.P.34, 78533 Buc Cedex
France
TO GEHC: TECHNICAL PUBLICATION GROUP – MR
FAX NO: 86–510–522–6688
REPORT ON TECHNICAL PUBLICATIONS – MR
(OMISSION, ERRORS, SUGGESTIONS)

PRODUCT PUBLICATION # .
PAGE # PAGE REV DATE REPORTED .
SERVICE RECORD SUBMITTED □ YES □ NO
PLEASE BE SPECIFIC IN YOUR CORRECTIONS AND SUGGESTIONS

IF ADDITIONAL PAGES ARE INCLUDED, ATTACH THEM TO THIS SHEET.

SERVICE REP
(Please Print)

PHONE
FAX
COUNTRY
CITY

REV 5

DIRECTION NO. 15214


GE HEALTHCARE
TECHNICAL PUBLICATION GROUP – MR

19# Changjiang Road


Wuxi, Jiangsu Province, 214028, P.R.China
FAX No. 86 - 510 - 5226688

(TEAR AT PERF)
GE Healthcare GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

WARNING
THIS SERVICE MANUAL IS AVAILABLE IN ENGLISH ONLY.
IF A CUSTOMER’S SERVICE PROVIDER REQUIRES A LANGUAGE
OTHER THAN ENGLISH, IT IS THE CUSTOMER’S RESPONSIBILITY
TO PROVIDE TRANSLATION SERVICES.
DO NOT ATTEMPT TO SERVICE THE EQUIPMENT UNLESS THIS
SERVICE MANUAL HAS BEEN CONSULTED AND IS UNDERSTOOD.
FAILURE TO HEED THIS WARNING MAY RESULT IN INJURY TO
THE SERVICE PROVIDER, OPERATOR OR PATIENT FROM
ELECTRIC SHOCK, MECHANICAL OR OTHER HAZARDS.

AVERTISSEME

CE MANUEL DE MAINTENANCE N’EST DISPONIBLE QU’EN


ANGLAIS.
SI LE TECHNICIEN DU CLIENT A BESOIN DE CE MANUEL DANS UNE
AUTRE LANGUE QUE L’ANGLAIS, C’EST AU CLIENT QU’IL
INCOMBE DE LE FAIRE TRADUIRE.
NE PAS TENTER D’INTERVENTION SUR LES ÉQUIPEMENTS TANT
QUE LE MANUEL SERVICE N’A PAS ÉTÉ CONSULTÉ ET COMPRIS.
LE NON-RESPECT DE CET AVERTISSEMENT PEUT ENTRAÎNER
CHEZ LE TECHNICIEN, L’OPÉRATEUR OU LE PATIENT DES
BLESSURES DUES À DES DANGERS ÉLECTRIQUES, MÉCANIQUES
OU AUTRES.

WARNUNG
DIESES KUNDENDIENST–HANDBUCH EXISTIERT NUR IN
ENGLISCHER SPRACHE.
FALLS EIN FREMDER KUNDENDIENST EINE ANDERE SPRACHE
BENÖTIGT, IST ES AUFGABE DES KUNDEN FÜR EINE
ENTSPRECHENDE ÜBERSETZUNG ZU SORGEN.
VERSUCHEN SIE NICHT, DAS GERÄT ZU REPARIEREN, BEVOR
DIESES KUNDENDIENST–HANDBUCH NICHT ZU RATE GEZOGEN
UND ERSTANDEN WURDE.
WIRD DIESE WARNUNG NICHT BEACHTET, SO KANN ES ZU
VERLETZUNGEN DES KUNDENDIENSTTECHNIKERS, DES
BEDIENERS ODER DES PATIENTEN DURCH ELEKTRISCHE
SCHLÄGE, MECHANISCHE ODER SONSTIGE GEFAHREN KOMMEN.
GE Healthcare GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

ATENÇÃO
 
ESTE MANUAL DE SERVICIO SÓLO EXISTE EN INGLÉS.
SI ALGÚN PROVEEDOR DE SERVICIOS AJENO A GEHC SOLICITA
UN IDIOMA QUE NO SEA EL INGLÉS, ES RESPONSABILIDAD DEL
CLIENTE OFRECER UN SERVICIO DE TRADUCCIÓN.
NO SE DEBERÁ DAR SERVICIO TÉCNICO AL EQUIPO, SIN HABER
CONSULTADO Y COMPRENDIDO ESTE MANUAL DE SERVICIO.
LA NO OBSERVANCIA DEL PRESENTE AVISO PUEDE DAR LUGAR A
QUE EL PROVEEDOR DE SERVICIOS, EL OPERADOR O EL
PACIENTE SUFRAN LESIONES PROVOCADAS POR CAUSAS
ELÉCTRICAS, MECÁNICAS O DE OTRA NATURALEZA.

AVVERTENZA

ESTE MANUAL DE ASSISTÊNCIA TÉCNICA SÓ SE ENCONTRA


DISPONÍVEL EM INGLÊS.
SE QUALQUER OUTRO SERVIÇO DE ASSISTÊNCIA TÉCNICA, QUE
NÃO A GEHC, SOLICITAR ESTES MANUAIS NOUTRO IDIOMA, É DA
RESPONSABILIDADE DO CLIENTE FORNECER OS SERVIÇOS DE
TRADUÇÃO.
NÃO TENTE REPARAR O EQUIPAMENTO SEM TER CONSULTADO
ECOMPREENDIDO ESTE MANUAL DE ASSISTÊNCIA TÉCNICA.
NÃO CUMPRIMENTO DESTE AVISO PODE POR EM PERIGO A
SEGURANÇA DO TÉCNICO, OPERADOR OU PACIENTE DEVIDO
A‘ CHOQUES ELÉTRICOS,MECÂNICOS OU OUTROS.

AVISO
IL PRESENTE MANUALE DI MANUTENZIONE È DISPONIBILE
SOLTANTO IN INGLESE.
SE UN ADDETTO ALLA MANUTENZIONE ESTERNO ALLA GEHC
RICHIEDE IL MANUALE IN UNA LINGUA DIVERSA, IL CLIENTE È
TENUTO A PROVVEDERE DIRETTAMENTE ALLA TRADUZIONE.
SI PROCEDA ALLA MANUTENZIONE DELL’APPARECCHIATURA
SOLO DOPO AVER CONSULTATO IL PRESENTE MANUALE ED
AVERNE COMPRESO IL CONTENUTO.
NON TENERE CONTO DELLA PRESENTE AVVERTENZA POTREBBE
FAR COMPIERE OPERAZIONI DA CUI DERIVINO LESIONI
ALL’ADDETTO ALLA MANUTENZIONE, ALL’UTILIZZATORE ED AL
PAZIENTE PER FOLGORAZIONE ELETTRICA, PER URTI
MECCANICI OD ALTRI RISCHI.
GE Healthcare GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

警 告

注意
本安装手册仅存英文本。
非 GEHC 公司的安装/维修人员要求非英文本的手册时,客户需自行负责翻
译。
未详细阅读和完全了解本手册之前,不得进行安装/维修。
忽略本文注意事项会对安装人员, 操作维修人员或病人造成触电,机械或
其他伤害。
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

REVISION HISTORY
REV . DATE PRIMARY REASON FOR CHANGE

1 ……… Dec. 15th, 2004 …… Initial release.

1
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

CAUTION
DO NOT USE THE FOLLOWING DEVICES NEAR THE MAGNET. THE USE OF
THESE DEVICES NEAR THE MAGNET MIGHT BE DAMAGED BY, OR
CAUSES DAMAGES TO THE MAGNET.

DEVICES NOT TO BE USED NEAR THIS EQUIPMENT

DEVICES WHICH INTRINSICALLY TRANSMIT RADIO WAVES SUCH AS:


CELLULAR PHONE, MOBILE RADIO TRANSMITTER, RADIO TRANSCEIVER,
RADIO CONTROLLED TOY, ECT.

KEEP POWER TO THESE DEVICES TURN OFF WHEN NEAR THE MAGNET.

MEDICAL STAFF IN CHARGE OF THIS EQUIPMENT IS REQUIRED TO


INSTRUCT TECHNICIANS, PATIENTS AND OTHER PEOPLE WHO MAY BE
AROUND THIS EQUIPMENT TO FULLY COMPLY WITH THE ABOVE
REGUALTION.

Warning
Never use the following devices near this equipment. They
may cause erratic function of this equipment.

Cellular phone, transceiver, mobile radio transmitter and


radio-controlled toy, etc

Keep those devices power-off near this equipment.

a
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

INTRODUCTION
This manual contains the procedure and information necessary for
unpacking a GEHC M20552PY 0.2T Profile Permanent Magnet (in the
following part of this manual, all the “magnet” refers to the GEHC
M20552PY 0.2T Profile Permanent Magnet) and installing the magnet
into the scan room. The followings are instructed in this manual:

Magnet transportation and storage.

Magnet rigging and delivering to the scan room.

Magnet leveling.

Magnetic Field Shimming.

All preliminary site plans and concepts must be reviewed by GE


Healthcare MR Installation Group prior to installation.

Unless specifically mentioned, GE Healthcare Group does not provide


the rigging and leveling tools or equipments, nor install facility input
power lines to the power distribution panel or the power lines required
in the magnet room, raised flooring, conduit, junction boxes, ducts,
plumbing, or RF shielded room illustrated in the document. This work
should be performed by licensed contractors.

The products involved are highly sophisticated, and special


engineering competence is required. During unpacking and
delivering the magnet, operators should be certified and checked for
any medical and physical contraindication to this operation, and this
should be mandatorily performed by licensed contractors. In
performing shimming work on the magnet, GE will use its own
specially trained field engineers. The purchaser of GE equipment shall
only utilize qualified personnel (i.e., GE’s field engineers, personnel of
third–party service companies with equivalent training) to perform
shimming servicing on the equipment.

All the contents of this Magnet Installation Manual are used for the
GEHC 0.2T Profile Permanent Magnet only.

All works must be in compliance with national and local building and

b
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

safety codes.

c
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

TABLE OF CONTENT
REVISION HISTORY ............................................................................................. 1

SECTION 1. MAGNET COMPONENTS ............................................................... 1

1-1. COMPONENTS OF THE PERMANENT MAGNET......................................... 1

1-2. MAGNET ACCESSORIES ................................................................................. 2

1-3. INSTALLATION EQUIPMENT ......................................................................... 3

1-4. MAGNET SAFETY CONSIDERATIONS ......................................................... 4

SECTION 2. MAGNET SHIPPING AND STORING............................................... 7

2-1. MAGNET SHIPPING ........................................................................................... 7

2-2. STORAGE REQUIREMENTS............................................................................ 8

2-4. MAGNET DATA ............................................................................................... 11

2-5. MAGNETIC FRINGE FIELD MAP DATA.............................................. 12

SECTION 3. MAGNET RIGGING AND DELIVERING......................................... 14

3-1. SAFETY CONSIDERATIONS .......................................................................... 14

3-2. MAGNET RIGGING AND DELIVERING ..................................................... 15

SECTION 4. MAGNET LEVELING..................................................................... 30

4-1 CAUTIONS IN LEVELING THE MAGNET. .................................................... 30

4-2 MAGNET LEVELING PROCEDURE FLOWCHART..................................... 31

4-3 MAGNET LEVELING ....................................................................................... 32

SECTION 5. MAGNETIC FIELD SHIMMING....................................................... 36

5-1 MAGNETIC FIELD SHIMMING PREPARATION........................................... 36

5-2 FIELD SHIMMING SPECIFICATIONS ........................................................... 39

5-3. MAGNETIC FIELD SHIMMING PROCEDURES .......................................... 40

5-4 FIELD SHIMMING SOFTWARE INTRODUCTION ....................................... 45

I
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

SECTION 1. MAGNET COMPONENTS

1-1. COMPONENTS OF THE PERMANENT MAGNET

The Permanent Magnet consists of the following major components.

1. 2 kilogauss (0.2 Tesla) permanent magnet.

2. Magnet peripherals and accessories.

The key magnet peripherals include magnet Heater, Thermal Insulator, RF Shield,
Shim Holder and Gradient Coil. And all the peripherals have been installed well before
shipping the magnet to the customer.

1
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

1-2. MAGNET ACCESSORIES

The followings are some of the key accessories of the GEHC 0.2T Profile Permanent Magnet.
For the accessories detail, please refer to the magnet shipping collector checklist.

1· Hard Rubber Pedestal Plate: This will go under the magnet pedestals.

Material: Natural Rubber


Number: 4 pieces
Thickness: 20mm

2· Stainless Leveling Plate: Used for Leveling of the Magnet.

1mm x 6 pieces
2mm x 6 pieces

3· Tools for Mechanical Shimming:

T-Wrench
Shaft
Extension Rod
Handle

4· Shim Assy:
Shimming kits
Shimming Plate
Shimming Holders

5· Permanent Magnet Installation Manual:


Including instructions for magnet shipping/storage, unpacking leveling and field shimming.

6· Locator Card.

7· Warning Labels: 3 pieces

8· Packing Check List.

2
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

1-3. INSTALLATION EQUIPMENT

Besides the magnet enclosures and accessories, special equipments which should be
prepared by the licensed installation contractors are also needed for a successful magnet
installation. The following equipments are some of the key equipments must be prepared for
the magnet installation, while they are not all the equipments might be used in the installation.

1. Ramp for removing cabinets from pallets.

2. Wrecking bar and claw hammer.

3. Magnet moving fixture for lifting and moving the magnet.

4. Four dual–life ratchet jacks and handles. Five ton (4,536 kg) capacity, 368 mm travel,
maximum starting height 38 mm toe, 533 mm head.

5. Free standing posts, plastic chain and warning signs for roping off site during
installation and service activity.

6. Fixture for field plotting equipment (SV calibration kit).

7. Aluminized polyester tape.

8. Clear plastic hose with 3m of length for magnet leveling.

9. Non-ferrous Wire Rope.

WARNING

THE STRONG MAGNETIC FIELD COULD CAUSE MALFUNCTION OR DAMAGE TO THE


MAGNETIC SENSITIVE DEVICES; TOOLS AND EQUIPMENTS MADE FROM FERROUS
MATERIALS COULD BE ATTRACTED BY THE STRONG MAGNETIC FIELD AND FLY TO THE
MAGNET. THESE KIND OF FLYING OBJECTS COULD CAUSE SERIOUS INJURIES TO THE
PEOPLE AROUND THE MAGNET. DURING THE MAGNET INSTALLATION PROCESS, THESE
KIND OF DANGERS SHOULD TRY TO BE AVOID WITH THE BEST EFFORT.

3
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

1-4. MAGNET SAFETY CONSIDERATIONS

WARNING

THERE IS A MAGNETIC FIELD PRESENT AROUND THE MAGNET AT TIME OF DELIVERY,


EVEN WHEN THE MAGNET IS FIXED INSIDE THE SHIPPING CRATE. UP MOST CARES
MUST BE TAKEN WHEN WORKING AROUND THE MAGNET. BE SURE TO KEEP ALL
FERROUS OBJECTS AND EQUIPMENT THAT MAY BE INFLUENCED BY THE MAGNET
AWAY FROM THE MAGNET. DETAILS ARE ILLUSTRATED IN TABLE 1-1.

WARNING

THE MAGNETIC AND ELECTRO-MAGNETIC FIELD CAN CAUSE MALFUNCTION OF ACTIVE


IMPLANTS, SUCH AS CARDIAC PACEMAKER, NEUROSTIMULATORS, AND
BIOSTIMULATION DEVICES. PERSONS WITH CARDIAC PACEMAKERS,
NEUROSTIMULATORS AND BIOSTIMULATION DEVICES MUST REMAIN OUTSIDE THE
RECOMMENDED FIVE GAUSS EXCLUSION ZONE AS ILLUSTRATED IN TABLE 1-1.

WARNING

THE STRONG FORCE GENERATED BY MAGNETIC FIELD CAN CAUSE DISPLACEMENT OF


INTRACRANIAL ANEURYSM CLIPS IN PATIENTS’ BRAIN, AS WELL AS ANY OTHER METAL
PROSTHESIS. PERSONS WITH THESE KIND DEVICES MUST REMAIN OUTSIDE THE
RECOMMENDED FIVE GAUSS EXCLUSION ZONE AS ILLUSTRATED IN TABLE 1-1.

WARNING

THE STAFF, WHO WILL BE INVOLVED IN THE MAGNET OPERATIONS, MUST BE CHECKED
FOR ANY MEDICAL / PHYSICAL CONTRAINDICATION TO THIS OPERATION (AS
DESCRIBED ABOVE), AND THIS SHOULD BE PERFORMED BY LICENSED CONTRACTORS,
AS WELL AS MEDICAL STAFF IN CHARGE OF THIS EQUIPMENT.

WARNING

THE 5 GAUSS LINE WIDENS ABOUT 1m WHEN THE CRATE IS OPENED AND THE MAGNET
IS RIGGED OUT FROM THE SHIPPING CRATE. BE CAUTIONS ABOUT THIS CHANGE IN
THE SCOPE OF THE MAGNETIC FRANGE FILED.

4
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

CAUTION
DO NOT USE THE FOLLOWING DEVICES NEAR THE MAGNET. USE OF THESE
DEVICES NEARS THE MAGNET MIGHT BE DAMAGED BY, OR CAUSE DAMAGES TO
THE MAGNET.

DEVICE WHICH INTRINSICALLY TRANSMIT RADIO WAVES SUCH AS: CELLULAR


PHONE, MOBILE RADIO TRANSMITTER, RADIO TRANSCEIVER, RADIO CONTROLLED-
--TOY, ETC. KEEP POWER TO THESE DEVICES TURN OFF WHEN NEAR THE MAGNET.

PLEASE CLEAR FERROUS OBJECTS OUT OF THE DELIVERY ROUTINE AND SCAN
ROOM BEFORE DELIVERING THE MAGNET TO THE SCAN ROOM. THESE FERROUS
OBJECTS COULD CAUSE DAMAGES TO THE MAGNET, OR INJURIES TO PERSON
AROUND THE MAGNET. AT THE SAME TIME, PREVENT PEOPLE WITH MEDICAL AND
PHYSICAL CONTRAINDICATIONS FROM ENTERING THE MAGNET OPERATION AREA.

CONSIDERATION MUST BE GIVEN TO THE MAGNET DELIVERY ROUTE TO ENSURE


THAT THE FLOOR CAN SUPPORT THE WEIGHT OF THE MAGNET, TOGETHER WITH
ANY RIGGING/DELIVERING EQUIPMENT REQUIRED TO MOVE IT. A STRUCTURAL
ANALYSIS SHOULD BE PERFORMED BY A PROFESSIONAL STRUCTURAL ENGINEER.

SPECIAL RIGGING EQUIPMENT MUST BE CONSIDERED IF THE MAGNET IS TO BE


LOWERED TO A DIFFERENT FLOOR LEVEL THAN THE RECEIVING LOCATION LEVEL.

THE TEMPERATURE FOR THE MAGNET IN THE DELIVERY PROCESS MUST BE


BETWEEN –30℃ AND 45℃ AND HUMIDITY UP TO 80%. IF MAGNET TEMPERATURE
HAS EXCEEDED 45℃, PLEASE CONTACT THE LOCAL SERVICE SUPPORT CENTER.

GAUSS (mT) LIMIT EQUIPMENT


(See Notes 1&2)
0.5 GAUSS(0.05mT) OR LESS Nuclear cameras
1 GAUSS(0.1mT) OR LESS Positron Emission Tomography scanner
Video display (color, B/W, monochrome)
Linear Accelerator CT scanner
Cyclotrons Ultrasound
Accurate Measuring scale Lithotriptor
Image intensifiers Electron microscope
Color TV/Color Monitor( without shielding)
Advantage Windows Workstation
3 GAUSS(0.3mT) OR LESS
5 GAUSS(0.5mT) OR LESS Cardio pacemakers Bio stimulation devices
Neurostimulators Uncontrolled public and staff access
10 GAUSS(1mT) OR LESS Shielded Color Monitor for Profile II
30 GAUSS(3mT) OR LESS MR equipment (Scan Control, Scan Room Unit, System
Gradient Unit, Step Down Transformer
50 GAUSS(5mT) OR LESS MR equipment (Regulated Step Down Telephones
Transformer or Main Disconnect Device Metal detector for screening
(in U.S.A). )
100 GAUSS(10mT) OR LESS Service Tool Magnet Power Supply Cabinet Smoke/heat detector
SM Laser Camera (without film processor)

5
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

Note 1 Refer to SECTION 3-2-2 for magnetic field plots.


2 Recommended limits given above are based on general MR site planning guidelines. Actual susceptibility of
specific devices may vary significantly depending on electrical design orientation of the device relative to the
magnetic field and the degree of interference considered unacceptable.
3 Verify operating limits with provider of helicopter service.

Table 1-1 MAGNETIC SENSITIVE DEVICE PROXIMITY LIMITS

6
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

SECTION 2. MAGNET SHIPPING AND STORING

2-1. MAGNET SHIPPING


Domestic transportation for the permanent magnet will be via spring–ride moving van. Export
transportation for the MR system overseas will be via air shipment in a pressurized cargo hold,
and the magnet via boat within an open-top ocean container.

MAGNET HANDLING INFORMATION:


MAXIMUM WEIGHT: 12000KG (magnet and shipping crate)
MAXIMUM GRADIENT: 30 °
MAXIMUM SHOCK LOADS: 7 G
MAXIMUM TEMPERATURE: 45℃

If the magnet is shipped to the hospital within the open-top ocean container, the first step
should be unpacking the shipping crate from the ocean container. Open the top cover of the
shipping ocean container and lift the shipping crate out by crane. Please prepare the crane
for this kind of operation.

NOTE: PLEASE PREPARE THE CRANE WITH ENOUGH CAPABILITY TO LIFT THE
SHIPPING CRATE OUT. PLEASE REFER TO SECTION 2-3 FOR DETAIL
INFORMATION ABOUT THE SHIPPING CRATE DATA. THE DIMENSION OF
THE OCEAN CONTAINER IS: WIDTH 2330MM, LENGTH 5900MM AND HEIGHT
2245MM.

If the shipping crate is unpacked from the ocean container at the seaport, the magnet should
be shipped to the hospital within the shipping crate only. The crane should also be prepared
to move the magnet within the shipping crate to the ground for a more safety magnet shipping
crate unpacking.

When lifting the shipping crate, please attach the stainless wire rope at the indicated point.
See illustration 2-1.

7
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

LIFT THE SHIPPING CRATE


Illustration 2-1

2-2. STORAGE REQUIREMENTS

If the system needs to be stored before installation, it must be stored in a warehouse, covered
by waterproof materials and protected from weather. The storage temperature for the system
should be between –34 ℃and 60℃ and the relative humidity between 0 and 90 % (non–
condensing).

NOTE: THE TEMPERATURE FOR THE MAGNET TRANSPORTATION AND STORAGE


MUST BE BETWEEN –30℃AND 45℃ AND HUMIDITY UP TO 80%.

WARNING

IF THE MAGNET TEMPERATURE IS OUT OF THE RANGE FROM – 30℃ TO 45℃, THE
MAGNETIC PROPERTY COULD HAVE IRREVERSIBLE DAMAGE. UPTO MOST CARES
SHOULD BE PAID TO THE MAGNET TRANSPORATION AND STORAGE TEMPERATURE.

8
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

2-3. SHIPPING CRATE DATA

Basic data of the shipping crate.


L x W x H: 2400mm x 2100mm x 1950mm
Weight: 1000kg (weight of the shipping crate only)

Refer to the Illustration 2-1


2100

2400 1950

1MM thick steel

9
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

2100 MM

2400MM

MAGNET SHIPPING CRATE


Illustration 2-1

To lift the shipping crate, please attach the stainless wire to the position, and just the outside
of the supporting squire bar as shown in Illustration 2-2.

APPLY THE WIRE ROPE AT THE CORRECT POSITION


Illustration 2-2

10
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

2-4. MAGNET DATA

L x W x H: 2040mm x 1300mm x 1380mm


Weight: 11000 Kg

Refer to the Illustration 2-3 for magnet dimension data in details.

2040

MAGNET DATA
Illustration 2-3

11
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

2-5. MAGNETIC FRINGE FIELD MAP DATA

This section provides the magnetic fringe field layout of the GEHC 0.2T Profile magnet before
being unpacked out from the shipping crate.

WARNING

UP MOST CARES MUST BE TAKEN WHEN WORKING AROUND THE MAGNET. BE SURE TO
KEEP ALL FERROUS OBJECTS, EQUIPMENT AND PEOPLE WITH MAGNETIC FIELD
SENSITIVE IMPLANTS OUT OF THE MAGNETIC FIELD. DETAILS ARE ILLUSTRATED IN
TABLE 1-1.

X-Y magnetic fringe field, with the magnetic field shield from the shipping crate, as shown in
illustration 2-4.

Illustration 2-4

12
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

WARNING

THE 5 GAUSS LINE WIDENS ABOUT 1m WHEN THE CRATE IS OPENED. BE SURE TO
KEEP ALL FERROUS OBJECTS AND EQUIPMENTS THAT MAY BE INFLUENCED BY THE
MAGNETIC FIELD, AS WELL AS PEOPLE WITH MAGNETIC FIELD SENSITIVE IMPLANTS,
AWAY FROM THE MAGNET.

Y-Z magnetic fringe field, with the magnetic field shield from the shipping crate, as shown in
illustration 2-5
Y

Illustration 2-5

13
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

SECTION 3. MAGNET RIGGING AND DELIVERING

The procedures of unpacking and rigging the magnet out from the shipping crate, as well as
the procedures of delivering the magnet into the scan room are described in this section. Due
to the strong magnetic field, up most care should be paid to all the operations instructed in this
chapter. And all the procedures are used for GEHC 0.2T Profile M20552PY magnet only.

HANDLING INFORMATION:

MAXIMUM WEIGHT: 11000KG (without shipping crate)


MAXIMUM TILT: 30 DEGREES
MAXIMUM SHOCK LOADS: 7G
MAXIMUM TEMPERATURE: -30℃ ~ 45℃
SIZE OF THE CRATE: 2400mm x 2100mm x 1950mm ( LxWxH)
SIZE OF THE MAGNET: 2040mm x 1300mmx 1380mm ( LxWxH)

3-1. SAFETY CONSIDERATIONS

This section contains safety measures that must be done when rigging the permanent magnet.
All these requirements must be done and checked before and during magnet rigging and
delivering operation.

The magnetic field could cause damages to the other devices (please refer to table 1-1). Be
sure to take actions to avoid all these potential damages.

There may be possibility that ferromagnetic tools may be get attracted to the magnet and
cause injury. Use nonmagnetic tools when working around the magnet. If nonmagnetic tools
(such as a bar or jack) cannot be prepared, up most care must be taken to make sure they do
not get drawn to the magnet, and put these tools far away (4m) from the magnet after the
usage.

DO NOT use steel wire to lift the Magnet. Use approved non–magnetic lifting straps or wire
rope.
DO NOT take off the OVERALL MASK covering the magnet before the magnet is placed in
the correct position in the scan room for leveling.

14
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

3-2. MAGNET RIGGING AND DELIVERING

3-2-1. Check the Shock Indicator and Tip Indicator.

The Shock Indicator and Tip Indicator are pasted on the surface of the shipping crate. The
position and the examples are shown in the illustration 3-1(a).

Tilt Indicator

Shock Indicator

a.

b. c.
THE SHOCK INDICATOR AND TIP INDICATOR
Illustration 3-1
The Shock Indicator, as shown in illustration 3-1(b), is used to indicate whether the maximum
shock load is out of the 10G, which is specified in the magnet transportation requirements.
The Tilt Indicator, please refer to illustration 3-1(b), is used to show whether the maximum tilt
is out of the 30° specification set to the shipment.

15
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

NOTE: IF INDICATORS SHOW THAT THE SHOCK LOAD OR TILT DURING THE
TRANSPORTATION PROCESS ARE OUT OF THE SPECIFICATION, PLEASE
CONTACT THE GE REPRESENTATIVE OR LOCAL SERVICE SUPPORT
CENTER IMMEDIATELY.

3-2-2. Open the Shipping Crate.

Connect crane (forklift) to the rings in top of the crate and open the top plate of the shipping
crate. See Illustration 3-2.

OPEN THE SHIPPING CRATE


Illustration 3-2

NOTE: THE FOUR IRON RINGS WELDED ON THE TOP OF THE SHIPPING COVER IS
DESIGNED FOR LIFTING THE TOP COVER ONLY. DO NOT USE THESE RINGS
TO LIFT THE WHOLE SHIPPING CRATE.

NOTE: THERE IS A 3MM STEEL PLATE ATTACHED TO THE TOP OF THE CRATE. THE
WEIGHT OF THE SHIPPING CRATE TOP COVER IS ABOUT 150KG.

16
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

NOTE: IF THE WIRE ROPE CONTAINS FEROMAGNETIC MATERIAL, PAY GREAT


ATTENTION TO PREVENT WIRE ROPE FROM BEING ATTRACTED TO THE
MAGNET.

There are another 4 screws used to fix the crossing bar inside of the shipping crate. Please
release these 4 screws also before open the shipping crate sidewalls, refer to Illustration 3-3
(a). And Release screws used to fix the shipping crate and remove the 4 sidewalls of the
crate. See Illustration 3-3 (b). The Illustration 3-4 shows the unpacked shipping crate with
the magnet covered by the Aluminum waterproof membrane. Cut open this waterproof
aluminum membrane at the bottom seal edge carefully and remove this coverage. The
vacuum coverage should be removed immediately after the unpacking the shipping crate.

WARNING

PLEASE CONFIRM AGAIN THAT THERE ARE NO FERROUS OBJECTS AROUND (2.5
METERS) THE MAGNET BEFORE OPEN THE SHIPPING CRATE.

PLEASE USE THE NON-MAGNETIC SENSITIVE TOOLS TO CUT THE VACUUM PACKAGE
OPEN.

IF THERE ARE ANY OBJECTS ATTRACTED TO THE MAGNET, PLEASE GET IN TOUCH
WITH SERVICE REPRESENTITIVE IMMEDIATELY. DO NOT TRY TO TAKE THE OBJECT
OFF FROM THE MAGNET BY HAND; DO NOT TRY TO TAKE THE OBJECT OFF FROM THE
MAGNET BY YOURSELF.

Release the screws (x4) Remove the Crossing Bars (x2)

(a)

17
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

Sidewall Screws: 44
Crossing Bar Screws, Screws material: Steel
2x2=4 Material Steel

(b)

REMOVE THE SCREWS IN THE EDGE OF THE SHIPPING CRATE


Illustration 3-3

NOTE: THE MATERIAL OF THE SCREWS IS CABON STEEL WHICH COULD BE


ATTRACTTED TO THE MAGNET AND CAUSE DAMAGES AND INJURIES. THESE
SCREWS MUST BE COLLECTED AND PLACED FAR AWAY FROM THE MAGNET
BEFORE OPEN THE SHIPPING CRATE. THE EYEBOLTS IS MADE BY NON-
MAGNETIC FIELD SENSITIVE STAINLESS STEEL.

THE UNPACKED SHIPPING CRATE ---WATERPROOF ALUMINUM MEMBRANE

18
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

OVERALL MASK

EYE BOLTS

RING PROTECTOR SET

MAGNET COVERED BY THE OVERALL MASK


Illustration 3-4

NOTE: THE MAGNET IS COVERED BY AN OVERALL MASK WHICH IS MADE OF STRONG


CLOUTH AND NON-MAGNETIC FIELD SENSITIVE STAINLESS STEEL MASH. THE
MAGNET RING IS COVERED BY THE RING PROTECTOR SET SET. THE
OVERALL MASK AND THE RING PROTECTOR SET ARE USED TO PROTECT THE
MAGNET FROM DAMAGING BY THE OBJECTS ATTRACTED BY THE MAGNETIC
FORCE. DO NOT TAKE OFF THE OVERAL MASK AND THE PROTECTOR
DURING ALL THE RIGGING AND DELIVBRING OPERATION. AFTER THE
SYSTEM INSTALLATION ACCOMPLISHED, PLEASE PRESERVE THE OVERALL
MASK AND THE RING PROTECTOR SET PROPERLY FOR THE FUTURE USAGE,
SUCH AS IN THE MAITENANCE.

WARNING

THE 5 GAUSS LINE WIDENS ABOUT 1m WHEN THE CRATE IS OPENED. BE SURE TO
KEEP ALL FERROUS OBJECTS AND EQUIPMENTS THAT MAY BE INFLUENCED BY THE
MAGNETIC FIELD, AS WELL AS PEOPLE WITH MAGNETIC FILED SENSITIVE IMPLANTS,
AWAY FROM THE MAGNET. REFER TO ILLUSTRATION 3-5, THE MAGNETIC FRINGE FIELD
WITHOUT THE MAGNETIC SHIELDING FROM THE SHIPPING CRATE.

19
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

MAGNETIC FIELD VIEW: Y-Z

MAGNETIC FIELD VIEW: X-Y

20
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

MAGNETIC FIELD VIEW: X-Z

MAGNETIC FRINGE FIELD PLOT


( without the shielding from the shipping crate)
Illustration 3-5

3-2-3.Check the temperature indicators.

Open the zippers and check the temperature indicators pasted on the magnet. Confirm
whether the magnet temperature exceeds 45℃ or not.

The Illustration 3-6 indicates the position of the zippers.

21
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

Side wall of the crate

Open the zippers and check


the temperature indicators

TEMPERATURE CHECKING AND EYE BOLTS INSTALLATION


Illustration 3-6

The temperature indicators record the highest temperature ever reached in the shipping
process. Individual indicator changes color from silver white, orange to black correspondingly
according to the different temperature. For example, as shown in illustration 3-7, the
temperature indicator shows that the highest temperature has exceeded 47℃ in the shipping
process.

Temperature indicator

The highest temperature ever


reached is 47℃

TEMPERATURE INDICATOR
Illustration 3-7,

22
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

NOTE: IF MAGNET TEMPERATURE HAS EXCEEDED 45℃, PLEASE CONTACT THE GE


REPRESENTATIVE OR SERVICE SUPPORT CENTER IMMEDIATELY.

WARNING

THE 5 GAUSS LINE WIDENS ABOUT 1m WHEN THE CRATE IS OPENED. BE SURE TO
KEEP ALL FERROUS OBJECTS AND EQUIPMENTS THAT MAY BE INFLUENCED BY THE
MAGNET, AS WELL AS PEOPLE WITH MAGNETIC FIELD SENSITIVE IMPLANTS, AWAY
FROM THE MAGNET. SEE ILLUSTRATION 2-3, 2-4 ISOGAUSS LINE PLOTS FOR THE
MAGNETIC FRINGE FIELD.

3-2-4. Check the magnet for damages in the shipping process, and
confirm that no parts are missing.

NOTE: REPORT ALL THE DAMAGES OR MISSING, IF ANY, IMMEDIATELY TO THE LOCAL
GE REPRESENTATIVES. AT THIS TIME BE READY TO DESCRIBE TYPE OF
DAMAGE, TYPE OF EQUIPMENT, SERIAL NUMBERS AND IF POSSIBLE THE
ORDER NUMBER.

3-2-5. Release the eyebolts from the wooden lock, as shown in illustration
3-8.

NOTE: THE EYEBOTLS ARE MADE FROM STAINLESS STEEL, AND ARE DESIGNED TO
LIFT THE GEHC 0.2T PROFILE MAGNET ONLY. PLEASE MAKE SURE TO USE
THE STAINLESS STEEL EYEBOLTS TO LIFT THE MAGNET.

1. THE EYEBOLTS ARE FIXED 2. REMOVE THE PLASTIC BELTS


TO THE WOODEN BLOCKS
BY PLASTIC BELTS.
Illustration 3-8

23
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

3-2-6. Open the zippers again and install the eyebolts to the magnet on the
top of the magnet. See illustration 3-9.

EYE BOLTS INSTALLATION


Illustration 3-9

3-2-7. Release the wooden locks, which are used to fix the magnet legs to
the shipping crate. Please see illustration 3-10.

RELEASE THE WOODEN BLOOK


Illustration 3-10

24
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

3-2-8. Prepare four Roller Skids on the ground. The magnet will be
delivered to the scan room on these Roller Skids.

3-2-9.Hook the eyebolts, lift the magnet and place it on the Roller Skids.
And then take off the eyebolts. Refer to illustration 3-11.

Please see Illustration 3-4 and Illustration 3-8 for the position and installation of the eyebolts.

NOTE: DO NOT REMOVE THE OVERALL MASK COVERING THE MAGNET DURING THE
RIGGING AND DELIVERING PROCESS. ALL THE SAFETY RISKS, SUCH AS
MAGNET FALLING DOWN, DURING THE LIFTING PROCESS SHOULD BE
CAREFULLY EVALUATED AND PREVENTED BEFORE THE LIFEING OPERATION.

NOTE: REPORT ALL THE DAMAGES OR MISSING, IF ANY, IMMEDIATELY TO THE


LOCAL GE REPRESENTATIVES. AT THIS TIME BE READY TO DESCRIBE TYPE
OF DAMAGE, TYPE OF EQUIPMENT, SERIAL NUMBERS AND IF POSSIBLE THE
ORDER NUMBER.

HOOK THE EYEBOLTS, LIFT THE MAGNET AND PLACE IT ON THE ROLLER SKIDS
Illustration 3-11

NOTE: MAKE SURE TO CHECK THE DIRECTION OF THE MAGNET BEFORE DELIVERING
SO THAT THE FRONT AND BACK SIDES WILL BE CORRECTLY ORIENTED. THE
MAGNET MIGHT BE IMPOSSIBLE TO REVERSE INSIDE THE SCAN ROOM.

3-2-10. Deliver the magnet to the Scan Room.

After the magnet is uncrated, the maximum magnet dimension is L2040 x W1300 x H1380.
The minimum dimensions for a straight path and a path with 90°turns for magnet delivery are
shown in Illustration 3-12 respectively.

25
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

Illustration 3-12

The magnet is completely enclosed by an OVERALL MASK to prevent ferrous objects from
being drawn to the magnet and damaging the magnet. DO NOT take off this OVERAL MASK
when delivering the magnet to the scan room.

Use the free standing posts, plastic chain to rope off the operation site during installation and
service activity. Warning signs should also be pasted to prevent people from entering the
magnetic fringe field area.

And the magnet room should be swept clear from ferrous objects before the arrival of the
magnet. This must be done and checked before delivery.

NOTE: CONFIRM THE DELIVERY ROUTE AND MAGNET ROOM ARE SWEPT
THOROUGHLY FROM FERROUS OBJECTS AND MAGNETIC FIELD SENSITIVE
DEVICES.

NOTE: PEOPLE WITH ANY IMBEDDED MEDICAL DEVICES SHOULD BE STRICTLY


FORBIDDEN FROM ENTERING THE DELIVERY ROUTE. THIS MUST BE DONE
AND CHECKED BEFORE AND DURING RIGGING/DELIVERING WITH
RESPONSIBLE PERSON ATTENDANCE.

NOTE: THE MAGNET IS 10000KG IN WEIGHT, PLEASE CONFIRM WHETHER THE FLOOR
CAN SUPPORT THE MAGNET AND THE ROLLER SKIDS WON’T DAMAGE THE
FLOOR SURFACE. ADDITIONAL PADDING STEEL PLATE MAY BE NEEDED TO
PROTECT THE FLOOR, AND THIS PADDING PLATE SHOULD BE PREPARED BY
THE CUSTOMER. PLEASE REFER TO DIRECTION 2186839, SIGNA PROFILE III,
PRE-INSTALLATION MANUAL FOR DETAILS ABOUT ARCHITECHURAL
REQUIREMENT.

26
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

NOTE: PLEASE CHECK THE FRONT AND REAR SIDE OF THE MAGNET BEFORE
ENTERING THE SCAN ROOM. IT MIGHT BE VERY HARD TO ADJUST THE
FRONT AND REAR SIDES OF MAGNET INSIDE THE MAGNET ROOM.

3-2-11.When reaching the scan room, check the temperature indicators on


the magnet again. Make sure magnet temperature does not exceed
45℃ during the delivery process.

NOTE: IF MAGNET TEMPERATURE HAS EXCEEDED 45℃, PLEASE CONTACT THE GE


REPRESENTATIVE OR LOCAL SERVICE SUPPORT CENTER.

NOTE: AN AIR CONDITIONER SHOULD BE INSTALLED IN THE MAGNET ROOM. THE


TEMPERATURE SHOULD BE CONTOLED WITHIN THE RANGE FROM 22℃ TO
26 ℃ AND KEEP STABLE.

3-2-12.Anchor the magnet in the scan room.

There is a steel plate fixed under the floor (invisible from the surface), but its position is clearly
marked on the floor. See Illustration 3-13. Please refer to Direction 2186839, Signa Profile III
Pre-Installation Manual, SECTION 6 for details about FLOOR SHIELDING.

Illustration 3-13

27
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

Place the rubber plates, which will be padded under the feet of the magnet, at the correct
position. The co-geometric center of the four rubber plates should be within ±50mm of the
steel plate center. Jack the magnet up sufficiently at the jack applicable position and remove
the magnet from the roller skids. And then, anchor the magnet on these rubber plates. See
illustration 3-14 as examples (the floor shielding arrangement could be different in each
hospital). Please apply the jack at the correct position indicated by the labels on the overall
mask. Refer to Illustration 3-15.

Rubber Plate
Leg of the magnet
Steel sheet Floor (Plywood or concrete)

ANCHOR THE MAGNET ON THESE RUBBER PLATES


Illustration 3-14

Illustration 3-15

WARNING

SPECIAL CARE MUST BE CONSIDERED TO USE THE JACK SINCE THERE IS STRONG
MAGNETIC FIELD UNDER THE MAGNET PRODUCED BY THE PERMANENT MAGNET.
WHEN MOVING THE JACK, KEEP THE DISTANCE FROM MAGNET MORE THAN 1 METER.

28
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

WHEN MOVING THE JACK AROUND THE MAGNET, YOUR BACK MUST FACE ON MAGNET
TO KEEP THE DISTANCE, AS ILLUSTRATION 3-16 SHOWS.

≥1m

CORRECT WAY TO USE THE JACK


Illustration 3-16

Check and verify that the magnet center comes within ±50mm of the center of steel plate,
see Illustration3-13. If the center of the magnet is out of this specification, the location of the
magnet MUST be adjusted. Jack the magnet again, adjust the position of the rubber plates
and then anchor the magnet on these rubber plates until the magnet center within the
specification.

3-2-14. Take the Ring Protector Set and Overall Mask off and then begin
leveling the Magnet.

Please refer to SECTION 4 “MAGNET LEVELING” for details.

WANING

DO NOT TAKE THE OVERALL MASK AND THE RING PROTECTOR SET OFF FROM THE
MAGNET UNTIL THE LAST MINUTES YOU ARE READY TO DO THE MAGNET LEVELING.
WITHOUT THE PROTECTION FROM THE OVERALL MASK, THE FERROUS OBJECTS
COULD BE EASILY STUCK TO THE MAGNET TIGHTLY, WHICH COULD CAUSE DAMAGES
TO THE MAGNET AND IT COULD BE HARD TO TAKE OFF FROM THE MAGNET ( IF THERE
IS ANY THING STUCK TO THE MAGNET, PLEASE GET INTOUCH WITH THE INSTALLATION
REPRESENTIVE, OR CONSULT WITH GEHC SERVICE CENTER. DO NOT TRY TO REMOVE
THE STUCK OBJECTS BY YOURSELF. THE REMOVING OF THESE OBJECTS FROM THE
MAGNET MAY CAUSE SERIOUS PERSONNEL INJURIES).

29
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

SECTION 4. MAGNET LEVELING

4-1 CAUTIONS IN LEVELING THE MAGNET.

WARNING

PEOPLE WITH ANY KINDS OF MAGNETIC FIELD SENSITIVE IMPLANT, SUCH AS CARDIAC
PACEMAKER, NEUROSTIMULATORS, BIOSTIMULATION DEVICES, ARE STRICTLY
FORBIDDEN FROM ENTERING THE SCAN ROOM.

NOTE: THE OVERALL MAGNET MASK AND THE RING PROTECTOR SET SET, WHICH
ARE UESD TO PROTECT THE MAGNET FROM BEING DAMAGED BY FLY-IN
OBJECTS, WILL BE REMOVED FROM THE MAGNET. PLEASE PREVENT THE
JACK AND OTHER FERREOUS OBJECTS FROM BEING ATTRACED TO THE
MAGNET AND CAUSING INJURUES TO OPERATOR AS WELL AS DAMAGES
TO THE MAGNET ITSELF.

NOTE: PLEASE PRESERVE THE OVERALL MASK AND THE RING PROTECTOR SET
PROPERLY AFTER FINISHING THE MAGNET LEVELING. THE OVERAL MASK
AND THE RING PROTECTOR SET MAY BE USED AGAIN LATER BY SERVICE
ENGINEERS.

NOTE: USE THE JACK AT THE CORRECT POINTS, WHICH ARE MARKED BY LABELS
ON THE OVERALL MASK. REFER TO ILLUSTRATION 3-15.

30
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

4-2 MAGNET LEVELING PROCEDURE FLOWCHART

START

REMOVE THE MAGNET OVERALL MASK


AND THE RING PROTECTOR SET

MAGNET LEVEL CHECK

WITHIN THE Y
SPEC ? RE-INSTALL THE OVERALL
MASK

RE-INSTALL THE OVERALL MASK AND END


THE RING PROTECTOR SET

MAGNET LEVELING

MAGNET LEVELING PROCEDURE FLOWCHART

Flowchart 4-1

31
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

4-3 MAGNET LEVELING

. 4-3-1. Sweep the Magnet Room thoroughly for Iron powder and other
ferrous objects.

4-3-2. Take off the stainless steel OVERALL MASK and Ring Protector Set,
as shown in the illustration 4-1, which covers the Magnet.
TEMPERATURE INDICATOR

REMOVE THE WARNING


OVERALL MASK

MAGNET FRONT VIEW

TAKE OFF THE OVERALL MASK


Illustration 4-1

4-3-3. Cut an opening in the Plastic Cover near the slit. See Illustration 4-2.

Slit

OPEN THE PLASTIC COVER


Illustration 4-2

32
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

Illustration 4-3

4-3-4. Fill 3/4 of the Invisible tube with water.

4-3-5. See Illustration 4-3. Measure the difference between the surface line
of the slit and water level height in the tube. There should be no
more than 2 mm difference between any of the 4 slits (at 4 corners
of the Magnet).

4-3-6. If there is more than 2mm difference, reinstall the Ring Protector Set
and the Overall Mask, tighten the rope and make the Overall Mask
cover the Magnet tightly, and then jack up the Magnet and insert
steel plates under the Magnet feet to level the magnet. Please refer
to Illustration 4-4.

33
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

Illustration 4-4

WARNING

THE JACK CONTAINS FERROMAGNETIC MATRIAL. SPECIAL CARE MUST BE


CONSIDERED TO USE THE JACK SINCE THERE IS STRONG MAGNETIC FIELD UNDER THE
MAGNET PRODUCED BY COUNTER MAGNET.

NOTE: PLEASE USE THE JACK AT THE POINT INDICATED BY LABELS IN CASE OF ANY
DAMAGES TO THE MAGNET.

Illustration 4-5

34
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

WARNING

SPECIAL CARE MUST BE CONSIDERED TO USE THE JACK SINCE THERE IS STRONG
MAGNETIC FIELD UNDER THE MAGNET PRODUCED BY COUNTER MAGNET. WHEN
MOVING THE JACK, KEEP THE DISTANCE FROM MAGNET MORE THAN 1 METER. WHEN
MOVING THE JACK AROUND THE MAGNET, YOUR BACK MUST FACE ON MAGNET TO
KEEP THE DISTANCE, AS ILLUSTRATION 3-15 SHOWS.

≥1m
Illustration 4-6

4-2-7. If the difference is within 2mm specification, the leveling process is


successfully finished. The next step will be the magnet field
shimming. If there is any interval between leveling and shimming,
put the Overall Mask and Ring Protector Set back on and tighten the
rope of the Overall Mask.

NOTE: THE OVERALL MASK AND THE RING PROTECTOR SET SHOULD BE PROPERLY
PRESERVED FOR THE POTENTIAL USAGE IN THE FUTURE. DURING ANY
MAGNET RELATED MAINTENANCE OR MANGET TRANSPORTATION, THE
OVERALL MASK AND THE RING PROTECTOR SET MUST BE USED TO PROTECT
THE MAGNET TO PREVENT ANY DAMAGES AND INJURIES.

NOTE: BEFORE THE SYSTEM INTEGRATION AND MAGNETIC FIELD SHIMMING, THE
MAGNET TCU SHOULD BE SET TO 32.5℃. IT IS STRONGLY SUGGESTED TO
SET THE MAGNET TUC AT THE SAME NUMBER WITH THE FACTORY MAGNET
SHIMMING TCU SETTING. THE FACTORY SHIMMING TCU SETTING NUMBER
COULD BE FOUND FROM THE MAGNET “ ACCEPTANCE TEST REPORT “,
WHICH COULD BE FOUND IN THE PLASTIC FOLDER PLACED IN THE TOP OF
BOTTOM POLE FACE. AND IT IS IN THE SAME FOLDER USED FOR THE
LOCATOR CARD.

35
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

SECTION 5. MAGNETIC FIELD SHIMMING


This section is an introduction to the GEHC 0.2T Profile Magnet field shimming method. The
GEHC 0.2T Profile Magnet uses PC type filed shimming software, which is a different field
shimming software from the shimming software mounted in the Operation Console. To shim
the GE magnet, the PC type field shimming software should be used.

The purchaser of GE magnet shall only utilize qualified personnel (i.e., GE’s field engineers,
personnel of third–party service companies with equivalent training) to perform shimming
servicing on the equipment.

5-1 MAGNETIC FIELD SHIMMING PREPARATION

5-1-1 Tools & Equipments:

For field shimming, special tools and equipments are needed. And these tools/devices, which
are not included in the shipping collector, should be prepared by service engineers who
preside the shimming. Please refer to table 5-1 for tools/equipments details.

Tools/Equipments Quantity
MetroLab Teslameter PT2025 46-255839P2 1
Single Probe#2 or #3 46-255839P102 1
Single Probe Mapping Fixture 2141561 1
Operation System:Windows 2000 N/A 1
Service CD SV manual 2124628 / 2359125 (Class A) 1
Non-Magnetic Tools 1 set
Thermometer 1
Field Shimming kit 2370113 1
Wrench 2370034 1

TOOLS/EQUIPMENTS

Table 5-1

5-1-2 Pre-Requirements before starting the shimming

a. The temperature of the shimming room should be within the range from 22 to 26℃, which
can be measured by the Temperature Meter. Please keep the room temperature sable
and avoid fluctuating during the shimming process.

36
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

b. The magnet TCU should be set to 32.5℃ and heating the magnet up (This should have
began after the magnet leveling). It is suggested to set the magnet TCU at the same
number with the factory magnet shimming TCU setting. The factory shimming TCU
setting number could be found from the magnet “ Acceptance Test Report “, which could
be found in the plastic folder placed in the top of bottom pole face, and is the same folder
used for the locator card.

c. Take off the Overall Mask and the Ring Protector Set from the magnet.

d. Make sure the insulators are installed properly on the magnet.

e. Heat the magnet no less than 24 hours until the temperature of the magnet is stable
enough.

Field Drift Inspection Specification:


Set the teslameter to Frequency mode as show in Illustration 5-1.

Frequency
mode

SET THE TESLA METER TO FREQUENCY MODE


ILLUSTRATION 5-1

Drift of the central field should be less than 150Hz per 10 minutes and this situation should
last at least two hours. To measure drift, the probe should be placed in the ISO center of the
magnetic field ( in Plane 7 ). Please find more detail procedures from: Service CD (Part NO.
2124628/2359125), Profile 3/4 Installation (For Profile 4b) 6th STEP Sumitomo 2.
Setup (Mapping Fixture setup).

Drift of the central field less than 150Hz per 10 minutes.


Inspection method: Measure the central field and calculate the drift every 10 minutes. Say

37
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

initial central field B1, and drifts to B2 10 minutes later, then the drift can get by: |B2-B1| ;
(|B2-B1|<150 Hz).

e. Confirm the table was installed properly (temporarily). Please refer to Service CD (Part
NO. 2124628/2359125), Profile 3/4 Installation (For Profile 4b) 3th STEP Table
Installation (Temporarily).
the table installation instruction for reference and turn the System Gradient Unit off.

NOTE: THE FIELD SHIMMING CANNOT START UNTIL ALL THE ABOVE REQUIREMENTS
ARE MET.

38
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

5-2 FIELD SHIMMING SPECIFICATIONS

5-2-1: Magnet field homogeneity specification.

The field shimming cannot be finished until the following two specifications are met.

a. At the FOV of 35*40*40cm, the inhomogeneity of the field must be less than 45 ppm
(135 pts Pk-Pk measurement).

b. Magnet Final Central Frequency Range is within 8.50 ~ 8.58 MHz.

In the next section there is an introduction about how to shim the magnetic field to meet above
specifications.

5-2-2. Magnet field shimming specification quick guide.

Parameter Specification Notes


Magnet Drift Rate ≤150Hz per 10 minutes In 2 hours period.
Room Temperature 24.0° C ±2.0° C Keep the temperature stable.
Suggest to be the same setting with
the factory TCU setting. Refer to
Magnet TCU setting 32.5 ° C ± 1.0 ° C
“Acceptance Test Report” for the
factory temperature setting number*.
Magnet Final Frequency 8.50 ~ 8.58 MHz
Range
40x40x35 cm DSV ≤ 45.0 ppm Compensated
Specification
18 cm DSV ** ≤ 3.3ppm Compensated

* The “Acceptance Test Report” could be found in the plastic folder placed on the pole face of the bottom Yoke.
** Only required to the Profile/I system, or EPI function.

SPECIFICATGION QUICK GUIDE

Table 5-2

39
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

5-3. MAGNETIC FIELD SHIMMING PROCEDURES


This section introduces the magnetic filed shimming procedures of the GEHC 0.2T Profile
Magnet. The basic magnetic field shimming procedures for GEHC 0.2T Profile Magnet are as
shown in flowchart 5-1.
Start

Hardware setup
(Mapping fixture, probe)

Run shimtool A, setup the working


directory for the first time

Shim Data Gathering , Section 5-4 -2 2.1)

Homogeneity in Spec? Yes

No
Mechanical shimming Calculation
Section 5-4 -2 3.1)

Yes
|Z1| < 7ppm & |Z2| < 7
ppm
No
Execute Mechanical
shimming
Section5-4-2, 3.1).

Scan passive shimming


Section 5-4-2, 3.4)

Choose a good result & execute


Passive Shimming
Section 5-4-2, 3.5)

Data Sheet

End

* There is no Transverse shimming in the GEHC 0.2T Profile Magnet mechanical shimming.
** By the “B0 Keep Mechanical shimming” method, the B0 field will be unchanged while harmonics could be
adjusted.
MAGNETIC FIELD SHIMMING PROCEDURE FLOWCHART
Flowchart 5-1

40
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

5-3-1. Hardware setup

1. Release the Overall Mask and the Ring Protector completely from the Magnet.

2. Fix the single probe fixture onto the bottom ring of the pole piece.

For the detail procedures, please refer to Service CD (Part NO. 2359125 or
2124628), Profile 3/4 Installation (For Profile 4b) 6th STEP
Sumitomo Chapter 2 (Mapping Fixture setup).

3. Fix the single probe on the Horizontal bar, and mount it on the single probe fixture.

For the detail procedures, please refer to Service CD (Part NO. 2359125 or
2124628), Profile 3/4 Installation (For Profile 4b) 6th STEP
Sumitomo Chapter 2 Setup (Mapping Fixture setup).

4. Power on the Tesla-meter PT2025; set and confirm the Teslameter to the
Frequency Mode.

Please refer to section 5-1-2. “ Pre-Requirements before starting the shimming”


for setting teslameter direction.

5-3-2. Shimtool software installation and Working directory setup.

Confirm the Shimming Software is installed successfully in your Laptop. The


software CDROM (GEHC 0.2T Profile Magnet Field Shimming Service CDROM, PN
5122078) could be found in the Magnet shipping accessories packing Box.

After running the software, please specify a directory for the software before a new
magnet is to be shimmed. Once it is specified, all the files the software generates
in the shimming process will save to this folder automatically. Please follow Section
5-4-2 1.1) Set Working Dir…..

5-3-3. Record the magnetic field point by point;

41
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

Please refer to Service CD (Part NO. 2359125 or 2124628), Profile 3/4


Installation (For Profile 4b) 6th STEP Sumitomo Chapter 5 Data
Gathering for the guidance about the data gathering procedures.

Please refer to section 5-4 2.1 “Data Gathering” to details about how to gather
the magnetic filed data with the shimming software.

NOTE: THE ORIENTATION SYSTEM OF THE GEHC 0.2T PROFILE MAGNET IS


DIFFTERENT FROM THE SUMITOMO / SHINETSU MAGNET AS SHOWN
IN THE ILLUSTRATION 5-1. PLEASE PAY ATTENTION TO MEASURE
THE MAGNETIC FIELD DATA FROM THE GEHC 0.2T MAGNET
SHIMMING COORDINATION 0 DEGREE.

180
210 150
240 120

270 90

300 60
330 30
0
Start Measuring the Data from here
Table Side
GEHC 0.2T Profile Magnet Coordinate System

Illustration 5-1 MAGNET OREINTATION SYSTEM.

5-3-4. Compare the Pk-Pk PPM value with spec.

Magnetic Field inhomogeneity pk-pk ppm value can be read from the field map in the
software main window (refer to illustration 5-3). If the overall inhomogeneity is in spec
(see section 5-2), field shimming could be finished, otherwise, please go to the Step 5-
6.

42
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

5-3-5. Mechanical shimming


If either Z1 or Z2 is more than 7 ppm, please perform the Mechanical Shimming, Axial
only. Please refer to the Section 5-4-2 3.1) to calculate “mechanical shimming”.

Please refer to Profile Service CD (Part NO. 2359125 or 2124628), Profile 3/4
Installation (For Profile 4b) 6th STEP Sumitomo 9. Mechanical Shim (Axial)
for detail procedures about how to implement the mechanical shimming. The GEHC
0.2T Profile does not designed to implement the Mechanical Shimming, transverse
method

After the mechanical shimming is finished, please go back to 5-3-3. Only when Z1 or
Z2 are all less than 7 ppm, could the passive shimming begin as introduce in section
5-3-6.

5-3-6. Passive Shimming.

If the mechanical shimming meets the spec of all Z1& Z2 less than 7ppm, one can
execute “scan passive shimming”, which is described in Section 5-2 3.4, to execute
the passive shimming. After scan and calculation finished, one can go through the
results and select a relative good shimming plan from the shimming calculation results
by evaluating the number of the shimming chips would be used, the location number to
be used and the anticipated homogeneity value to be achieved. Run the passive
shimming with the parameter setting as the selected target ppm and shims number to
get the passive shimming plan. Please refer to Section 5-2 3.5 as instruction to the
passive shimming.

For locate the shim chips, please refer Profile Service CD (Part NO. 2359125 or
2124628), Profile 3/4 Installation (For Profile 4b) 6th STEP Sumitomo.

NOTE: THE ORIENTATION OF THE SHIM HOLDER SOCKET POSITION is


DIFFERENT FROM THE SUMITOMO OR SHINETSU MAGNET (REFER
TO ILLUSTRATION 5-1 AS EXAMPLE). PLEASE FOLLOW THE GEHC
0.2T MAGNET SHIM SOCKET ORIENTATION TO PLACE THE SHIMMING
CHIPS.

43
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

5-3-7. Go to 5-3-3.

Compare with the specs in section 5-2. Do the passive shimming or mechanical
shimming again until the specs are met.

5-3-8. Make the Data Sheet.

After the magnet shimming is accomplished, please fill the magnet installation
spreadsheet which could be found in the Shimming software Service CDROM (5122078).

After fill this form, please send the spreadsheet to ProfileMagnet.GEMSC@GE.COM, or


you can ask engineering via this email account if you have any question in the installation
process.

NOTE: THE OVERALL MASK AND THE RING PROTECTOR SET SHOULD BE PROPERLY
PRESERVED FOR THE USAGE IN THE FUTURE. DURING ANY MAGNET
RELATED MAINTENANCE OR MANGET TRANSPORTATION, THE OVERALL
MASK AND THE RING PROTECTOR SET MUST BE USED TO PROTECT THE
MAGNET FROM ANY DAMAGES AND PERSONNEL INJURIES.

NOTE: IN ANY KIND OF MRI SYSTEM MAINTENANCE, PLEASE BE CAUTIONS TO THE


STRONG MAGNETIC FIELD. PLEASE REMOVE ALL THE FERROUS OBJECT
FROM THE MAINTENANCE OPERATION SITE. PLEASE REFER TO THE SECTION
1 FOR THE SAFETY MEASURES.

44
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

5-4 FIELD SHIMMING SOFTWARE INTRODUCTION

There is a specific shimming software is prepared to the field shimming of GEHC 0.2T Profile
magnet. This shimming software is mounted in the GEHC 0.2T Profile Magnet field shimming
Service CDROM (5122078), which is sent to the installation site with the magnet. To
implement the field shimming to the GEHC 0.2T Profile Magnet, the service engineers must
install this software to the Laptop first.

5-4-1: Software Installation.


Please find the CDROM (5122078) in the magnet-shipping collector. In this CDROM there
are one folder and two independent files, including GEHC shimming software, two files of
GEHC 0.2T profile Magnet Installation Manual (PN:2370115-100) and the Magnet installation
data investigation excel sheet. Please copy all the content of the CDROM to your Laptop hard
disk.

5-4-2: Software Execution and Interface introduction.


Go into the shimming software directory, and [double click] the executive file named
“ShimToolA”. The shimming software User Interface will jump out. In the shimming software
operation window there are Manual Bar, Tool Bar, Main Window and output window.
Please refer to illustration 5-3 for details. All the calculation results and the shimming direction
will be instructed in the Output Window.

45
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

SHIMTOOL SOFTWARE INTERFACE


ILLUSTRATION 5-3

46
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

1. File Menu

File Manual
Illustration 5-4

1.1 Set Working Dir...


Before one begins to shim a new magnet, please set the Working Directory first. After this
step, all the maps, passive shimming analysis result files, B0 drift monitoring records and
output log files will be saved to this directory by default. Please refer to the Illustration 5-5 &6.

SET WORKING DIR…


ILLSTRATION 5-5

When click on Button, one can select the target directory as desire. One is also allowed to
establish a new directory and rename it at this step.

47
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

SET WORKING DIR…


ILLSTRATION 5-6

NOTE: THE BUTTON OFTEN APPEARS IN THIS SOFTWARE. IT IS ALWAYS USED


TO SELECT THE TARGET FILE OR DIRECTORY.
NOTE: TO SPECIFY A DIRECTORY FOR THE MAGNET SHIMMING IS STRONGLY
SUGGESTED BEFORE EXECUTING THE FIELD SHIMMING. PLEASE SPECIFY A
NEW DIRECTORY EACH TIME ONE SHIM A NEW MAGNET.

1.2. New
To generate a new empty map file which is waiting for data gathering.

1.3 Open...

To open an existing map file or to open an existing psm file (passive shimming result file).
The software will automatically search the base map and show the result with the predicted
map. (The base map’s location is described in the psm file. If the base map file is moved to
another place, it should be edited with notepad or other ASCII editors).

1.5 Save
To save current map file.

48
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

1.6 Save As...


To save current map file with a different name.

1.7 Most Recent Used Files

MOST RECENT USERD FILES


ILLUSTRATION 5-7

The most recent used file will be listed for the convenience as shown in the illustration 5-7.

49
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

2. Data Menu

2.1 Manual Data gathering

MANUAL DATA GATHERING DIALOG.


ILLUSTRATION 5-8

Manual Data gathering Dialog is used for recording the magnetic field value on each specified
points to generate a map. There are totally 135 points in 13 planes should be measured one
map: in plane 2~ plane 12, each plane has 12 angles (0 degree to 330 degree, 30 degree for
each step) and 1 point in each of plane 1 and plane 13. The B0 strength on the ISO center
(plane7) should be measured twice, firstly in the beginning of taking the map and secondly in
the end, for compensating the field drifting in the field mapping process. Please refer to
illustration 5-9 for the arrangement of the planes. The angles are marked on the shimholders
of the magnet. One can also have a look at the Profile Service CD (Part NO. 2359125 or
2124628 ), Profile 3/4 Installation (For Profile 4b) 6th STEP Sumitomo for more
details about the measuring and recording the data.

50
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

B0 FIELD PLANE LAYOUT


ILLUSTRATION 5-9

When using the software, the field value is combined by Prefix and suffix. For example, if the
read value for teslameter is 8.5012345MHz, the prefix 8.5 could be specify, and only “ suffix”
number is need to type in. After typing, please click "Accept" key or type “Enter” key on the
keyboard to accept the data.

"Overall Warning Bound(PPM +/-)" edit box lets you to set the overall warning boundary, if
field value typed in is outside of it, a warning will occur to let you be aware that abnormal data

is inputted. You can click right side to change the current point and recollect the data if you
made a mistake.

"Short Term Warning Bound (PPM +/-)" edit box is similar with the above edit box, which
focuses on the difference between the current field value and the last, theoretically there
should be no big difference between the two consequent data.

51
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

3. Analysis Menu

ANALYSIS MANUAL
ILLUSTRATION 5-10

This menu contains all commands related with shimming a magnet.

3.1 Mechanical Shimming


After run “mechanical shimming” command to an existing map, the program will bring out the
mechanical shimming plan in the output Window similar as the following Illustration 5-26. If
the software gives the result that either |Z1| or |Z2| is larger than 7, one should use the
shimming plan and perform the mechanical shimming physically.

OUTPUT WINDOW
ILLUSTRATION 5-11

One can watch Teslameter’s display value and tune the Top B0 plug and then Bottom B0 plug,
Sequentially until it is the same as what is in the output window, respectively.

Please refer to Profile Service CD (Part NO. 2359125 or 2124628), Profile 3/4
Installation (For Profile 4b) 6th STEP Sumitomo for detail procedures about implement
the mechanical shimming.

52
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

3.2 Decompose Field

This function allows you to read the axial harmonic of the magnetic field. Only harmonics from
(0,1)~(0,12) are listed. The most important items are (0,1) and (0,2), or say Z1 and Z2.

3.3 Set Scan Shim Param...

SET SCAN SHIM PARAM


ILLUSTRATION 5-12

The dialog shown in Illustration 5-12 is used to set the parameters for the “ Scan Shim”.
“ Scan Shim” function is used to assisting service engineer to find the best Passive Shimming
“ target “. Please refer to 3.5) Passive shimming for more intruction.

With Only ONE time setting here, one can execute passive shimming analysis and get a
series of simulation results simply and easily (Please refer to Illustration 5-15). The target
PPM can be from 0.01 to 100. “Shims limit” stands for the shimming chips number limitation
in each socket, and it should be less than or equal to 4. The default values are suitable to
most cases. You can also change them before run “scan passive shimming” in 3.4) to see if
better result could be generated (The better result comes from the comprehensive
consideration of the “Anticipated PPM” value and “Shims Number”. Both this two
parameter should be as small as possible).

Target PPM from 12 to 36, with delta of 0.2 and Shims Limit from 4 to 4 with delta of 1 is set
as the default value. When the best analysis results occur at the boundary, the Target PPM
value scanning range could be extended in have more analysis result around the target PPM
value. For Example: If the target is set as from 12ppm to 36ppm, while the best analysis

53
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

result is just near 12ppm, you could change the Target PPM range as from 5PPM to 14PPM
and run scanning again to see if there is an even better result could be found.

3.4 Scan passive shimming

“ Scan Shim” function is used to assisting you to find the best Passive Shimming “ target “. To
execute scanning of passive shimming with the specified target ppm and shims number limit
in each pocket, which is described in 3.4), to get the simulated shimming. A dialog, as
shown in Illustration 5-28, will appear after clicking the “scan passive shimming” command. If
this is the first-time passive shimming (NO shims in the shimholders), keep “2nd or More
Iteration shim” unchecked, and the “Shim File of Last Time” Edit Box is therefore disabled.
Please refer to Illustration 5-13.

SCAN PASSIVE SHIMMING – FIRST TIME


ILLUSTRATION 5-13

Otherwise (there are shims already on shim holders), please check the “2nd or More Iteration
shim” and specify the psm file of last time before the current loaded map is taken. Please
refer to Illustration 5-14 for reference.

54
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

SCAN PASSIVE SHIMMING – 2nd or More Iteration shim


ILLUSTRATION 5-14

After click on “OK” Button, the brief summary of analysis result for each target is shown in
Output Window as shown in illustration 5-15. ( One can click on “Scan Para.” Button to set
parameters for scanning pass shimming. Please refer to 3.4 ”set scan shim param. ”for
reference).

SCAN PASSIVE SHIMMING – 2nd or More Iteration shim


ILLUSTRATION 5-15

After scanning is finished, one can look through the brief summary, choose the best
anticipated result and execute 3.5) “the passive shimming” with the corresponding “ settings”.
Please consider the “shims number” and the “Anticipated PPM” comprehensively with main
principle the smaller “shims number” with smaller “Anticipated PPM” the better.

55
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

3.5 Passive shimming

A dialog will appear and let you specify the passive shimming parameter as shown in
Illustration 5-16.

PASSIVE SHIMMING –FIRST TIME


ILLUSTRATION 5-16

PASSIVE SHIMMING – 2nd or More Iteration shim


ILLUSTRATION 5-17

If this is the first-time passive shimming (NO shims on shim holders), then keep “2nd or More
Iteration shim” unchecked, as shown in the illustration 5-16, and the “Shim File of Last Time”

56
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

edit-box is therefore disabled, as shown in illustration 5-17. Otherwise (There are shims
already on shim holders), please check the “2nd or More Iteration shim” and specify the psm
file of last time before the current loaded map is taken. These items are the same as what in
3.4.

Please specify the “Target PPM” value and “Shim Limit” value you choose from “Scan passive
shimming” results (refer to 3.4). After click on “OK” button, the program will analysis, generate
and open the result with notepad. ( for detail information about the analysis result file, please
refer to section 4).

The string in the “Output Shim File” edit box is the name of the output shim file, which is
consistent with the name of the base map file as default ( Specified by user or assigned by the
software). The program will automatically add the anticipated inhomogeneity ppm value, target
ppm value to the file name, which has an extension of “.psm”. For example, if you type
“afterMech1” in this edit box as the file name, the target ppm is 13.6, the anticipated ppm is
22.5 after calculation, then the actually output shim file will be saved as the name
“afterMech1_22.5(13.6).psm” automatically. If one accept the result of the psm file and
perform shimming with it, please specify this name for the next iteration of passive shimming
if needed. (Please put this name in the “Shim File of Last Time”, as shown in Illustration 5-17.)

3.6 Stop Scan.

To interrupt the current scanning. Do not close the shimtool software when scanning passive
shimming is running.

3.7 Calculate 18 DSV ppm.

This function only for Profile / I system.


To calculate the inhomogeneity on 18 cm diameter sphere. Result will be shown in Output
Window. DATA to record down is the value after zeroing (1,0), (1,-1)and (1,1). Please refer
to the information in the illustration 5-18.

57
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

18CM DSV HOMOGENEITY CACULATION


ILLUSTRATION 5-18

4. Passive shimming analysis result file.

Please try to find the section of "Display Shim Direction and Shim Location - Lower Delta
","Display Shim Direction and Shim Location - Upper Delta", "Display Shim Direction
and Shim Location - Lower Total “and "Display Shim Direction and Shim Location -
Upper Total" and execute passive shimming (put small shims into pockets on shim holders).
Note: the Ring index is from 1 to 6.

In the file, L1~L6 means ring1~ ring 6 on shim holder of lower side, U1~U6 means ring 1~ring
6 on shim holder of upper side. “”Nx (Sx)” means x shims(x=1,2,3,4) with North(South) pole up
putting into the specified pocket of the shim holder after the shim holder is lay on the ground
for both lower and upper shim holder. Please refer to illustration 5-19 for information.

58
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

PASSIVE SHIMMING RESULTS


ILLUSTRATION 5-19

Please refer to Profile Installation Service Profile Service CD (Part NO. 2359125 or
2124628), Profile 3/4 Installation (For Profile 4b) 6th STEP Sumitomo Passive
Shimming for detail procedures about implement the passive shimming, including remove the
shim holder, place the shimming chips, install the shim holder, etc.

NOTE: THE SHIMHOLDER ORIENTATION OF THE GEHC 0.2T PROFILE MAGNET IS


DIFFTERENT FROM THE SUMITOMO MAGNET SHIMHOLDER. PLEASE
FOLLOW THE SHIMMING PROCEURES IN THE CDROM, WHILE USING THE

59
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100

GEHC 0.2T PROFILE SHIMHOLDER ORIENTATION TO PLACE THE SHIM


CHIPS.

5. Magnet Shimming Feedback.

After the magnet shimming is accomplished, please fill the magnet installation spreadsheet
which could be found in the Shimming software Service CDROM (5122078).

After fill this form, please send the spreadsheet to ProfileMagnet.GEMSC@GE.COM, , or

you can ask engineering via this email account if you have any question in the installation
process.

60