GE Healthcare GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
DAMAGE IN TRANSPORTATION
If shipment was handled by moving van service uncrate, call Traffic Milwaukee
immediately when any damage is found. At this time be ready to describe type
of damage, type of equipment, serial numbers and if possible the order number.
20-September-2003
GE Healthcare
GE Healthcare: Telex 3797371
P.O. Box 414, Milwaukee, Wisconsin 53201 U.S.A.
(Asia, Pacific, Latin America, North America)
PRODUCT PUBLICATION # .
PAGE # PAGE REV DATE REPORTED .
SERVICE RECORD SUBMITTED □ YES □ NO
PLEASE BE SPECIFIC IN YOUR CORRECTIONS AND SUGGESTIONS
SERVICE REP
(Please Print)
PHONE
FAX
COUNTRY
CITY
REV 5
(TEAR AT PERF)
GE Healthcare GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
WARNING
THIS SERVICE MANUAL IS AVAILABLE IN ENGLISH ONLY.
IF A CUSTOMER’S SERVICE PROVIDER REQUIRES A LANGUAGE
OTHER THAN ENGLISH, IT IS THE CUSTOMER’S RESPONSIBILITY
TO PROVIDE TRANSLATION SERVICES.
DO NOT ATTEMPT TO SERVICE THE EQUIPMENT UNLESS THIS
SERVICE MANUAL HAS BEEN CONSULTED AND IS UNDERSTOOD.
FAILURE TO HEED THIS WARNING MAY RESULT IN INJURY TO
THE SERVICE PROVIDER, OPERATOR OR PATIENT FROM
ELECTRIC SHOCK, MECHANICAL OR OTHER HAZARDS.
AVERTISSEME
WARNUNG
DIESES KUNDENDIENST–HANDBUCH EXISTIERT NUR IN
ENGLISCHER SPRACHE.
FALLS EIN FREMDER KUNDENDIENST EINE ANDERE SPRACHE
BENÖTIGT, IST ES AUFGABE DES KUNDEN FÜR EINE
ENTSPRECHENDE ÜBERSETZUNG ZU SORGEN.
VERSUCHEN SIE NICHT, DAS GERÄT ZU REPARIEREN, BEVOR
DIESES KUNDENDIENST–HANDBUCH NICHT ZU RATE GEZOGEN
UND ERSTANDEN WURDE.
WIRD DIESE WARNUNG NICHT BEACHTET, SO KANN ES ZU
VERLETZUNGEN DES KUNDENDIENSTTECHNIKERS, DES
BEDIENERS ODER DES PATIENTEN DURCH ELEKTRISCHE
SCHLÄGE, MECHANISCHE ODER SONSTIGE GEFAHREN KOMMEN.
GE Healthcare GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
ATENÇÃO
ESTE MANUAL DE SERVICIO SÓLO EXISTE EN INGLÉS.
SI ALGÚN PROVEEDOR DE SERVICIOS AJENO A GEHC SOLICITA
UN IDIOMA QUE NO SEA EL INGLÉS, ES RESPONSABILIDAD DEL
CLIENTE OFRECER UN SERVICIO DE TRADUCCIÓN.
NO SE DEBERÁ DAR SERVICIO TÉCNICO AL EQUIPO, SIN HABER
CONSULTADO Y COMPRENDIDO ESTE MANUAL DE SERVICIO.
LA NO OBSERVANCIA DEL PRESENTE AVISO PUEDE DAR LUGAR A
QUE EL PROVEEDOR DE SERVICIOS, EL OPERADOR O EL
PACIENTE SUFRAN LESIONES PROVOCADAS POR CAUSAS
ELÉCTRICAS, MECÁNICAS O DE OTRA NATURALEZA.
AVVERTENZA
AVISO
IL PRESENTE MANUALE DI MANUTENZIONE È DISPONIBILE
SOLTANTO IN INGLESE.
SE UN ADDETTO ALLA MANUTENZIONE ESTERNO ALLA GEHC
RICHIEDE IL MANUALE IN UNA LINGUA DIVERSA, IL CLIENTE È
TENUTO A PROVVEDERE DIRETTAMENTE ALLA TRADUZIONE.
SI PROCEDA ALLA MANUTENZIONE DELL’APPARECCHIATURA
SOLO DOPO AVER CONSULTATO IL PRESENTE MANUALE ED
AVERNE COMPRESO IL CONTENUTO.
NON TENERE CONTO DELLA PRESENTE AVVERTENZA POTREBBE
FAR COMPIERE OPERAZIONI DA CUI DERIVINO LESIONI
ALL’ADDETTO ALLA MANUTENZIONE, ALL’UTILIZZATORE ED AL
PAZIENTE PER FOLGORAZIONE ELETTRICA, PER URTI
MECCANICI OD ALTRI RISCHI.
GE Healthcare GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
警 告
注意
本安装手册仅存英文本。
非 GEHC 公司的安装/维修人员要求非英文本的手册时,客户需自行负责翻
译。
未详细阅读和完全了解本手册之前,不得进行安装/维修。
忽略本文注意事项会对安装人员, 操作维修人员或病人造成触电,机械或
其他伤害。
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
REVISION HISTORY
REV . DATE PRIMARY REASON FOR CHANGE
1
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
CAUTION
DO NOT USE THE FOLLOWING DEVICES NEAR THE MAGNET. THE USE OF
THESE DEVICES NEAR THE MAGNET MIGHT BE DAMAGED BY, OR
CAUSES DAMAGES TO THE MAGNET.
KEEP POWER TO THESE DEVICES TURN OFF WHEN NEAR THE MAGNET.
Warning
Never use the following devices near this equipment. They
may cause erratic function of this equipment.
a
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
INTRODUCTION
This manual contains the procedure and information necessary for
unpacking a GEHC M20552PY 0.2T Profile Permanent Magnet (in the
following part of this manual, all the “magnet” refers to the GEHC
M20552PY 0.2T Profile Permanent Magnet) and installing the magnet
into the scan room. The followings are instructed in this manual:
Magnet leveling.
All the contents of this Magnet Installation Manual are used for the
GEHC 0.2T Profile Permanent Magnet only.
All works must be in compliance with national and local building and
b
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
safety codes.
c
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
TABLE OF CONTENT
REVISION HISTORY ............................................................................................. 1
I
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
The key magnet peripherals include magnet Heater, Thermal Insulator, RF Shield,
Shim Holder and Gradient Coil. And all the peripherals have been installed well before
shipping the magnet to the customer.
1
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
The followings are some of the key accessories of the GEHC 0.2T Profile Permanent Magnet.
For the accessories detail, please refer to the magnet shipping collector checklist.
1· Hard Rubber Pedestal Plate: This will go under the magnet pedestals.
1mm x 6 pieces
2mm x 6 pieces
T-Wrench
Shaft
Extension Rod
Handle
4· Shim Assy:
Shimming kits
Shimming Plate
Shimming Holders
6· Locator Card.
2
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
Besides the magnet enclosures and accessories, special equipments which should be
prepared by the licensed installation contractors are also needed for a successful magnet
installation. The following equipments are some of the key equipments must be prepared for
the magnet installation, while they are not all the equipments might be used in the installation.
4. Four dual–life ratchet jacks and handles. Five ton (4,536 kg) capacity, 368 mm travel,
maximum starting height 38 mm toe, 533 mm head.
5. Free standing posts, plastic chain and warning signs for roping off site during
installation and service activity.
WARNING
3
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
WARNING
WARNING
WARNING
WARNING
THE STAFF, WHO WILL BE INVOLVED IN THE MAGNET OPERATIONS, MUST BE CHECKED
FOR ANY MEDICAL / PHYSICAL CONTRAINDICATION TO THIS OPERATION (AS
DESCRIBED ABOVE), AND THIS SHOULD BE PERFORMED BY LICENSED CONTRACTORS,
AS WELL AS MEDICAL STAFF IN CHARGE OF THIS EQUIPMENT.
WARNING
THE 5 GAUSS LINE WIDENS ABOUT 1m WHEN THE CRATE IS OPENED AND THE MAGNET
IS RIGGED OUT FROM THE SHIPPING CRATE. BE CAUTIONS ABOUT THIS CHANGE IN
THE SCOPE OF THE MAGNETIC FRANGE FILED.
4
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
CAUTION
DO NOT USE THE FOLLOWING DEVICES NEAR THE MAGNET. USE OF THESE
DEVICES NEARS THE MAGNET MIGHT BE DAMAGED BY, OR CAUSE DAMAGES TO
THE MAGNET.
PLEASE CLEAR FERROUS OBJECTS OUT OF THE DELIVERY ROUTINE AND SCAN
ROOM BEFORE DELIVERING THE MAGNET TO THE SCAN ROOM. THESE FERROUS
OBJECTS COULD CAUSE DAMAGES TO THE MAGNET, OR INJURIES TO PERSON
AROUND THE MAGNET. AT THE SAME TIME, PREVENT PEOPLE WITH MEDICAL AND
PHYSICAL CONTRAINDICATIONS FROM ENTERING THE MAGNET OPERATION AREA.
5
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
6
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
If the magnet is shipped to the hospital within the open-top ocean container, the first step
should be unpacking the shipping crate from the ocean container. Open the top cover of the
shipping ocean container and lift the shipping crate out by crane. Please prepare the crane
for this kind of operation.
NOTE: PLEASE PREPARE THE CRANE WITH ENOUGH CAPABILITY TO LIFT THE
SHIPPING CRATE OUT. PLEASE REFER TO SECTION 2-3 FOR DETAIL
INFORMATION ABOUT THE SHIPPING CRATE DATA. THE DIMENSION OF
THE OCEAN CONTAINER IS: WIDTH 2330MM, LENGTH 5900MM AND HEIGHT
2245MM.
If the shipping crate is unpacked from the ocean container at the seaport, the magnet should
be shipped to the hospital within the shipping crate only. The crane should also be prepared
to move the magnet within the shipping crate to the ground for a more safety magnet shipping
crate unpacking.
When lifting the shipping crate, please attach the stainless wire rope at the indicated point.
See illustration 2-1.
7
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
If the system needs to be stored before installation, it must be stored in a warehouse, covered
by waterproof materials and protected from weather. The storage temperature for the system
should be between –34 ℃and 60℃ and the relative humidity between 0 and 90 % (non–
condensing).
WARNING
IF THE MAGNET TEMPERATURE IS OUT OF THE RANGE FROM – 30℃ TO 45℃, THE
MAGNETIC PROPERTY COULD HAVE IRREVERSIBLE DAMAGE. UPTO MOST CARES
SHOULD BE PAID TO THE MAGNET TRANSPORATION AND STORAGE TEMPERATURE.
8
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
2400 1950
9
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
2100 MM
2400MM
To lift the shipping crate, please attach the stainless wire to the position, and just the outside
of the supporting squire bar as shown in Illustration 2-2.
10
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
2040
MAGNET DATA
Illustration 2-3
11
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
This section provides the magnetic fringe field layout of the GEHC 0.2T Profile magnet before
being unpacked out from the shipping crate.
WARNING
UP MOST CARES MUST BE TAKEN WHEN WORKING AROUND THE MAGNET. BE SURE TO
KEEP ALL FERROUS OBJECTS, EQUIPMENT AND PEOPLE WITH MAGNETIC FIELD
SENSITIVE IMPLANTS OUT OF THE MAGNETIC FIELD. DETAILS ARE ILLUSTRATED IN
TABLE 1-1.
X-Y magnetic fringe field, with the magnetic field shield from the shipping crate, as shown in
illustration 2-4.
Illustration 2-4
12
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
WARNING
THE 5 GAUSS LINE WIDENS ABOUT 1m WHEN THE CRATE IS OPENED. BE SURE TO
KEEP ALL FERROUS OBJECTS AND EQUIPMENTS THAT MAY BE INFLUENCED BY THE
MAGNETIC FIELD, AS WELL AS PEOPLE WITH MAGNETIC FIELD SENSITIVE IMPLANTS,
AWAY FROM THE MAGNET.
Y-Z magnetic fringe field, with the magnetic field shield from the shipping crate, as shown in
illustration 2-5
Y
Illustration 2-5
13
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
The procedures of unpacking and rigging the magnet out from the shipping crate, as well as
the procedures of delivering the magnet into the scan room are described in this section. Due
to the strong magnetic field, up most care should be paid to all the operations instructed in this
chapter. And all the procedures are used for GEHC 0.2T Profile M20552PY magnet only.
HANDLING INFORMATION:
This section contains safety measures that must be done when rigging the permanent magnet.
All these requirements must be done and checked before and during magnet rigging and
delivering operation.
The magnetic field could cause damages to the other devices (please refer to table 1-1). Be
sure to take actions to avoid all these potential damages.
There may be possibility that ferromagnetic tools may be get attracted to the magnet and
cause injury. Use nonmagnetic tools when working around the magnet. If nonmagnetic tools
(such as a bar or jack) cannot be prepared, up most care must be taken to make sure they do
not get drawn to the magnet, and put these tools far away (4m) from the magnet after the
usage.
DO NOT use steel wire to lift the Magnet. Use approved non–magnetic lifting straps or wire
rope.
DO NOT take off the OVERALL MASK covering the magnet before the magnet is placed in
the correct position in the scan room for leveling.
14
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
The Shock Indicator and Tip Indicator are pasted on the surface of the shipping crate. The
position and the examples are shown in the illustration 3-1(a).
Tilt Indicator
Shock Indicator
a.
b. c.
THE SHOCK INDICATOR AND TIP INDICATOR
Illustration 3-1
The Shock Indicator, as shown in illustration 3-1(b), is used to indicate whether the maximum
shock load is out of the 10G, which is specified in the magnet transportation requirements.
The Tilt Indicator, please refer to illustration 3-1(b), is used to show whether the maximum tilt
is out of the 30° specification set to the shipment.
15
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
NOTE: IF INDICATORS SHOW THAT THE SHOCK LOAD OR TILT DURING THE
TRANSPORTATION PROCESS ARE OUT OF THE SPECIFICATION, PLEASE
CONTACT THE GE REPRESENTATIVE OR LOCAL SERVICE SUPPORT
CENTER IMMEDIATELY.
Connect crane (forklift) to the rings in top of the crate and open the top plate of the shipping
crate. See Illustration 3-2.
NOTE: THE FOUR IRON RINGS WELDED ON THE TOP OF THE SHIPPING COVER IS
DESIGNED FOR LIFTING THE TOP COVER ONLY. DO NOT USE THESE RINGS
TO LIFT THE WHOLE SHIPPING CRATE.
NOTE: THERE IS A 3MM STEEL PLATE ATTACHED TO THE TOP OF THE CRATE. THE
WEIGHT OF THE SHIPPING CRATE TOP COVER IS ABOUT 150KG.
16
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
There are another 4 screws used to fix the crossing bar inside of the shipping crate. Please
release these 4 screws also before open the shipping crate sidewalls, refer to Illustration 3-3
(a). And Release screws used to fix the shipping crate and remove the 4 sidewalls of the
crate. See Illustration 3-3 (b). The Illustration 3-4 shows the unpacked shipping crate with
the magnet covered by the Aluminum waterproof membrane. Cut open this waterproof
aluminum membrane at the bottom seal edge carefully and remove this coverage. The
vacuum coverage should be removed immediately after the unpacking the shipping crate.
WARNING
PLEASE CONFIRM AGAIN THAT THERE ARE NO FERROUS OBJECTS AROUND (2.5
METERS) THE MAGNET BEFORE OPEN THE SHIPPING CRATE.
PLEASE USE THE NON-MAGNETIC SENSITIVE TOOLS TO CUT THE VACUUM PACKAGE
OPEN.
IF THERE ARE ANY OBJECTS ATTRACTED TO THE MAGNET, PLEASE GET IN TOUCH
WITH SERVICE REPRESENTITIVE IMMEDIATELY. DO NOT TRY TO TAKE THE OBJECT
OFF FROM THE MAGNET BY HAND; DO NOT TRY TO TAKE THE OBJECT OFF FROM THE
MAGNET BY YOURSELF.
(a)
17
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
Sidewall Screws: 44
Crossing Bar Screws, Screws material: Steel
2x2=4 Material Steel
(b)
18
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
OVERALL MASK
EYE BOLTS
WARNING
THE 5 GAUSS LINE WIDENS ABOUT 1m WHEN THE CRATE IS OPENED. BE SURE TO
KEEP ALL FERROUS OBJECTS AND EQUIPMENTS THAT MAY BE INFLUENCED BY THE
MAGNETIC FIELD, AS WELL AS PEOPLE WITH MAGNETIC FILED SENSITIVE IMPLANTS,
AWAY FROM THE MAGNET. REFER TO ILLUSTRATION 3-5, THE MAGNETIC FRINGE FIELD
WITHOUT THE MAGNETIC SHIELDING FROM THE SHIPPING CRATE.
19
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
20
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
Open the zippers and check the temperature indicators pasted on the magnet. Confirm
whether the magnet temperature exceeds 45℃ or not.
21
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
The temperature indicators record the highest temperature ever reached in the shipping
process. Individual indicator changes color from silver white, orange to black correspondingly
according to the different temperature. For example, as shown in illustration 3-7, the
temperature indicator shows that the highest temperature has exceeded 47℃ in the shipping
process.
Temperature indicator
TEMPERATURE INDICATOR
Illustration 3-7,
22
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
WARNING
THE 5 GAUSS LINE WIDENS ABOUT 1m WHEN THE CRATE IS OPENED. BE SURE TO
KEEP ALL FERROUS OBJECTS AND EQUIPMENTS THAT MAY BE INFLUENCED BY THE
MAGNET, AS WELL AS PEOPLE WITH MAGNETIC FIELD SENSITIVE IMPLANTS, AWAY
FROM THE MAGNET. SEE ILLUSTRATION 2-3, 2-4 ISOGAUSS LINE PLOTS FOR THE
MAGNETIC FRINGE FIELD.
3-2-4. Check the magnet for damages in the shipping process, and
confirm that no parts are missing.
NOTE: REPORT ALL THE DAMAGES OR MISSING, IF ANY, IMMEDIATELY TO THE LOCAL
GE REPRESENTATIVES. AT THIS TIME BE READY TO DESCRIBE TYPE OF
DAMAGE, TYPE OF EQUIPMENT, SERIAL NUMBERS AND IF POSSIBLE THE
ORDER NUMBER.
3-2-5. Release the eyebolts from the wooden lock, as shown in illustration
3-8.
NOTE: THE EYEBOTLS ARE MADE FROM STAINLESS STEEL, AND ARE DESIGNED TO
LIFT THE GEHC 0.2T PROFILE MAGNET ONLY. PLEASE MAKE SURE TO USE
THE STAINLESS STEEL EYEBOLTS TO LIFT THE MAGNET.
23
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
3-2-6. Open the zippers again and install the eyebolts to the magnet on the
top of the magnet. See illustration 3-9.
3-2-7. Release the wooden locks, which are used to fix the magnet legs to
the shipping crate. Please see illustration 3-10.
24
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
3-2-8. Prepare four Roller Skids on the ground. The magnet will be
delivered to the scan room on these Roller Skids.
3-2-9.Hook the eyebolts, lift the magnet and place it on the Roller Skids.
And then take off the eyebolts. Refer to illustration 3-11.
Please see Illustration 3-4 and Illustration 3-8 for the position and installation of the eyebolts.
NOTE: DO NOT REMOVE THE OVERALL MASK COVERING THE MAGNET DURING THE
RIGGING AND DELIVERING PROCESS. ALL THE SAFETY RISKS, SUCH AS
MAGNET FALLING DOWN, DURING THE LIFTING PROCESS SHOULD BE
CAREFULLY EVALUATED AND PREVENTED BEFORE THE LIFEING OPERATION.
HOOK THE EYEBOLTS, LIFT THE MAGNET AND PLACE IT ON THE ROLLER SKIDS
Illustration 3-11
NOTE: MAKE SURE TO CHECK THE DIRECTION OF THE MAGNET BEFORE DELIVERING
SO THAT THE FRONT AND BACK SIDES WILL BE CORRECTLY ORIENTED. THE
MAGNET MIGHT BE IMPOSSIBLE TO REVERSE INSIDE THE SCAN ROOM.
After the magnet is uncrated, the maximum magnet dimension is L2040 x W1300 x H1380.
The minimum dimensions for a straight path and a path with 90°turns for magnet delivery are
shown in Illustration 3-12 respectively.
25
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
Illustration 3-12
The magnet is completely enclosed by an OVERALL MASK to prevent ferrous objects from
being drawn to the magnet and damaging the magnet. DO NOT take off this OVERAL MASK
when delivering the magnet to the scan room.
Use the free standing posts, plastic chain to rope off the operation site during installation and
service activity. Warning signs should also be pasted to prevent people from entering the
magnetic fringe field area.
And the magnet room should be swept clear from ferrous objects before the arrival of the
magnet. This must be done and checked before delivery.
NOTE: CONFIRM THE DELIVERY ROUTE AND MAGNET ROOM ARE SWEPT
THOROUGHLY FROM FERROUS OBJECTS AND MAGNETIC FIELD SENSITIVE
DEVICES.
NOTE: THE MAGNET IS 10000KG IN WEIGHT, PLEASE CONFIRM WHETHER THE FLOOR
CAN SUPPORT THE MAGNET AND THE ROLLER SKIDS WON’T DAMAGE THE
FLOOR SURFACE. ADDITIONAL PADDING STEEL PLATE MAY BE NEEDED TO
PROTECT THE FLOOR, AND THIS PADDING PLATE SHOULD BE PREPARED BY
THE CUSTOMER. PLEASE REFER TO DIRECTION 2186839, SIGNA PROFILE III,
PRE-INSTALLATION MANUAL FOR DETAILS ABOUT ARCHITECHURAL
REQUIREMENT.
26
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
NOTE: PLEASE CHECK THE FRONT AND REAR SIDE OF THE MAGNET BEFORE
ENTERING THE SCAN ROOM. IT MIGHT BE VERY HARD TO ADJUST THE
FRONT AND REAR SIDES OF MAGNET INSIDE THE MAGNET ROOM.
There is a steel plate fixed under the floor (invisible from the surface), but its position is clearly
marked on the floor. See Illustration 3-13. Please refer to Direction 2186839, Signa Profile III
Pre-Installation Manual, SECTION 6 for details about FLOOR SHIELDING.
Illustration 3-13
27
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
Place the rubber plates, which will be padded under the feet of the magnet, at the correct
position. The co-geometric center of the four rubber plates should be within ±50mm of the
steel plate center. Jack the magnet up sufficiently at the jack applicable position and remove
the magnet from the roller skids. And then, anchor the magnet on these rubber plates. See
illustration 3-14 as examples (the floor shielding arrangement could be different in each
hospital). Please apply the jack at the correct position indicated by the labels on the overall
mask. Refer to Illustration 3-15.
Rubber Plate
Leg of the magnet
Steel sheet Floor (Plywood or concrete)
Illustration 3-15
WARNING
SPECIAL CARE MUST BE CONSIDERED TO USE THE JACK SINCE THERE IS STRONG
MAGNETIC FIELD UNDER THE MAGNET PRODUCED BY THE PERMANENT MAGNET.
WHEN MOVING THE JACK, KEEP THE DISTANCE FROM MAGNET MORE THAN 1 METER.
28
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
WHEN MOVING THE JACK AROUND THE MAGNET, YOUR BACK MUST FACE ON MAGNET
TO KEEP THE DISTANCE, AS ILLUSTRATION 3-16 SHOWS.
≥1m
Check and verify that the magnet center comes within ±50mm of the center of steel plate,
see Illustration3-13. If the center of the magnet is out of this specification, the location of the
magnet MUST be adjusted. Jack the magnet again, adjust the position of the rubber plates
and then anchor the magnet on these rubber plates until the magnet center within the
specification.
3-2-14. Take the Ring Protector Set and Overall Mask off and then begin
leveling the Magnet.
WANING
DO NOT TAKE THE OVERALL MASK AND THE RING PROTECTOR SET OFF FROM THE
MAGNET UNTIL THE LAST MINUTES YOU ARE READY TO DO THE MAGNET LEVELING.
WITHOUT THE PROTECTION FROM THE OVERALL MASK, THE FERROUS OBJECTS
COULD BE EASILY STUCK TO THE MAGNET TIGHTLY, WHICH COULD CAUSE DAMAGES
TO THE MAGNET AND IT COULD BE HARD TO TAKE OFF FROM THE MAGNET ( IF THERE
IS ANY THING STUCK TO THE MAGNET, PLEASE GET INTOUCH WITH THE INSTALLATION
REPRESENTIVE, OR CONSULT WITH GEHC SERVICE CENTER. DO NOT TRY TO REMOVE
THE STUCK OBJECTS BY YOURSELF. THE REMOVING OF THESE OBJECTS FROM THE
MAGNET MAY CAUSE SERIOUS PERSONNEL INJURIES).
29
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
WARNING
PEOPLE WITH ANY KINDS OF MAGNETIC FIELD SENSITIVE IMPLANT, SUCH AS CARDIAC
PACEMAKER, NEUROSTIMULATORS, BIOSTIMULATION DEVICES, ARE STRICTLY
FORBIDDEN FROM ENTERING THE SCAN ROOM.
NOTE: THE OVERALL MAGNET MASK AND THE RING PROTECTOR SET SET, WHICH
ARE UESD TO PROTECT THE MAGNET FROM BEING DAMAGED BY FLY-IN
OBJECTS, WILL BE REMOVED FROM THE MAGNET. PLEASE PREVENT THE
JACK AND OTHER FERREOUS OBJECTS FROM BEING ATTRACED TO THE
MAGNET AND CAUSING INJURUES TO OPERATOR AS WELL AS DAMAGES
TO THE MAGNET ITSELF.
NOTE: PLEASE PRESERVE THE OVERALL MASK AND THE RING PROTECTOR SET
PROPERLY AFTER FINISHING THE MAGNET LEVELING. THE OVERAL MASK
AND THE RING PROTECTOR SET MAY BE USED AGAIN LATER BY SERVICE
ENGINEERS.
NOTE: USE THE JACK AT THE CORRECT POINTS, WHICH ARE MARKED BY LABELS
ON THE OVERALL MASK. REFER TO ILLUSTRATION 3-15.
30
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
START
WITHIN THE Y
SPEC ? RE-INSTALL THE OVERALL
MASK
MAGNET LEVELING
Flowchart 4-1
31
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
. 4-3-1. Sweep the Magnet Room thoroughly for Iron powder and other
ferrous objects.
4-3-2. Take off the stainless steel OVERALL MASK and Ring Protector Set,
as shown in the illustration 4-1, which covers the Magnet.
TEMPERATURE INDICATOR
4-3-3. Cut an opening in the Plastic Cover near the slit. See Illustration 4-2.
Slit
32
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
Illustration 4-3
4-3-5. See Illustration 4-3. Measure the difference between the surface line
of the slit and water level height in the tube. There should be no
more than 2 mm difference between any of the 4 slits (at 4 corners
of the Magnet).
4-3-6. If there is more than 2mm difference, reinstall the Ring Protector Set
and the Overall Mask, tighten the rope and make the Overall Mask
cover the Magnet tightly, and then jack up the Magnet and insert
steel plates under the Magnet feet to level the magnet. Please refer
to Illustration 4-4.
33
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
Illustration 4-4
WARNING
NOTE: PLEASE USE THE JACK AT THE POINT INDICATED BY LABELS IN CASE OF ANY
DAMAGES TO THE MAGNET.
Illustration 4-5
34
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
WARNING
SPECIAL CARE MUST BE CONSIDERED TO USE THE JACK SINCE THERE IS STRONG
MAGNETIC FIELD UNDER THE MAGNET PRODUCED BY COUNTER MAGNET. WHEN
MOVING THE JACK, KEEP THE DISTANCE FROM MAGNET MORE THAN 1 METER. WHEN
MOVING THE JACK AROUND THE MAGNET, YOUR BACK MUST FACE ON MAGNET TO
KEEP THE DISTANCE, AS ILLUSTRATION 3-15 SHOWS.
≥1m
Illustration 4-6
NOTE: THE OVERALL MASK AND THE RING PROTECTOR SET SHOULD BE PROPERLY
PRESERVED FOR THE POTENTIAL USAGE IN THE FUTURE. DURING ANY
MAGNET RELATED MAINTENANCE OR MANGET TRANSPORTATION, THE
OVERALL MASK AND THE RING PROTECTOR SET MUST BE USED TO PROTECT
THE MAGNET TO PREVENT ANY DAMAGES AND INJURIES.
NOTE: BEFORE THE SYSTEM INTEGRATION AND MAGNETIC FIELD SHIMMING, THE
MAGNET TCU SHOULD BE SET TO 32.5℃. IT IS STRONGLY SUGGESTED TO
SET THE MAGNET TUC AT THE SAME NUMBER WITH THE FACTORY MAGNET
SHIMMING TCU SETTING. THE FACTORY SHIMMING TCU SETTING NUMBER
COULD BE FOUND FROM THE MAGNET “ ACCEPTANCE TEST REPORT “,
WHICH COULD BE FOUND IN THE PLASTIC FOLDER PLACED IN THE TOP OF
BOTTOM POLE FACE. AND IT IS IN THE SAME FOLDER USED FOR THE
LOCATOR CARD.
35
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
The purchaser of GE magnet shall only utilize qualified personnel (i.e., GE’s field engineers,
personnel of third–party service companies with equivalent training) to perform shimming
servicing on the equipment.
For field shimming, special tools and equipments are needed. And these tools/devices, which
are not included in the shipping collector, should be prepared by service engineers who
preside the shimming. Please refer to table 5-1 for tools/equipments details.
Tools/Equipments Quantity
MetroLab Teslameter PT2025 46-255839P2 1
Single Probe#2 or #3 46-255839P102 1
Single Probe Mapping Fixture 2141561 1
Operation System:Windows 2000 N/A 1
Service CD SV manual 2124628 / 2359125 (Class A) 1
Non-Magnetic Tools 1 set
Thermometer 1
Field Shimming kit 2370113 1
Wrench 2370034 1
TOOLS/EQUIPMENTS
Table 5-1
a. The temperature of the shimming room should be within the range from 22 to 26℃, which
can be measured by the Temperature Meter. Please keep the room temperature sable
and avoid fluctuating during the shimming process.
36
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
b. The magnet TCU should be set to 32.5℃ and heating the magnet up (This should have
began after the magnet leveling). It is suggested to set the magnet TCU at the same
number with the factory magnet shimming TCU setting. The factory shimming TCU
setting number could be found from the magnet “ Acceptance Test Report “, which could
be found in the plastic folder placed in the top of bottom pole face, and is the same folder
used for the locator card.
c. Take off the Overall Mask and the Ring Protector Set from the magnet.
e. Heat the magnet no less than 24 hours until the temperature of the magnet is stable
enough.
Frequency
mode
Drift of the central field should be less than 150Hz per 10 minutes and this situation should
last at least two hours. To measure drift, the probe should be placed in the ISO center of the
magnetic field ( in Plane 7 ). Please find more detail procedures from: Service CD (Part NO.
2124628/2359125), Profile 3/4 Installation (For Profile 4b) 6th STEP Sumitomo 2.
Setup (Mapping Fixture setup).
37
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
initial central field B1, and drifts to B2 10 minutes later, then the drift can get by: |B2-B1| ;
(|B2-B1|<150 Hz).
e. Confirm the table was installed properly (temporarily). Please refer to Service CD (Part
NO. 2124628/2359125), Profile 3/4 Installation (For Profile 4b) 3th STEP Table
Installation (Temporarily).
the table installation instruction for reference and turn the System Gradient Unit off.
NOTE: THE FIELD SHIMMING CANNOT START UNTIL ALL THE ABOVE REQUIREMENTS
ARE MET.
38
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
The field shimming cannot be finished until the following two specifications are met.
a. At the FOV of 35*40*40cm, the inhomogeneity of the field must be less than 45 ppm
(135 pts Pk-Pk measurement).
In the next section there is an introduction about how to shim the magnetic field to meet above
specifications.
* The “Acceptance Test Report” could be found in the plastic folder placed on the pole face of the bottom Yoke.
** Only required to the Profile/I system, or EPI function.
Table 5-2
39
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
Hardware setup
(Mapping fixture, probe)
No
Mechanical shimming Calculation
Section 5-4 -2 3.1)
Yes
|Z1| < 7ppm & |Z2| < 7
ppm
No
Execute Mechanical
shimming
Section5-4-2, 3.1).
Data Sheet
End
* There is no Transverse shimming in the GEHC 0.2T Profile Magnet mechanical shimming.
** By the “B0 Keep Mechanical shimming” method, the B0 field will be unchanged while harmonics could be
adjusted.
MAGNETIC FIELD SHIMMING PROCEDURE FLOWCHART
Flowchart 5-1
40
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
1. Release the Overall Mask and the Ring Protector completely from the Magnet.
2. Fix the single probe fixture onto the bottom ring of the pole piece.
For the detail procedures, please refer to Service CD (Part NO. 2359125 or
2124628), Profile 3/4 Installation (For Profile 4b) 6th STEP
Sumitomo Chapter 2 (Mapping Fixture setup).
3. Fix the single probe on the Horizontal bar, and mount it on the single probe fixture.
For the detail procedures, please refer to Service CD (Part NO. 2359125 or
2124628), Profile 3/4 Installation (For Profile 4b) 6th STEP
Sumitomo Chapter 2 Setup (Mapping Fixture setup).
4. Power on the Tesla-meter PT2025; set and confirm the Teslameter to the
Frequency Mode.
After running the software, please specify a directory for the software before a new
magnet is to be shimmed. Once it is specified, all the files the software generates
in the shimming process will save to this folder automatically. Please follow Section
5-4-2 1.1) Set Working Dir…..
41
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
Please refer to section 5-4 2.1 “Data Gathering” to details about how to gather
the magnetic filed data with the shimming software.
180
210 150
240 120
270 90
300 60
330 30
0
Start Measuring the Data from here
Table Side
GEHC 0.2T Profile Magnet Coordinate System
Magnetic Field inhomogeneity pk-pk ppm value can be read from the field map in the
software main window (refer to illustration 5-3). If the overall inhomogeneity is in spec
(see section 5-2), field shimming could be finished, otherwise, please go to the Step 5-
6.
42
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
Please refer to Profile Service CD (Part NO. 2359125 or 2124628), Profile 3/4
Installation (For Profile 4b) 6th STEP Sumitomo 9. Mechanical Shim (Axial)
for detail procedures about how to implement the mechanical shimming. The GEHC
0.2T Profile does not designed to implement the Mechanical Shimming, transverse
method
After the mechanical shimming is finished, please go back to 5-3-3. Only when Z1 or
Z2 are all less than 7 ppm, could the passive shimming begin as introduce in section
5-3-6.
If the mechanical shimming meets the spec of all Z1& Z2 less than 7ppm, one can
execute “scan passive shimming”, which is described in Section 5-2 3.4, to execute
the passive shimming. After scan and calculation finished, one can go through the
results and select a relative good shimming plan from the shimming calculation results
by evaluating the number of the shimming chips would be used, the location number to
be used and the anticipated homogeneity value to be achieved. Run the passive
shimming with the parameter setting as the selected target ppm and shims number to
get the passive shimming plan. Please refer to Section 5-2 3.5 as instruction to the
passive shimming.
For locate the shim chips, please refer Profile Service CD (Part NO. 2359125 or
2124628), Profile 3/4 Installation (For Profile 4b) 6th STEP Sumitomo.
43
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
5-3-7. Go to 5-3-3.
Compare with the specs in section 5-2. Do the passive shimming or mechanical
shimming again until the specs are met.
After the magnet shimming is accomplished, please fill the magnet installation
spreadsheet which could be found in the Shimming software Service CDROM (5122078).
NOTE: THE OVERALL MASK AND THE RING PROTECTOR SET SHOULD BE PROPERLY
PRESERVED FOR THE USAGE IN THE FUTURE. DURING ANY MAGNET
RELATED MAINTENANCE OR MANGET TRANSPORTATION, THE OVERALL
MASK AND THE RING PROTECTOR SET MUST BE USED TO PROTECT THE
MAGNET FROM ANY DAMAGES AND PERSONNEL INJURIES.
44
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
There is a specific shimming software is prepared to the field shimming of GEHC 0.2T Profile
magnet. This shimming software is mounted in the GEHC 0.2T Profile Magnet field shimming
Service CDROM (5122078), which is sent to the installation site with the magnet. To
implement the field shimming to the GEHC 0.2T Profile Magnet, the service engineers must
install this software to the Laptop first.
45
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
46
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
1. File Menu
File Manual
Illustration 5-4
When click on Button, one can select the target directory as desire. One is also allowed to
establish a new directory and rename it at this step.
47
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
1.2. New
To generate a new empty map file which is waiting for data gathering.
1.3 Open...
To open an existing map file or to open an existing psm file (passive shimming result file).
The software will automatically search the base map and show the result with the predicted
map. (The base map’s location is described in the psm file. If the base map file is moved to
another place, it should be edited with notepad or other ASCII editors).
1.5 Save
To save current map file.
48
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
The most recent used file will be listed for the convenience as shown in the illustration 5-7.
49
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
2. Data Menu
Manual Data gathering Dialog is used for recording the magnetic field value on each specified
points to generate a map. There are totally 135 points in 13 planes should be measured one
map: in plane 2~ plane 12, each plane has 12 angles (0 degree to 330 degree, 30 degree for
each step) and 1 point in each of plane 1 and plane 13. The B0 strength on the ISO center
(plane7) should be measured twice, firstly in the beginning of taking the map and secondly in
the end, for compensating the field drifting in the field mapping process. Please refer to
illustration 5-9 for the arrangement of the planes. The angles are marked on the shimholders
of the magnet. One can also have a look at the Profile Service CD (Part NO. 2359125 or
2124628 ), Profile 3/4 Installation (For Profile 4b) 6th STEP Sumitomo for more
details about the measuring and recording the data.
50
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
When using the software, the field value is combined by Prefix and suffix. For example, if the
read value for teslameter is 8.5012345MHz, the prefix 8.5 could be specify, and only “ suffix”
number is need to type in. After typing, please click "Accept" key or type “Enter” key on the
keyboard to accept the data.
"Overall Warning Bound(PPM +/-)" edit box lets you to set the overall warning boundary, if
field value typed in is outside of it, a warning will occur to let you be aware that abnormal data
is inputted. You can click right side to change the current point and recollect the data if you
made a mistake.
"Short Term Warning Bound (PPM +/-)" edit box is similar with the above edit box, which
focuses on the difference between the current field value and the last, theoretically there
should be no big difference between the two consequent data.
51
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
3. Analysis Menu
ANALYSIS MANUAL
ILLUSTRATION 5-10
OUTPUT WINDOW
ILLUSTRATION 5-11
One can watch Teslameter’s display value and tune the Top B0 plug and then Bottom B0 plug,
Sequentially until it is the same as what is in the output window, respectively.
Please refer to Profile Service CD (Part NO. 2359125 or 2124628), Profile 3/4
Installation (For Profile 4b) 6th STEP Sumitomo for detail procedures about implement
the mechanical shimming.
52
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
This function allows you to read the axial harmonic of the magnetic field. Only harmonics from
(0,1)~(0,12) are listed. The most important items are (0,1) and (0,2), or say Z1 and Z2.
The dialog shown in Illustration 5-12 is used to set the parameters for the “ Scan Shim”.
“ Scan Shim” function is used to assisting service engineer to find the best Passive Shimming
“ target “. Please refer to 3.5) Passive shimming for more intruction.
With Only ONE time setting here, one can execute passive shimming analysis and get a
series of simulation results simply and easily (Please refer to Illustration 5-15). The target
PPM can be from 0.01 to 100. “Shims limit” stands for the shimming chips number limitation
in each socket, and it should be less than or equal to 4. The default values are suitable to
most cases. You can also change them before run “scan passive shimming” in 3.4) to see if
better result could be generated (The better result comes from the comprehensive
consideration of the “Anticipated PPM” value and “Shims Number”. Both this two
parameter should be as small as possible).
Target PPM from 12 to 36, with delta of 0.2 and Shims Limit from 4 to 4 with delta of 1 is set
as the default value. When the best analysis results occur at the boundary, the Target PPM
value scanning range could be extended in have more analysis result around the target PPM
value. For Example: If the target is set as from 12ppm to 36ppm, while the best analysis
53
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
result is just near 12ppm, you could change the Target PPM range as from 5PPM to 14PPM
and run scanning again to see if there is an even better result could be found.
“ Scan Shim” function is used to assisting you to find the best Passive Shimming “ target “. To
execute scanning of passive shimming with the specified target ppm and shims number limit
in each pocket, which is described in 3.4), to get the simulated shimming. A dialog, as
shown in Illustration 5-28, will appear after clicking the “scan passive shimming” command. If
this is the first-time passive shimming (NO shims in the shimholders), keep “2nd or More
Iteration shim” unchecked, and the “Shim File of Last Time” Edit Box is therefore disabled.
Please refer to Illustration 5-13.
Otherwise (there are shims already on shim holders), please check the “2nd or More Iteration
shim” and specify the psm file of last time before the current loaded map is taken. Please
refer to Illustration 5-14 for reference.
54
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
After click on “OK” Button, the brief summary of analysis result for each target is shown in
Output Window as shown in illustration 5-15. ( One can click on “Scan Para.” Button to set
parameters for scanning pass shimming. Please refer to 3.4 ”set scan shim param. ”for
reference).
After scanning is finished, one can look through the brief summary, choose the best
anticipated result and execute 3.5) “the passive shimming” with the corresponding “ settings”.
Please consider the “shims number” and the “Anticipated PPM” comprehensively with main
principle the smaller “shims number” with smaller “Anticipated PPM” the better.
55
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
A dialog will appear and let you specify the passive shimming parameter as shown in
Illustration 5-16.
If this is the first-time passive shimming (NO shims on shim holders), then keep “2nd or More
Iteration shim” unchecked, as shown in the illustration 5-16, and the “Shim File of Last Time”
56
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
edit-box is therefore disabled, as shown in illustration 5-17. Otherwise (There are shims
already on shim holders), please check the “2nd or More Iteration shim” and specify the psm
file of last time before the current loaded map is taken. These items are the same as what in
3.4.
Please specify the “Target PPM” value and “Shim Limit” value you choose from “Scan passive
shimming” results (refer to 3.4). After click on “OK” button, the program will analysis, generate
and open the result with notepad. ( for detail information about the analysis result file, please
refer to section 4).
The string in the “Output Shim File” edit box is the name of the output shim file, which is
consistent with the name of the base map file as default ( Specified by user or assigned by the
software). The program will automatically add the anticipated inhomogeneity ppm value, target
ppm value to the file name, which has an extension of “.psm”. For example, if you type
“afterMech1” in this edit box as the file name, the target ppm is 13.6, the anticipated ppm is
22.5 after calculation, then the actually output shim file will be saved as the name
“afterMech1_22.5(13.6).psm” automatically. If one accept the result of the psm file and
perform shimming with it, please specify this name for the next iteration of passive shimming
if needed. (Please put this name in the “Shim File of Last Time”, as shown in Illustration 5-17.)
To interrupt the current scanning. Do not close the shimtool software when scanning passive
shimming is running.
57
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
Please try to find the section of "Display Shim Direction and Shim Location - Lower Delta
","Display Shim Direction and Shim Location - Upper Delta", "Display Shim Direction
and Shim Location - Lower Total “and "Display Shim Direction and Shim Location -
Upper Total" and execute passive shimming (put small shims into pockets on shim holders).
Note: the Ring index is from 1 to 6.
In the file, L1~L6 means ring1~ ring 6 on shim holder of lower side, U1~U6 means ring 1~ring
6 on shim holder of upper side. “”Nx (Sx)” means x shims(x=1,2,3,4) with North(South) pole up
putting into the specified pocket of the shim holder after the shim holder is lay on the ground
for both lower and upper shim holder. Please refer to illustration 5-19 for information.
58
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
Please refer to Profile Installation Service Profile Service CD (Part NO. 2359125 or
2124628), Profile 3/4 Installation (For Profile 4b) 6th STEP Sumitomo Passive
Shimming for detail procedures about implement the passive shimming, including remove the
shim holder, place the shimming chips, install the shim holder, etc.
59
GE HEALTHCARE GEHC 0.2T Profile Magnet Installation Manual
REV 1 DIRECTION 2370115-100
After the magnet shimming is accomplished, please fill the magnet installation spreadsheet
which could be found in the Shimming software Service CDROM (5122078).
you can ask engineering via this email account if you have any question in the installation
process.