SPECIFICATION
FOR
XRR-3332X

Seiji Tanaka

General Manager
Sales Department

Specification Number: T160093-S963 Rev. 0

Issued : 2016-04-15
Revised :
SPECIFICATION
XRR-3332X

To : Samsung Electronics Co., Ltd.

SPECIFICATION
FOR
XRR-3332X

Toshiba Rotating Anode X-ray Tube Assembly


for
Diagnostic Radiography

Contents Page

1. General Data ................................................................................................. 2

2. Maximum and Minimum Ratings ................................................................... 4

3. Dimensional Outline ...................................................................................... 6

4. Outgoing Test ............................................................................................. 12

5. Warning ...................................................................................................... 14

6. Cautions ...................................................................................................... 15

7. Export Control ............................................................................................. 19

8. Limitation of Product Liability ...................................................................... 20

9. Warranty on X-ray tubes ............................................................................. 21

10. Seasoning of Tubes .................................................................................... 22

Hereinafter, the company name "Toshiba Electron Tubes & Devices Co., Ltd" is
stated "TETD".

Issued : 2016-04-15
Revised :
1/25
SPECIFICATION
XRR-3332X

General Data

IEC Classification ......................................................................................... Class Ⅰ ME EQUIPMENT

Applicable Standards:
IEC60336:2005 IEC60522:1999 IEC60601-1:2005 IEC60601-1-2:2007 IEC60601-1-3:2008
IEC60601-2-28:2010 IEC60613:2010 IEC/TR60788:2004

Attachment:
Technical Data ................................................................................................................................. 1p
Tube Service Report ....................................................................................................................... 1p
Seasoning of Tubes ........................................................................................................................ 1p
Operation Manual ............................................................................................................................ 1p
Safety Notice on X-ray Tube ........................................................................................................... 1p
Information and Instruction for Assemblers and Users .................................................................. 1p
Appendix (Regarding compliance with DHHS CEF 21) ................................................................. 1p
Silicon Grease ................................................................................................................................. 2p

Electrical:
Circuit:
High Voltage Generator ............................................. Constant Potential High-Voltage Generator
Grounding ........................................................................................................... Center-Grounded
Nominal X-ray Tube Voltage:
Radiographic ....................................................................................................................... 150 kV
Nominal Focal Spot Value:
Large Focus ................................................................................................................................ 1.2
Small Focus ................................................................................................................................ 0.6
Nominal Anode Input Power (at 0.1s): (As shown in Fig.3)
60 Hz 50 Hz
Large Focus ......................................................................................................... 46 kW 42 kW
Small Focus ......................................................................................................... 20 kW 18 kW
Nominal Radiographic Anode Input Power:
60 Hz 50 Hz
Large Focus ......................................................................................................... 43 kW 40 kW
Small Focus ...................................................................................................... 19.5 kW 17.5 kW
Filament Characteristics .......................................................................................... As shown in Fig.4
Filament Frequency Limits ................................................................................................. 0 ~ 25 kHz
Emission Characteristics ......................................................................................... As shown in Fig.4

Issued : 2016-04-15
Revised :
2/25
SPECIFICATION
XRR-3332X

Motor Ratings:
Stator: XS-AL
Starting Running
Driven Frequency [Hz] 50/60 50/60
Input Power [W] 910 83
1) 3)
Voltage [V] 130 40
2)
Current [A] 7.8 2.3
4)
Min. Speed Up [s] 0.8 -
Capacitor [µF] 44 44
Note: 1) Applied voltage between common and main terminal.
2) Common current.
3) The every applied voltage must be never exceeded 110% of the above specification.
4) The speed-up time is allowed up to 110% of the above specification.

Anode Speed:
60 Hz ............................................................................................................... Minimum 3200 min-1
50 Hz ............................................................................................................... Minimum 2700 min-1
Stator Resistance:
Common-Main Winding .......................................................................................................... 9.4 Ω
Common-Auxiliary Winding .................................................................................................. 28.3 Ω
Resistance between Housing and Low Voltage Terminals ........................................ Minimum 2 MΩ
Normal Operating Range of the Housing Temperature ..................................................... 16 ~ 75 ℃
Mode of Operation ............................................................................................................. Intermittent

Mechanical:
Dimensions: (As shown in Fig.1)
Overall Length ................................................................................................................... 385 mm
Maximum Diameter ........................................................................................................... 140 mm
Target:
Anode Angle ................................................................................................................. 14 degrees
Diameter .............................................................................................................................. 74 mm
Construction ..................................................................... Rhenium-Tungsten faced Molybdenum
Permanent Filtration ................................................................... 1.0 mm Al / 75 kV IEC60522:1999
Radiation Protection:
Leakage Technique Factor .................................................................................... 150 kV, 3.4 mA
X-ray Coverage .................................................................................... 354 × 354 mm at SID 710 mm
Weight (Approx.) ......................................................................................... 12.5 kg(Without end cup)
13.1kg(With end cup)
High Voltage Receptacle ................................................................................. CLAYMOUNT MINI-75
Cooling Method ................................................................................................... Natural or forced air
Insert Model ......................................................................................................................... XRR-3332
Tube Housing Model Number ............................................................................................... XH-1023
Finish .......................................................................................................... Munsell 1.44G 9.20 / 0.13
Wiring:
Do not connect stator coil and temperature relay line in series circuit (see Fig.1).

Issued : 2016-04-15
Revised :
3/25
SPECIFICATION
XRR-3332X

Maximum and Minimum Ratings


(At any time, these values must not be exceeded.)

Maximum X-ray Tube Voltage:


Radiographic ....................................................................................................................... 150 kV
Between Anode (or Cathode) and Ground ................................................................................ 75 kV
Minimum X-ray Tube Voltage ..................................................................................................... 40 kV
Maximum X-ray Tube Current: (As shown in Fig.4)
Large Focus ........................................................................................................................ 600 mA
Small Focus ........................................................................................................................ 300 mA
Maximum Filament Current:
Large Focus ............................................................................................................................ 5.5 A
Small Focus ............................................................................................................................ 5.3 A
Filament Voltage:
Large Focus (At maximum filament current 5.5A) .................................................... 12.7 ~ 17.1 V
Small Focus (At maximum filament current 5.3A) ........................................................ 7.0 ~ 9.4 V
Filament Frequency Limits ................................................................................................. 0 ~ 25 kHz
Continuous Anode Input Power 1) .......................................................................... 142 W (200 HU/s)
Thermal Characteristics (As shown in Fig.5):
Anode Heat Content ........................................................................................... 210 kJ (300 kHU)
Maximum Anode Heat Dissipation .................................................................. 870 W (1226 HU/s)
X-ray Tube Assembly Heat Content ................................................................. 750 kJ (1056 kHU)
Nominal Continuous Input Power: 2)
Without Air-circulator ................................................................................ 150 W (13 kHU/min)

Note: 1) “Continuous Anode Input Power” is the X-ray tube anode input power.
Therefore, when tube is used for long time exposure or frequent exposures, the total
input power, which includes stator input power, filament heating power and average
input power, needs to be calculated not to exceed the “Nominal Continuous Input
Power”.
2) “Nominal Continuous Input Power” is specified at 750 kJ of full housing heat storage
which is generated by X-ray tube anode input power, stator input power and filament
heating input power.

Issued : 2016-04-15
Revised :
4/25
SPECIFICATION
XRR-3332X

Environmental Limits

Operating Limits:
Temperature .................................................................................................................. 10 ~ 40 °C
Humidity .......................................................................................................................... 30 ~ 85 %
(No condensation)
Atmospheric Pressure ............................................................................................... 70 ~ 106 kPa
Shipping and Storage Limits:
Temperature ................................................................................................................. -20 ~ 70 °C
Humidity .......................................................................................................................... 20 ~ 90 %
(No condensation)
Atmospheric Pressure ............................................................................................... 50 ~ 106 kPa

Issued : 2016-04-15
Revised :
5/25
SPECIFICATION
XRR-3332X

Fig.1 Dimensional Outline


Unit mm
E
G
C

63

F

193

4-M6 8 DEEP 130 D


92

140
4-M5 8 DEEP
O

41.27
385
45°

11.1
116
192

D
C S

76


L C

H A B C CLAYMOUNT MINI-75

140
53
81

62

FOCAL SPOT
C

CLAYMOUNT MINI-75

Issued : 2016-04-15
Revised :
6/25
SPECIFICATION
XRR-3332X

Fig.2 Rupture Disc

The Rupture Disc is mounted on anode side housing wall.


When external cover is removed, it is visibly recognized as labeled with label G) in next page.

Rupture Disc
Anode side wall

Followings are important notice of the Rupture Disc:

Function: When housing inner pressure reaches to the working pressure of the Rupture Disc, the
Rupture Disc collapses and vaporized oil and hot oil flow through the collapsed Rupture
Disc.
The external cover can protect the initial impact of oil flow, however, it can not protect oil
drop or oil leakage.

Caution: 1. Never touch the Rupture Disc or remove the label and the screws.
Otherwise, it could cause collapse of Rupture Disc.
2. Please note the Rupture Disc is not the device to protect X-ray tube from destruction
when over rated power input but just to reduce the impact of housing rupture.

Issued : 2016-04-15
Revised :
7/25
SPECIFICATION
XRR-3332X

Drawing of Labels

A) Main Label B) Regulation Label

C) Caution Label D) Polarity Labels

E) Caution Label F) Pointer Label


Inside of anode side ornamental cover

G) Caution Label
Labeled on Rupture Disc

H)Barcode Label
Model name
Serial Number and Tube Number
Print Day

Issued : 2016-04-15
Revised :
8/25
SPECIFICATION
XRR-3332X

Fig.3 Maximum Rating Charts


(Absolute Maximum Rating Charts)
2015-05-09_水村さんデータより作成 2015-05-09_水村さんデータより作成
2015-05-12_水村さん修正データより

Conditions: Tube Voltage


XRR-3332X Maximum Rating
Constant XRR-3332X Maximum Rating
Potential High-Voltage Generator
60Hz 1.2mm Stator Power Frequency 60Hz 60Hz 0.6mm

Nominal Focal Spot Value: 1.2 Nominal Focal Spot Value: 0.6
80kV 70kV60kV50kV 40kV 70kV 60kV 50kV
600 300

90kV 80kV 40kV


500 100kV 250 90kV
110kV 100kV

TUBE CURRENT [mA]


TUBE CURRENT [mA]

400 125kV 200 110kV


125kV
150kV
300 150 150kV

200 100

100 50

0 0
0.01 0.02 0.05 0.1 0.2 0.3 0.5 0.7 1 2 3 5 7 10 0.01 0.02 0.05 0.1 0.2 0.3 0.5 0.7 1 2 3 5 7 10
0.03 0.07 0.03 0.07
EXPOSURE TIME [s] EXPOSURE TIME [s]

2015-05-09_水村さんデータより作成 2015-05-09_水村さんデータより作成
2015-05-12_水村さん修正データより
Conditions: Tube Voltage
XRR-3332X Maximum Rating
Constant Potential High-Voltage XRR-3332X
GeneratorMaximum Rating
50Hz 1.2mm Stator Power Frequency 50Hz
50Hz 0.6mm
Nominal Focal Spot Value: 1.2 Nominal Focal Spot Value: 0.6
70kV 60kV 50kV 40kV 60kV 50kV
600 300
80kV 70kV 40kV

500 90kV 80kV


250
100kV
TUBE CURRENT [mA]

90kV
TUBE CURRENT [mA]

110kV
400 200 100kV
125kV 110kV
125kV
300 150kV 150
150kV

200 100

100 50

0 0
0.01 0.02 0.05 0.1 0.2 0.3 0.5 0.7 1 2 3 5 7 10 0.01 0.02 0.05 0.1 0.2 0.3 0.5 0.7 1 2 3 5 7 10
0.03 0.07 0.03 0.07
EXPOSURE TIME [s] EXPOSURE TIME [s]

Issued : 2016-04-15
Revised :
9/25
SPECIFICATION
XRR-3332X
2000-11-09_E7823FX_10%DOWN
2015-05-09_XRR-3332X_1000mA→600mA
Fig.4 Emission & Filament Characteristics 2015-05-11_水村さんデータにより見直し

THREE-PHASE
Constant Potential High-Voltage Generator
Focal Spot : 1.2 mm
Nominal Focal Spot Value: 1.2
600 18
80kV

500 100kV 16

FILAMENT VOLTAGE [V]


125kV
TUBE CURRENT [mA]

400 14
150kV

300 12
Ef

200 10

60kV
100 8
40kV

0 6
3.9 4.1 4.3 4.5 4.7 4.9 5.1 5.3 5.5

FILAMENT CURRENT [A]

2000-11-09_E7823FX_10%down
For Reference2015-05-09_XRR-3332X_440mA→300mA
Only
2015-05-11_水村さんデータより

THREE-PHASE

Focal Spot : 0.6 mm


Nominal Focal Spot Value: 0.6

300 9

80kV
250 8
100kV
FILAMENT VOLTAGE [V]
TUBE CURRENT [mA]

125kV
200 7
150kV

150 6
Ef

100 5
60kV
40kV
50 4

0 3
3.6 3.8 4.0 4.2 4.4 4.6 4.8 5.0 5.2 5.3

FILAMENT CURRENT [A]

For Reference Only

Issued : 2016-04-15
Revised :
10/25
2015-05-09_E7239Xベース SPECIFICATION
     180W→150W XRR-3332X
     E72369Xの値*0.833

Fig.5 Thermal Characteristics


Housing

X-ray Tube Assembly Heating / Cooling Curve


750
700 150W
COOLING
600
HEAT STORAGE [kJ]

500

400

300

200

100
HEATING
0
0 20 40 60 80 100 120 140 160 180 200

TIME [min]

2007-10-22_E7886Xの60Wなし

ANODE
Anode Heating / Cooling Curve
210
200

180
COOLING
160
HEAT STORAGE [kJ]

140

120

100

80
142W
60
HEATING
40

20

0
0 1 2 3 4 5 6 7 8 9 10
TIME [min]

The heating curves are showing example of average


input power to anode in operation.

Issued : 2016-04-15
Revised :
11/25
SPECIFICATION
XRR-3332X

Outgoing Test

The following table shows the outgoing test conditions and screening criteria, not contain the contract
specification. Therefore, TETD can revise them anytime and they may not be necessarily applied to
every delivered product. However, even if they are changed, TETD shall guarantee all specified items
mentioned in this contract specification.

Test Condition Limits


Test Item Ep Ip t N ti If F Sym Note
Cir Unit Min. Max.
[kV] [mA] [s] [s] [A]
High Voltage S,FW 160 1.5 60 1 - Adj L - - - - 1,2,3

Emission & Tube 75 307 - - - - L If A 4.12 5.11


S,FW 1
Current Characteristics 75 133 - - - - S If A 4.14 5.06
- - - - - 5.5 L Ef V 12.67 17.13
Filament Characteristics DC
- - - - - 5.3 S Ef V 7.14 9.66
Width - 1.7
S,FW 75 307 Adj - - Adj L - mm 1,4
Focal Spot Size Length - 2.4
(Slit Method) Width - 0.9
S,FW 75 133 Adj - - Adj S - mm 1,5
Length - 1.3
Coasting Time - (AC 100V, 50Hz) - s 20 - 10

Noise - (AC 40V, 50Hz) - db - 67 7,10

X-ray Leakage - 150 3.4 - - - Adj L - Note8 - 0.44 6

Permanent Filtration - 75 1 - - - Adj L - Note9 1.0 1.3 6

Appearance - - - - - - - - - Note 2
Terms (Abbreviated)
Cir : Circuit Ep : Tube Voltage Ip : Tube Current
t : Duration N : Number of Exposure ti : Interval Time
If : Filament Current F : Focus Sym : Symbol
Adj : Adjust L : Large Focus S : Small Focus
Ef : Filament Voltage S,FW : Single-Phase Full-Wave Rectified

Issued : 2016-04-15
Revised :
12/25
SPECIFICATION
XRR-3332X

Note
1. Anode, rotating (AC 100V, 50Hz)
2. It should not be abnormal.
3. Run without a kick during at least 1 minute.
4. Slit Magnification factor ×1.
5. Slit Magnification factor ×2.
6. Anode, rotating (AC 40V, 50Hz)
7. Noise level measuring conditions.
(1) Environment: acoustic shield room.
(Background noise level : under 40 dB)
(2) Place X-ray tube assembly on the soft mat.
(3) The distance between X-ray port surface and microphone is 25 cm.
(Precision sound level meter: Comply with IEC Publ. 179A characteristics A.)

Microphone of sound level meter

Radiation Port
25cm

8. Unit: mGy/h
X-ray leakage test condition is measured at tube current of 2mA.
Judgment is done by converting measured value at tube current of 3.4mA.
(Specification: Max. value is 0.44mGy/h)
9. Aluminum equivalent, unit: mm
10. Put on X-ray tube axis horizontally.

Issued : 2016-04-15
Revised :
13/25
SPECIFICATION
XRR-3332X

Warning

Warning to Interface with X-ray Generator


1. Housing Rupture
Never input over rated power to X-ray tube assembly.
If the input power exceeds tube specification, it causes the over temperature of anode, insert tube
glass shatter and ultimately the following serious problems due to generating over-pressure by oil
vaporization inside housing assembly.
In such a critical condition causing housing rupture by over load, the safety thermal switch can not
protect X-ray tube even if it works.
*Housing sealing parts rupture.
*Human injury including burns due to hot oil escape.
*Fire accident due to flaming anode target.
The X-ray generator should have a protective function which manages input power to be within
tube specification.

2. Over Load Protection


X-ray tube assembly has a thermal protection device to notice the generator to terminate the input
power. However it only works for gradual temperature increase when the tube is operated within
the specified X-ray condition (but too long repetitive use). And the protection is not effective in
such a case that rapid increase of target temperature destroys insert tube under overrated input.
In this sense, over load protection is definitely required on X-ray generator.
From the point of effective and reliable operation, the followings are recommended:
* Software control:
CPU calculates the total input power and controls the target temperature.
This protection is effective to human error.
* Independent protection circuit:
Independent shut down circuit separated from X-ray control unit is effective and reliable.

3. Rupture Disc
Never touch the Rupture Disc located on anode side wall and the screws.
And when rotor cable is attached, be careful that the rotor cable should not be located on or over
the Rupture Disc for the following mechanical reason.
The Rupture Disc is the device which protects X-ray tube assembly from harmful destruction by its
mechanical collapse.
When extreme over rated power is input to X-ray tube assembly, the Rupture Disc activates and
hot oil escapes through the collapsed Rupture Disc for safety purpose.
For the reliable mechanical process, nothing should be located on or over the Rupture Disc.

4. Electric Shock
To avoid the risk of electric shock, this equipment must only be connected to a supply with
protective earth.

Issued : 2016-04-15
Revised :
14/25
SPECIFICATION
XRR-3332X

Cautions

Caution to Interface with X-ray Generator


1. Over Rating
X-ray tube assembly can be broken with applying just one over rated shot.
Please read the technical data sheets carefully and follow the instructions.

2. Permanent Filtration
The total filtration and the distance between X-ray focal spot and human body are regulated
legally. They should be complied with the regulation.

3. Safety Thermal Switch


X-ray tube assembly has safety thermal switch to prohibit further input power when the tube
housing reaches to the temperature of switch-open.
The switch should be hooked up with the X-ray generator which control output power to X-ray tube
assembly.
The switch is not recommended connecting stator coil in series circuit.
Even if the switch works, never turn the system power off. The cooling unit should be activated if
used with the system.

4. Unexpected Malfunction
X-ray tube assembly may have the risk to be unexpectedly malfunctioning due to life termination
or failure. If the serious problems caused by the above risk is expected, you are requested to have
a contingency plan to avoid such a case.

5. New Application
If you use the product with new application not to be mentioned in this specification or with
different type of X-ray generator, please contact to us for confirming its availability.

Issued : 2016-04-15
Revised :
15/25
SPECIFICATION
XRR-3332X

Caution for Installation, Adjustment and Maintenance


1. Qualified Persons
Only qualified persons who have technical training and professional knowledge can handle X-ray
tube assembly.

2. Fragile Glass
X-ray tube is assembled with glass, therefore, it can be broken with the mechanical vibration or
pulsed shock over 19.6m/s2 (2G).
Careful handling is required to treat or transport.

3. Ground Terminal
X-ray tube assembly has ground terminal. Ground cable should be connected.

4. High Voltage
All X-ray tubes operate at voltages high enough to kill through electrical shock. Never touch the
high voltage delivered plugs or terminals.
When direct access to such parts is required, the primary circuit should be disabled and high
voltage capacitors/cables discharged.

5. High Voltage Plug


High voltage plug should be cleaned up and free from any physical damages. Silicon compound
application is required for high voltage stability.

6. Rotating Anode
(1) Test rotation of the anode with the large focus filament at the preheat condition. At this time, do
not apply the tube voltage.

(2) When the hand switch "READY" button is pressed, the filament will become brighter and the
anode target will start rotating. The anode for normal speed will reach the predetermined
number of revolution in 0.4 to 1 sec., and it will continue rotation for some time after the stator
drive input is cut off.
The anode for high speed shall be automatically braked to reduced speed and continue
coasting after the drive input to the stator is cut off.

Issued : 2016-04-15
Revised :
16/25
SPECIFICATION
XRR-3332X

(3) Ensure that the anode target rotates in the proper direction.
X-ray tube should always rotate counter-clockwise, viewing from the cathode side.

- Confirm direction of rotation -

(4) If the inside of any tube unit cannot be observed (for example, the tube unit with a beryllium
window, etc.), check the tube unit to ensure again that the terminal numbers of the drive circuit
of the low-voltage terminals are properly connected.

7. Operation Atmosphere
X-ray tube assembly is not allowed to be used in the atmosphere of flammable or corrosive gas.

8. Protective Cover
X-ray tube assembly is not allowed to be used without the protective cover attached.

9. Handling
Appropriate jig or tools are required for tube installation to avoid physical damages.

10. Returning Tube


X-ray tube assembly should be repackaged with the original material when it is returned back for
quality examination in our factory. Be careful to put the tube upside cathode. If the packaging is
not proper, the tube may not be correctly examined. TETD does not warrant it if returned X-ray
tube assembly is in damage at receiving.

Issued : 2016-04-15
Revised :
17/25
SPECIFICATION
XRR-3332X

Caution in Operation
1. X-ray Radiation
X-ray tube assembly should have the beam limiting equipment mounted on the X-ray port to
protect unnecessary radiation.

2. Dielectric Oil
X-ray tube assembly has dielectric oil contained for high voltage stability. As it is poisonous for
human health, if it is exposed to the non-restricted area, it should be disposed as following to the
local regulation.

3. Operation Atmosphere
X-ray tube assembly is not allowed to be used in the atmosphere of flammable or corrosive gas.

4. Disposal
X-ray tube housing is lined with lead to protect unnecessary radiation.
The lead powder or vapor is harmful for human health.
Dispose the scrapped products according to the requirement of local regulation.

5. Adjust the Tube Current


Depending on the operating conditions, the filament characteristics might be changed.
This change might lead to the over rate exposure to X-ray tube assembly.
To prevent the X-ray tube assembly from being damaged, adjust the tube current regularly.
Besides when "ROTANODE" has arcing problem in a long time use, adjustment of the tube
current is required.

6. X-ray Tube Housing Temperature


Do not touch on X-ray tube housing surface just after operation due to high temperature.
Stay X-ray tube to be cooled.

7. Operating Limits
Prior to usage, please confirm the environmental condition is within the operating limits.

8. Any Malfunction
Please contact to TETD immediately, if any malfunction of the X-ray tube assembly is noticed.

Issued : 2016-04-15
Revised :
18/25
SPECIFICATION
XRR-3332X

Export Control

1. US Export Administration Regulations


This product is free from the U.S. Export Administration Regulation.

2. Others
Distribution of the products and/or devices which incorporate the products may require prior
approval of or notification to the regulatory authorities and/or the relevant government authorities.

Issued : 2016-04-15
Revised :
19/25
SPECIFICATION
XRR-3332X

Limitation of Product Liability

In case you receive a claim from a third party that any loss or damage to property, bodily injury or death
of a person was caused by a defect of the Component, you shall immediately notify TETD of such
claim and consult with TETD for any actions to be taken. In any event, TETD's liability shall be confined
to the extent reasonably foreseeable and proximately caused by the defect of Component with a
limitation of aggregated amount paid by you for the Components.
Provided, however, TETD shall not be liable in the cases, where,

1. it was impossible for TETD to discover the defect based upon the state of scientific or technical
knowledge at the time of delivery to you,

2. the defect is due to the compliance with your instruction regarding the specification or design,

3. you failed to incorporate fail-safe design to your products in consideration of the reasonably
expected failure ratio/pattern of the Components incorporated therein,

4. the defect is due to the compliance with mandatory regulations/standards issued by the public
authorities; or

5. the defect did not exist at the time of delivery.

Issued : 2016-04-15
Revised :
20/25
SPECIFICATION
XRR-3332X

Warranty on X-ray tubes

1. TETD warrants the X-ray tube to be free of manufacturing defects which will impair their normal
operational life during the warranty period, provided that the X-ray tubes are used within the
ratings and in accordance with TETD’s instructions and specifications.

2. The warranty period extends for sixteen (16) months from the date of the original Bill of Lading or
twelve (12) months from the date of original installation, whichever occurs first.
If the X-ray tube shall prove to be defective during the warranty period, TETD shall refund or
replacement he original amount.

3. Notification of the claim shall be received by TETD within three (3) months after discovery of
failure, filling in TETD’s report form. If return of the defective X-ray tube is required by TETD, each
return shall be made without delay, and in accordance with TETD instruction.

4. The warranty shall not apply to defects resulting from accidents, alterations, abuse or misuse, or
improper installation.

Issued : 2016-04-15
Revised :
21/25
SPECIFICATION
XRR-3332X

Seasoning of Tubes

WARNING
All radiation safety rules must be observed. In order to protect all personnel from harmful
X-ray radiation exposure, close the collimator and / or block the tube assembly radiation
port with at least a six (6) mm thick piece of lead.

This instruction has been prepared for personnel who install or service the TOSHIBA Rotating Anode
X-ray tube Assemblies - “ROTANODE”

This procedure establishes the proper method to enable the ROTANODE X-ray tube to be 'seasoned'
when it is put into operation for the first time, or when it has not been used for more than one week, or
when the daily kV used is greater then 120 kV. The seasoning procedures must be carried out to
assure subsequent trouble-free operation and to maximize X-ray tube life.

Do not perform an exposure with no X-ray tube current flows. Be sure to have X-ray tube
current, at any time during exposures.

REQUIREMENTS:
1. Use the larger focal spot of the tube for radiographic exposures. Use the circuit specified in
technical data.
2. The tube voltage and current must be monitored for all exposures during this seasoning
operation.
3. The anode rotation should be done with normal speed (50Hz or 60Hz motor drive).
4. Do not exceed the ratings as detailed in the technical data provided with the ROTANODE X-ray
tube.
5. Do not exceed the rating of the Starter or High voltage Generator as detailed in the technical data
provided.

WARNING
All radiation safety rules must be observed. In order to protect all personnel from harmful
X-ray radiation exposure, close the collimator and / or block the tube assembly radiation
port with at least a six (6) mm thick piece of lead.

PROCEDURES:
Select the applicable procedure for each X-ray tube from the procedures described below.

CAUTION - If any abnormality in kV or mA is observed during these procedures, reduce the kV until no
abnormality is observed and then increase the kV at a slower rate than detailed in the
applicable table.
(The maximum tube testing voltage must be strictly limited to the maximum rated tube
voltage of the unit.)

NOTE - Even if arcing occures during the seasoning process, by complying with the CAUTION
detailed above, and by completing the seasoning procedure, then the arcing will be eliminated
and the X-ray tube may be operated normally without damage to the ROTANODE.

Issued : 2016-04-15
Revised :
22/25
SPECIFICATION
XRR-3332X

1. Seasoning Procedure at Installation for ROTANODE when Continuous Fluoroscopy is


available
a. Perform continuous fluoroscopy at 50 kV and Approx. 1 mA for 2 minutes and then gradually
increase the tube voltage as detailed in the 'Continuous Load' section of the 'Seasoning Schedule
at Installation' table below.
b. After a pause of 4 minutes, perform a 100 msec radiographic exposure every minute, increase the
tube voltage according to 'Short Time Load' section of the 'Seasoning Schedule at Installation'
table below.

NOTE - The mA to be used for this procedure is 50 % of the maximum mA available at maximum kV
for the larger focal spot.

2. Seasoning Procedure at Installation for ROTANODE when Coninuous Fluoroscpy is not


available.
Perform a 100 msec radiographic exposure every minute beginning at 50 kV and then increase the
tube voltage according to 'Short Time Load ' section of the 'Seasoning Schedule at Installation' table
below.

NOTE - The mA to be used for this procedure is 50 % of the maximum mA available at maximum kV
for the larger focal spot.

3. Seasoning Procedure when the ROTANODE has not been used for a week, or when the
maximum kV available is greater than 120 kV, and when Continuous Fluoroscpy is available.
a. Perform continuous fluoroscopy at 50 kV and Approx. 1mA and then gradually increase the tube
voltage as detailed in the 'Continuous Load' section of the 'Seasoning Schdule at Starting' table
below.

NOTE - The maximum tube voltage after 1 minute should be the maximum daily kV used.

b. After a pause of 2 minutes, perform a 100 msec radipgraphic exposure every 30 seconds,
increase the tube voltage according to 'Short Time Load' section of the 'Seasoninng schedule at
Starting' table below.

NOTE - The maximum kV should be limited to the maximum daily used kV.
NOTE - The mA to be used for this procedure is 50 % of the maximum mA available at maximum kV
for the larger focal spot.
NOTE - This procedure should be performed daily when the ROTANODE is operated regularly above
120 kV.

4. Seasoning Procedure when the ROTANODE has not been used for a week, or when the
maximum kV available is greater than 120 kV, and when Continuous Fluoroscopy is not
available.
Perform a 100 msec radiographic exposure every 30 seconds, beginning at 50 kV and then increase
the tube voltage according to 'Short Time Load ' section of the 'Seasoning Schedule at Starting'
table below.

NOTE - The maximum kV should be limited to the maximum daily used kV.
NOTE - The mA to be used for this procedure is 50 % of the maximum mA Available at maximum kV
for the larger focal spot.

Issued : 2016-04-15
Revised :
23/25
SPECIFICATION
XRR-3332X

NOTE - This procedure should be performed daily when the ROTANODE is operated regularly above
120 kV.

Seasoning Schedule at Installation

A: Max. Rated Tube Voltage 150kV A


150 B: Max. Rated Tube Voltage 125kV
C: Max. Rated Tube Voltage 100kV

130 B
A. B
TUBE VOLTAGE [kV]

110
C C

90

70

50

2
0
0 4 8 12 16 20 24 28 32 36 TIME [min]

Continuous Load Short Time Load


(One time / min.)

Recess

Seasoning Schedule at Starting


150
Daily Operating Tube
140 Voltage

120
TUBE VOLTAGE [kV]

100

80

60

1
0
0 2 4 6 8 TIME [min]

Continuous Short Time Load


Load (One time / 30 s)

Recess

Issued : 2016-04-15
Revised :
24/25
SPECIFICATION
XRR-3332X

History of Revision

Date Page Revised Contents Revised Reason

Issued : 2016-04-15
Revised :
25/25