Professional Documents
Culture Documents
PREFACE 3 MAINTENANCE 12
Scope 3 Calorimeter Maintenance 12
Related Instructions 3 General Cleaning Instructions 12
Important Information 3 Fuse Ratings 12
Intended Usage 3 Electric Ignition Problems 12
Explanation of Symbols 4
Safety Information 4 6772 CALORIMETRIC THERMOMETER 13
General Specifications 4
Environmental Conditions 4
DRAWINGS 14
Provisions for Lifting and Carrying 4
1341 with 6775 & 2901 Electrical Hookup Diagram 14
1341 with 6772 Electrical Hookup Diagram 14
ASSEMBLE THE CALORIMETER 5
Parts for the 1341 Calorimeter 15
Related Instructions
No. Description
205M Operating Instructions for the 1108 Series
Oxygen Combustion Vessel
207M Analytical Methods for Oxygen Bombs
545M Operating Instructions for 6775/6775A
Digital Thermometer
201M Limited Warranty
www.parrinst.com 3
Parr Instrument Company
PREFACE (CONTINUED)
1. Unpack the calorimeter carefully and check 5. Check the calorimeter bucket, noting the three
the individual parts against the packing list. If dimples in the bottom of the bucket which rest
shipping damage is discovered, report it im- on supporting pins in the bottom of the jacket.
mediately to the delivering carrier. Unpack The single dimple must always face forward
the jacket first and set it on a sturdy bench or when setting the bucket in the jacket.
table in a location that is reasonably free from
drafts and is protected from sources of radiant 6. Connect the ignition unit. While any electrical
heat, preferably in an air conditioned room. system capable of furnishing approximately
Temperature changes in the room should be 23 V can be used to ignite the fuse in the oxy-
minimal. gen combustion vessel, most users will prefer
to use a Parr 2901 Ignition Unit for this purpose.
There should be convenient access to running This unit operates from any standard electrical
water, to a drain and to an appropriate grounded outlet to provide the proper low voltage firing
electrical outlet. About 0.75 m2 (8 ft2) of work- current, providing also a convenient push
space will be required. Access to a chemical switch, indicating lamp and connecting termi-
balance sensitive to 0.1 mg and to a solution, nals.
trip or torsion balance capable of weighing up to
3.0 kg with 0.1 g sensitivity will be required. The Connect one of the lead wires from the calo-
1341 Calorimeter requires availability of oxygen, rimeter jacket to the terminal labeled “10CM”.
99.5 % purity, with appropriate connection, Connect the second lead to the terminal labeled
2500 psig, maximum. “Common”. Plug the power cord into an appro-
priately grounded electrical outlet. The power
2. Set the cover on the jacket. Remove the two supply cord is the main electrical disconnect
screws from the side of the jacket and attach device for the ignition unit. After the unit has
the stirrer drive motor. Check the stirrer shaft to been plugged into an outlet do not press the
be sure that it turns freely and slip the drive belt firing button unless the lead wires inside the
onto the motor and stirrer pulleys. jacket are connected to the combustion vessel.
www.parrinst.com 5
Parr Instrument Company
OPERATING THE CALORIMETER
Operating the 1108 Oxygen Combustion 5. Let the stirrer run for 5 minutes to reach equi-
Vessel librium before starting a measured run. At the
end of this period record the time on the timer
Detailed instructions for preparing the sample and
of the 6775 Digital Thermometer and read the
charging the 1108 Oxygen Combustion Vessel are
temperature.
given in Instruction Manual, No. 205M. Follow these
instructions carefully, giving particular attention to the 6. Read and record temperatures at one-minute
precautions to be observed in charging and handling intervals for 5 minutes. Then, at the start of the
the oxygen combustion vessel. 6th minute…
the rate of the temperature change becomes 13. Wash all interior surfaces of the combustion
constant over a period of 5 minutes. vessel with a jet of distilled water and collect the
washings in a beaker.
11. After the last temperature reading, stop the mo-
tor, remove the belt and lift the cover from the 14. Remove all unburned pieces of fuse wire from
calorimeter. Wipe the thermistor shaft and stirrer the combustion vessel electrodes; straighten
with a clean cloth and set the cover on the A37A them and measure their combined length in
support stand. Lift the combustion vessel out of centimeters. Subtract this length from the initial
the bucket; remove the ignition leads and wipe length of 10 cm and enter this quantity on the
the vessel with a clean towel. data sheet as the net amount of wire burned.
Alternatively, the correction in calories is located
12. Open the knurled knob on the oxygen combus- on the card of the 45C10 fuse wire.
tion vessel head to release the gas pressure be-
fore attempting to remove the cap. This release 15. Titrate the combustion vessel washings with a
should proceed slowly over a period of not less standard sodium carbonate solution using meth-
than one minute to avoid entrainment losses. yl orange or methyl red indicator. A solution of
After all pressure has been released, unscrew the 0.0709 N sodium carbonate is recommended for
cap; lift the head out of the cylinder and place it this titration to simplify the calculation. This is
on the support stand. Examine the interior of the prepared by dissolving 3.76 g of Na2CO3 in water
vessel for soot or other evidence of incomplete and diluting to 1L. NaOH or KOH solutions of the
combustion. If such evidence is found, the test same normality may be used.
will have to be discarded.
16. Analyze the combustion vessel washings to
determine the sulfur content of the sample if it
CAUTION! Do not have any part of exceeds 0.1 %. Methods for determining sulfur
the body in the pressure relief path. are discussed in Instruction Manual No. 207M.
www.parrinst.com 7
Parr Instrument Company
CALCULATING THE HEAT OF COMBUSTION
t = tc - ta - r1 · (b-a) - r2 · (c-b)
a = time of firing, min
b = time (to nearest 0.1 min) when the tem- Thermochemical Corrections
perature reaches 60 % of the total rise, min Compute the following for each test:
r1 = rate (temperature units per minute) at e2 = correction for heat of formation of sulfuric
which the temperature was rising during acid (H2SO4), cal
the 5 min period before firing, °C/min
e2 = 13.7 · c2 · m
r2 = rate (temperature units per minute) at which
the temperature was rising during the 5 min
period after time c. If the temperature was e3 = correction for heat of combustion of fuse
falling instead of rising after time c, r is wire, cal
negative and the quantity -r · (c-b) becomes
e3 = 2.3 · c3 when using Parr 45C10 NiCr
positive and must be added when comput-
fuse wire
ing the corrected temperature rise, °C/min
e3 = 2.7 · c3 when using 34 AWG iron fuse
c1 = volume of standard alkali solution used in
wire
the acid titration, mL
m = mass of sample, g
W = 2426 cal/°C
m = 0.9936 g
= 3.234 °C
e1 = 23.9 cal
= 7841 cal/g
www.parrinst.com 9
Parr Instrument Company
STANDARDIZING THE CALORIMETER
= 2392 cal/°C
www.parrinst.com 11
Parr Instrument Company
DISCUSSION OF CALCULATIONS (CONT.) MAINTENANCE (CONT.)
To improve the precision and simplify the procedure for measuring and recording the temperature
rise in a 1341 Calorimeter, Parr offers the 6772 Calorimetric Thermometer. Readings are taken with
a thermistor sealed in a stainless steel probe. This two-channel precision calorimetric thermometer
features Parr’s newest generation microprocessor control. The thermometer uses a Linux operating
system with communication and file management systems comparable to a PC. A bright, color,
touch screen display and data input system utilizes a graphical interface for easy instrument setup
and operation. A USB port is available for use with a balance, printer, or barcode reader. (For more
information on barcode capability, please contact Parr’s sales staff.) An Ethernet port is available for
data transfer using TCP/IP communications. No additional ignition source is required as the 6772
Calorimetric Thermometer acts as an ignition unit.
Specifications:
• Thermistor Probe
• One Probe – Standard
• Two Probes – Optional
• 0.0001 °C Temperature Resolution
• Absolute Accuracy:
±0.1 °C (without calibration)
±0.05 °C (with calibration)
• Repeatability:
±0.002 °C (Single Point)
• Linearity:
±0.002 °C (10 °C Span)
• Ethernet network communications
• Balance & Printer Port: USB
• Updates via the Internet
www.parrinst.com 13
Parr Instrument Company
DRAWINGS
1341 1168E2
CALORIMETER THERMISTOR PROBE
(BUCKET)
POWER CORD
34E2 (115V)
1203EEE (10A/220V CONT EU)
2198EEE (10A/250V BRAZIL)
1981EEE (10A/220V CHINA)
1202EEE (13A/250V BS PLUG)
POWER CORD
34E2 (115V) 6775
1203EEE (10A/220V) DIGITAL THERMOMETER
1341
CALORIMETER 1168E2 THERMISTOR PROBE (BUCKET)
1759
PRINTER
6772
A2170E USB PORT CALORIMETRIC
PRINTER CABLE THERMOMETER
POWER CORDSET
A719E (115V)
ETHERNET 1200EEE (250V, CONT EU)
PORT 1456EEE (220V)
2121EEE (250V, BRAZIL)
EXTENSION CORD
A1877E (115V)
A1877E2 (230V)
A1877E3 (230V EURO)
3 4
2 10
5
1
7 9
www.parrinst.com 15
204M R10 10/08/2021