Document Information:
Document Author: Gaub, Dawn {DSSD~MANNHEIM}
Business Area / Unit: Global Platforms & Support
Confidentiality: Confidential
Document Class: Device Master Record
Document Type: Device Master Record
Document Creator: Gaub, Dawn {DSSD~MANNHEIM}
Document Lifecycle Status: Effective
Valid From: 12-Jun-2023 11:26:42 (UTC)
Valid To:
Document Title: cobas_pro_QRG_V3-3_EN
Document Number: 0000000000001200000680782
Document Version: 01
Template: No
Electronic Signatures:
Signed By: hoffma15 (Andre Hoffmann {DSSDCS})
Role: Author
Signature Differentiation: Technical Writer
Signed Date: 05-Jun-2023 07:39:05 (UTC)
Publication information
Where to find information The User Assistance contains all information about the
product, including the following:
• Routine operation
• Maintenance
• Safety
• Troubleshooting information
• Configuration information
• Background information, for example, about analytical
principles
General attention
Incident reporting
r Inform your Roche representative and your local
competent authority about any serious incidents
which may occur when using this product.
Table of contents
Main workflow ........................................................................ 8
Quick reference: Before routine operation ................. 9
Quick reference: Pre-routine workflow .............. 9
Quick reference: Replacing reagents.................. 11
Quick reference: During routine operation.................. 16
Quick reference: Running patient samples....... 16
Quick reference: Checking test results .............. 18
Quick reference: Performing calibration............ 19
Quick reference: Performing QC........................... 20
Quick reference: After routine operation...................... 22
Quick reference: Post-routine workflow ............ 22
Reference information.......................................................... 24
Quick reference: Reagents and consumables
- ISE.................................................................................. 24
Quick reference: Reagents and consumables
– c 503 ............................................................................ 26
Quick reference: Reagents and consumables
– e 801 ............................................................................ 30
Quick reference: cobas e flow HIV Duo ........... 32
Maintenance logs ....................................................... 33
Main workflow
Main workflow
Software Instrument
1 Logging on
7 Downloading parameters
To verify QC results:
u Related topics
• Quick reference: Replacing reagents (11)
3 Loading reagent packs at 1. Press the Reagent Loader button and wait
the e 801 analytical unit until it stops flashing.
2. If reagent packs were unloaded, check the
status indicators:
: Current bottle
2. Pull up the pipe lifter until it locks.
3. Dry each aspiration pipe with a separate
lint-free cloth.
4. Replace the bottle.
: Current bottle
3. Lift the aspiration pipe and hold it.
4. Dry each aspiration pipe with a separate
lint-free cloth.
5. Replace the bottle.
y Replacing reagents
8 Replacing solid waste 1. When the status indicator on top is lit, open
boxes – e 801 the consumables drawer.
2. Remove the full solid waste box and
replace it with a new one.
3. Close the consumables drawer.
4. Open the transparent door and press the
status button corresponding to the
replaced solid waste box.
5. Close the transparent door.
y Replacing reagents
– : Test is selected.
– : Result is available.
≥20 mm ≥20 mm
2 Viewing test result details 1. On Routine > Results, choose the sample
in question.
: Calibration is recommended.
y Post-routine workflow
6 Cleaning ECL sipper nozzles 1. Open the cover of the sipper station.
and liquid level sensors –
2. Clean the ECL sipper nozzles and liquid
e 801
level sensors from top to bottom.
– Clean with alcohol.
– Then clean with deionized water.
3. Close the cover of the sipper station and
the main cover.
y Post-routine workflow
Reference information
In this section
Quick reference: Reagents and consumables - ISE (24)
Quick reference: Reagents and consumables –
c 503 (26)
Quick reference: Reagents and consumables –
e 801 (30)
Quick reference: cobas e flow HIV Duo (32)
Maintenance logs (33)
ISE Reference Electrolyte REF Refer to Reference solution for each ISE
Method Sheet measurement enabling measurements on
08392013190
the ISE analytical unit
Bottle (2 L)
Dead volume 250 mL
ISE Diluent DIL Refer to Dilution of sample material
Method Sheet
04880480190
Bottle (2 L)
Dead volume 100 mL
ISE Internal Standard IS Refer to IS measurement before and after each
Method Sheet sample measurement
04880455190
Bottle (2 L)
Dead volume 100 mL
ISE Cleaning Solution/ - Pour fresh Cleaning the ISE electrode flow path and
Elecsys SysClean dilution vessel
11298500316 (in position 2 of daily wash rack)
(in position 1 and 2 of weekly wash rack)
Bottle (100 mL)
y Reagents and consumables of the ISE analytical unit
Bottle (9 x 12 mL)
ISE Deproteinizer - Refer to Cleaning the ISE reagent flow path using
Method Sheet maintenance action 28 ISE Reagent Flow
20763071122
Path Wash
Bottle (6 x 21 mL)
EcoTergent - Refer to Prepare 2% EcoTergent solution (1:50
Method Sheet dilution with deionized water) to clean the
06544410190
rinse stations and the racks.
Glass ampoule
(10 x 3 mL)
ISE Standard High ISE HIGH Refer to Calibrating of ISE
(S2); Method Sheet
11183982216 (in position 2 of calibrator rack)
ISE S3
Calibrating of ISE
Refer to Method Sheet (in position 3 of calibrator rack)
Glass ampoule
(10 x 3 mL)
Na electrode (yellow) - Refer to
Method Sheet
10825468001
Cleaning wire - -
CLEANING WIRE 3 ASSY
08906238001
(Ø 0.2 mm for ISE probe)
Bottle (2 L)
(back of drawer)
EcoTergent ECO-D 28 days Additive for the incubation bath to reduce
onboard formation of bubbles, inhibit growth of micro
08063354190
organisms, increase water conductivity
needed for level detection.
NaOHD NAOHD 84 days Alkaline wash solution for reagent probes and
onboard reaction cells for maintenance action 17
08063451190
System Wash.
Material number
Photometer lamp
LAMP HALOGEN ASSY 12V/50W
04813707001
Reaction cells(a)
Reaction Cell for c 503 / c 513
07700814001
Spring holder
SPRING HOLDER
11709933001
Spring
SPRING L 913561
11228854001
(a) Storage temperature: 7-40 °C (44.6-104.0 °F). For details about expiration date, refer to the product label.
y Consumables of the c 503
Material number
Cleaning wire
CLEANING WIRE 3 ASSY
08906238001
(Ø 0.2 mm for S1 probe)
(Ø 0.26 mm for S2 probe)
(Ø 0.5 mm for wash nozzles)
S2 Sample probe incl. 1 seal piece
PROBE S2 (ST) ASSY
08762660001
Rack sticker
LABEL FOR STD-RACK
12024128001 (orange)
12024144001 (dark blue)
12024152001 (light blue)
12024136001 (light green)
12024098001 (yellow)
12024101001 (pink)
12024080001 (brown)
Can be used on racks for all analyzers
(a) Storage temperature: 7-40 °C (44.6-104.0 °F). For details about expiration date, refer to the product label.
y Consumables of the c 503
Print Date: 18-Nov-2015 16:48:09 (UTC)
Print Comment:
Roche Diagnostics
Title: cobas_pro_QRG_V3-3_EN Document Number: 0000000000001200000680782
Version: 01
cobas® pro integrated solutions · Software version 02-03Status:
· QuickEffective Valid3.3
Reference Guide · Publication version from: 12-Jun-2023 11:26:42 (UTC)
Confidentiality: Confidential Content Page 29 (40)
30
Product names
Material number
AssayTip/AssayCup tray
including Solid waste containers (WasteLiner)
05694302001
SysClean cup
LFC Cup ASSY
07485425001
Subresult Subresult
Host
Final result
Main result
Subresults
Maintenance logs
Maintenance logs are checklists that you can use to
document your maintenance.
Version: 01
SSU SSU name: Serial number: Month: Year:
Print Comment:
Monthly O1 S2 Date/ID
Cleaning the water container (for SSU Mat. No. 08464502001 only) 11 9
Confidentiality: Confidential
Roche Diagnostics
As required at your discretion O1 S2 Date/ID Date/ID Date/ID
Title: cobas_pro_QRG_V3-3_EN
Cleaning the surface of the 2D barcode reader 2 –
· Quick
Cleaning the water container (for SB Mat. No. 09205675001 only) 11 9
Effective
SB4 SB name: Serial number:
Cleaning racks 3 –
Valid3.3
SSU – SB – General 1 Operator time, in minutes. 2 System time, in minutes.
Print Comment:
Daily O1 S2 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31
Confidentiality: Confidential
Roche Diagnostics
Running a weekly wash rack 3 40
Title: cobas_pro_QRG_V3-3_EN
Every 2 weeks O1 S2 Date/ID Date/ID Date/ID
Monthly O1 S2 Date/ID
· Quick
Eliminating clogging in a sample probe 26 4
Effective
Replacing a sample probe 26 4
Version: 01
AU: c 503 AU name: Serial number: Month: Year:
Print Comment:
Weekly O1 S2 Date/ID Date/ID Date/ID Date/ID Date/ID
Confidentiality: Confidential
Roche Diagnostics
Cleaning the reagent probes 2 –
Title: cobas_pro_QRG_V3-3_EN
Cleaning the wash nozzles 2 –
Cleaning the reaction cell covers and the ultrasonic mixer covers 5 –
Monthly O1 S2 Date/ID
Effective
Eliminating clogging in a sample probe (S1) 26 4
Print Comment:
Daily O1 S2 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31
Confidentiality: Confidential
Roche Diagnostics
Cleaning the rinse and the wash stations 5 –
Title: cobas_pro_QRG_V3-3_EN
Cleaning the ProCell aspiration pipes 6 12
Effective
Cleaning the consumables drawer 5 –