Muhammad Adnan Ahmed; Awais Iqbal; Engineering Team Leader Zulfiqar Shaikh; Tahir 8aig; Hasan Amir, Sabir Ali Calibration procedure of weigh brdge. Dear Adnan/Awais,
The following is the calibration procedure of weight indicator of weigh bridge.
8142 PRO+ Metier Toledo Digital Weight indicator Calibration Procedure. Step#1: Open front cover by four screws & find four dip switches. Step#2 Switch on dip switchttl & FI will appear on display. Step#3 Hold enter button till F1.6 appeared then pressed 0 button to display Step#4 Pressed enter twice & count down will start from 15 to 0 then "Add Ld" will appear on display. Stetp#5 Placed test weight on Weigh Bridge & pressed enter then put test weight value 41040kg then pressed enter & count down will start from 15 to 0. Step#6Pressed exit & now calibration is completed so revert dip switch position & if found difference then repeat procedure. 8142Pro+ Specifications Excitation Voltage: +10VDC, power up to eight 3500 analog load cells. Span range: 3 ~ 32m V Zero range: 0 ~ 25mV Regards, Amjad Akeel From: Sent: To: Cc: Subject: