Professional Documents
Culture Documents
2. Applicable system
MRT-1503, MRT-2003, MRT-2001, MRT-1501
3. Applicable date
This change is applicable to systems shipped in April 2014 and later
4. Contents
The 100M-HUB in EXCELART-series systems is changed as follows due to part
discontinuation.
100M-HUB: BSM43-1379 (FXC ES108M2) → BSM43-1783 (FXC ES1008MTP)
Conventional New
The BSM43-1379 and BSM43-1783 are fully compatible with each other.
FIELD SERVICE MEMO
Ref. No. :
FSM-MR3265
Page :
2 of 7
[New 100M-HUB (FXC ES1008MTP) setup and installation procedure]
The procedure for installing the new 100M-HUB (FXC ES1008MTP) in the control cabinet of
an EXCELART-series system is described below.
It is necessary to change the port setting to "Auto Negotiation" mode or "100M HALF" mode
depending on the system and to remove the mounting bracket for the FXC ES108M2 if it is
present.
(1) If the mounting bracket for the FXC ES108M2 is present at the 100M-HUB installation
location, remove it. The FXC ES1008MPT can be installed using the magnets on the bottom.
<Button setting>
• OPERATION mode • CONFIG mode