Professional Documents
Culture Documents
PD Pro
Techimp
USER MANUAL
notification.
Neither this manual nor any of its parts shall not be reproduced or transmitted in any form either
electronically or physically, including photocopying and recording, without the express written
1 INTRODUCTION....................................................................................................................................... 7
INTRODUCTION ................................................................................................................................. 15
LOGIN .............................................................................................................................................. 15
CHOOSE LANGUAGE.......................................................................................................................... 19
LOGOUT ........................................................................................................................................... 19
DATABASE........................................................................................................................................ 22
ACQUISITION .................................................................................................................................... 36
PROCESSING .................................................................................................................................... 59
INTERNAL PD ................................................................................................................................ 77
SURFACE PD ................................................................................................................................ 81
CORONA PD .................................................................................................................................. 83
The following user’s manual describe the usage of the Techimp PD Pro software in all its features,
highlighting the best practices and the standard usage for a correct detection of the signal relevant
to Partial Discharges (PD) acquisitions, the management of such data and the analysis for a correct
evaluation of the eventual phenomena occurring and their harmfulness for an insulation system.
This Page Was Intentionally Left Blank
2 PRELIMINARY ACTIONS
This section describes all the preliminary actions to be taken prior to the usage of the Techimp PD
Pro software.
In order to run the Techimp PD Pro in an optimal way, it is necessary to have it installed on a laptop
- RAM: 8GB
In order to correctly install the Techimp PD Pro the following two software constraints must be
satisfied:
Non complying with the minimum hardware or software requirements listed here above can result in
experiencing errors during the installation, standard operation or in a loss of performance of the
Techimp PD Pro.
SOFTWARE INSTALLATION
Before starting the Techimp PD Pro installation procedure, it is necessary to make sure to satisfy
the hardware and software requirements listed in sections 2.1 and 2.2.
The installation can be started by double-clicking on the Techimp PD Pro installer provided by
In the following figures (from Figure 2-1 to Figure 2-11), screenshots of the dialog windows for the
installation are reported. During a standard installation, the parameters reported in the dialog
Figure 2-1
Figure 2-2
Figure 2-3
As shown in the following Figure 2-4, some additional software is installed along with the installation
of the Techimp PD Pro. Such software is required for the correct operation of the features of the
Techimp PD Pro.
Figure 2-4
Figure 2-5
Figure 2-6
Figure 2-7
Figure 2-8
Figure 2-9
Figure 2-10
Figure 2-11
At the end of the installation procedure, Techimp PD Pro can be run from the shortcut located on
the desktop or from the Start menu. The software is now ready for use.
3 SOFTWARE DESCRIPTION
INTRODUCTION
In the following sections, the Graphical User Interface (GUI) windows of Techimp PD Pro will be
LOGIN
Once the user has double-clicked on the shortcut icon on the desktop, the first GUI windows of the
• Username: admin
• Password: admin
The login page allows managing the access to the data and the functionalities of the Techimp PD
Pro software in a shared computer, allowing only the pre-configured users to access the main page.
Once the login procedure has been successfully completed, the window in Figure 3-2 will be
presented. The window in Figure 3-2 is the main page of the Techimp PD Pro where all the main
MAIN PAGE
The main page of Techimp PD Pro allows recalling the main features of the software plaT-Form. As
depicted in Figure 3-2, the main modules of the software are represented by means of hexagonal
icons.
When the mouse pointer is on one of such icons, the relevant sub-features are highlighted and
activated.
In the upper-right side, some icons related to settings management (users, language, license, etc.)
can be found. All of these icons together with the related features and tools will be described in the
next paragraphs.
Acquisition
Database
USER MANAGEMENT
The user management section of the Techimp PD Pro is composed of two different icons highlighted
in Figure 3-3. The first one is the Add User icon: an operator can create a new user profile. The
second is the Users editing icon: a user profile can be created or deleted.
Add User
Edit Users
By clicking the Add User icon, the window in Figure 3-4 is displayed. All the fields (username, email,
password) of the window in Figure 3-4 shall be filled in to create a new user profile.
When the mandatory fields are filled in the Add user icon in the center-right part of the window is
enabled and, by clicking on it, the user is created and it can be used later on.
Figure 3-4 Add new user wizard
From the main page, by clicking on the Users editing icon, the window in Figure 3-5 is displayed.
Delete User
Edit User
In this window, Figure 3-5, all the existing users are listed and the operator can edit or delete them
By clicking on the Choose language icon the drop-down menu, Figure 3-6, will appear and the
operator can select the desired Techimp PD Pro language. Additional languages may be included
in the software.
LOGOUT
By clicking on the Logout icon, Figure 3-7, the user session in use will be terminated and a new login
Logout
In order to use the Techimp PD Pro software, it must be activated by means of an activation key. A
license key has been provided by Techimp with the software. In order to activate it, it is necessary
to recall the License window of Figure 3-8 from the Main page icon, located on the upper right part
of Figure 3-2.
Offline Activation
License
.xml file Online Activation
Properties
.xml file
Upload Icon
Software version
Figure 3-8 Manage licenses window
Based on how the keys can be activated the license activation procedure can be divided into two
typologies: offline and online. In the following, (3.7.1 and 3.7.2) these two procedures are described.
In the bottom part of the Manage Licenses window, the Techimp PD Pro software version together
The off-line procedure has to be used when it is not possible to connect to the Internet the computer
where the Techimp PD Pro software is installed. In such cases, the first part of this procedure can
be done on a machine that has internet access, while it can be concluded on the target machine with
no Internet connection.
The first part of the procedure is composed of the steps listed below.
• Click on the Manage license icon, in the upper-right side of the Techimp PD Pro main page;
• Click on the Activation offline icon (represented by the wrench in Figure 3-8). The software
will create a .dat file located in “C:/User/«Logged-in user»/Documents” which must be sent
to Techimp.
Once the file that was created has been sent to Techimp, the first part of the procedure is concluded.
After the first part is completed, Techimp will send back two license files: QlmLicense.xml and
QlmProductProprieties.xml. Such files are required in order to finalize the second part of the
procedure on the target computer. The steps to follow are in the following:
• Click on the Manage license icon, in the upper-right side of the Techimp PD Pro main page;
• Use the License and Properties slots to import the two .xml files sent by Techimp;
The procedure is then finalized and the required Techimp PD Pro features are now enabled.
This simplified procedure has been designed for those machines (PC, laptop, server, …) that have
an Internet connection.
In order to complete the Techimp PD Pro activation, follow the below procedure:
• Open the Techimp PD Pro software;
• Click on the Manage license icon, in the upper-right side of the Techimp PD Pro main page;
• Click on the Activation on-line icon (represented by the network icon, Figure 3-8)
The procedure is then finalized and the required Techimp PD Pro features are now enabled.
DATABASE
Techimp PD Pro exploits a database structure to efficiently manage, store and organize the
acquisition data. In the following sections, all the features and tools related to the data managing will
be described. All the database management functionalities can be accessed via the dedicated icons
available on the center-left side of the main page window as shown in Figure 3-9. The database
Database Module
Query Registry
In sections 0, Error! Reference source not found., 3.8.3 and 3.8.4 the database sub-modules are d
escribed.
3.8.1 CREATE NEW REGISTRY
A registry can be considered as a folder in which all the data acquisitions related to a detection point
in an electrical asset are stored. Moreover, general and technical information on the measurement
campaign is related to the registry. Hence, in this way a link will be always present between the PD
A new registry can be created by means of the Create new Registry icon in the Techimp PD Pro
main page of Figure 3-9. By clicking on the icon, a specific wizard procedure will help the operator
in the creation of a new registry. In Figure 3-10 the first step of the wizard is reported.
In this first step, these following characteristics of the electrical asset are requested:
• Rated Voltage
• Manufacturer
• Vintage
• Asset type, where the available asset typology are: cable, GIS, rotating machine, switchgear
The more information is filled in, the more focused the queries in a second time can be. Once at
least the mandatory fields are filled in, the Next button in the bottom-right side becomes active and
the operator can proceed to the second step. The technical information that can be inserted in the
second step window of the wizard is dependent on the selected asset. As an example, in Figure
• Detection points number and the identification labels per each detection point;
all the asset types. Moreover, per each asset, there are also specific data, as an example: the cable
length for cables, the machine typology for rotating machines or the bushing type for transformers.
To confirm and save the new registry item, click on the Save icon in the bottom-right side of the
window (Figure 3-11), otherwise click on the Previous icon to return to the previous page. Finally,
the information window of Figure 3-11 informs the operator that the registry creation has been
finalized.
The Query: registry sub-module can be exploited to verify and find the last registry created (see
Filters
Plus Icon
An existing registry can be retrieved by means of the Query: Registry icon in the Techimp PD Pro
main page of Figure 3-9. By clicking on the icon, the window in Figure 3-13 appears where all the
saved registries are listed, and an operator can apply query filters, find a registry and edit it.
By clicking on the Plus icon, present on the left side of a row, more detailed information is shown
and new icons appear in the bottom-left side of the exploded row (Figure 3-14).
Import
Acquisition
Copy
Modify Acquisition
Delete
Moreover, if there are acquisitions linked to the selected registry, the Techimp PD
Pro software does not allow the operator to directly delete both the registry and all
the data associated with it (Figure 3-16). But, in this case, the software will force the
operator to delete the acquisitions at first and then it is possible to delete the registry
Figure 3-16 Warning: the registry contains acquisitions are you sure you want to delete it?
In the upper part of the Query Registry window (Figure 3-13), a light-blue label Filters is available.
By clicking on it, its row will be exploded as shown in Figure 3-17. In the Filters area there are four
static filters, organized in four different rows, and also a dynamic filter represented by the drop-down
menu in the fifth row and indicated by the More filters label. The available static filters, useful for
• Registry ID;
• Customer name;
• Asset type;
• Location name.
If more than one filter is used, they ones used are combined together using the AND logic condition.
In order to apply the required filters, it is necessary to click on the Apply icon on the right side of the
Filters area. In order to clear all the filter fields, it is necessary to click on the Reset all icon (aside
The filter settings can also be saved by clicking on the Save icon and recalled or displayed the recent
searches by clicking on the respective icons on the upper-right side of the Query Registry window
(Figure 3-13).
Saved Queries
Queries
history
Filters Area
Apply, Delete
or Save filters
icons
Customized
Filters
Concerning the dynamic filters, Figure 3-18 displays some quantities that can be chosen as a search
filter. Even for the dynamic filters, more than one filter can be applied and combined using the AND
logic condition.
Figure 3-18 Dynamic filters
An existing acquisition can be retrieved by means of the Query: Acquisition icon on the Techimp PD
Pro main page of Figure 3-9. By clicking on this icon, the window in Figure 3-19 appears. Here all
the acquisitions saved are listed and, an operator can apply research filters, find acquisition and
manage it.
Plus
Icon
corresponding acquisition are shown, as depicted in Figure 3-20. The four icons beside the T-F Map
Saved
Queries Queries
Filters history
Open in
Processing Export
acquisition
Delete
acquisition
Show more
detail
On the right side of the Plus button, there is a checkbox. By means of it, an operator can select more
than one acquisition and it can apply to them the same command as an example, it can send them
In the upper part of the Query acquisition window (Figure 3-19), as for the Query registry window,
there is the Filter zone by which an operator can apply search filters. For further detail about the
filters and their property refer to section Error! Reference source not found..
3.8.4 QUERY: TEST PLAN
An existing Test Plan can be searched by means the Query: Test Plan icon in the Techimp PD Pro
main page of Figure 3-9. By clicking on this icon, the window in Figure 3-21 appears. Here all the
An acquisition in Test Plan mode can be started by clicking on the Acquire icon that is present on
An operator can edit or delete a Test Plan by means of the respective icons placed on the right side
Saved
Queries Queries
Filter
history
s
Edit Start an
Delete Acquisition
Regarding the search filters in the upper part of the Query Test Plan window (Figure 3-21), they have
the same features already explained in the Par. Error! Reference source not found..
TEST PLAN EDITOR
In this section, the Techimp PD Pro Test Plan Editor module and its functionalities are described.
The main purpose of the Test Plan module is to suit as much as possible the factory quality control
measurement procedures. It allows an operator to define different test procedures having each one
of them its own measurement steps defined in terms of supply voltage, warning thresholds (in mV
Given a quality control test, the Test Plan Editor has to be used at the beginning for the customization
of the test procedures. In order to create a new Test Plan click on the Test Plan Editor icon in the
Techimp PD Pro main page of Figure 3-22. After that, when an operator wants to perform the test
according to the procedure, the Test Plan Acquisition modality has to be used (see section 0).
The Test Plan configuration window of Figure 3-23 is composed of five areas:
• Measurement plan it allows to create and define the steps of the measurement procedure;
• Sync where all the settings related to the sync channel are present;
• Calibration where the desired PD channels are selectable together with the selection of
IEC60270 calibrations if they are available or valid for the equipment under test;
• Additional Test Plan keys where it is possible to add additional fields, mandatory or not to be
Sync area
Measurement
Plan area
Acquisition
parameters
area
Calibration
area
Additional Test
Plan Keys area
Channels selector
Import Calibration
Figure 3-23 Test Plan editor window
The first row in the upper part of the window allows to define the name to associate with the Test
Plan. Beside the box name, a selector allow to perform the Test Plan according to the IEC60270.
In the Measurement Plan area, each row of the table represents a step of the measurement
• Name: a meaningful name is required (i.e. a progressive number or the test voltage);
• Warning threshold [V or pC]: a threshold value for the acceptance of the result, visualized
The area dedicated to the Acquisition settings includes the following parameters:
• Time length [s]: it is the total length of the displayed and acquired waveform for each pulse;
• Dead time [s]: the short time after a Trigger event when a detector cannot process another
event;
• Max pulses number and Acquisition Timeout [s]: are the parameters relevant to each
acquisition. The acquisition stops when the Max pulse number is reached. In case the set
number of pulses is not reached, the acquisition stops when the timeout is reached.
• Phase Shift: if the synchronization signal in use has a different phase with respect to the
voltage applied to the EUT, it is possible to shift the software phase to the desired value by
• Sync Ratio: it is the scale factor between the HV (applied to the EUT - Equipment Under
• Through the checkboxes it is possible to select the channels that are going to be used in the
tests;
• By clicking on the magnifying glass icon, a calibration that has been previously done can be
imported.
In the Additional Test Plan keys area additional information related to the Test Plan can be configured
for the operator to be filled when starting the plan. Such information can be defined as mandatory to
In the following section, Techimp PD Pro acquisition module and all its functionalities are described.
In order to start a new acquisition, it is necessary to click on the Acquisition icon in the Techimp PD
Acquisition
Module
An acquisition wizard, composed of two steps, will appear. In Figure 3-25 the first step is represented
and, in it, the operator can choose an existing registry in which the new acquisitions are saved or he
can create a new one (see section 3.8 for more information regarding the registry features).
Detail of the
selected
Registry
Create a new
Registry
Skip future
Query an acquisition
existent Registry association
with the next step. At this step (Figure 3-26) the operator can select:
1. The acquisition mode, choosing between Base mode, Test Plan, Advanced.
2. The instrument, choosing between PDCheck, PDScope, Aquila, Micro Aquila, and
PDBase.
PD Instruments
Available
Acquisition
modalities
In the following paragraphs, the three acquisition mode will be presented and explained.
The Base mode is a simplified acquisition modality. In this working condition the software, based on
the ongoing PD activity, automatically chooses the better acquisition parameters as the Full Scale
Connect
Channel
Acquire
• Insert the instrument IP address and the Port number of the instrument.
• Click on the Connect icon (on the upper-left side of the window).
• Set the acquisition information by means of the wizard (Figure 3-28). In this window it is
possible to choose the phase under test, the detection point and the sensors utilized. Click
To simply perform tests and avoid these associations in the window in Figure 3-25 the “Skip
• Phase shift: if the synchronization signal in use has a phase that is different from the voltage
applied to the insulating system, it is possible to shift the phase to the desired value by means
of this command.
Phase ref
Detection point
in use PD sensor in use
PD Test type
• In the window in Figure 3-27, in the bottom-left side, click on the Play button to start the
automatic acquisition process. The software automatically defines the optimal values of Full
Scale and Trigger and will complete the acquisition. If more than one acquisition needs to be
performed, an operator can easily click the Play button again and a new acquisition will be
done.
In Figure 3-29 an example of multiple acquisitions done in Base mode is reported, as can be seen,
all acquisitions are organized in a table together with some practical information. An acquisition can
be selected and then, using the icons above the table (on the right side), it is possible to add a
comment, delete the acquisition, change the association between the acquisition and the detection
In order to terminate the acquisition session, the Finish button (on the bottom-right side of the
window) can be clicked and all the acquisition, linked to the selected registry, are saved.
The Reset button (on the bottom-right side of the window) can be used to restart a new acquisition
session.
Change Set
Acquisition data
Delete
Start the
Acquisition Terminate the
acquisition session
The Advanced mode is the modality especially designed for the PD skilled operator. In this
acquisition modality, all parameters can be manually modified and more powerful features, as the T-
In Figure 3-30 the advanced acquisition window is shown, that interface layout is 3-phases oriented
In order to establish a connection to the instrument the IP address and the port number must be
inserted in the corresponding fields and then the Connect button has to be clicked.
Connect Port number
IP address
Once the instrument has been reached and the connection has been established, the interface is
Start streaming
Channel 1
Channel 2
Channel 3
Connection
Status
Connect/
Disconnect Waveform
Graph
Three Channel
Streaming Part
Acquisition
Settings
Acquisition
Log
In the highest part of the window in Figure 3-32 a status bar (Figure 3-33) is available, where there
are:
• The icons dedicated to the Instrument connection, to Start/ Stop the streaming and to perform
Simultaneous acquisitions (Figure 3-33). In particular, by clicking the latter icon that is enable
• The fields for the definition of the IP address and port number are needed for connecting the
• The indication of the sync frequency (in Hz) and the amplitude of the sync signal (in Volt).
• The indication of the enable/disable of the monitoring session, the status of the synchronizing
On the right side, of the advanced acquisition window (Figure 3-32), and starting from the above
• The waveforms graph that represents the simultaneous streaming of the signals coming into
the three PD channels (Figure 3-34). Exploiting the channel selector (“All”, “1”, “2”, or “3”) an
With the selector below the waveform graph, the desired value of the pretrigger can be
chosen.
Channel
selector
Pretrigger
selector
• The part dedicated to the acquisition parameters (Figure 3-35) where a skilled operator can
fix the settings according to the acquisition that he would like to perform. To clarify the
Table 1
Parameter Description
The pattern and the T-F Map will be reset when the number of
Buffer dimension
collected pulses reaches the buffer dimension.
• The acquisition log (Figure 3-36), where all the acquisitions that have been done are
organized in a table. Each row of the table and thus each acquisition, can be selected and
a comment can be added, they can be deleted or sent to the processing module.
In the list above the main tools of this area are described:
• Full scale: is the maximum value (positive and negative) displayed in the graphs (PRPD
Pattern and waveform graph), it is possible to change its value by writing the desired one in
• Trigger threshold: is the minimum amplitude that enables a waveform acquisition. Pulses
that do not reach the trigger value will be neither displayed nor acquired. It is possible to
change the trigger value by writing the desired value in the trigger text box or by moving the
• Phase shift: if the synchronization signal in use has a phase that is different from the voltage
applied to the insulating system, it is possible to shift the phase to the desired value by
the phase of the synchronization signal (sinusoidal waveform). When the amount of data
reaches the maximum pulse number the software resets the pattern automatically.
• T-F Map: the points of the graph are related to the equivalent time length and the equivalent
frequency of the detected pulses. Each point in the Pattern corresponds to a point in the T-
F Map. Refer to ANNEX 5 for additional information about Time – Frequency mapping.
• The Circuit type in use: due to the fact that the PD identification algorithm uses also the pulse
polarity for an enhanced automatic diagnosis, it is necessary to set the right kind of
measurement circuit used for the selected channel. It can be Direct (e.g. a high-frequency
current transformer wrapped around the ground connection of the EUT, Equipment Under
• APTWTi acquisition: it generates a dataset in which the amplitudes and phases, the
equivalent time lengths and equivalent frequencies, and the time intervals are contained. To
• Figure 3-38 (where the sinusoid and the floppy disk are represented).
• Waveform acquisition: it generates a dataset with the same quantities of the APTWTi
acquisition plus the waveform per each acquired pulse. To perform this acquisition click the
second icon of
• Figure 3-38 (where a waveform and the floppy disk are represented).
• Link Set Acquisition Data: it recalls the window in Figure 3-28 where an operator can define
the association between the channel, the phase under test and the detection point. In this
way, the software can correctly correlate the acquisitions and the relevant phase and
detection points. If dataset are linked, the queries fo the acquisitions are easier and smarter
• IEC60270 acquisition: the software enables the IEC60270 compliance. Since the peaks
Figure 3-38
• Reset buffer: reset the pulses present in the PD Pattern and T-F Map
• Show T-F Map Filter: recall the T-F Map filter window (see Par. 3.10.2.2)
• Disable T-F Map filters: it allows to disable the T-F Map filters without deleting them
Once the IEC60270 icon is clicked, the Calibration wizard, in Figure 3-39 and Figure 3-40, is
displayed. This procedure allows the operator to set the calibration pulse in order to obtain the PD
Calibration
file name
Trigger
Full scale selector
Streaming
Waveform
Injected pC
After having connected the calibrator to the EUT, insert the correct value in pC injected in the Injected
pC box. In order to save the calibration pulse select the related icon and click on the check icon. The
calibration will be saved with the name written in File Name box. If the Save calibration button is not
selected, the calibration will be applied only to the actual measurement session but not saved: the
It is also possible to load a previously saved calibration by clicking on the Load icon in the first step
of the wizard.In the second step of the Calibration wizard, the following warning could be exposed
This advises the operator that the saving feature is not allowed because the association between the
acquisition channel, the phase under test and the detection point has not been done yet. Hence, to
save the Calibration is required to the operator to previously make these associations by clicking on
the Link Set Acquisition Data icon (
Figure 3-38). In this way, the software can correctly correlate a Calibration with a phase of the
The T-F Map filter (see Figure 3-42) allows filtering signals through the T-F Map. In order to open
the window: right-click on the T-F Map >>> Show T-F Map filters.
Drop-down
menu
Delete filters
Apply filters
By clicking and dragging the mouse on the Map it is possible to select the areas in the T-F Map
which have to be filtered. A letter indicating each selected area will appear in the drop-down menu
and the content of the list as default will be automatically discarded in the acquisition pulses. If more
filters have a common area overlapped, the total area will be processed (discarded) as a union of
With a double click on its area, a filter can be selected then it can be:
• Deleted: by clicking on the Canc button on the keyboard or clicking on the dedicated icon.
Between the acquisition window for the PDCheck, PDScope, Micro Aquila and Aquila and the ones
for the PDBase there are some differences explained in the following section. In Figure 3-43 the
advanced acquisition window for the PDBase is shown. As can be noticed the interface is very similar
to the ones presented in section Error! Reference source not found.. The main difference is the p
resence of the Channel selector (see Figure 3-43) through which an operator can choose on which
channel activate the streaming. In fact, with a PDBase, only one single channel can be active at a
time, the others can be maintained open but they will not have the streaming active.
Sync management
Channel selector
Channel selector
Channel selector
Another difference is the Sync management window. By clicking on Sync in the upper-right side of
the window in Figure 3-43 the Sync management interface in Figure 3-44 is presented. In the
Synchronization window, it is possible to visualize the waveforms of all the synchronization signals
Full Scale
Time length
Filter bandwidth
Applied voltage
It is possible to select each of the six different channels in the top of window in order to visualize the
• Time Length: it sets the waveform length of the synchronization signal. It can be changed using
• Filter Band [Hz]: it allows to filter, via software, the synchronization signal. It is a low pass filter,
and the value of the cut frequency can be selected by the drop-down menu;
• Applied Voltage: it is possible to set the voltage applied to the EUT. The software automatically
set the ratio in accordance with the value read in the Sync Amplitude.
3.10.4 TEST PLAN MODE
The Test Plan mode is a Techimp PD Pro acquisition modality focused on quality tests or, in general,
on the operators that have to repeat the same test on the same asset typology several times. In this
working condition the software, based on the ongoing PD activity, automatically chooses the better
acquisition parameters as the Full Scale and the Trigger threshold and autonomously performs the
acquisition.
A measurement session is composed of different steps and of the related warning thresholds that
are completely customizable through the Techimp PD Pro Test Plan Editor module (see section
3.9). The operator has to simply check, during each step of the test, that the PD activity is below
such values.
Connect
Phase Shift
Step number
• Click on the Connect icon (on the upper-left side of the window).
• Select the channel number to utilize (upper-left side of the window, below the Connect icon).
• Set the acquisition information by means of the appeared wizard (Figure 3-28). In this window
it is possible to choose the phase under test, the detection point and the sensors utilized.
To simply perform tests and avoid these associations, in the window in Figure 3-25 the “Skip
• Set the Phase Shift: if the synchronization signal in use has a phase that is different from the
voltage applied to the insulating system, it is possible to shift the phase to the desired value
Phase analyzed
Detection point
in use PD sensor
in use
Sync
sensor
in use
PD Test type
Apply
• To start the automatic acquisition click on the Play button on the bottom-left side of the
window in Figure 3-45. The software automatically defines the better values of Full Scale and
Trigger and will complete the acquisition. Per each step of the measurement procedure, if
more than one acquisition wants to be done, an operator can easily press the play button
continue in the test and goes in the next step by clicking the Next icon on the bottom-left side
of the window.
• Per each step, as shown in Figure 3-47, the software asks the operator the result that can
be positive, doubt or negative in relation to the warning threshold of the step (for simplicity
the thresholds, per each step, are shown in the upper-right side of Figure 3-45).
• Figure 3-48 shows as the window appear when all steps have been done and the operator
Finish Reset
button button
Steps
results
• Finally, an operator can or click on the Reset icon in the bottom-right side of the window and
so starting a new identical measurement session or he can simply close the window. In both
cases, the software will ask the operator if a report has to be saved (Figure 3-49).
In the central part of the window (Figure 3-48) the acquisitions performed in the test are listed in the
Acquisition log. An operator can select one or more acquisition in order to:
(Figure 3-50).
Set
acquisition
data
Delete
Open the
Add a Processing
Acquisition comment module
log
3.10.4.1 IEC60270
In the Test Plan Editor (Par. 3.9), during the definition of the Test Plan characteristics, it can be
expressed the requirement of performing the tests according to the IEC 60270. Since the peaks
Before starting the measurement process the software will display the Calibration wizard (Figure
3-52). This procedure allows the operator to set the calibration pulse in order to obtain the PD
amplitude in pC values.
After having connected the calibrator to the EUT, insert the correct value of the pC injected in the
injected pC box. Change the value of Full Scale to avoid a saturated pulse displayed in the Waveform
graph and select a Trigger threshold higher enough to trigger just on the injected pulse and not on
the noise.
Load
calibration
Figure 3-51
Calibration
file name
Full scale
Trigger
selector
Streaming
Waveform
Injected pC
To save the calibration pulse select the related icon and then click on the Check icon. The calibration
If the Save calibration button is not selected, hence the operator has selected simply the Check icon,
the calibration will be applied only to the actual measurement session but not saved: the calibration
It is also possible to load a previously saved calibration by clicking on the Load icon in the first step
of the wizard.
During the definition of the IEC 60270 Test Plan, in the Test Plan Editor (Par. 3.9), a Calibration can
be loaded and, in this way, an operator will not be required to perform any calibration before the test.
PROCESSING
The Techimp PD Pro processing module has the purpose of elaborating PD data exploiting several
In Figure 3-53 the main page of the processing module is reported. This window is composed of up
to six rows (being the representation of the instrument channels) in which the acquisitions can be
• The first part, on the left, in which all the acquisitions opened are listed. The acquisitions
saved with all the waveforms of the acquired pulses are highlighted by the green font.
• The second part, in the center of the window, in which the PRPD Pattern and the T-F Map
are shown together with information as the Full Scale, Trigger and phase shift. The previous
and next arrows, allow passing the data files in visualization one by one.
• The third part, on the right, in which the principal information related to the selected
acquisition is reported.
Add Advanced
acquisitions Processing
Acquisition
detail
Load new
acquisitions
Acquisition
selectors
In the window (Figure 3-53) on the upper-left side of each channel three icons are present:
• Plus: other acquisitions can be added to those already present by clicking on this icon;
• Load New Acquisition: by clicking this icon the previously listed acquisitions will be replaced
• Advanced Processing: by clicking on the mechanical gear icon the sub-module dedicated
In the Processing main page, the operator can change the Full Scale value by writing the value in
its box or by clicking on the arrows buttons on the right as well as change the Phase Shift value by
writing the desired value or using the arrow buttons. These parameter changes are not permanent
(i.e. the PD file is not modified), but only temporarily applied to the PD Pattern graph in order to have
With a right-click on the Pattern or on the T-F Map, the exposed data can be copied or exported, the
first command allows the operator to make a print screen, while the latter one will save the data in a
.csv file.
3.11.1 ADVANCED PROCESSING: MAIN PAGE
The Techimp PD Pro advanced processing (Figure 3-54) allows performing a deeper analysis of
PD stored data.
Main
page icon
PD tools
Waveform
slide bar Pointer cursor
The advanced processing window is composed of multiple PD tools, i.e. Separation, Pattern Filter,
Identification, TDR location, Pattern Movie, Trending and Report. Each tool has a dedicated icon
that can be activated and visualized in the Advanced window by clicking the relevant icon on the
On the opposite side to the PD tools, the Class list allows us to select and display on the PD Pattern
one of the Classes defined through the Separation tool (see 3.11.2 for more details about the logical
operation of the class definition). When no PD tool is activated the Advanced window shows the PD
Pattern and the T-F Map, moreover below of it, but only for waveform files, the waveform and pulse
spectrum graphs.
Moving the waveform slide bar in the central part of the window it is possible to plot all the waveforms
The pointer cursors enabled shows the Amplitude and the Phase values of the selected pulse/dot
on the PD Pattern. If the pointer cursor is activated on T-F Map too each movement of one cursor
will move the other one accordingly over the corresponding dot on its Pattern graph.
The Full Scale and Phase Shift controls, in Figure 3-54, allow changing the Full Scale and phase
shift values, in order to get a better visualization of the pattern. It is possible to change these values
either by clicking the up and down arrows or writing the value in the relevant box. It is important to
set the proper Full Scale and Phase Shift because the automatic identification algorithm relies on
this information.
On the Pulse Spectrum and Waveform graphs, zooming can be performed by clicking with the left
mouse button on the pulse spectrum graph and dragging the cursor (or using the mouse scroll wheel)
in order to have a better visualization. To reset the visualization, right-click the mouse button and
Thanks to the Separation tool it is possible to select different clusters in the T-F Map in order to
separate different PD phenomena and analyze generated sub-patterns. To activate the Separation
tool select the Separation icon in the upper part of the window as shown in Figure 3-55.
Class
selector
Separation Area
icon list
Separation
Map
Separation
result
Sub Patterns
In this window, an operator can see the Separation Map, the sub-patterns in the lower part, the class
list on the upside and the area list on the central part.
To perform the separation drag the mouse cursor on the Separation Map graph to define a
rectangular area that includes the desired portion of the Map. After the mouse button has released
a class is created and the software is ready for the definition of a new class. If an area needs to be
added in an existing Class it is sufficient to select the class with the Class selector, click on the
Separation Map graph and move the cursor to create a rectangle area on the portion of the
To apply the separation click the check icon in the center part of the window. After it has applied the
cluster separation the generated sub-patterns appear in the lower part of the window.
Each class created by the Separation procedure is indicated by a letter and listed in the Area list.
A rectangular in the Separation Map can be selected by a double click on it (or selecting it in the
Area list), then it can be deleted by clicking on the trash bin icon (in the central part of the window),
it can be modified through the cursor when it is positioned onto its borders or moved when the cursor
is inside it. Once a rectangular is selected it can be deleted also clicking the Canc keyboard button.
Class #0 includes all the pulses which do not belong to any other class, i.e. the dots that have not
been included in any rectangular area. If a certain area of the Map is selected for different classes,
then the pulses located in this area belong to the last defined class.
When all the desired classes are defined, it is possible to go back to the main page by clicking on
the house icon on the upper side of the window. As indicated in Figure 3-56 the sub-pattern relevant
Class
selected
With the Pattern Filter tool, it is possible to perform manual filtering of PD Pattern. This is useful
when two or more phenomena are concentrated in the same region on the T-F Map and it is not
possible to perform a proper separation. To activate the Pattern filter tool click on the Pattern Filter
icon as shown in see Figure 3-57. This function is very important in order to get the correct
phenomenon identification.
Class
selector
Apply
filter
First of all, the right setting of the Acquisition circuit (in the upper part of this window) must be
checked. On top of this window, there is a PD Pattern with colored lines that allow defining amplitude
and phase ranges, for positive and negative pulses. It is possible to move those lines using the
Red lines define a region for positive pulses, while blue lines define a region for negative pulses. It
is noteworthy that the dots corresponding to positive pulses are plotted above the horizontal axis of
the PD Pattern if the Direct measurement circuit is selected. On the contrary, positive pulses are
plotted below the horizontal axis of the PD Pattern, if the Indirect measurement circuit is selected.
Once the filtering cursors have been set it is possible to view, on the low part of the window, a
preview PD Pattern, that shows only pulses inside regions defined through colored lines. To apply
If a separation has been done it is possible to perform a different pattern filtering for each class. To
do this select a class through the Class selector, then define the filtering for that class.
• Trigger: to show the Trigger level set during the file acquisition.
• Copy data: it allows to perform a print screen of the PD Pattern picture on your clipboard.
• Copy data: takes a snapshot of the T-F Map and saves it in the clipboard.
Pro software. As a default, this tool is enabled when the Processing module is opened and, by
opening the Filter page (Figure 3-57), it is possible to check the output of the automatic filtering
function. First of all the right setting of the Acquisition circuit (in the upper part of this window) must
be checked then the output of the automatic filter algorithm could be modified by moving the cursors
in the PD Pattern.
It is possible to restore automatic filtering results by disabling and enabling the auto filtering again
By clicking on the Identification icon as shown in Figure 3-58, the operator can get a final response
in terms of PD Identification.
Identification
icon
Drop-down
menu
The system takes advantages of an architecture that evaluates the behavior of the PD phenomena
1. Corona Discharges: the PD’s are recognized as discharges generated in open air or gas
2. Surface Discharges: the PD’s are recognized as discharges that develop on the surface of
solid insulation materials, due to the cut component of the electric field to the surface;
3. Internal Discharges: the PD’s are recognized as discharges occurring in an air gap
completely surrounded by a solid dielectric, otherwise between the solid dielectric and a
metallic electrode; these discharges are generated by the normal component of the electric
do any inference because of a large amount of unclassifiable data acquired; in this case, try
to filter again the pattern and retry to run the identification procedure;
On the lower-right side of the Identification window, a table of Diagnostic Indexes is given.
The red and the blue columns display respectively the parameter values for positive and negative
Table 2
Parameter Description
By expanding the drop-down menu showed in Figure 3-58 an operator can select (if the bought
license includes this feature) what kind of Equipment Under Test (EUT) the analyzed data are
referred to.
Through the TDR location tool (if the bought license includes this feature) it is possible to locate the
PD source along a cable circuit affected by PD activity using the Time Domain Reflectometry (TDR)
technique. The distance between the measuring point and the PD source is obtained from the time
delay among the PD pulse and its reflections generated at cable terminations.
To activate the TDR Location select the TDR Location icon as shown in Figure 3-60.
Third
First part
part
TDR location
icon
Second
Slide bar
part
Average
pulse
Fourth
part
• In the upper part of the window, the waveform of each acquired pulse is reported. Moving the
horizontal slide bar below the upper waveform graph it is possible to plot all the waveforms
• The second section (on the central left) shows the plot of the average pulse of the acquisition.
To use the average pulse for location purposes is useful because background noise
amplitude is reduced.
• If the average pulse waveform is still affected by background noise it is possible to modify
low and high cut-off frequency through the Waveform Filtering (third section on the right part
of Figure 3-60).
• The fourth section (on the bottom side) shows the location of PD source with respect to the
measuring point after the cursor positions and the cable data have been properly provided,
PD pulse (t1), the red one indicates the first reflection (t2), and the grey represents the second
reflection (t3). Using the mouse scroll wheel a zoom of the waveforms can be made to easily adjust
NOTE: To adequately perform the location the cursors must be carefully positioned:
the blue one on the PD pulse, the red one on the first reflection and the green one
place the cursors on the rising point of the pulse not on the peak point as shown in
Figure 3-61.
Figure 3-61 Advanced processing: TDR location with proper positioning of the cursors
Once the three cursors have been placed, the user can proceed to the PD source location. In the
left part of the fourth section, the propagation speed of the pulse and total cable length are indicated
(input by the user) while on the right there is the difference between t3 and t1 and between t2 and
t1. These parameters, as well as L1 and L2, are calculated automatically according to propagation
Pattern movie tool allows visualizing the real-time sequence of detected pulses that occurred during
the measurement. To activate the Pattern movie tool select the Pattern Movie icon as shown in
Figure 3-62.
Identification
icon
Black
cursor
Progress
bar
Speed cursor
regulator
On the PD Pattern, it is possible to follow the movement of a black cursor that constantly spots the
position of the last pulse that occurred during the acquisition time sequence, like a movie. A vertical
cursor is reported on the right side of the pattern by which the refresh speed of the pattern may be
changed.
On the bottom of the PD Pattern there is a progress bar that indicates the movie progress.
With a right-click on the PD Pattern an operator can choose:
• Trigger: to show the Trigger level set during the file acquisition.
• Copy data: it allows to perform a print screen of the PD Pattern picture on your clipboard.
• Copy data: takes a snapshot of the T-F Map and saves it in the clipboard.
Thanks to the trending tool it is possible to plot the evolution of PD phenomena over time.
To activate the Trending tool select the Trending icon as shown in Figure 3-63.
Trending
icon
Selectable
quantities
acquired file, that we would like to add in the trend, must be opened in the same row of the
It is possible to choose what variables to plot by selecting the desired ones in the bottom part of the
• Nw: the repetition rate as pulses per period of the applied voltage;
when we look after the trend of a specific class the program looks only after the
trend of this specific region in the T-F Map. This logic of calculation may lead to
For each point of the trend, it is possible to visualize the relevant PD Pattern and T-F Map by clicking
on it. Using the mouse scroll wheel a zoom of the graph can be made.
• Trigger: to show the Trigger level set during the file acquisition.
• Copy data: it allows to perform a print screen of the PD Pattern picture on your clipboard.
• Copy data: takes a snapshot of the T-F Map and saves it in the clipboard.
• Copy data: takes a snapshot of the Trend graph and saves it in the clipboard.
Thanks to the Report tool it is possible to automatically generate a summary report in which all the
analyses results are inserted. An operator can click on the Report icon to lunch the generation of the
Report forwarding
through mail
are reported while, in the following parts, the entire PD Pattern and T-F Map together with the results
NOTE: all patterns represented in the following pictures are acquired in indirect polarity.
INTERNAL PD
Spherical void in epoxy (low overvoltage) Spherical void in epoxy (low overvoltage)
Spherical void in epoxy (low overvoltage) Spherical void in epoxy (large overvoltage)
Spherical void in epoxy, sharp electrodes Internal PD with strongly divergent field
machine with taped grading system rotating machine with painted grading system
Needle on high voltage electrode (low Needle on high voltage electrode (low
overvoltage) overvoltage)
Needle on high voltage electrode (large Needle on high voltage electrode (large
overvoltage) overvoltage)
Needle on high voltage electrode (large Needle on high voltage electrode (large
overvoltage) overvoltage)
Needle on ground voltage electrode (low Needle on ground voltage electrode (low
overvoltage) overvoltage)
This measurement approach allows the signals detected, sampled and acquired to be gathered into
homogeneous clusters based on the shapes of the signals. Signals coming from the same sources
are characterized by similar shape parameters. Such shape parameters are the equivalent time
length (T) and the equivalent frequency (F) and they are calculated according to the following
L
(1)
s (t ) = s(t ) s( ) d
2
L
(3)
T= (t − t ) s (t )
2 2
0 dt
0
2
(4)
F= f
2
S( f ) d f
0
The decomposed PD signals are then plotted in an XY graph, the T-F map. By means of this graph,
it is possible to separate the different phenomena and relate them to the sources that can be noise,
clusters, each one belonging to a different signal source relevant to the color of the cluster. The
resulting PRPD sub-patterns are built according to the separation carried out in the T-F map and are
relevant to different phenomena. Such separation (that can be automatic or manual) allows an easier
Figure 5-1: an example of separation of signals by means of the T-F map clusters.
Moreover, the T-F map separation allows tracking the time behavior of each single phenomenon,
avoiding the raising of false alarms caused by the superposition of contributions from different
sources.
Notes
…………………………………………………………………………………………………………………
…………………………………………………………………………………………………………………
…………………………………………………………………………………………………………………
…………………………………………………………………………………………………………………
…………………………………………………………………………………………………………………
…………………………………………………………………………………………………………………
…………………………………………………………………………………………………………………
…………………………………………………………………………………………………………………
…………………………………………………………………………………………………………………
…………………………………………………………………………………………………………………
…………………………………………………………………………………………………………………
…………………………………………………………………………………………………………………
…………………………………………………………………………………………………………………
…………………………………………………………………………………………………………………
…………………………………………………………………………………………………………………
…………………………………………………………………………………………………………………
…………………………………………………………………………………………………………………
…………………………………………………………………………………………………………………
…………………………………………………………………………………………………………………
…………………………………………………………………………………………………………………
…………………………………………………………………………………………………………………
…………………………………………………………………………………………………………………
…………………………………………………………………………………………………………………
…………………………………………………………………………………………………………………
…………………………………………………………………………………………………………………
…………………………………………………………………………………………………………………
…………………………………………………………………………………………………………………
…………………………………………………………………………………………………………………
…………………………………………………………………………………………………………………
…………………………………………………………………………………………………………………
…………………………………………………………………………………………………………………
…………………………………………………………………………………………………………………
…………………………………………………………………………………………………………………
…………………………………………………………………………………………………………………
…………………………………………………………………………………………………………………
…………………………………………………………………………………………………………………
…………………………………………………………………………………………………………………
…………………………………………………………………………………………………………………
…………………………………………………………………………………………………………………
…………………………………………………………………………………………………………………
…………………………………………………………………………………………………………………
…………………………………………………………………………………………………………………
…………………………………………………………………………………………………………………
…………………………………………………………………………………………………………………
…………………………………………………………………………………………………………………
…………………………………………………………………………………………………………………
…………………………………………………………………………………………………………………
…………………………………………………………………………………………………………………
…………………………………………………………………………………………………………………
…………………………………………………………………………………………………………………
…………………………………………………………………………………………………………………
…………………………………………………………………………………………………………………
…………………………………………………………………………………………………………………
measurement automation development energy innovation measurement automation development energy innovation measurement automation development
info@techimp.com
www.techimp.com
Via Toscana, 11/C
Tel.: +39.051.19986050
Fax: +39.051.19986099
Techimp – Altanova Group S.r.l.