Professional Documents
Culture Documents
02 DATE: 17-11-01
TITLE/TAG FUNCTION CHECK BTEX ANALYZER
EQUIPMENT TYPE BTEX ANALYZER
INTERVAL 1 MONTHLY
PRECAUTIONS
Note: Tick boxes to indicate completion.
1. Obtain an approved Work Permit before commencing activities.
PREPARATION/ISOLATION
2. Check that the equipment tags are as detailed on work order sheet/work permit.
CAUTION: ONLY QUALIFIED PERSONS ARE AUTHORISED TO PERFORM THIS
PROCEDURE. REFER TO WARNINGS AND PRECAUTIONS IN VENDOR
MANUAL.
MAINTENANCE ACTIVITY
3. Check the unit for any signs of damage.
4. Check the status of the analyzer control panel.
5. Check the condition of emitter, receiver & optical-fibre.
6. Check and adjust as necessary reference calibration of the analyzer.
7. Check hardware requirements needed for reference calibration of the analyzer.
8. Check and adjust as necessary analyzer’s if zero response deviates more than 0.025ppm from zero.
Perform reference calibration (Refer to section 5.2.2 of vendor’s manual)
9. Back up data from the hard disk to Compact Disk.
10. Keep maintenance records of the analyzer at the monitoring station.
11. Sign off the work permit.
REPAIR WORK
12. Is repair work or corrective action required? YES NO
13. Has a Work Order been raised? YES NO
REFERENCE INFORMATION
Opto-Analyzer for Air Quality Monitoring Hardware Operating Manual
Appendix D of Document Number 5-A-AA-EC-E-0001