Professional Documents
Culture Documents
Unit Conversion Protocol of A2000Plus Client (Version1.1.2)
Unit Conversion Protocol of A2000Plus Client (Version1.1.2)
client (Version1.1.2)
Take the FT4 as a
instance
2019-7-10
1. Unit conversion function(1/9)
Under the maintainer level, entry test item. Find ft4 and change “Unit
conversion factor A"
2
1. Unit conversion function(2/9)
After changing “Unit conversion factor A“, there will be a messages to prompt modify the second sets parameter ,Click
“OK” first, and the second sets parameter will be changed in 7-9 steps.
3
1. Unit conversion function(3/9)
After changing “Unit conversion factor A“, then change “User supplier"
4
1. Unit conversion function(4/9)
Modify User supplier parameter
5
1. Unit conversion function(5/9)
After changing User supplier, there will be a same prompt. Click OK.
6
1. Unit conversion function(6/9)
click Next to switch the second sets parameter.
7
1. Unit conversion function(7/9)
After switch the second sets, the details shown below:
8
1. Unit conversion function(8/9)
Modify the Unit conversion factor A and User supplier like step1-5.
9
1. Unit conversion function(9/9)
After changing Unit conversion factor A and User supplier the details shown below:
10
2. Some parameters of unit conversion function(1/2)
Under the maintainer level, entry Parameter setting, and set the value of Enable Print User
Result as 0.
11
2. Some parameters of unit conversion function(2/2)
In the result interface, click “Customarizion”, select the User Result and marking it to make
the user result shown in the software.
12
www.autobio.com.cn