Professional Documents
Culture Documents
net/publication/350053930
0.2T - 0.23T Open MRI systems, Issue 5. Comparative report of four MRI
systems
CITATIONS READS
0 1,999
3 authors, including:
All content following this page was uploaded by Ioannis Delakis on 14 March 2021.
December 2005
Our evaluations are usually of products supplied by the manufacturer. We expect these
products to be representative of those on the market but cannot guarantee this.
Prospective purchasers should satisfy themselves about any modifications that might have
been made after our evaluation.
The Centre for Evidence-based Purchasing (formerly the Device Evaluation Service)
transferred from the Medicines and Healthcare products Regulatory Agency to NHS PASA
on 1 September 2005. We are currently undergoing extensive redesign to help us provide
the information that purchasers want in the way they want it presented. Please visit our
website to keep updated.
E-mail: cep@pasa.nhs.uk
All evaluation reports published since 2002 are available in full colour to download from
our website: www.pasa.nhs.uk/cep
MagNET
Bagrit Centre
Department of Bioengineering
Imperial College London
Exhibition Road
London SW7 2AZ
Apart from any fair dealing for the purposes of research or private study, or criticism, or review,
as permitted under the Copyright, Designs & Patents Act, 1998, this publication may only be
reproduced, stored, or transmitted in any form or by any means with the prior permission, in
writing, of the Controller of Her Majesty’s Stationery Office (HMSO).
Information on reproduction outside these terms can be found on the HMSO website
(www.hmso.gov.uk) or e-mail: hmsolicensing@cabinet-office.x.gsi.gov.uk
Contents
Summary ....................................................................... 1
Purpose of this report ............................................... 1
Comparative specifications....................................... 1
Technical evaluation................................................. 1
Introduction................................................................... 2
Structure................................................................... 2
Evaluated systems ................................................... 3
Comparative specification ........................................... 4
Magnet system ......................................................... 4
RF system ................................................................ 7
Gradient system ....................................................... 8
Sequence information............................................... 9
Computer system ................................................... 11
Radio-frequency (RF) coils..................................... 13
Technical evaluation................................................... 18
Details of assessment ............................................ 18
Quadrature head coil evaluation............................. 19
Body coil evaluation................................................ 24
Spatial resolution .................................................... 27
Geometric tests ...................................................... 29
Slice profile and slice width .................................... 34
Imaging speed measurements ............................... 36
Acoustic noise ........................................................ 39
Acknowledgements .................................................... 41
References .................................................................. 42
Appendix ..................................................................... 43
Manufacturers’ comments ...................................... 43
2D imaging speed................................................... 44
3D imaging speed................................................... 48
Summary
Comparative specifications
The comparative specification is presented as a side-by-side summary
comparison of the specifications of each scanner and related equipment. It is
grouped into a series of sub-sections relating to different aspects of the
scanner, such as magnet, gradients, coils etc. Specification data were supplied
by the manufacturers. The data have not been verified by CEP.
Technical evaluation
The technical evaluation section presents a detailed analysis of the head and
body coils available from each manufacturer. Parameters such as signal-to-
noise ratio, uniformity, resolution are included in the evaluation. The data are
published after consultation with manufacturers. Their comments are included in
the appendix.
Report 05097: 0.2T-0.23T open MRI systems - Issue 5
1
Introduction
The reader is also referred to the following NHS document for the built-
environment requirements of an open MR suite: Extremity and open MRI,
magnetic shielding and construction for radiation protection (HBN 6-volume 3.
This volume is available via the Department of health website:
http://www.dh.gov.uk/PublicationsAndStatistics/Publications/fs/en).
This report does not attempt to explain MRI. Readers who are unfamiliar with
this modality may have difficulty understanding the results presented in this
report. In this case, advice should be sought from a suitably qualified MRI
specialist.
Structure
The main body of the report is divided into two sections, the first presents
Report 05097: 0.2T-0.23T open MRI systems - Issue 5
2
Evaluated systems
This report presents system specifications and technical evaluation data for the
MRI systems listed below and shown in Figure 1.
3
Comparative specification
The specification data were supplied by the manufacturers and have not been
verified by CEP.
Magnet system
Table 2: Magnet specification
GE Hitachi Philips Siemens
Parameter
Profile Airis Mate Panorama Concerto
RF frequency MHz 8.54 8.5 9.8 8.25
4
User information
Vertical (z) 0.5 mT fringe field m 2.0 1.9 Top 2.5 x 1.9
bottom 2.5
Minimum installed area m2* 23 16 17 30
Minimum ceiling height cm* 230 240 234 225
To include 0.5 mT fringe field
5
User information
6
User information
RF system
Table 7. RF system specification
GE Hitachi Philips Siemens
Profile Airis Mate Panorama Concerto
Name/type/ Data not Data not Synergy Data not
version of the system supplied by supplied by platform supplied by
manufacturer manufacturer manufacturer
Number of Data not Data not 4/4 2
independent RF supplied by supplied by
receiver channels manufacturer manufacturer
(standard/
optional)
Bandwidth of each Data not Data not 330 kHz 0.5
independent RF supplied by supplied by
receiver channel manufacturer manufacturer
(MHz)
Number of Analog-to- Data not Data not 1 x 18bits 1 (reads out
Digital Converters for supplied by supplied by amplitude and
each independent RF manufacturer manufacturer phase)
channel
Sampling frequency of Data not Data not 40 MHz 10
each Analog-to-Digital supplied by supplied by
Converter (MHz) manufacturer manufacturer
Report 05097: 0.2T-0.23T open MRI systems - Issue 5
7
User information
Gradient system
Table 8: Gradient system specification
GE Hitachi Philips Siemens
Profile Airis Mate Panorama Concerto
Shielding Self shielded Passive Active Active
Single axis maximum amplitude
x mT/m 15 15 18 20
y mT/m 15 15 18 20
z mT/m 15 15 21 20
Single axis slew rate
x mT/m/ms 30 30 30 40
y mT/m/ms 30 30 30 40
z mT/m/ms 30 30 50 40
Duty cycle at max amplitude 100% 50% 50% 100%
Amplitude @100% duty cycle 15 10 8 (x,y) 20
mT/m 9 (z)
Report 05097: 0.2T-0.23T open MRI systems - Issue 5
8
User information
Sequence information
Quoted values correspond to imaging with no interpolation, no asymmetric
echo, no partial Fourier, no parallel imaging, no fat-sat, no rectangular matrix,
no rectangular field-of-view.
Minimum FOV mm 30 80 25 5
Maximum imaging matrix 512×512 1024 x 512 1024×1024 512×512
Minimum 2D slice thickness mm 1.4 2.0 0.4 1.7
Minimum 3D slice thickness mm 0.2 0.5 0.1 0.05
9
User information
Segmented and
2D/3D phase contrast
angiography
2D timing protocols
Diffusion Data not DWI DWI SE Advance turbo
supplied by Turbo DWI SE package: HASTE
manufacturer
Orthopaedic Data not Data not Comprehensive ortho Comprehensive ortho
supplied by supplied by package package in standard
manufacturer manufacturer configuration
Cardiac Data not Data not Cardiac Imaging Cardiac package:
supplied by supplied by Package special sequences
manufacturer manufacturer and scan protocols for
MR studies of the
heart
3D imaging Data not Data not Standard: 3D SE, 3D 3D DESS (for
supplied by supplied by TSE, 3D FFE, 3D B- orthopaedic imaging)
manufacturer manufacturer FFE ,3D-DFFE1st,3D-
DFFE 2nd,MRA 3D 3D CISS (for very high
FFE resolution studies eg:
inner-ear
examinations)
Report 05097: 0.2T-0.23T open MRI systems - Issue 5
10
User information
Computer system
Table 12: Main computer system - architecture
GE Hitachi Philips Siemens
Profile Airis Mate Panorama Concerto
Type RISC SUN 64bit RISC Hewlett Dual
Ultra SPARC 2 CPU Packard Pentium 4TM/
XW8200 Intel XeonTM
Operating system UNIX UNIX Windows XP Windows XP
CPU speed MHz 167x2, 400, 440 Dual CPU 3.6 2x 3600
440 GHz
Word length bit 64 64 32 32
Memory size MB 256 512 3096 2048
512 1024
Hard disk 9 34.6 73 36
Software GB (+36 for
database)
Hard disk Included above Included above Same HD 76
Images GB
Image capacity 30000 27000 750 000 ~ 190000
2562 images*
Archive drive MOD MOD 5.25” MOD CD-R
Size GB 2.3 2.6 9.1 650 MB
Image capacity 30000 17000 130000 ~ 4000
2562 images*
* Image capacity for uncompressed 2562 images
Report 05097: 0.2T-0.23T open MRI systems - Issue 5
11
User information
12
User information
Type QR LR PA R PA R
1 element
Dimensions cm 33×40×30.9 21 15.5 × 15.5 (me) 16 (me)
(h x w x l) (dia) 24 × 17.5 (la) 20 (la)
(h x w) (dia)
R=receive, T/R=transmit/receive, PA=phased array, Q=quadrature, L=linear, w=width, h=height, d=depth,
l=length, dia=diameter, circ=circumference, xs=extra small, sm=small, me=medium, la=large, xl=extra
large, xxl=extra extra large
13
User information
14
User information
Spine coils
Name CTL array CTL array Synergy As above
body/spine coil No extra coil
required
Type PA Q R PA R PA R -
Dimensions cm 57.1x86.3 x41.2 40x104 Data not supplied -
Report 05097: 0.2T-0.23T open MRI systems - Issue 5
(w x l x d) (w x l) by manufacturer
R=receive, T/R=transmit/receive, PA=phased array, Q=quadrature, L=linear, w=width, h=height, d=depth,
l=length, dia=diameter, circ=circumference, xs=extra small, sm=small, me=medium, la=large, xl=extra
large, xxl=extra extra large
15
User information
(w × l × d)
Name Wrist array No other coil No other coil No other coil
Type PA Q R - - -
Dimensions cm 10.4x19x17 - - -
(w x l x d)
R=receive, T/R=transmit/receive, PA=phased array, Q=quadrature, L=linear, w=width, h=height, d=depth,
l=length, dia=diameter, circ=circumference, xs=extra small, sm=small, me=medium, la=large, xl=extra
large, xxl=extra extra large
16
User information
Others
Name Neuro vascular Shoulder TMJ No other coil
array
Type PA Q R Q R/PA R LR -
Dimensions cm 37.8x51.7x33.8 22x22.4 Data not -
(w x l x d) (w x l) supplied by
manufacturer
R=receive, T/R=transmit/receive, PA=phased array, Q=quadrature, L=linear, w=width, h=height, d=depth,
l=length, dia=diameter, circ=circumference, xs=extra small, sm=small, me=medium, la=large, xl=extra
large, xxl=extra extra large
17
Technical evaluation
Details of assessment
Table 20 provides information on the open MRI systems included in the
technical evaluation. The data from the type-tested systems have been
published in previous evaluation reports unless otherwise stated. We ask that
all images be acquired without any pre-reconstruction or post-processing filters.
This is not possible for some manufacturers where pre-reconstruction filters are
part of the reconstruction process and cannot be switched off. In addition, some
manufacturers of low field systems recommend the use of post-processing
filters in clinical imaging to improve signal-to-noise ratio (SNR). These filters
however, may affect spatial resolution. Consequently, only unfiltered images are
analysed for SNR and resolution tests.
We aim to include new technical evaluation data for the GE, Hitachi and
Siemens systems in the next issue of this report.
Table 20: MR model abbreviation codes for evaluated open MRI systems
Manufacturer GE Hitachi Philips Siemens
Model Profile Airis Mate Panorama MAGNETOM
Concerto
Key GE-PRO HI-AIR PH-PAN SI-CON
Assessment date February 1998 November 2000 June 2005 July 2001
18
Technical evaluation
Table 21: Experimental conditions for quadrature head coil SNR test
System GE-PRO HI-AIR PH-PAN SI-CON
Bandwidth kHz ± 7.81 ± 15.00 ± 5.0 ± 10.24
Test object T01A-L T01A-L MAGFF-Loaded TO1A_S
Loading MAG-ANN None In test object In test object
Q phantom 328 183.3 54 166.78
Q head 386 139.5 60 167.75
Q factor 1.08 0.87 1.05 1.00
Analysis
The SNR is calculated from the signal mean of five regions of interest (ROI) in
one image and the standard deviation of five ROIs in the subtracted image. The
SNR values are normalised for voxel size, scan time, sampling bandwidth and
Q-factor.
Results
Results are presented in Table 22. Graph 1 presents a comparison of the
normalised SNR (NSNR) results.
Report 05097: 0.2T-0.23T open MRI systems - Issue 5
19
Technical evaluation
HI-AIR
SI-CON
GE-PRO*
PH-PAN
*GE systems apply a pre-reconstruction Fermi filter as standard. This could not be switched off during this
type-testing.
Report 05097: 0.2T-0.23T open MRI systems - Issue 5
20
Technical evaluation
Scan conditions
The head coil uniformity test is carried out using our standard type-test
protocols (as for the head coil SNR test). A flood field test object is placed at the
centre of the coil and is imaged at the iso-centre in all three imaging planes.
Analysis
Uniformity measurements are made in all three directions. The fractional
uniformity is calculated for each direction from the percentage of pixels that lie
within 10% of the mean value of a central ROI.
Results
The optimum value for uniformity is 1.00. The results for all three directions are
presented in Table 23 and Graph 2.
HI-AIR
Report 05097: 0.2T-0.23T open MRI systems - Issue 5
GE-PRO
PH-PAN
SI-CON
21
Technical evaluation
Analysis
Ghosting measurements are made from the four images acquired in each signal
average set. Ghosting is calculated as the ratio of the maximum ghost found in
the image (minus the background noise) to the signal x 100.
Results
The optimum value for ghosting is zero, hence the lower the ghosting, the better
the performance of the system. Table 24 shows the maximum ghosting for each
echo. Graph 3 and Graph 4 show comparisons of maximum ghosting for 1 and
2 NSA respectively.
NSA = 1
NSA = 2
Report 05097: 0.2T-0.23T open MRI systems - Issue 5
22
Technical evaluation
GE-PRO
HI-AIR
PH-PAN
SI-CON
GE-PRO
HI-AIR
PH-PAN
Report 05097: 0.2T-0.23T open MRI systems - Issue 5
SI-CON
23
Technical evaluation
Analysis
The SNR is calculated from the signal mean of five ROIs in one image and the
standard deviation of five ROIs in the subtracted image. The values obtained for
SNR are normalised for voxel size, scan time, sampling bandwidth and Q-
factor. Q measurements are presented in Table 25.
Report 05097: 0.2T-0.23T open MRI systems - Issue 5
Results
The SNR results for the body coils of the evaluated systems are shown in Table
26 and Graph 5.
24
Technical evaluation
GE-PRO*
PH-PAN
SI-CON
HI-AIR
0 20 40 60 80 100
Normalised SNR (NSNR)
*GE systems apply a pre-reconstruction Fermi filter as standard. This could not be switched off during this
type-testing.
Report 05097: 0.2T-0.23T open MRI systems - Issue 5
25
Technical evaluation
Analysis
Uniformity measurements are made in all three directions. The fractional
uniformity is calculated for each direction from the percentage of pixels that lie
within 10% of the mean value of a central ROI.
Results
The optimum value for uniformity is 1.00. The results are presented in Table 27.
Graph 6 gives a comparison of the measured fractional uniformity for open
systems.
GE-PRO
PH-PAN
SI-CON
HI-AIR
26
Technical evaluation
Spatial resolution
Scan conditions
The resolution tests are carried out using our standard type-test protocols (as
for the head coil SNR test). The resolution test object is placed at the centre of
the head coil. The test object contains a set of four angled Perspex MTF
(modulation transfer function) blocks. The image nominal pixel size of 0.98 mm
is defined by the FOV divided by the matrix size. The test object is scanned at
the iso-centre in the transverse, sagittal and coronal planes.
Analysis
The resolution is analysed using in-house resolution software which calculates
the 50% point on the MTF plot. This frequency is converted into pixel resolution.
Results
The optimum value for resolution is 0.98. Table 28 shows the results for the
mean spatial resolution in both the PE and FE directions for the open systems
tested. Graph 7 shows a comparison of the spatial resolution for the different
systems.
27
Technical evaluation
PH-PAN
SI-CON
HI-AIR
GE-PRO*
28
Technical evaluation
Geometric tests
Geometric linearity
Scan parameters
The geometric linearity tests are carried out using our standard type-test
protocols (as for the head coil SNR test, except FOV=256 mm). The geometry
test object is placed at the centre of the head coil and is imaged at the iso-
centre in transverse, sagittal and coronal planes.
Analysis
The linear distances are measured three times in each direction. These
measurements are converted from pixels to mm and compared to the actual
distance of 120 mm for linearity.
Results
Table 29 shows a comparison of the mean geometric linearity for the three
directions with different type-tested open systems. Graph 8 compares the mean
geometric linearity error.
HI-AIR
SI-CON
PH-PAN
GE-PRO
*The optimum value is 0 ± 1 mm. Systems are ranked in order of the absolute deviation from optimum.
29
Technical evaluation
Geometric distortion
Scan parameters
The geometric distortion tests are carried out using our standard type-test
protocols (as for the head coil SNR test, except FOV=256 mm). The geometry
test object is placed at the centre of the head coil and is imaged at the iso-
centre in transverse, sagittal and coronal planes.
Analysis
Geometric distortion is measured as the Coefficient of Variation (CV) of the
linear distance measurements in each direction. The CV indicates the degree of
variation of the distance measurements from one another. The coefficient of
variation is defined as:
standard deviation
CV = x 100%
mean
Results
The optimum value for the coefficient of variation is zero, hence the lower the
coefficient of variation, the lower the in-plane distortion and the better the
performance of the system. Table 30 shows a comparison between open
systems for the mean coefficient of variation in three gradient directions. Graph
9 presents the coefficient of variation as a measure of distortion for each
gradient direction.
30
Technical evaluation
GE-PRO
PH-PAN
HI-AIR
SI-CON
0 0.25 0.5
distortion (%)
Report 05097: 0.2T-0.23T open MRI systems - Issue 5
31
Technical evaluation
Analysis
Distortion is assumed to be minimal for the nine disks within the central Region
of Interest (ROI) 1, as shown in Figure 2. The positions of these disks are used
to calculate the actual pixel size. The predicted positions of the remaining 36
disks can then be calculated. The distortion is defined as the difference
between the predicted and actual disk positions in both x and y directions within
ROI 2 and ROI 3. ROI 2 consists of the 16 disks around ROI 1 whilst ROI 3 is
defined as the 25 disks around the edge of the phantom.
Results
The optimum value for distortion is zero, hence the lower the distortion the
better the performance of the system. Table 31 shows the results for the
evaluation on the PH-PAN system. No results are available from systems
evaluated before 2005. Graph 10 and Graph 11 present the distortion results in
each ROI.
32
Technical evaluation
Table 31: Large FOV distortion measurements (in mm) for x and y
directions*
x-direction y-direction Mean
ROI 2
PH-PAN 3.43 (2.69) 3.12 (2.99) 3.27 (2.84)
ROI 3
PH-PAN 6.08 (4.73) 7.26 (5.27) 6.67 (5.00)
*Values in brackets correspond to measurements taken on images with geometric correction switched on.
with geometric
correction
without geometric
correction
PH-PAN
0 2 4 6 8 10
distortion (mm)
with geometric
correction
without geometric
correction
PH-PAN
0 2 4 6 8 10
distortion (mm)
33
Technical evaluation
Analysis
The slice width is measured from the full-width-half-maximum (FWHM) of the
slice intensity profile.
Results
Table 32 shows a comparison between open systems for the mean measured
slice width for three imaging planes. Graph 12 and Graph 13 present a
comparison of the error in the measured slice width.
34
Technical evaluation
Graph 12: Comparison of measured slice widths for a nominal slice width
of 3 mm (mean of three planes)*
PH-PAN
SI-CON
HI-AIR
GE-PRO
* The optimum value is 0 ± 0.3 mm. Systems are ranked in order of the absolute deviation from optimum.
Graph 13: Comparison of measured slice widths for a nominal slice width
of 5 mm (mean of three planes)†
PH-PAN *
SI-CON
HI-AIR
Report 05097: 0.2T-0.23T open MRI systems - Issue 5
GE-PRO
35
Technical evaluation
Analysis
Both the 2D and 3D imaging speed tests require a fixed volume to be acquired
using 2D and 3D fast imaging sequences. The aim of these tests is to measure
data acquisition speed in voxels/second. The voxel size is defined by the image
matrix and the number of slices in the fixed range. Both the 2D and 3D imaging
speed tests allow for the use of parallel imaging techniques.
Imaging speed has been calculated as the ratio of imaged voxels to total scan
time. The total scan time is the time displayed by the scanner, rather than the
number of phase-encoding steps × repetition time (TR).
Results
The numerical results of the 2D and 3D imaging speed tests are presented in
the Appendix. No results are available for Fast Spin Echo (FSE) and Echo
Planar Imaging (EPI) sequences for systems evaluated before 2005. Graph 14
and Graph 15 show a comparison of 2D imaging speed with Gradient Echo
(GRE), EPI and FSE sequences, respectively. Graph 16 and Graph 17 show a
comparison of 3D imaging speed with GRE and FSE sequences, respectively.
Report 05097: 0.2T-0.23T open MRI systems - Issue 5
36
Technical evaluation
PH-PAN
SI-CON
GE-PRO
HI-AIR
FSE EPI
Report 05097: 0.2T-0.23T open MRI systems - Issue 5
PH-PAN
*No results are available for FSE and EPI-type 2D sequences for systems evaluated before 2005.
37
Technical evaluation
PH-PAN
SI-CON
GE-PRO
HI-AIR
PH-PAN
*No results are available for FSE-type 3D sequences for systems evaluated before 2005.
38
Acoustic noise
The acoustic noise levels measured in this assessment are provided for
indication only. Variations in factors such as room acoustics may mean that
similar noise levels may not be reproduced at a different site - even with an
identical pulse sequence. Acoustic noise levels are given in terms of the
continuous equivalent level, LAEQ, which is A-weighted root-mean-square sound
pressure level (SPL) averaged over the measurement period of 1 minute. The
relevant safety levels follow.
MHRA
Hearing protection is recommended for all patients even when exposure is less
than 99 dB(A). Where sites can demonstrate noise levels significantly below 85
dB(A) then this requirement may be relaxed. Please refer to Guidelines for
Magnetic Resonance Diagnostic Equipment in Clinical Use: Medical Devices
Agency 2002.
Staff
NOISE AT WORK REGULATIONS 1989 (UK)
Employers have legal duty to protect the hearing of their employees. Hearing
protection must be available for workers exposed to 85 dB(A) and must be worn
if levels exceed 90 dB(A). Employers are responsible for performing risk
assessments for employees exposed to noise. This would include staff present
in the MR scan room during imaging.
EU DIRECTIVE 2003/10/EC
This directive becomes effective in the UK in 2006. Compared to the regulations
above action levels will be reduced by 5 dB(A). Substitution and control of noisy
Report 05097: 0.2T-0.23T open MRI systems - Issue 5
39
Technical evaluation
Table 34. Acoustic noise levels for comparative pulse sequences (LAEQ)
Pulse sequence with acoustic noise level dB(A)
System SE FSE 3D-GRE
GE-PRO* 74.2 70.6 79.7
HI-AIR 76.1 68.1 77.0
†
PH-PAN 88.6 80.3 (76.5) 86.3 (84.5)†
SI-CON## 78.2 67.0 78.4
*TR=30 ms for 3DGE sequence
†Value in brackets was measured with SofTone on
##TE=14, ETL=3 fro FSE sequence; TE=9.02, slices=16 for 3DGE sequence
HI-AIR
SI-CON
GE-PRO
Report 05097: 0.2T-0.23T open MRI systems - Issue 5
with softone
PH-PAN
40
Acknowledgements
We would like to acknowledge the support of GE, Hitachi, Philips and Siemens
during this evaluation. In particular we would like to thank GE, Hitachi, Mr Jari
Erkkila, Dr Thomas Andreae from Philips, and Siemens for their expert
assistance during the technical tests.
Report 05097: 0.2T-0.23T open MRI systems - Issue 5
41
References
42
Appendix
Manufacturers’ comments
GE
The manufacturer had no additional comments.
Hitachi
The manufacturer had no additional comments.
Philips
The manufacturer had no additional comments.
Siemens
The manufacturer had no additional comments.
Report 05097: 0.2T-0.23T open MRI systems issue
43
Appendix
2D imaging speed
2D imaging speed: sequence parameters
Table 35: Standard imaging protocol for 2D imaging speed evaluation.
Parameter Standard protocol
Sequence Gradient echo (GRE), echo planar imaging
(EPI), fast spin echo (FSE)
TE (ms) Manufacturer’s choice
TR (ms) Manufacturer’s choice
Flip angle (degrees) Manufacturer’s choice
NSA Manufacturer’s choice
Bandwidth Manufacturer’s choice
Echo train length (where applicable) Manufacturer’s choice
FOV (mm) 250 x 250
Matrix (PE x FE) 256 x 256
Slice width (mm) <5
Range (mm) I To equal 200
Contiguous slices II Maximise
Scan time (min:sec) III Minimise
The Roman numerals I, II and III represent the order in which the sequence parameters are set.
Report 05097: 0.2T-0.23T open MRI systems - Issue 5
44
Appendix
45
Appendix
Table 37: Scan parameters for 2D imaging speed evaluation on the Hitachi
Airis Mate MRI system.
Parameter Standard protocol
Sequence SARGE
TE (ms) 7.5
TR (ms) 740
Flip angle (degrees) 1
NSA 1
Bandwidth Not measured
Echo train length (where applicable) Not applicable
FOV (mm) 250 x 250
Matrix (PE x FE) 256 x 256
Slice width (mm) 5
Range (mm) I 200
Contiguous slices II 40
Scan time (min:sec) III 3:09
The Roman numerals I, II and III represent the order in which the sequence parameters are set.
Table 38: Scan parameters for 2D imaging speed evaluation on the Philips
Panorama MRI system.
Parameter Standard protocol
Sequence T1-TFE2D TSEFAST2D IR-EPI
TE (ms) 3.8 192 117
TR (ms) 7.6 2225.3 13490
Flip angle (degrees) 15 90 90
Report 05097: 0.2T-0.23T open MRI systems - Issue 5
NSA 1 1 1
Bandwidth 578.7Hz/px 385.8Hz/px 375.9Hz/px
Echo train length (where applicable) Not applicable 256 64
FOV (mm) 250 x 250 250 x 250 250 x 250
Matrix (PE x FE) 256 x 256 256 x 256 256 x 256
Slice width (mm) 5 5 5
Range (mm) I 200 200 200
Contiguous slices II 40 48 40
Scan time (min:sec) III 1:05 1:29 0:54
The Roman numerals I, II and III represent the order in which the sequence parameters are set.
46
Appendix
47
Appendix
3D imaging speed
3D imaging speed: sequence parameters
Table 43: Standard imaging protocol for 3D imaging speed evaluation
Parameter Standard protocol
Sequence Gradient echo (GRE), fast spin echo (FSE)
TE (ms) Manufacturer’s choice
TR (ms) Manufacturer’s choice
Flip angle (degrees) Manufacturer’s choice
NSA Manufacturer’s choice
Bandwidth Manufacturer’s choice
Echo train length (where applicable) Manufacturer’s choice
FOV (mm) 250 × 250
Matrix (PE x FE) 128 × 128
Slice width (mm) <5
Range (mm) I 250 × 250 x To equal 200
Contiguous slices II Maximise
Scan time (min:sec) III Minimise
The Roman numerals I, II and III represent the order in which the sequence parameters are set.
TR (ms) 23
Flip angle (degrees) 30
NSA 1
Bandwidth Not measured
Echo train length (where applicable) Not applicable
FOV (mm) 250 × 250
Matrix (PE x FE) 128 × 128
Slice width (mm) 1.6
Range (mm) I 250 × 250 × 198.4
Contiguous slices II 124
Scan time (min:sec) III 6:26
The Roman numerals I, II and III represent the order in which the sequence parameters are set.
48
Appendix
Table 45: Scan parameters for 3D imaging speed evaluation on the Hitachi
Airis Mate MRI system
Parameter Standard protocol
Sequence SARGE
TE (ms) 7.9
TR (ms) 25
Flip angle (degrees) 1
NSA 1
Bandwidth Not measured
Echo train length (where applicable) Not applicable
FOV (mm) 250 × 250
Matrix (PE x FE) 128 × 128
Slice width (mm) 1.6
Range (mm) I 250 × 250 × 200
Contiguous slices II 125
Scan time (min:sec) III 6:40
The Roman numerals I, II and III represent the order in which the sequence parameters are set.
Table 46: Scan parameters for 3D imaging speed evaluation on the Philips
Panorama MRI system
Parameter Standard protocol
Sequence T1-FFE3D TSE
TE (ms) 2 192
TR (ms) 4 2225.3
Flip angle (degrees) 20 90
Report 05097: 0.2T-0.23T open MRI systems - Issue 5
NSA 1 1
Bandwidth 1041.7 Hz/px 500 Hz/px
Echo train length (where applicable) Not applicable 64
FOV (mm) 250 × 250 250 × 250
Matrix (PE x FE) 128 × 128 128 × 128
Slice width (mm) 1.4 4.2
Range (mm) I 250 × 250 × 201.6 250 × 250 × 201.6
Contiguous slices II 144 48
Scan time (min:sec) III 1:38 1:29
The Roman numerals I, II and III represent the order in which the sequence parameters are set.
49
Appendix
50