The DAQSTATION DX100P/DX200P pharmaceutical models provide electronic record keeping functions that comply with the
requirements of FDA regulation 21CFR Part 11.
DAQSTATION pharmaceutical models display measured data in real-time on a high resolution color TFT liquid crystal display.
Data can be saved to CompactFlash memory card or ZIP disk storage media. Electronic signatures can be added to the saved
data records at the DAQSTATION itself or the included PC application software.
Application Software
• PC Software designed to be used in conjunction with the DX100P/DX200P can also apply electronic signatures in the same
manner as the DX100P/DX200P.
• Electronic signature information is stored as an attachment to the measurement file in order to protect the original data.
• Configuration change is supported via Ethernet.
High Reliability
• Internal non-volatile flash memory does not require battery backup. Data and configuartion settings are saved during any power
outages.
• Conforming to the IEC529-IP65 and NEMA No. 250 TYPE 4 standards, the front bezel protects against dust and water wash-
down intrusion.
Improved efficiency, and quality control are keywords that companies focus on in todays manufacturing environment. As demand
for these goals increases, the information required to make decisions that affects them also increases.
Until now, industrial recorders were used primarily to observe and record batch measurement data, but in order to quickly extract
the precise information needed in a given situation from this expance of data, recorders with a high degree of information pro-
cessing ability have become a necessity.
YOKOGAWA, the on-going world leader in recording technology, introduces its newest data acquisition station, the DX100P/
DX200P DAQSTATION to all members of pharmaceutical related industries. DX100P DX200P
2 3
Currently, recorders are used in a wide range of applications
during the manufacture of pharmaceutical products.
Ethernet
4
Complies with Electronic Recording Regulation (21 CFR Part 11)
Saving Data in Binary Format
The DX100P/DX200P saves measured data, measured
settings, and the operation log into one, binary file.
Binary data is tamper proof, offering a high level of security.
DX100P/DX200P and the PC software can provide an alarm if
the file is damaged or altered in any way.
Batch Function
A batch name (batch number + lot number) can be assigned in
operation mode which is accessible to users with appropriate
access rights. It is possible to configure automatic
incrementing of lot numbers at the start of each batch if
necessary. Assigning batch names provides a reference that
you can use to retrieve historical measurement data.
You can also store comments related to the measured data
along with 3 lines of User information.
Log In Function
Administrators can assign up to 90 user names in the system
mode, and by configuring log in modes, can specify which
functions are available to each user, as well as limiting the
number of people operating the DX100P/DX200P. With log in
mode settings the recorder operation can be determined on a
user by user basis.
Audit Trail
The configuration of the DX100P/DX200P cannot be changed
whilst the DX100P/DX200P is storing data. Configuration
changes made are automatically stored to the recorder media
once any changes have been completed.
5
Complies with Electronic Recording Regulation (21 CFR Part 11)
Electronic Signature Function
After a batch is complete it is displayed in the DX100P/DX200P’s
historical display mode, or by use of the associated PC software.
After checking the historical data ,it is possible to sign the batch
record. Information concerning the batch review such as, Pass/
Fail determination and comments, can be added as the record is
signed. Three levels of signature are possible such as operator
level, supervisor level, and quality control level. The original data
is in no way affected. Signing a record involves inputting a
password, or user ID and password. With the sign record func-
tion, you no longer need paper copies for document control.
Sign Record
Log In Function
A user name and password, or user name, user ID, and
password are necessary to log in. The DX100P/DX200P
checks each user name against previous user IDs and
passwords, and prevents duplicate password registration.
Additionally the DX100P/DX200P checks for duplicate user
names.
Also, with the automatic password expiration function, persons
attempting to log in with passwords exceeding previously
entered expiration dates will be prompted to have their
passwords renewed.
Password Renewal
Validation Documentation
Documents such as IQ (Installation Qualification) and OQ (Operational Qualification) must be completed as part of an
FDA compliant system within a pharmaceutical manufacturing plant.
Validation Documentation (sold separately) is a tool that can help you simplify the process of validating the DX100P/
DX200P within an FDA validated process.
Accessories
DX100P (Electronic file) 438221 DX200P (Electronic file) 438224
DX100P (A4 sized paper) 438222 DX200P (A4 sized paper) 438225
DX100P (Letter sized paper) 438223 DX200P (Letter sized paper) 438226
6
DAQSIGNIN (21 CFR Part 11 Compliance Application Software)
DAQSIGNIN (standard accessory software package) allows you to display batch records and measured data along
with operation logs and configuration data that have been stored on the DX100P/DX200P. Additionally electronic
signatures can be added to the files, allowing for example quality control to sign records within a network environ-
ment at some time after the records have been completed.
Management Software
Viewer Software
By entering the correct password or user ID/passwords, you
can sign data files generated by the DX100P/DX200P that you
have redisplayed or checked in trend, digital, circular, alarm
table display, message table display, and other formats to an
attached data file without changing the original data. If
someone signed the file previously, you can confirm the
signature status, check the comments, and then sign it
yourself under your own log in.
Along with checking the configuration settings and operation
log on the DX100P/DX200P, you can also make printouts for
Electronic Signature
each batch record. Also, you can read in numerical values from
the displayed data using the cursor, perform interval arithmetic,
and convert files to ASCII, Excel, and Lotus 1-2-3 format.
Operation Log
Settings Software
You can display, edit, and save configuration settings related to
the measurement and calculation channels, and settings related
to the screen display and other items, and transfer them via
Ethernet or external medium to the DX100P/DX200P. Only
Administrator can change the configuration via Ethernet. (When
recording or MATH is stopped.) Also, by opening several setting
files, you can compare how the configuration changes were
carried out, and obtain an audit trail.
You can also print out the configuration, and control them as
paper documents.
Settings Software
Communicator
7
Pursuing good operability with new functions
Easy Text Entry Option
A new wireless remote control option greatly simplifies text entry operations on DX100P/DX200P models.
Control and setting parameters can now be input by remote control !
ID31
ID00 ID01
All operations can be performed by the remote The remote control can:log into
control terminal. a DX100P/DX200P
One remote control can operate up to 32 enter long text messages enter
DX100P/DX200P units. setting paramete
Time Sync
DX100P/DX200P PC
(SNTP client) (SNTP client)
8
Specifications
See the DX100P/DX200P General Specifications documents (GS 04L05A01-00E, 04L06A01-00E) for complete product specifications.
9
Specifications
See the DX100P/DX200P General Specifications documents (GS 04L05A01-00E, 04L06A01-00E) for complete product specifications.
Event data: Measurement data: 2 bytes/record Web server function: Displays the DX100P/DX200P's screen, alarm information,
Calculation data: 4 bytes/record instantaneous values, etc. on a browser.
Sampling time: Example (for DX106P: 6 measurement channels, 0 calcula- FTP client function: Automatic file transfer from DX100P/DX200P unit (display
tion channels) data files, event files, report file snap shot file)
Display data files only FTP server function: Manual file transfer of information on removable storage media,
directory editing, file deletion, and checking free space on
Display updating removable storage media, working through a host computer
(min/div) 1 minutes 5 minutes 20 minutes 30 minutes 60 minutes 240 minutes
Monitor function: Real-time remote monitoring of DX100P/DX200P
Saving interval measurement data (special protocol)
2 seconds 10 seconds 40 seconds 60 seconds 120 seconds 480 seconds
(seconds) Setting function: Configuration of DX100P/DX200P via communication
Approx. Approx. Approx. Approx. Approx. Approx. (special protocol)
Sampling Time
86 hours 18 days 72 days 108 days 217 days 868 days SNTP client function: The time on the DX100P/DX200P can be synchronized to
the time of a SNTP server.
Event data files only
SNTP server function: The DX100P/DX200P can operate as a SNTP server.
Display updating
(min/div) 1 minutes 5 minutes 10 minutes 30 minutes 60 minutes 120 minutes ■ Power Supply
Rated supply voltage: 100–240 VAC (automatic switching)
Sampling Time Approx. Approx. Approx. Approx. Approx. Approx. Operating supply voltage range: 90–132, 180–264 VAC
69 hours 14 days 28 days 86 days 173 days 347 days Rated supply frequency: 50/60 Hz (automatic switching)
File saving method: Auto save DX100P power consumption: Power supply voltage With LCD saver ON Normal mode Maximum
Display data file: Saved to removable storage media at 100 VAC Approx. 30 VA Approx. 32 VA Approx. 45 VA
fixed intervals (10 minutes to 31 days).
Event file: Saved to removable storage media at 240 VAC Approx. 42 VA Approx. 47 VA Approx. 62 VA
fixed intervals (3 minutes to 31 days) DX200P power consumption: Power supply voltage With LCD saver ON Normal mode
when using free trigger. Maximum
100 VAC Approx. 50 VA Approx. 53VA Approx. 75 VA
■ Log in Function: Not all operations are allowed when starting login mode
with the power ON (only the monitor display can be 240 VAC Approx. 78 VA Approx. 80 VA Approx. 106 VA
switched (on/off selectable)) . Enter user name, user ID,
and password to login to the DX100P/DX200P.
System administrator: 3 names can be registered, access to all keys available
Normal operating requirements
General user: 90 names can be registered, and access to key operations Supply voltage ranges : 90 to 132, 180 to 250 V AC
and limitations on the sign record function can be assigned Supply frequencies : 50 Hz ± 2%, 60 Hz ± 2%
using the login mode settings. Ambient temperature : 0–50°C
Login mode settings: 30 patterns Ambient humidity : 20 to 80% RH (at 5 to 40°C )
Password expiration: Select Off, 1 month, 3 months, or 6 months
■ Batch Function: In operation mode, you can input a batch name (a batch Reference performance specifications
number plus an 8 digit lot number for a maximum of 32 Measurement and display accuracy:
characters) and a comment (up to 3 lines, 32 characters each). (reference operating conditions: temperature of 23 ± 2°C,
Automatic incrementing of lot numbers at each batch start. humidity 55 ± 10% RH, supply voltage of 90 to 132 or 180
Preset headers 1, 2, and 3 (each having a maximum of 64 to 250 V AC, supply frequency of 50/60 Hz ± 1%, minimum
characters) can be viewed on the batch input screen. 30 minutes warmup time; no vibrations or other which
Display event data files: The following information is added to the attached data file: would adversely affect the performance of measuring
• User name instruments)
• Header 1 (can be used for the application description etc.)
• Header 2 (can be used for the supervisor name etc.) Maximum digital
Input type Range Measurement accuracy (digital reading) reading resolution
• Header 3 (can be used for the manager name etc.)
• Batch name (a batch number plus an 8-digit lot number 20 mV 10 µV
for up to 32 characters) 60 mV 10 µV
• Comments (up to 32 characters, 3 lines each) 100 µV
200 mV ±(0.1% of rdg + 2 digits)
DC voltage
■ Sign Record Function: After checking the measured data, electronic signatures on 2V 1 mV
3 levels, pass/fail determinations, and comments (up to 32
6V 1 mV
characters) can be recorded.
20 V 10 mV
■ Alarm Functions
Number of settings: Maximum 4 per channel 50 V ±(0.1% of rdg +3 digits) 10 mV
Alarm types: Upper/lower limits, difference upper/lower limits, change R ±(0.15% of rdg + 1°C)
rate increase/decrease limits, delay upper/lower limits
S R and S are ±3.7°C for 0 to 100°C, and
(alarm delay)
Change rate alarm interval: Measurement period × 1–15 ±1.5°C for 100 to 300°C
Hysteresis: Switched between ON (0.5% of display span) and OFF B B is ±2°C for 400 to 600°C; accuracy not
(same for all channels/levels)
guaranteed for less than 400°C
Display: Status (alarm type) display and common alarm display
shown on digital display unit when alarm occurs. ±(0.15% of rdg +0.7°C)
K
Switching between display holding/non-holding. Thermocouple ±(0.15% of rdg +1°C) for -200 to -100°C
Notification: Email notification (without
reference E ±(0.15% of rdg +0.5°C) 0.1°C
Storage: Stored information: Alarm occurrence/clear time, alarm type junction
Number of stored records: Most recent 240 records maximum J ±(0.15% of rdg +0.5°C)
compensation
ACK: Specific or global alarms can be controlled ccuracy) T ±(0.15% of rdg + 0.7°C) for -200 to -100°C
Output:
N ±(0.15% of rdg +0.7°C)
Output points: DX100P (with option): 2, 4, or 6 points
DX200P (with option): 2, 4, 6, 12, or 24 points W ±(0.15% of rdg +1°C)
Operations: Switching between excitation/non-excitation, holding/non-holding L ±(0.15% of rdg +0.5°C)
■ Communications Interface U ±(0.15% of rdg + 0.7°C) for -200 to -100°C
Media: 10BASE-T Pt100
Protocol: SMTP, HTTP, FTP, TCP, UDP, IP, ARP, ICMP, SNTP RTD ±(0.15% of rdg + 0.3°C)
Email sending function: JPt100
Notification types: The following information is presented by email:
Reference junction compensation: INT (internal)/EXT (external) switching (common to all channels)
Alarm notification: Alarm information is presented when an alarm occurs or is cleared
Reference junction compensation accuracy:
System notification: Notification of time when power is interrupted/restored
Types R, S, B, W: ±1°C
Notification of time remaining when internal memory
Types K, J, E, T, N, L, U: ±0.5°C (for measurement at 0°C or higher)
overwriting starts
Maximum input voltage: 2 VDC or lower voltage range and thermocouple: ±10 VDC (continuous)
Notification of remaining free space when remaining space
6 VDC or higher voltage range: ±60 VDC (continuous)
in storage media falls to 10% or 6 MB.
Input resistance: 2 VDC or lower voltage range and thermocouple:10ΩMor higher
Periodic notification: Periodic notification of instantaneous values at preset times
6 VDC or higher voltage range: approximately 1 MΩ
or intervals
Input external resistance: DC voltage, thermocouple input:2 kΩ or lower
Report notification: Notification of report data when report time-out occurs
RTD input:1 wire, 10 Ω or less (all three wires equal)
(with /M1 option)
Input bias current: 10 nA or less
User invalid: Notification of user becoming invalid because of entering
Maximum common mode noise voltage:
the wrong password three times.
250 VAC rms (50/60 Hz)
Notification addressee: 2 address groups
Common mode rejection ratio (CMRR):
(multiple addresses may be specified in each group, with a
120 dB (50/60 Hz ±0.1%, 500Ω unbalanced, across minus
maximum of 150 characters per group) terminal and ground)
10
Normal mode rejection ratio (NMRR): Constants DX100P: Up to 12 constants can be set.
40 dB (50/60 Hz ±0.1%) DX200P: Up to 30 constants can be set.
Maximum noise voltage across channels: Online digital communications input: Can be used for calculation formulas other than statistical
250 VAC rms (50/60 Hz) calculations.
Interference across channels:120 dB (for 500Ω input external resistance and 60 V input DX100P: 12 channels
to other channel) DX200P: 30 channels
Remote inputs: Up to 8 remote inputs can be used. Remote statuses (0/1)
can be used in calculation formulas.
Reporting functions:
Option specifications Report types: Hourly reports, daily reports, hourly + daily reports, daily +
■ Easy Test Entry weekly reports, daily + monthly reports
Number of units under control : Up to 32 units by ID setting Calculation types: Average values, maximum values, minimum values,
Max. communication distance : Up to 8m, depending on battery strength area of use totalized values
Operational functions ■ Cu10/Cu25 RTD Input/3-Wire Isolated RTD Input (/N1)
• User Name/User ID/Password input for logging in This option enables Cu10 and Cu25 inputs in addition to the standard inputs.
• Message input
• Engineering mode setting ■ 3-Wire Isolated RTD Input (/N2)
• System mode setting With this option, all RTD input points are isolated (A, B, and b are all isolated).
• Trend/Digital/Bar Graph display change *Only available with the DX106P, DX112P, DX210P, DX220P, and DX230P.
■ Calibration Correction
■ 24 VDC/AC Power Driven Model (/P1)
Functions: Input value correction with linearization
Rated supply voltage: 24 VDC or 24 VAC (50/60 Hz)
Points: Selectable from off, 2 to 16
Operating supply voltage range: 21.6 to 26.4 VDC/AC
Target channel: Measurement channel
Target range: All range mode DX100P power consumption: Power supply voltage With LCD saver ON Normal mode Maximum
■ Alarm Relay Contact Output (/AR1, /AR2, /A3, /A4*, /A5*) 24 VAC Approx. 17 VA Approx. 19 VA Approx. 30 VA
Functions: Relay output through back side when alarm occurs
Outputs: 2, 4, 6, 12* or 24* 24 VAC(50/60 Hz) Approx. 28 VA Approx. 32 VA Approx. 45 VA
Relay contact capacitance: 250 VDC/0.1 A (resistance load), 250 VAC (50/60 Hz)/3 A DX200P power consumption: Power supply voltage With LCD saver ON Normal mode
Output form: NO-C-NC (switching between excitation/non-excitation, Maximum
AND/OR, holding/non-holding) 24 VAC Approx. 34 VA Approx. 35 VA Approx. 54 VA
* /A4 and /A5 are for DX200P only. 24 VAC(50/60 Hz) Approx. 50 VA Approx. 53 VA Approx. 76 VA
■ Serial Communications (/C2, /C3)
■ Remote Control (/R1)
Functions: Allows the host computer to control (available control
The remote control can be used to control the following through contact input (as many
commands are limited) the DX100P/DX200P as well as
as 8 points can be set)
receive data from the DX100P/DX200P.
• Memory start/stop (level)
Media: EIA RS-232 (/C2) or RS-422-A/485 (4-wire) (/C3) compliant
• Time setting (time set to reference time through contact; trigger; 250 ms or greater)
Protocol: Special protocol or Modbus
• Calculation start/stop (level)
Synchronization method: Start-stop synchronization
• Calculation data reset (trigger; 250 ms or greater)
Communication method (RS-422-A/485):
• Manual sampling (trigger; 250 ms or greater)
4-wire half-duplex multi-drop connection (1:N, where N is 1–32)
• Message writing (as many as 8 types can be set; trigger; 250 ms or greater)
Transfer rate: 1200, 2400, 4800, 9600, 19200, 38400 bps
• Alarm ACK (trigger; 250 ms or greater)
Data length: 7/8 bits
• Snapshot (trigger; 250 ms or greater)
Stop bit: 1 bit
Parity: ODD, EVEN, NONE ■ 24 VDC Transmitter Power Supply Output (/TPS2*, /TPS4, /TPS8*)
Maximum communication distance: 1.2 km (RS-422-A/485) Output voltage: 22.8–25.2 VDC (for rated load current)
Communication mode: Control and settings I/O are in ASCII mode. Rated output current: 4–20 mA DC
Measurement data are output in ASCII or binary mode. Maximum output current: 25 mA DC (overcurrent assured operation current:
Modbus communication: Operation mode: RTU MASTER or RTU SLAVE approximately 68 mA DC)
RTU MASTER: Capable of data acquisition for 8 packet groups.
Registers of a continuous data type in the same slave can
be registered in a single packet group.
RTU SLAVE: Outputs measurement/calculation data and alarm statuses.
Evaluated Barcode Scanner Metrologic Inc.
MS 9540-RS (RS-232 interface)
Symbol Technologies Inc.
LS 1902-RS (RS-232 interface)
■ VGA Output (/D5, DX200P only)
Enables connection to external display device.
■ FAIL/Memory End Output (/F1)
Select FAIL output, memory output, or batch start/stop output on 2 relay outputs.
FAIL Output: Relay output when system error occurs
Memory mode output: Relay output a specified number of hours before internal
memory overwriting starts (1, 2, 5, 10, 20, 50, or 100
hours), or when available space on the external memory
medium falls below 10% or 6 MB.
Batch start/stop: Batch start/stop status relay output
User invalid: When a user becomes invalid because of entering the
wrong password three times.
Login status: When the login function is enabled and there is a user
logged in the DX100P/DX200P.
Relay contact capacitance: 250 VDC/0.1 A (resistance load), 250 VAC (50/60 Hz)/3 A
■ Clamped Input Terminal (/H2)
A clamped input terminal is used as an input terminal.
■ Desktop Type (/H5[ ], /H5)
Includes carrying handle and power cord (model /H5 does not include power cord)
■ Mathematical Functions (/M1)
These functions enable the calculations listed below, as well as displaying and
recording trends and digital values on calculation channels.
Number of calculation channels:DX102P, DX104P: 8 channels
DX106P, DX112P: 12 channels
DX204P, DX208P: 8 channels
DX210P, DX220P, DX230P: 30 channels
Calculation types:
General calculations: Arithmetic calculations (+, -, *, /), square roots, absolute values,
common logarithms, exponents, powers, relational calculations
(<, >, =, ≠), logical calculations (AND, OR, NOT, XOR)
Statistical calculations: Time-series data averages, maximum values, minimum
values, totalized values
Moving averages: Moving averages are determined for calculation results.
11
Model Code Accessories
DX100P Accessories (Sold separately)
Suffix Optional Product Model(Part number) Specifications
Model code Description
code code
438221 Electronic file for DX100P
DX102P DAQSTATION DX100P (2ch)
438222 A4 sized paper for DX100P
DX104P DAQSTATION DX100P (4ch)
438223 Letter sized paper for DX100P
DX106P DAQSTATION DX100P (6ch) Validation Document
DAQSTATION DX100P (12ch)
438224 Electronic file for DX200P
DX112P
External -2 100MB Zip (with medium) 438225 A4 sized paper for DX200P
Memory -3 CompactFlash memory card (CF+Adapter) 438226 Letter sized paper for DX200P
-5 250MB Zip (with medium) 415920 25Ω±00.1%
Shunt resistor for
Display Language -1 Emglish, deg F & Summer/winter time (with English DAQSIGNIN) 415921 100Ω±0.1%
screw terminal
Option Specifications /AR1 Alarm output 2 points/Remote control *1*2 415922 10Ω±0.1%
/AR2 Alarm output 4 points/Remote control *1*2 438920 250Ω±0.1%
/A3 Alarm output 4 points *1*3 Shunt resistor for
438921 100Ω±0.1%
clamped terminal
/C2 RS-232 interface (including MODBUS) *4*5 438922 10Ω±0.1%
/C3 RS-422-A/485 interface (including MODBUS) *4*5 A1053MP 100MB
/F1 FAIL/memory end output *3 Zip disk
A1056MP 250MB
/H2 Clamped input terminal
CompactFlash memory card (CF+Adapter) B9968NL 32MB or more
/H5 Desktop type(without power cord, screw type power terminal) *6
A1347EF(DX100P) 250V, 1ATL
/H5[ ] Desktop type (with power cord)*7
A1352EF(DX100P/P1) 250V, 4ATL
/M1 Mathematical function (with report function) Fuse
/N1 Cu10,Cu25 RTD input/3 legs isolated RTD
A1423EF(DX200P) 250V, 1.25ATL
/N2 3 legs isolated RTD *8 A1463EF(DX200P/P1) 250V, 6.3ATL
/P1 24V DC/AC power supply Mounting bracket B9900BX –
/R1 Remote control Module removal handle 790581 –
/TPS2 24V DC Power Supply for Transmitter(2 loop) *9 Input terminal 438227 For /KB1, /KB2 option
/TPS4 24V DC Power Supply for Transmitter (4 loop) *10
/KB1 Easy Text Entry (with input terminal) *11*12 Dimensions
/KB2 Easy Text Entry (without input terminal) *11
/CC1 Calibration Correction +0.4
136.50
*1 /AR1,/AR2 and /A3 cannot be specified together. *2 If /AR1 or /AR2 is specified, /R1 cannot be specified. *3 If /A3 is specified, 137 0
+2
Unit:mm
/F1 cannot be specified. *4 /C2 and /C3 cannot be specified together. *5 In case that Modbus master function is utilized, /M1 must DX100P
be specified. *6 In case that 24 VDC/AC power supply(/P1) and desktop type are specified, /H5 must be specified. /P1 and /H5[ ]
cannot be specified together. *7 /H5[ ](D-Power cord UL, CSA st'd, F-Power cord VDE st'd, R-Power cord SAA st'd, J-Power cord
BS st'd, H-Power cord GB st'd) *8 /N2 cannot be specified for DX102P and DX104P. *9 In case that /TPS2 is specified, /TPS4, +2
/AR2, /A3 or /F1 cannot be specified. *10 In case that /TPS4 is specified, /TPS2, /AR1, /AR2, /A3 or /F1 cannot be specified. *11 137 0
/KB1 and /KB2 cannot be specified together. *12 In case that /KB1 is specified, input terminal(4382 27) is attached.
9.3
code 40.8 thickness: 2 to 26
+0.4
151.5
136.50
144
DX208P DAQSTATION DX200P (8ch)
101.8
DX210P DAQSTATION DX200P (10ch)
42.3
7.5
103.3 mounting hardware)
144
DX230P DAQSTATION DX200P (30ch) 151.5
Min. 361
Option Specifications /AR1 Alarm output 2 points/Remote control *1*2
+2
2810
/H5 Desktop type (without power cord, screw type power terminal)*6
180
mounting hardware)
/N2 3 legs isolated RTD *8 180
288
/P1 24 VDC/AC power supply 294.6