Professional Documents
Culture Documents
1
3/2/2024
We can study
• Defects, porosity
Electrons in
Electrons out
or: x-rays out
Interaction of electrons
with the specimen surface
produces electrons signals
• Secondary electrons
• Backscattered electrons
2
3/2/2024
Depth of focus
MENA3100
3
3/2/2024
Elemental mapping
Ce
Fe Sr
In-situ imaging
• A modern SEM can be equipped with various
accessories, e.g. a hot stage
4
3/2/2024
5
3/2/2024
6
3/2/2024
Resolution
Resolution
7
3/2/2024
Abbe’s Equation
Abbe’s Equation describes resolution in perfect optical system
8
3/2/2024
• No energy loss
Inelastic Scattering
9
3/2/2024
Interaction volume
Interaction volume
is like hemisphere
is like Tear drop
Use of a high accelerating voltage will result in deep penetration length and a
large primary excitation region resulting in loss of detailed surface information
of sample
Secondary Electrons
• Emission of loosely bound electron from the sample surface upon
interaction of energetic electrons (ionisation).
• SE have low energy 3-5 eV, escape only from a region within few
nanometers of the material surface.
10
3/2/2024
Secondary Electrons
Effect of surface topography and position of detector on the secondary
electron detection
Secondary Electrons
Effect of Accelerating voltage
11
3/2/2024
Secondary Electrons
Effect of Accelerating voltage
Secondary Electrons
Effect of Accelerating voltage and Atomic number on yield of SE and BSE
The number of SE and BSE emitted from the specimen for each incident electron are
known as the secondary electron coefficient (δ) and the backscattered electron
coefficient (η), respectively.
12
3/2/2024
Secondary Electrons
Intensity of Secondary Electrons decreases as the distance
increases from incident beam
• A BSE is one which has undergone single or multiple scattering events and which
escapes from the surface with an energy greater than 50 eV.
• Roughly 10-50% of the beam electrons are backscattered toward their source,
and on average these retain 60-80 % of their initial energy.
• BSEs yield ~6% of light element carbon (dark will be image) but ~50% for heavier
element like gold or tungsten (brighter will be image as more electrons are
deflected).
• Lateral resolution of BSEs image is worse (1.0 μm) than SEs (10 nm)
• BSEs carry information about features that are deep beneath the surface.
13
3/2/2024
14
3/2/2024
• Only those BSE will reach detector which are travelling along the
direct line of sight towards the detector , resulting in lower yield.
15
3/2/2024
• By applying voltage to the detector, electron-hole pair will not recombine and can
be separated, thus producing current which can be amplified.
• The detector is a thin flat plate which is mounted on the objective pole piece, and
thus does not interfere with normal operation of the instrument.
16
3/2/2024
SEM Optics
SEM purpose is to produce a fine beam of electrons incident on the specimen
SEM Optics
Condenser Lens
The condenser lens of focal length (fc) produces demagnified image of the filament.
𝑣1
𝑑1 = 𝑑𝑜 ×
𝑢1
do diameter of the filament
Objective lens
The objective lens of focal length (fo) further demagnify the filament image, producing a probe
of diameter d on the surface of the specimen.
𝑣2 𝑤𝑑
𝑑 = 𝑑1 × = 𝑑1 × The probe size can be
𝑢2 𝑢2
controlled by altering
wd or 𝑣2 is the working distance of the mircoscope. the strength of the
condenser lens.
17
3/2/2024
SEM Optics
2
𝛼1
𝐼1 = 𝐼𝑜 ×
𝛼𝑜
SEM Optics
Spherical Aberration
18
3/2/2024
Electromagnetic Lens
Condenser Lens
Condenser lens converges the cone of electron beam to a spot. The closer the
spot to the condenser lens smaller the spot diameter. Condenser lens current
control initial spot size and is referred to as the spot size control.
19
3/2/2024
Aperture
The objective aperture arm fits above the objective lens in the SEM.
Apertures help to
minimise spherical
aberrations.
Aperture
The objective aperture arm fits above the objective lens in the SEM.
0.2 𝑊𝐷
𝐷𝑒𝑝𝑡ℎ 𝑜𝑓 𝐹𝑖𝑒𝑙𝑑 = mm
𝐴𝑀
20
3/2/2024
Aperture
Decreasing the aperture size will reduce the beam angle α for the same WD,
resulting in an enhancement of the depth of field
Aperture
21
3/2/2024
Aperture
Aperture
22
3/2/2024
Astigmatism
Astigmatism
The image in the centre shows a correctly focused image that has also been corrected for
astigmatism. At left and right are examples of poorly corrected astigmatism, seen as
streaking of the image.
23
3/2/2024
24
3/2/2024
Scanning Coils
• Scanner coils or pair of deflector plates (small coils of wire carrying the
controlling current) in the electron column , typically , in the final lens, deflect
the beam in x and y axes so that it scans in a raster fashion over a rectangular
area of sample surface.
• The beam is scanned over the specimen in a series of lines and frames called
a raster. A raster can be thought of as arrays of picture points or pixels.
• Scanning action is produced by altering the strength of the current in the scan
coils as a function of time, so that the beam is moved through a sequence of
positions on the specimen (e.g., locations 1, 2, 3, 4, etc. in Fig. 4.1).
• Slower scan rates allow more electrons to be collected at each point along the
line of beam scan. This produces better quality image.
Scanning Coils
25
3/2/2024
• Signal detection
• Signal processing (amplification, filtering, and digitization of the signals
to improve signal-to-noise ratio and enhance image quality.)
• Image Reconstruction (mapping the detected signals to corresponding
pixels on the screen, where each pixel represents a specific point on the
specimen surface).
• Display on Computer Screen
• The amplified signal from the detector is output to a high quality cathode ray
tube (CRT), and the minimum size of spot which may be obtained on such a
CRT is about 0.1 mm.
• A 100 mm2 CRT contain 1000 x 1000 discrete picture elements or pixels.
• The working resolution of the instrument can not be better than pixel size which
is given by
100 𝜇𝑚
𝑝=
𝑀
Scanned 10 μm x 10 μm
on the specimen, then on
100 mm x 100 mm CRT
screen, the linear
magnification will be
10,000x.
26
3/2/2024
• If probe larger than specimen pixel, the signal from adjacent pixels is merged,
and the resolution degraded.
• If probe size is exactly same as the diameter of the pixels, high resolution
achieves (required for optimum performance)
Resolution
27
3/2/2024
Electron Guns
Requirements
• Adjustable energy
Modern SEM use “Field Emission” source which produces high current and low
energy dispersion
28
3/2/2024
(0.1-30 kV)
29
3/2/2024
• 5 to 10x greater brightness and longer life time compared with conventional
tungsten guns
• LaB6 single crystal 100–200 μm in diameter and 0.5 mm long and is mounted on
graphite which act as the resistive heater to elevate the temperature of the
crystal.
Advantage Disadvantage
-Smaller emission Readily
area reduces spot oxidised at
size. high
-Smaller energy temperature.
spread resulting in
high resolution and To increase life
smaller chromatic time high
aberration vacuum (10-8
-Can be replaced Torr) required
easily in conventional
SEM
30
3/2/2024
• Schottky Emitters
To get emission from the source it is
operated at an elevated K) in a
vacuum of ~10-9 mbar and the
electric field at the tip end is
of the order of ½-1 V/nm.
SE and CFE have superior performance in terms of brightness, source size and life
time. SE has 100 x larger area than CFE and is capable of delivering more than 50x
higher emission current than CFE. Probes smaller than 2 nm can be obtained which
provides higher resolution for SEM image. Field emitters must operate under
ultrahigh vacuum (10-9 Torr) to stabilise emission and to prevent contamination.
Vacuum Systems
• Ultrahigh vacuum is must for SEM (10-6 Torr or better)
• Mechanical pump and diffusion pumps are employed to pump down the
chamber from atmospheric pressure.
31
3/2/2024
Vacuum Systems
Specimen preparation
32
3/2/2024
Sputtering
• A technique used to deposit thin films of a material onto a surface ( “substrate").
• First a gaseous plasma is created and ions are accelerated from this plasma into
some source material ("target").
• The source material is eroded by the arriving ions via energy transfer and is ejected
in the form of neutral particles - either individual atoms, clusters of atoms or
molecules.
• As these neutral particles are ejected they will travel in a straight line unless they
come into contact with something - other particles or a nearby surface. If a
"substrate" such as a Si wafer (Sample to be imagined) is placed in the path of these
ejected particles it will be coated by a thin film of the source material.
SEM image of Emiliania huxleyi Cells SEM image of virus (T4 Bacteriophage)
Mag: 25,000X
33
3/2/2024
SEM of Stress corrosion cracking in 300 Pitting corrosion in 300 series stainless
series steel steel backscattered SEM image
34
3/2/2024
SEM micrographs for the Fe-10Ni-15Al alloy aged at 750°C for (a) 75, (b) 250, and (c) 500 h,
and at 920°C for (d) 25, (e) 100 and (f) 200 h.
http://www.intechopen.com/books/scanning-electron-microscopy/sem-
analysis-of-precipitation-process-in-alloys
35
3/2/2024
36