Professional Documents
Culture Documents
Gupta 2016
Gupta 2016
Texture Analyser
Gourab Sen Gupta, Weizhong Shi Ken Mercer, John Bronlund
School of Engineering and Advanced Technology School of Engineering and Advanced Technology
Massey University Massey University
Palmerston North, New Zealand Palmerston North, New Zealand
Email: g.sengupta@massey.ac.nz, shwzh1990@gmail.com Email: {k.a.mercer, j.e.bronlund}@massey.ac.nz
Abstract—Existing texture analysers are expensive and offer [2]. The textual properties of foods have the following
limited programming functionalities to control the instrument for characteristics:
repetitive measurements. Moreover, measurement data is logged
1. They are a group of physical properties
into a file for post processing and visualization, with no provision
for it to be displayed in real-time. This paper details the 2. They derive from the structure of the food
electronic design of a low-cost load cell based food texture 3. Belong under the mechanical or rheological subheading
analyser targeted for use in scientific laboratories to evaluate of physical properties.
texture of fruits. Sophisticated programmable devices, such as 4. Sensed by the feeling of touch
high-precision ADC and stepper motor controller/driver have 5. Objective measurement is usually by using functions of
been incorporated in the design. A LCD based input device mass, force, distance and time.
provides a user friendly graphical interface for interaction with
Food scientists should consider the whole characteristics of
the instrument. Further, a novel command based language has
been developed to control the instrument and a user can write his
food and choose to measure the most appropriate property for
own script to control the instrument for an experiment. The the product of interest [6]. It is necessary to select an
instrument will facilitate food science research and product instrument that is capable of measuring these characteristics.
development. The system has been tested on several fruits and The resultant data is required to be logged in a file and
measurement data compared with a standard texture analyser. displayed on the screen to help scientists to do their analysis.
A fruit texture measurement system focuses on measuring
Keywords- food texture, command script, load cell, real-time
the properties of fruits. It includes the analyser, which is
data acquition and display
capable of measuring the tested food material and collecting
I. INTRODUCTION data [7]. The simple structure of a food texture measurement
system is illustrated in Figure 1. While the analyser is taking
Food texture measurement is a way to know the physical measurements, the data is transferred simultaneously to the
structure of food [1]. Malcolm Bourne has elaborated how food computer through a standard protocol. Processed by the
texture properties are perceived by the human senses [2]. software running on the computer, the correlation of the force,
Textures affect consumer preference of accepted foods and time, and speed is displayed on the screen at the end of the test.
every food manufacturer strives to make his product of the
highest quality in terms of texture and flavour [3]. These days,
special instruments are used to imitate the feeling and
perception of human by using sophisticated sensors that
measure the properties of texture in a scientific way. For food
product developers, it can help them transfer the target texture
Figure 1. Functional blocks of a texture measurement system
from the lab to the consumer’s kitchen in a scientific way.
Food texture measurement is a well-established technique In general, a texture analyser consists of four core elements
that can evaluate the mechanical and physical properties of the [6]:
food structure and raw ingredients for pre- and post-quality 1. A probe which is in contact with the food sample. For
control checks [4]. The application of food texture example: penetrating probe, spindle, or cutting blade.
measurement is across a wide range of food types, including 2. A driving mechanism that can provide motion (vertical
fruit, meat, confectioneries, snack bars, cheese, etc. Since the and horizontal) to the probe.
texture is a property related to the sense of touch, it can be 3. A sensory instrument that can detect the resistance
measured easily by using a mechanical method such as force offered by the food sample
[4]. The standard tests in food texture measurement, such as 4. A readout system that can display the result of the test.
compression, puncture and penetration, cutting and shearing,
and extrusion are used to measure the hardness, softness, Currently available texture analysers are quite expensive, in
crispness and other properties of food [5]. The user can collect excess of $10,000. In this paper we propose a low cost solution
the data of the mechanical food texture measurement in a detailing the electronic design and a novel command language
variety of ways. Bourne indicates that the texture is not a for control. The system is evaluated for its accuracy against an
single-dimensional attribute; it is a multidimensional attribute industry standard instrument by doing measurements on a
variety of fruits.
Figure 2. Block diagram of a food texture analyser
Figure 5. The result of tests done on squeeze ball using standard texture
analyser (force in Newton vs. time in sec)
20 Test1(New
Instrument) Touch
0 screen Load cell
under the arm
0.005
0.9
1.065
1.245
1.425
1.605
0.18
0.36
0.54
0.72
Protection
Figure 10. Combining the results of the standard texture analyser and the new system
instrument for kiwifruit (force in gm vs. time in sec, both scaled by 10)
Shaft Ball
TABLE IV. THE MAXIMUM, MINIMUM, MEAN AND STANDARD DEVIATION Screw
OF ERROR FOR MEASUREMENTS DONE ON KIWIFRUIT WITH THE STANDARD
AND THE NEW INSTRUMENT
Figure 11. The new texture analyser
Sample fruit: Kiwi fruit (Force range 5~10 kg)
Maximum Error 7.797502 N The comparison of experimental results obtained using a
standard texture analyser and the newly designed instrument
Minimum Error -6.18065 N has been presented. The avocado, kiwifruit, and persimmon
Mean Error 0.429691 N were used as examples to test the performance of the new
Standard Deviation 2.786301 N equipment. Results show that the machine designed for this
project can be as accurate as the standard one. The complete
range of force of the load cell, from 0 to 20 kg, has been tested.
Table IV summarises the indicators of the accuracy of the
measurement data for the tests done on kiwifruit using the However, there is still room for improvement. For example,
standard and the new instrument respectively. From the data, the effect of noise generated by the stepper motor on the
the minimum, maximum, and mean errors have been calculated measurement data needs to be addressed. On a few occasions,
which are very low. The standard deviation of the error is more noise has resulted in inaccurate readings. Moreover the nature
than that for avocado but still very low (~2.78 N or 0.283 kg) of the noise has been very unpredictable. As future work,
for the range 5 to 10 kg, which is larger than avocado. software code for filtering and noise reduction can be added.
The touch screen will be enhanced to show more information
Table V summarises the indicators of the accuracy of the that includes not only visual buttons but also the real time
measurement data for the tests done on persimmon using the graph generated from the data transferred from the load cell.
standard and the new instrument respectively. From the data,
the minimum, maximum, and mean errors have been calculated The novel contributions of the project are:
which are very low. The standard deviation of the error is very 1. Implementation of a new programmable system that a
low (~0.934 N or 0.095 kg) for the range 10 to 20 kg. user can use to debug and control the whole system.
The programming method uses a new command
TABLE V. THE MAXIMUM, MINIMUM, MEAN AND STANDARD DEVIATION
OF ERROR FOR MEASUREMENTS DONE ON PERSIMMON WITH THE STANDARD language written in the form of a script. The language is
AND THE NEW INSTRUMENT generic and can be used in any other application where
stepper motor control is required.
Sample fruit: persimmon (Force range 10~20 kg)
2. Using advanced technology, both hardware schematic
Maximum Error 2.420821 N design and PCB design were done. The new PCB board
Minimum Error -3.26856 N is much smaller than the original one and combines the
Mean Error 0.239707 N motor driver, analogue-to-digital convertor, power
Standard Deviation 0.934364 N supply and the embedded microcontroller into one.
3. A touch screen based user interface has been
developed. The touch screen has visual buttons to instrument. This is shown in Figure 12. Various speed values
control the basic operations of the instrument. can be set and pushbuttons used to interact with the system.
4. The real time graph of the measurement data can be The text box on the right of the GUI can be used to enter
displayed on the computer screen while the testing is in individual control commands presented in Section III.
progress.
REFERENCES
5. An infra-red sensor has been added to detect the
position of the robotic arm, so that the robotic arm can [1] Wilkinson, C., Dijksterhuis, G., and Minekus, M., From food structure
to texture. Trends in Food Science & Technology, 2000. 11(12): p. 442-
be slowed down from pre-test speed to test speed before 450.
compressing/penetrating the test sample, which could [2] Bourne, M.C., Food texture and viscosity: concept and measurement.
otherwise cause damage to the food. 2002: Academic Press.
[3] Surmacka, S.A., The whys and whats of objective texture measurements.
Canadian Institute of Food Technology Journal, 1969. 2(4): p. 150-156.
[4] Brown, R.D., Food Texture Analysis. American Laboratory, 2010.
http://americanlaboratory.com/914-Application-Notes/485-Food-
Texture-Analysis/
[5] Rosenthal, A.J., Food texture: measurement and perception. 1999.
[6] Bourne, M.C., Basic principles of food texture measurement, in Dough
rheology and baked product texture. 1990, Springer. p. 331-341.
[7] Singh, H., et al., The Analysis of Stress Relaxation Data of Some
Viscoelastic Foods Using a Texture Analyser. Journal of texture studies,
2006. 37(4): p. 383-392.
[8] Arduino Nano. Available from:
http://www.arduino.cc/en/Main/ArduinoBoardNano
[9] L6472 Datasheet, ST Microelectronics. Available from:
http://www.farnell.com/datasheets/1922760.pdf
Figure 12. The Graphical User Interface (GUI) for the instrument [10] AD7192 Datasheet, Analog Devices. Available from:
http://www.analog.com/static/imported-files/data_sheets/AD7191.pdf.
Furthermore, a simple intuitive graphical user interface
(GUI) has been developed to directly interact with the