You are on page 1of 60

DFWL PLUS • DFWLI PLUS

www.diniargeo.com
TABLE OF CONTENTS

1. Introduction and warnings 4


2. Technical features 5
3. Approval 6
4. Connections 7
Single-channel 7
Multichannel 8
5. Programming 9
How to access the programming menu 9
How to save the programming and exit the menu 9
Programming menu 10
6. Communication strings 44
7. Communication commands 49
8. Wiring diagrams 51
CPU board 51
232 serial port with RJ11 connector 51
Optional IN/OUT DFIO board 51
Optional analogue DAC16OSER output board 52
9. Programming errors 53
10. Summary of the parameters 54
11. FAQ - Frequently asked questions 57

CONTENTS BY TOPIC

Calibration
Quick calibration 11
Complete calibration 40

Equalisation
Multi-channel connection 8
Equalisation procedure 39

Safety
Protection of the configuration menu via PIN 43
Protection of the user menu via PIN 43

Communication
Configuration of the serial port for PC 13
Configuration of the serial port for printer 15
Wiring diagrams 50
Communication strings 44
Communication commands 49

Approval
Approval seal 6
Viewing the metrological version 6

Reset
Factory configuration reset 36
Complete memory reset 43

3
TECH_MAN_ENG_DFWLPLUS
1. INTRODUCTION AND WARNINGS

This product is the best solution for multi-function weighing applications, offering ease of use, high precision in reading the weight,
and many functions to speed up and simplify everyday work.

This manual provides an overview of the potentials of the product. The configuration menu can be used to adapt the product
functionality to the required weighing application.

WARNINGS:
• Do not make repairs or replace electronic components of the instrument boards.
• Only use original spare parts.
• Any tampering with the equipment or use of non-original spare parts voids the warranty and relieves the
manufacturer of any liability.
• Before any installation or repair that involves access to electronic parts, turn off the device and disconnect any source
of power supply (battery, 230V network or other).
• Always use network power supply sources regulated within ± 10% of the rated voltage.
• In applications in connection with third parties, always follow the specifications given on the approval decree of the
equipment.
• Do not immerse in water.
• Do not wash with water jets (except versions with specific IP protection degree).
• Protect from direct rainfall (except versions with specific IP protection degree).
• Do not use aggressive cleaning solvents or substances.
• Do not install in potentially explosive environments.
• Earth connect any earth socket located on the equipment casing, using a cable with a diameter of at least 16 mm2.

4
TECH_MAN_ENG_DFWLPLUS
2. TECHNICAL FEATURES

DFWL PLUS DFWLI PLUS

Case ABS AISI 304 stainless steel

Backlit LCD
Screen
digit height = 25 mm

Brightness 5 intensity levels

Keypad 5 keys, waterproof

IP protection rating - IP68

Standard 1 PG9, plastic 2 PG9, steel


Cable gland
Extra - 1 PG9, steel

Load cell inputs 4

Number of scales 1

Maximum number of connectable load


8 x 350 Ω
cells

Maximum input current to load cells 120 mA

Internal use -10°C /+40°C


Temperature range
Type-approved -10°C /+40°C
Battery 4 x AA -
Battery duration up to 40 h -

Power supply External Internal


Power supply IN: 110/240 Vac IN: 88/264 Vac
OUT: 12 Vdc OUT: 12 Vdc
Available plugs EU, AU, UK, US EU, AU, UK, US, CH
RS232 1 (RJ11 connector)
RS232 + CTS 1 (internal, not usable) 1 (internal)
RS485 1 (internal, not usable) 1 (internal, optional)
Serial ports
TTL 1 (internal, not usable) 1 (internal)
Sensor 1 (internal, not usable) 1 (internal)
USB 1 Mini-USB (internal, for manufacturer use only)
Internal use from 100 to 800,000
Number of divisions
Type-approved 10,000 / 3 x 3,000
Max. number 6
Digital outputs (DFIO, 48 Vac or 60 Vdc
-
optional) Features 15 mA
10 Ω Max
Max. number 2
Digital inputs (DFIO,
- 12 / 24 Vdc
optional) Features
5 / 20 mA
Analogue output Voltage 0/5V 0 / 10 V -5 / 5 V -10 / 10 V
(DAC16OSER, -
optional) Current 0 / 20 mA 0 / 20 mA

5
TECH_MAN_ENG_DFWLPLUS
3. APPROVAL
a. b. c.

d. e.

How to display the metrological version of the instrument

1. Turn off the scale 2. Follow the procedure:

C Turn on the scale and press the key while the startup messages
are being displayed.
0.0
888888
2 sec C

xx.xx Metrological version

xx.xx.xx Weighing program version


-off- max

0.0 0.0

6
TECH_MAN_ENG_DFWLPLUS
4. CONNECTIONS

Single-channel

Cell 0.0
Connect the scale to the main terminal block using the first reading channel
of the A/D converter.

Cell 4 Cell 1

Junction 0.0
box

Cell 3 Cell 2

Reference terminal block for 1-channel connection

CELL3 CELL4
NOTES:
SIG+ • For 6-wire connection with “Sense”, set the dip switches to OFF.
• For 4-wire connection, set the dip switches to ON.
EXC+

EXC+
EXC-

EXC-
SIG+

SIG+
SIG-

SIG-

SIG-

SEN+
CELL2
SEN-
WARNING:
EXC+

EXC-

SIG+

SIG-

EXC+
COM4 - TTL
Make connections with indicator off and power supply disconnected.
SW2
EXC-
SENSE
+5V GND0 10 2
Comply with the electronic specifications indicated in the table on page
CELL1 ON 4
SENSOR

SENSOR

ON ON

SEN+ EXC+ SEN- EXC-

7
TECH_MAN_ENG_DFWLPLUS
0.0 Multi-channel with digital equalisation

The 4 channels of the converter can be used to connect 2, 3 or 4 cells, digital-


ly equalising them without using junction boxes.
Cell 4 Cell 1

Cell 3 Cell 2

Reference terminal blocks for 4-channel connection

CELL3 CELL4
NOTES:
SIG+ • Set the dip switches to ON
EXC+

EXC+
EXC-

EXC-
SIG+

SIG+
SIG-

SIG-

SIG-

SEN+
CELL2
SEN-
EXC+

EXC-

SIG+

SIG-

EXC+
COM4 - TTL

EXC- SW2 +5V GND0 10 2


SENSE
CELL1 ON
SENSOR

SENSOR

ON ON

SEN+ EXC+ SEN- EXC-

8
TECH_MAN_ENG_DFWLPLUS
5. PROGRAMMING

How to access the programming menu

1. Turn off the scale 2. Follow the procedure:

C Press while the power messages are displayed.

0.0
888888
2 sec C

-off- max

0.0

How to save the programming and exit the menu

To save the programming changes made, press button C repeatedly while navigating the menu backwards, until the message
SAVE? appears: press to save or C to exit without saving.

Example (read from right to left):

Save and exit

888888 save? C  C ADVANC C MoDE


C
max Cancel and exit

0.0

9
TECH_MAN_ENG_DFWLPLUS
PROGRAMMING MENU

Cal Calibration 14

 Zeroing the pre-tare (zero calibration) 15

 Area of gravity of the place of use 15

 Configuration of the serial ports 16

 Print customisation 23

 Weighing filter 31

 Adjusting the display 32

 Battery use 33

C Energy saving 33

 Auto switch-off 34

o Using the remote control 34

 Analogue output 35

 Digital inputs 37

o Digital outputs 38

 Factory configuration reset 39

 Diagnostics 39

 Advanced 40

Parameter visible only under certain conditions. Parameter or menu subject to approval.
10
TECH_MAN_ENG_DFWLPLUS
MENU
CAL Quick calibration
How to enter How to How to save
browse and exit
1. Off = Start of the calibration procedure:
2. On =
3. =
A-1 div Set the division and press

i Page 9 = C i Page 9 0.001


How to set the value

A
Cal ...
B 1
 div
C

D
 200
E

F

G
 Set the capacity
H CapaC 000.000
 and press

I How to set the value


C
J

or
K
o 0 ... 9

L
 unload
M kg

N
o -ok-
kg kg
O

Enter the calibration
P
 000000 weight and press
Q
 How to set the value

or
0 ... 9
kg

Load kg

-oK-

unload
kg

Cal.ok kg kg
If an advanced calibration was performed
previously, step Cal will lead directly to
steps Q-1 Q-2 Q-3 .
0.0 (See page 37)

11
TECH_MAN_ENG_DFWLPLUS
MENU
0.Cal Zeroing the pre-tare
How to enter How to How to save
browse and exit
1. Off = Acquisition of the zero point
2. On =
3. =
B 0Cal sure? -ok-
i Page 9 = C i Page 9 Grav
A
Cal
B
Cal
0
C

Grav
D
 Grav Area of gravity of the place of use
E

F Once the calibration is completed, for proper operation set the area of use in
 this pitch (if different from that of calibration).
G

H
C Grav 9.80390 Area of gravity
 (9.75001...9.84999)
I SeriaL
C How to set the value

J

or
K
o 0 ... 9
L

M

N
o
O

P

Q


12
TECH_MAN_ENG_DFWLPLUS
MENU
serial Configuration of the serial ports
How to enter How to How to save
browse and exit
1. Off =
Com.PC Communication with PC, PLC or Repeater
2. On =
3. =
Selection of the communication mode
i Page 9 = C i Page 9
D-1-1 mode
A
Cal
B
 On demand (*)

C

D

serial  On demand with code 485
(0...99)
E 1 00
 CompC
2
F
 Comprn 1
mode  Continuous transmission
2 (8 tx /sec)
G
 3
Aux1 ProtoC
H 4 3
 Aux2 Comsel
Automatic stability trans-
4 
I
C 5
usb baud mission
J 6 5
 advanC bit
K
o  Transmission at the pres-
sure of
L

M

N
  For Dini Argeo repeater
o
O

P
 modbus See dedicated manual

Q


Com.SEL

Selection of the protocol:

D-1-2 ProtoC Visible only if


Mode = Contin,
extend Stable, PRINT.K

short
Custom

Com.SEL
For available protocols see page 44.

13
TECH_MAN_ENG_DFWLPLUS
MENU COM port selection for PC / PLC connection

How to enter How to


browse
How to save
and exit
D-1-3 Com.SEL
1. Off =
2. On

=
3. =
i Page 9 = C i Page 9
...
A
Cal
B
 4
C

D

serial baud
E 1
 CompC
2
F
 Comprn 1
mode Communication speed (Baud rate)
2
G
 3
Aux1 ProtoC D-1-4 baud
H 4 3
 Aux2 Comsel 1200
4
I
C 5
usb baud
J 6 5
 advanC bit ...
K
o
L

9600
M

N
o
O
115200

P

Q
 bit

Configuration of the serial protocol

D-1-5 bit
n.8.1

n.8.2

...

e.7.2

mode
14
TECH_MAN_ENG_DFWLPLUS
MENU
Com.Prn Communication with printer or repeater or PC
How to enter How to How to save
browse and exit
1. Off = Selection of the communication mode
2. On
3.
= D-2-1 mode
=
tiCket Thermal printer
i Page 9 = C i Page 9

A
Cal label Labeller
B

C
 Continuous transmission
 (8 tx/sec)
D

serial
 Automatic stability trans-
E 1
 CompC mission
2
F
 Comprn   Manual transmission of
G 3 1 string for PC when press-
 Aux1 Mode ing
2
H
 4
Aux2 protoC e. For Dini Argeo repeater
I 5 3
C usb baud
J 6 4
 advanC bit protoC
K 5
o Cts
Selection of the protocol:
L 6
 powerp
M
D-2-2 protoC Visible only if
 Mode = Contin or
N extend Stable
o
O short

P
 Custom
Q
 baud
For available protocols see page 44.

Communication speed (Baud rate)

D-2-3 BAUD
1200

...

9600

115200

BIT
15
TECH_MAN_ENG_DFWLPLUS
MENU
Configuration of the serial protocol
How to enter How to
browse
How to save
and exit
D-2-4 bit
1. Off =
2. On =
n.8.1
3. =
i Page 9 = C i Page 9
n.8.2
A
Cal
B
 ...
C

D

serial e.7.2
E 1
 CompC
2
F
 Comprn
G
 3
Aux1 1
Mode Cts
2
H
 4
Aux2 protoC Printer control signal
I 5 3
C usb baud D-2-5 Cts
J
 6
advanC 4
bit no
K 5
o Cts
L 6 
 powerp
M
 
N
o
O
  Free adjustment
P
 nChar 000 Characters
Q sent
 (0...999)

time 000
Sending timeout (0...9999)

powerp

Printer power supply / Radio-frequency module

D-2-6 powerp
no

For use by the



manufacturer

Printer power supply from


indicator via Vaux connec-
pwr.int tor.

See wiring diagram


(see page 50).
Mode
16
TECH_MAN_ENG_DFWLPLUS
MENU
aux.1 - aux.2 Auxiliary ports
How to enter How to How to save
browse and exit
1. Off = Selection of the communication mode
2. On D-3-1

3.
=
D-4-1 Mode
=
i Page 9 = C i Page 9
onde On demand (*)

A
Cal 485 On demand with code 485
B (0...99)

C
  Continuous transmission
(8 tx /sec)
D

serial
Automatic stability
E 1 
 CompC transmission
2
F
 Comprn Transmission at the
print.k
G 3 pressure of
 Aux1
H 4
 Aux2 1
Mode repe.6 For Dini Argeo repeater
2
I
C 5
usb protoC
J 6
 advanC 3
baud protoC
K 4
o bit
Selection of the protocol:
L
 D-3-2
M D-4-2 protoC

N extend
o
O short

P
 Custom
Q
 baud
For available protocols see page 44.

Communication speed (Baud rate)


D-3-3
D-4-3 baud
1200

...

9600

115200

bit
17
TECH_MAN_ENG_DFWLPLUS
MENU
Configuration of the serial protocol
D-3-4
How to enter How to
browse
How to save
and exit D-4-4 bit
1. Off = n.8.1
2. On =
3. =
i Page 9 = C i Page 9
n.8.2

A ...
Cal
B

C e.7.2

D

serial
E 1 mode
 CompC
2
F
 Comprn
G 3 usb USB port
 Aux1
H 4
 Aux2 1
Mode For manufacturer use only.
2
I
C 5
usb protoC
J 6
 advanC 3
baud
K 4
o bit
L

M

N
o
O

P

Q


18
TECH_MAN_ENG_DFWLPLUS
MENU
AdvanC Advanced configurations
How to enter How to How to save
browse and exit Connection port of radio-frequency module (for use by the manufacturer)
1. Off =
2. On = D-6-1 radio
3. =
=
off
i Page 9 C i Page 9

A Refer to the radio


Cal module manual for CoM1
B selecting the radio
 channel (QSG_ ...
C 0...37
* * Selection
of the radio
RF2G4).
C
 CoM4
channel.

D

serial
E 1
 CompC
2
ttltil
F
 Comprn
G 3 TTL port / inclinometer activation (for use by the manufacturer)
 Aux1
H 4
D-6-2 ttltil Only visible on
 Aux2 DFWL models.
I 5
no
C usb
J 6
 advanC
K
yes
o 1
radio
L
 2
ttltil
M
 3
term term
N
o 4
iGnore Closing character of each print line

O
 D-6-3 term
P
 Cr
Q

Crlf

iGnore
Ignore unknown commands

D-6-4 iGnore When an unknown


command is sent:
No - Selecting NO
will result in the
response “ERR04”.
- Selecting
yes YES ignores
the command
(no response).

radio

19
TECH_MAN_ENG_DFWLPLUS
MENU
layout Print customisation
How to enter How to How to save
browse and exit
1. Off = Parameters for ticket/label mode
2. On =

HEADER
3. =
i Page 9 = C i Page 9
header HEADER
A
HEADER
Cal HEADER
B

 WEIGHS N. 1
C
  TICKET N. 15
D
 GROSS
E

layout  TARE
F 1 NET
 lanG
2
G
 Char C 24/12/2015 10:30
H 3
 header 
4 
I
C data

J 5
 weiGhs


K 6
o tiCket 
7
L
 CloCk
M 8
 barC.39
N 9
o barC.up Additional parameters for label mode
O 10
 barC.l
P 11
 barC.h
L.MArG
12
Q
 barC.dt
13
Copies HEADER
Up.MArG

14
end.tiC HEADER
15
HEADER
b.line HEADER
16 ...
label
17
lb.save
18
test



20
TECH_MAN_ENG_DFWLPLUS
MENU Print language settings

How to enter How to


browse
How to save
and exit
E-1 lanG ital
1. Off =
EnGl enGl
2. On =
deut
3. =
fran
i Page 9 = C i Page 9 ...
espa
A Chines
Cal port
B Grek
 Grek
C

D
Char

E
layout
 Font dimensions
F 1
 lanG E-2-1 Char1 Label mode Ticket mode

2
G
 Char Main font

H 3 font.1 normal
 header 1
Char1 1 x 1,5 mm

4 2
I
C data Char2
J 5 font.1d double
 weiGhs 1 x 3 mm

K 6
o tiCket
7 font.2
L
 CloCk 1,5 x 2,5 mm

M 8
 barC.39
N
o 9
barC.up font.2d
1,5 x 5 mm

O 10
 barC.l
P
 11
barC.h font.3
2 x 3 mm
12
Q
 barC.dt
13 font.3d
Copies 2 x 6 mm
14
end.tiC
15
b.line font.4
3 x 4 mm
16
label
17
lb.save font.4d
3 x 8 mm
18
test
font.5
4 x 6 mm

font.5d
4 x 12 mm

Char2
E-2-2 Char2 See Char 1

21
TECH_MAN_ENG_DFWLPLUS
MENU Enables header printing

How to enter How to


browse
How to save
and exit
E-3-1 enable
no Disabled
1. Off =
2. On =
3. =
yes Only 1st weight totalisation
i Page 9 = C i Page 9

A
Cal total Only upon each weight
B
totalisation

C
 always Also in total
D

E

layout line1
F 1
 lanG Contents of the header lines
2
G
 Char E-3-2 line1 First row of header
H
 3
header No Visible only on line 2, line 3 and
line 4
4 1
I
C data enable
J 5 2 Char1
 weiGhs line1
K 6 3
o tiCket line2 Char 2
7 4
L
 CloCk line3
M 8 5
 barC.39 line4 01 032 032*
N 9 Enter
o barC.up the first
O 10 character
 barC.l 02 032 032* Enter the
P 11
 barC.h second
12
character
Q
 barC.dt ...
13 032*
Copies
14
end.tiC
15
b.line 24 032 032* Enter
the last
16 character
label
17
lb.save line1
18
test How to set the value

or
0 ... 9

Repeat the operation to program line 2, line 3 and line 4. Select no


to disable them.

22
TECH_MAN_ENG_DFWLPLUS
MENU How to print / delete the row being programmed

How to enter How to How to save


browse and exit 01 032
1. Off =
2. On =
3. = 02 032 delete -ok-
Delete the header
row
i Page 9 = C i Page 9
Print the header
A
Print -ok- row
Cal 03 032
B

...
C

D
 Programming example

072 *
E
layout
 01 032 H
F 1 HELLO
 lanG
G

2
Char 02 032 069 * E
H 3
 header
I
C
4
data 1
enable 03 032 076 * L
J 5 2
 weiGhs line1
K 6 3 04 032 076 * L
o tiCket line2
7 4
L
 CloCk line3
M 8 5 05 032 079 * O
 barC.39 line4
How to set the value
N 9
o barC.up
O 10
06 032 032 * or
 barC.l 0 ... 9
P 11
 barC.h ...
Q

12
barC.dt 24 032 032 *
13
Copies
14 (*) List of characters
end.tiC
15 32 47 / 62 > 77 M 92 \ 107 k 122 z
b.line 33 ! 48 0 63 ? 78 N 93 ] 108 l 123 {
16
label 34 “ 49 1 64 @ 79 O 94 ^ 109 m 124 |
35 # 50 2 65 A 80 P 95 _ 110 n 125 }
17
lb.save 36 $ 51 3 66 B 81 Q 96 ‘ 111 o 126 ~
18 37 % 52 4 67 C 82 R 97 a 112 p
test 38 & 53 5 68 D 83 S 98 b 113 q
39 ‘ 54 6 69 E 84 T 99 c 114 r
40 ( 55 7 70 F 85 U 100 d 115 s
41 ) 56 8 71 G 86 V 101 e 116 t
42 * 57 9 72 H 87 W 102 f 117 u
43 + 58 : 73 I 88 X 103 g 118 v
44 , 59 ; 74 J 89 Y 104 h 119 w
45 - 60 < 75 K 90 Z 105 i 120 x
46 . 61 = 76 L 91 [ 106 j 121 y

23
TECH_MAN_ENG_DFWLPLUS
MENU Selection of the weight data

How to enter How to


browse
How to save
and exit
E-4 data
1. Off =
Gross
yes
2. On =
3. =
i Page 9 = C i Page 9
no

A
Cal tare
B
yes

C
 no
D

E net

layout yes
F 1
 lanG
2 no
G
 Char
H 3
 header weiGhs
4
I
C data Progressive weighed
J
 5
weiGhs E-5 weiGhs
K 6 no
o tiCket
7
L
 CloCk
yes On each weighing
M 8
 barC.39 operation
N 9
o barC.up
O 10 sum On total only
 barC.l
P 11
 barC.h
always Both on each weighing
12
Q
 barC.dt operation / weight
13 totalisation and total
Copies
14
end.tiC tiCket
15 Ticket/label progressive
b.line
16
label
E-6 tiCket
no
17
lb.save
18
test yes On each weighing
operation

SUM On total only

Both on each weighing


always operation / weight
totalisation and total

CloCk
24
TECH_MAN_ENG_DFWLPLUS
MENU Date and time

How to enter How to


browse
How to save
and exit
E-7 CloCk
1. Off = no
2. On =
3. =
On each weighing
i Page 9 = C i Page 9 yes
operation

A
Cal
B SUM On total only

C
 Both on each weighing
D always
 operation / weight
totalisation and total
E
layout

F
 1
lanG barC.39
2
G
 Char
H 3
 header Bar code 39
4
I
C data E-8 barC.39
J 5 no
 weiGhs
K 6
o tiCket
7 yes On each weighing
L
 CloCk operation
M 8
 barC.39
N 9 SUM On total only
o barC.up
O 10
 barC.l
P 11 always Both on each weighing
 barC.h operation / weight
12 totalisation and total
Q
 barC.dt
13
Copies barC.UP
14
end.tiC Barcode top margin (mm) Visible only if barC.39 (E-8)
15
b.line is active
If mode=tiCket (0...9)
16 E-9 barC.up 000
label Se mode=label (0...999)

17
lb.save barC.L
18
test Barcode left margin (mm) Visible only if barC.39 (E-8)
is active
If mode=tiCket (1...99)
E-10 barC.l 00 Se mode=label (0...99)

barC.H

Barcode height (mm) Visible only if barC.39 (E-8)


is active
If mode=tiCket (1...99)
E-11 barC.h 00 Se mode=label (0...255)

barC.DT
25
TECH_MAN_ENG_DFWLPLUS
MENU Selection of the weight data Visible only if barC.39 (E-8)
is active
How to enter How to
browse
How to save
and exit
E-12 barC.dt
1. Off =
Gross
2. On =
3. =
i Page 9 = C i Page 9
net
A
Cal
B
 tare
C

D

E Copies

layout
F 1
 lanG
2 Multi-copy prints
G
 Char
H
 3
header E-13 Copies n 1...3
I
C
4
data end.tiC
J 5
 weiGhs
K 6
o tiCket Paper outlet for end of ticket/receipt
7
L
 CloCk E-14 end.tiC
M
 8
barC.39 no
N 9
o barC.up
On each weighing
O
 10
barC.l yes operation
P 11
 barC.h
12 On total only
Q
 barC.dt SUM
13
Copies
Both on each weighing
14
end.tiC always
operation / weight
15 totalisation and total
b.line
16
label B.line
17
lb.save
18
test White print head preheating line (thermal printer only)

E-15 b.line
no

yes

lABEL
26
TECH_MAN_ENG_DFWLPLUS
MENU Label configuration Only visible if mode (D-2-1) = Label

How to enter How to


browse
How to save
and exit
E-16 label
1. Off =
2. On
width 063 Size (w)
= (0...999)
3. =
i Page 9 = C i Page 9
heiGht 080 Size (h)
(0...999)

A
Cal up.marG 010 Upper margin
B (mm) - (0...999)

C
 l.MarG 06 Left margin
D (mm) - (0...99)

E
layout
 peel
F 1 yes Peeler
 lanG
2
G
 Char
H 3
 header no
4
I
C data
J 5
 weiGhs Gap 3 Gap between labels
(mm) - (0...7)
K 6
o tiCket
7
L
 CloCk
M 8
 barC.39 lb.save
N 9
o barC.up
O 10
 barC.l Saving labels in printer memory (only for label mode)
P 11
 barC.H Visible only if
12
E-17 lb.save L.SAVE? mode (D-2-1) = Label
Q
 barC.dt
13 TEST
Copies
14
end.tiC Test print ALL FORMATS
15
b.line E-18 test
16
label LANG
17
lb.save
18
test

27
TECH_MAN_ENG_DFWLPLUS
MENU
filter Weighing filters
How to enter How to How to save
browse and exit
1. Off = To change the responsiveness of the scale.
2. On This is useful to adjust the scale according to your needs.
=
3. = With an approved instrument, it is not possible to change the filter.

i Page 9 = C i Page 9

A Foreword:
Cal “F 10” represents the lowest filtering incidence.
B By increasing the incidence, the weight becomes more stable.
 It is advisable to carry out several weighing operations by changing the
C
 incidence until the best compromise between responsiveness and stability
is achieved.
D

E

F F-1 f 1 Filter at 5 Hz.

filter
G 1
 f 1
2
F-2 f 2 Filter at 10 Hz.
H
 f 2
I 3
C f 3 F-3 f 3 Filter at 20 Hz.
J 4
 f 4
K
o 5
f 5 F-4 f 4 Filter at 40 Hz.

L
 6
f 6
7
F-5 f 5 Filter at 80 Hz.
M
 f 7
N
o 8
f 8 F-6 f 6 Filter at 160 Hz.
O 9
 f 9
P
 10
f 10 F-7 f 7 Filter at 325 Hz.

Q 11
 Custom Only visible if
F-8 f 8 Filter at 650 Hz. N.CHAN < 3

Only visible if
F-9 f 9 Filter at 1300 Hz.
N.CHAN < 2

Only visible if
F-10 f 10 Filter at 2600 Hz.
N.CHAN < 2

F-11 Custom For use by the manufacturer.

28
TECH_MAN_ENG_DFWLPLUS
MENU
sCreen Adjusting the display
How to enter How to How to save
browse and exit
1. Off = Backlighting
2. On = G-1 bak.lit
3. =
=
no
i Page 9 C i Page 9

A
Cal yes Always on
B

C
 auto On when weight is
D unstable

E
 briGht
F
 Brightness
G
sCreen
 G-2 briGht
H
 1
bak.lit brGt 1
2
I
C briGht
J
 3
loCk
K 4
o Colour
L
 brGt 5
M

N
o loCk
O
 Display lock (for use by the manufacturer)

P
 G-3 loCk
Q no


yes

Colour
Backlight colour

G-4 Colour Only visible in


version with colour
White display.

Green

bak.lit
29
TECH_MAN_ENG_DFWLPLUS
MENU
batt Battery power supply
How to enter How to How to save
browse and exit
1. Off =
2. On =
H batt
3. = no
i Page 9 = C i Page 9

A
Cal aa 4 AA Batteries

B

C
 pb Lead rechargeable battery
D

E
  niMh rechargeable
F battery

G
 pb.pw For manufacturer use only.
H
batt

I
eco.bat
C niMh.pw For manufacturer use only.
J

K
o
L eco.bat

M
 WARNING:
N use only original rechargeable batteries.
o
O

P
 eCo.bat Energy saving for battery operation
Q

Visible only if batt (H)
is active

I eco.bat
ECo.no Maximum consumption,
maximum reactivity

Reduced reactivity,
eCo.min reduced consumption

Reduced reactivity,
ECo.med reduced consumption

Minimum consumption.
ECo.max Scale in standby and pow-
er at the touch of a button
and automatic switch-off
at the end of weighing.
autoff

30
TECH_MAN_ENG_DFWLPLUS
MENU
autoff Auto switch-off
How to enter How to How to save
browse and exit
1. Off =
2. On =
J autoff
3. = no
i Page 9 = C i Page 9

A
Cal yes M 001
How to set the value
B

C
 or
D 0 ... 9

E
 remote
F

G
 remote Remote control
H

I
C K remote
J
autoff
 no
K
remote
o 6-Key radio-frequency
remote control. All keys
L
 perform the tare.
M
 

N 6-Key radio-frequency
o remote control. Multi-
O
  6 function mode.

P
 6-Key radio-frequency
Q remote control in
   broadcast mode. All keys
perform the tare.

  6-Key radio-frequency


remote control in
broadcast mode. Multi-
function mode.

an.out

The broadcast mode allows sending the control to multiple scales


simultaneously.

31
TECH_MAN_ENG_DFWLPLUS
MENU
An.out Analogue output
How to enter How to How to save
browse and exit
1. Off = Only visible when the optional analogue board is present.
2. On =
Selecting the COM port
3. =
i Page 9 = C i Page 9
L-1 Com
none
A
none Com.pC
Cal ... Com.prn
B
 aux.1
aux.2
C aux.2

D
 net
E
 Operation proportional to net/gross weight
F
 L-2 net
G
 yes Proportional to net weight
H

I
C no Proportional to gross
J weight

K
o pnt.und
L

An.out
M 1
L-3 pnt.und 00000 Analogue value for weight in
 Com underload (0...65535)
N 2
o net
O 3
L-4 wGt.1 000.000 First linearisation weight val-
 pnt.und ue (-99999...999999)
P 4
 wGt.1
Q
 5
pnt.1
L-5 pnt.1 01300 Analogue value related to
wGt.1 (0...65535)
6
wGt.2
Second weight value
7
pnt.2
L-6 wGt.2 000.000
of linearisation
8 (-99999...999999)
wGt.3
9
pnt.3 L-7 pnt.2 63250 Analogue value related to
wGt.2 (0...65535)
10
pnt.ovr Third weight value
L-8 wGt.3 000.000 of linearisation
(not mandatory)
(-99999...999999)

L-9 pnt.3 00000 Analogue value related to


wGt.3 (not mandatory) (
0...65535)

L-10 pnt.ovr 65535 Analogue value relative


to weight in “overload”
(0...65535)

Using the real time update of the output, it is possible to check the
entered value with a tester (see example on page 33).
32
TECH_MAN_ENG_DFWLPLUS
MENU Programming example:

How to enter How to How to save the analogue output provides 10V to program a linearisation point so that at
browse and exit 1000kg.
1. Off =
2. On =
3. =
L-4 wGt.1 001000 Programming 1000kg
i Page 9 = C i Page 9

A
Cal
B
L-5 pnt.1

C

D
 Enter 62300 (the reference value given in the table) and check the analogue
E output using a tester.

F

G i The output is updated in real time.
 9.98
H

I
C
J

K
o
L
An.out


A
m
M 1
 Com

t l
Vo
N 2
o net
O 3
 pnt.und Adjust the analogue output by increasing or decreasing the value.
P 4 Minimum variations of at least 10 points are recommended, (62310, 62320,
 wGt.1 62330, etc.)
Q 5
 pnt.1 How to set the value
6
wGt.2
7 or
pnt.2 0
...
9
8
wGt.3
9 The output is updated in real time.
pnt.3 10.00
10
pnt.ovr
A
m
tl
Vo

Once the desired adjustment has been made, confirm the value with .

33
TECH_MAN_ENG_DFWLPLUS
V / mA
MENU
How to enter How to How to save
browse and exit pnt.ovr
1. Off =
2. On = pnt.3
3. =
i Page 9 = i Page 9
C
pnt.2
A
Cal pnt.1
B

C

0
D
 pnt.und

Max
Max + 9d
kg

kg
E

F

G
 Value to be entered Output volts mA output

H
1200 ~0V ~ 0 mA
 11250 ~ 4 mA
I
C 52200 ~ 20 mA
J 62300 ~ 10 V

K
o
L

M inputs Digital inputs

inputs
N
o 1
inp.b.1 Input configuration 1
O 2
 inp.b.2 M-1 inp.b.1
P
 no Disabled
Q

zero Repeat key

tare Repeat key

mode Repeat key

print Repeat key

loCk.kb Keypad deactivation

inp.b.2

Repeat the same operation for inp.b.2.

34
TECH_MAN_ENG_DFWLPLUS
MENU
output Digital outputs
How to enter How to How to save
browse and exit
1. Off = Operation on net weight, gross weight or batching
2. On =
3. =
N-1-1 funCt For dosing / filling:

i Page 9 = C i Page 9
no • Activate mode dosaGe
• Set tare unloCk.

A The output is activated


Cal Gross only after the container has
B been calibrated (by key
 or external button) and is
deactivated when the set
C
 net target (setpoint) is reached.
D To carry out two-speed
 filling, two outputs must be
programmed with dosaGe
E
 dosaGe mode.
F

G
mode
 Normally open operation (n.o.) or closed (n.c.)
H
 N-1-2 mode
I
C n.o.
J

K
o n.C.
L

M
 direCt
N
output
o Output activation mode
O 1
 rel.b.1 N-1-3 direCt
P 2
 rel.b.2 1
funCt yes Direct, stable or unstable
weight
Q 3 2
 rel.b.3 mode
4
rel.b.4 3
direCt no Only at stable weight
5 4
rel.b.5 hister
6
rel.b.6 hister
Double threshold operation
(activation weight threshold ≠ from output deactivation weight threshold)

N-1-4 hister Operation:


YES (Double threshold)
yes ON
OFF
ON
OFF

no no (Single threshold)
ON
OFF
ON
OFF

funCt
Repeat the same operation for rel.b.2, rel.b.3, rel.b.4, rel.b.5
and rel.b.6.
35
TECH_MAN_ENG_DFWLPLUS
MENU
RESET Factory configuration reset
How to enter How to How to save
browse and exit
1. Off = Function for restoring default
O reset sure?
2. On = configurations keeping the
stored calibration.
3. = diaG
i Page 9 = C i Page 9

A diaG Diagnostics
Cal
B
 Converter. Check of input signal in µV.

C
P-1 adC.uv In case of multiple equalised channels, press
 buttons or to examine all the channels
selected.
D

E

P-2 displa Display. Integrity check of all segments and icons.

F
 Keypad. Press any key to check its correct
G
P-3 keyb operation, with beep and code on display.

H
 CTS. Check of status of the control signal from the
I
P-4 Cts printer.
C
J Optional digital outputs. Check the activation and
 P-5 output deactivation of each contact.
K Example:  activates output 1. Press button
o to select the next output.
L
 WARNING: before entering the step 
M check that the activation of the output does not
 cause conditions of danger to persons, animals or
N property.
o
Optional digital inputs. Check the activation and
O

reset P-6 inputs deactivation of each input.
P Example:  input not active

diaG Example:  active input
Q 1 Press button to select the next input.
 adC.uv
2
displa Analogue output. Enter the digital value and check
3
P-7 an.out the analogue output response with a tester.
keyb
4
Cts P-8 ser.num Scale serial number.
5
output
Reviewing hardware (e.g.  ) followed by
6
inputs P-9 prG.ver software version (e.g. ).
7
an.out
8
P-10 div.int Number of internal divisions.
ser.num
Converter. Checking signal in ADC points.
9
prG.ver P-11 adC.pnt In case of multiple equalised channels, press
buttons or to examine all the channels
10
div.int selected.
11
adC.pnt
12
P-12 bt.adC Battery voltage value.
bt.adC
13
P-13 pw.adC Mains power supply voltage value.
pw.adC
14
P-14 serial Scale serial number.
serial 36
TECH_MAN_ENG_DFWLPLUS
MENU
AdvanC Advanced
How to enter How to How to save
browse and exit
1. Off = Calibration parameters
CAL.PAR
2. On =
3. =
Configuration of the decimal point (0...3)
i Page 9 = C i Page 9
Q-1-1 deCim
A
Cal 0
B

C
 0.0
D

E 0.00

F

G
0.000

H

I div
C
J Reading division

K
o
Q-1-2 div
L
1

M
 2
N
o
O

P

Q
advanC
 200
1
Cal.par
2 1
equal.p deCim
3 2 u.m.
Cal.adv div
4 3 Unit of measure
stb.flt u.m.
5
peak.fl 4
ranGe1 Q-1-3 u.m.
6 5 G
metrol ranGe2
7 6
keyb ranGe3 kG
8 7
tilt equal
9 8
reaCt n.Chan t
10
loCk.kb
11
alibi.r lb
12
pin.teC
13
pin.use
14 ranGe1
dflt.t 37
TECH_MAN_ENG_DFWLPLUS
MENU Scale capacity. Set Max or Range 1 (Max range = 800.000)

How to enter How to


browse
How to save
and exit
Q-1-4 ranGe1 003.000 How to set the value
1. Off =
2. On =
ranGe2 or
3. = 0 ... 9

i Page 9 = C i Page 9
Range 2
A For multirange scales, set the second weighing range.
Cal
B Q-1-5 ranGe2 000.000 How to set the value

C ranGe3
 or
...
0 9
D

E Range 3
 For multirange scales, set the third weighing range.
F
 Q-1-6 ranGe3 000.000 How to set the value
G
 equal
H or
 0
...
9
I
C
J Example of multirange configuration at 1500/3000 kg, division 0,5/1 kg.
 Set: deCi = 0.0
K
o div = 5
L ranGe 1 = 1500.0

ranGe 2 = 3000.0
M

N
o Equalisation function

O
 Q-1-7 equal
P
no

Q

advanC yes
1
Cal.par
2 1
equal.p deCim
3 2 n.Chan
Cal.adv div
4 3 Connection diagram on page 8.
stb.flt u.m. Equalisation procedure on page 39.
5 4
peak.fl ranGe1
6 5 Only visible if Equal(Q-1-7)= Yes
metrol ranGe2 Equalised analogue channels

7
keyb 6
ranGe3 Q-1-8 n.Chan
8
tilt 7
equal Ch 2
9 8
reaCt n.Chan
10
loCk.kb Ch 3
11
alibi.r Ch 4
12
pin.teC
13
pin.use deCim
14
dflt.t 38
TECH_MAN_ENG_DFWLPLUS
MENU
equal.p Equalisation
How to enter How to How to save
browse and exit
1. Off = .p is only visible if the function l (Q-1-7) is activated in the
2. On =
menu  (Q-1).
3. =
The equalisation wizard asks to acquire the zero point with scale unloaded
i Page 9 = C i Page 9 and to later place a weight of about 1/8 of the maximum capacity (Max) on
each individual cell in the required order. The message oK will appear
A after the procedure.
Cal
B Proceed with the calibration.

C

D
 0.0
E
 0.0

F
 Cell 1 Cell 4
G Cell 4 Cell 1

H

I
C
J
 Cell 2 Cell 3
K Cell 3 Cell 2
o
L

M

N
o Q-2-1 eq.0
O
 -ok-
P
eq.1

Q
advanC

1
Cal.par Q-2-2 eq.1 kg

-ok-
2
equal.p 1
eq.0 eq.2
3 2 = 1/8 Max (kg)
Cal.adv eq.1
kg

4 3
stb.flt eq.2 Q-2-3 eq.2
5
peak.fl 4
eq.3 -ok-
5
eq.3 kg
6
metrol eq.4
7
keyb
8
tilt Q-2-4 eq.3
9 -ok-
reaCt eq.4 kg

10
loCk.kb
11
alibi.r
12
pin.teC
Q-2-5 eq.4 kg

-ok-
13
pin.use Cal.adv
14
dflt.t 39
TECH_MAN_ENG_DFWLPLUS
MENU
CAL.adv Complete calibration
How to enter How to How to save
browse and exit
1. Off = Before calibrating, configure the decimals (deCi - Q-1-1), the division
(div - Q-1-2) and the capacity (ranGe - Q-1-4,5,6).
2. On =
3. = Start of the calibration procedure:
= Unload the pan, press and wait for
i Page 9 C i Page 9 Q-3-1 zero the message -ok-.
A
Cal kg kg
B -ok-

C
 kg kg
Now acquire the calibration points (up
D
 Cal.pnt to 3):
E
 n 1 Ι. Set the calibration points
F (1...3)

G How to set the value

H

I
C
000000 ΙΙ. Enter the calibration
J
 weight
K
o How to set the value

L

or
M
 kg kg 0 ... 9

N
o
O kg kg ΙΙΙ. Load the weight and
 wait
P

Q

advanC -oK-
1
Cal.par
2 000000 Repeat the steps from ΙΙ for
equal.p the next points
3
Cal.adv
4
stb.flt 1
zero ...

5
peak.fl CAL.ok
6
metrol CAL.PNt
7
keyb
8
tilt
9
reaCt stb.flt Additional stability filter
10
loCk.kb
11
alibi.r
12
pin.teC
Q-4 stb.flt For use by the manufacturer.

13
pin.use
14
dflt.t 40
TECH_MAN_ENG_DFWLPLUS
MENU
peak.fl Anti-peak filter
How to enter How to How to save
browse and exit
1. Off =
2. On =
Q-4 peak.fl For use by the manufacturer.

3. =
i Page 9 = C i Page 9

A
Cal metrol Metrological parameters

B
 Resetting percentage using the key
C
 02 With approved scale (0...2)
D
Q-5-1 0.perC With non-approved scale

E
div.stb (0...50)


F

Sensitivity of the weight stability control
G
 (0...99)
H
Q-5-2 div.stb n 02

I
0.trk
C
J

Stability detection time
K
o With approved scale (500)
L
Q-5-3 tim.stb 500 With non-approved scale
 (10...10000)
M 0.trk

N
o
Zero hold function (tracking)
O

P
Q-5-4 0.trk
 tr no
Q
advanC

1
Cal.par tr 1/4
2
equal.p
3
Cal.adv tr 1/2
4
stb.flt
5 ...
peak.fl
6
metrol 1
0.perC
7
keyb 2
div.stb tr 10
8
tilt 3
tim.stb
9
reaCt 4
0.trk 0.trk.sp
10
loCk.kb 5
0.trk.sp
11
alibi.r 6
on.zero Zero tracking speed
12 With approved scale
pin.teC 7
Cal.adj Q-5-3 0.trk.sp 1000 (1000)
13 8 With non-approved scale
pin.use Cal.man on.zero (100...5000)
14 9
dflt.t d.sale 41
TECH_MAN_ENG_DFWLPLUS
MENU Reset at power and reset percentage

How to enter How to


browse
How to save
and exit
Q-5-6 on.zero
1. Off =
no
2. On =
3. =
=
i Page 9 C i Page 9 yes
A
Cal C.perC 10
B
 Cal.adj
C
 Re-acquisition / change of the calibration points in memory.
D
 Q-5-7 Cal.adj
E Change of first calibration
 zero point
F I. Enter the
 point.1 000000 calibration
G
 point.2 weight
(0...999999)
H
 point.3 load
I II. Load the
C weight on
J the scale and
 wait
K
o cal.ok
L
 Cal.man
M Repeat the same operation for Point 1, Point 2 and point 3

N
o Q-5-8 Cal.man For use by the manufacturer.

O
 Q-5-9 d.sale For use by the manufacturer.

P

Type of keypad
Q

advanC Visible depending on model
1
Q-6 keyb
Cal.par norm
2
equal.p
3
Cal.adv
4 ext
stb.flt
5
peak.fl
6
metrol 1
0.perC tilt
7 Inclinometer (for use by the manufacturer)
keyb 2
div.stb
8
tilt 3
tim.stb Q-7 tilt
9
reaCt 4
0.trk n.o. Normally open
10
loCk.kb 5
0.trk.sp
11 Normally closed
alibi.r 6
on.zero n.C.
12
pin.teC 7
Cal.adj
13 8
pin.use Cal.man reaCt
14 9
dflt.t d.sale 42
TECH_MAN_ENG_DFWLPLUS
MENU Reactivation of the print or weight totalisation function

How to enter How to


browse
How to save
and exit
Q-8 reaCt
1. Off = After unloading the pan
2. On
zero
=
3. =
i Page 9 = C i Page 9 When weight is unstable
inst
A
Cal
B
always Function always active

C

D
 loCk.kb
E
 Permanent keypad lock (excluding key C )
F
 Q-9 loCk.kb
G
 no
H

I
C yes
J

K
o
L
alibi.r

M
 Reset of fiscal memory (alibi memory, optional)

N
o Q-10 alibi.r sure? Only visible if alibi
memory option is
O
 pin.teC present
Reset is not possible
P if instrument is
 certified
Q Access PIN to programming menu
advanC

1
Q-11 pin.teC
Cal.par no
2
equal.p
3
Cal.adv Enter the access PIN
4 yes
stb.flt How to set the value
5 123456
peak.fl
6
metrol pin.use or
7 0 ... 9
keyb
8
tilt Access PIN to user menus
9
reaCt Q-12 pin.use See Pin.teC
10
loCk.kb
11 Total reset of memory and of calibration, with reset of the factory settings.
alibi.r
12
pin.teC Q-13 dflt.t sure?
13
pin.use Cal.par
14
dflt.t 43
TECH_MAN_ENG_DFWLPLUS
6. COMMUNICATION STRINGS

Short string

01ST,GS, 0.0,kg<CR><LF>

where

01 Code 485 of the instrument (2 characters), only if communication mode 485 is enabled
ST Scale status (2 characters):
US - Unstable weight
ST - Stable weight
OL - Weight overload (out of range)
UL - Weight underload (out of range)
TL - Scale not level (inclinometer active)
, ASCII 044 character
GS Type of weight data (2 characters)
GS - Gross
NT - Net
, ASCII 044 character
0.0 Weight (8 characters including the decimal point)
, ASCII 044 character
kg Unit of measurement (2 characters)
<CR><LF> Transmission terminator, characters ASCII 013 and ASCII 010

Extended string

011, ST, 0.0,PT 20.8, 0,kg<CR><LF>

where

01 Code 485 of the instrument (2 characters), only if communication mode 485 is enabled
1 ASCII 049 character
, ASCII 044 character
ST Scale status (2 characters):
US - Unstable weight
ST - Stable weight
OL - Weight overload (out of range)
UL - Weight underload (out of range)
TL - Scale not level (inclinometer active)
, ASCII 044 character
0.0 Net weight (10 characters including the decimal point)
, ASCII 044 character
PT Indication of pre-set manual tare (2 characters)
20.8 Tare weight (10 characters including the decimal point)
, ASCII 044 character
0 Number of pieces (10 characters)
, ASCII 044 character
kg Unit of measurement (2 characters)
<CR><LF> Transmission terminator, characters ASCII 013 and ASCII 010

44
TECH_MAN_ENG_DFWLPLUS
Custom string

Foreword:

The configuration can only be carried out using DiniTools in the “CUSTOM STRING FORMATTING” section.
The maximum length of the string configuration field is 100 characters. Fixed alphanumeric characters or variable macros may be
used inside.
There are also “tokens” that define the characters that a variable (<>) will become in the string when the indicator is in weighing
mode.
The custom string can be composed using the macros on page 49

The default custom string setting is shown below:

<2><P><W7.><U><M><S><CR><LF>

where

<2> ASCII STX character (start of text)


<P> Weight polarity
<W7.> Weight (formatted to 7 digits with decimal point)
<U> Unit of measure
<M> Type of weight (gross, net, tare)
<S> Weight status
<CR><LF> Transmission terminator, characters ASCII 013 and ASCII 010

Default token:

Gross mode token “<M>” G


Net mode token “<M>” N
Tare mode token “<M>” T
Preset weight token “<M>” T
Unit token “<U>” K
Positive polarity token “<P>” SPACE
Negative polarity token “<P>” -
Invalid status token “<S>” I
Motion status token “<S>” M
Valid status token “<S>” SPACE
Coz status token “<S>” Z
Overload status token “<S>” O
Underload status token “<S>” O
Overload char token “<S>” &
Underload char token “<S>” =

The default command to be sent to the weight indicator to request the custom string is:

nnSF#1<CR><LF>

45
TECH_MAN_ENG_DFWLPLUS
Available Macro:

Macro Format Description Token value (standard)


Weight sign Sign of the displayed The string will show the character
<P>
weight set in tokens:
<PG> Sign of the gross weight
Positive polarity token “<P>” “SPACE” if the weight is positive
<PN> Sign of the net weight
Negative polarity token “<P>” “-” if the weight is negative
<PT> Sign of the tare
Unit of measure The string will show the character
<U> set in token:
Unit token “<U>” “K”, to indicate kg
Weight type <M> Displayed weight The string will show the character
set in tokens:
<MG> Gross weight
Gross mode token “<M>” “G“ if the weight is gross
<MN> Net weight
Net mode token “<M>” “N“ if the weight is net
Tare mode token “<M>” “T“ if the weight is a tare
<MT> Tare
Preset tare weight token “<M>” “T“ if the weight is a preset tare
Weight status The string will show the character
set in tokens:
Invalid status token “<S>” “I” if the weight is not valid
Motion status token “<S>” “M” if the weight is not stable
Valid status token “<S>” “SPACE” if the weight is valid
Coz status token “<S>” “Z” if the weight is equal to zero
<S>
Overload status token “<S>” “O” if the weight is in overload
Underload status token “<S>” “O” if the weight is in underload
Overload char token “<S>” “&” if the weight is in overload, all the weight
digits are replaced by &
Underload char token “<S>” “=” if the weight is in underload, all the wei-
ght digits are replaced by =
Weight value Displayed weight The string will show the weight.
<W-0x.y>
[left align ] Depending on the parameters set, the format will change accordingly:
Displayed weight - Show the sign “-” if the weight is negative
<w-0x.y>
[ right align]
Gross weight 0 Fills the empty spaces with “0”
<G-0x.y>
[left align ]
Gross weight x Indicates the number of digits by which the weight is shown
<g-0x.y>
[ right align] (max 15, from 1 to F)
Net weight . Show the decimal point
<N-0x.y>
[left align ] (If present in the weight)
Net weight y Indicates the fixed number of digits after the decimal point
<n-0x.y>
[ right align] (If the decimal point is present)
Tare
<T-0x.y>
[left align ]
-, 0, ., y characters are optional.
Tare
<t-0x.y>
[ right align]
Bit sequence Indicate a sequence of See bit table
<Bn,Bn,...>
exactly 8 bits
ASCII character Add the ASCII character. “CR” <CR> add the carriage return
<X> and “LF” are recognised as ASCII <LF> add the line feed
character 13 and 10 <2> add the STX character

46
TECH_MAN_ENG_DFWLPLUS
Bit:

Bit Bit used Type Name Description


B0 1 Fixed Bit value = 0 Used to complete the byte
B1 1 Fixed Bit value = 1 Used to complete the byte
B3 1 Variable Net weight 1 if the displayed weight is net
0 if the displayed weight is gross
B4 1 Variable Weight equals to zero 1 if the gross weight is equal to zero (zero indication on the screen)
0 if the gross weight is not equal to zero (no zero indication on the
screen)
B5 1 Variable Stable weight 1 if the weight is stable (no motion indication on the screen)
0 if the weight is unstable (motion indication on the screen)
B6 1 Variable Negative gross weight 1 if the gross weight is negative
0 if the gross weight is positive
B7 1 Variable Weight in underload / 1 if the weight is in underload / overload
overload 0 if the weight is not in underload / overload
B9 1 Variable Active tare 1 if a tare is active
0 if there isn’t a tare active
B10 1 Variable Active preset tare 1 if a preset tare is active
0 if there isn’t a preset tare active
B11 2 Variable Weight type 00 if the displayed weight is gross
01 if the displayed weight is net
10 if the displayed weight is a tare

47
TECH_MAN_ENG_DFWLPLUS
Example of custom string configuration to replicate the standard Dini Argeo string:

String definition 1,<S>,<M>,<P><W7.>,<U><CR><LF>


Gross mode token “<M>” GS
Net mode token “<M>” NT
Tare mode token “<M>” T
Preset tare weight token “<M>” PT
Unit token “<U>” Depending on the unit of measure
Positive polarity token “<P>” SPACE
Negative polarity token “<P>” -
Invalid status token “<S>” SPACESPACE
Motion status token “<S>” US
Valid status token “<S>” ST
Coz status token “<S>” ZR
Overload status token “<S>” OL
Underload status token “<S>” UL
Overload char token “<S>” (null)
Underload char token “<S>” (null)

Example of custom string configuration to replicate the extended Dini Argeo string:

String definition 1,<S>,<WA.>,<MT><TA.>, ,<U><CR><LF>


Gross mode token “<M>” GS
Net mode token “<M>” NT
Tare mode token “<M>” SPACESPACE
Preset tare weight token “<M>” PT
Unit token “<U>” Depending on the unit of measure
Positive polarity token “<P>” SPACE
Negative polarity token “<P>” -
Invalid status token “<S>” SPACESPACE
Motion status token “<S>” US
Valid status token “<S>” ST
Coz status token “<S>” ZR
Overload status token “<S>” OL
Underload status token “<S>” UL
Overload char token “<S>” (null)
Underload char token “<S>” (null)

48
TECH_MAN_ENG_DFWLPLUS
7. COMMUNICATION COMMANDS
Foreword:

in the serial commands and the respective responses

nn Address 485 of the instrument (2 characters), only if communication mode 485 is enabled
<CR> ASCII terminator character 13 (0D) (1 character)
<LF> ASCII terminator character 10 (0A) (1 character)

Simple weight reading


Command nnREAD<CR><LF>
Response Short string (see page 44)

Complete weight reading


Command nnREXT<CR><LF>
Response Long string (see page 44)

Reading custom string


Command SF#1<CR><LF> (configurable from Dinitools)
Custom string (see page 45)
Response

Performing a semi-automatic tare


Command nnTARE<CR><LF>
Response OK<CR><LF&gt indicates that the command has been received correctly

Setting a tare value (PT)


Command nnTMANtttttttt<CR><LF>
Where t...t is the tare value, with decimal points, max 8 characters.

Response OK<CR><LF&gt indicates that the command has been received correctly

Examples TMAN1.56<CR><LF>
sets a tare value of 1.56
TMAN100<CR><LF>
sets a tare value of 100

Clearing the stored tare


Command nnCLEAR<CR><LF>
Response OK<CR><LF&gt indicates that the command has been received correctly

Zeroing the scale (ZERO key function)


Command nnZERO<CR><LF>
Response OK<CR><LF&gt indicates that the command has been received correctly

49
TECH_MAN_ENG_DFWLPLUS
SPECIFIC COMMANDS FOR ALIBI MEMORY (OPTIONAL)

Save requests

Command nnPID<CR><LF> request to save the weighing

Response successful registration


nnPIDss,c,wwwwwwwwwwuu,ppttttttttttuu,xxxxx-yyyyyy<CR><LF>

no registration
nnPIDss,c,wwwwwwwwwwuu,ppttttttttttuu,NO<CR><LF>
where:
ss weight status (2 characters)
TL Tilt error condition (NO RECORDING)
OL Overload condition (NO RECORDING)
UL Underload condition (NO REGISTRATION)
ST Stable weight
US Unstable weight (NO RECORDING)

c Scale number (1 character)

w…w Gross weight (10 characters)

uu Unit of measurement (2 characters)

pp Type of tare: double space “ “ if semi-automatic, “PT” if preset (2 characters)

t…t Tare value (10 characters)

xxxxx Number of rewrite (5 characters)

yyyyyy Progressive weighing (6 characters)

Example PIDST,1, 1500,0kg,PT 2,8kg,00000-000158<CR><LF>

Reading a stored weighing

Command nnALRDxxxxx-yyyyyy<CR><LF>
Where xxxxx is the rewrite number, yyyy is the weighing sequence.

Response s , w w w w w w w w w w u u , p p t t t t t t t t t t u u<CR><LF>

where:
s Number of scales (always 1)

w...w Gross weight (10 characters)

uu Unit of measurement (“g”, “kg”, “ t”, “lb”)

pp Type of tare: double space “ “ if semi-automatic, “PT” if preset (2 characters)

t...t Tare value (10 characters)

Example ALRD00000-000158<CR><LF>
1, 1500,0kg, 2,8kg<CR><LF>

50
TECH_MAN_ENG_DFWLPLUS
8. WIRING DIAGRAMS

CPU board (DFWL-1x, DFWLI-1x, TPWNBT-1x)

IR SENSOR
(front side)
POWER IN: 3,3...12Vdc

+ Vin

JP2 - GND
Power from JP2
power supply
unit + BAT

Power from +Vaux


USB port

BUZZER GND

+5V
USB 2.0
CTS
CELL3 CELL4 RX
SIG+
TX
EXC+

EXC+
EXC-

EXC-
SIG+

SIG+
SIG-

SIG-

SIG- COM 2 COM 1


RS232 RS232

SEN+
CELL2 J1
Legal
SEN-
J1
EXC+

EXC-

SIG+

SIG-

Internal use

TERM 485
ON BOOT
EXC+
COM4 - TTL CAL
RS485 RS232
EXC- SW2 +5V GND0 10 2 485
SENSE COM 3 COM 1
SOCKET
CELL1 ON
BOOT
ON JP3
SW3

GND

CTS
A[+]
B[-]
SENSOR

RX

TX
SENSOR

ON ON
I/O BOARD - CLOCK - ALIBI MEMORY RS485 Termination resistance
SEN+ EXC+ SEN- EXC-
Auto switch on

RS232 serial port with RJ11 connector Optional IN/OUT DFIO board
INPUTS
12 ÷ 48 V

OUTPUTS
48 Vac / 60 Vdc
0.5 A Max
VDC

2 4 5

PIN MEANING
VDC
2 TX

4 GND
EXTERNAL BOARD
5 RX
Alibi Clock
JP1
JP2

51
TECH_MAN_ENG_DFWLPLUS
Optional analogue DAC16OSER output board

Board
CPU

to GND
to RX
to TX

to A+
to B-
JP2
GND
A+
RX
TX

B-
SW1 GND
JP2
V+
PLC
GND
A+
RX
TX

B-

GND
JP2 SW1 I+
GND
V+
GND
A+
RX
TX

B-

Board to +Vin +
JP2 5 ÷ 12 V
SW1 GND GND
DC
CPU to GND -
V+ I+
GND
A+
RX

B-

JP2
GND GND SW1
+ I+
V+
A+

5 ÷ 12 VDC
B-

OFF ON OFF ON OFF ON OFF ON

-
1 1 1 1

SW1 GND GND


2 2 2 2

+
OPERATION RANGE (only for live analogue output)
I+
5 ÷ 12 VDC V+
-
GND
+ OFF ON OFF ON OFF ON OFF
ION+
5 ÷ 12
1 VDC 1 0 / 10 V1 1
- 2 2 2 2

OFF ON OFF ON OFF ON OFF ON


÷ 12 VDC
1 1 1
-10 / 101 V
2 2 2 2

ON OFF ON OFF ON OFF ON


1 1 1
0/5V
2 2 2

ON OFF ON OFF ON
1 1
-5 / 5 V
2 2

52
TECH_MAN_ENG_DFWLPLUS
9. PROGRAMMING ERRORS

MESSAGE DESCRIPTION SOLUTION


Check the presence of the board inside the
al.err “Alibi memory” board (optional) not detected. indicator. If present, check it is not damaged and is
installed correctly.

er.1.b.x Check the presence of the board inside the


indicator.
If missing, deactivate any inputs or outputs
“inputs / outputs” board (optional) not detected.
(parameter “inputs” or “output”, see page 37-38).
If present, check it is not damaged and is installed
Er.r.b.x correctly.

Check the cells are connected properly.


Eq.err Impossible to perform equalisation. Check the signal of each cell in the diagnostic menu
(menu diaG, parameter adC.uv, see page 39).

First calibrate the zero point, then proceed with the


preC. Calibration error.
next points.

Check the connection of the load cell.


Err.pnt Calibration error. Check that the cell signal is stable, valid and greater
than that of the previously acquired point.

Er 11 Calibration error. Increase the calibration weight.

Check that the signal coming from the cell increases


upon the increasing of the weight loaded on the
Er 12 Calibration error.
scale. When acquiring the calibration points, use the
increasing calibration weights.

Repeat the calibration, checking that the capacity


Er 37 Calibration error.
and division have been correctly set.

Reset the factory configurations (menu advanC,


Er 39 Instrument not configured.
parameter dflt.t, see page 46).

Er 85 Instrument configured but not calibrated. Perform calibration.

Check that the signal coming from the load cell is


C.er. 36 Calibration error.
not negative.

Check in the menu diaG, parameter AdC.uv (see


page 39) that the signal is stable and try again.
Err.mot Unstable weight
If the connection of the cells is with 4 wires, check
that the sense jumpers are inserted.

53
TECH_MAN_ENG_DFWLPLUS
10. SUMMARY OF THE PARAMETERS

Cal Calibration 11
div Division 11
0.Cal Zeroing the pre-tare (zero calibration) 12
Grav Area of gravity of the place of use 12
serial Configuration of the serial ports 13
Com.PC PC port configuration (communication with PC, PLC or repeater) 13
mode Selection of the communication mode 13
protoC Communication protocol 13
Com.sel COM port selection for use as PC port 14
baud Communication speed (baud rate) 14
bit Configuration of the serial protocol 14
Com.Prn PRN port configuration (communication with printer or repeater) 15
mode Selection of the communication mode 15
protoC Communication protocol 15
baud Communication speed (baud rate) 15
bit Configuration of the serial protocol 16
Cts Printer control signal 16
power.P Printer power supply 16
aux.1 Communication with PC, PLC or repeater 17
mode Selection of the communication mode 17
protoC Communication protocol 17
baud Communication speed (baud rate) 17
bit Configuration of the serial protocol 18
aux.2 Communication with PC, PLC or repeater 17
mode Selection of the communication mode 17
protoC Communication protocol 17
baud Communication speed (baud rate) 17
bit Configuration of the serial protocol 18
usb Communication with PC, PLC or repeater 18
advanC Advanced configurations 19
radio Connection port of radio-frequency module 19
ttl.til TTL port / Inclinometer activation 19
term Closing character of each print line 19
iGnore Ignore unknown commands 19
layout Print customisation 20
lanG Print language settings 21
Char Setting the font 21
header Print header 22
data Selection of the weight data 24
weiGht Progressive weighed 24
tiCket Ticket/label progressive 24
CloCk Date and time 25
barC.39 Bar code 39 25
barC.up Barcode top margin (mm) 25
barC.l Barcode left margin (mm) 25
barC.h Barcode height (mm) 25
barC.dt Selection of the weight data 26
Copies Multi-copy prints 26

54
TECH_MAN_ENG_DFWLPLUS
end.tiC Paper outlet for end of ticket/receipt 26
b.line White print head preheating line (thermal printer only) 26
label Label configuration 27
lb.save Saving labels in printer memory 27
test Test print all formats 27
filter Weighing filters 30
sCreen Adjusting the display 29
bak.lit Backlighting 29
briGht Brightness 29
loCk Display lock (for use by the manufacturer) 29
Colour Backlight colour 29
batt Battery power supply 30
eCo.bat Energy saving for battery operation 30
autoff Auto switch-off 31
remote Remote control 31
an.out Analogue output 32
inputs Digital inputs 34
output Digital outputs 35
reset Factory configuration reset 36
diaG Diagnostics 36
adC.uv Converter (uV) 36
displa Display 36
keyb Keypad 36
Cts CTS status 36
output Digital outputs 36
inputs Digital inputs 36
an.out Analogue output 36
ser.num Serial number 36
prG.ver Firmware release 36
div.int Internal divisions 36
adC.pnt Converter (ADC points) 36
bt.adC Battery voltage 36
pw.adC Mains power supply voltage 36
serial Bridge between serial ports 36
ADVANC Advanced 37
Cal.par Calibration parameters 37
deCim Configuration of the decimal point 37
div Reading division 37
u.m. Unit of measure 37
ranGe 1 Scale capacity (maximum capacity / first weighing range) 38
ranGe 2 For multirange scales (second weighing range) 38
ranGe 3 For multirange scales (third weighing range) 38
equal Equalisation function 38
n.Chan Equalised analogue channels 38
equal.p Equalisation 39
Cal.adv Complete calibration 42
stb.flt Additional stability filter 40
peak.fl Anti-peak filter 41

55
TECH_MAN_ENG_DFWLPLUS
metrol Metrological parameters 41
0.perC Reset percentage via key 41
div.stb Sensitivity of the weight stability control 41
tim.stb Stability detection time 41
0.trk Zero hold function (tracking) 41
0.trk.sp Zero tracking speed 41
on.zero Reset at power and reset percentage 42
Cal.adj Re-acquisition / change of the calibration points in memory 42
Cal.man Manual calibration (via insertion of ADC points) 42
d.sale Limitations for direct sales 42
keyb Type of keypad 42
tilt Inclinometer (for use by the manufacturer) 42
reaCt Reactivation of the print or weight totalisation function 43
loCk.kb Permanent keypad lock (excluding key C ) 43
alibi.r Reset of fiscal memory (alibi memory, optional) 43
pin.teC Access PIN to programming menu 43
pin.use Access PIN to user menus 43
dflt.t Total reset of memory and calibration 43

56
TECH_MAN_ENG_DFWLPLUS
11. FAQ - Frequently Asked Questions

Calibration

Can I change the maximum capacity without recalibrating?


Yes, you must change the parameters RANGE 1,2,3 (Q-1-4,5,6). (See page 41)

Can I change the division without recalibrating?


Yes, you must change the parameter diV (Q-1-2). (See page 40)

Can I change the position of the decimal point without recalibrating?


Yes, you must change the parameter deCim (Q-1-1) and the value of the calibration points using
step CAL.MAn (Q-5-6). (See page 40 and 45)

Can I calibrate the instrument in “multi-division” mode?


Yes, using the advanced configuration function from PC and the Dinitools program.

Communication

Scale doesn’t answer


• Check that the cable is in good condition and that there are no faults (use a multimeter).
• Check that the communication port of the PC or device being used is not compromised. If necessary, try another device / PC.
• Check that you have connected the cable to the correct serial port.
• Check step configuration Baud and bits. (See page 17)
• Temporarily activate continuous communication and retry string reception. If the string has been received correctly, carefully
check the syntax of the command sent, the communication timeouts and the presence of the terminator.

Generic

The scale does not turn on


• Check that the input voltage level to the motherboard is correct.
• Try the forced power by inserting the “ON BOOT” jumper present on the motherboard. If the indicator lights up, check the cor-
rect operation of the keypad, using the diagnostics menu diaG. (See page 39)
• Possible failure of the internal rechargeable battery (if present).

57
TECH_MAN_ENG_DFWLPLUS
Notes

This publication, or portions thereof, may not be duplicated without written permission from the Manufacturer. All information in this manual is based on the data
available at the time of its publication; the Manufacturer reserves the right to make changes to its products at any time, without notice and without incurring any penalty.
We therefore recommend that you always check for updates.
The person responsible for the use of the scale must ensure that all safety regulations in force in the country of use are applied, ensure that the scale is employed in
accordance with the intended use and avoid any dangerous situation for the user.
The Manufacturer declines all responsibility for any weighing errors.

58
TECH_MAN_ENG_DFWLPLUS
HEAD OFFICE
Via Della Fisica, 20

www.diniargeo.com
Via Dell’Elettronica, 15
SERVICE ASSISTANCE

41042 Spezzano di Fiorano, Modena - Italy


41042 Spezzano di Fiorano, Modena - Italy
Tel. +39 0536 843418 - Fax +39 0536 843521

Tel. +39 0536 921784 - Fax +39 0536 926654


Stamp of the authorised service centre
TECH_MAN_ITA_DFWL PLUS The information in this document is approximate and can be subject to variations without prior notice by Dini Argeo, in compliance with the norms in force. The official
technical data are available in the updated version on the following website www.diniargeo.com or by contacting Dini Argeo Customer Service.
Rev. 30.11.2022