You are on page 1of 12

elemental analysis

The Thermo Scientific iCE 3000

Product Specifications
Series AA Spectrometers are
fitted with highly efficient
background correction systems
for flame and furnace use. Their
important design features are
described in this article along
with discussions on the
advantages and disadvantages of
Design Considerations for other techniques.

High Performance Background


Correction Systems in
Atomic Absorption Spectrometry

What is Background Absorption? The monochromator is tuned to the


The technique of atomic absorption resonance wavelength, but will pass a
spectrometry depends on the formation of range of wavelengths (typically 0.1 - 1.0 nm) -
an atomic vapor from a sample solution and much greater than the atomic absorption
on the measurement of element specific line width.
radiation absorbed by the vapor. When only the analyte atomic species
Any metallic element in atomic form are present, the amount of light absorbed
will absorb light over a narrow range of will depend only on the concentration of the
wavelengths centred on the resonance analyte element in the light path.
wavelength whose value is specific to the Materials other than the analyte,
element. The absorption line width is about derived from the sample matrix and the
0.002 nm at normal atomization atomization system itself, may also be
temperatures. In an atomic absorption present in the light path. In some
spectrometer, the light source is usually a circumstances, these can absorb or scatter
hollow cathode lamp, which produces the radiation, thus increasing the total
resonance radiation of about 0.001 nm line absorbance and hence the apparent
width. The lamp also generates light at concentration of the analyte. This effect is
wavelengths other than the resonance known as non-specific, or background,
wavelength, and this makes it necessary to absorption and, if present, can reduce the
pass the light through a monochromator. accuracy of the analysis unless steps are
taken to correct for it.

Part of Thermo Fisher Scientific


Product Specifications Molecular Absorption Scatter Structured Absorption
Many sample matrices contain halide salts Some types of sample matrix are not Other components of the sample matrix will
at relatively high concentrations. In the completely vaporized in the atomizer, but be atomized, and their atomic absorption
atomizer, these will form molecular vapors, form small particles or clusters. These spectra may contain lines close to the analyte
resulting in molecular absorption of the scatter the light from the hollow cathode resonance line. Cases of direct line overlap
radiation. Molecular absorption spectra are lamp, preventing it from reaching the are very rare, but it is possible that lines
broad band, with band widths of the order detector, and so cause an apparent derived from the matrix components will fall
of tens of nanometres, and for alkali metal absorption signal. The degree of scatter within the monochromator band pass.
halide salts, absorption increases sharply in depends strongly on both the wavelength of In most cases, the broadband
the deep ultra-violet region (below 220 nm) the light and the size of the particles, and background absorption is a relatively smooth
of the spectrum. Some typical vapor phase increases markedly at low wavelengths. function of wavelength, and may be regarded
spectra of alkali metal halide salts are Background absorption caused by as virtually constant over the bandpass of the
shown in Figure 1 below. scatter is, therefore, also most significant spectrometer monochromator.
for elements with their resonance lines Occasionally, examples are encountered
below 220 nm, and depends both on the of more complex background absorption
total dissolved solids content of the spectra with line structure superimposed on
samples, and the nature of the dissolved this relatively constant background. An
solids - refractory metals and those forming example of these so-called structured
very stable oxides, such as the alkaline backgrounds is the determination of
earths, cause the greatest problems. selenium in whole blood, where the iron
content of the matrix exhibits such structure
within the monochromator bandpass. Should
such structure occur within the
monochromator bandpass of a deuterium
background correction system, systematic
errors will be present in the corrected
atomic absorption signal, and hence the
analytical result. Such examples are
relatively uncommon in AA analysis but,
when present, may advantageously be dealt
with using the Zeeman correction technique.

Figure 1: Molecular spectra of simple halide salts

Background absorption caused by


molecules is thus particularly troublesome
when measuring elements with their main
resonance line below 220 nm and in sample
matrices containing alkali metal halides. This
includes many of the toxic heavy metals,
such as lead, cadmium and arsenic in
biological and environmental samples, such
as natural waters, effluents and body fluids.
Methods of Compensating Continuum Method It is possible to use continuum source
for Background Absorption The second (background) measurement in this background correction with sequential
technique is made using radiation from a measurements of total (line source) and
Introduction source with a broad spectral output, such as a background only (continuum source)
Instrumental methods of correcting for deuterium arc lamp. The radiation intensity absorbance, followed by manual subtraction
background absorption depend upon making from such lamps is almost independent of to give the corrected atomic signal. It is also
at least two separate measurements of the wavelength over the monochromator band possible to design the spectrometer in such a
sample absorption. The first is the normal pass (i.e. over a wavelength range of about 1 way as to automatically make both
atomic absorption measurement made at nm), and so ‘fills’ the monochromator band measurements and subtract the results, and
the analyte resonance line, and gives the pass. The analyte atomic absorption takes to do this sufficiently rapidly that many pairs
total (atomic plus background) absorption. A place over a wavelength range of about 0.002 of measurements can be made and
second measurement is then made, in such nm; the amount of continuum source radiation subtracted over the normal sampling period,
a way as to reduce the effect of the absorbed by the analyte atoms is, therefore, providing fully automatic background
absorption from the analyte atoms to a negligible. Background absorption is largely a correction. This is the basis of almost all
negligible amount, while not affecting the broad band phenomenon, and is also commonly used background correction
background absorption. This second reasonably independent of wavelength over systems, and most modern atomic absorption
measurement gives only the background the monochromator band pass. Only the spectrometers are capable of making
absorbance, so that subtraction of the two background absorption will, therefore, be measurements that are automatically
measurements will give the true atomic measured by the continuum source. This corrected for background absorption.
absorbance. situation is shown diagrammatically in figure 2. Well designed background correction
All instrumental background correction systems based upon the continuum source
systems are based on this principle. They technique are capable of correcting the
differ in the means used to reduce the atomic majority of non-specific absorption problems
signal to negligible levels for the background encountered during both flame and graphite
measurement, and in the frequency with furnace atomic absorption measurements.
which the two measurements are made. Only in situations in which the background
absorption is NOT independent of
wavelength over the monochromator band
pass, or where direct spectral overlap of the
analyte atomic line and a narrow band
absorption due to a matrix component
occurs, will it give incorrect results.
Commonly used continuum sources,
such as deuterium arc lamps, emit most of
their radiation in the ultra-violet region of
the spectrum, below 350 nm. Background
correction in the visible region is, therefore,
not possible, unless an additional source is
used. However, almost all background
absorption phenomena occur in the ultra-
violet region - scatter and molecular
absorption are usually negligible at
wavelengths above 350 nm - so this is not a
serious practical limitation.

Figure 2: Atomic absorption from a hollow


cathode lamp and a continuum source.
Product Specifications Self Reversal Method The hollow cathode lamp is, run with
alternating low and high current pulses.
Zeeman Method
This technique (also known as the Smith- This technique does not rely on a continuum
Hieftje technique) relies upon a broad band Total absorbance is measured during the or pseudo-continuum source for the
source to measure the background absorption. low current pulse and the background measurement of the background absorption.
Rather than employ a separate source, the absorbance is measured during the high Instead, the absorption of the analyte
hollow cathode lamp itself is used. current pulse. The two signals are atomic absorption line is cancelled, by
Hollow cathode lamps are normally automatically subtracted to give the applying an intense magnetic field to the
operated at lamp currents up to 25 mA, corrected atomic signal. atomizer. The effect of the magnetic field on
when they produce the narrow line emission The self-reversal technique is capable the absorption spectra of free atoms is
spectra required for atomic absorption of correcting for most background complex, but essentially causes the single
measurements. If the peak current is absorption phenomena over the full sharp line to split into a family of lines
increased to several hundreds of milliamps, wavelength range of the spectrometer. spread around the wavelength of the original
the emission lines broaden and self reverse Although the emission line is broadened line (some examples are shown in Figure 4).
(Figure 3). Under these conditions, the during the high current pulse, it does not Furthermore, the split absorption lines
narrow atomic absorption line can only 'fill' the monochromator band pass in the are polarised in directions either parallel or
absorb a small part of the broadened same way as a true continuum source. The perpendicular to the magnetic field. If the
emission line, while the broad band or system can, therefore, correct for some hollow cathode radiation is also polarized,
continuous background absorption will occur cases of structured background. It will not, by passing it through a suitable polarizer,
as usual. Absorbance measurements made however, correct for actual line overlap. measurements made with the magnetic
at very high lamp currents, will be mainly The disadvantages of the technique are field on will not contain any analyte atomic
due to the non-specific absorption. the effect of the high current pulses on the absorption, and so will be due solely to
hollow cathode lamp - unless specially background absorption taking place at the
designed lamps are used, the lamp lifetime analyte resonance wavelength.
will be significantly reduced - and the
breakdown of the assumption that the
atomic absorption is negligible during the
high current measurement.
The degree to which the lines for the
different elements broaden and self reverse,
and hence the validity of this assumption,
will depend on the element and on the
detailed construction of the lamp,
particularly the cathode material and the fill
gas pressure. When the assumption breaks
down, the background-only signal measured
Figure 3: Hollow cathode lamp emission line will actually contain a significant
broadening at high lamp currents. contribution from the atomic signal. When
this is subtracted from the total signal, the
apparent atomic signal that results will be
over-corrected. The effect is to reduce
analytical sensitivity (by as much as 70% in
some cases) and increase the curvature of
the calibration graph.

Figure 4: Zeeman splitting of atomic lines


The Zeeman technique is the only one With Zeeman background correction, Instrumental Design
in which measurement of the background the magnet must be placed around the
absorption takes place at exactly the same atomizer. It is impractical, therefore, to Range
wavelength, and over exactly the same apply this technique to flame Irrespective of the physical technique
band pass, as measurement of the total measurements, and commercially available employed to derive the background
signal. It is, therefore capable of correcting systems are usually limited to graphite absorbance signal, all systems have to
for highly structured background spectra furnace analyses. It is also possible to apply subtract this signal from the total signal
and even cases of spectral overlap. A the Zeeman effect to the radiation source, to obtain the corrected atomic absorption.
disadvantage is the relatively complex causing the emission lines, rather than the To obtain an accurate result, both the total
magnet and its control electronics; magnetic absorption lines, to split. Measurements and the background absorbance signal
field strengths up to almost 1 tesla (10 made with the split lines (for elements must be within the measuring range of
kilogauss) are required to obtain optimum exhibiting the ‘normal’ Zeeman effect, at the spectrometer.
line splitting for some elements. least), will not include a contribution from
Some elements, including many of the analyte atomic absorption. However, the Accuracy
those most commonly determined by atomic background absorption is not then measured As the analytical signal (from which the
absorption spectrometry, exhibit a at exactly the same wavelength as the final result is derived) is the difference
phenomenon known as ‘anomalous’ Zeeman atomic absorption and, furthermore, is between the total and the background-
splitting. Here a component of the split line measured with a highly structured source. If only signal, it will contain errors
remains at exactly the same wavelength as the background absorption is also originating in both of these signals, in
the unsplit line. The atomic absorption is structured, large errors can be introduced. addition to any errors introduced by the
not, therefore, reduced completely to zero Normal hollow cathode lamps will not signal subtraction itself. The accuracy
when the magnetic field is on and the operate in intense magnetic fields and with which the primary signals are
background-only measurement contains a specially designed sources are required. measured, therefore, needs to be
contribution from the analyte atomic Consequently, this arrangement has only significantly greater than the accuracy
absorption. This, when subtracted from the been used in a limited way, and in special required for the analytical signal.
total, magnet-off, measurement, results in purpose instrumentation. It can be shown from statistical
over-correction of the atomic signal, The magnetic field may be applied as considerations that the error in the
reducing the apparent chemical sensitivity either a DC (permanent) magnet system or corrected absorbance signal will be
and increasing the calibration curvature, in as an AC (electromagnet) system. In the minimized when the errors in the total
a similar way to the self-reversal technique. case of a DC system, the magnetic field is absorbance and background-only
always present. Differentiation of total and absorbance signals are of similar
background signals is obtained by switching magnitude. This implies that the
the plane of polarization of the hollow intensities of the radiation sources used to
cathode radiation. This is usually achieved measure the absorbance signals should
by physically rotating the polarizer itself also be similar.
which requires a complex, high precision
motor driver mechanism. The AC Zeeman
method where differentiation is obtained by
modulating the magnetic field, is preferable,
as the field can be modulated by driving the
magnet with a simple AC waveform.
Furthermore, the higher the frequency of
magnet modulation, the closer the two
signals of total and background absorption
are measured in time. With fast transient
furnace absorption peaks, this reduces
analytical errors.
Product Specifications Emssion Background Correction Systems Optical System
Flame, and particularly graphite furnace
atomizers, can produce intense emissions Quadline Source Location and Alignment
of radiation as a result of the high The continuum source technique used in the Although the deuterium arc lamp has a long
temperatures used. Thermo Scientific AA Series of useful lifetime, it will eventually require
The spectrometer must be capable of spectrometers offers the performance replacement.
accurately measuring the source lamp required by the vast majority of analysts in For accurate operation a continuum
intensities in the presence of the atomizer a cost effective, and theoretically sound, source background correction system, it is
emission, in order to derive the two package. Detailed improvements in the essential that the two sources are precisely
absorbance signals. Bright source lamps, source design, the optical configuration of aligned relative to each other and to the
and an optimized optical system are, the spectrometer and the signal processing atomizer. In the Thermo Scientific iCE 3000
therefore, essential. functions overcome many of the limitations Series AA Spectrometers, the deuterium arc
The internal dimensions of the Thermo traditionally associated with the technique. lamp, once initially aligned, does not require
Scientific cuvette have been considered in This results in a high performance system further adjustment. The hollow cathode
the design of the spectrometer optical capable of handling the most demanding lamps are automatically aligned to the
systems, to ensure that thermal emission applications. deuterium beam using the spectrometer's
from the cuvette walls is, as far as possible, automatic lamp alignment facilities.
screened from the detector without Source Lamp
compromising the high optical transmission A high intensity deuterium arc lamp is used iCE 3500 Optical System Layout
efficiency and low noise characteristics of as the continuum source, with a novel 4 The novel echelle optical system layout of
the spectrometer. electrode design that allows the lamp to be the Thermo Scientific iCE 3500 AA is shown
This is particularly useful when electronically modulated at four times the in Figure 5. Front and rear beam selector
determining refractory elements requiring mains frequency, without the use of mirrors act together to direct the light beam
high atomization temperatures and with mechanical choppers. This modulation through either the flame or the furnace
main resonance lines above 320 nm. frequency, together with the digital signal atomizer. The optics have been optimized
processing described below, permits accurate for furnace use so that the “pencil-beam”
Transient Response correction of signals changing at rates of up optics eliminate potential emission
Atomic and background signals generated by to about 70 absorbance units per second. problems and half height slits (traditionally
flame atomization systems are normally The novel lamp design, coupled with an used but with an attendant loss of energy)
stable, steady state measurements. Graphite innovative power supply, allow the arc are not required.
furnace atomization, however, can generate current, and hence the
rapidly changing signals. A modern, fast intensity of the
heating design, such as the Thermo Scientific emitted radiation, to
GFS35, can generate total and background be varied over a wide
signals that are varying independently at range while
rates of up to 50 absorbance units per maintaining excellent
second. Pairs of measurements must, stability at any
therefore, be taken at a sufficiently fast rate particular setting. The
to enable the background correction system to continuum source
accurately track these two signals and intensity can,
subtract them to give an accurate, undistorted therefore, be
corrected signal. accurately matched
with the hollow
cathode lamp
intensity over a wide range of elements and Figure 5: iCE 3500 optical system showing the
wavelengths, ensuring that the corrected echelle design.
signal is measured as accurately as possible.
Echelle technology is the only solution to
providing high energy throughput and pencil
beam optics, without going to a very large
traditional Czerny-Turner design (which would
need a 500 mm focal length and an 1800
lines/mm grating). The echelle reciprocal
linear dispersion of 0.5 nm/mm is the best in
this class of instrument.
Signal Processing Digital systems are also more versatile. Errors from this source can be reduced
Operations such as averaging successive by using fast modulation of the sources and
Analogue versus Digital Processing signal values, which are difficult to perform detector system, so that many pairs of
Automatic background correction systems with analogue circuits, can be simply measurements are made rapidly and the
require significant amounts of signal achieved by programming the amount that the signals change between
processing to derive the analytical signal. microprocessor. For these reasons, the signal individual measurements is reduced. The
The basic measurements taken at the processing functions in the iCE 3000 Series versatility offered by microprocessor based
detector give the intensities of radiation are carried out by a digital system. Radiation digital signal processing allows a further
from the hollow cathode lamp and the intensities are measured by a photomultiplier, refinement to be made, further improving the
continuum source lamp, each with, and and converted immediately to digital form accuracy of the background correction (as
without, the presence of sample in the using a device known as an Analogue-to- shown in Figure 6b for bracketing correction).
atomizer. From these signals, the total and Digital convertor. All subsequent processing,
background absorbance values are derived including logarithmic conversion to give the
by taking the logarithm of the measured absorbance values and subsequent
radiation intensities from each source with, subtraction of the background-only signal,
and without, sample. Finally, the two are performed by the microprocessor.
absorbance values are subtracted to give The microprocessor also optimizes both
the corrected atomic absorbance. the gain of the photomultiplier and the
Traditional instruments performed this intensity of the deuterium arc lamp, ensuring
processing using analogue electronics. that the raw intensity measurements are
They inevitably have a number of variable made under optimum conditions.
elements which have to be set up
individually to compensate for small Signal Bracketing
differences in the electronic components. It is not possible in a practical background
This adds to the initial cost of the correction system to measure the total and
instrument, and means that regular background absorbance signals at exactly the
same time. If the signals are changing Figure 6b: Measurement sequence for bracketing
maintenance and adjustment is needed to correction.
maintain the original performance. rapidly, the actual value of the total signal
With the development of digital when the background signal is measured If we wait to perform the calculation
microprocessor based electronics, an could be slightly different, resulting in an until tT2, when the next measurement of
alternative has become available. In a digital error in the corrected signal. Consider the the total signal T2 is taken, we can use the
system, the values of the different quantities sequence of measurements shown in Figure values of T1 and T2, together with the
are represented by numbers, not voltages, 6a for simple correction, where T1, B1 are known times tT2, tB1 and tT2, to estimate
and are manipulated by a microprocessor. the first pair of total and background the actual value, TE, of the total
The major advantage of this approach is that absorbance measurements, taken at times tT1 absorbance at time tB1.
there are no variable components in the and tB1, and T2, B2 are the second pair, taken TE = T1 + ( T2 -T1 )x(tB1 - tT1)/(tT2 - tT1)
system that need to be set up, and that can at times tT2 and tB2. Simple subtraction of B1 Subtracting B1 from TE will then give
later go out of adjustment. The long term from T1 gives the result C for the corrected a very much better estimate of the true
performance is more stable and little, or no, atomic absorbance, underestimating the true atomic absorbance C1.
maintenance is required. value C1 by the amount that the total signal This technique is known as signal
has changed in the time from tT1 to tB1. bracketing, and only becomes practicable
when digital signal processing is used. In
combination with the fast modulation
system used in the iCE 3000 Series, it will
reduce errors in the corrected signals
caused by timing effects, to less than 1%
for signals changing at up to 70 absorbance
units/second.

Figure 6a: Measurement sequence for simple


correction.
Product Specifications Quadline Conclusions QuadLine correction may be employed
for the highest measurement sensitivity,
The design of the Zeeman magnet
system must also achieve other vitally
The key aspects of the QuadLine method of
background correction may be summarized whereas Zeeman correction may be selected important objectives. The system must be
as follows: when encountering a structured background. readily demountable for routine service and
• highest possible measurement sensitivity An additional benefit is the ability to employ user maintenance operations. It must be
compared with alternative techniques both techniques in combination within a extremely durable in the presence of the
• novel wideband source allowing fumace analysis. Since any modulated potentially aggressive chemicals employed
continuous intensity variation and fast Zeeman system is limited to correction during in the AA technique and, of course, must
modulation frequency
the atomize phase only, due to potential minimize extraneous stray fields due to their
• optical matching of the beam profiles
overheating of the magnet system, vitally detrimental effects on nearby, sensitive
from both the background correction
source and HCLs important method development information electronics and displays. Finally, the entire
• fast modulation frequencies of 200/240 Hz may be lost in the earlier pre-treatment system must be comprehensively equipped
• beam combining techniques ensuring furnace phases due to the inability to monitor with safety interlocks which, together with
homogeneity within the beam profile the corrected absorption signal. By selecting similar furnace devices, ensure operator and
• bracketing correction of both background the combined mode, QuadLine correction may equipment safety.
and dark level signals to overcome time be employed in all but the atomize phase,
dependence of measurement periods during which the Zeeman technique is used. Operation
• dark level correction to zero every 10 mS By this unique means, much valuable method On selection of the Zeeman-based
to improve dynamic range of ADC correction system, measurements are
development information may be obtained.
conversion
alternately taken in both the magnet on and
• intensity matching of background source Zeeman Technique
and HCL to improve dynamic magnet off conditions. With the magnet
measurement range The magnet, placed around the furnace head, off, the instrument provides a measure of
is modulated at a high 100/120Hz frequency the total (atomic + background signal). With
Zeeman Correction in order to achieve minimal time separation the magnet on, the instrument provides a
Employing the alternative major background of total and corrected signals. The magnet measure of the background-only signal at
compensation technique of Zeeman drive circuit achieves a maximum field the analyte wavelength. This is achieved by
correction can be shown to be of analytical strength of almost 0.9 Tesla (9 kilogauss) - effectively “removing” the analyte
benefit in a number of specific situations as optimum for all AA analytes. A trapezoidal absorption profile from the measurement
outlined earlier, in particular those of wavefom is used for the drive current, wavelength, as shown in Figure 7.
structured backgrounds and spectral ensuring that the measurement period is Subtraction of these signals then provides
overlaps. Furthermore, some analysts may performed totally during maximum and the displayed, corrected absorption profile.
find the technique fundamentally less prone constant magnetic field conditions, unlike A further benefit of the Zeeman
to operator variability, since no source sinusoidal field modulation techniques. correction system is that effectively the two
alignment is required, beyond maximising The magnet pole pieces are so designed measurements (magnet on and magnet off)
the HCL energy throughput. to provide a magnetic field constant to provide a referenced measurement and
For these reasons, Thermo Fisher within 5% in the key central working volume source drift correct ion without the
Scientific provide the iCE 3500 (dual of the graphite cuvette in the furnace head. necessity of conventional double beam
atomiser), and iCE 3400 Zeeman-only Due to the inhomogeneity of atomic and optics. Since only a single optical beam is
furnace systems. The system employ the molecular species within the graphite employed, alignment of the measurement
electromagnetic, or AC, Zeeman technique cuvette volume, such a constant magnetic source and atomization unit is also
applied at the atomizer. Uniquely, all these field is vital in providing reliable data. relatively straightforward.
systems also provide the QuadLine
background correction technique as
standard within the same spectrometer.
Thus, the technique of preference may be
chosen for a particular analysis by a simple
option within the operating software.

Figure 7: Inverse AC Zeeman method of correction.


Zeeman Conclusions Correction Performance Emission Breakthrough
The key aspects of this Zeeman method of The background corrected determination of
background correction may be summarized Steady State chromium in a nitrous oxide/acetylene flame
as follows: It is difficult to generate large background is a severe test of the background
• accurate correction of structured signals using conventional flame correction systems ability to reject flame
backgrounds and spectral overlaps atomization. Metal gauzes held in the light emission signals. Not only is the chromium
• correction at the exact analyte path are, therefore, used to generate large resonance line at 357.9 nm extremely bright,
wavelength
background-only signals. Since there can but the deuterium arc lamp intensity is also
• not limited to a wavelength range, as
be no atomic absorption in this experiment, beginning to fall in this region of the
with a deuterium source
any deviation of the background corrected spectrum. The optimum flame conditions for
• self-referenced to reduce drift
result from zero reflects the residual errors the determination are such that an intense
• less alignment necessary, due to single
source in the system. cyanide molecular emission, centred around
• efficient, optimized magnetic field The results from a typical experiment 355 nm, is produced from the flame. The
modulation ensures that the complete performed on an iCE 3000 Series AA effects of these phenomena on the
measurement is at full magnetic field spectrometer set up for nickel analysis background correction signal can be
strength using the default parameters and making illustrated by comparing the chromium flame
• field strength of almost 0.9 Tesla to 10 replicate measurements, each of 2 detection limits with and without, background
ensure complete line separation
seconds duration, are shown in Table 1. correction. The results obtained from an
• rapid field modulation of 100/120Hz to
They show that, even with a earlier, analogue instrument are compared
ensure faithful tracking of fast furnace
signals background signal approaching 2 with the digital spectrometer in Table 2 .
• relatively constant magnetic field strength absorbance units, residual background The chromium detection limits of the
along the length of the furnace cuvette to correction errors are of a similar order to earlier analogue instruments are clearly
achieve efficient spatial correction the optical noise in the system. In a real degraded when the background correction
• alternative QuadLine method for the sample measurement, these errors would system is used, while the Thermo Scientific
highest sensitivity requirements and be lost completely in flame and AA maintains its performance with no
combined QuadLine/Zeeman mode for
atomization noise. significant change.
method development work

GAUZE ABSORBANCE CORRECTED STANDARD


ABSORBANCE DEVIATION

1 0.353 0.0019 0.0006


2 0.506 0.0017 0.0009
3 0.860 0.0027 0.0007
4 1.285 0.0031 0.0028
5 1.803 0.0059 0.0044

Table 1: Background correction tests with gauzes

INSTRUMENT BC OFF BC ON
DETECTION LIMIT DETECTION LIMIT
(µg/mL) (µg/mL)

Analogue 1 0.033 0.115


Analogue 2 0.028 0.211
Digital Thermo Scientific AA 0.033 0.029

Table 2: Rejection of Emission effects


Product Specifications Transient Accuracy and
Transient Response
Figure 12 shows the corrected signal
generated by atomising 10 µL of a
Large transient background signals are reference sea water containing less than
relatively easy to generate by vaporizing 0.2 µg L-1 of chromium, while the same
relatively large amounts of a typical matrix sample when spiked with 4 µg L-1 gives the
material such as sodium chloride from a signal shown in Figure 13. The chromium
graphite furnace cuvette. Furthermore, if the signal is clean and undistorted, and can be
spectrometer is set up on a non-atomic readily quantified.
emission line, the ‘atomic’ signal must be
zero, so that any errors produced by the Figure 8: Background signal from 100 µg of NaCl
background correction system will become
immediately apparent.
Figure 8 shows the result of atomising
100 µg of sodium chloride at 1500 °C, with
the spectrometer set up on the non-
absorbing lead line at 220.4 nm. A
background signal approximately 1.4
absorbance units high has been produced,
but the corrected signal (shown in Figure 9) Figure 12: Corrected chromium signal and
shows no significant deviation from the background for unspiked seawater
baseline. Increasing the amount of sodium
chloride to 200 µg increases the background
Figure 9: Corrected signal under 1.4A background
peak height to around 2.6 absorbance units
peak
(Figure 10). Expanding the corrected signal
(Figure 11) reveals a significant increase in
noise at the background peak maximum (a
2.6 absorbance signal will reduce the
intensity of light reaching the detector to
approximately 0.25 % of the value at zero
absorbance), but, as before, there is no
indication of a systematic error present. Figure 13: Corrected chromium signal and
By expanding the time axis of these background for spiked seawater
signals, it is possible to measure the actual
rate at which the background signal is
changing. At the point at which it is
Figure 10: Background signal from 200 µg of NaCl
changing most rapidly, on the leading edge
of the peak between 0.7 and 0.9 absorbance
units, the measured rate of rise is slightly
over 25 absorbance units per second.

Figure 11: Corrected signal under 2.6A


background peak
Correction Method Zeeman Summary
Continuum Source When structured backgrounds, spectral The QuadLine and Zeeman background
It will normally be the case that overlaps or correction in the visible part of correction systems designed for the Thermo
advantageous measurement sensitivity the spectrum are encountered, Zeeman Scientific iCE 3000 Series range of atomic
improvements can be obtained by employing correction will often solve the problem. absorption spectrometers are capable of
the QuadLine correction method in Figure 15 illustrates the example of a cobalt correcting for background signals up to at least
preference to Zeeman. Figure 14 illustrates a interferent at the gold wavelength, an two absorbance units, which can be changing at
typical improvement achieved in the case of analysis that would prove impossible if only rates exceeding 25 absorbance units per second.
the analysis of copper. This may often prove the continuum source method were available. They provide a level of performance likely to
to be the difference between being able to meet the needs of all analysts performing flame
actually perform a given analysis, or not, or furnace atomic absorption measurements.
when working at the extreme trace levels.

Figure 14: Improved sensitivity for QuadLine (- - -)


compared with Zeeman (----) for 200pg Cu

Figure 15:The improved accuracy of Zeeman


analysis for Au in the presence of 150µg Co

Combination Mode
Since the Zeeman technique can only be
applied during the relatively short atomization
phase of the furnace, due to magnet
overheating restrictions, adequate knowledge
of the efficiency of other pretreatment
phases, or clean-out phases, of the furnace
cycle will not normally be available.
Using the unique Thermo Scientific iCE
3500 and iCE 3400 Combination mode, it is
now possible to view the progress of the
analysis in all phases by employing Zeeman
for the main measurement phase and using
the Quadline technique in all other phases.
Product Specifications

iCE 3300

iCE 3400

iCE 3500

Thermo Electron Manufacturing Ltd


©2010 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. Specifications, terms (Cambridge) is ISO Certified.
and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. PS40690_E 01/10C

Africa-Other +27 11 570 1840 Denmark +45 70 23 62 60 India +91 22 6742 9434 South Africa +27 11 570 1840
Australia +61 2 8844 9500 Europe-Other +43 1 333 50 34 0 Italy +39 02 950 591 Spain +34 914 845 965
Austria +43 1 333 50 34 0 Finland / Norway / Sweden Japan +81 45 453 9100 Switzerland +41 61 716 77 00
Belgium +32 53 73 42 41 +46 8 556 468 00 Latin America +1 608 276 5659 UK +44 1442 233555
Canada +1 800 530 8447 France +33 1 60 92 48 00 Middle East +43 1 333 50 34 0 USA +1 800 532 4752
China +86 10 8419 3588 Germany +49 6103 408 1014 Netherlands +31 76 579 55 55 www.thermo.com

You might also like