SPECTROPHOTOMETER
AND
COMPONENTS
SPECTROPHOTOMETER
S
S
Components of Spectrophotometer
4
Medium & high pressure Incandescent Tungston or
Mercury lamp, emits continum Tungston-iodide
Spectral distribution
Temperature
2 Monochromator & its function
Colored Filters
Interference Filters
Prism
Diffraction gratings
1. Colored filters
Colored
Light ray enters Low transmittance
filter
colored filter
Destructive Interference
waves cancel each other
3. Prism: Refraction of light waves
Diffraction grating mechanism of operation
Degree of bending
depends on wavelength
In phase wave front
Wave fronts form reinforce each other
& come out
Not in phase wave fronts
cancel each other
Results in complete
spectra
Diffraction grating consisting
of many parallel groves
3 Sample cell/ Flowcell
4 Photodetector
1. Photocell: Structure & Function
Silver
Selenium
4 Advantages of Photocell
3 Inexpensive
4 Durable
3 Disadvantages of Photocell
1 Photocurrent is linear
1 Excellent linearity
2 Good Speed
3 Small size
QUALITY ASSURANCE
OF
SPECTROPHOTOMETER
QUALITY ASSURANCE
S
5 Checks for Quality Assurance
1 Absorbance Correction
2 Wavelength Accuracy
3 Stray Light
4 Linearity
5 Lipemia
Absorbance must be blanked with a reference solution for single
beam spectrophotometers
4 reference solutions for blanking single beam
spectrophotometers
Water
Air (enzymes)
Sample (bilirubin)
Double beam spectrophotometers permit automatic
correction of sample and reference absorbance
Wavelength indicated on control dial must match
actual wavelength passed by monochromator
PM Tube
Max Abs
Display
340 nm
Max Wavelength matched
Control Dial
Stray light refers to wavelengths outside the band
transmitted by monochromator
Grating
Absorbance
PM Tube
Display
Control Dial
3 most common causes of stray light are;
Cutoff filter
Grating
Absorbance
PM Tube
Display
0.9
0.8
0.7
0.6
0.5
0.4
0.3
0.2
0.1
0
0 2 4 6 8 10 12 14 16 18 20 22 24 26 28
Concentration g/dL
Use of diluted colored solutions to check linearity
Filter
Grating
PM Tube
Max Abs
Display
340 nm
Control Dial
High centrifugation clears lipemia of a turbid sample
(14000 RPM)
14000 RPM
25 min