Digitoxin
Digitoxin
REF 03002659 122 100 tests R2 Digitoxin~Ru(bpy)2+3 (black cap), 1 bottle, 9 mL:
• Indicates analyzers on which the kit can be used Digitoxin, labeled with ruthenium complex 4.0 µg/L; phosphate
buffer 100 mmol/L, pH 6.8; preservative.
MODULAR
Elecsys 2010 ANALYTICS cobas e 411 cobas e 601 cobas e 602 Precautions and warnings
E170 For in vitro diagnostic use.
Exercise the normal precautions required for handling all laboratory reagents.
• • • • • Disposal of all waste material should be in accordance with local guidelines.
Safety data sheet available for professional user on request.
English Avoid the formation of foam with all reagents and sample types
(specimens, calibrators, and controls).
Intended use
Immunoassay for the in vitro quantitative determination of digitoxin in human Reagent handling
serum and plasma. The measurement is used for identifying and treating The reagents in the kit have been assembled into a ready-for-use
digitoxin intoxication and monitoring digitoxin levels to support a therapy. unit that cannot be separated.
The electrochemiluminescence immunoassay “ECLIA” is intended for All information required for correct operation is read in via the
use on Elecsys and cobas e immunoassay analyzers. respective reagent barcodes.
Summary1,2,3,4,5,6,7 Storage and stability
Digitoxin belongs to the family of cardio-active glycoside drugs. Its exact Store at 2-8 °C.
mode of action has not been fully explained. A dose of digitoxin eventually Store the Elecsys Digitoxin reagent kit upright in order to ensure complete
increases the intracellular concentration of Ca2+, which is extremely important availability of the microparticles during automatic mixing prior to use.
for cardiac muscle function. As a result, the heart’s ability to contract Stability:
increases and its contraction amplitude increases, resulting in a greater
volume of blood being transported with each beat of the heart. unopened at 2-8 °C up to the stated expiration date
Digitoxin is used to treat cardiac insufficiency. Additional indications include the after opening at 2-8 °C 12 weeks
treatment of pronounced arterial hypertension in older patients, in preoperative on the analyzers 4 weeks
settings for hypertension patients with coronary insufficiency, and for treatment
Specimen collection and preparation
of angina pectoris patients with an enlarged heart and a tendency for distolic
Samples for digitoxin analyses should preferably be collected
ventricular pressure to increase. Digitoxin in contraindicated in patients with
8-24 hours after administration of the drug.1
severe ventricular arrhythmia and certain forms of pericarditis.
Only the specimens listed below were tested and found acceptable.
The therapeutic range of digitoxin is approx. 13-39 nmol/L (10-30 ng/mL).3 In
Serum collected using standard sampling tubes. Some tubes containing
this case, digitoxin exhibits much stronger protein binding than digoxin. Since
separating gel are not suitable for use in therapeutic drug monitoring;
the digitalis effect depends on a number of factors, the therapeutic and the
note the data provided by the manufacturer.
toxic concentration range can overlap.2 Serum values should therefore be
interpreted only within the context of the entire clinical picture. Li-, Na-, NH+4-heparin, K3-EDTA, and sodium citrate plasma.
Digitoxin is eliminated with a half life of 6-8 days. It is metabolized largely in the Criterion: Recovery within 90-110 % of serum value or slope 0.9-1.1 + intercept
liver. In this process, digoxin is produced out of about 10 % of the administered within < ± 2 x analytical sensitivity (LDL) + coefficient of correlation > 0.95.
dose. Approximately 30 % of the digitoxin administered is eliminated renally. Stable for 1 week at 2-8 °C, 6 months at -20 °C.8,9 Freeze only once.
Determination of the serum concentration is indicated, e.g. for therapy The sample types listed were tested with a selection of sample collection tubes
monitoring, for monitoring a patient’s dosing regimen, for confirming that were commercially available at the time of testing, i.e. not all available
suspicion of intoxication, in patients with changed pharmacokinetics tubes of all manufacturers were tested. Sample collection systems from
or unknown premedication, and in patients in which a digitalis effect various manufacturers may contain differing materials which could affect
was not detected in an ECG (electrocardiogram). the test results in some cases. When processing samples in primary tubes
(sample collection systems), follow the instructions of the tube manufacturer.
Test principle Centrifuge samples containing precipitates before performing the
Competition principle. Total duration of assay: 18 minutes. assay. Do not use heat-inactivated samples. Do not use samples
• 1st incubation: By incubating the sample (15 µL) with a digitoxin-specific and controls stabilized with azide.
biotinylated antibody, an immunocomplex is formed, the amount of which Ensure the patients’ samples, calibrators, and controls are at ambient
is dependent upon the analyte concentration in the sample. temperature (20-25 °C) before measurement.
• 2nd incubation: After addition of streptavidin-coated microparticles Because of possible evaporation effects, samples, calibrators, and controls
and digitoxin labeled with a ruthenium complexa, the still-vacant sites on the analyzers should be measured within 2 hours.
of the biotinylated antibodies become occupied, with formation of an
antibody-hapten complex. The entire complex becomes bound to the Materials provided
solid phase via interaction of biotin and streptavidin. See “Reagents - working solutions” section for reagents.
• The reaction mixture is aspirated into the measuring cell where the Materials required (but not provided)
microparticles are magnetically captured onto the surface of the • REF 03002667122, Digitoxin CalSet, for 4 x 1 mL
electrode. Unbound substances are then removed with ProCell.
• REF 04917049190, PreciControl Cardiac II, for 2 x 2 mL each
Application of a voltage to the electrode then induces chemiluminescent
of PreciControl Cardiac II 1 and 2
emission which is measured by a photomultiplier.
• REF 11732277122, Diluent Universal, 2 x 16 mL sample diluent or
• Results are determined via a calibration curve which is
REF 03183971122, Diluent Universal, 2 x 36 mL sample diluent
instrument-specifically generated by 2-point calibration and a
master curve provided via the reagent barcode. • General laboratory equipment
a) Tris(2,2’-bipyridyl)ruthenium(II)-complex (Ru(bpy)2+ • Elecsys 2010, MODULAR ANALYTICS E170 or cobas e analyzer
3 )
Accessories for Elecsys 2010 and cobas e 411 analyzers:
Reagents - working solutions
• REF 11662988122, ProCell, 6 x 380 mL system buffer
M Streptavidin-coated microparticles (transparent cap), 1 bottle, 6.5 mL:
Streptavidin-coated microparticles 0.72 mg/mL; preservative. • REF 11662970122, CleanCell, 6 x 380 mL measuring cell cleaning solution
R1 Anti-digitoxin-Ab~biotin (gray cap), 1 bottle, 9 mL: • REF 11930346122, Elecsys SysWash, 1 x 500 mL washwater additive
Biotinylated polyclonal anti-digitoxin antibody (sheep) 0.5 mg/L; • REF 11933159001, Adapter for SysClean
phosphate buffer 100 mmol/L, pH 6.8; preservative. • REF 11706802001, Elecsys 2010 AssayCup, 60 x 60 reaction vessels
Digitoxin
Digitoxin
Elecsys 2010 and cobas e 411 analyzers References
Repeatabilityb Intermediate precision 1. Oellerich M. Pharmaka (Drug monitoring). In: Thomas L (ed.). Labor
Sample Mean SD CV SD CV und Diagnose, TH-Books, Frankfurt, 5. edition, 1998:1174. Englisch:
nmol/L ng/mL nmol/L ng/mL % nmol/L ng/mL % Clinical Laboratory. 1st English Edition 1998:1151.
HSc 1 4.15 3.17 0.236 0.180 5.7 0.271 0.207 6.5 2. Jortani SA, Valdes R Jr. Digoxin and Its Related Endogenous Factors
HS 2 33.9 25.8 0.769 0.587 2.3 0.827 0.631 2.4 Critical Reviews. Clin Lab Sci 1997;34(3):225-274.
3. Valdes R, Jortani SA, Gheorghiade M. Standards of laboratory practice:
HS 3 73.6 56.2 1.262 0.963 1.7 1.936 1.478 2.6
cardiac drug monitoring. Clin Chem 1998;44(5):1096-1109.
PC Ud1 26.5 20.3 0.676 0.516 2.6 0.804 0.614 3.0
4. Smith TW. Pharmacokinetics, Bioavailability and Serum Levels of
PC U2 53.0 40.4 0.968 0.739 1.8 1.592 1.215 3.0 Cardiac Glycosides. J Am Coll Cardiol 1985;5:43A-50A.
b) Repeatability = within-run precision 5. Doherty JE. Digitalis glycosides. Pharmacokinetics and their clinical
c) HS = human serum
d) PC U = PreciControl Universal implications. Ann Internal Med 1973;79:229-238.
6. Perrier D, Mayerson M, Marcus FI. Clinical pharmacokinetics of
MODULAR ANALYTICS E170, cobas e 601 and cobas e 602 analyzers digitoxin. Clin Pharmacokin 1977;2:292-311.
Repeatability Intermediate precision 7. Beller GA, Smith TW, et al. Digitalis Intoxication: A Prospective Clinical
Sample Mean SD CV Mean SD CV Study with Serum Level Correlations. New Engl J Med 1971;284:989-997.
nmol/L ng/mL nmol/L ng/mL % nmol/L ng/mL nmol/L ng/mL % 8. Young DS. Effects of Preanalytical Variables on Clinical Laboratory
Tests, 2nd editon, 1997:3-180.
HS 1 8.45 6.45 0.19 0.15 2.3 8.43 6.43 0.34 0.26 4.0
9. Guder WG, Narayanan S, Wisser H, Zawta B. List of Analytes;
HS 2 29.2 22.3 0.62 0.47 2.1 29.2 22.3 0.76 0.58 2.6
Preanalytical Variables. Brochure in: Samples: From the Patient to the
HS 3 79.0 60.3 1.20 0.91 1.5 76.9 58.7 1.90 1.45 2.5 Laboratory. GIT-Verlag, Darmstadt;1996:8. ISBN 3-928865-22-6:32/33.
PC U1 22.0 16.8 0.52 0.39 2.3 22.2 17.0 0.61 0.47 2.8 10. Passing H, Bablok W, et al. A General Regression Procedure for Method
PC U2 45.9 35.0 0.78 0.60 1.7 47.0 35.9 1.33 1.02 2.8 Transformation. J Clin Chem Clin Biochem 1988;26:783-790.
Method comparison For further information, please refer to the appropriate operator’s manual
A comparison of the Elecsys Digitoxin assay (y) with CEDIA Digitoxin (x) for the analyzer concerned, the respective application sheets, the product
using clinical samples gave the following correlations (ng/mL): information, and the package inserts of all necessary components.
Number of samples measured: 82
Passing/Bablok regression10
Slope: 0.95 (95 % confidence range: 0.93-1.11)
Intercept: -0.30 (95 % confidence range: -2.88-0.08) COBAS, COBAS E, ELECSYS and MODULAR are trademarks of Roche. Other brand or product names
are trademarks of their respective holders. INTRALIPID is a trademark of Fresenius Kabi AB.
Coefficient of correlation: 0.912 Significant additions or changes are indicated by a change bar in the margin. Changes to reagent barcode
test parameters which have already been read in should be edited manually.
The sample concentrations were between approx. 6.5 and © 2010, Roche Diagnostics
55 nmol/L (approx. 5 and 42 ng/mL).
Analytical specificity
For the polyclonal antibody used, the following cross-reactivities were found: Roche Diagnostics GmbH, Sandhofer Strasse 116, D-68305 Mannheim
www.roche.com
Concentrations tested Cross reactivity
ng/mL %
Cortisol 2000 n.d.e
Corticosterone 2000 n.d.
DHEA-S 2000 n.d.
Prednisolone 2000 n.d.
Ouabain 2000 1.1
Prednisone 1000 n.d.
Estradiol 200 n.d.
Progesterone 200 n.d.
Proscillaridin 5000 3.6
k-strophanthin 500 40
Testosterone 200 n.d.
Digitoxigenin-mono-digitoxoside 40 153
Digitoxigenin-bis-digitoxoside 40 146
Digitoxigenin 40 220
Dihydrodigoxin 25 n.d.
Digoxigenin-mono-digitoxoside 25 n.d.
Digoxigenin-bis-digitoxoside 25 n.d.
Digoxin 25 n.d.
Digoxigenin 25 n.d.
e) n.d. = not detectable; cross-reactions below the lower detection limit