Professional Documents
Culture Documents
Introduction of
optical instrument
Remarks:
1.Source
2.λ selector
3.Container
4.Radiation
detector
5.Readout
To select emission λ
Sample fixed
at right angle / Fluorescence
To select excitation λ
to avoid
scattering
Prepared by Dr.Kavirajaa, Ph.D., AMRSC
Sources of Radiation
Must be
optically transparent in wavelength range of interest
chemically inert
UV
quartz
GOAL
To allow only specific λ to reach the detector at
any given time
Monochromator (Prism)
Allow selection of a range of wavelength
An alternative to filters
Capable of producing many different wavelengths
Wavelength of interest is selected by proper refraction
angle from a prism or reflection angle from a grating.
Allows wavelengths to be scanned as function of time.
a continuous measurement
Refraction at
the two faces
results in
angular
dispersion of
the radiation
k = calibration sensitivity
Real transducers
Exhibits a small, constant response in the absence
of radiation
Dark current (kd)
S = k P + kd
S = current/voltage output of transducer
k = calibration sensitivity
kd = dark current which is usually constant over
short measurement periods.
Equips with a compensating circuit
Reduces kd to zero
S=kP
Prepared by Dr.Kavirajaa, Ph.D., AMRSC
Types of transducers
1. Two types:
a.Photon transducers (photoelectric or quantum
detectors
i. Responds to photons
b.Thermal
ii. Responds to heat
Note: The distinction between photon and heat transducers is important because shot
(Skoog, Pg: 112) noise often limits the behavior of photon transducer and thermal noise
often limits thermal transducers.
For measurement of
UV possess enough energy
to cause photoemission
Visible
to occur
Near IR
Phototube (PT)
Radiation causes emission of electrons from a photosensitive solid surface
1. Consists of semicylindrical
cathode: Supports a layer of
photoemissive material that tend to
emit electron when it is radiated.
Variety of photoemissive materials:
highly sensitive, red sensitive, UV
sensitive, flat response
2. Wire anode:
Sealed in an evacuated transparent
envelope
Thermocouples
Bolometers
Pyroelectric transducers
Speed
Ease and convenience
Skill required by operators
Cost
Availability of equipment
Per sample cost
Absolute Error = (X – m)
(X m)
Relative Error (%) = 100
m
where: X = The experimental result
m = The true result
All Methods, except counting, contain errors – do not know “true” value
Absolute
n Standard Deviation (SD):
SD ( X
i 1
i X ) 2
/(n 1)
Calibration Curve: A plot of the result or signal vs. the known amount of a
known compound or property (standard) being measured.
1000000
peak area
800000
600000
400000
200000
0
0 10 20 30 40 50 60 70 80 90
conce ntration (ppm )
40
30
Method B
20
10
0
0 2 4 6 8 10 12
Prepared by Dr.Kavirajaa, Ph.D., AMRSC
Concentration (mM)
Selectivity: degree to which the method is free from interference by other
species in the sample
70
No method is totally free from
60 interference from other species.
50
Selectivity coefficient (k):
40
Species A
kB,A = mB/mA
30
20
Species B Relative slopes of calibration
10
curves indicate selectivity:
0
0 2 4 6 8 10 12
S = mA(cA + kB,Acb) + Sbl
Concentration (mM)
1) Multiple determination of
blank samples.
noise 2) Estimation of best-fit to
calibration curves
Sbl
2.0 0.150
4.0 0.294
6.0 0.434
8.0 0.570
10.0 0.704
A serum sample gave an absorbance of 0.350. Find the glucose concentration and its
standard deviation, calibration sensitivity, detection limit and dynamic range.