Professional Documents
Culture Documents
and Instrumentation
Maria Margarette A. Dolor, RMT
Electromagnetic Energy
• Described as photons of
energy travelling in waves
characterized by their
frequency and
wavelength.
The distance between two successive peaks
WAVELENGTH
Frequency
• Is the number of vibrations of wave motion
per second
• The lower the wave frequency, the longer
the wavelength
• The wavelength is inversely related to
frequency and energy.
Spectrometry
spectrophotometry, Luminescence
atomic absorption, fluorescence and
and mass chemiluminescence
FOUR BASIC spectrometry [MS]
DISCIPLINES OF
ANALYTICAL
CHEMISTRY Electroanalytic
methods
Chromatography
gas, liquid, and thin
electrophoresis,
layer
potentiometry, and
amperometry
Kinds of Colorimetry
• Visual Colorimetry– uses the eyes in
determining end point
• Photoelectric Colorimetry
–2 Categories
✓Spectrophotometry
COLORIMETRY –(Spectrophotometric measurements)
✓Filter photometry
–(Photometric measurements:
measurement of light intensity
independent of wavelength)
SPECTROPHOTOMETRY
• is a technique used to measure solute concentration by measuring
light intensity in a narrower wavelength.
a. Power supply e. Exit slit
COMPONENTS OF A b. Light source f. Cuvet/sample cell
SPECTROPHOTOMETER c. Entrance slit g. Photodetector
d. Monochromator h. Readout device
BEER’S LAW
A = abc = 2-log%T
Light Source
• Provides polychromatic light and must generate sufficient
radiant energy or power to measure the analyte of interest.
• Light is divided into visible and invisible light. Visible light
falls in between, with the color violet at 400 nm and red at
700 nm wavelengths being the approximate limits of the
visible spectrum.
• The most common source of light for work in the visible and
near-infrared regions is the incandescent tungsten or
tungsten-iodide lamp.
• The lamps most commonly used for ultraviolet (UV) work
are the deuterium discharge lamp and the mercury arc
lamp.
Two Types of Light Source
a. Continuum source- emits radiation that changes in intensity. It is
widely used in the laboratory.
oTungsten light bulb- commonly used light source in the visible and
near infrared region. Does not supply sufficient radiant energy for
measurements below 320nm.
oDeuterium lamp- routinely used to provide UV radiation in
analytic spectrometers. More stable and has a longer life than a
hydrogen lamp.
oXenon Discharge lamp- produces a continuous source of
radiation. It cover both the UV and visible range
Two Types of Light Source
b. Line source- emits limited radiation and wavelength
oMercury vapor lamps and Sodium vapor lamps- exits narrow bands
of energy at well defined places in the spectrum (UV and visible)
oLight Amplification by Stimulated Emission of Radiation (LASER)-
transform light of various frequencies into an intense, focused, and
nearly non-divergent beam of monochromatic light.
ENTRANCE SLIT
• Minimizes unwanted or stray light and prevents the entrance of
scattered light into the monochromator system
MONOCHROMATOR
• It isolates specific or individual or desired wavelength of light.
• Monochromatic light - light radiation of a single wavelength
Types of Monochromators
• Colored Glass Filter
• least expensive, pass a relatively wide band of radiant energy and have a low
transmittance of the selected wavelength. Although not precise, they are
simple, inexpensive, and useful.
Types of Monochromators
Interference Filters
• produce monochromatic light based on the principle of constructive
interference of waves
• Two pieces of glass, each mirrored on one side, are separated by a
transparent spacer that is precisely one-half the desired wavelength.
Prism
• Wedge-shaped pieces of
glass, quartz or sodium
chloride
• A narrow beam of light
focused on a prism is
refracted as it enters the
more dense glass.
• The prism can be rotated,
allowing only the desired
wavelength to pass through
an exit slit.
Types of Monochromators
Diffraction Gratings
• most commonly used,
consists of many parallel
grooves (15,000 or 30,000
per inch) etched onto a
polished surface.
• based on the principle that
wavelengths bend as they
pass a sharp corner
• better resolution than
prism
Types of Monochromators
EXIT SLIT
• Controls the width (bandpass) of light beam- allows only a narrow
fraction of the spectrum to reach the cuvette
SAMPLE CELL
• absorption cell/ analytical cell/ cuvette, may be round or square
• Holds the solution whose concentration is to be measured
• Path length is 1 cm, however, to increase sensitivity, some cuvettes
are designed to have path length of 10 cm, increasing the absorbance
for a given solution by a factor of 10
Different Types of Cuvette
• Alumina silica glass
• Most commonly used (350 to 2000nm)
• Quartz/ Plastic
• Used for the measurement of solution requiring visible and ultraviolet
spectra
• Borosilicate glass
• Soft Glass
PHOTODETECTOR
• convert the transmitted radiant energy into an equivalent amount of
electrical energy
Types of Photodetector
Barrier Layer Cells / Photocell / Photovoltaic Cell
• least expensive, composed of a film of light-sensitive material,
frequently selenium, on a plate of iron
• require no external voltage source but rely on internal electron
transfer
• used mainly in filter photometers with a wide bandpass
• inexpensive and durable; however, it is temperature sensitive and
nonlinear at very low and very high levels of illumination.
Types of Photodetector
Phototube
• contain a negatively charged cathode and a positively charged
anode enclosed in a glass case
• Has a photosensitive material that gives off electron when light
energy strikes it
• Requires external voltage
Types of Photodetector
Photomultiplier Tube
• Most commonly used detector- measures visible and UV regions, 200
times more sensitive than the phototube
• Has excellent sensitivity and has rapid response- detects very low
level of light
• Should never be used to room light because it will burn out
Types of Photodetector
Photodiode
• Not as sensitive as PMT but with excellent linearity, speed and small
size
• Measures light at a multitude of wavelength- detects less amount of
light
• Most useful as a simultaneous multichannel detector
METER / READ- OUT DEVICE
• It displays output of the detection system
• E.g. Galvanometer, ammeter, LED
2 Types of
Spectrophotometer
A. SINGLE BEAM SPECTROPHOTOMETER
• Simplest type of spectrophotometer
• Designed to make one measurement at a time at one specified
wavelength
B. DOUBLE BEAM SPECTROPHOTOMETER
• Instrument that splits the monochromatic light into two components:
• One beam passes throughout the sample
• Another beam passes through a reference solution or reagent blank
• The additional beam corrects for variation in light source intensity
Two Types of
A. DOUBLE-BEAM IN SPACE- uses
Double-Beam two(2) photodetectors
Spectrophotometer
b. Double-beam in time- uses one
Two Types of photodetector and alternately passes the
Double-Beam monochromatic light throughout the sample
Spectrophotometer cuvet and the reference cuvet using a chopper
or rotating sector mirror
SPECTROPHOTOMETER QUALITY ASSURANCE
• Performing at least the following checks should validate instrument
function: wavelength accuracy, stray light, and linearity.
A. Wavelength Accuracy
• means that the wavelength indicated on the control dial is the actual
wavelength of light passed by the monochromator
• Most commonly checked using standard absorbing solutions or filters
(Didymium or holmium oxide in glass). Some instruments with narrow
bandpass use a mercury vapor lamp to verify wavelength accuracy.
SPECTROPHOTOMETER QUALITY ASSURANCE
B. Stray Light
• any wavelengths outside the band transmitted by the monochromator
• most common causes of stray light are reflection of light from scratches on optical
surfaces or from dust particles
• major effect is absorbance error
• detected by using cutoff filters or liquids, such as NiSO4, NaNO2, and acetone,
absorb strongly at short wave- lengths and can be used in the same way to detect
stray light in the UV range.
C. Linearity
• demonstrated when a change in concentration results in a straight line calibration
curve
• Colored solutions may be carefully diluted and used to check linearity
ATOMIC ABSORPTION SPECTROPHOTOMETRY
• used to measure concentration by detecting the absorption of electromagnetic
radiation by atoms rather than by molecules, measures the light absorbed by atoms
dissociated by heat
• Principle
• Element is not excited but merely dissociated from its chemical bond and place in an unionized,
unexcited, ground state
• Light source:
• Hollow Cathode Tube (a separate lamp is required for each metal, example, a copper hollow-
cathode lamp is used to measure Cu)
• Electrodeless discharge lamps are a relatively new light source for atomic absorption
spectrophotometers.
• Interferences: chemical, matrix and ionization
ATOMIC ABSORPTION SPECTROPHOTOMETRY
• DISADVANTAGES
• Inability of the flame to dissociate samples into free atoms (phosphate may interfere
with calcium analysis by formation of calcium phosphate)
• SOLUTION: Lanthanum or strontium is added to samples to form stable complexes with phosphate
• ionization of atoms following dissociation by the flame
• SOLUTION: reduce the flame temperature
• Matrix interference, due to the enhancement of light absorption by atoms in organic
solvents or formation of solid droplets as the solvent evaporates in the flame
• SOLUTION: pretreatment of the sample by extraction
COMPONENTS OF THE AAS
• BURNER – uses flame to dissociate the chemical bonds and form free,
unexcited atoms.
TYPES:
a. Total consumption burner – flame is more concentrated and can be made
hotter, thus lessening chemical interferences.
• produces large droplets in the flame and produces a high acoustical noise.
b. Premix burner –gases are mixed and the sample is atomized before entering
the flame and the large droplets go to waste and not in the flame.
• It has less noisy signals with longer path length and greater absorption and
sensitivity.
• flame is less hot and therefore cannot dissociate metal complexes.
• MONOCHROMATOR – selects the desired
COMPONENTS wavelength from a spectrum of wavelength which
could either be a prism or diffraction gratings.
OF THE AAS • ATOMIZER(NEBULIZER/GRAPHITE FURNACE) is
used to convert ions to atoms
• CHOPPER- is used to modulate the light source
• DETECTOR – uses photomultiplier tubes to
measure the intensity of the light signal.
• READ OUT DEVICE
FLAME EMISSION PHOTOMETRY
• measures light emitted by excited atoms, was widely used to
determine concentration of excited atoms such as Na+, K+, or Li+.
• Principle: Excitation of electrons from lower to higher state energy
• Light Source: Flame
• Method: Indirect Internal Standard Method
• Internal Standard: Lithium/Cesium (corrects variations in flame and
atomizer characteristics)
COMPONENTS OF THE FLAME PHOTOMETER
GASES – using a mixture of hydrogen and oxygen gas,
(acetylene, propane or natural gas)
ATOMIZER OR BURNER – breaks up the solution into finer
droplets so that the atom will absorb heat energy from the
flame and get excited.
COMPONENTS OF THE FLAME PHOTOMETER
INTERFERENCE FILTERS AS MONOCHROMATOR
• Na filter – transmit yellow light (589 nm)
• K filter – transmit violet light (767 nm)
• Lithium – transmit red light (761 nm)