Professional Documents
Culture Documents
Presented by
Mounik Rout
M.Pharm (Pharmaceutical technology)
Guided by
Dr. Sasmita Kumari Acharjya
• 1.Wavelength(λ) - The distance in between both the crests and troughs is wavelength .
• The unit used for wavelength are - Å , nm , μm .
• 2.Wavenumber(̄) – Number of waves passing through the space of 1cm is called wavenumber .
• ̄= 1/λ
• The unit is Cmˉ¹
Fig4. frequency
• 4.Energy(E) :- Electromagnetic radiation have some energy .
• When wavelength increases , energy of EMR decreases , & when wavelength
decreases energy of EMR increases .
• E = h (where h is planck’s constant=6.626 x 10ˉ²⁷ erg sec)
• E= h x C/λ (=C/λ)
Different EMR and their wavelength
Electromagnetic Radiation Wavelength
Table.1
1.30-
1.20-
1.10-
1.00-
0.90-
0.80-
0.70-
0.60-
0.50-
0.40-
0.30-
0.20-
0.10-
0.00-
4. n * - When electron from n molecular orbital get excited in * orbital , the transition
called n * transition .
• It requires lowest energy as comparison of all 4 transition .
• Chromophore- These covalently bonded moieties with any compound which are responsible for
absorption of electromagnetic radiation in the UV-Visible region .
Example of chromophore : Aldehyde , Ethylene, Carbonyl .
1. Chromophore with * : ex - >C=C< , -CC-
2. Chromophore with n * : ex - -N=N-
Whenever chromophore compound will added to any compound the absorption wavelength will
increase .
• Auxochrome – It is defined as any moieties which does not shows any absorption when separated
but when it combines with any chromophore it increases the wavelength towards longer wavelength
by
formation of a new chromophore .
ex: -OH , NH₂ ,OR , etc .
• Absorption :-
• It is a quantitative evaluation .
• Spectroscopy is based upon absorption theory.
Incident Radiation(Iₒ)
Lambert’s Law :- when a beam of monochromatic radiation is passed through the absorbing medium
then the decrease in the intensity of radiation will directly proportional to the thickness of solution .
• Equation A Log(Iₒ/Iₜ) L
A L L= Thickness of solution / Thickness of cuvette
A= L = Molar absorptivity co-efficient (unit= l/mol.cm)
Beer’s law :- When a beam of monochromatic radiation is passed through the solution then the
decrease in the intensity of radiation will be directly proportional to the concentration of the solution .
• Equation A Log(Iₒ/Iₜ) C
A C C = Concentration of the solution
A= C
• Beer Lambert’s Law –
• When a beam of monochromatic radiation is passed through the absorbing medium ,
then the decrease in the intensity of radiation is directly proportional to the thickness
as well as the concentration of the solution .
Detector 2
Sample
Lens
Amplifier
Gratings
Collimating lens
Radiation Recorder
Source Entrance slit
Monochromator
• Instrumentation
• There are 2 type of spectrophotometer used in UV-Visible spectroscopy .
1. Single beam
2. Double beam
• Double beam spectrophotometer – The instruments used in double beam
spectrophotometer are ;
1. Radiation source
2. Monochromator : Entrance slit
Collimating lens
Gratings
Reflecting lens
Exit slit
3. Chopper 5. Cuvette
4. Mirror 6. Detector
7. Amplifier
8. Recorder
• Radiation source – 1. Hydrogen-Deuterium lamp (ranges from 200-375nm)
2. Tungsten lamp (ranges from 375 -800nm for uv-visible spectroscopy)
• Monochromator – used to convert polychromatic radiation beam to monochromatic
radiation .
• Collimating lens – This converts the non-parallel beams to parallel beam .
• Gratings – This is used in place of prism now-a-days because it contains 1200-1400 groves
per mm .
• Exit slit – through this the required wavelength of beam passes to Chopper .
• Chopper – It equally divide the radiating beam in 2 parts in to the mirror .
• Mirror – it reflects the radiation in to both Cuvette containing reference and sample .
• Cuvette – it is made up of Quartz and glass material .
1. Quartz is used because it have very smooth surface and it don’t absorb the UV
radiation .
2. Glass is used in visible spectroscopy & it has UV absorption capacity .
• Detector – Detector used to detect the wavelength coming through the cuvette & it is of 3 types :
1. Photomultiplier tube – It multiplies the photons rays of the radiation .
c at h
ode
Photon Anod
e
Electron
Dynode
Amplifier connected from the cathode and anode and amplifies the radiation .
2. Photodiode Array – In this detector there is presence of diodes , which converts
radiation into electronic signal .
• It is mainly used in HPLC , and its range is up to 200-800nm .
• 3. Charged coupled diode – Silicon surface is used to catch photon from the radiation of
UV-Visible source & converts radiation in to digital signal .
• Recorder – It records the amplified wavelength and shows the UV-Visible spectrum .