Professional Documents
Culture Documents
X-ray Spectroscopy
2015
Thomas Slater
School of Materials
List of Contents
ABSTRACT .............................................................................................................................................. 8
Declaration ............................................................................................................................................... 9
Acknowledgements ................................................................................................................................ 11
Publications ............................................................................................................................................ 12
1. Introduction ..................................................................................................................................... 14
1
2.4. Spatial Resolution in the STEM ........................................................................................ 38
................................................................................................................................................. 47
4. Electron Tomography...................................................................................................................... 64
................................................................................................................................................. 87
2
6.1. Introduction ..................................................................................................................... 124
............................................................................................................................................... 126
3
List of Figures*
Figure 2.1. Illustration of the components of a scanning transmission electron microscope and
the associated signals, energy dispersive X-ray (EDX), bright field (BF), high angle annular dark
Figure 2.2. Diagram illustrating the principal of reciprocity between a) CTEM and b) STEM for
elastic scattering. The source and detection planes are interchanged and the position of the
Figure 2.3. Example EELS spectrum showing the characteristic features of the zero-loss peak,
the plasmon-loss peaks and core-loss edges that have had their intensity multiplied by 50.
Figure 2.4. a). Diagram of the process in which an incident electron causes a characteristic X-
ray to be emitted from an atom. b) Diagram of the possible electron transitions in the atom with
Figure 2.5. Spectra from a sample containing elements Al, O and Co from vacuum (blue),
carbon film (green) and the sample (red). There are few counts from vacuum, demonstrating
Figure 2.6.. Schematics of cross-sections of EDX detectors. a) Si(Li) detector, in which the
anode covers the whole back face. b) SDD detector, in which concentric p-type rings promote
Figure 2.7. Plot of normalised Au Mα peak intensity from an AgAu nanocube as a function of tilt
Figure 4.1. Diagram of two coordinate systems related by a set of Euler angles. The blue
coordinate system is equivalent to the coordinate system of the object and the red coordinate
4
j
Figure 4.2.a) Illustration of the projection of a small dot f(x,z) on to a coordinate system (x ,θ) to
j
form a projection pj(x ,θj). b)The sinogram associated with a tilt series obtained for a point; the
j
slice highlighted in the sinogram corresponds to the projection in (a), p j(x ,θj).
Figure 4.3. Representation of methods for tilt axis determination, using a sample of Au and Pd
nanoparticles on graphene. a) Projection at 0° tilt displaying the lines (in red) at which
reconstructions are performed to display the arcing in (c-e). b) Summation of the entire tilt series,
displaying streaking in the direction perpendicular to the tilt axis at an angle α to the horizontal.
c-e) Reconstructions of individual slices (i-iii) represented in (a) for c) a tilt axis shifted from the
correct orientation by 40 pixels, d) a tilt axis tilted from the correct orientation by 15° and e) a tilt
axis approximately at the correct orientation. The sample represented here is Au and Pd
Figure 4.4. Backprojection of an object along a varying number of directions. a) Image of a two-
j
Figure 4.5. a) The projected intensity p (x) of an object f(x,z) at an angle of θj. b) An illustration
of the angular sampling of an object in Fourier space under a single tilt acquisition scheme with
a missing wedge. The red line in (b) corresponds to the Fourier transform of the projected
j
intensity p (x) at the same angle θj.
Figure 4.6. Reconstructions of the object in (a) from projections at 1° intervals using b) an
unweighted backprojection operator from projections over the full tilt range, c) a weighted
backprojection operator from projections over the full tilt range and d) a weighted backprojection
operator with projections from between ± 70° displaying the missing wedge.
Figure 7.1. a) SEM image within the plasma-FIB dual beam instrument showing the pillar
sample after annular milling. b) The 30 μm diameter pillar attached to a pin for further analysis.
c) An SEM image of the cross-sectional slab cut from the nano-CT pillar before liftout. d) An ion-
beam image of one polished face of the slab attached to a TEM grid. e) Ga ion-beam image of
the “block” cut from the cross-sectional slab and attached to the lift-out needle. f) SEM image of
5
Figure 7.2. X-ray nanotomography of a pillar of stainless steel 316. a) Volume visualisation
displaying an isosurface rendering of the outer surface of the pillar, traces along the three grain
boundaries and the surface of the segmented cavities. b) Slice through the X-ray
nanotomography reconstruction displaying the three grain boundaries and the intensity changes
Figure 7.3. STEM-EDX elemental maps of the STEM needle displaying the intensity of X-ray
counts associated with the a) Fe Kα, b) Cr Kα, c) Ni Kα, d) Mo Lα, e) Nb Kα, f) O Kα, g) C Kα
and h) Si Kα peaks. The austenite phase at top and bottom is associated with Fe, Ni and to a
lesser extent Mo and Mn, the ferrite phase is predominantly Fe, the M23C6 carbides contain Cr,
Mn, Mo, C and O, the G-phase precipitate contains Ni, Mo and Si. The surface of the needle
has high carbon and silicon intensities that are associated with organic contamination.
Figure 7.4. Visualisation of the three dimensional reconstruction of elemental distributions and
the associated segmented phases. Volume visualisations of a) the Fe Kα signal (red), b) the Cr
Kα signal (yellow) and c) the Ni Kα signal (blue). Surface visualisations of the segmented
phases showing the d) ferrite phases (green), e) carbide precipitates (purple) and f) austenite
Figure 7.5. Images displaying the faces of the extracted block and the correlation between a)
SEM images and b) slices through the CT data of faces i-iv. c) Faces (i) and (ii) displayed in
three dimensions from the CT data with the X-ray CT coordinate system indicated.
Figure 7.6. a) Visualisation of the STEM-EDX pillar with respect to the SEM image of face (iii)
(flipped) and b-c) with respect to the top-down SEM image taken after annular milling with (c)
and without (b) segmented cavities. The coordinate systems indicated all correspond to the X-
ray CT coordinates.
Figure 7.7. Correlation of the orientation of the STEM-EDX needle embedded in to the micro-CT
data (a) with that from the SEM image of the final pillar (b) at the same orientation in the SEM
as in Figure 5ai.
*Please note that the figures for each manuscript are self-contained and
6
List of Abbreviations
2D – Two dimensional
3D – Three dimensional
BF – Bright field
CT – Computed tomography
IR - Infrared
7
Three Dimensional Chemical Analysis of Nanoparticles Using Energy
Dispersive X-ray Spectroscopy
A thesis submitted to The University of Manchester for the degree of Doctor of
Philosophy in the Faculty of Engineering and Physical Sciences
2015
Thomas Slater
School of Materials, The University of Manchester
ABSTRACT
8
Declaration
No portion of the work referred to in the thesis has been submitted in support of an application
for another degree or qualification of this or any other university or other institute of learning.
9
Copyright Statement
i. The author of this thesis (including any appendices and/or schedules to this thesis)
owns certain copyright or related rights in it (the “Copyright”) and s/he has given The
administrative purposes.
ii. Copies of this thesis, either in full or in extracts and whether in hard or electronic copy,
may be made only in accordance with the Copyright, Designs and Patents Act 1988 (as
amended) and regulations issued under it or, where appropriate, in accordance with
licensing agreements which the University has from time to time. This page must form
iii. The ownership of certain Copyright, patents, designs, trade marks and other intellectual
property (the “Intellectual Property”) and any reproductions of copyright works in the
thesis, for example graphs and tables (“Reproductions”), which may be described in this
thesis, may not be owned by the author and may be owned by third parties. Such
Intellectual Property and Reproductions cannot and must not be made available for use
without the prior written permission of the owner(s) of the relevant Intellectual Property
and/or Reproductions.
iv. Further information on the conditions under which disclosure, publication and
commercialisation of this thesis, the Copyright and any Intellectual Property and/or
Reproductions described in it may take place is available in the University IP Policy (see
10
Acknowledgements
Firstly, I would like to express the sincerest gratitude to my supervisor, Dr Sarah Haigh, for all of
her help, support and guidance throughout my studies. I have been lucky to have a supervisor
who takes such genuine interest and care in the progress and well-being of her students.
I would also like to thank all members of the Haigh group who have provided me with a great
deal of support and friendship over my time in the group. In particular, I would like to thank
Edward Lewis and Yiqiang (Kelvin) Chen, with whom I have shared the entirety of my studies
and who have helped me on uncountable occasions. In addition, I would like to thank Dr Eric
Prestat for many helpful discussions on the STEM and his ever present optimism, particularly to
do with the Manchester weather. I am thankful to both Dr Alan Harvey and Matthew Smith for
Thank you to all those who I have had the pleasure of collaborating with throughout the course
of my studies, including but not limited to Dr Pedro Camargo, Dr Nestor Zaluzec, Prof Grace
I would like to extend my thanks to all of my family and friends for all of their support throughout
my studies. Thank you to my parents and sister for their continued support and unconditional
love. My sincerest thanks go to my girlfriend Sarah for her everlasting patience and for her
generosity, care and support. Thank you also to Fred and Henry for their help, and hindrance, in
Thank you to the EPSRC and the School of Materials for their support in funding my studies.
11
Publications
SLATER, T.J.A., CAMARGO, P.H.C., BURKE, M.G., ZALUZEC, N.J. & HAIGH, S.J. (2014).
Function Of Specimen Tilt Angle For Tomographic Data Acquisition In The S/TEM. Journal of
SLATER, T.J.A., MACEDO, A., SCHROEDER, S.L.M., BURKE, M.G., O'BRIEN, P., CAMARGO, P.H.C. &
HAIGH, S.J. (2014). Correlating Catalytic Activity Of Ag-Au Nanoparticles With 3D Compositional
SLATER, T.J.A., CHEN, Y.Q., AUTON, G, ZALUZEC, N.J., HAIGH, S.J. X-Ray Absorption Correction
SLATER, T.J.A., JANSSEN, A., CAMARGO, P.H.C., BURKE, M.G., ZALUZEC, N.J., HAIGH, S.J. STEM-
NEWTON, L., SLATER, T., CLARK, N. & VIJAYARAGHAVAN, A. (2013). Self Assembled Monolayers
(SAMs) On Metallic Surfaces (Gold And Graphene) For Electronic Applications. Journal Of
BARTON, C.W., SLATER, T.J.A., ROWAN-ROBINSON, R.M., HAIGH, S.J., ATKINSON, D. & THOMSON, T.
(2014). Precise Control Of Interface Anisotropy During Deposition Of Co/Pd Multilayers. Journal
BURNETT, T.L., MCDONALD, S.A., GHOLINIA, A., GEURTS, R., JANUS, M., SLATER, T., HAIGH, S.J.,
ORNEK, C., ALMUAILI, F., ENGELBERG, D.L., THOMPSON, G.E. & W ITHERS, P.J. (2014). Correlative
CHEN, Y.Q., SLATER, T.J.A., LEWIS, E.A., FRANCIS, E.M., BURKE, M.G., PREUSS, M. & HAIGH, S.J.
(2014). Measurement Of Size-Dependent Composition Variations For Gamma Prime (Gamma ')
12
HARRIS, P.J.F., SLATER, T.J.A., HAIGH, S.J., HAGE, F.S., KEPAPTSOGLOU, D.M., RAMASSE, Q.M. &
LEWIS, E.A., HAIGH, S.J., SLATER, T.J.A., HE, Z., KULZICK, M.A., BURKE, M.G. & ZALUZEC, N.J.
LEWIS, E.A., SLATER, T.J.A., PRESTAT, E., MACEDO, A., O'BRIEN, P., CAMARGO, P.H.C. & HAIGH,
S.J. (2014). Real-Time Imaging And Elemental Mapping Of Agau Nanoparticle Transformations.
VAN DE LOCHT, R., SLATER, T.J.A., VERCH, A., YOUNG, J.R., HAIGH, S.J. & KROEGER, R. (2014).
13
1. Introduction
Nanoparticles display vastly different properties to their bulk counterparts due to their incredibly
small size (1 – 100 nm). The size of nanoparticles results in a large surface to volume ratio,
which is of particular interest in catalysis, where active catalytic sites are solely related to the
surface of the material (Serp & Philippot, 2013; Xia, et al., 2013). Size effects also promote
properties such as localised surface plasmon resonances in metallic nanoparticles (Hutter &
Fendler, 2004; Raether, 1988; Willets & Van Duyne, 2007) and quantum confinement in
semiconductor quantum dots (Wroggen, 1997), both of which have seen great interest for
optical purposes, such as new generation solar materials (Hillhouse & Beard, 2009; Kim, et al.,
The precise size and shape of nanoparticles has a large impact on many of their properties. As
well as an increase in catalytic performance in general for decreasing nanoparticle size (Laoufi,
et al., 2011; Thiele, 1939), the precise morphology of nanoparticles is important for their
catalytic properties. Nanoparticle catalysis is highly facet dependent, i.e. the catalytic activity is
dependent on the crystal facets available at the surface of nanoparticle (Cuenya, 2013;
Narayanan & El-Sayed, 2004), which provides a key dependence on nanoparticle shape. The
size and shape of plasmonic nanoparticles also has a large effect on their plasmon resonance
wavelength, which offers the potential to tune the wavelength to a desired value (Liz-Marzan,
2006; Willets & Van Duyne, 2007). Thus, accurate characterisation of both size and shape of
In nanoparticles of more than one element, the precise composition and the distribution of
elements both play key roles in the particles’ properties. Nanoparticles of more than one
constituent element have become particularly important for catalytic applications, where the
combination of properties from different elements can improve catalytic properties (Ferrando, et
al., 2008; Wang & Li, 2011). For example, there has been great interest in combining additional
metals with Pt nanoparticles for their use in hydrogen fuel cells where the addition of Ni or Co to
form bimetallic particles acts to reduce the cost of expensive Pt catalysts, enhances the oxygen
reduction reaction and improves the stability through many thousands of reaction cycles (Liu, et
14
al., 2013; Mukerjee & Srinivasan, 1993). In synthesizing multi-element nanoparticles it is vital to
ensure that the surface is catalytically active, either through segregation of the more catalytically
active element or precise control of a core-shell structure so that the electronic distribution is
optimised (Jiang, et al., 2011; Toshima & Yonezawa, 1998). As the crystallography of surface
facets is important for catalytic properties, the composition of each facet is therefore directly
relevant to the catalytic activity of multi-element nanoparticles (Cui, et al., 2012; Shao, et al.,
2006). The distribution of elements within nanoparticles is therefore another key property for
characterisation, although typically more challenging to determine than the size and shape
analysis.
The transition from single-element to multi-element nanoparticles also has large effects on the
plasmonic properties of noble metal nanoparticles. The wavelength and linewidth of localised
surface plasmon resonances are tuneable through the addition of a second metal (Cortie &
McDonagh, 2011; Link, et al., 1999), such that the resonance wavelength can be tuned to
wavelengths of lower absorption in key materials, such as tissue (Jain, 2008). The distribution of
the elements within nanoparticles plays a key role in their plasmonic properties; fully alloyed
nanoparticles display sharper resonance peaks than core-shell systems for example (Prodan, et
al., 2003).
Extending the definition of nanoparticles to include those encapsulated in a solid matrix, often
termed nanoprecipitates, vastly extends the scope of nanoparticle studies. Any secondary
phase of material, distinct from the matrix surrounding it, with dimensions of 1 -100 nm can be
considered a nanoparticle. In fact, characterisation of this type of particle often involves removal
from the matrix via etching, in order to obtain unambiguous information on the particles’
structure and composition without contribution from the surrounding matrix, at which point the
Nanoprecipitates play an important role in many of the properties of bulk materials. For example,
in the case of carbide formation in stainless steels, the formation of intragranular carbides
greatly increases the strength of the steel due to improved creep resistance (Sourmail, 2001;
Yamamoto, et al., 2007). Carbides formed at grain boundaries also play an important role in the
corrosion behaviour of stainless steels, causing Cr depletion in the matrix which lowers the
corrosion resistance (Armijo, 1968; Joshi & Stein, 1972; Tedmon, et al., 1971). Thus,
15
characterising the number, distribution and the composition of precipitates has an important
The size, shape and elemental distribution of nanoparticles can be characterised by a wide
variety of analytical methods. For example, Ultra Violet – Visible (UV-Vis) and Infrared (IR)
spectroscopies (Amendola & Meneghetti, 2009; Haiss, et al., 2007; Higgins, et al., 2003) and X-
ray diffraction (Dorofeev, et al., 2012; Patterson, 1939) can be used to determine nanoparticle
size and shape. X-ray photoelectron spectroscopy (Baer, et al., 2010) and X-ray absorption
spectroscopy (Hwang, et al., 2005) are used to probe the surface and bulk composition of
nanoparticle populations respectively. However, these techniques are only able to obtain
average values from a large ensemble of nanoparticles and do not offer direct visualisation of
Scanning probe techniques, which include atomic force microscopy (AFM) and scanning
tunnelling microscopy (STM), have shown great utilisation in examining individual nanoparticle
size and morphology (Grobelny, et al., 2011; Henry, 2005; Kano, et al., 2015). Characterisation
of individual particles is particularly important due to the large dispersity in properties that is
often found amongst samples of synthesized nanoparticles. However, the nature of scanning
probe techniques means that it is not possible to characterise nanoparticles embedded inside a
material. Examination of chemical variations is possible also with STM, but it is not
nanoparticle size and shape determination straightforward in the scanning electron microscope
(SEM) (Bootz, et al., 2004; Liu, 2005). However, the use of energy dispersive X-ray (EDX)
spectroscopy in the SEM is limited in its resolution, due to the large interaction volume of X-rays
from nanoparticles mounted on a thick substrate (Goldstein, et al., 2003). Techniques utilising
the transmission electron microscope (TEM) and scanning transmission electron microscope
(STEM) are incredibly versatile in characterising the size, morphology and composition of
nanoparticles. In comparison to the use of scanning probe techniques, mapping the extent of
nanoparticle composition is becoming a routine process in the STEM due to the complementary
16
spectroscopic techniques of electron energy loss spectroscopy (EELS) and EDX spectroscopy
The aim of this thesis is to thoroughly investigate the methodology of STEM-EDX tomography in
order to provide three dimensional elemental imaging of nanoparticles. The results included in
this thesis have been published in, or have been submitted to, a number of peer-reviewed
journals and the submitted manuscripts are included in full. These chapters therefore have self-
contained figure numbering and referencing. The information presented in this thesis is the
result of work with a number of collaborators and their contributions are made clear at the start
of each chapter where appropriate. Contributions not assigned to collaborators were performed
by the author.
number of factors influencing the robustness of the technique. In particular, the effect of the
sample holder shadowing the X-ray detectors is investigated and compensation methods are
discussed and implemented. EDX tomography is applied to understand the catalytic behaviour
of AgAu nanoparticles in chapter 6. The three dimensional elemental analysis reveals a change
in the distribution of Ag and Au with overall composition that correlates to their catalytic
properties.
17
2. The Scanning Transmission Electron Microscope
Electron microscopy has become one of the most important materials characterisation
techniques since its inception in the 1930s (Knoll & Ruska, 1932). Its strength comes from the
very small wavelength of the electron, particularly in comparison to that of light. The minimum
resolvable distance between two objects is limited by diffraction to the order of the wavelength
of the illuminating radiation, as was first proposed by Lord Rayleigh in 1879 (Rayleigh, 1879).
wavelengths (on the order of 100 nm). Electron wavelengths are typically on the order of 1 x
-3
10 nm for typical transmission electron microscope energies (100 - 300 keV) and therefore
have the potential to resolve not just nanometre size features but features down to the atomic
level.
Types of TEM are generally split in to two categories, the conventional TEM (CTEM) and the
scanning TEM (STEM). The more well-known technique, the CTEM, is akin to a conventional
light microscope in that a broad parallel illumination is used that can be treated as a plane wave.
In the scanning transmission electron microscope a very finely focussed probe is scanned
across a sample in a pixel-by-pixel manner (Fig 2.1). The two techniques are interrelated by the
principle of reciprocity, which will be explained in the next section, but leads to direct
comparison between imaging modes in the two techniques. In this chapter, all concepts will be
introduced in relation to the STEM but relation of imaging modes back to those of the CTEM is
18
Figure 2.1. Illustration of the components of a scanning transmission electron microscope and
the associated signals: energy dispersive X-ray (EDX) spectroscopy, bright field (BF), high
angle annular dark field (HAADF) and electron energy loss (EEL) spectroscopy.
In short, in the STEM, electrons are extracted from a source in the electron gun and an image of
this source is effectively demagnified to a small spot on the sample by the condenser lens
system and the objective lens. This spot can be raster scanned across the sample by deflection
coils located immediately before the specimen. After undergoing interactions with the sample,
scattering angle. Electron detectors include bright field and annular dark field detectors, as well
as spectrometers for electrons and X-rays (Fig 2.1). One of the great advantages of the STEM
is that a number of different signals can be simultaneously collected by different detectors, with
the detected signals having ideally originated at the scanned position of the beam. This type of
raster scanned approach is particularly useful in spectroscopy, where a full spectrum can be
This chapter introduces the concept of scanning transmission electron microscopy with relation
and the concepts of elastic and inelastic scattering, as well as coherent and incoherent imaging
are introduced. Secondly, imaging modes are discussed, with a particular focus on nanoparticle
19
analysis. As the focus of this thesis is on energy dispersive X-ray spectroscopy, the key
developments of the technique are reviewed that have allowed spectrum imaging of
nanoparticle systems. At the end of the chapter, factors limiting STEM imaging are discussed, in
sections detailing the resolution of the STEM and electron beam damage.
In the STEM, practically all of the detected signals are due to the interaction of the electron
probe with the sample. Electrons interact strongly with matter, with a very short mean free path
compared to photons (on the order of nanometres or tens of nanometres depending on the
electron energy and material (Powell & Jablonski, 2010; Sondheimer, 2001)), which limits the
thickness of the sample that electrons may be transmitted through. Electron interactions are
termed either elastic interactions, in which the transmitted electron does not lose an appreciable
appreciable amount of energy. In reality, elastic interactions also result in a transfer of energy
but, particularly for the small-angle scattering considered in the CTEM or STEM, this energy is
so small when compared to the energy of the beam electrons (< 1eV compared to 100 – 300
keV) that it is not considered (Egerton, 2011a). Both elastic and inelastic interaction
mechanisms are important to different imaging modes in the STEM and are discussed here
independently.
Elastic scattering can be described by considering the Coulombic interaction of the probe
electrons with the potential distribution of the sample. The electron is both a particle and a wave
and therefore attempts to describe this interaction may make use of either the particle or wave
If the electron is considered as a particle, the elastic interaction can be described in the simplest
terms by considering the scattering of the probe electrons by the point potentials of the atomic
nuclei. This scattering is known as Rutherford scattering and has a differential cross section, the
20
where e is the unit electric charge, Z is the atomic number of the nucleus and E0 is the initial
electron beam energy (Williams & Carter, 2009b). Thus, in this scheme, the scattered intensity
2
at a scattering angle θ is proportional to Z of the nucleus. In practice, shielding of the nuclear
potential/charge by atomic electrons modifies this scattering to some degree. This can be
𝜃 2 𝜃 2 𝜃 2
represented by replacing (sin ) with (sin ) + ( 0) in the Rutherford cross section to give
2 2 2
𝑑𝜎(𝜃) 𝑒 4 𝑍2 (2.2)
= 2 ,
𝑑Ω 𝜃 2 𝜃 2
16(𝐸0 )2 ((sin ) +( 0 ) )
2 2
where θ0 is the screening parameter and can be thought of as the scattering angle below which
the electron screening affects the scattering cross section. The screening parameter is also
dependent on Z and in practice acts to reduce the power of the Z dependence, particularly for
low-angle scattering.
The particle description of electron scattering is useful to a certain degree, where the differential
cross section can be thought of as the probability of scattering a single electron by a single
atom to certain scattering angles. In this description we would expect this intensity distribution to
sum for many electrons scattered from an array of atoms so that if each electron could only
some scattering angle it should therefore be possible to include only electrons scattered to a
particular range of angles, the intensity of which will show a clear dependence on the atomic
In actuality, in order to fully describe the intensity distribution at detectors the electron probe
must be treated as a wave, in which the scattering from atomic potentials is described by a
scattering factor. When considering the electron as a wave, we have to take in to account that
the scattering process may be coherent or incoherent in nature. Coherent scattering involves
the summing of the amplitudes of scattered waves, which, in a crystalline lattice, will result in
constructive and destructive interference from points of scatter. This is the basis of both phase
contrast and diffraction contrast in bright field and dark field STEM and CTEM.
21
The intensity of coherent imaging in the STEM, Icoh, can be described by a convolution of the
wavefunction of the probe ψp with the transmission function of the sample φ, which describes
how the sample alters the electron wavefunction, for a probe position Rp (Nellist, 2011):
2
𝐼𝑐𝑜ℎ (𝑹𝑝 ) = |𝜓𝑝 (𝑹𝑝 )⨂𝜙(𝑹𝑝 )| . (2.3)
If the imaging mode can be shown to be incoherent, i.e. the scattering intensities from individual
atoms can be summed, we can approximate the scattering using the Rutherford description.
This is the basis for contrast in bright field imaging of non-crystalline materials, so called mass-
thickness contrast and, as we will expedite later, in high-angle annular dark field imaging. In
contrast to coherent imaging, the intensity of incoherent imaging is given by a convolution of the
intensity of the STEM probe and the object function of the specimen, that describes how the
2 (2.4)
𝐼𝑖𝑛𝑐𝑜ℎ (𝑹𝑝 ) = |𝜓𝑝 (𝑹𝑝 )| ⨂𝑂(𝑹𝑝 ).
The intensity of incoherent imaging is simply the convolution of the square of the object function
with the point spread function of the electron probe, which is the intensity distribution in the
probe.
Another key point to make about purely elastic scattering is that it can be viewed as reversible
with respect to time, i.e. the scattering process is independent of the order of events. This leads
to the principle of reciprocity in the STEM, that the same scattering process occurs, and
therefore the same images are obtained, if the source and detector are exchanged (illustrated in
Fig 2.2). This allows the transfer of the detailed mechanistic understanding of contrast formation
22
Figure 2.2. Diagram illustrating the principal of reciprocity between a) CTEM and b) STEM for
elastic scattering. The source and detection planes are interchanged and the position of the
objective aperture with respect to the sample is reversed. Adapted from (Brown, 1981).
In addition to the elastic scattering described above, probe electrons may also undergo inelastic
the sample. Inelastic interactions include those with phonons, plasmons and the ionisation of
core-level electrons. All of these interactions have the potential to be used spectroscopically,
however, the energies associated with phonon or plasmon modes have historically been too
therefore the most intensely scrutinised inelastic scattering process in the STEM.
Ionisation of atomic electrons is usually described by the Bethe cross-section σI (Bethe, 1930),
as given in equation 2.4, or one of a number of variants of the Bethe cross-section (Powell,
1976):
𝜋𝑒 4 𝑏𝑠 𝑛𝑠 𝑐𝑠 𝐸0 (2.5)
𝜎𝐼 = ( ) log ( ),
𝐸0 𝐸𝑐 𝐸𝑐
where Ec is the energy required to ionise the atom, ns is the number of electrons in the ionised
subshell and bs and cs are known as Bethe parameters and are constant for ionisations of the
23
same sub-shell. Here, a total cross section is presented as opposed to the differential cross
section presented for elastic scattering. The total cross section is generally a reasonable
description of spectroscopic processes in the STEM, for EDX spectroscopy (EDXS) due to the
fact that there is no angular dependence on characteristic X-ray production (Bethe, 1930) and
for EELS when the collection aperture is large enough to collect almost all inelastically scattered
Whilst the interaction of the electron beam with phonons has only recently become a
spectroscopically measurable quality (Krivanek, et al., 2014), it does play an important role in
electron imaging modes in the STEM. Thermal diffuse scattering (TDS), which is usually defined
as interaction of the beam electrons with pre-existing phonons in the crystal lattice, is
particularly important in HAADF imaging where it has been shown that the majority of the
electrons impinging on the detector are from thermal diffuse scattering if a large detector inner
angle is used (Jesson & Pennycook, 1995). TDS in CTEM has conventionally been treated as
“absorption” of electrons, after which the absorbed electrons do not contribute to diffraction or
image formation. Similarly, in the treatment of HAADF images the effect of TDS can be
considered using an absorptive potential approach (Allen, et al., 2003; Findlay, et al., 2003), in
Bright field (BF) images in the STEM are acquired by collecting electrons with a detector
mounted on the optic axis (Fig 2.1). Due to the principle of reciprocity, the bright field technique
in the STEM can be treated in the exact same manner as the more commonly used bright field
imaging in the CTEM. In the strictest sense, true BF imaging in the STEM, in which only
undiffracted electrons are collected, is only achieved when the convergence angle of the probe,
determined by the objective aperture, is smaller than the scattering angle of the first Bragg
scattered disk. When the convergence angle matches this criterion there is no contribution from
phase contrast, just as in CTEM BF imaging. Typically, however, STEM BF imaging is operated
with a large convergence angle that includes information from at least the first Bragg diffracted
disks and phase contrast is generated in a similar manner to high resolution TEM (HRTEM)
24
imaging in the CTEM. Contrast from the bright field detector is therefore due to diffraction
contrast, phase contrast and mass thickness contrast, which will each be discussed here briefly
in turn.
The coherent summing of electron wave amplitudes in a crystalline material results in so-called
diffraction contrast. Changes in contrast due to coherent scattering come from changing sample
thickness or orientation (including effects due to defects or bending) and all come under the
umbrella term of diffraction contrast. Whilst diffraction contrast is a very important contrast
mechanism for traditional CTEM and STEM of thin foils and has extensive associated literature
(Hirsch, et al., 1977; Williams & Carter, 2009a), in the study of nanoparticles it has made little
conducive to diffraction contrast as it may obscure information from mass thickness contrast.
The coherent nature of the electrons impinging on the BF detector also lead to Fresnel fringes
at the edge of structures in the presence of defocus in a similar manner to BF CTEM (Colliex, et
al., 1977). Fresnel fringes can obscure the analysis of the surfaces of nanoparticles, especially
function that has a phase, at a given wavevector, given by the aberration phase shift –χ(Kp),
where Kp is the transverse component of the wavevector of the partial plane wave included by
the aperture (Nellist, 2011). In the weak phase object description of phase contrast it is
assumed that the transmission function of the object results in only a small phase shift with no
change in amplitude and thus the intensity of coherent scattering is dependent on the factor
sin(χ(Kp)). This factor is known as the phase contrast transfer function and determines the
intensity of contributions from features of certain spatial frequencies. The aberration phase shift
is dependent on all aberrations, including defocus, and therefore small changes in focus can
result in changes in the intensity of periodic features, including reversals in contrast. The
complex relationship between the properties of the objective lens and the contrast in a phase
contrast image mean that interpretation of these images relies on comparison with accurate
Bright field imaging in the STEM or CTEM can be used to distinguish between different
elements through mass thickness contrast, particularly when the mass thickness difference is
25
large. For instance, BF imaging has been used to verify the encapsulation of La in carbon
nanoparticles (Ruoff, et al., 1993) and Pt in carbon nanotubes (Kyotani, et al., 1997). With
The high-angle annular dark field (HAADF) imaging mode is that in which an annular detector is
used to collect electrons scattered above a threshold scattering angle (see Fig 2.1). HAADF is
Incoherence of HAADF imaging is assumed either due to exclusion of Bragg scattered beams
and domination of thermal diffuse scattering (Howie, 1979; Jesson & Pennycook, 1995) or
through averaging over a large number of Bragg scattered discs (Nellist, 2011). Thus, the
intensity of the HAADF signal is often described by Rutherford scattering and is largely
dependent on the Z of the material and its thickness. The exact exponent of the Z dependence
depends upon the element in question and the angles that the detector subtends (Treacy, 2011).
Through the principle of reciprocity the HAADF mode is related to the use of a large annular or
tilted electron source in the CTEM, which would provide a largely incoherent source of electrons
The strong dependence of HAADF imaging on the atomic number of the sample has led to its
wide-spread use in assessing the elemental distribution within nanoparticles. While a slowly
a core-shell nature via HAADF imaging. A wide range of bimetallic nanoparticles have been
studied in this way, including AgAu (Belic, et al., 2011), FeAu (Cho, et al., 2005), AgSi (Cassidy,
et al., 2013) and AuRh (Chantry, et al., 2012). The interface between regions of different
elements may even be probed down to atomic resolution, as has demonstrated in AuPd (Ferrer,
et al., 2008; Mayoral, et al., 2012) and GaPd (Leary, et al., 2013). Although HAADF imaging
may distinguish between different regions of known composition, it cannot determine the
elements present within a nanoparticle, for that a truly spectroscopic technique, such as EELS
or EDXS, is needed.
26
2.2.3. Electron Energy Loss Spectroscopy
Electron energy loss spectroscopy involves the collection and analysis of electrons that have
lost energy due to inelastic scattering processes within the sample (Egerton, 2009). The EEL
spectrometer works in much the same way as a prism of light, separating electrons of different
energy spatially on to a CCD camera by a magnetic field normal to the electron beam. The
electron energy loss spectrum is generally split in to two regions, the low-loss region that
contains energy losses associated with phonons and plasmons and the core-loss region that
contains energy losses associated with ionisation events (Fig 2.3). At a core-loss edge, after
subtracting a background at each pixel, integrating over the intensity of a number of channels
will give an intensity related to the elemental composition (Egerton, 2011b). The elemental
composition can be quantified in a similar manner to the EDXS quantification that is described
later. One of the great benefits of EELS is the high-energy resolution of the technique (down to
10s of meV in the most recent cases (Krivanek, et al., 2013)) in comparison to EDX
spectroscopy (≈ 130 eV for silicon drift detectors (Eggert, et al., 2004)). The ability to resolve
differences in energy loss on the order of nanometres allow visualisation of the fine structure of
core-loss edges that reveals information on the chemical bonding of elements (Batson, 1993;
Silcox, 1998). Similarly, the energy resolution allows measurement of subtle shifts in edge
energy that can be associated to bonding environment (Potapov & Schryvers, 2004).
Figure 2.3. Example EEL spectrum showing the characteristic features of the zero-loss peak,
the plasmon-loss peaks and core-loss edges, the latter have had their intensity multiplied by 50
times to ensure they are visible at this scale. Elemental identification is possible through the
27
A similar imaging method to EELS is energy filtered TEM (EFTEM), which is the broad-beam
CTEM equivalent of EELS (Reimer, 1995). In EFTEM a slit is placed in the dispersion plane of
the electron spectrometer so that only electrons of a certain energy are allowed to form an
Elemental mapping via EELS can provide direct confirmation of the core-shell structure of
nanoparticles, for example in CoMn (Lee, et al., 2011). In terms of elemental segregation, EELS
has directly shown Pt surface segregation in electrochemically dealloyed CuPt (Wang, et al.,
2012) and acid leached PtNi nanoparticles (Gan, et al., 2013), as well as Pt segregation to
certain edge sites in PtNi nanoparticles (Cui, et al., 2013), helping to explain each system’s
nanoparticles can also be observed, with an ordered structure seen in PtCo nanoparticles
(Wang, et al., 2013). Through observation of shifts in the edge energy, it is also possible to
monitor phase changes within a single nanoparticle for a single element, such as that in Pd
Atomic resolution EELS mapping, routinely performed in bulk oxides (D'Alfonso, et al., 2010;
Kimoto, et al., 2007), has also been demonstrated in an oxide nanoparticle (Rossell, et al.,
2012). In this study atomic resolution EELS mapping of Ba-doped SrTiO3 nanoparticles was
used to reveal the extent of dopant clustering within these nanoparticles. Dopants down to the
single atom level can be detected via EELS in bulk materials (Varela, et al., 2004), which should
As discussed previously, inelastic scattering of probe electrons through interaction with atomic
electrons may result in the ionisation of atoms within the sample, in which a core-level electron
is ejected. The decay of atomic electrons as the electronic configuration of the atom proceeds
back to the ground state may result in the production of X-rays of a specific energy (Fig 2.4).
This energy is dependent on the energy difference between electronic states and is therefore
28
elements within a sample, which is the underlying principle of EDX spectroscopy. Characteristic
X-rays are labelled based on the final and initial energy levels of the transitioning electron, these
Figure 2.4. a). Diagram of the process in which an incident electron causes a characteristic X-
ray to be emitted from an atom. b) Diagram of the possible electron transitions in the atom with
electron scattering. It is important to note that X-ray generation is always in competition with
another secondary process, namely the emission of Auger electrons. The fraction of ionised
atoms which decay through X-ray emission is known as the fluorescence yield, ω, and can be
-3
as low as 10 for low energy X-rays (Krause, 1979). Characteristic X-ray detection is therefore
known to be a highly inefficient detection method for light elements, particularly when compared
to electron energy loss spectroscopy (EELS), in which the primary process is detected.
However, there are advantages to characteristic X-ray generation over the collection of
inelastically scattered electrons. X-rays are known to be emitted isotropically over all angles,
allowing a somewhat arbitrary position of X-ray detectors and no need for consideration of
differential cross sections (Bethe, 1930). Characteristic X-ray generation is also an inherently
One of the drawbacks of X-ray detection is that the X-rays that reach the detector are not just
produced through the radiative decay of ions. X-rays are also produced through the
29
deceleration of the electron beam as it undergoes inelastic scattering, emitting X-rays that are
continuously over an energy spectrum, as electrons in the beam may undergo any amount of
where E0 is the energy of the incident electrons and K is Kramers’ constant. The continuous
nature of Bremsstrahlung X-rays mean that characteristic X-rays in an EDX spectrum are
distribution of Bremsstrahlung radiation is strongly forward peaked and X-ray detectors are
generally placed at large scattering angles, therefore the Bremsstrahlung background does not
In the detection of electrons in the CTEM/STEM, the projection system, including the pre- and
post-specimen apertures and the position of the detectors, ensures that only electrons
associated with small angle scattering from the electron beam are detected. For EDX detectors,
a collimator is used to try to ensure the detection of X-rays only from the illuminated specimen
area, but invariably X-rays generated from outside of this area are also detected. X-rays
generated from outside of the illuminated area are known as spurious or system X-rays, these
terms have slightly differing definitions, and can compromise accurate spectroscopic analysis.
Spurious X-rays may be generated either by stray electrons exciting characteristic X-rays away
from the intentionally illuminated area, or through emitted X-rays (either characteristic or
Bremsstrahlung) fluorescing X-rays away from the illuminated area. X-rays generated by stray
electrons can often be determined through collection of a ‘hole’ spectrum from a region in which
Typically, system X-rays may occur from the support grid and holder (typically Cu), the pole-
piece (Fe and Co), the detector (Si and Pb) and, counterproductively, the X-ray collimator (Zr in
the case of the Super-X detector system presented here). It is possible to avoid, or at least to
limit, the production of spurious X-rays through the use of holders and grids manufactured from
30
Be, which causes production of low energy X-rays which are not resolved by the EDX detector
and cannot fluoresce higher energy X-rays elsewhere. Limiting the tilt of the sample if mounted
on to a grid also helps to limit spurious X-ray production from grid bars, as large sample tilts
lead to grid bars interacting with a larger number of stray electrons and Bremsstrahlung X-rays,
although tilting the sample is a necessity for some detector-holder setups in order to avoid
Figure 2.5. Spectra from a sample containing elements Al, O and Co from vacuum (blue),
carbon film (green) and the sample (red). There are few counts from vacuum, demonstrating
that stray electrons are not a major contributor to spurious X-rays in this particular case. Sample
The proportion of generated X-rays that are ultimately collected is dependent on a number of
factors, including the solid angle that the detector subtends, the absorption of X-rays in the
sample, and at the detector, and the proportion of X-rays that pass all the way through the
detector. A large amount of detector development has aimed to increase this value, often known
as the detection efficiency, as EDX spectroscopy in the STEM has typically been limited by low
All commercially available X-ray detectors for the CTEM/STEM work on the principle of electron-
probability of exciting a number of electron-hole pairs with energy equal to the bandgap of the
detection material (1.1 eV in the most commonly used case of Si). The generated electrons are
collected at an anode where they are then detected as a sharp step in voltage from a field effect
31
transistor. Noise, whether electronic or thermal noise, is detected as small steps in voltage in
much the same way as characteristic X-rays. This noise may be detected in isolation, where it is
assigned to channels around 0 eV and leads to a large peak at this value, or it may be detected
on top of a step due to a characteristic X-ray, which leads to broadening of the detected
energies of the characteristic X-rays, thus limiting the energy resolution of the technique. The
detection mechanism of the detector may also result in artefacts in the spectra generated. For
example, if two X-rays are incident on the detector within the read time of the detector a sum
peak is generated at an energy that is the sum of the two X-rays. Detectors may also suffer
from incomplete charge collection which results in a shoulder on the characteristic peak at the
low energy side of the peak. For most commercial detectors the acquisition software
Most modern EDX detectors in the TEM are now silicon drift detectors (SDDs), as opposed to
the Li drifted silicon detectors (Si(Li)), which were standard until recently. The SDD features a
large n-doped silicon crystal with concentric p-doped rings embedded on the back side of the
detector (Fig 2.6b). The design of the SDD, with a small area anode (typically less than 100 μm
in diameter) to collect electrons, enables shorter discharge times than the Si(Li) detector, in
which the anode covers the whole back face (Fig 2.6a). This leads to much lower processing
times in the SDD which has allowed much higher X-ray count rates. It also means that the
larger and larger detectors (Newbury, 2006). The small anode also removes an important
source of noise in the detector, which allows good energy resolution at relatively modest
temperatures (above -100°C). This removes the necessity to cool with liquid nitrogen and
32
Figure 2.6. Schematics of cross-sections of EDX detectors. a) Si(Li) detector, in which the
anode covers the whole back face. b) SDD detector, in which concentric p-type rings promote
When high count rates are present (> 100,000 counts), it is noted that increasing the detector
size has a large negative impact on the resolution of the detector (Eggert, et al., 2004). For the
low count rates that are characteristic of the STEM however, large area SDDs can play an
important role in increasing the solid angle of the detector system. The SDD also lends itself to
the fabrication of arrays of single detectors (Fiorini, et al., 2002; Zaluzec, 2004). It seems
entirely possible that arrays of SDDs that are shaped to fit specifically within the STEM stage
One drawback of SDDs is that they are typically much thinner than the Si(Li) detector, which
means that they do not detect all incident X-rays above a certain X-ray energy. Thus, the
detection efficiency of an SDD decreases above a certain threshold energy. The detection
efficiency of detectors is also affected by the positioning of a window in front of the EDX
detector. Traditionally, a thin window (10 – 25 μm) of Be was placed in front of the EDX detector
in order to prevent ice or contamination build up on the cold detector. Over the years these
windows have had their thickness gradually reduced to less than 100 nm through the
incorporation of other materials such as polymer films (Williams, et al., 1995). It is now common
for SDDs to be “windowless” in that they do not have a traditional window but are shuttered
when not in use (Anderhalt & Sandborg, 2011). Even when detectors are windowless they will
typically have a very thin (on the order of 100 nm) light shield to reflect/absorb incident light and
thus have a degraded detection efficiency for low energy X-rays. The detection efficiency of a
single detector is therefore largely dependent on the X-ray energy, with detection efficiency
33
degraded at both low and high X-ray energies. The precise detection efficiency can be vastly
different for each type of detector and highly dependent on X-ray energy (Fladischer, 2013).
The traditional geometry of an EDX detector in the STEM is a single detector, inserted from
outside the column to as close a distance as possible to the sample, typically subtending a solid
angle of approximately 0.1 - 0.3 sr. A solid angle of between 0.1 – 0.3 sr results in detection of
only 0.8 – 2.4 % of characteristic X-rays. When we consider this poor collection efficiency in
combination with potentially low fluorescence yields described in section 2.3.1, it is obvious that
EDX spectroscopy is a very inefficient method to detect inelastic electron scattering. For
example, a typical EDX detector of 0.2 sr is able to detect a maximum of approximately 0.005 %
of inelastic electron scattering events from carbon atoms, without consideration of X-ray
absorption, compared to a value often larger than 50 % for EELS. It should be noted, as the
positive case for EDX, that the nature of EDX spectroscopy means that peak to background
ratios are often far greater than those of EELS and heavy element detection is often much more
favourable in EDXS than EELS (von Harrach, et al., 2010). A large part of EDX spectroscopy
development in the STEM in recent years has therefore focussed on increasing the solid angle
There have been many suggestions of detector geometries that increase the solid angle of the
EDX detector. In most cases, the introduction of the SDD has allowed increases in detector size
or efforts have been made to position the detector closer to the sample (Williams & Carter,
2009c). However, novel detector geometries have also been suggested to increase detector
solid angle. These include a dual detector setup (Myers, et al., 2010), an annular detector
positioned above the sample (Kotula, et al., 2008) and a π-sr detector in a post-specimen
position (Zaluzec, 2009). The design that has found arguably the greatest commercial success,
and the one used almost exclusively here, is the Super-X detector, as included in FEI’s
ChemiSTEM technology (von Harrach, et al., 2009). This detector is in fact four separate SDDs
that are positioned symmetrically within the polepiece of the STEM. Each detector is similar to a
standard SDD, although positioned very close to the sample, but the fact that there are four
detectors results in an increased solid angle of up to 0.9 sr (von Harrach, et al., 2009). The
precise solid angle achievable depends on the gap within the objective lens, as this determines
34
how close to the sample the detectors can be positioned. The positioning of the detectors at
regular 90° azimuthal intervals around the optic axis also allow for X-ray detection when the
2.3.3. Quantification
straightforward process that can yield important quantitative information at near atomic
resolution (Kothleitner, et al., 2014). Quantification in CTEM/STEM has principally followed the
method set out by Cliff and Lorimer, using their Cliff-Lorimer k-factors that relate the intensity of
a pair of X-ray peaks (IA and IB) to their composition, in this case the ratio in weight fraction (CA
and CB) (equation 2.6) (Cliff & Lorimer, 1975), where A and B are two elements of interest:
𝐶𝐴 𝐼
= 𝑘𝐴𝐵 𝐴 . (2.7)
𝐶𝐵 𝐼𝐵
Historically, this type of intensity ratio was preferred to absolute intensities due to instabilities in
the CTEM/STEM, primarily instability in the beam current. Cliff-Lorimer factors are dependent
the accelerating voltage and the collection/detection efficiency of the detector. Cliff-Lorimer
factors can be determined in a number of ways. The most robust method, in terms of accurate
composition (Sheridan, 1989). Standard samples are available that contain a large number of
elements but invariably it can be difficult to obtain accurate standards for all elements. A less
accurate but widely used method is to calculate approximate k-factors from microscope
parameters. This calculation is often performed via equation 2.7, which is obtained through a
(σ ωaε)B AA (2.8)
k AB = (σI ,
I ωaε)A AB
where ω is the fluorescence yield, aX is the relative transition probability of the peak, εX is the
detection efficiency of the detector and AX is the atomic weight of the element X.
Another quantification method gaining increasing popularity is the zeta-factor method proposed
by Watanabe (Watanabe & Williams, 2006). This method uses single element factors to
determine composition and has become possible due to the stable beam currents achieved in
35
the modern day STEM. The intensity of X-rays of an element A are related to the composition of
𝜁𝐴 𝐼𝐴 (2.9)
𝐶𝐴 = ,
𝜌𝑡
where ρ is the density of the sample and t is the sample thickness. There are many practical
benefits to the zeta-factor method that derive from its use of only one X-ray peak. Firstly,
experimental determination of zeta-factors can be done from single element standards of known
thickness or volume, which are easy to obtain for a large number of elements. Secondly, the
zeta-factor offers a method to determine sample thickness or inelastic cross-sections when the
In both the Cliff-Lorimer and zeta-factor methods an important initial step is to separate
background subtraction and there are a number of well-developed methods to do this (Williams
& Carter, 2009c). The most basic of background subtraction methods is the simple window
method. Here, the intensity of the background in two windows, one above and one below the
characteristic peak, is integrated and subtracted from the intensity integrated over a window of
the same width on the characteristic peak. This very simple technique is very effective for well
separated peaks on a slowly varying background, typical of the high-energy regime, but it is
often not possible to perform this type of background subtraction in the low-energy regime.
A slightly more advanced method is to model the Bremsstrahlung background using Kramers’
law (equation 2.5), so that it is possible to subtract the modelled counts in the peak regions.
Kramers’ law can be fit to background regions of spectra by varying the constant K to match the
intensity and the modelled intensity can simply be subtracted from the regions underneath the
peaks.
An entirely different approach to background subtraction comes in the form of filtering out the
smooth background. A top-hat filter can be convoluted with the EDX spectrum and produces a
function that is similar to a second differential of the spectrum. Thus, the slowly varying portions
of the spectrum are close to zero and the peaks remain. However, this form of filtering does not
36
intensity present at the base of peaks, and careful comparison with element standards is
needed to extract peak intensities. Nevertheless, this method is still used in some commercial
In the quantification of EDX spectra from nanoparticles the absorption of X-ray intensity through
the sample is often neglected. This is often a reliable assumption, although not always,
particularly when considering low energy X-rays. The absorption of X-rays in matter follows the
exponential attenuation law put forward by Beer and Lambert for light, where X-ray intensity is
given by
𝜇
−𝜌 ]𝑙 (2.10)
𝐼 = 𝐼0 𝑒 𝜌 ,
𝜇
where I0 is the intensity of X-rays before absorption, ρ is the density of the sample, ] is the
𝜌
mass attenuation coefficient of the sample for X-rays of the energy in question and l is the path
length through the material. The values of both density and mass attenuation coefficient are
strongly dependent on the atomic number of the material so that higher atomic number
materials typically absorb X-rays more strongly, although the exact dependence of the mass
attenuation coefficient on Z is quite complicated due to absorption edges. The mass attenuation
coefficient is also highly dependent on X-ray energy; lower energy X-rays are absorbed more
For an analysed point the absorption of X-rays must be considered for all X-ray paths passing
from all points on the electron beam to the detector. Assuming the detector is a point detector,
consideration of all X-ray paths is performed by integrating all X-ray paths along the electron
beam. This is straightforward in thin film samples where the X-ray path length is given by
𝑙 = 𝑧 csc(𝛼) (Philibert, 1963), where z is the depth of X-ray production in the sample and α is the
take-off angle. This leads to an expression for X-ray absorption of a single energy of X-ray from
a thin film of
𝜇
𝜇 −( ]𝜌𝑡 csc(𝛼)) (2.11)
𝐴𝐶𝐹 = ] (1 − 𝑒 𝜌 ).
𝜌
The absorption correction factor (ACF) can be combined with the zeta-factor in the form above
(equation 2.11) or two factors can be combined in ratio with k-factors. Calculating the X-ray
37
absorption from free standing nanoparticles is certainly not so straightforward and so far has not
been examined in the STEM. However, X-ray beam paths have been determined for a variety of
particle morphologies for electron probe microanalysis (EPMA) (Armstrong & Buseck, 1975;
Zreiba & Kelly, 1988) and it seems reasonable that these may be applied in a similar manner to
STEM studies. The application of absorption correction to spherical particles is the subject of
chapter 3.
The improved collection efficiency of the latest EDX detector systems has generated an
single nanoparticles. For example, EDX spectrum imaging has revealed the structure of AuPd
nanoparticles synthesized by microbes (Tran, et al., 2012) and after calcination (Wilson, et al.,
2013). EDXS elemental mapping of catalysts after electrochemical aging has also been applied
to reveal Pt surface segregation in Pt/PdAu nanoparticles (Sasaki, et al., 2012) and the
Elemental mapping can also provide key information to aid understanding of the long term
nanoparticles have been observed by EDX spectroscopy to form a Cr 2O3 layer, a result which is
driven by the greater ability of Cr to oxidise compared to Ni (Wang, et al., 2014). Similarly, in
core-shell CuSe-CuS, the CuSe component has been found to preferentially oxidise, with a
similar result found in the inverse core-shell CuS-CuSe structure (Miszta, et al., 2014).
Elemental redistributions during in situ heating and beam induced oxidation of AgAu
nanocrystals have been mapped with in situ EDX spectrum imaging, demonstrating alloying
during heating and the formation of hollow Au-core Ag2O-shell nanostructures under electron
The spatial resolution of techniques in the STEM is not an entirely straightforward topic. The
resolution of coherent imaging modes is particularly complex, with the angles of both the probe
forming aperture and the detector playing an important role in defining the resolution limits
(Nellist, 2011). The situation is somewhat less complex for incoherent modes, and in particular
38
inelastic scattering, where the resolution of the technique is simply related to the size of the
electron beam within the sample. For this type of inelastic scattering it is pertinent to split the
topic into two parts, factors influencing the size of the electron beam before the sample and
The size of the electron probe is generally known to be influenced by three factors (Craven,
2 𝐼
The diameter of the probe due to the finite source size, 𝑑𝑠 = √ , where α is the
𝜋𝛼 𝐵
The diameter of the first Airy disk due to diffraction from the probe-forming aperture,
1.22𝜆
𝑑𝑑 = , where λ is the wavelength of the probe electrons.
𝛼
The diameter of the disk of least confusion caused by lens aberrations, where each
1
The first two terms are dependent on , with the prefactor determining which term will be the
𝛼
more significant. For thermionic sources, the low brightness of the source leads to a large d s
term which dominates all others and has meant that recent advances in the STEM have been
coupled to the availability of the field emission gun (FEG) source. For cold FEGs in particular,
this term tends to be outweighed by the diffraction term, dd, although for a Schottky FEG when
using beam currents associated with analytical work (100s of pA) it is often the ds term that
dominates at low values of α. The finite source size and diffraction terms diminish with
increasing α, so that for standard imaging conditions spherical aberrations of the third order
angle will result in poorer resolution. This description is largely simplified from discussions of the
electron probe diameter, which require a full wave optical treatment for accurate probe size
Aberration correction of electron optical systems has produced rapid improvements in STEM
imaging by reducing the size of the electron probe via the addition of non-spherical electron
39
lenses (Bleloch & Ramasse, 2011; Haider, et al., 1998; Krivanek, et al., 1999). With the advent
of electron optical aberration correction in the STEM the third order spherical aberration is
essentially reduced to zero. Thus, the probe forming angle can be increased until the limits of
th
aberrations of higher order, which are typically 5 order spherical aberrations or chromatic
th
aberrations, depending on the specifics of the microscope. At this point, either 5 order
the probe size further (Krivanek, et al., 2013). Aberration correction is particularly useful for
analytical electron microscopy studies, as it allows an increased probe current inside a smaller
probe size.
Through aberration correction the electron probe size can be reduced to sub-Ångstrom size at
the surface of the sample. However, as the electron beam traverses the specimen it does not
stay at the same size. Firstly, the convergent nature of the beam ensures that the beam
intensity spreads after the focal point even if there is no sample present, although the
convergence angles associated with the STEM tend to be so small (10s of mrads) that this
effect is relatively small. Assuming a 100 nm sample with the focal point at the midpoint of the
In a sample at most orientations, interactions of the electron beam with the sample atoms will
result in the scattering of the beam that leads to beam broadening as the beam traverses the
sample. This beam broadening is dependent on the density and thickness of the sample and
can be estimated using the single-scattering equation of Goldstein (Goldstein, et al., 1977)
1
𝑍̅ 𝜌 2 3 (2.12)
𝑏 = 7.21 × 105 ( ̅) 𝑡 2 ,
𝐸0 𝐴
where 𝑍̅ is the average atomic number of the sample, E0 is the energy of the probe electrons, ρ
is the sample density, 𝐴̅ is the average atomic weight of the sample and t is the sample
thickness. This beam broadening term takes in to account 90% of the beam intensity and for
100 nm thick Ag and Au samples, and a beam energy of 200 kV, would result in broadening to
diameter, such that in thick samples the initial beam diameter has little impact on the overall
40
EDX spectroscopy, electron scattering is a largely stochastic process and therefore cannot be
accurately described using fixed terms. Many treatments of beam broadening involve use of
Monte Carlo codes that generate random scattering processes to estimate beam broadening
However, when the electron beam direction is oriented parallel to a major crystallographic zone
axis of a material the major consideration is no longer beam broadening but is instead a
phenomenon known as electron channelling (Fertig & Rose, 1981; Pennycook & Jesson, 1990).
Here, the line of positive potentials along the beam direction confines the electron beam to the
atomic columns, with the intensity distribution “beating” down the column in the “Pendelossung”
manner. Electron channelling is most often associated with atomic resolution imaging in the
STEM, as it a key component to inhibiting electron beam broadening that would remove any
atomic resolution features. However, electron channelling is also an important consideration for
STEM imaging in which pixel sizes are above that at which atomic columns could be resolved,
as it alters the effective interaction parameter of electron scattering processes. That is, electron
channelling will act to increase the number of electron interactions within the sample over a
given distance. This can be seen as an increase in the intensity of electron scattering (Tafto &
Lehmpfuhl, 1982) and characteristic X-ray production (Duncumb, 1962; Liao & Marks, 2013)
along certain atomic columns in nanoparticles oriented along a major zone axis. This can be
seen from EDX spectra collected from an AgAu nanocube tilted about its <100> zone axis,
where the intensity of the Au Mα peak increases by 50% on the zone axis compared to a 5°
41
Figure 2.7. Plot of normalised Au Mα peak intensity from a 50 nm size AgAu nanocube as a
function of tilt angle away from the <100> zone-axis. Sample courtesy of Pedro Camargo
Whilst the strong interaction of the electron beam with the sample is of great use for detecting
signals from very small sample volumes, it can also lead to relatively high levels of damage
(Egerton, et al., 2004). If sufficient damage occurs to provide structural changes to the
nanoparticle, then the nanoparticles may no longer be representative of the initial state. It is
therefore important to find methods to mitigate against beam damage, firstly by considering the
processes under which beam damage can occur. Electron beam damage in inorganic
nanoparticle samples becomes a particular problem when using techniques that require a high
Beam damage in inorganic samples can occur via both the elastic and inelastic scattering
processes outlined in section 2.1. Rutherford scattering has up until this point been considered
as elastic scattering, as at the small scattering angles considered in electron imaging there is
immeasurable energy transfer. However, this is not the case for large angle scattering which, for
a 200 keV electron beam scattering from an Al atom, can transfer energy of up to 20 eV
(Egerton, et al., 2010). Rutherford scattering gives rise to “knock-on” damage, where
backscattering of an electron can result in a large enough transfer of energy to remove an atom
42
from its position in the atomic lattice. This occurs for incident electron energies up to a threshold
energy, which is the energy required to break atomic bonds at specific lattice sites.
Knock-on damage is element dependent, in that a greater amount of energy can be transferred
to lighter atoms. For the vast majority of elements the energy transferred through high-angle
elastic scattering at CTEM/STEM energies is not sufficient to displace samples in the bulk
material. Knock-on damage in inorganic thin films is known to be primarily through sputtering
from the exit surface of the electrons, where the energy required to break surface bonds is
lower than that in the bulk (Egerton, et al., 2010). The lowest knock-on energies are found at
defect or kink sites, where it is expected that sputtering rates are greatest (Egerton, et al., 2010).
Considering that nanoparticles possess a large number of these sites, it is anticipated that
sputtering from nanoparticles occur at lower energies than that of defect free surfaces, as has
been seen for the sputtering of Ag from AgAu nanoparticles (Braidy, et al., 2008). The addition
(Krakow, et al., 1994; Marks, 1994), in which the crystallinity of nanoparticles can be temporarily
lost. In order to reduce the effect of knock-on damage it is possible to reduce the energy of the
incident electrons below the threshold energy, potentially eliminating damage altogether, as is
seen in the study of carbon nanomaterials at low electron energies (Bell & Erdman, 2013;
Suenaga & Koshino, 2010). It is also possible to inhibit knock-on damage via sputtering by
encapsulating the sample in a thin layer of a conducting material such as carbon, so that it is
the carbon that is preferentially sputtered (Egerton, 2013; Muller & Silcox, 1995).
Decreasing the incident energy of the electron beam to a few 10s of keV may limit damage from
Rutherford scattering, however, for typical S/TEM beam energies this increases the cross
section of inelastic scattering, which peaks at approximately five times the critical ionisation
energy (Williams & Carter, 2009b). Damage due to the ionisation of atoms through inelastic
scattering is often termed radiolysis, and the precise damage mechanism after initial bond
breaking depends on the material in question (Egerton, et al., 2004). It has been seen, however,
that radiolysis does not proceed in conducting samples (Egerton, et al., 2006), which it is
suggested is due to the presence of conduction band electrons that are able to rapidly fill any
empty core-levels. In insulating materials where radiolysis is often the primary damage
mechanism, coating with a conductive layer has been shown to provide some protection to
43
damage (Egerton, 2013), although it is not known whether this is due to the supply of free
beam heating or charging. Charging due to the electron interactions, primarily inelastic
scattering, has been shown to cause significant damage in inorganic samples at high beam
currents (Cazaux, 1995). Similarly, inelastic collisions may cause deposition of thermal energy
in a sample. However, it is known that for thin samples at typical STEM accelerating voltages
that the small fraction of the energy deposited results in only minor temperature rises (Egerton,
et al., 2004). Reducing the dose rate, i.e. the beam current, has been shown to be an effective
method to limit electron beam induced charging (Cazaux, 1995) and should also reduce
For all damage mechanisms a reliable method of limiting degradation is to lower the total dose,
although this may preclude a sufficient signal to noise ratio at a low dose. Particularly in the
techniques used here, EDXS and its extension to electron tomography (reviewed in Chapter 4),
a high dose is currently a necessity, although every effort should be made to reduce the dose
as far as possible.
2.6 Summary
The scanning transmission electron microscope is a key tool to probing the elemental
majority of imaging modes are able to distinguish regions of different elements, although BF and
HAADF modes do not provide the truly spectroscopic signal of EELS and EDXS. Advances in
EDX detector systems over the past decade have led to a vastly improved capability for
challenges to performing EDX spectrum imaging, particularly the low count rates typically
associated with the technique. To improve count rates, thicker samples or higher electron doses
are favourable, albeit with inherent associated drawbacks such as decreased resolution or
beam damage.
44
3. X-ray Absorption Correction in Spherical Nanoparticles
3.1. Introduction
This chapter details a method for the determination and application of X-ray absorption
correction to spherical nanoparticles. There are many studies which describe quantitative EDX
analysis of nanoparticles in the STEM, all of which assume that the particles are small enough
that X-ray absorption may be ignored. This is a good assumption for many nanoparticles but
should be carefully considered when analysing larger nanoparticles, particularly when using low
The presented manuscript deals specifically with X-ray absorption in spherical nanoparticles.
There are many examples of spherical or near-spherical nanoparticles in the literature and the
specific motivation for this work came from the analysis of spherical nanoprecipitates extracted
from Ni-based superalloys (Chen, et al., 2014). The correction factors here for spherical
nanoparticle morphologies, although care should be taken in the quoted precision of the
Derivation of X-ray path lengths could be extended to a number of nanoparticle shapes but X-
ray absorption correction in the traditional sense is only possible for nanoparticles with well-
the nanoparticle is not discernible from a two dimensional STEM image, then a full
determination of X-ray path length could be obtained from HAADF electron tomography.
fashion to determine the composition at each voxel. Chapter 5 attempts to provide a platform for
this type of quantitative work through thorough investigation of the methodology of STEM-EDX
tomography, although full quantification on a voxel by voxel basis is beyond the scope of this
thesis.
45
A portion of the experimental data was collected by Yiqiang Chen (approximately half of that
collected on the Tecnai), who also assisted in the derivation of the X-ray path length. The Au
The results of this project have been submitted to Microscopy and Microanalysis. The original
46
Paper: X-Ray Absorption Correction for Quantitative STEM-EDX of
Spherical Nanoparticles
1 1 2 1,3 1
Thomas Slater , Yiqiang Chen , Gregory Auton , Nestor Zaluzec , Sarah Haigh
60439, US
Abstract
A new method to perform X-ray absorption correction for spherical particles in quantitative
experimental data of X-ray counts from Au nanoparticles to verify the derived methodology. The
effect of detector elevation angle is considered and a comparison with thin film absorption
correction is included.
Introduction
The physical and chemical properties of nanoparticles are known to be sensitive to their precise
composition has a large effect on their localised surface plasmon resonance wavelength
(Hostetler, et al., 1998; Kariuki, et al., 2004) and the stoichiometry of iron oxide nanoparticles
plays a vital role in determining their magnetic properties (Carvalho, et al., 2013; Salazar, et al.,
2011). Properties can also depend on the elemental segregation within a particular nanoparticle.
For example, the segregation of certain elements to the surface or to particular facets of
nanoparticles has been shown to greatly affect their catalytic performance (Cui, et al., 2012;
Slater, et al., 2014b). The average composition of large ensembles of nanoparticles can be
accurately obtained through time of flight mass spectrometry or flame atomic absorbance
spectroscopy and average surface composition through X-ray photoelectron spectroscopy (Baer
47
& Engelhard, 2010; Baer, et al., 2010). However, nanoparticle samples are often polydisperse in
Energy dispersive X-ray (EDX) spectroscopy within the scanning transmission electron
spectroscopy in the STEM has traditionally been based on the use of Cliff-Lorimer factors (k-
factors) (Cliff & Lorimer, 1975) that relate the relative intensity of X-ray peaks for two elements
cA I (1)
=kAB A ,
cB IB
standard samples (Sheridan, 1989; Wood, et al., 1984), or can be theoretically calculated
(Goldstein, et al., 1977; Zaluzec, 1979), and can provide accurate quantitative information of
thin samples. The relationship in equation 1 was derived assuming that the specimen region of
interest is sufficiently thin so that the combined effects of electron energy loss, x-ray absorption
and x-ray fluoresence can be ignored. While the effects of energy loss on the ionization cross-
section are negligible for typical TEM specimens, the same is not generally true for x-ray
absorption effects, particularly for low energy x-rays in high atomic number materials. In order
to correct for the effect of X-ray absorption on relative peak intensities, an absorption correction
factor (ACF) was introduced which attempts to correct for the differences in the X-ray absorption
of the two X-ray peaks analysed. This absorption correction factor is given by (Philibert, 1963;
μB (2)
- ] ρl(z)
∫e ρ dz
ACF= μA
,
- ] ρl(z)
∫e ρ dz
where l(z) is the X-ray path length through the sample to the detector as a function of the
𝜇
distance, z, along the optic axis, ρ is the density of the sample, is the mass attenuation
𝜌
coefficient of a particular X-ray peak within the sample and the integral is performed over all
possible X-ray paths to the detector. The ACF term can be simply multiplied by the initial Cliff-
48
Lorimer factor to provide a modified Cliff-Lorimer factor, that takes absorption into account (Qiu,
detector through the sample and thus varies with the experimental configuration. Historically,
thin film analysis has dominated EDX spectroscopy due to the prevalence of polished metal and
semiconductor foil samples and the simple geometry of this system. For a thin film, the x-ray
path length to the detector for an x-ray generated at a depth z in the specimen is determined by
sin α
the specimen-detector geometry. This path length (Fig 1a) is given by l(z)=z , where α is
cos(α-θE )
the beam incidence angle, i.e. the angle between the electron beam and the surface of the
sample, in the vertical plane subtending the centre of the detector, θE is the elevation angle of
the detector, defined as the angle between the horizontal plane subtending the specimen and
the point at the centre of the detector face and t is the sample thickness (Zaluzec, 1981).
Inserting this expression into equation 2 and assuming a point detector allows closed form
evaluation of the integral in the ACF (equation 4), giving the result expressed in equation 5,
μB sin α (4)
(-ρ] ρz )
t cos(α-θE ) dz
∫0 e
ACF = μA sin α
,
(-ρ] ρz )
t cos(α-θE ) dz
∫0 e
μB sin α (5)
μ A (-ρ] ρt )
] 1-e cos(α-θE )
ρ
ACF = μ B
( μA sin α
).
] (-ρ] ρt )
ρ 1-e cos(α-θE )
Figure 1. Schematics illustrating the geometry of X-ray detection from a) a thin film sample of
𝑡
thickness t and b) a spherical particle of radius R= .
2
49
This absorption correction equation has been widely used for many years in thin film analysis
Spherical nanoparticles can often have dimensions and compositions such that X-ray
absorption cannot be ignored during quantitative EDX analysis. Whilst the X-ray path length in a
uniform thin film is constant for any position of the beam, for a spherical geometry this is not the
case, as illustrated in Fig 1. For this reason, the path length in a sphere must be defined using a
three-coordinate system. X-ray path lengths for particles of a number of different geometries
have been previously calculated considering detector geometries for both EDX spectroscopy
(EPMA) (Armstrong & Buseck, 1975; Zreiba & Kelly, 1988). Here, we present a derivation of X-
ray path length in a sphere similar to that of Zreiba and Kelly, using a cartesian coordinate
system in which z is the direction of the electron beam, x is the direction towards the X-ray
detector in the plane perpendicular to the path of the electron beam and y is the direction
perpendicular to the plane in which the X-ray path sits (Fig 2). Unlike for the thin film geometry
the specimen thickness t in a spherical particle depends strongly on the position (x,y).
50
Figure 2. Schematics illustrating the geometry of an X-ray path through a spherical particle of
radius R at spatial coordinates x and y, where x is the direction towards the X-ray detector in the
horizontal plane and y is the direction perpendicular to the plane in which the X-ray path sits. a)
schematic subtending the xy plane, i.e. viewed along the optic axis, b) schematic subtending
the xz plane, i.e. within the plane in which the X-ray travels from the sample to the detector.
An X-ray is generated at point P with coordinates (x,y,z). Variations in the value of y, the
distance away from the centre of the particle perpendicular to the plane in which the X-ray path
sits, act only to change the effective radius of a circular slice of the sphere in which the X-ray
travels, so that the modified radius, Rm, is simply calculated as a chord a distance y from the
(6)
Rm =√R2 -y2 .
The X-ray path through the spherical particle, l(z), is then calculated similarly as a chord in a
circle of radius Rm at a distance of |OQ| from the circle centre, minus the distance |PQ|,
(7)
l(z) =√Rm 2 -|OQ|2 − |PQ|.
The distances |OQ| and |PQ| are found through simple trigonometry to be,
t (8)
|OQ| = (z- +x tan θE ) cos θE
2
t (9)
|PQ|= (x-(z- ) tan θE ) cos θE .
2
51
2
(10)
𝑡 t
l(z) =√Rm 2 - (𝑥 sin 𝜃𝐸 − ( − 𝑧) cos θE ) -x cos θE - ( -z) sin θE ,
2 2
t=2√Rm 2 -x2 =2√R2 -y2 -x2 . Substituting our result from equation 10 into equation 2 we arrive at an
2 (11)
μB t t
-ρ] ρ(√Rm 2 -((z- +x tan θE ) cos θE ) -x cos θE -( -z) sin θE )
2 2
t
∫0 e dz
ACF = 2
.
μA √ t t
-ρ] ρ( Rm 2 -((z- +x tan θE ) cos θE ) -x cos θE -( -z) sin θE )
2 2
t
∫0 e dz
Equation 11 is not analytically solvable and therefore must be solved numerically for each
individual case. We note that the final form of the ACF has no dependence on the tilt of the
sample, due to the rotational symmetry of a sphere, although changing the tilt angles may act to
partially shadow the detector and therefore change the effective elevation angle of the detector.
In the special case of x = y = 0, i.e. when the beam is positioned on the centre of the particle,
t
Rm = R = and the absorption correction factor becomes
2
μB 2 (12)
-ρ] ρ(√R2 -((z-R) cos θE ) -(R-z) sin θE )
2R
∫0 e dz
ACF = .
μA 2
-ρ] ρ(√R2 -((z-R) cos θE ) -(R-z) sin θE )
2R
∫0 e dz
For the case in which the detector is positioned at zero elevation angle, 𝜃𝐸 = 0, the ACF at x = y
= 0 is simplified to
μB (13)
2R -ρ] ρ√z(2R-z)
∫0 e dz
ACF= μA
.
2R -ρ] ρ√z(2R-z)
∫0 e dz
If more than one detector is used, then the numerator and denominator in the ACF simply
become a sum of the X-ray absorption terms for each detector using a coordinate system
2 (14
μB t t
-ρ] ρ(√R2 -yi 2 -((z- +xi tan θE ) cos θE ) -xi cos θE -( -z) sin θE )
2 2
t
∑d ∫0 e dz
i
ACF= 2
. )
μA t t
-ρ] ρ(√R2 -yi 2 -((z- +xi tan θE ) cos θE ) -xi cos θE -( -z) sin θE )
2 2
t
∑d ∫0 e dz
i
52
The X-ray absorption due to the derived path lengths is examined for three specific cases here,
to show the effect on nanoparticles with largely different ratios in mass attenuation coefficient.
For the case of the absorption of Lα and Mα X-rays in Au nanoparticles, experimental data is
acquired in order to verify the distribution and extent of X-ray absorption in this particular case.
Au nanoparticles were chosen as they have a known fixed composition, in comparison to multi-
element nanoparticles, and they have X-ray peaks at energies that provide appreciable
absorption correction (2.1 keV for Au Mα and 9.7 keV for Au Lα).
Experimental Methods
Theoretical calculations were performed using Python, with numerical integration performed
using the quad function from the SciPy package (Jones, et al., 2001). Experimental net count
maps were rotated by 45° using the rotate function in the SciPy package. Mass-attenuation
coefficients were taken from (Henke, et al., 1993) where available or through linear interpolation
when the value was not listed and are presented in the results section. Densities for each
system were taken as a weighted average of the densities of constituent elements and the
-3 -3 -3
values used were ρAu = 19.32 gcm , ρAuPt = 20.39 gcm and ρNiO = 6.67 gcm .
(10 nm, 30 nm, 50 nm, 80 nm, 100 nm, 200 nm and 400 nm radius). Nanoparticles were
deposited dropwise on to holey carbon coated copper TEM grids and subsequently washed
dropwise with methanol. Au films were deposited on to 50 nm thick SiN membranes (Agar
S171-2T) using a Moorfield MiniLab evaporator. Au films of thickness 23 nm, 27 nm, 41 nm, 64
EDX spectrum, single spectra, HAADF images and low-loss EELS spectrum images of Au
nanoparticles and Au thin films were acquired on a probe corrected Titan G2 80-200 (S)TEM
microscope operated at 200 kV with a beam current of 600 pA. EDX spectrum images of size
256x256 pixels were acquired for a single 200 nm diameter particle for approximately 30 mins
using multiple passes with a 100μs per pixel dwell time. Single spectra were acquired for 10 -
60 s for 10 nanoparticles of each nanoparticle size, with the electron beam positioned in the
centre of each nanoparticle and drift correction applied. EDX spectra of Au nanoparticles were
also acquired on a Tecnai TF30 microscope operating at 300kV with a beam current of 500pA.
53
HAADF STEM images for each particle were acquired before spectrum acquisition and used to
measure the nanoparticle radius. Au thin film thickness was measured from low-loss EELS
spectrum images using a t/λ method in Digital Micrograph and the measurement was averaged
over a 2 μm x 2μm area for each film. The contribution from the SiN support membrane was
measured independently and its contribution subtracted to obtain the net t/λ value for the Au film.
The t/λ ratio was calibrated using measurements from low-loss spectrum images of Au
nanoparticles where the spherical shape allowed the thickness to be accurately known.
Analysis of EDX spectra and spectrum images was conducted using Bruker Esprit software
(version 1.9.4.3448). Net counts were extracted using the physical TEM model background with
constant fit ranges. Net count maps were acquired using the same settings but with a 2x binning.
All errors in counts and sample sizes are the standard deviation of the measurements taken in
each case.
Detector elevation angles were found by acquiring EDX maps of spherical nanoparticles over a
range of sample tilt values and selecting the angle at which there was a minimum in
The proposed equations for X-ray absorption correction in spherical particles were used to
model the variation in the ratio of counts of different nanoparticle systems. For simulation of the
absorption behaviour, three different test cases are presented: AuPt, pure Au and NiO. The test
cases have been chosen as examples for which the difference in mass attenuation coefficient
𝜇 2 -1 𝜇 2 -1
can be characterised as small (Au Lα ( = 125 cm g ) vs Pt Lα ( = 137 cm g )), medium (Au
𝜌 𝜌
𝜇 2 -1 𝜇 2 -1 𝜇 2 -1 𝜇
Lα ( = 128 cm g ) vs Au Mα ( = 1020 cm g )) and large (Ni Kα ( = 48 cm g ) vs O Kα ( =
𝜌 𝜌 𝜌 𝜌
2 -1
4650 cm g )). A value of normalised count ratio is used to represent the extent of X-ray
absorption throughout, where a value of 1 for the count ratio corresponds to no difference in
Simulations of the ratio in counts across a 200 nm spherical particle for each of the three
systems show the same characteristic distribution of counts but at different intensity scales (Fig
3). The values displayed in Fig 3 are calculated with a detector elevation angle of 18° in all
54
cases. Simulation of the ratio of counts with one detector present reveals the largest variation in
counts across a particle (Fig 3(a-c)i), with the absorption of the lower energy X-ray at its
greatest furthest from the detector. A single detector setup such as this is employed in the
majority of S/TEM systems. While the difference in the ratio of counts between Au Lα and Pt Lα
reaches a low of only 0.995 (Fig 3ai), the difference in the ratio of counts between Ni Kα and O
Kα has a minimum value of 0.543 (Fig 3bi). This ratio equates to the fact that in a 200 nm
diameter particle of NiO, quantification without the use of absorption correction in this case
would lead to identification of the composition of one side of the nanoparticle as close to Ni 2O.
particle, where all particles are 200 nm diameter spheres. Maps are shown for i) a single
detector, ii) two detectors at 180° to each other azimuthally and iii) four detectors at an
azimuthal interval of 90° around the optic axis. The detector configuration is shown
schematically in (d) for each detector setup (i-iii). A detector elevation angle of 18° was used for
When two detectors on opposite sides of the particle are employed, a detector geometry
previously employed by Wu et al (Wu, et al., 2010), the maximum variation is from points at the
surface closest to the detector to points at the surface at 90° to the X-ray path (Fig 3(a-c)ii). The
55
difference in the absorption of Au Lα and Pt Lα is very low (a minimum of the ratio in counts of
0.998) in this case (Fig 3aii). In the case of NiO the ratio in counts reaches a minimum of 0.771
Finally, a simulation was performed for four detectors that are at 90° to one another azimuthally
around the optic axis, as is the case in the Super-X detector (von Harrach, et al., 2009) (Fig 3(a-
c)iii). Here, the difference in the ratio of counts is smaller than that for a one or two detector
setup. In the most severe absorption case considered here, that of NiO, the normalised count
ratio varies from 0.788 in the centre of the particle to 0.869 at the points closest to each detector.
This demonstrates that care should be taken when analysing EDX maps of large nanoparticles
when using a four detector setup that variations in X-ray absorption are not misattributed to a
rise in the percentage composition of the element with the lower energy X-ray at the edge of the
particle. Although, we note that the errors are much smaller, even for the most extreme NiO
case presented here, lack of absorption correction would only lead to a rise in the apparent
atomic percentage of O in NiO from 44 % in the centre of the particle to 46 % at the edge.
These values are still somewhat different from the 50 % true value if no absorption occurred.
In order to experimentally verify the proposed models, the simulated maps of the Au Mα vs Au
Lα count ratios are compared to experimentally acquired equivalent data from a ~ 200 nm
have been normalized by fitting a linear equation to the data points and dividing all of the count
ratios by the value found for x = 0 for each microscope. A Titan microscope with a Super-X EDX
detector was used, in which four detectors are positioned at 90° azimuthal intervals around the
optic axis, rotated 45 degrees with respect to the primary tilt axis of the specimen holder. This
also allows us to investigate the effect of different detector configurations on X-ray absorption
within a nanoparticle. The agreement between experimental and simulated data is excellent (Fig
4b). Each detector configuration shows the same qualitative variation across a particle as the
simulated data (Fig 4a) and, in addition, the values of the normalised count ratio in the
56
Figure 4. Maps of the normalized count ratio between Au Mα and Au Lα X-ray counts acquired
from an approximately 200 nm diameter Au particle showing a) simulated absorption maps and
b) experimental maps using i) one detector, ii) two detectors facing each other and iii) four
detectors at 90° to each other. c) Schematics showing the position of detectors either on (black)
or off (white).
In addition to investigating how the X-ray absorption in spherical particles varies across a single
particle, the dependence on particle size has also been investigated. The theoretical models for
AuPt, Au and NiO, considering the same three pairs of X-ray peaks are plotted in Fig 5 for both
spherical particles and parallel sided thin films. While the difference between the particle and
thin film cases is small for the Pt Lα vs Au Lα case, 1 % difference even at a thickness of 400
nm, the difference between the particle and thin film absorption is large when considering Ni Kα
verification was again provided through the use of pure Au samples. The count ratios obtained
from point spectra at a central beam position on a number of spherical Au nanoparticles were
acquired for a range of sizes of nanoparticle (Fig 5), with all four detectors of the Super-X
detector. The normalized count ratios measured at the centre of the nanoparticles (black points)
57
appear to qualitatively match the theoretical relationship defined in equation 12 (black line). The
error bars represent the standard deviation of values for each point. Point spectra were
acquired at an arbitrary position on untilted thin films of a range of thicknesses and the
normalized ratio between Au Mα and Lα X-ray counts is represented by red points in Fig 5. The
experimental data points appear to qualitatively provide a good fit to the theoretical calculation
(red line) and confirm the predicted difference in X-ray absorption between the particle and thin
Figure 5. Comparison of X-ray absorption at the centre of spherical nanoparticles and that of
thin films. Experimental data points of the ratio between Au Mα and Lα X-ray counts are
The effect of detector elevation angle on the absorption correction behaviour has also been
investigated for spectra acquired at the centre of spherical nanoparticles (Fig 6). The change in
X-ray absorption at the centre of a nanoparticle due to a change in elevation angle is a relatively
small change over the range of detector elevation angles commonly used, represented here by
simulations at elevation angles of 4° and 18° to match experimentally acquired data. For the
extreme case of NiO particles, the difference in count ratio for the two elevation angles reaches
the Au Mα/Lα count ratio obtained from point spectra collected from the centre of Au
nanoparticles, using two different detector geometries (four detectors mounted at approximately
18° elevation angle, Super-X detector on Titan, and one detector mounted at approximately 4°
58
elevation angle, Oxford XMAX 80T on Tecnai F30). The change in elevation angle has little
effect on the absorption of our test case of Au Mα/Lα X-rays, the theoretical absorption
correction in this case only reaching a 1% variation at a radius of 250 nm, which results in no
elevation angles of 4° (blue) and 18° (black). Experimental data points of the ratio between Au
These results, while illustrating the correction factors for a number of test cases, should not be
used to generalize the magnitude of absorption effects in spherical particles, as the degree of a
specific correction is a function of the composition and importantly, the relative mass absorption
coefficients for the x-ray lines analyzed in the particle. The formulae presented in the
individual case.
Summary
A method for performing absorption correction in the EDX analysis of spherical nanoparticles
has been presented and experimentally verified. In many cases nanoparticles are small enough
and X-rays of sufficiently high energy that X-ray absorption can be neglected when trying to
characterising large nanoparticles (the difference in absorption is larger than 10% for radii
59
greater than 80 nm for the Au particles in this study) or when considering low energy X-ray
peaks (for example carbon or oxygen peaks) neglecting X-ray absorption effects has the
potential to introduce large errors. We also note that errors in absorption correction will be
particularly serious when analyzing nanoparticles containing both light and heavy elements
significantly different to that required for thin film samples. The standard thin film analysis is
therefore not appropriate when performing quantitative analysis of nanoparticle composition and
This absorption correction should allow accurate quantification of the elemental composition of
systems this theory could also be applied to cylindrical nanorods with the correct orientation
(long axis perpendicular to the vertical plane subtending the sample and the X-ray detector). For
through the application of electron tomography to provide a full three dimensional shape,
although this process is relatively long and not entirely straightforward. The absorption
correction factor given here can be automatically applied with knowledge of the detector
geometry and the sample density. The method outlined here can also be incorporated in to the
ζ-factor method (Watanabe & Williams 2006), which would also provide a measure of the
sample density.
Acknowledgments
T.J.A. S and S.J.H and gratefully acknowledge support from the North-West Nanoscience
Doctoral Training Centre, EPSRC grant EP/G03737X/1, EPSRC grant EP/M010619 and the
acknowledges support from the Electron Microscopy Center at the Center for Nanoscale
Office of Basic Energy Sciences User Facility under Contract No. DE-AC02-06CH11357, as well
as, a visiting appointment in the School of Materials at the University of Manchester. The
authors wish to acknowledge the support from HM Government (UK) for the provision of the
60
funds for the FEI Titan G2 80-200 S/TEM associated with research capability of the Nuclear
Advanced Manufacturing Research Centre. The data associated with the paper is openly
References
ARMSTRONG, J.T. & BUSECK, P.R. (1975). Quantitative chemical analysis of individual
BAER, D.R. & EENGELHARD, M.H. (2010). XPS analysis of nanostructured materials and
BAER, D.R., GASPAR, D.J., NACHIMUTHU, P., TECHANE, S.D. & CASTNER, D.G. (2010).
Application of surface chemical analysis tools for characterization of nanoparticles. Anal Bioanal
CARVALHO, M.D., HENRIQUES, F., FERREIRA, L.P., GODINHO, M. & CRUZ, M.M. (2013).
Iron oxide nanoparticles: the Influence of synthesis method and size on composition and
CLIFF, G. & LORIMER, G.W. (1975). Quantitative-analysis of thin specimens. J Microsc (Oxford,
CUI, C., GAN, L., LI, H.-H., YU, S.-H., HEGGEN, M. & STRASSER, P. (2012). Octahedral PtNi
nanoparticle catalysts: exceptional oxygen reduction activity by tuning the alloy particle surface
GOLDSTEIN, J.I., COSTLEY, J.L., LORIMER, G.W. & REED, R.J.B. (1977). Quantitative X-ray
analysis in the electron microscope. In Scanning Electron Microscopy, JOHARI, O. (Ed.), pp.
HENKE, B.L., GULLIKSON, E.M. & DAVIS, J.C. (1993). X-ray interactions - photoabsorption,
scattering, transmission and reflection at E=50-30,000 eV, Z=1-92. At Data Nucl Data Tables 54,
181-342.
61
HOSTETLER, M.J., ZHONG, C.J., YEN, B.K.H., ANDEREGG, J., GROSS, S.M., EVANS, N.D.,
PORTER, M. & MURRAY, R.W. (1998). Stable, monolayer-protected metal alloy clusters. J Am
JONES, E., OLIPHANT, T.O. & PETERSON, P.P. (2001). SciPy: Open Source Scientific Tools
for Python.
KARIUKI, N.N., LUO, J., MAYE, M.M., HASSAN, S.A., MENARD, T., NASLUND, H.R., LIN,
Y.H., WANG, C.M., ENGELHARD, M.H. & ZHONG, C.J. (2004). Composition-controlled
microanalysis. In 3rd International Congress on X-ray Optics and Microanalysis, PATTEE, H. H.,
QIU, Y., NGUYEN, V.H., DOBBIE, A., MYRONOV, M. & WALTHER, T. (2013). Calibration of
spectroscopy of SiGe layers in analytical transmission electron microscopy. J Phys Conf Ser
471, 012031.
SALAZAR, J.S., PEREZ, L., DE ABRIL, O., LAI TRUONG, P., IHIAWAKRIM, D., VAZQUEZ, M.,
SHERIDAN, P.J. (1989). Determination of experimental and theoretical KαSi factors for a 200-kV
SLATER, T.J.A., CAMARGO, P.H.C., BURKE, M.G., ZALUZEC, N.J. & HAIGH, S.J. (2014a).
Understanding the limitations of the Super-X energy dispersive X-ray spectrometer as a function
of specimen tilt angle for tomographic data acquisition in the S/TEM. J Phys: Conf Ser 522,
012025.
SLATER, T.J.A., MACEDO, A., SCHROEDER, S.L.M., BURKE, M.G., O’BRIEN, P., CAMARGO,
P.H.C. & HAIGH, S.J. (2014b). Correlating catalytic activity of Ag–Au nanoparticles with 3D
62
VON HARRACH, H.S., DONA, P., FREITAG, B., SOLTAU, H., NICULAE, A. & ROHDE, M.
(2009). An integrated silicon drift detector system for FEI Schottky field emission transmission
WALTER, T. & WANG, X. (2015). Self-consistent method for quantifying indium content from X-
WATANABE, M. & WILLIAMS, D.B. (2006). The quantitative analysis of thin specimens: a
review of progress from the Cliff-Lorimer to the new zeta-factor methods. J Microsc (Oxford, UK)
221, 89-109.
WOOD, J.E., WILLIAMS, D.B. & GOLDSTEIN, J.I. (1984). Experimental and theoretical
determination of kαFe factors for quantitative X-ray microanalysis in the analytical electron
WU, J., KIM, A., MARVIN, R., MYERS, B., WOODRUFF, T., O'HALLORAN, T., DRAVID, V.,
MCILWRATH, K. & LI, S. (2010). Imaging and elemental mapping of biological specimens with
the hitachi HD-2300A dual-EDS scanning transmission electron microscope. Microsc Microanal
16(S2), 884-885.
GOLDSTEIN, J. I. and JOY, D. C. (Eds.), pp. 121-155. New York: Plenum Press.
ZALUZEC, N.J. (1981). On the geometry of the absorption correction in analytical electron
ZREIBA, N.A. & KELLY, T.F. (1988). Absorption and fluorescence corrections of characteristic
63
4. Electron Tomography
Images collected in the STEM are two dimensional (2D) projections of a three dimensional (3D)
volume, with information only collected in directions perpendicular to the electron beam. When
determining the structure of objects in the STEM, often assumptions are made about the
distribution of information in the direction parallel to the electron beam, which may or may not
be accurate. In order to fully determine the structure of a 3D object in the STEM information is
needed in the third dimension, this is where electron tomography plays a role.
dimensional images of the object. Typically, this involves reconstructing a three dimensional
volume from two dimensional images. This can take the form of reconstruction from a series of
projections through an object at a number of different angles (as is the case in X-ray computed
tomography (CT) (Hsieh, 2009) and electron tomography (Midgley, 2007)), or from a series of
slices through the object in question (as is the case in focussed ion beam (FIB) scanning
electron microscope (SEM) slice and view tomography (Dunn & Hull, 1999; Sakamoto, et al.,
1998)).
The form of tomography in which an object is reconstructed from its projections has its roots in a
1918 paper by Johann Radon (reprinted in (Deans, 2007)), in which he provides a proof that an
n dimensional object can be fully reconstructed from an infinite set of n-1 dimensional
projections. The Radon transform, R, of a two-dimensional object f(x) is a series of line integrals
To reconstruct the object an inverse Radon transform must be found and implemented on
collected data. In practice, a finite number of line integrals are collected at a finite number of
angles. Implementations of this form of tomography therefore offer an approximation to the full
solution of the object as defined by Radon. Projections in STEM are usually similarly assumed
to be formed from a series of line integrals through the object to each pixel in an image. In this
64
case, the intensity (I) in an image pixel is equivalent to the projection pj which, in the notation of
j j j j
where (x , y , z ) = r is the rotated coordinate system of the projection. The coordinate system of
the projection, rj, is related to the coordinate system of the object, r, through the three Euler
angles (θj, φj, ψj) of the projection, shown in Fig 4.1, via
1 0 0 (4.6)
and 𝐷𝜓𝑗 = (0 cos 𝜓𝑗 − sin 𝜓𝑗 ).
0 sin 𝜓𝑗 cos 𝜃𝑗
65
Figure 4.1. Diagram of two coordinate systems related by a set of Euler angles. The blue
coordinate system is equivalent to the coordinate system of the object and the red coordinate
The line integral assumption is a simplification of the true case, in which a projection is formed
via an integral over a double-cone through the object with its apex at the focal point of the probe.
The line integral approach is valid for small probe convergence angles and will be assumed
throughout this work. However, the double-cone approach may be necessary as the probe
convergence angle increases, which is becoming practically achievable due to the influence of
aberration correction, and has been successfully employed when combining tilt and focal series’
(Dahmen, et al., 2014). For thicker specimens the effect of beam spreading has an even greater
effect on the probe distribution than the probe forming angle, as discussed in section 2.4. This
effect could become important in obtaining accurate reconstructions from thick specimens,
Assuming that projections are acquired with the sample only rotated along a single axis so that
only θ is not zero, then the Radon transform can be seen to transform an object in a Cartesian
j j
coordinate system (x,y,z) to that of a coordinate system (x ,y ,θ) that is a representation of the
66
object in Radon space. A representation of an object in this way is often termed a sinogram, as
a point in Cartesian coordinates is translated to a sinusoidal curve in Radon space (Fig 4.2).
j
Figure 4.2.a) Illustration of the projection of a small dot f(x,z) on to a coordinate system (x ,θ) to
j
form a projection pj(x ,θj). b)The sinogram associated with a tilt series obtained for a point; the
j
slice highlighted in the sinogram corresponds to the projection in (a), pj(x ,θj).
Electron tomography in the STEM begins with the acquisition of a tilt series using one of the
imaging modes detailed in sections 2.2 and 2.3. Subsequent to acquisition, the tilt series’
images must be accurately aligned, as is detailed in section 4.4. After alignment, the tomogram
is reconstructed via one of a number of different algorithms (section 4.5). Finally, the tomogram
Visualisation is performed either (i) through extraction of two dimensional slices through the
tomogram, termed orthoslices if they are orthogonal to one another, (ii) through creating a
surface across a number of connected voxels, termed an isosurface if all voxels of the surface
have the same intensity, or (iii) through direct visualisation of the entire volume with different
67
4.3. Tilt Series Acquisition
Electron tomography is most often performed through acquisition of a series of images with the
sample tilted around a single axis, with angular increments typically between 1° and 10°,
depending on the imaging mode and sample limitations. The tilt increments are almost always
determined by the dose that the specimen can tolerate, due to the damage mechanisms
described in section 2.5, with imaging modes that require a larger dose per projection (EELS
and EDXS) often requiring a larger tilt increment to limit damage (Mobus, et al., 2003; Saghi, et
al., 2007).
For samples mounted on a standard TEM grid, the size of the holder within the polepiece of the
microscope and the position of grid bars often limits the extremes of the tilt to ± 70° or even
lower angles. This limit in the extreme tilt angles restricts the possible data collected and results
minimise the missing wedge have led to specimen holders which allow novel acquisition
schemes, primarily the dual-axis acquisition scheme that reduces the missing wedge to a
‘missing pyramid’ (Arslan, et al., 2006; Penczek, et al., 1995). The achievable angular range
can be extended to the full 180° when using a needle shaped sample attached to an on-axis
each imaging mode must conform to the projection requirement in order to obtain an accurate
reconstruction. The projection requirement states that the intensity detected must be a
monotonic function of the physical quantity to be reconstructed (Frank & Hawkes, 2006). Ideally,
the intensity should be a linear summation, as defined in equation 4.2, but nonlinearity can be
corrected if the function is monotonic (Yamasaki, et al., 2014). The use of a number of imaging
modes within electron tomography is outlined in this section, with a particular focus on
68
4.3.1.1. HAADF Tomography
The rapid growth of electron tomography studies in materials science over the last decade has
coincided with the development of HAADF tomography (Midgley & Dunin-Borkowski, 2009;
Midgley & Weyland, 2003). This is due to the extreme reduction in diffraction contrast when
using HAADF STEM in comparison with BF TEM. Thus, electron tomography of crystalline
specimens has been made possible without violating the projection requirement. Due to the Z
contrast of HAADF imaging, it has been used to distinguish between sample volumes of
different composition.
catalysts (Midgley & Weyland, 2003) and has seen widespread use for distinguishing heavy
metal nanoparticles on a lighter support material (Friedrich, et al., 2009; Weyland, 2002; Ziese,
et al., 2004). HAADF tomography of individual nanoparticles can be used to separate discrete
elemental phases, for example in well separated core-shell nanoparticles (Gomez-Grana, et al.,
nanoparticles has even been extended to atomic resolution (Chen, et al., 2013; Scott, et al.,
2012), where the application of separating discrete elemental phases is still valid, for example in
a core-shell PbSe/CdSe nanoparticle (Bals, et al., 2011) and an AgAu nanorod (Goris, et al.,
2013a) using the advanced reconstruction algorithms that are discussed in section 4.5.3.
A problem of HAADF STEM is the “cupping” artefact that arises due to the non-linear damping
of intensity (Van den Broek, et al., 2012). This has the effect of making the interior of objects
appear less intense than the exterior and can cause problems when separating regions of
different elements, for example, when segmenting core-shell nanoparticles. Van den Broek et al.
have proposed a method to correct this artefact, but it still remains as a problem in elemental
To date, the most widely used truly spectroscopic imaging technique is EFTEM tomography.
EFTEM was first applied to biological tomography studies (Koster, et al., 1997), using the zero-
loss image, to enhance image contrast. The first EFTEM tomography studies in the context of
materials science were performed by Möbus (Mobus, et al., 2003) and Midgley (Midgley &
69
Weyland, 2003) somewhat simultaneously. These two papers outline the distinct methods in
Single EFTEM image – Taking a single EFTEM image may be sufficient enough to
image free standing particles but there are stringent conditions as discussed by Möbus
Jump ratio – This involves taking two images, one at the maximum of the ionisation
peak and one just before the edge and then dividing the former by the latter. This
method is free from diffraction contrast but is not able to give quantitative information on
composition.
three images that are below, on and above the associated ionisation edge. The
although these images may still display diffraction contrast that affects the
reconstruction.
As discussed earlier, inelastic electron scattering should be particularly good for observing low-
Z elements. For example, nitrogen doping can be clearly resolved in carbon nanotubes (Florea,
et al., 2012). EFTEM tomography is not just useful for compositional analysis that complements
HAADF tomography. As well as filtering core loss electrons that give distinct elemental
fingerprints it is also possible to filter electrons that have losses associated with the formation of
possible to image different materials containing the same element. Plasmon loss images can
provide contrast between, for example, carbon nanotubes in nylon (Gass, et al., 2006) or silicon
nanoparticles in silica (Yurtsever, et al., 2006) that would not provide sufficient contrast from
mass-thickness based imaging modes. This method can also provide spectroscopic
measurements for materials of different elements in which HAADF contrast may be low, such as
Electron tomography using full spectral imaging in EELS was first reported in 2009,
2009). Subsequently, further studies have examined the feasibility of EELS tomography (Yedra,
et al., 2012) and compared imaging modes using EELS and EFTEM (Goris, et al., 2011).
70
Extension of 3D elemental reconstructions to a 4D spectrum volume that includes a full energy
loss dimension has been proposed (Goris, et al., 2014b; Yedra, et al., 2014), but it is unclear
physical meaning or meets the projection requirement. The full 4D spectrum volume has been
used to track changes in the oxidation state of Ce in 3D, to provide a 3D distribution of oxidation
states in a ceria nanoparticle (Goris, et al., 2014b). Plasmon imaging has also been utilised in
nanocube (Nicoletti, et al., 2013), although careful consideration of the specimen is needed in
this case.
Whilst EFTEM tomography has been reported regularly over the past decade, the majority of
reports of EDX tomography have come over the past two years. This is largely due to the
difficulties associated with tilt series acquisition that result from the EDX detector geometry
(Mobus, et al., 2003). Specifically, the detector is usually positioned at the same height as the
sample, or slightly above, and maximum counts are recorded when the sample is tilted to 15-
20°. Thus for most microscopes, in tilting through a large range of angles, as is required for
electron tomography, the number of counts is greatly reduced at angles away from the optimum.
In addition, as discussed in section 2.3, the restricted TEM polepiece means that EDX detectors
typically have a limited solid angle of 0.1 - 0.3 sr and therefore only collect a few percent of
The first report of EDX tomography was in the study of nanoparticles in a polycrystalline FeAl
matrix by Möbus (Mobus, et al., 2003). In this report, the difficulties of performing EDX
tomography were clearly illustrated. Images were acquired at tilt intervals of 10° owing to the
long acquisition times required and the shadowing of the detector by the sample holder led to
overall intensity variations over the range of tilt angles, therefore significant artefacts were
large intervals have been shown to be useful in a process known as ‘hybrid tomography’
(Mobus, et al., 2010; Saghi, et al., 2007) in which EDX data is combined with HAADF
71
compressed sensing approaches discussed in section 4.5.3, now offer the possibility to reduce
artefacts in reconstructions from projections at 10° angular increments (Leary, et al., 2013).
Other early EDX tomography studies utilised a needle shaped sample to avoid missing
projections associated with X-ray detector shadowing (Kotula, et al., 2007; Yaguchi, et al.,
2008). However, these studies still required long acquisition times (> 50ms per pixel dwell
times) due to the small solid angle of detectors and were undertaken at relatively large tilt
increments (5 or 10°).
Recent advances in EDX detector design have meant that EDX tomography has become a truly
viable alternative to EFTEM tomography with the advantage that the full spectral range is
collected simultaneously. The most pervasive detector design in these studies is the Super-X
detector, introduced in section 2.3.2. The four detectors increase the solid angle of collection to
around 0.7 - 0.9 sr and their symmetrical position around the optic axis means that X-ray counts
do not fall as sharply when tilting the sample. Developments in field emission gun (FEG)
technology and aberration correction have also led to a potential increase in beam current for
the same probe size. Therefore, it is possible to achieve higher counts in EDX at the same
resolution and a higher signal to background than in initial EDX tomography studies.
An early report of EDX tomography using the Super-X detector system investigated the
structure of a high-k dielectric transistor (Lepinay, et al., 2013). Using a needle shaped
specimen it was possible to acquire EDX maps at 2° for a full 180° tilt range and this provided a
reconstruction with largely reduced artefacts compared to reconstructions from far fewer
situated on carbon films on TEM supports grids (Genc, et al., 2013; Goris, et al., 2014a;
Liakakos, et al., 2014). In these papers no account was made for the detector shadowing due to
the sample holder or grid. However, there has been some interest in how this shadowing may
be characterised and compensated for in the literature, with a model to determine detector
shadowing put forward recently (Yeoh, et al., 2015). Additionally, an EDX tomography tilt series
has been acquired simultaneously to an EELS tilt series, paving the way for complementary
72
Even with advances in detector design, EDX spectrum images will continue to require a higher
overall electron dose than that required in HAADF STEM imaging. When studying beam
sensitive specimens high electron doses result in morphological changes over a tilt series that
invalidate reconstructions. For this reason, reconstruction algorithms such as the discrete
tomography and compressed sensing algorithms discussed in section 4.5.3, that allow
reconstruction from a limited number of projections, will be particularly useful. EDX tomography
is particularly suited to these algorithms as its data is easily separated in to discrete elemental
data.
thicknesses than HAADF STEM and particularly EFTEM/EELS. This is particularly useful when
considering that the thickness of standard “thin foil” TEM samples increases by up to (3x) in the
beam direction when tilting. Another advantage of EDX over EELS could be in atomic-resolution
spectroscopy due to contrast reversal and non-locality in EELS (Wang, et al., 2008). This
advantage will become particularly useful if atomic resolution discrete tomography (Van Aert, et
Invariably, after acquisition the acquired images in the tilt series must be aligned so that they
share a common coordinate system. Firstly, all projections must be aligned spatially so that the
coordinate system of each projection can be described simply by a combination of Euler angles,
j j
so that the origin of each projected coordinate system is the same (x ’,y ’ = 0). Secondly, in
single-axis tomography, the orientation of the tilt axis must be determined so that the
Spatial alignment of images in the tilt series in the most basic sense can be achieved through
simple cross-correlation (Frank & Brandt, 2006). The cross-correlation is almost always
performed between pairs of adjacent tilt images, as the similarity is largest between adjacent
images. Alignment of adjacent images can lead to the accumulation of errors, as small errors
can be propagated and accumulated through the tilt series (Frank, et al., 1987), and for this
reason cross-correlation often uses the central image as the initial reference and proceeds to
the tilt extremes in both directions. It has also been suggested to use an iterative method to
73
determine the projection to use as the initial reference that provides the best alignment (Jones &
Haerting, 2013), rather than simply starting from the central image. Alignment of images of
nanoparticles, or images that do not differ in the objects that they contain, may also be
performed via a variant of centre of mass determination (Scott, et al., 2012), although this
In bright-field TEM tomography of biological samples, often the image contrast is not good
enough for cross-correlation to provide an accurate result (Frank & Mastronarde, 2006). In
many cases, metal nanoparticles are used as fiducial markers which can be individually tracked
through the tilt series (Frank & Mastronarde, 2006). This not only allows spatial alignment of
images but allows accurate determination of the thickness of the sample and the tilt axis
orientation. Similarly, if image contrast is good enough, “patches” or features of the image itself
can be used as fiducials in much the same way (Brandt & Ziese, 2006).
If image alignment has not used fiducial markers, or a similar technique, to track the tilt axis
then the tilt axis orientation must be subsequently defined. The orientation of the tilt axis can be
found in the most straightforward manner by observing a sum of the tilted images, in which
image intensity forms a streak along the direction perpendicular to the tilt axis (Fig 4.3a). The
“series summation” technique will not work for spherical nanoparticles, or nanoparticles in which
all three dimensions are similar, however, as there will be no elongation of features at different
tilt angles. Determination of tilt axis orientation is also often performed using an arc minimisation
technique (Midgley, 2007). If the tilt axis is misoriented in the reconstruction, the reconstruction
will show the intensity of objects smeared out as arcs along the true tilt axis. A shift in the tilt
axis is therefore present as arcing in the same direction for all slices, whereas a misoriented tilt
axis that has its centre in the same position results in arcing in different directions above and
below the midpoint (Fig 4.3c-e). By performing reconstructions of discrete 2D slices along the tilt
axis it is possible to quickly visualise the presence of arcing and correct the alignment of the tilt
axis accordingly.
74
Figure 4.3. Representation of methods for tilt axis determination, using a sample of Au and Pd
nanoparticles on graphene. a) Projection at 0° tilt displaying the lines (in red) at which
reconstructions are performed to display the arcing in (c-e). b) Summation of the entire tilt series,
displaying streaking in the direction perpendicular to the tilt axis at an angle α to the horizontal.
c-e) Reconstructions of individual slices (i-iii) represented in (a) for c) a tilt axis shifted from the
correct orientation by 40 pixels, d) a tilt axis tilted from the correct orientation by 15° and e) a tilt
axis approximately at the correct orientation. The sample represented here is Au and Pd
75
4.5. Reconstruction
4.5.1. Backprojection
Subsequent to the alignment of a tilt series, the 2D data is then reconstructed into a 3D
tomogram. Reconstruction methods in tomography can be split into two categories, real space
exclusively begin with a backprojection operation. In real space, backprojection simply involves
projecting the intensity in a pixel of a projection back along a line at the same angle in which the
projection was taken. The intensity of voxels due to backprojection from a single projection is
defined by Frank and Radermacher (Frank & Radermacher, 2006) as multiplying the projected
pixel intensity by a 3D spread function (𝑙(𝑥 𝑗 − 𝑥 ′𝑗 , 𝑦 𝑗 − 𝑦 ′𝑗 , 𝑧 𝑗 )), related to the point spread
However, in many situations the spread function is ignored such that 𝑝𝑗𝑏 (𝑥 𝑗 , 𝑦 𝑗 , 𝑧 𝑗 ) = 𝑝𝑗 (𝑥 𝑗 , 𝑦 𝑗 ),
which can result in misrepresented intensity in nanoparticles, although this can be corrected
through measurement and subsequent application of the point spread function (Heidari Mezerji,
et al., 2015). Backprojection from a full set of projections is then performed by simply summing
the intensity from all projections, such that the backprojected intensity, b(x,y,z), is given by
series’ of a limited number of projections (> 2° intervals) through backprojection can result in
streaking artefacts (observed to different extents in Fig 4.4(b-d)), although it may be possible to
76
Figure 4.4. Backprojection of an object along a varying number of directions. a) Image of a two-
theorem. This theorem states that the Fourier transform of a projection of an object corresponds
to a central slice through the full three dimensional Fourier transform of the object at the
equivalent tilt angle (Fig 4.5). Considering this theorem it is therefore possible to perform a
can then be applied a 3D inverse Fourier transform to retrieve the reconstruction (Gilbert,
1972b).
77
j
Figure 4.5. a) The projected intensity p (x) of an object f(x,z) at an angle of θj. b) An illustration of
the angular sampling of an object in Fourier space under a single tilt acquisition scheme with a
missing wedge. The red line in (b) corresponds to the Fourier transform of the projected
j
intensity p (x) at the same angle θj.
The projection slice theorem also illustrates a problem with the standard backprojection
algorithm. A finite number of projections sample low frequency information at smaller intervals
than that of high frequency information and therefore low frequency information dominates
reconstructions using backprojection alone (Fig 4.6b). In order to compensate for this sampling
a weighting is often applied as a ramp filter in Fourier space, i.e. points in Fourier space are
multiplied by the distance from the tilt axis, that acts to give greater weighting to higher
(WBP/FBP) and is a very commonly used reconstruction technique, the effect of which is
illustrated in Fig 4.6c. In ‘filtered’ backprojection a low-pass filter is also applied to the Fourier
transform of the reconstructed object in order to limit high frequency noise which would be
enhanced by the ramp filter. Fig 4.5 also displays the missing wedge of information in Fourier
space that occurs as a result of the tilt limits imposed on some single-axis tilt series’. The effect
of this on the reconstructed intensity is a smearing of intensity at the extreme tilt angles and
elongation of the object in the z-direction, equivalent to the beam direction at 0° tilt (Fig 4.6d).
78
Figure 4.6. Reconstructions of the phantom object in (a) from projections at 1° intervals using b)
an unweighted backprojection operator from projections over the full tilt range, c) a weighted
backprojection operator from projections over the full tilt range and d) a weighted backprojection
operator with projections from between ± 70° displaying the missing wedge.
It has become more and more commonplace in electron tomography to improve on an initial
series data. This type of reconstruction uses successive reconstructions to produce theoretical
projections that are then compared to the initial acquired projections. The difference in
projections can then be projected back to obtain a difference reconstruction which can be
The first class of algebraic reconstruction techniques, generally known as the algebraic
reconstruction technique (ART) (Gordon, et al., 1970), compared a single projection from the
reconstruction with the corresponding original projection and made corrections only in that
direction, before moving on to the next projection. Whilst this method is simple and fast to
implement computationally it has been shown to produce results that diverge in the presence of
noise.
One of the most widely used classes of reconstruction algorithm is the simultaneous iterative
reconstruction technique (SIRT) (Gilbert, 1972a). Instead of making corrections after the
analysis of each individual projection, SIRT algorithms simultaneously compare all projections in
order to ensure any corrections made do not cause the reconstruction to diverge from the most
likely solution. With the advance of computer processing speeds, and the introduction of multi-
79
4.5.3. Advanced Reconstruction Techniques
Recently, there has been a lot of interest in using prior information to assist in the reconstruction
of electron tomography datasets. This prior information could be that there are only a few
intensity levels expected in the reconstruction (Batenburg, et al., 2009) or that the intensity in
the reconstruction should be found on a regular lattice (Goris, et al., 2013a; Van Aert, et al.,
basis, i.e. the reconstruction is assumed to be sparse in some basis (Leary, et al., 2013).
Enforcing sparsity is the equivalent of maximising the number of zero voxels in a reconstruction.
For example, a subset of the compressed sensing approach (including discrete tomography and
total variation minimisation) assumes that the reconstruction is made up of a small number of
grey levels and enforces this condition on the reconstruction (Goris, et al., 2013b). Total
variation minimisation assumes sparsity in the gradient domain, i.e. minimizes the number of
pixels in which there is a change in intensity. Discrete tomography can be viewed as an extreme
case of sparsity in the gradient domain, where not only sparsity is enforced but the intensity of
voxels in this domain is limited to a small number of values. These types of reconstruction
should be particularly useful for spectroscopic electron tomography in which only intensity from
a single element is reconstructed as a single grey level could be applied to assign voxels either
A compressed sensing approach can also be performed in the image domain, assuming that
the number of voxels containing intensity is sparse. This approach is particularly suitable for
imaging modes in which background pixel intensities are zero, such as EDX tomography. This
type of algorithm has been shown to produce high fidelity reconstructions from a few as 10
projections (Leary, et al., 2013), which will be particularly useful for EDX tomography in which
The resolution of a reconstruction from an electron tomography tilt series is dependent upon
many factors, including but not limited to: the initial image resolution (discussed in section 2.4),
the reconstruction volume, tilt angle interval, noise characteristics and reconstruction routine
(Midgley & Weyland, 2003). In all cases, the resolution along the tilt axis, which is defined here
80
as the y-axis with a resolution of dy, should be that of the initial projections if the tilt series is
aligned perfectly. In directions perpendicular to the tilt axis it is commonplace to define the
resolution using the Crowther criterion (Crowther, et al., 1970), which gives the resolution of a
backprojected reconstruction perpendicular to the tilt axis, dx and dz, equation 4.9.
𝜋𝐷
𝑑𝑥 = 𝑑𝑧 = , (4.9)
𝑁𝑝
where D is the volume of the reconstruction and Np is the number of projections. In many single
tilt-axis studies, and as displayed in Fig 4.6d, the missing wedge results in elongation of objects
in the z-direction. The resolution in z, dz, is elongated due to the missing wedge by a factor e xz
where θmax is the maximum tilt angle away from 0° at which projections are acquired in radians.
𝑑𝑧 = 𝑑𝑥 𝑒𝑥𝑧 , (4.11)
For a somewhat standard value of θmax of 70°, this results in an elongation in dz by a factor of
1.31.
In practice, equations 4.9 – 4.11 apply to backprojected reconstructions, but are known to
typically overestimate the resolution in real space of reconstructions using iterative techniques
(Heidari Mezerji, et al., 2011). The resolution of reconstructions using SIRT or other more
advanced techniques is determined only through measurement of the final reconstruction (Chen,
In section 4.3.1 it was outlined how STEM tomography is able to provide morphological and
spectroscopic information of samples at the nanometre scale. In addition, the use of both bright
field and weak beam dark field imaging in conjunction with electron tomography can provide 3D
reconstructions visualizing dislocations in materials (Barnard, et al., 2006; Liu, et al., 2014),
although there are stringent conditions to ensure the projection requirement is met. It is also
81
possible to scan a finely focussed precessed beam over the sample and hence to obtain the
three dimensional distribution of crystallographic grains in a sample via tilt series acquisition,
although this requires comparison of obtained diffraction patterns to a standard library (Gemmi
Three-dimensional information in the TEM does not necessarily require the acquisition of a
tomographic tilt series. A method related to tilt series electron tomography is single particle
images of particles of an identical structure are used as input projections for a 3D reconstruction
(Frank, 2002; van Heel, et al., 2000). This method has been recently applied in the physical
sciences to image the 3D structure of a single nanoparticle in a liquid, taking advantage of the
fact that the nanoparticle rotates to different orientations due to the motion of the fluid
surrounding it (Park, et al., 2015). Other three dimensional imaging methods in the STEM
include scanning confocal electron microscopy, using a large probe-forming angle to reduce the
depth of focus of the probe and extract information through a sample (Frigo, et al., 2002; Nellist
& Wang, 2012), which can also be used spectroscopically in combination with STEM-EELS (Xin,
et al., 2013).
(Blavette, et al., 1993; Kelly & Miller, 2007; Seidman, 2007). APT uses an applied voltage to
positionally sensitive detector, with the position of the atom determined to nanometre resolution
in all dimensions. The time-of-flight of the evaporated atom is used to determine the chemistry,
so that each atom is defined as a particular element or, in some cases, isotope. 3D information
from atom probe tomography has been successfully combined with data from STEM
tomography in a number of cases (Herbig, et al., 2015; Xiong & Weyland, 2014). For example,
provide 3D reconstructions of the same sample volume in an Al-Ag alloy (Arslan, et al., 2008).
A major drawback of all studies in the TEM (and APT studies) is the small sample volumes
analysed (sample thicknesses of a few hundred nanometres as a maximum). The small scale of
volumes analysed raises the question of whether the volume chosen is representative of the
bulk samples used in industrially significant processes. In the case of nanoparticle analysis,
82
while electron tomography fully characterises individual nanoparticles or small-scale
nanoparticle ensembles, questions may arise whether these are representative of the full
nanoparticle population. The use of characterisation techniques that probe a larger sample
volume at a lower resolution, alongside electron tomography, allows verification that the STEM
Three dimensional imaging in the SEM probes a larger sample volume albeit at a lower
resolution and is possible via sequential sectioning and imaging of one face of the material.
Serial removal of sample sections can be achieved either through use of an ultramicrotome
(Denk & Horstmann, 2004; Zankel, et al., 2009) or a focussed ion beam (FIB) (Uchic, et al.,
2007; Zankel, et al., 2014). The spatial resolution of the two-dimensional slices is limited by the
resolution of the imaging technique in the SEM and the resolution in the slicing direction is
limited by the slice thickness (approximately 10 nm in the FIB) (Mobus & Inkson, 2007; Villinger,
et al., 2012). Three-dimensional information in the SEM can include morphological information
through EDX spectroscopy and crystallographic information via electron backscatter diffraction
The FIB is also a useful tool for performing sample preparation for electron and atom probe
tomography, allowing preparation of both thin TEM lamellae (Giannuzzi, et al., 1998; Mayer, et
al., 2007) and needle shaped samples for TEM and APT (Thompson, et al., 2007). Additionally,
in contrast to sample preparation via electropolishing, the FIB allows accurate ‘site-specific’
At larger length scales X-ray tomography has become a highly versatile tool for 3D imaging of
materials (Maire & Withers, 2014; Stock, 2008). X-ray tomography is a similar technique to TEM
tomography, with a series of projections taken of a sample rotated at different angles using X-
rays as the illuminating radiation rather than electrons. The contrast in images formed by X-rays
is typically due to attenuation of X-rays, via the Beer-Lambert law expressed in equation 2.9 in
section 2.3.3.2, although other imaging mechanisms such as phase contrast (Cloetens, et al.,
1999; Momose, et al., 1996; Wilkins, et al., 1996) provide alternative contrast formation
mechanisms.
83
The collection of attenuated images in X-ray tomography is, in general, far quicker than that in
STEM tomography and the rotation of the sample can be performed with a higher degree of
accuracy in the rotated angle. For these reasons, it is possible to acquire a much greater
nanotomography, in which the resolution is below 1 μm, the time to acquire a single projection is
typically limited to minutes and therefore the total number of projections is typically no more
than a few hundred. In lower resolution studies high speed acquisition has allowed in-situ
studies of processes such as fatigue (Buffiere, et al., 2010) and porosity evolution (Babout, et
al., 2001).
The use of 3D imaging modes at larger length scales alongside STEM studies allows
verification that the STEM sample is representative of a bulk sample and puts any STEM study
in a wider material context (Mobus & Inkson, 2007). For instance, Tariq et al combined X-ray
techniques is a rapidly developing field (Lopez-Haro, et al., 2013; Thiele, et al., 2013; Yazzie, et
al., 2012) with great promise for answering many problems in materials science, geology and
4.9. Summary
transmission electron microscopy into three dimensional information. Through the use of
HAADF, EELS or EDXS imaging modes the distribution of elements can be characterised within
a single nanoparticle in three dimensions. The geometry of the holder within the polepiece of
the STEM and the placement of detectors limits the allowable angles at which an image can be
acquired. The maximum allowable dose also places limitations on the ability to acquire a large
number of images, particularly for the spectroscopic imaging modes. Novel algorithms present
an opportunity to overcome dose limitations, as they allow high fidelity reconstructions from a
84
5. Investigating STEM-EDX Tomography Methodology
5.1. Introduction
This chapter is concerned with examining the methodology of energy dispersive X-ray
tomography in the STEM and exploring the conditions under which the technique provides an
accurate reconstruction. The inclusion of large solid angle silicon drift detectors within the STEM
has led to a renewed interest in the technique and to a number of studies over the past few
years (Genc, et al., 2013; Goris, et al., 2014; Haberfehlner, et al., 2014; Lepinay, et al., 2013;
The key effects to consider are shadowing of the X-ray detectors and absorption of X-rays
within the sample. The extent of detector shadowing varies as a function of the sample tilt angle,
depending on the geometry of the sample holder and EDX detector, and is shown to introduce
artefacts into the reconstruction of a tilt series in a similar manner to the missing wedge effect
outlined in section 4.5.1. In order to compensate for detector shadowing, previous studies have
suggested multiplication by a correction factor (Yeoh, et al., 2015), or utilising a sample holder
in which detector shadowing is reduced to zero or negligible levels (Goris, et al., 2014; Lepinay,
apply an experimentally determined correction to the acquisition time of each spectrum image.
This chapter also considers the extent of X-ray absorption in a particular AgAu nanoparticle and
presents general levels of X-ray absorption in elemental media. X-ray absorption may cause
intensity variations across an elemental map that affect the validity of its use in tomographic
reconstruction. Similarly, variations in detector shadowing across an image area may cause an
overall intensity variation and the extent of this effect is considered in this chapter.
The AgAu nanoparticles used in this study were synthesized by Alexandra Macedo and Pedro
Camargo (Department of Chemistry at the University of Sao Paulo). The optical profiling of the
85
The results of this project have been submitted to Ultramicroscopy. The original manuscript for
presented below.
86
Paper: STEM-EDX Tomography of Bimetallic Nanoparticles: A
Methodological Investigation
a a c a a,b
Thomas JA Slater , Arne Janssen , Pedro HC Camargo , M Grace Burke , Nestor J Zaluzec ,
a
Sarah J Haigh
a
School of Materials, University of Manchester, Manchester, M13 9PL, UK
b
Nanoscience and Technology Division, Argonne National Laboratory, Argonne, Illinois 60439,
US
c
Departamento de Química Fundamental, Instituto de Química, Universidade de São Paulo,
Highlights:
dispersive X-ray (EDX) tomography within the scanning transmission electron microscope,
bimetallic AgAu nanoparticles. The detector collection efficiency when using a standard
tomography holder is characterised using a tomographic data set from a single nanoparticle and
compared to a standard low background double tilt holder. Optical depth profiling is used to
investigate the angles and origin of detector shadowing as a function of specimen field of view.
A novel time-varied acquisition scheme is described to compensate for variations in the intensity
of spectrum images at each sample tilt. Finally, the ability of EDX spectrum images to satisfy the
projection requirement for nanoparticle samples is discussed, with consideration of the effect of
1. Introduction
87
The characterisation of nanoparticles has been greatly enhanced over the past decade by the
tomography [1,2]. This technique involves tilting the sample to different angles along one or
more axes and collecting a series of images at each tilt angle. This data can be used to
reconstruct the three dimensional sample volume via an established tomographic algorithm.
Electron tomography reconstructions using tilt series data sets of transmission electron
microscope (TEM) or scanning transmission electron microscope (STEM) images have allowed
is more difficult. Using electron energy loss spectroscopy (EELS) or energy dispersive X-ray
(EDX) spectroscopy in the STEM it is possible to obtain elemental maps that show features
down to the atomic scale [6,7]. This two dimensional spectrum imaging, in which a spectrum is
collected for every pixel, is now routine but, the combination of spectroscopic imaging with
electron tomography, has proved far more challenging experimentally. Practically, the principle
limitation when acquiring both EELS and EDX tomographic spectrum image tilt series are the
long acquisition times associated with even a single spectrum image [8].
New EDX detector geometries in the transmission electron microscope are capable of EDX
spectrum imaging at a wide range of tilt angles and with a significantly improved solid angle for
X-ray collection, greatly increasing the ability to acquire EDX tomography data sets [9,10]. In
particular, the Super-X detector configuration, composed of four separate silicon drift detectors
(SDDs) arranged symmetrically around the optic axis, has proved capable of performing EDX
tomography with much shorter acquisition times at each specimen tilt angle [11,12]. However,
even for these new generation large solid angle detectors, the geometrical percentage of X-rays
detected is low (approximately 6% for 0.7 sr), such that the signal to noise ratio of EDX
spectrum images is often poor. The low signal detection means that very high electron doses
are often required, through the use of high probe currents and/or long acquisition times. In the
9 2
tilt series discussed in this work a total dose of approximately 7x10 electrons/nm was
employed, similar to that reported for a previous study of a single transistor published by
Lepinay et al [12]. This will often limit the application of the technique, as many nanoparticles
are insufficiently robust to withstand this high total dose without significant structural change. An
important aim of experimental procedures in EDX tomography should therefore be to limit the
88
overall electron dose whilst maximising the signal at each angle. Limiting the number of
projections to as few as possible has the advantage of improving the quality of the spectrum
image data for each tilt angle and is therefore highly desirable, although requiring the use of
advanced reconstruction algorithms [13,14] to realise high fidelity reconstruction when few
Historically, one of the key challenges for EDX tomography has been that traditional single
detector EDX systems are often limited to a very narrow range of tilt angles [8]. At other angles
the penumbra of the specimen holder prevents an X-ray signal being detected by ‘shadowing’
the EDX detector [8]. The design of new detector geometries such as the Super-X detector
system [15] means that X-rays may be detected for a wider range of specimen tilt angles (+/-
70°), although some detector shadowing does occur for samples that are deposited on standard
TEM grids or that use traditional high-tilt tomography holders [16]. Shadowing at all tilt angles is
only eliminated through the use of 360° rotation tomography holders [12] and these are not
compatible with all types of specimen. Shadowing of X-ray detectors will cause intensity
variations as a function of tilt angle and will lead to systematic errors in the intensity contribution
This paper investigates a novel methodology for EDX tomography of isolated nanoparticles. As
an example we have used AgAu nanoparticles for which EDX tomography has already been
demonstrated to give important insights into the different elemental distributions as a function of
overall composition [17]. The details of detector shadowing are investigated and a varied-time
acquisition scheme is introduced to compensate changes in detector solid angle. Finally, the
ability of EDX spectra to satisfy the projection requirement for nanoparticle samples is
2. Experimental Methods
High angle annular dark field (HAADF) scanning transmission electron microscope imaging was
performed on an FEI probe-corrected Titan G2 80-200 S/TEM with a high brightness X-FEG
electron source and Super-X energy dispersive silicon drift detectors (SDDs). The microscope
was operated at an accelerating voltage of 200 kV with a beam current of 0.5 nA, a
89
convergence angle of 21 mrad and a HAADF acceptance inner angle of 50 mrad. EDX
spectrum images were acquired in the Bruker Esprit software using the Titan’s Super-X detector
system with a dwell time of 30 μs and a typical image size of 512x512 pixels. This detector has
2
a total solid angle of approximately 0.7 sr distributed over four 30 mm SDDs equally separated
at 45° from the holder tilt axis at an elevation angle of approximately 18° from the horizontal [15]
(Fig 1). Full spectrum image datacubes were exported from Esprit to Gatan’s DigitalMicrograph
software. As individual pixels were too noisy for accurate quantitative analysis, elemental X-ray
counts for each tilt angle were extracted from spectra obtained from a summation over all pixels
in the full spectrum image. Background subtraction was performed via a window method using
energy windows (of the same width as that of the signal window) before and after the peaks of
interest, taking care to ensure that no other peaks are included in these regions [18]. Elemental
maps were extracted at energies of 2.92 keV – 3.06 keV (Ag Lα) and 9.58 keV – 9.82 keV (Au
Lα) and are shown after applying a 5-pixel smoothing (as detailed in section 3.2).
A Fischione 2020 single tilt tomography holder was compared to an FEI high-visibility low-
background double-tilt specimen holder. EDX tomography used an angular increment of 10°
and a tilt range of ±70°. Traditional tilt series data sets were obtained with a constant acquisition
time of 300s for each spectrum image while in our novel time-varied tomographic scheme
acquisition times were varied between 236s and 895s in order to compensate for the calibrated
detector shadowing at the particular specimen tilt. Manual alignment was used to centre the
specimen at each tilt angle and HAADF images were acquired simultaneously with spectrum
images at each tilt angle. Additional intermediate HAADF images were collected every 5° to
assist with subsequent alignment of the data set via cross-correlation of HAADF images. Spatial
image alignment was performed using cross-correlation of HAADF images in FEI’s Inspect3D
software package. Tilt axis alignment was also undertaken in Inspect3D followed by using a
ImageJ [20] and FEI’s Avizo software platforms were used for visualisation of reconstructions.
To investigate the effect of variable intensity data sets a phantom was simulated in MATLAB
radon function was used to simulate projections between ±70° at 10° intervals. The inverse
90
radon function was then used to perform a filtered backprojection, firstly, on projections of the
same intensity and, subsequently, with projections multiplied by the same factor as the intensity
differences provided by shadowing from a single tilt tomography holder on all four Super-X
detectors (Fig 2c), as well as multiplication factors taken from the variation in intensity from only
Three-dimensional optical images of the Fischione 2020 holder were acquired with a Keyence
VK-X210 3D laser scanning confocal microscope. The microscope is equipped with a violet
laser (wavelength 408 nm) and a 16-bit photomultiplier for accurate detection of reflected light
on surfaces. High-resolution three-dimensional (3D) maps of the top and bottom surfaces of the
holder were obtained by stitching 54 single images of the top side and 25 single images of the
bottom side together. The assembled three-dimensional map of each surface was analysed with
AgAu nanoparticles were chosen as a suitable test system for this study because of their
robustness to the structural changes induced by the electron beam and from the presence of
two elements with characteristic X-rays in different parts of the X-ray spectrum. These particles
-
were synthesised via a galvanic replacement reaction between Ag nanoparticles and AuCl 4 (aq),
as detailed in previous work [21]. This process leads to nanoparticles with a range of
morphologies including hollow donut-like structures like that shown in Fig 2a. At Au contents
less than approximately 20 at% the gold is segregated to the nanoparticle surface while at
higher Au contents the nanoparticle surface is enriched in Ag. A nanoparticle sample of average
deposited from solution onto continuous carbon-coated copper 200-mesh TEM grids (from Agar
Scientific, product code AGS160). Care was taken to acquire tilt-series’ of nanoparticles in the
In order to account for detector shadowing, when using a standard tomography holder, the
extent of shadowing at each angle must first be accurately calibrated. Doing this requires a
sample that will itself give a constant X-ray signal independent of tilt angle. Robust single
91
nanoparticles like that shown in Fig 2a present a suitable specimen from which X-ray counts
associated with the elements present within the nanoparticles should not significantly vary with
specimen tilt angle. Any variations in X-ray peak intensity from spectral images of single
Figure 1. Schematic illustrating the geometry of the Super-X detectors in the FEI Titan G2 80-
200 (S)TEM. (a) viewed in the direction of the electron beam and (b) viewed for a 2D cross-
section passing through detectors 2 and 4 at 45° to the long axis of the specimen holder.
Figure 2. Calibration of Super-X detector shadowing in the Titan G2 for the Fischione 2020
single tilt tomography holder. a) EDX Au and Ag elemental map for a single AgAu nanoparticle
at 0 degrees showing the particle that was used to perform the detector shadowing calibration,
b) background subtracted Au Lα X-ray counts (9.7 keV) at each pair of detectors as a function
of tilt angle, taken from EDX spectrum images acquired for the same AgAu nanoparticle using
an acquisition time of 5 mins, 10° tilt increments and an angular range of ±70°, c) background-
subtracted summed X-ray counts for Au Lα (9.7 keV) and Au Mα (2.1 keV) peaks.
92
In order to characterise the variation in total X-ray count rates as a function of tilt angle,
spectrum images were acquired with a Fischione 2020 single-tilt tomography holder using the
Titan’s Super-X EDX detector system (Fig 1) for a single AgAu nanoparticle. Spectrum images
were acquired for 5 minutes every 10° for a tilt range of ±70° (Fig 2b,c). Detectors mounted on
either side of the holder tilt axis will show similar behaviour and the response of detectors 3 + 4
is symmetrical to that of detectors 1 + 2 (Fig 2b). The response of these pairs of detectors is
similar to that of a standard geometry single SDD detector [16]. The full detector response (Fig
2c) consists of the sum of detectors 1-4. Perhaps surprisingly, this data demonstrates that for
this specimen holder the poorest X-ray detection efficiency occurs at low specimen tilts. At 0°
the Super-X detector collects only 30% of the X-rays measured at a specimen tilt of 60° for the
same nanoparticle, due to the penumbra of the Fischione 2020 tomography holder. The Au Lα
(9.7 keV) and Au Mα (2.1 keV) characteristic X-ray peaks are found to display an almost
identical relationship with specimen tilt angle (Fig 2c), demonstrating that detector shadowing
does not vary with X-ray energy for this energy range. This suggests that the same
compensation of shadowing can be used for all characteristic X-ray peaks above 2.1 keV.
Shadowing of the X-ray detectors is possibly due to either the penumbra of the sample holder or
the bars of the copper TEM grid, or a combination of both on different sides of the holder. In
order to determine the source of shadowing, and the angles at which X-rays are shadowed in tilt
series, we have taken an optical profile of the top and bottom surfaces of the 2020 holder. A
three dimensional image of the surface of the Fischione holder is produced (Fig 3a,c), in which
line profiles of height variations of the holder give the angles at which the holder shadows any
emitted X-rays. The detectors are located at 45° angles azimuthally from the tilt axis (Fig 1a)
and therefore we have extracted line profiles at an azimuthal angle of 45° from both sides (Fig
3b,d). Initially we assume the nanoparticle specimen is located on the carbon film in the centre
of a grid square in the middle of the holder. We also assume that the carbon film of the grid lies
flat on top of the grid, the square holes have a width of 90 μm, the grid bar height is 20 μm and
the grid is oriented at 45° to the tilt axis. We neglect bowing of the carbon film; a phenomenon
which is known to occur for a number of different grids from a range of suppliers and which may
increase with the large dose supplied to the area surrounding the nanoparticle. The line profiles
in Fig 3 reveal a polar angle of 21° from the specimen to the top surface of the holder and 18° to
the bottom surface. In comparison, the 200 mesh copper TEM support will provide no
93
shadowing above the sample and shadowing up to an elevation angle of 24° below the sample.
Thus for the somewhat idealised situation described above, shadowing above the sample
occurs at a polar angle of 21° due to the sample holder and shadowing below the sample
occurs at a polar angle of 24° due to the grid bars (Fig 4). Use of these angles for shadowing
assumes point detectors at 45° azimuthal angles to the tilt axis, but we note that this is an
approximation as the finite detector size will cause shadowing to vary over the width of the
detector.
94
Figure 3. Optical profiles of the top and bottom surfaces of the Fischione 2020 holder. a) 3D
profile of the top surface of the holder with the direction of line profile, b, illustrated. b) Line
profile of plane indicated in a, displaying a polar angle of approximately 21° from the centre of
the grid to the highest point of the top surface of the holder. c) 3D profile of the bottom surface
of the holder with line profile d illustrated. d) Line profile of plane indicated in c, from which a
polar angle of approximately 18° from the centre of the grid to the bottom surface of the holder
is found.
95
Figure 4. Schematic diagram showing the elevation angles from the sample at which the holder
and grid bars will shadow emitted X-rays above and below the specimen in the direction of a
The angle intersected by each detector can be calculated through prior knowledge of the total
solid angle and the detector areas. Given the Super-X detector system is characterised by a
2
total solid angle of approximately 0.7 sr and a single detector area of 30 mm , the distance of
the detectors from the optic axis is calculated from simple geometry as approximately 12 mm.
where ra is the radius of a single detector and d is the radial distance of the detector from the
area of interest. Equation 1 can be used to estimate the distance from the sample to the
detector and therefore allows determination of the polar angle between the uppermost and
lowermost parts of the detector. This polar angle is approximately 27° in this case, although this
neglects detector tilt which is known to be present and which means that the angular extent of
the detector is not equal when considering the polar and azimuthal subtending angles. We have
determined the elevation angle of the detector using the minimum in X-ray counts observed for
our tilt series. Fitting Gaussians to the data in Fig 2b gives minimum counts at an average α-tilt
of 18° for each pair of detectors. A polar angle, θ, in the plane of the detectors, can be
translated to a polar angle, α, in the plane normal (perpendicular) to the α-tilt axis by the
equation:
96
𝑡𝑎𝑛𝜃
𝛼 = tan−1 . (2)
𝑐𝑜𝑠45°
Thus where θ = 27°, 𝛼 =36° and given the elevation angle of the centre of the detector normal to
the α-tilt axis is 18° this predicts that the detector subtends an angle from 𝛼 =0° to 𝛼 =36 °. This
suggests that each pair of detectors should be fully shadowed when the sample holder is tilted
to the angular range of ± 8° to ± 32° (given the holder shadowing shown in Fig 4 and converting
the θ angles (21 and 24°) to 𝛼 angles in the detector plane (28 and 32°)). These values appear
to qualitatively fit the tilt-series data corresponding to normalised detector counts of less than
10%.
The tilt series data is also compared to the detector shadowing model proposed by Yeoh et al
[23] (Fig 5a). We have used a value of δ = 5° for the detector tilt angle and values of d = 10.5
mm and r = 2.9 mm for the distance to the detector and the radius of the detectors respectively.
The data here fits best with their proposed model, for the shadowing angles we have
determined, when a detector elevation angle of 16° is used (Fig 5a). The counts at negative tilt
angles appear to fit well qualitatively to the detector model, although the counts at positive tilt
angles do not. This may be due to a discrepancy between the detection efficiencies of the two
detector pairs. This model also allows consideration of the effect of analysing a nanoparticle
away from the centre of a grid square as shown in Fig 5. If the nanoparticle is away from the
centre of the grid square the tilt series becomes asymmetric, although not in the same manner
as the data acquired, with the counts of tilts in one direction raised whilst the counts of tilts in
97
Figure 5. Modelling detector shadowing using the model of Yeoh et al [23]. a) Plot of normalised
counts with respect to tilt angle for the positions within a grid square indicated in (b).
Experimental data points for Au Lα (9.7 keV) X-rays (reproduced from Fig 2c) are also shown
grid square with black in the centre, blue 10 μm from the edge towards detector 1 and 45 μm
from the edge towards detector 2 and red 10 μm from both edges towards detector 1 and
detector 2.
A tilt series of a similar AgAu nanoparticle using an FEI high-visibility low-background double-tilt
holder was also acquired for comparison of shadowing using a standard specimen holder (Fig
6). In comparison, the non-tomographic holder displays the reversed dependence on tilt angle,
i.e. low tilt angles give the least detector shadowing over the full tilt range of this holder (±30°).
This is due to the low-profile of the top of the holder that is designed specifically not to shadow
the Super-X detectors and is consistent with similar measurements [24]. Whilst this holder
displays a lower variation in counts around 0° tilt, its limited tilt range (±30°) would lead to large
98
Figure 6. Calibration of Super-X detector shadowing in the Titan G2 for the FEI high-visibility
low-background double tilt holder using a similar particle to that shown in Fig 2a. a) Au Lα
keV) and Au Mα (2.1 keV) X-ray counts extracted from EDX spectrum images acquired for the
same AgAu nanoparticle using an acquisition time of 5 mins and 5° tilt increments for angles of
±30°.
reconstruction
Variations in the intensity of different projections at different tilt angles can lead to artefacts
simulated two-phase object. Fig 7b shows the object reconstructed using 14 simulated
projections each with the same total intensity (10° tilt intervals over an angular range of ±70°). In
comparison, Fig 7c shows the same object reconstructed from projections where the total
intensity varies in a similar way to that predicted by the shadowing variations revealed in Fig 2c
for the 2020 single-tilt tomography holder, when used with the Super-X detector system. To
further illustrate the effect, the projections were backprojected with intensity variations following
those of only one pair of Super-X detectors (1+2) (Fig 7d), as calibrated in Fig 2b. This
illustrates the effect of using a conventional single side-mounted EDX detector. All
reconstructions show streaking artefacts associated with a missing wedge of projections and
large angular intervals respectively, causing the spherical phantom to become “lemon-shaped”
(with “lumps” in the direction of the missing wedge). However, the varied intensity
reconstructions also show noticeable variations in intensity, where the intensity of the outer ring
is diminished in the plane of the projections with lower intensity (as indicated by the white
99
arrows in Fig 7b,c). When considering this effect in Fourier space, it can be thought of as
weighting more strongly features in central slices away from the slice at k z = 0 (as represented
in Fig 7 e-f), thus reducing contrast for features in the central slices around that at k z = 0. Where
there is sufficient signal within the shadowed regions, these artefacts can be avoided by
100
Figure 7. Demonstration of artefacts associated with variations in projection intensities. a)
simulated projections of the two phase image in (a) for a tilt range of ±70° and angular intervals
of 10°. b) Reconstruction using projections of the same intensity, displaying missing wedge and
streaking artefacts only. c) Reconstruction applying the intensity variations due to shadowing
(measured when employing all 4 Super-X detectors with a single tilt tomography holder, as
shown in Fig 2c). d) Reconstruction applying the intensity variations from only a pair of Super-X
detectors on one axis, as shown in Fig 2b. (e-f) Representation in Fourier space of the
projections. f) Variable intensity projections weighted for the full Super-X detector geometry. g)
101
3.3. Methods for compensating detector shadowing
We propose two methods by which the shadowing of the Super-X detector system can be
compensated and variations in the count rates at each specimen tilt angle can be reduced. The
first is to acquire spectrum images for a constant time at each tilt angle and then multiply each
spectrum image by a ‘shadowing compensation’ factor that normalises the total counts, as
suggested in the early work of Möbus [8]. This is straightforward for samples in which the total
counts do not change with specimen tilt, such as single nanoparticles, but for samples in which
the total counts change with tilt angle, such as thin films, a calibration sample would be needed.
The other drawback of this approach is that the signal-to-noise ratio across projections will vary
and simply multiplying the lowest intensity projections, or alternatively scaling down the highest
intensities, will retain the low signal-to-noise ratios that are present for these projections.
Therefore, although this approach has the advantage of simplicity, it will not present the most
accurate acquisition scheme. An alternative is to adjust the acquisition time at each tilt angle so
as to achieve a constant total specimen X-ray signal for each projection. This allows the
maximum tolerable electron dose to be optimally distributed over all projections or can be used
to minimise total acquisition time. The disadvantage of this approach is that it requires prior
knowledge of the detector shadowing for a particular sample holder and microscope
combination: obtained using a separate tomographic data set. This calibration can be obtained
either by measuring the time taken to acquire a fixed number of X-ray counts for a specific peak
within summed spectrum images at each tilt angle or by comparing the different X-ray signals
for summed spectrum images obtained with a constant acquisition time as a function of holder
tilt (as illustrated in Figs 2 and 4). We note that as different sample holders have different
geometries and as there are variations in the position of EDX detectors on different instruments,
calibration of the detector shadowing will need to be performed for each sample holder and
microscope combination in order to derive specific time varied acquisition schemes. This
calibration will also differ for different positions within a grid square, as illustrated in Fig 5, so the
We have acquired EDX spectrum images every 10° using variable tilt-dependent acquisition
times as shown in Fig 8a. The Au Lα, Au Mα and Ag Lα specimen X-ray counts displayed far
smaller variations than for a fixed-time acquisition scheme for the whole range of specimen tilt
102
angles (intensity variations were within ±15 %) demonstrating the success of our time-
dependent acquisition scheme for reducing variations in X-ray signals (Fig 8b). However, even
these remaining small variations in summed X-ray counts are larger than can be explained by
statistical noise fluctuations alone and could be caused by two possible effects. The pixel dwell
time and spectrum image size are constant for all tilt angles and this dictates the time increment
to which the acquisition time can be specified for each tilt angle. For example, a dwell time of 30
μs and an image size of 512x512 pixels results in a discrete minimum time increment of 8s.
Small variations in the X-ray counts are therefore likely to be due to the discrete time
increments enforced in the data acquisition. Proportionally, this will have the largest effect for
spectrum images acquired with the shortest acquisition times, so for our data will cause the
greatest errors at high tilt angles. The accuracy of the shadowing calibration will also affect the
accuracy of the X-ray count rates obtained in the time-dependent data series. Low X-ray signals
in the initial fixed time data are likely to cause larger errors when predicting optimal acquisition
times. For our holder-microscope combination the spectrum images around 0° tilt have the
lowest X-ray signals (approximately 6000 total Au-Lα counts at 0° tilt compared to
approximately 19000 counts at 70°). Our calibration shows the largest deviation at small tilt
angles (especially at -10° as shown in Fig 8b, c) this latter effect is likely to be the more
significant for our data. We also note that for very high count rates the ratios of pixel dwell time
to detector processing time and count rate can result in lost data, thus also potentially
contributing to this variability. The combination of dwell time (approximately a few tens of μs),
detector time constant (approximately 2 μs) and count rates (< 10Kcps) in our experiments did
103
900
800
700
600
Time (s)
500
400
300
200
Ag L
60000 Au L
Au M
50000
40000
Counts
30000
20000
10000
0
-80 -60 -40 -20 0 20 40 60 80
Angle ()
10000 1.00
0.99
9000
0.98
Proportion of time
Ag L Counts
8000 0.97
0.96
7000 0.95
0.94
6000
0.93
5000 0.92
-80 -60 -40 -20 0 20 40 60 80
Angle ()
Figure 8. a) Total acquisition time at each angle employed in the time-varied acquisition scheme
reported in this study. b) Au Lα, Au Mα and Ag Lα X-ray counts at each specimen tilt angle
under the varied acquisition time scheme. The variation in counts of each elemental peak using
the time-varied acquisition scheme are significantly reduced (variation < ±15%). c) Comparison
of Ag Lα counts at each specimen tilt angle and to the proportion of the calibrated time the
104
The accuracy of the shadowing calibration is poorer where the collection efficiency is low. This
accuracy could be improved by repeating the calibration using longer acquisition times or by
measuring the time required to reach a certain value of specimen X-ray counts.
The choice of tilt increment for the EDX tilt series data set is a compromise determined by
factors such as sample stability, achievable tilt range and X-ray count rates. For a fixed
tolerable electron dose, larger tilt increments, such as 10°, allow longer acquisition times for
each specimen tilt angle, resulting in improved signal to noise ratios within the individual
spectrum images. However, large tilt increments have been shown to limit the fidelity of
reconstruction algorithms may allow high fidelity reconstructions even when using large tilt
intervals of 10° or more [13], but this then poses a problem in the alignment of tilt series’.
Alignment procedures based on cross-correlation are less accurate when aligning images taken
every 10° than for smaller tilt intervals due to the larger differences between subsequent images
[27]. Difficulties with alignment of low signal-to-noise ratio spectrum images can be overcome by
performing image registration using the simultaneously acquired HAADF images and
subsequently applying this registration data to the elemental maps. Supplementary HAADF
images can be acquired at smaller tilt increments to aid registration without significantly
increasing the total acquisition time for the data set. To aid alignment we have acquired HAADF
images at every 5° to assist in the alignment of EDX maps acquired at every 10°.
The majority of EDX tomography studies have employed smoothing functions to the obtained
elemental maps to partially compensate for low signal intensity and a poor signal-to-noise ratio
[11,12]. To investigate the effect of smoothing filters on EDX elemental maps it is desirable to
work with a simulated data set. In this work we have simulated EDX maps using a test object
consisting of a pair of concentric rings; representing a hollow two-phase nanoparticle with the
outer phase having a higher concentration of the elemental being mapped (Fig 9a(i)). To match
the signal to that observed in typical experimental data, the intensity of the image was set to
zero for 30% of pixels selected at random from the whole field of view. An equivalent proportion
of zero value pixels are typically observed in experimental Au Lα elemental maps acquired over
105
300 s for the nanoparticles used in this study (Fig 9c). Noise was then added, equivalent to 20
counts per 100x100 pixels, producing a simulated EDX map (Fig 9a(ii)) which provides a good
match to the experimental data shown in Fig 9c. A 5-pixel smoothing window (Fig 9a(iii)) is
shown to suppress noise and give a signal which closely resembles the original test object.
Lepinay et al report that an edge-preserving smoothing filter suppressed noise well for their test
images [12]. However, using this filter on our test image we have found no significant
improvement from a generalised smoothing filter and also the production of negative intensity
artefacts (Fig 9a(iv) and 9b(iv)). We have therefore employed a 5-pixel smoothing to the
106
Figure 9. Producing a simulated EDX elemental map which closely resembles the experimental
data. a) i) Initial test image of object representing a two phase nanoparticle. ii) Simulated
elemental maps in which 30% of data points have been set to zero and Poisson noise added. iii)
Image b processed with 5-pixel smoothing window showing effective suppression of noise in the
image. iv) Image b processed with an edge preserving filter demonstrating the appearance of
negative intensity artefacts within the nanoparticle. All images are shown with the same
intensity scale and are 512x512 pixels. b) Line profiles (i-iv) through the images (a(i-iv))
107
respectively at the positions indicated by the arrows. c) Line profile taken from representative
experimental dataset with the corresponding Au Lα map of an AgAu nanoparticle shown inset.
We note that the signal-to-noise ratio for individual spectra can be improved by binning spectral
images or acquiring smaller-sized spectrum images at lower magnification. This has the
advantage that each pixel contains a more reliable spectrum that can undergo background
subtraction and may also assist with other spectral processing procedures such as multivariate
statistical analysis (MSA) [28,29]. The disadvantage is that binning of spectral images is likely
to degrade spatial resolution and may also reduce the ability to perform advanced
The time-dependent acquisition scheme illustrated in Fig 8 has been successfully applied to
Lα are shown in Fig 10(i-ii) at tilt angles of -60°, 0° and 60°. The spectra from the full spectrum
108
Figure 10. Projections at a) -60°, b) 0° and c) 60° for i) Au Lα counts and ii) Ag Lα counts. iii)
Fig 11 shows orthoslices and a volume render for one of these reconstructions, demonstrating
the segregation of Ag to the surface at a composition of 60 at% Ag and 40 at% Au. A discussion
of the correlation between compositional segregation and catalytic activity for this system can
be found in Slater et al [17]. The reconstructions here represent counts from only one peak in
each case (Au Lα and Ag Lα) but the counting statistics could be improved by using the full
family of peaks. However, care should be taken to avoid overlapping peaks, as is the case for a
109
Figure 11. Example of a tomographic reconstruction using a time-dependent acquisition scheme
performed for a AgAu nanoparticle. a) Orthoslice through the Ag reconstruction normal to the
optic axis, displaying clear segregation of Ag to the surface of the nanoparticle. b) Orthoslice
through the Ag reconstruction parallel to the optic axis. c) Orthoslice through the Au
reconstruction normal to the optic axis. d) Orthoslice through the Au reconstruction parallel to
the optic axis. e) Volume rendering of the Ag and Au reconstructions, Ag volume is green and
Au volume is red.
Previous studies have also successfully applied a standard ‘constant time’ acquisition approach
systems [11,31]. However, there has been little discussion of the ability of the EDX signal
The projection requirement of tomography states that the intensity of the signal used to perform
the reconstruction must be a monotonic function of the quantity to be reconstructed [32], and
direct proportionality between intensity and the physical quantity is desirable. A key question for
EDX tomography is therefore whether the characteristic X-ray intensity generated in the STEM
meets this requirement for the constraints of a particular sample and detector geometry.
The intensity of the characteristic X-ray signal for element A is related to the probability of
ionisation. Where nA is the number density of atoms of element A within the sample, QA is the
110
ionisation cross section and t is the sample thickness, the probability of ionisation for an atom of
The expected characteristic X-ray generation can then be calculated by consideration of the
competition between X-ray emission and other de-excitation methods, such as Auger electron
emission, defined as the fluorescence yield (ωA) [18]. The intensity of the X-ray signal for a
Ω (3)
𝐼𝐴 = n𝐴 Q 𝐴 t ω𝐴 𝑎𝐴 𝐷𝑒 ( ) 𝜀𝐴 ,
4𝜋
where aA is the relative transition probability, De is the total number of electrons incident on the
The only quantities in equation 3 which vary as a function of the two dimensional position within
a spectrum image are the number density of atoms of element A (nA) and the thickness of the
sample. The intensity of the characteristic X-rays described in equation 3 therefore fully satisfies
the projection requirement; the detected signal is directly proportional to the mass-thickness of
element A. The exception to this is when crystals are oriented with the electron beam passing
along a major zone-axis, resulting in the electron beam being more tightly bound to atomic
columns due to electron channelling [28]. For this reason, a quantitative analysis should avoid
channelling conditions or exclude projections of crystals close to major zone axes from
reconstructions.
X-rays are produced not only through emission of characteristic X-rays from atoms within the
sample but also through the deceleration of the electron beam in the sample. Bremsstrahlung,
with an increase in emitted X-ray energy. A sufficient signal-to-background ratio at each pixel
simple two-window method [18]. If background subtraction is not performed, the contribution of
Bremsstrahlung will affect the intensity of elemental maps produced at each tilt angle and
therefore potentially influence the final tomographic reconstruction. The Bremsstrahlung signal
scales proportionally with atomic number, Z, and therefore could cause several artefacts. For
example, low energy X-rays may contain erroneous intensity in regions of high Z. For
111
reconstruction of high energy X-ray peaks the Bremsstrahlung background is low and the
Equation 3 also neglects the effect that X-ray absorption and fluorescence within the sample will
have on the intensity of characteristic X-rays emitted. Here, we will consider the implications
that absorption will have on the projection requirement. For absorption to negatively affect the
ability of the sample to accurately satisfy the projection requirement, the absorption must vary
across the spectrum image. The absorption of X-rays is described by an exponential attenuation
law [18]:
𝜇
−𝜌𝐿( ) (4)
𝐼 = 𝐼0 𝑒 𝜌 ,
where I/I0 is the fraction of X-rays not absorbed at a particular energy, ρ is the density of the
𝜇
element and ( ) is the mass attenuation coefficient of the element. Equation (4) can be
𝜌
rearranged to calculate a maximum path length through the sample, L, for a fixed amount of
𝐼
𝑙𝑛 𝐼0 (5)
𝐿= 𝜇 .
𝜌( )
𝜌
Fig 12 uses elemental mass attenuation coefficients and standard specimen densities [33] to
predict the maximum allowable X-ray path length for a single element sample as a function of
atomic number in the limit of 1% absorption (I/I0 =0.99) at two different X-ray energies (2 keV
and 10 keV). This figure demonstrates that absorption is less than 1% and can safely be
neglected for characteristic X-ray energies of 10 keV or above for a path length less than 50 nm
and for an atomic number of less than 60. For less energetic X-rays (2 keV), a less stringent
limit of 10% absorption (I/I0 =0.9) yields similar acceptable sample constraints (specimen size
less than 50 nm and atomic number less than 60). The values displayed in Fig 12 provide a
guide to understanding the size of specimens acceptable for EDX tomography and which X-ray
energies are likely to provide elemental maps that most accurately satisfy the projection
requirement.
112
Figure 12. Maximum X-ray path length in a sample as a function of atomic number for a single
element sample when considering a limit of 1% absorption (I/I0 =0.99) at two different X-ray
energies (red, 2 keV and blue, 10 keV) and for a less stringent 10% absorption limit (I/I0 =0.9)
for an X-ray energy of 10 keV (green). Mass attenuation coefficients and the densities of
elements used to calculate this data have been taken from NIST [33].
Taking a 1% absorption limit, pure elemental Ag has a maximum allowable path length through
the specimen of 80 nm for 10 keV X-rays and 7 nm for 2 keV X-rays. The same absorption limit
for pure elemental Au gives a maximum path length through the specimen of 44 nm for 10 keV
X-rays and 5 nm for 2 keV X-rays. In this study we focus on bimetallic AgAu nanoparticles with
diameters of approximately 40 nm (Fig 11), although the hollow morphology of the particles
means that for most particles the maximum projected thickness is in practise typically 10 nm or
less. For the AgAu nanoparticle used in this study (Fig 11), the particle is roughly 50 at% of
each element and the maximum X-ray path length is approximately 5 nm. The attenuation
𝜇 𝜇 (6)
= ∑𝑖 𝑤𝑖 ( ) ,
𝜌 𝜌 𝑖
where wi is the weight fraction of the ith atomic constituent [34]. Absorption of 2 keV X-rays
through the nanoparticle of interest (Fig 11) was found to be at maximum 1%, validating the use
113
of X-ray peaks above 2 keV in the reconstruction of this nanoparticle (Au Mα = 2.1 keV).
However, 1 keV X-rays will undergo maximum absorption of approximately 4% within the
nanoparticle which suggests that carbon (0.277 keV) and oxygen (0.525 keV) Kα X-rays will
show significant variations in absorption even in this small sample. The use of a maximum path
length differs from standard absorption correction procedures [35] which integrate the X-ray
path lengths along the electron beam within the sample. The complex geometry of these
particles makes integration challenging and hence we choose to consider the maximum path
length as this will provide an overestimate to X-ray absorption within an image and so a
Another possible violation of the projection requirement comes from the fact that the extent of
detector shadowing varies as a function of sample position, and consequently it is feasible for
shadowing to vary within a single spectrum image. This situation is illustrated in Fig 13 which
shows a 2D cross-section subtending the sample, the shadowing object and the middle of a
single Super-X detector (top right). The difference in angle, Δθ, at which X-rays are shadowed
for two points at the extremes of the spectrum image (separated by a distance Δl), due to the
presence of an object with height H (typically the holder or grid bars) at a distance I from the
𝐻 𝐻 (7)
Δ𝜃 = tan−1 ( ) − tan−1 ( ).
𝑙 𝑙+Δ𝑙
114
Figure 13. Diagram illustrating the variation in detector shadowing between two points on the
sample separated by a distance Δl, shadowed by an object of height H at a distance l from the
image area.
across an EDX spectrum image area can be estimated as a function of α-tilt. We use the
variation in the unshadowed detector area as an estimate of the variation in X-ray counts across
an image area. This is an approximation that does not take in to account the change in X-ray
flux density across the detector but provides an accurate order of magnitude estimate. The
detector area, A, shadowed by an object casting a shadow of height h on the detector, which is
a function of θT, the polar angle due to tilt of the sample, is given by
𝑟−ℎ(𝜃𝑇 ) (8)
𝐴(𝜃𝑇 ) = 𝑟 2 cos −1 ( ) − (𝑟 − ℎ(𝜃𝑇 ))√2𝑟ℎ(𝜃𝑇 ) − ℎ(𝜃𝑇 )2 ,
𝑟
For the Super-X detector system the variation in shadowing for each pair of detectors (1+2 or
3+4 as shown in Fig 1a) must be considered separately but, within a pair, the shadowing can be
assumed to behave identically, due to their equivalent position with respect to the α-tilt axis. The
proportion of the detector area that is partially shadowed across the image, compared to the
detector area that is entirely unshadowed is termed P12 for detector pair 1+2 and P34 for
115
We begin by discussing the shadowing resulting from the top side of the sample holder, for the
situation where both pairs of detectors are partially shadowed simultaneously (Fig 14a). From
simple trigonometry the height, h, of the shadow cast upon the detector is given by
where d is the distance to the centre of the detectors from the sample, θE is the elevation angle
of all detectors (assumed to be the same), θB is the angle to the bottom of the detector pair, θS
is the angle over which the object shadows the detector pair, and δ is the polar tilt angle of the
detector. All angles are defined in the plane of the sample and two Super-X detectors (at a 45°
azimuthal angle to the α-tilt axis) as show in Fig 14a. An α-tilt of the sample can be translated to
In the range of tilt angles in which detector shadowing from the top of the sample holders occurs
116
Figure 14. Diagram illustrating the geometry of angles used in determination of the extent of
detector shadowing from (a) the top side of the sample holder and (b) the bottom side of sample
Equations 10 and 11 provide the variation in shadowing when using only one detector pair,
either detectors 1 and 2 or detectors 3 and 4. The variation in shadowing can reach 100%
across the image area when the detector pair is fully shadowed at one side of the area, i.e. P 12
detector located only on one side of the tilt axis, care should be taken to avoid acquiring tomographic
However, when both detector pairs are used, as is the case in this study, the shadowing effect is largely
mitigated by the greater effective detector area provided by the complementary detector pair on the
other side of the tilt axis. To calculate the contribution from both detectors we have to multiply the
variation in shadowing of a single detector pair Pij by the angular fraction of the total detector area that
117
Thus, the maximum variation in shadowing across the specimen distance Δl when considering
We have determined the corresponding angles approximately from our tilt-series data (Fig 2b)
and from the optical data of the sample holder (Fig 3). Fitting a Gaussian to our experimental
tilt-series data in Fig 2b reveals an average minimum in counts at an α-tilt of approximately 18°,
equivalent to a polar angle of θT = 13°, which we have related to the elevation angle of all four
detectors. We have used a detector tilt of δ = 0-15° and note that this has little effect on our
calculations, with no change in shadowing variation to 1 significant figure. The angles θ U and θB
are determined from the elevation angle, tilt and radius of the detectors. From the optical profile
data (Fig 3) we know that on the top side of the sample holder the height of the shadowing is h
= 600 μm and the distance to the shadowing is l = 1.6 mm, from which we calculate values of θS
= 21° and Δθ = 0.001° when Δl = 100 nm. Note in this and subsequent equations we have also
tacitly assumed that there are no contributions to this shadowing from any collimators, which
are absent in our system. However, for different configurations, should collimators be present,
then their presence will slightly modify the preceding equations which are based purely on the
Using these values, over a distance of 100 nm the percent of shadowing due to the top side of
-3
the holder reaches a maximum of 3x10 %. P12 and P34 will have opposite signs, meaning the
variation in shadowing is in the opposite direction for each detector pairs and acts to somewhat
Considering shadowing from the underside of the sample the shadowing height, h, on the
detector is given by
On the underside of the sample grid, our optical measurements have demonstrated that the
sample only causes shadowing for a polar angle of less than 18°, while the middle of a grid
square to the height of the grid bars in the TEM grids used in this study was 24°. This suggests
that the maximum shadowing angle on the underside of the sample is due to the grid bars (θS =
118
24°), as shown in Fig 4. However, bowing of the carbon film across a grid square could lead to
an increase or decrease in this angle. Assuming the carbon film lays flat, shadowing due to grid
bars can be estimated using values of width and depth obtained from suppliers in each
particular case, or through measurement of grid dimensions. In this case, we have used grids
with a hole of width 90 μm and a grid bar depth of 20 μm. Assuming the sample area is located
in the middle of a grid square, the angular difference between points 100 nm from each other
gives Δθ = 0.05°. For the underside of the sample P12 and P34 are calculated in a similar manner
However, unlike for the top side of the specimen holder both detector pairs are never
simultaneously shadowed (Fig 14b) and therefore the variation in shadowing can be given
simply by
This reaches a maximum of 0.1% and hence can be largely ignored for high spatial resolution
tomographic data sets. Importantly, for larger image areas, images taken close to the grid bars
or for grids with smaller grid areas this variation may be large enough so that projections at a
number of angles should be removed from the tilt series. For instance, if the nanoparticle was
located at a 10 μm distance to the grid bars closest to detectors 1 and 2, as represented by the
red dot in Fig 5b, the shadowing angle due to grid bars would increase to θS = 63°, increasing
Δθ to 0.2° over a 100 nm image area. In this case the variation in shadowing across the image
area would reach approximately 30%. Care should always be taken to ensure that the sample is
situated away from any grid bars when using a standard TEM grid and the use of fine-mesh grid
entirely avoided. In fact, the use of continuous films may be preferable to entirely remove the
contribution of grid bars to shadowing. We further note that where data is acquired from
specimen areas close to the grid bars or from a grid with a smaller mesh size the shadowing
from the top side may also be due to grid bars rather than the sides of the sample holder.
119
5. Conclusions
New designs of large solid angle energy dispersive X-ray detector systems using multiple
detectors have made EDX tomography more readily accessible. However, careful consideration
of sample and acquisition parameters is required to optimize data acquisition schemes and to
ensure that artefacts associated with low signal to noise ratio data do not affect the fidelity of
the effect of detector shadowing for the Titan’s Super-X EDX system, maximizing the available
X-ray signal when using a penumbra-limited single-tilt tomography holder. This compensated
acquisition scheme provides approximately constant X-ray counts in the characteristic peaks
directly proportional to the number of atoms per unit volume of the element in question.
Absorption of X-rays limits the size of samples in which this proportionality holds true and, at
some thickness of sample, absorption will result in violation of the projection requirement,
particularly for high atomic number samples and reconstructions of low energy X-ray signals.
We have considered the potential for shadowing variations within a single spectrum image field
of view and determine that these are insignificant and can be largely ignored for small sample
areas (<100 nm) when using all four Super-X EDX detectors.
Acknowledgements
S.J.H. and T.J.A.S. acknowledge funding from multiple research grants including the
Engineering and Physical Sciences Research Council (EPSRC) UK Grants EP/G035954/1 and
would like to thank the North West Nanoscience Doctoral Training Centre (NOWNano DTC) for
supporting his work. N.J.Z. also acknowledges support from the Electron Microscopy Center at
the Center for Nanoscale Materials of Argonne National Laboratory, a U.S. Department of
Energy, Office of Science, Office of Basic Energy Sciences User Facility under Contract No.
University of Manchester. The authors wish to acknowledge the support from HM Government
120
(UK) for the provision of the funds for the FEI Titan G2 80-200 S/TEM associated with research
References
121
[16] Slater, T.J.A., Camargo, P. H. C., Burke, M. G., Zaluzec, N. J. & Haigh, S. J. Understanding
the limitations of the Super-X energy dispersive x-ray spectrometer as a function of
specimen tilt angle for tomographic data acquisition in the S/TEM. JPCS 522 (2014)
012025.
[17] Slater, T. J. A. et al. Correlating Catalytic Activity of Ag–Au Nanoparticles with 3D
Compositional Variations. Nano Lett. (2014) 1921-1926.
[18] Williams, D. B. & Carter, C. B., Transmission Electron Microscopy, second ed., Springer,
New York, 2009.
[19] Kremer, J. R., Mastronarde, D. N. & McIntosh, J. R. Computer visualization of three-
dimensional image data using IMOD. J. Struct. Biol. 116 (1996) 71-76.
[20] Rasband, W. S., ImageJ, U. S. National Institutes of Health, Bethesda, Maryland, USA,
1997-2014.
[21] Petri, M. V., Ando, R. A. & Camargo, P. H. C. Tailoring the structure, composition, optical
properties and catalytic activity of Ag-Au nanoparticles by the galvanic replacement
reaction. Chem. Phys. Lett. 531 (2012) 188-192.
[22] Zaluzec, N. J. Analytical Formulae for Calculation of X-Ray Detector Solid Angles in the
Scanning and Scanning/Transmission Analytical Electron Microscope. Microsc.
Microanal. 20 (2014) 1318-1326.
[23] Yeoh, C. S. M. et al. The Dark Side of EDX Tomography: Modeling Detector Shadowing to
Aid 3D Elemental Signal Analysis. Microscopy and Microanalysis FirstView (2015) 1-6.
[24] Schlossmacher, P., Klenov, D., Freitag, B. & von Harrach, H. S. Enhanced Detection
Sensitivity with a New Windowless XEDS System for AEM Based on Silicon Drift
Detector Technology. Microscopy Today 18 (2010) 14-20.
[25] Thomas, J. M. et al. The chemical application of high-resolution electron tomography:
Bright field or dark field? Angew. Chem. Int. Edit. 43 (2004) 6745-6747.
[26] Batenburg, K. J. et al. 3D imaging of nanomaterials by discrete tomography.
Ultramicroscopy 109 (2009) 730-740.
[27] Frank, J. & Brandt, S., Electron Tomography, second ed., Springer, New York, 2008.
[28] Kotula, P., Brewer, L., Michael, J. & Giannuzzi, L. Computed Tomographic Spectral
Imaging: 3D STEM-EDS Spectral Imaging. Microsc. Microanal. 13 (2007) 1324-1325.
[29] Yaguchi, T., Konno, M., Kamino, T. & Watanabe, M. Observation of three-dimensional
elemental distributions of a Si device using a 360 degrees-tilt FIB and the cold field-
emission STEM system. Ultramicroscopy 108 (2008) 1603-1615.
[30] Goris, B., Polavarapu, L., Bals, S., Van Tendeloo, G. & Liz-Marzan, L. M. Monitoring
galvanic replacement through three-dimensional morphological and chemical mapping.
Nano Lett. 14 (2014) 3220-3226.
[31] Liakakos, N. et al. Co-Fe Nanodumbbells: Synthesis, Structure, and Magnetic Properties.
Nano Lett. 14 (2014) 2747-2754.
[32] Hirsch, P. B., Whelan, M. J. & Howie, A. On Production Of X-Rays In Thin Metal Foils.
Philos. Mag. 7 (1962) 2095.
122
[33] Seltzer, S.M. and Hubbell, J.H. (1995), Tables and Graphs of Photon Mass Attenuation
Coefficient and Mass Energy-Absorption Coefficients for Photon Energies 1 keV to 20
MeV for Elements Z = 1 to 92 and Some Dosimetric Materials, Appendix to invited
plenary lecture by J.H. Hubbell ``45 Years (1950-1995) with X-Ray Interactions and
Applications'' presented at the 51st National Meeting of the Japanese Society of
Radiological Technology, April 14-16, 1995, Nagoya, Japan.
[34] Seltzer, S. M. Calculation of Photon Mass Energy-Transfer and Mass Energy-Absorption
123
6. STEM-EDX Tomography of AgAu Nanoparticles
6.1. Introduction
This chapter describes the application of STEM-EDX tomography to investigate the elemental
particular, the catalytic yield of propargylamines, a building block in many anti-apoptotic drugs
(Maruyama, et al., 2002a; Maruyama, et al., 2002b; Tatton, et al., 2002), in a three component
coupling reaction is correlated to changes in elemental distribution within the AgAu nanoparticle
catalyst.
The galvanic replacement reaction allows the synthesis of hollow nanoparticles using a
sacrificial template to determine the morphology of the final nanoparticle (Sun & Xia, 2004; Xia,
et al., 2013). A metallic nanoparticle template undergoes a redox reaction with a metal salt, with
the salt reducing at the surface of the nanoparticle and the metal template oxidising. The use of
an Au metal salt, HAuCl4, with an Ag nanoparticle template is widely reported and results in a
predominantly Au hollow shell structure (Au, et al., 2008; Sun & Xia, 2004; Zhang, et al., 2008).
These previous studies claim that all of the AgAu nanostructures formed possess a fully alloyed
structure. The precise distribution of Ag and Au is difficult to determine due to the very similar
Hollow nanoparticles, such as those synthesized through galvanic replacement, are of particular
interest to catalysis due to their increased surface area compared to solid nanoparticles. The
galvanic replacement reaction offers a facile method to create hollow structures through a single
reaction step, but the synthesized nanoparticle has a composition and structure dependent on
the precise reaction conditions (Au, et al., 2008; Sun & Xia, 2003). Here, the effect of varying
the concentration of the Au salt is investigated and is shown to result in large changes in
observed are a prime candidate for investigation using STEM-EDX tomography and the three
dimensional elemental reconstructions obtained provide a clear picture of how the elemental
124
The AgAu nanoparticles used in this study were synthesized by Alexandra Macedo and Pedro
Camargo (University of Sao Paulo), who also performed the catalytic testing of the samples.
XPS analysis of the samples was performed by Sven Schroeder (University of Leeds).
The results of this project were published in Nano Letters in 2014. The original manuscript for
presented below. The supplementary information associated with this manuscript is provided
125
Paper: Correlating Catalytic Activity of Ag-Au Nanoparticles with 3D
Compositional Variations
1 2 1 1,3
Thomas J. A. Slater , Alexandra Macedo , M. Grace Burke , Paul O’Brien , Pedro H. C
2* 1*
Camargo , Sarah J. Haigh
1
School of Materials, University of Manchester, Manchester, M13 9PL, UK
2
Departamento de Química Fundamental, Instituto de Química, Universidade de São Paulo,
3
School of Chemistry, University of Manchester, Manchester, M13 9PL, UK
electron tomography
ABSTRACT: Significant elemental segregation is shown to exist within individual hollow silver-
gold (Ag-Au) bimetallic nanoparticles produced from the galvanic reaction between Ag particles
-
and AuCl4 . Three dimensional compositional mapping using energy dispersive X-ray (EDX)
tomography within the scanning transmission electron microscope (STEM) reveals that
Maximum Au surface coverage was observed for nanoparticles with approximately 25at% Au
important fundamental insights towards optimizing the properties and reducing the cost of
The exceptional catalytic properties of Au nanoparticles as compared to the inert nature of the
1
bulk element are well known . The emergence of catalytic activity as particles’ critical
dimensions approach nanometer scale is in part due to the increase in the particle’s surface to
1
volume ratio . Synthetic routes that maximize this property, especially in distinct geometries
2,3
such as hollow shells or cages , have been extensively investigated. Bimetallic noble metal
4
nanoparticles are known to have excellent catalytic performance in a wide range of reactions .
The extent of alloying in these systems is known to influence catalytic activity and selectivity but
126
5
an understanding of the origin of this effect is lacking . Hollow bimetallic nanoparticles of
accurately tunable size, shape, and composition can be readily prepared by a galvanic
6-8
replacement reaction . The Ag-Au bimetallic system offers the opportunity to both reduce the
9,10
cost of the Au catalysts and enhance catalytic performance . Hollow Ag-Au nanostructures of
various compositions and shapes have been synthesized by galvanic reactions between Au
- 10,11
metal salts, such as AuCl4 , and Ag nanoparticle sacrificial templates . Previous low
resolution TEM imaging and optical spectroscopy observations suggest that this process
initiates with the creation of Ag ions at the surface and the formation of a gold surface layer
which immediately transforms into an Au-Ag alloy because of the fast interdiffusion between
7,12
gold and silver at 100 °C . As the replacement reaction proceeds, dealloying and processes
such as Ostwald ripening induce morphological reconstruction, as well as the formation, and
enlargement, of pinholes in the walls. However, the alloying in individual particles, as well as the
elemental distribution have now been made clear by analytical EDX analysis within the STEM
that has only become possible due to recent advances in EDX detector design.
Ag-Au nanoparticles with controlled compositions and structures (solid vs. hollow interiors) were
spectrometry were Ag93Au7, Ag82Au18, Ag78Au22 and Ag66Au34. The catalytic activity of the Ag-Au
nanoparticles as a function of composition and structure was investigated for the three-
3
component (A ) coupling reaction between cyclohexanecarboxyaldehyde, piperidine, and
the synthesis of heterocyclic compounds and represent structural elements of several natural
products and drugs, which has led to their use as intermediates in the total synthesis of such
13,14
molecules . Recently, it has been shown that this transformation can be catalyzed by
15,16
nanostructured Au via a C-H activation mechanism . The yield of propargylamines as a
function of the Au atomic percentage in the Ag-Au nanoparticles is given in Fig.1b. All the
bimetallic and hollow Ag-Au nanoparticles produce higher yields of propargylamine than their
with the increased surface area in the bimetallic nanoparticles due to the formation of hollow
127
Figure 1. Ag-Au bimetallic nanoparticles improve propargylamine yield. a, Three component
composition of 18at%Au (numerical %yields are noted on graph). c, TEM images of the
compositions plotted in b, (i) Ag93Au7, (ii) Ag82Au18, (iii) Ag78Au22 and (iv) Ag66Au34.
More surprising is an exceptionally high peak in yield of 60% at a catalyst composition of 18 at%
Au (Fig.1b). TEM imaging shows that the sizes and morphologies of the nanoparticles are
broadly similar for all bimetallic compositions (Fig.1c), suggesting that the observed peak in the
catalytic results are not explained by size, shape or surface area differences alone.
Nanoscale compositional variations have been observed to strongly affect the catalytic activity
17,18
of platinum (Pt) alloy nanoparticles . Au and Ag form a miscible solid solution over the full
range of compositions, the lattice parameter of Au is just 0.2% larger than that of Ag. Therefore,
structurally sensitive techniques: X-ray diffraction, electron diffraction and high resolution TEM
been used to elucidate surface composition for bimetallic nanoparticles but cannot provide
information on the local distribution of elements or on the variation of alloying inside the volume
128
of a nanoparticle. High angle annular dark-field (HAADF) scanning transmission electron
microscope (STEM) imaging is able to reveal local compositional variation if there are
sufficiently large differences in the average atomic number of constituent elements. However,
the non-uniform cross section of our particles meant that HAADF STEM imaging of the
Through the combination of STEM imaging with electron energy loss (EEL) or EDX
aberration corrected STEM imaging, high resolution chemical spectrum images for >30 of the
Ag-Au bimetallic nanoparticles of different compositions were obtained (Fig.2). These images
are the first observation of a reversal in the surface segregation within individual nanoparticles
with increased Au content. Nanoparticles with a low Au content (average composition Ag93Au7)
have an approximately 1nm Au-rich surface layer, whereas those from the sample with the
appeared homogeneous in composition but a small fraction (~20%) showed an Au-rich surface
layer. Nanoparticles with intermediate average composition Ag78Au22 were also mostly
homogeneously alloyed except for a small fraction that displayed Ag surface segregation. The
apparent surface segregation behavior for an individual nanoparticle correlates well with its
actual Au composition measured by processing the EDX sum spectra obtained over the entire
field-of-view for each spectrum image (Fig.2e). In order to gain a better understanding of the
surface segregation behavior, line profiles have been extracted perpendicular to the surface of
the elemental maps (Fig.2a-d). Such two dimensional elemental images can be difficult to
interpret because different compositions overlap when projected along the electron beam
direction. For example, Fig.3c shows a line compositional profile extracted from the 2D
elemental maps in Fig.3a,b in which Ag appears to be clearly segregated to the surfaces for the
left-hand side of the nanoparticle ring but not for the right-hand side.
129
Figure 2. STEM EDX mapping showing the inversion of surface segregation behaviour for Ag-
from populations with average stoichiometry (a) Ag93Au7, (b) Ag82Au18, (c) Ag78Au22 and (d)
Ag66Au34. The actual composition of each nanoparticle shown inset was measured using the
summed EDX spectrum for the whole field of view of each spectrum image. The chemistry of 20
nanoparticles with a range of compositions was mapped and the structure classified as to
whether the particle appeared to show a gold-rich surface, a silver rich surface or no surface
segregation (alloyed composition). The average nanoparticle composition was then measured
from the summed spectral image and the results are shown in e. The change from gold to silver
surface segregation is found to correlate with the gold content of the nanoparticle.
130
Figure 3. 3D chemical mapping of an Ag-Au nanoparticle with a high Au content showing Ag
surface segregation. Two dimensional EDX maps acquired at zero degrees of tilt showing the
distribution of a, Au and b, Ag. c, Compositional variation extracted right to left from the 2D
maps at the position of the line shown in a and b. 2D EDX maps extracted from the 3D
tomograms showing the distribution of d, Au and e, Ag. f, Compositional variation extracted right
to left from the 2D slices shown in d and e. The line profile extracted from the tomographic data
slice more clearly reveals the silver surface segregation due to the improved resolution in the z
direction. g, Surface visualization of the Au (red) and Ag (green) tomograms revealing the silver
surface segregation (see supplementary movie S1). The average composition measured for the
whole nanoparticle is given inset. The scale bars in a, b, d and e are 10nm.
method similar to X-ray tomography. Electron tomography using HAADF STEM imaging has
23
proven a useful tool in the analysis of nanoparticle size, shape and distribution . Energy filtered
24 24,25
transmission electron microscopy (EFTEM) and STEM imaging with EDX spectroscopy
can provide chemical sensitivity to electron tomographic reconstructions but both approaches
are challenging. New EDX detector designs (see Supplementary Information) combined with the
increased currents and small probe size obtainable in aberration corrected STEM instruments
mean EDX tomography is now an increasingly viable tool for nanoparticle characterization.
131
In order to better understand the surface segregation behavior of the nanoparticles, EDX
tomograms were compared for a nanoparticle exhibiting Ag-surface segregation with another
reveals a fairly uniform surface layer approximately 2 nm thick (Fig.3d, e). The compositional
line profile from the tomogram (Fig.3f) reveals the benefit of the three dimensional data set for
nanoparticle surface analysis, more clearly demonstrating the segregated surface compared to
the profiles obtained from the two dimensional data (Fig.3c). This is due to the improved
resolution in the z-direction of the three dimensional data, removing the contribution of the top
and bottom surfaces from the line profile data. Processing of the EDX sum spectrum revealed
an average composition of Ag60Au40 for the particle shown in Fig.3, close to the average
composition of the whole sample (Ag66Au34). Fig.3f shows a surface visualization of the Au and
taken from the low Au sample Ag93Au7 shows clear evidence of a thinner (~1nm) non-uniform
Au surface segregation present on both the inner and outer surfaces of the hollow particle
Figure 4. EDX Tomogram of nanoparticle with average composition Ag93Au7 with linescan. a,
Au and b, Ag slices through the respective tomograms. Surface segregation of Au can be seen
in the slice through the Au tomogram. c, Linescan through slices a and b showing the
thin surface of Ag outside of the Au. d, Surface visualization of the Au (red) and Ag (green)
tomograms displaying the surface segregation of Au. The scale bars in a and b are 10nm.
132
The 2D compositional mapping and 3D tomographic data show a clear inversion in the surface
composition for these particles as the Au content is increased. Previous theoretical and lower
spatial resolution experimental studies have suggested that particles are either homogeneously
6,7,26,27
alloyed (due to the lower Ag-Au bond energy compared to either Au-Au or Ag-Ag ) or
28
have Ag segregated to the surface as a result of this element’s lower surface energy .
Our results demonstrate that the Ag and Au distribution in the nanoparticles strongly depends
upon the composition of the particle. At low Au content (<18 at%) galvanic replacement initially
more Ag is replaced by Au, there is an increased driving force to minimize the particle’s
intermetallic bond energy and surface energy which produces alloying and finally, at higher
at% Au, close to the point at which the particles change from Au surface segregation to a
homogeneously alloyed composition (Fig. 2e). This is consistent with the observation that Au
3 29
provides a greater yield than Ag for A coupling reactions . It is feasible that the greater
increasing the electron density of the surface relative to that which would be found in either Ag
Our results provide direct 3D observation of an inversion in the segregation of silver and gold for
the first time. This approach is not confined to galvanic replacement reactions but could be
applied to nanomaterials synthesized via many other chemical and physical processing routes.
There are other examples of bimetallic systems which have shown a peak in catalytic
information a peak in mass specific activity has been attributed solely to a change in the bulk
30
composition of the nanoparticle . Our results demonstrate that segregation in bimetallic
particles is influenced by the synthetic route and is not simple to predict. Furthermore, the
experimental results from this study provide direct experimental evidence that nanoscale
surface segregation can influence nanoparticle properties, providing insights for the improved
133
ASSOCIATED CONTENT
Supporting Information.
coupling-reaction, STEM imaging and tomography. This includes figures demonstrating the
single/poly-crystalline nature of nanoparticles and EDX spectra. Videos of aligned tilt series and
rotating views of isosurface renders are also available. This material is available free of charge
AUTHOR INFORMATION
Corresponding Author
Author Contributions
PHCC and AM synthesized the nanoparticles, performed preliminary TEM imaging and
measured catalytic data. TJAS and SJH performed the STEM imaging and EDX analysis and
wrote the paper. TJAS performed the 3D reconstructions. All authors contributed to the
Notes
ACKNOWLEDGMENTS
TJAS, POB and SJH thank the UK Engineering and Physical Sciences Research Council,
NowNANO doctoral training centre for funding support. SJH thanks the USA Defense Threat
Reduction Agency (grant number HDTRA1-12-1-0013) and Gates Foundation for funding
support. PHCC and AM thank FAPESP and CNPq for funding support (grant numbers
2011/06847-0 and 471245/2012-7, respectively). The authors wish to acknowledge the support
from HM Government (UK) for the provision of the funds for the FEI Titan G2 80-200 S/TEM
associated with research capability of the Nuclear Advanced Manufacturing Research Centre.
ABBREVIATIONS
134
EDX, energy dispersive X-ray; STEM, scanning transmission electron microscope; HAADF,
high-angle annular dark field; EFTEM, energy filtered transmission electron microscopy.
REFERENCES
(1) Herzing, A. A.; Kiely, C. J.; Carley, A. F.; Landon, P.; Hutchings, G. J.,
Identification of active gold nanoclusters on iron oxide supports for CO oxidation. Science 2008,
(2) Kim, S. W.; Kim, M.; Lee, W. Y.; Hyeon, T. Fabrication of hollow palladium
spheres and their successful application to the recyclable heterogeneous catalyst for Suzuki
(3) Lou, X. W.; Archer, L. A.; Yang, Z. C., Hollow Micro-/Nanostructures: Synthesis
(4) Sankar, M.; Dimitratos, N.; Miedziak, P. J.; Wells, P. P.; Kiely, C. J.; Hutchings,
G. J., Designing bimetallic catalysts for a green and sustainable future. Chem. Soc. Rev. 2012,
41 (24), 8099-8139.
(5) Toshima, N.; Yonezawa, T., Bimetallic nanoparticles - novel materials for
(6) Petri, M. V.; Ando, R. A.; Camargo, P. H. C., Tailoring the structure,
composition, optical properties and catalytic activity of Ag-Au nanoparticles by the galvanic
(7) Sun, Y. G.; Xia, Y. N., Mechanistic study on the replacement reaction between
silver nanostructures and chloroauric acid in aqueous medium. J. Am. Chem. Soc 2004, 126
(12), 3892-3901.
(8) Skrabalak, S. E.; Chen, J.; Sun, Y.; Lu, X.; Au, L.; Cobley, C. M.; Xia, Y., Gold
Nanocages: Synthesis, Properties, and Applications. Acc. Chem. Res. 2008, 41 (12), 1587-
1595.
(9) Kim, Y.; Hong, J. W.; Lee, Y. W.; Kim, M.; Kim, D.; Yun, W. S.; Han, S. W.,
(10) Bracey, C. L.; Ellis, P. R.; Hutchings, G. J., Application of copper-gold alloys in
catalysis: current status and future perspectives. Chem. Soc. Rev. 2009, 38 (8), 2231-2243.
135
(11) Cobley, C. M.; Xia, Y., Engineering the properties of metal nanostructures via
galvanic replacement reactions. Mater. Sci. Eng. R-Rep. 2010, 70 (3-6), 44-62.
(12) Sun, Y. G.; Xia, Y. N., Alloying and dealloying processes involved in the
preparation of metal nanoshells through a galvanic replacement reaction. Nano Lett. 2003, 3
(11), 1569-1572.
(13) Miura, M.; Enna, M.; Okuro, K.; Nomura, M., Copper-catalyzed reaction of
(14) Jenmalm, A.; Berts, W.; Li, Y. L.; Luthman, K.; Csoregh, I.; Hacksell, U.,
Stereoselective epoxidation of phe-gly and phe-phe vinyl isosteres. J. Org. Chem. 1994, 59 (5),
1139-1148.
(15) Gonzalez-Bejar, M.; Peters, K.; Hallett-Tapley, G. L.; Grenier, M.; Scaiano, J. C.,
Rapid one-pot propargylamine synthesis by plasmon mediated catalysis with gold nanoparticles
(16) Kidwai, M.; Bansal, V.; Kumar, A.; Mozumdar, S., The first Au-nanoparticles
(17) Cui, C.; Gan, L.; Heggen, M.; Rudi, S.; Strasser, P., Compositional segregation
in shaped Pt alloy nanoparticles and their structural behaviour during electrocatalysis. Nature
(18) Stamenkovic, V. R.; Mun, B. S.; Arenz, M.; Mayrhofer, K. J. J.; Lucas, C. A.;
Wang, G.; Ross, P. N.; Markovic, N. M., Trends in electrocatalysis on extended and nanoscale
(19) Chen, D. H.; Chen, C. J., Formation and characterization of Au-Ag bimetallic
(20) Wang, A. Q.; Chang, C. M.; Mou, C. Y., Evolution of catalytic activity of Au-Ag
bimetallic nanoparticles on mesoporous support for CO oxidation. J. Phys. Chem. B 2005, 109
(40), 18860-18867.
(21) Muller, D. A., Structure and bonding at the atomic scale by scanning
136
(22) D'Alfonso, A. J.; Freitag, B.; Klenov, D.; Allen, L. J., Atomic-resolution chemical
mapping using energy-dispersive x-ray spectroscopy. Phys. Rev. B 2010, 81 (10), 100101.
(24) Mobus, G.; Doole, R. C.; Inkson, B. J., Spectroscopic electron tomography.
(25) Genc, A.; Kovarik, L.; Gu, M.; Cheng, H.; Plachinda, P.; Pullan, L.; Freitag, B.;
Wang, C., XEDS STEM tomography for 3D chemical characterization of nanoscale particles.
(26) Deng, L.; Hu, W.; Deng, H.; Xiao, S.; Tang, J., Au-Ag Bimetallic Nanoparticles:
Surface Segregation and Atomic-Scale Structure. J Phys. Chem. C 2011, 115 (23), 11355-
11363.
(27) Wei, L. Y.; Qi, W. H.; Huang, B. Y.; Wang, M. P., Surface segregation of Au-Ag
(28) Gong, H. R., Electronic structures and related properties of Ag-Au bulks and
(29) Huang, B.; Yao, X.; Li, C.-J., Diastereoselective synthesis of alpha-oxyamines
alkynes, and amines in water. Adv. Synth. Catal. 2006, 348 (12-13), 1528-1532.
(30) Zhang, H.; Jin, M. S.; Wang, J. G.; Li, W. Y.; Camargo, P. H. C.; Kim, M. J.;
Yang, D. R.; Xie, Z. X.; Xia, Y. A., Synthesis of Pd-Pt Bimetallic Nanocrystals with a Concave
Structure through a Bromide-Induced Galvanic Replacement Reaction. J. Am. Chem. Soc. 2011,
137
Supporting Information - Correlating catalytic activity of Ag-
Methods
For the synthesis of the Ag nanoparticle templates, PVP (10 g, MW 10,000 g/mol) was
dissolved in 75 mL of ethylene glycol at room temperature, and 400 mg AgNO3 was added to
this solution. The resulting suspension was stirred at room temperature until the complete
dissolution of AgNO3. This mixture was heated up to 100 °C at a constant rate of 1 °Cmin-1
and the reaction was then allowed to proceed for 1.5 h. After, 175 mL of distilled water was
added and the reaction mixture was allowed to cool to room temperature. The suspension
containing the Ag nanoparticles was washed three times with water by successive
centrifugation at 7000 rpm and removal of the supernatant. After washing, the resulting Ag
nanoparticles were suspended in 250 mL of water. This suspension was then used for the
The synthesis of bimetallic and hollow AgAu nanoparticles used the galvanic reaction
(1 mg/mL, MW 55,000 g/mol) and 27.8 mL of the Ag nanoparticle suspension were added
into a 1000 mL round bottom flask. This system was heated up to 100 °C for 10 min. Then,
138
-
0.2 mM AuCl4 (aq) solution was added dropwise and the reaction allowed to proceed at
100 °C for another 10 min. 100, 200, 300, or 400 mL aliquots of a 0.2 mM AuCl4-(aq) solution
were used to prepare AgAu nanoparticles with average compositions of Ag93Au7 (Ag 93at%
Au 7at%), Ag82Au18 (Ag 82at% Au 18at%), Ag78Au22 (Ag 78at% Au 22at%), and
Ag66Au34 (Ag 66at% Au 18at%), respectively. The Au and Ag atomic percentages were
model AA-6300, equipped with an air-acetylene flame. After cooling down to room
temperature, each suspension of AgAu nanoparticles was washed three times with water by
successive rounds of centrifugation at 7000 rpm and removal of the supernatant. After
washing, the AgAu nanoparticles were suspended in 10 mL of water for the catalytic tests.
The catalytic activities of AgAu nanoparticles towards the three-component coupling reaction
3
(A ) were screened using cyclohexanecarboxyaldehyde, piperidine, and phenylacetylene as
round bottom flask and sonicated under N2 for 5 min. Then, cyclohexanecarboxyaldehyde,
(168 mg, 1.5 mmol), piperidine (85 mg, 1.0 mmol) and phenylacetylene (153 mg, 1.5 mmol)
were added to the suspension containing the AgAu nanoparticles and stirred under a nitrogen
atmosphere. The Au mol% in the reaction corresponded to 1.6, 1.5, 2.0, and 2.5 for Ag93Au7,
mixture was heated up to 90 oC and the reaction was allowed to proceed for 12 h under
vigorous stirring. After that, the reaction mixture was extracted with dichloromethane (3 x 20
mL), leaving the AgAu nanoparticles in the aqueous phase. The organic layer was dried over
MgSO4·XH2O and the solvent was removed under reduced pressure. The crude product was
purified by column chromatography using silica gel as the stationary phase and an
The catalytic activities of Ag and Au monometallic nanoparticles were also investigated for
comparison. In this case, a similar procedure was performed. For Ag nanoparticles, 25.4 mL
of the suspension after washing was centrifuged and suspended in 10 mL of distilled water
139
containing Au nanoparticles were centrifuged and suspended in 10 mL of distilled water (Au
dissolving 100 mg of HAuCl4·3H2O in 950 mL of distilled water. This solution was heated to
100 °C. A solution of sodium citrate (50 mL, 0.5 wt%) was then added and the reaction was
allowed to proceed for 60 min. The resulting suspension containing the Au nanoparticles was
cooled to room temperature and stored until use. The product, 1-(1-cyclohexyl-3-phenylprop-
1 13
2-yn-1-yl) piperidine, was identified by H and C NMR spectroscopy and GC-MS
1
spectrometry: H NMR (200 MHz, CDCl3): δ 7.54 – 7.19 (m, 5H), 3.10 (d, J = 9.9 Hz, 1H),
2.75 – 2.26 (m, 4H), 2.08 (t, J = 7.5 Hz, 1H), 1.85 – 0.76 (m, 16H) ppm; 13C NMR (50 MHz,
CDCl3): 131.83, 128.29, 127.72, 123.92, 87.88, 86.25, 64.50, 50.86, 39.70, 31.46, 30.55,
-83
26.93, 26.41, 26.23, 24.84 ppm. MS: m/z (%) 198 (M , 100), 153 (3), 128 (9), 115 (65), 103
(4), 91 (7), 77 (5), 69 (4), 55 (11), 41 (38). Analytical thin layer chromatography (TLC) was
carried out by using silica gel 60 F254 pre-coated plates. Visualization was accomplished
with vanillin (vanillin 0.01 g/mL in AcOH/H2SO4 (99:1) solution) or ninhydrin as color reagent
(ninhydrin 0.005 g/mL in EtOH (96%) solution). The gas chromatography analyses were
1
carrier gas and equipped with a DB-5 column (30m x 0.25mm x 0.25mm). The H NMR (200
MHz) and 13C NMR (50 MHz) spectra were recorded on Bruker AC 200 spectrometer using
tetramethylsilane (TMS) as the internal standard. Chemical shifts were reported in parts per
million (ppm, δ) downfield from the TMS. Low-resolution mass spectra were recorded on
Shimadzu GC-17A coupled with QP5050A MS, using HP-5MS column (30m x 0.25mm x
0.25mm).
Nanoparticles were prepared for (S)TEM imaging by placing a few drops of a suspension
onto a holey carbon coated copper support grid (200 or 300 mesh) and allowing to air dry.
TEM imaging was performed using a JEOL 1010 transmission electron microscope operated
at 80 kV. STEM and EDX spectrum imaging were performed using a probe-side aberration-
corrected FEI Titan G2 80-200 S/TEM operated at 200 kV. STEM images were collected
using a convergence angle of 18 or 26 mrad and a high angle annular dark-field (HAADF)
140
detector with an inner angle of 55 mrad. EDX compositional analysis was performed using
the Super-X detector configuration (4 x 30 mm2 SDDs) with a solid angle of ~0.7 srad and a
beam current of ~1.2 nA. The two dimensional elemental spectrum images shown in Figure 2
were acquired with a total dose of 1.0 μC (5.9x10 electrons), an acquisition time of ≈10
minutes and a dwell time of 15μs per pixel. The compositional analysis for each nanoparticle
was obtained from an EDX sum spectrum (obtained by a summation of all spectra contained
in the whole field-of-view for a spectrum image containing only a single nanoparticle) and
performing quantification for the gold/silver ratio with background fitting and standard Cliff-
Lorimer “k-factors”. For display purposes all spectrum images were processed using a 3-pixel
Tomographic reconstruction.
spectrum images) were simultaneously acquired using the above imaging conditions. Data
was acquired at 5° intervals using a Fischione 2020 single tilt tomography holder and a
specimen tilt range of ±70°. Acquisition times at each angle were varied to compensate for
the tilt induced shadowing of the detector with the acquisition time as a function of tilt angle
given in Fig. S3. Bruker Esprit software was used to acquire and process the EDX spectrum
images.
Compositional maps were extracted and processed using a 3-pixel smoothing window in
Esprit and exported as separate RGB files for the Au and Ag Lα maps. Maps were then
converted to 16-bit unsigned integer files for visualization using Gatan DigitalMicrograph™
and ImageJ software. Maps were formatted and aligned in FEI Inspect3D software with
alignment being performed via cross- correlation calculated from the simultaneously acquired
HAADF images. Tomographic reconstruction was performed using the Inspect3D SIRT
algorithm using 5 iterations. Line profiles of both 2D maps and slices through the 3D data
cubes were processed in ImageJ. Further visualization of the tomographic data was
141
HAADF STEM and EDX
Atomic resolution imaging of the nanoparticles using HAADF STEM has revealed that the
particles have a complex hollow morphology for the whole range of compositions studied in
this work. Nanoparticles were found to have the expected face centered cubic structure with
both single crystal (Fig. S1a,b) and polycrystalline structures found for all compositions. The
crystal or polycrystalline. Spectra obtained from the four particles shown in Fig. 2 are
average compositions (a) Ag93Au7 , (b) Ag82Au18 , (c) Ag78Au22, and (d) Ag66Au34 .
Fourier Transforms of the images shown inset illustrate that the nanoparticles are largely
trivial. (e)-(h) the corresponding HAADF images (a)-(d) with overlay shading illustrating the
different crystallographic regions and presence of both single crystal and polycrystalline
nanoparticles. The spots in the inset Fourier transforms illustrate which spots
142
Supporting Figure S2. Summed EDX spectra extracted from the EDX spectral data cubes
composition through the nanoparticles is illustrated through the change in the relative
heights of the Au and Ag peaks for each nanoparticle. The Cu peak is due to the copper
grid on which the particles are deposited. The Fe peak is due to the pole-piece of the
mapping of O matches the HAADF image rather than being localized at the surface. The
lack of a Cl peak in the spectra of nanoparticles confirms no AgCl present in the formed
nanoparticles.
We have used a scanning transmission electron microscope fitted with a high brightness
source, aberration corrected probe optics and symmetrical arrangement of large angle EDX
close to 1 nm. However, even for this microscope a Fischione 2020 high tilt tomography
holder suffers from significant detector shadowing due to the sides of the holder blocking the
path of X-rays emitted from the sample to the detector. Fig. S3a shows that shadowing
changes the detected sample counts by a factor of three when using standard TEM grids with
143
Supporting Figure S3. Time-varied acquisition scheme for recording tomographic STEM
EDX data. a, X-ray sample counts as a function of specimen tilt angle for the same
particle using a uniform acquisition time of 300 seconds for each tilt angle. b, Acquisition
time as a function of tilt angle necessary to correct for measured variation in detector
counts shown in (a). c, X-ray sample counts as a function of specimen tilt angle obtained
when the calculated non-uniform acquisition times (given in b) are applied to tomographic
data acquisition.
To overcome this limitation we have employed a novel acquisition scheme under which the
time to acquire each EDX map is varied in order to compensate for detector shadowing at
each angle. The time-varied approach can be justified considering the equation describing
the intensity of the EDX signal for element A, IA, (assuming no absorption or fluorescence
S2,S3
takes place within the sample) :
𝐼𝐴 = 𝑁𝐴 𝑄𝐴 𝜔𝐴 𝑎𝐴 𝑡𝐷𝑒 (Ω/4π)𝜀𝐴
fluorescence yield, aA is the relative transition probability of the peak in question, t is the
thickness of the sample, De is the electron dose, Ω/4π is the proportion of the whole 4π solid
angle that the detector subtends and εA is the detector efficiency. Shadowing of element A
can be included as a factor in detector efficiency εA. To reconstruct the number density of
element A (NA) accurately then IA should not vary as a function of any other parameter as a
function of specimen tilt angle. This requirement is clearly violated through variation in
detector shadowing as a function of tilt angle, i.e. εA varies with specimen tilt angle. To
counteract this variation one of the other parameters in equation 1 could be altered to render
the product of all parameters other than NA constant over all tilt angles. The only parameter
144
that can be altered independently from the sample to normalize the equation is De, the
electron dose, which can be varied straightforwardly by altering the acquisition times of each
spectral image. This approach assumes that the entire nanoparticle is shadowed equally at
each tilt angle; a reasonable approximation considering the small size of the nanoparticles
relative to the detector. Using the calibrated acquisition times shown in Fig.S3b signal counts
for Au and Ag X-ray lines of a single nanoparticle are observed to remain constant (±10%).
An example of the HAADF STEM images, Au and Ag elemental maps for one of the
tomographic data sets obtained is shown in Fig.S4. HAADF-STEM images of the AgAu
particle displayed in Fig.3 acquired before and after tilt-series acquisition are shown in Fig.S4
Supporting Figure S4. HAADF-STEM images of AgAu nanoparticles. a, before and b, after
tilt series acquisition showing little change in structure of the nanoparticle. c, atomic
resolution HAADF-STEM image (with inset FFT) acquired after tilt-series acquisition showing
XPS analysis was performed for (i) the pure Ag precursor nanoparticles and for
Ag93Au7, (iii) Ag82Au18, (iv) Ag78Au22 and (v) Ag66Au34, as measured by Flame
Absorbance Spectroscopy (FAS). XP spectra were recorded with a Kratos Axis Ultra
145
nanoparticle suspension into aluminium lids for differential scanning calorimetry
Experiments were performed while operating the X-ray source with a power of 180 W (15
kV and 12 mA), with a pressure of 10−8 mbar during analysis. The instrument was
operated in CAE (constant analysis energy) mode, with a pass energy of 20 eV for high
spectra were measured within the spectral range of interest with 0.1 eV steps and 600 ms
dwell time per data point. Repeats were carried out to check for radiation damage. Survey
spectra for elemental analysis were obtained using an analyser pass energy of 80 eV, with
Analysis of the data was carried out with Casa XPS software.S5 A Shirley background was
S5,S6
used in all curve-fitting along with a GL(80) lineshape (80% Gaussian, 20% Lorentzian
S5
using the Gaussian/Lorentzian product form). Samples were referenced to the
positions was within ±0.1 eV. The results of the analysis of the Ag and Au components in
the spectra are summarised in Supplementary Table S1. The high resolution Au 4f and Ag 3d
146
Elemental
analysis Surface
Bulk BE FWHM Area BE FWHM Area
Surface Au 4f7/2 Ag 3d5/2
Ag 0 0 - -(eV) -(eV) - (eV) 0.85
367.9 (eV) 1368
(at%
Ag93Au 7 (FAS)
7 Au) (XPS)
19 enrichment
2.7 83.21 0.85 (a.u.)
213 367.9 0.80 (a.u.)
1007
Ag82Au 1 18 35 1.9 83.44 0.78 626 367.9 0.80 1739
(Surface/Bulk
Ag78Au 2 22 48 2.2 83.69 0.85 4025 368.0 0.85 4998
8 66Au 3 34
Ag 48 1.4
) 83.66 0.81 740 368.1 0.90 1017
2
4
Supporting Table S1. Comparison of the elemental analysis by FAS (overall bulk
Au in the surface region of the particles is evident from the ratio between the Au
concentrations detected by XPS and FAS. The results of the chemical shift analysis via curve
fitting analysis of the high resolution Au 4f and Ag 3d photoemission signals are summarised
in the two main column on the right side of the table. For each photoemission line the core
level binding energy (BE), full width at half maximum (FWHM) and area of the Gauss-Lorentz
XPS data confirmed the expected presence of Au and Ag. Table S1 compares the
quantification for the Au/Ag ratio obtained via XPS to that measured using FAS. XPS is highly
surface sensitive, with the great majority of the signal coming from within a few nm of the
surface (the inelastic mean free paths for Au 4f and Ag 3d electrons are approximately
S9
1.6 nm and 1.5 nm, respectively). These results show a greater Au content than was
measured for the bulk, demonstrating that a significant amount of deposited Au remains at
or at least near the surface for mean compositions up to 34 at% Au. This Au enrichment is
most pronounced for the lowest Au composition (7 at%) and smallest for the highest Au
composition (34 at%). We did not obtain data for mean compositions above 34 at% Au,
our STEM EDX measurements, occurs for individual particles above ~40 at% Au (Fig. S2e),
but which due to the spread in composition is not likely to be observed for the
147
Supporting Figure S5. Au 4f (left) and Ag 3d (right) photoemission spectra of the
AgxAu100-x nanoparticle preparations and best fit models obtained using 80%
Gaussian/20% Lorentzian lineshapes and a Shirley background. The spectra are stacked
according to Au content, from 0 at% (top) to 34 at% (bottom). The FAS-determined overall
composition of the nanoparticles is indicated within the figure for each spectrum.
The core level binding energies (BEs) associated with the Au 4f7/2 and Ag 3d5/2 emission
lines of the nanoparticles indicate strongly that Au and Ag are present in the metallic state,
oxidation state of either element. This agrees with STEM-EDX results in which no
enhanced oxygen signal was observed for the surfaces of the nanoparticles. The
S4
observed Au 4f7/2 BEs are significantly lower than the bulk Au metal value (83.96 eV ),
which is in good agreement with previously reported values for Au, Ag and their alloys in
S10-S12
nanoparticulate form. The Au 4f7/2 BEs are lowest for the samples with the
smallest Au content, indicating that the observed catalytic activity is associated with a
significant electronic perturbation of the Au component. All observed Au 4f7/2 BEs as well as
the overall smaller variations in the Ag 3d5/2 BEs are in line with previously reported
S11,S13,S14 Overall,
chemical shifts associated with Au/Ag alloying and de-alloying.
expected variations in core level BEs due to alloying are of the same magnitude as surface
core level shiftsS15 and the shifts associated with nanoparticle size and morphology.S10
148
Supporting References
S1 von Harrach, H. S. et al. An integrated Silicon Drift Detector System for FEI Schottky
Field Emission Transmission Electron Microscopes. Microscopy and Microanalysis 15, 208-
209 (2009).
review of progress from the Cliff-Lorimer to the new zeta-factor methods. Journal of
electron spectrometers 5 - Re-evaluation of the reference energies. Surf. Interface Anal. 1998,
26, 642-649.
S5 Fairley, N.; Carrick, A. The CASA Cookbook: Recipes for XPS Data Processing, pt.
1: Auger and X-ray Photoelectron Spectroscopy, 2 ed.; Briggs, D., Seah, M. P., Eds.; Wiley:
Chichester, 1990.
S7 The XPS of Polymers Database; 1 ed.; Surface Spectra Ltd: Manchester, UK, 2000.
S9 Tanuma, S.; Powell, C. J.; Penn, D. R. Calculations of Electron Inelastic Mean Free
Paths II.Data for 27 Elements over the 50-2000 eV range. Surf. Interface Anal. 1991, 17,
911-926.
S10 Radnik, J.; Mohr, C.; Claus, P. On the origin of binding energy shifts of core levels of
supported gold nanoparticles and dependence of pretreatment and material synthesis. Phys.
149
S11 Bzowski, A.; Kuhn, M.; Sham, T. K.; Rodriguez, J. A.; Hrbek, J. Electronic structure
of Au-Ag bimetallics: Surface alloying on Ru(001). Phys. Rev. B 1999, 59, 13379-13393.
S12 Mason, M. G. Electronic Structure of Supported Small Metal Clusters. Phys. Rev. B
S13 Tyson, C. C.; Bzowski, A.; Kristof, P.; Kuhn, M.; Sammy-Naiken, R.; Sham, T. K.
Charge Redistribution in Au-Ag Alloys from a Local Perspective. Phys. Rev. B 1992, 45,
8924-8928.
S14 Watson, R. E.; Hudis, J.; Perlman, M. L. Charge Flow and d Compensation in Gold
S15 Egelhoff, J. Core-Level Binding-Energy Shifts at Surfaces and in Solids. Surf. Sci.
150
7. Summary and Future Work
that absorption and fluorescence have a negligible effect. This thesis demonstrates that when
X-ray absorption is not negligible, X-ray absorption correction can be applied for well-defined
spherical nanoparticles. In the future, this type of absorption correction could be easily
taking only inputs on the parameters defining the elements present and detector geometry.
tomogram to attempt to quantify composition at each voxel. Although there has been recent
progress in the area of STEM-EDX tomography, quantitative results have not yet been
demonstrated. Currently, EDX tomography can be used to provide qualitative data visualising
demonstrated in chapter 6 and (Genc, et al., 2013; Goris, et al., 2014). Accurate quantification
of STEM-EDX tomograms will require higher counts per pixel in spectrum image projections
than those presented here and are also complicated by artefacts introduced by the absorption
of X-rays. Absorption of X-rays within the sample will cause variations in elemental maps away
from the true elemental distribution, although the use of four detectors equally spaced around
the optic axis compensates for this somewhat if none of the detectors are shadowed. STEM-
correction in many samples. In the future, STEM-EDX tomography has the potential to calculate
151
the absorption of X-rays accurately through a sample, so that a more accurate quantification
can be achieved in two or three dimensions. Iteratively calculating the absorption of X-rays
through a STEM-EDX reconstruction and applying this to the projections acquired should
eventually converge to a solution of the true composition, neglecting the effect of X-ray
fluorescence.
X-ray absorption in the sample holder rather than the sample results in shadowing of the X-ray
detectors, which this thesis has demonstrated can be an important effect in the accurate
addition to the modelling of detector shadowing shown in previous literature (Yeoh, et al., 2015).
Here, the variations in intensity caused by variations in detector shadowing are compensated
for by varying the acquisition time of each projection, so that X-ray count rate is maintained
(Chapters 5 and 6). This thesis also demonstrated that variations in shadowing across an image
may be large if using detectors on only one side of the tilt axis, but when using detectors on
either side of the tilt axis can be largely ignored. Novel positioning of detectors, such as an
annular detector above the sample holder (Zaluzec, 2009), or one parallel to the tilt-axis have
Ultimately, beam damage of the sample may still limit the X-ray counts at each pixel in STEM-
EDX tomography as the cross-section for X-ray production is much lower than the cross-
reconstruction algorithms will allow good fidelity reconstructions from few projections (Goris, et
al., 2013; Leary, et al., 2013), which will allow a larger per image dose if the total dose is limited
due to beam damage of the sample. A larger dose per image will be particularly beneficial for
EDX tomography, as STEM-EDX spectrum imaging suffers from much poorer signal to noise
than imaging via EELS or HAADF-STEM. Another way to limit damage may be to use cryo-
holders that keep the sample at a temperature below -170 °C (Baker & Rubinstein, 2010; Unger,
2001), although these holders typically have high sides that shadow EDX detectors to a large
degree.
The optimal conditions for STEM-EDX tomography are highly dependent on the sample
152
images acquired. In all studies it is important to maximise the counts at each pixel (single pixel
spectra are displayed in Appendix A) to improve the signal to noise ratio. For this reason,
maximising pixel size without degrading resolution is highly recommended for each spectrum
factors discussed in section 2.4, so that pixel size can be set close to the beam diameter. The
maximum dose per each tilt angle should then be used, which is dependent on the number of
projections acquired.
The number of projections necessary is entirely dependent on the reconstruction algorithm used
but here it was found that larger than 5° tilt increments restricts the fidelity of reconstructions
using SIRT. The SIRT algorithm appears to enhance the inherent noise present in STEM-EDX
data when using successive iterations and choice of the number of iterations should be made
EDX spectrum imaging and tomography has been applied here to the study of AgAu
segregation of Au and Ag is observed with increasing Au content and the composition at the
point at which Au surface segregation ceases appears to correlate with that at which catalytic
nanoparticle and future work could apply the approach developed here to other bimetallic
systems containing heavy elements such as Au and Pt. It is also feasible in the future that
whether this is in-situ heating (Gontard, et al., 2014), mechanical testing (Dehm, et al., 2012; Yu,
et al., 2015) or the observation of reactions in liquids and gases (de Jonge & Ross, 2011;
Sharma, 2005; Wang, et al., 2015). In particular, the development of in-situ heating (Lewis, et
al., 2014a) and liquid cell (Lewis, et al., 2014b) holders with low-profile designs specifically for
153
7.1. Correlative Tomography
The previous chapters have detailed the application of EDX tomography to freestanding
nanoparticles, in which the length scale of interest is 1-100 nm, ideally suited to investigation via
the STEM-EDX technique. However, the extension of electron tomography to probe nanoscale
features in bulk material samples raises the questions of whether the volume analysed in
electron tomography is representative of the bulk, or whether the volume chosen is located at
Many material studies have used a range of three dimensional imaging techniques in order to
investigate structural features that span multiple length scales from the atomic to the
macroscopic (Li, et al., 2015; Lopez-Haro, et al., 2013; Tariq, et al., 2011). However, only
recently has it been possible to apply a range of techniques on the same sample volume at
multiple length scales, which can then be spatially correlated to ensure the high resolution
techniques are acquired at a precisely known location (Burnett, et al., 2014a; Burnett, et al.,
Over the past six months I have undertaken initial studies on the correlation of STEM-EDX
tomography data with that of X-ray nanotomography. The sample I have studied is that of a
cavitated grain boundary in a AISI-type 316H stainless steel extracted from within a weld region
of an AGR power station steam header (Burnett, et al., 2014a). Nanoparticles precipitated at the
grain boundary are characterised with STEM-EDX tomography and are located precisely at a
boundary was identified using X-ray microtomography data undertaken by Tim Burnett and
Robert Bradley (University of Manchester) (Burnett, et al., 2014a). The X-ray microtomography
2
was performed on a matchstick sample of length of 0.74 mm and cross-section of 0.36 mm .
The presence of a cavitated grain boundary was confirmed through extraction of a sub-surface
cross-section at the specified site. Subsequently, a ~120 μm high pillar of 30 μm diameter was
isolated from the matchstick sample through annular milling in a plasma-FIB (Fig 7.1a),
undertaken by Tim Burnett. The plasma-FIB provides a rapid method of machining volumes at
the scales of tens of micrometres, providing milling rates up to fifty times greater than
154
conventional Ga+ ion beam milling (Delobbe, et al., 2014). The pillar was attached to a pin
Figure 7.1. a) SEM image within the plasma-FIB dual beam instrument showing the pillar
sample after annular milling. b) The 30 μm diameter pillar attached to a pin for further analysis.
c) An SEM image of the cross-sectional slab cut from the nano-CT pillar before liftout. d) An ion-
beam image of one polished face of the slab attached to a TEM grid. e) Ga ion-beam image of
the “block” cut from the cross-sectional slab and attached to the lift-out needle. f) SEM image of
X-ray nanotomography reveals the presence of three grain boundaries within the pillar, two
decorated with cavities and one visible only through small intensity differences due to large
intergranular carbides (Fig 7.2). The X-ray nanotomography was performed by Robert Bradley
(University of Manchester). The three grain boundaries meet at a triple junction in the middle of
the pillar, which can be visualised in three dimensions through manual tracking of intensity
variations due to the intergranular carbides (Fig 7.2). The location and morphology of individual
cavities along the cavitated grain boundary are fully determinable through visualisation of the X-
ray CT data. The large carbides that are visible as slight changes in intensity along the grain
boundary are also present along the cavitated grain boundary but their morphology cannot be
155
Figure 7.2. X-ray nanotomography of a pillar of stainless steel 316. Volume visualisation
displaying an isosurface rendering of the outer surface of the pillar, traces along the three grain
boundaries and the surface of the segmented cavities. Also displayed is a slice through the X-
ray nanotomography reconstruction displaying the three grain boundaries and the intensity
Visualisation of the X-ray CT data allowed identification of a preferred orientation for extraction
of a cross-sectional “slab” from which to remove needle shaped samples for high resolution 3D
STEM tomography. This orientation was chosen as favourable because in this geometry the full
range of X-ray CT slices were seen to contain all three grain boundaries relaxing the
requirement for precise vertical positioning of the slab. A slab approximately 5μm thick was
lifted out from the pillar (Fig 7.1c) and attached to a “pronged” half-moon TEM grid (Fig 7.1d).
The sides of the slab were polished to allow the acquisition of SEM images, which could be
manually correlated to slices through the X-ray CT data. The removal of the slab sample was
A block with length 22 μm and a square 2 μm x 2 μm cross section was removed from the slab
(Fig 7.1e) using the FIB-SEM and attached to the top of an on-axis tomography pin. Each of the
four faces of the block were polished to allow SEM images to be taken of the grain boundary at
156
each face (Fig 7.5a), in order to identify the location of cavities and grain boundaries within the
block. This was needed in order to determine the optimal location for subsequent extraction of a
suitable STEM tomography needle sample of length greater than 1 μm and diameter 200 – 300
nm using annular milling in the FIB-SEM. Extraction of the block and the annular milling steps
different nanoscale secondary phases. Previous lower resolution imaging of this material
(Burnett, et al., 2014a) has identified the presence of secondary phases of ferrite, M 23C6
carbides and G-phase decorating similar cavitated boundaries within a different portion of the
same austenitic steel sample. STEM-EDX elemental mapping of the needle shaped sample
prepared here reveals the same secondary phases, sandwiched between two austenitic
domains (Fig 7.3). The prepared needle fully encapsulates a small (~ 50 nm diameter) carbide
precipitate.
Figure 7.3. STEM-EDX elemental maps of the STEM needle displaying the intensity of X-ray
counts associated with the a) Fe Kα, b) Cr Kα, c) Ni Kα, d) Mo Lα, e) Nb Kα, f) O Kα, g) C Kα
and h) Si Kα peaks. The austenite phase at top and bottom is associated with Fe, Ni and to a
lesser extent Mo and Mn, the ferrite phase is predominantly Fe, the M23C6 carbides contain Cr,
Mn, Mo, C and O, the G-phase precipitate contains Ni, Mo and Si. The surface of the needle
has high carbon and silicon intensities that are associated with organic contamination.
The reconstruction of STEM-EDX tomography involves the reconstruction of the X-ray intensity
associated with individual elements, such that there is a separate reconstruction of the intensity
distribution of each element. Here, we have reconstructed the distribution of Fe, Cr and Ni in
three dimensions (Fig 7.4 a-c), as these elements allow segmentation of the identified phases.
157
The regions in which there is a high Fe intensity with an associated Ni signal is associated with
the austenite phase and that of a high Fe intensity without an associated Ni intensity is
associated with ferrite. Similarly, a high intensity in the Cr reconstruction corresponds to the
carbide precipitates and a high intensity in the Ni reconstruction with no associated Fe intensity
is associated to the G-phase. Through segmentation of the elemental data using these
constraints the individual phases are segmented in three dimensions (Fig 7.4 d-f).
Figure 7.4. Visualisation of the three dimensional reconstruction of elemental distributions and
the associated segmented phases. Volume visualisations of a) the Fe Kα signal (red), b) the Cr
Kα signal (yellow) and c) the Ni Kα signal (blue). Surface visualisations of the segmented
phases showing the d) ferrite phases (green), e) carbide precipitates (purple) and f) austenite
Spatial correlation of the EDX tomography and X-ray nanotomography volumes is not possible
from the two volumes alone, as features present in the EDX tomography volume are not visible
in the X-ray CT volume due to the limited spatial resolution of X-ray CT. Instead, volume
registration is achieved through correlation of the position of SEM images acquired through the
158
FIB preparation steps with the X-ray CT volume. Registration of the X-ray CT volume and SEM
images is achieved through a manual feature matching, taking advantage of the cavities
An SEM image of one face of the extracted cross-sectional slab is initially matched to a slice
through the X-ray CT volume to determine the approximate position of the block. SEM images
of each face of the block are then manually positioned to match slices in the X-ray CT data.
Cavities present on each of the four faces display a distinctive morphology that allows
correlation to slices in the X-ray CT data. In this way, the location of the extracted block can be
determined to an accuracy of less than 100 nm in all dimensions, positioning the SEM
coordinates into that of the X-ray CT coordinate system. The viewing directions of the four faces
which we number from (i) to (iv), are thus also defined with respect to the X-ray CT coordinates.
159
Figure 7.5. Images displaying the faces of the extracted block and the correlation between a)
SEM images and b) slices through the CT data of faces i-iv. c) Faces (i) and (ii) displayed in
three dimensions from the CT data with the X-ray CT coordinate system indicated.
Correlation of the position of the STEM needle within the X-ray CT volume relies primarily on
the correlation of the block faces. Knowledge that the needle was fabricated in one of the
160
corners of the block, and that it does not intersect any cavities, positions the needle in an area
2
in the defined X and Y dimensions of less than 1 μm in the material coordinate system.
Correlation of the XY position of the needle with that of a top-down view SEM image further
increases the positional accuracy to within a few hundred nanometres (Fig 7.6b). The position
of the needle in the Z direction of the CT volume, that of the long axis of the needle, is harder to
determine and is achieved here through positioning the centre of the precipitates at a Z-
coordinate in the materials coordinate system that corresponds to the location of precipitates in
the correlated SEM image of face (iii) (Fig 7.6a). This assumes that the precipitates are located
at the same Z-coordinates at the XY position of the needle as they are found at the XY
Figure 7.6. a) Visualisation of the STEM-EDX pillar with respect to the SEM image of face (iii)
(flipped) and b-c) with respect to the top-down SEM image taken after annular milling with (c)
and without (b) segmented cavities. The coordinate systems indicated all correspond to the X-
ray CT coordinates.
correlation of an SEM image of the needle to the STEM-EDX 3D data. An SEM image of the
needle with a viewing direction in the same orientation as that of block face (i) is compared to
the final reconstruction. The SEM image of the needle shows bright regions associated with the
higher Z of the Nb containing G-phase precipitates. The viewing direction of block face (i)
specifies a direction in the material coordinate system at which the orientation of the segmented
161
G-phase can be matched to that in the SEM image of the needle, this defines the rotational
Figure 7.7. Correlation of the orientation of the STEM-EDX needle embedded in to the micro-CT
data (a) with that from the SEM image of the final pillar (b) at the same orientation in the SEM
as in Fig 5a(i).
The steps outlined above represent the first steps towards an automated workflow for spatially
correlated studies combining STEM tomography with larger scale techniques such as X-ray
such as atom probe tomography and FIB-SEM slice and view in the future. Correlative
tomography also has great promise for use in a number of industrially relevant material systems.
For instance, the location of individual nanoparticles that display a particular fluorescence
microscopy. Three dimensional imaging through X-ray nanotomography could reveal the
morphology of the site of the nanoparticle in the solar cell and further pinpoint the location of the
nanoparticle for extraction of a needle shaped sample via FIB-SEM. The 3D morphology and
chemistry of this identified single nanoparticle could then be obtained through STEM
tomography or APT.
Correlative tomography also has the potential to analyse in-situ processes at multiple length
scales in the future. In the instance of cracking of an engineering material due to mechanical
stress or a corrosive environment, it is already possible to investigate this type of process in-situ
162
at the macroscale via X-ray CT. If the in-situ X-ray CT is stopped at an important point in the
process of failure of the material, a correlative tomography approach will allow extraction of the
crack-tip for further micro- or nanoscale analysis via FIB-SEM or STEM tomography. Correlative
tomography offers an exciting opportunity to understand the links between nano- and
163
8. References
ALLEN, L.J., FINDLAY, S.D., OXLEY, M.P. & ROSSOUW, C.J. (2003). Lattice-Resolution Contrast
From A Focused Coherent Electron Probe. Part I. Ultramicroscopy 96(1), 47-63.
ANDERHALT, R. & SANDBORG, A. (2011). Windowless Silicon Drift Detectors. Microscopy And
Microanalysis 17(S2), 620-621.
ARMSTRONG, J.T. & BUSECK, P.R. (1975). Quantitative Chemical Analysis Of Individual
Microparticles Using The Electron Microprobe. Theoretical. Analytical Chemistry 47(13), 2178-
2192.
ARSLAN, I., TONG, J.R. & MIDGLEY, P.A. (2006). Reducing The Missing Wedge: High-Resolution
Dual Axis Tomography Of Inorganic Materials. Ultramicroscopy 106(11–12), 994-1000.
ARSLAN, I., MARQUIS, E.A., HOMER, M., HEKMATY, M.A. & BARTELT, N.C. (2008).
Towards Better 3-D Reconstructions By Combining Electron Tomography And Atom-Probe
Tomography. Ultramicroscopy 108(12), 1579-1585.
AU, L., LU, X. & XIA, Y. (2008). A Comparative Study Of Galvanic Replacement Reactions
Involving Ag Nanocubes And Aucl2- Or Aucl4. Advanced Materials 20(13), 2517.
BABOUT, L., MAIRE, E., BUFFIERE, J.Y. & FOUGERES, R. (2001). Characterization By X-Ray
Computed Tomography Of Decohesion, Porosity Growth And Coalescence In Model Metal
Matrix Composites. Acta Materialia 49(11), 2055-2063.
BAER, D.R., GASPAR, D.J., NACHIMUTHU, P., TECHANE, S.D. & CASTNER, D.G. (2010).
Application Of Surface Chemical Analysis Tools For Characterization Of Nanoparticles.
Analytical And Bioanalytical Chemistry 396(3), 983-1002.
BAKER, L.A. & RUBINSTEIN, J.L. (2010). Radiation Damage In Electron Cryomicroscopy. Methods
In Enzymology, Vol 481: Cryo-Em, Part A - Sample Preparation And Data Collection 481, 371-
388.
BALDI, A., NARAYAN, T.C., KOH, A.L. & DIONNE, J.A. (2014). In Situ Detection Of Hydrogen-
Induced Phase Transitions In Individual Palladium Nanocrystals. Nature Materials 13(12), 1143-
1148.
164
BALS, S., CASAVOLA, M., VAN HUIS, M.A., VAN AERT, S., BATENBURG, K.J., VAN TENDELOO, G. &
VANMAEKELBERGH, D. (2011). Three-Dimensional Atomic Imaging Of Colloidal Core-Shell
Nanocrystals. Nano Letters 11(8), 3420-3424.
BARNARD, J.S., SHARP, J., TONG, J.R. & MIDGLEY, P.A. (2006). High-Resolution Three-
Dimensional Imaging Of Dislocations. Science 313(5785), 319.
BATENBURG, K.J., BALS, S., SIJBERS, J., KUBEL, C., MIDGLEY, P.A., HERNANDEZ, J.C., KAISER, U.,
ENCINA, E.R., CORONADO, E.A. & VAN TENDELOO, G. (2009). 3D Imaging Of Nanomaterials By
Discrete Tomography. Ultramicroscopy 109(6), 730-740.
BATSON, P.E. (1993). Simultaneous STEM Imaging And Electron Energy-Loss Spectroscopy
With Atomic-Column Sensitivity. 366(6457), 727-728.
BELIC, D., CHANTRY, R.L., LI, Z.Y. & BROWN, S.A. (2011). Ag-Au Nanoclusters: Structure And
Phase Segregation. Applied Physics Letters 99(17), 171914.
BELL, D.C.. & ERDMAN, N. (2013). Low Voltage Electron Microscopy: Principles And
Applications. Chichester, UK: Wiley.
BETHE, H. (1930). Zur Theorie Des Durchgangs Schneller Korpuskularstrahlen Durch Materie.
Annalen Der Physik 397(3), 325-400.
BLAVETTE, D., BOSTEL, A., SARRAU, J.M., DECONIHOUT, B. & MENAND, A. (1993). An
Atom-Probe For 3-Dimensional Tomography. Nature 363(6428), 432-435.
BLELOCH, A. & RAMASSE, Q. (2011). Lens Aberrations: Diagnosis And Correction. In Aberration-
Corrected Analytical Transmission Electron Microscopy, Pp. 55-87. John Wiley & Sons, Ltd.
BOOTZ, A., VOGEL, V., SCHUBERT, D. & KREUTER, J. (2004). Comparison Of Scanning
Electron Microscopy, Dynamic Light Scattering And Analytical Ultracentrifugation For The Sizing
Of Poly(Butyl Cyanoacrylate) Nanoparticles. European Journal Of Pharmaceutics And
Biopharmaceutics 57(2), 369-375.
BRAIDY, N., JAKUBEK, Z.J., SIMARD, B. & BOTTON, G.A. (2008). Quantitative Energy Dispersive X-
Ray Microanalysis Of Electron Beam-Sensitive Alloyed Nanoparticles. Microscopy And
Microanalysis 14(2), 166-175.
BRANDT, S.S. & ZIESE, U. (2006). Automatic TEM Image Alignment By Trifocal Geometry.
Journal Of Microscopy-Oxford 222, 1-14.
BRYDSON, R. & HONDOW, N. (2011). Electron Energy Loss Spectrometry And Energy
Dispersive X-Ray Analysis. In Aberration-Corrected Analytical Transmission Electron
Microscopy, Brydson, R. (Ed.), Pp. 163-210. John Wiley & Sons, Ltd.
165
BUFFIERE, J.Y., MAIRE, E., ADRIEN, J., MASSE, J.P. & BOLLER, E. (2010). In Situ
Experiments With X Ray Tomography: An Attractive Tool For Experimental Mechanics.
Experimental Mechanics 50(3), 289-305.
BURNETT, T.L., GEURTS, R., JAZAERI, H., NORTHOVER, S.M., MCdONALD, S.A., HAIGH, S.J.,
BOUCHARD, P.J. & W ITHERS, P.J. (2014a). Multiscale 3D Analysis Of Creep Cavities In AISI Type
BURNETT, T.L., MCdONALD, S.A., GHOLINIA, A., GEURTS, R., JANUS, M., SLATER, T., HAIGH, S.J.,
ORNEK, C., ALMUAILI, F., ENGELBERG, D.L., THOMPSON, G.E. & WITHERS, P.J. (2014b). Correlative
Tomography. Scientific Reports 4, 4711.
CAO, M., ZHANG, H.B., LU, Y., NISHI, R. & TAKAOKA, A. (2010). Formation And Reduction Of
Streak Artefacts In Electron Tomography. Journal Of Microscopy 239(1), 66-71.
CASSIDY, C., SINGH, V., GRAMMATIKOPOULOS, P., DJURABEKOVA, F., NORDLUND, K. & SOWWAN, M.
(2013). Inoculation Of Silicon Nanoparticles With Silver Atoms. Scientific Reports 3, 7.
CHANTRY, R.L., SIRIWATCHARAPIBOON, W., HORSWELL, S.L., LOGSDAIL, A.J., JOHNSTON, R.L. & LI,
Z.Y. (2012). Overgrowth Of Rhodium On Gold Nanorods. Journal Of Physical Chemistry C
116(18), 10312-10317.
CHEN, C.C., ZHU, C., WHITE, E.R., CHIU, C.Y., SCOTT, M.C., REGAN, B.C., MARKS, L.D., HUANG, Y.
& MIAO, J.W. (2013). Three-Dimensional Imaging Of Dislocations In A Nanoparticle At Atomic
Resolution. Nature 496(7443), 74.
CHEN, Y.Q., SLATER, T.J.A., LEWIS, E.A., FRANCIS, E.M., BURKE, M.G., PREUSS, M. & HAIGH, S.J.
(2014b). Measurement Of Size-Dependent Composition Variations For Gamma Prime (Γ′)
Precipitates In An Advanced Nickel-Based Superalloy. Ultramicroscopy 144, 1-8.
CHO, S.J., IDROBO, J.C., OLAMIT, J., LIU, K., BROWNING, N.D. & KAUZLARICH, S.M. (2005). Growth
Mechanisms And Oxidation Resistance Of Gold-Coated Iron Nanoparticles. Chemistry Of
Materials 17(12), 3181-3186.
CLOETENS, P., LUDWIG, W., BARUCHEL, J., VAN DYCK, D., VAN LANDUYT, J., GUIGAY,
J.P. & SCHLENKER, M. (1999). Holotomography: Quantitative Phase Tomography With
166
Micrometer Resolution Using Hard Synchrotron Radiation X Rays. Applied Physics Letters
75(19), 2912-2914.
COLLIEX, C., CRAVEN, A.J. & WILSON, C.J. (1977). Fresnel Fringes In Stem. Ultramicroscopy 2(4),
327-335.
CORTIE, M.B. & MCDONAGH, A.M. (2011). Synthesis And Optical Properties Of Hybrid And
Alloy Plasmonic Nanoparticles. Chemical Reviews 111(6), 3713-3735.
CROWTHER, R.A., DEROSIER, D.J. & KLUG, A. (1970). Reconstruction Of 3 Dimensional Structure
From Projections And Its Application To Electron Microscopy. Proceedings Of The Royal
Society Of London Series A-Mathematical And Physical Sciences 317(1530), 319.
CUI, C., GAN, L., LI, H.-H., YU, S.-H., HEGGEN, M. & STRASSER, P. (2012). Octahedral Ptni
Nanoparticle Catalysts: Exceptional Oxygen Reduction Activity By Tuning The Alloy Particle
Surface Composition. Nano Letters 12(11), 5885-5889.
CUI, C., GAN, L., HEGGEN, M., RUDI, S. & STRASSER, P. (2013). Compositional Segregation In
Shaped Pt Alloy Nanoparticles And Their Structural Behaviour During Electrocatalysis. Nature
Materials 12(8), 765-771.
D'ALFONSO, A.J., FREITAG, B., KLENOV, D. & ALLEN, L.J. (2010). Atomic-Resolution Chemical
Mapping Using Energy-Dispersive X-Ray Spectroscopy. Physical Review B 81(10), 100101.
DAHMEN, T., BAUDOIN, J.-P., LUPINI, A.R., KUEBEL, C., SLUSALLEK, P. & DE JONGE, N. (2014).
Combined Scanning Transmission Electron Microscopy Tilt- And Focal Series. Microscopy And
Microanalysis 20(2), 548-560.
DE JONGE, N. & ROSS, F.M. (2011). Electron Microscopy Of Specimens In Liquid. Nature
Nanotechnology 6(11), 695-704.
DEANS, S.R. (2007). The Radon Transform And Some Of Its Applications. Mineola, NY: Dover
Publications Inc.
167
DEFRISE, M., VIERGEVER, M., TODD-POKROPEK, A. (1988). Possible Criteria for Choosing
Computer Science in Medical Imaging 39, Pp. 293-303. Springer Berlin Heidelberg.
DEHM, G., LEGROS, M. & KIENER, D. (2012). In-Situ TEM Straining Experiments: Recent
Progress In Stages And Small-Scale Mechanics. In In-Situ Electron Microscopy, Pp. 227-254.
Wiley-VCH Verlag GMBH.
DOROFEEV, G.A., STRELETSKII, A.N., POVSTUGAR, I.V., PROTASOV, A.V. & ELSUKOV,
E.P. (2012). Determination Of Nanoparticle Sizes By X-Ray Diffraction. Colloid Journal 74(6),
675-685.
DROUIN, D., COUTURE, A.R., JOLY, D., TASTET, X., AIMEZ, V. & GAUVIN, R. (2007). CASINO V2.42
- A Fast And Easy-To-Use Modeling Tool For Scanning Electron Microscopy And Microanalysis
Users. Scanning 29(3), 92-101.
DUNN, D.N. & HULL, R. (1999). Reconstruction Of Three-Dimensional Chemistry And Geometry
Using Focused Ion Beam Microscopy. Applied Physics Letters 75(21), 3414-3416.
EGERTON, R.F. (2009). Electron Energy-Loss Spectroscopy In The TEM. Reports On Progress
In Physics 72(1), 016502.
EGERTON, R.F. (2013). Control Of Radiation Damage In The TEM. Ultramicroscopy 127, 100-
108.
EGERTON, R.F., LI, P. & MALAC, M. (2004). Radiation Damage In The TEM And SEM. Micron
35(6), 399-409.
EGERTON, R.F., MClEOD, R., WANG, F. & MALAC, M. (2010). Basic Questions Related To
Electron-Induced Sputtering In The TEM. Ultramicroscopy 110(8), 991-997.
EGERTON, R.F., WANG, F. & CROZIER, P.A. (2006). Beam-Induced Damage To Thin Specimens
In An Intense Electron Probe. Microscopy And Microanalysis 12(01), 65-71.
168
EGGERT, T., BOSLAU, O., GOLDSTRASS, P. & KEMMER, J. (2004). Silicon Drift Detectors With
Enlarged Sensitive Areas. X-Ray Spectrometry 33(4), 246-252.
FERRER, D., BLOM, D.A., ALLARD, L.F., MEJIA, S., PEREZ-TIJERINA, E. & JOSE-YACAMAN, M. (2008).
Atomic Structure Of Three-Layer Au/Pd Nanoparticles Revealed By Aberration-Corrected
Scanning Transmission Electron Microscopy. Journal Of Materials Chemistry 18(21), 2442-2446.
FERRANDO, R., JELLINEK, J. & JOHNSTON, R.L. (2008). Nanoalloys: From Theory To
Applications Of Alloy Clusters And Nanoparticles. Chemical Reviews 108(3), 845-910.
FERTIG, J. & ROSE, H. (1981). Resolution And Contrast Of Crystalline Objects In High-Resolution
Scanning-Transmission Electron-Microscopy. Optik 59(5), 407-429.
FINDLAY, S.D., ALLEN, L.J., OXLEY, M.P. & ROSSOUW, C.J. (2003). Lattice-Resolution Contrast
From A Focused Coherent Electron Probe. Part II. Ultramicroscopy 96(1), 65-81.
FIORINI, C., LONGONI, A., PEROTTI, F., LABANTI, C., ROSSI, E., LECHNER, P., SOLTAU, H. &
STRUDER, L. (2002). A Monolithic Array Of Silicon Drift Detectors For High-Resolution Gamma-
Ray Imaging. IEEE Transactions On Nuclear Science 49(3), 995-1000.
FLADISCHER, S. (2013). Application of new EDXS quantification methods in TEM for organic
semiconductor devices. PhD thesis, University of Graz.
FLOREA, I., ERSEN, O., ARENAL, R., IHIAWAKRIM, D., MESSAOUDI, C., CHIZARI, K., JANOWSKA, I. &
CUONG, P.-H. (2012). 3D Analysis Of The Morphology And Spatial Distribution Of Nitrogen In
Nitrogen-Doped Carbon Nanotubes By Energy-Filtered Transmission Electron Microscopy
Tomography. Journal Of The American Chemical Society 134(23), 9672-9680.
FRANK, J., MCeWEN, B.F., RADERMACHER, M., TURNER, J.N. & RIEDER, C.L. (1987). 3-
Dimensional Tomographic Reconstruction In High-Voltage Electron-Microscopy. Journal Of
Electron Microscopy Technique 6(2), 193-205.
FRANK, J. & HAWKES, P. (2006). The Electron Microscope As A Structure Projector. In Electron
Tomography, Pp. 83-111. Springer New York.
FRANK, J. & MASTRONARDE, D. (2006). Fiducial Marker And Hybrid Alignment Methods For
Single- And Double-Axis Tomography. In Electron Tomography, Pp. 163-185. Springer New
York.
169
FRIEDRICH, H., DE JONGH, P.E., VERKLEIJ, A.J. & DE JONG, K.P. (2009). Electron Tomography For
Heterogeneous Catalysts And Related Nanostructured Materials. Chemical Reviews 109(5),
1613-1629.
FRIGO, S.P., LEVINE, Z.H. & ZALUZEC, N.J. (2002). Submicron Imaging Of Buried Integrated
Circuit Structures Using Scanning Confocal Electron Microscopy. Applied Physics Letters
81(11), 2112-2114.
GAN, L., HEGGEN, M., O'MALLEY, R., THEOBALD, B. & STRASSER, P. (2013). Understanding And
Controlling Nanoporosity Formation For Improving The Stability Of Bimetallic Fuel Cell Catalysts.
Nano Letters 13(3), 1131-1138.
GASS, M.H., KOZIOL, K.K.K., WINDLE, A.H. & MIDGLEY, P.A. (2006). Four-Dimensional Spectral
Tomography Of Carbonaceous Nanocomposites. Nano Letters 6(3), 376-379.
GENC, A., KOVARIK, L., GU, M., CHENG, H., PLACHINDA, P., PULLAN, L., FREITAG, B. & WANG, C.
(2013). XEDS STEM Tomography For 3D Chemical Characterization Of Nanoscale Particles.
Ultramicroscopy 131, 24-32.
GEMMI, M. & OLEYNIKOV, P. (2013). Scanning Reciprocal Space For Solving Unknown
Structures: Energy Filtered Diffraction Tomography And Rotation Diffraction Tomography
Methods. Zeitschrift Fur Kristallographie 228(1), 51-58.
GIANNUZZI, L.A., DROWN, J.L., BROWN, S.R., IRWIN, R.B. & STEVIE, F. (1998).
Applications Of The FIB Lift-Out Technique For TEM Specimen Preparation. Microscopy
Research And Technique 41(4), 285-290.
GILBERT, P.F.C. (1972b). Reconstruction Of A 3-Dimensional Structure From Projections And Its
Application To Electron-Microscopy .2. Direct Methods. Proceedings Of The Royal Society Of
London Series B-Biological Sciences 182(1066), 89.
GOLDSTEIN, J.I., COSTLEY, J.L., LORIMER, G.W. & REED, R.J.B. (1977). Quantitative X-Ray
Analysis In The Electron Microscope. In Scanning Electron Microscopy, Johari, O. (Ed.), Pp.
315. Chicago, IL: IITRI.
GOLDSTEIN, J., NEWBURY, D.E., JOY, D.C., LYMAN, C.E., ECHLIN, P., LIFSHIN, E.,
SAWYER, L. & MICHAEL, J.R. (2003). Scanning Electron Microscopy And X-Ray Microanalysis.
New York: Springer US.
GOMEZ-GRANA, S., GORIS, B., ALTANTZIS, T., FERNANDEZ-LOPEZ, C., CARBO-ARGIBAY, E.,
GUERRERO-MARTINEZ, A., ALMORA-BARRIOS, N., LOPEZ, N., PASTORIZA-SANTOS, I., PEREZ-JUSTE,
J., BALS, S., VAN TENDELOO, G. & LIZ-MARZAN, L.M. (2013). Au@Ag Nanoparticles: Halides
Stabilize {100} Facets. Journal Of Physical Chemistry Letters 4(13), 2209-2216.
170
GONTARD, L.C., DUNIN-BORKOWSKI, R.E., FERNANDEZ, A., OZKAYA, D. & KASAMA, T. (2014).
Tomographic Heating Holder For In Situ TEM: Study Of Pt/C And Ptpd/Al2O3 Catalysts As A
Function Of Temperature. Microscopy And Microanalysis 20(3), 982-990.
GORDON, R., BENDER, R. & HERMAN, G.T. (1970). Algebraic Reconstruction Techniques (ART)
For 3-Dimensional Electron Microscopy And X-Ray Photography. Journal Of Theoretical Biology
29(3), 471.
GORIS, B., BALS, S., VAN DEN BROEK, W., VERBEECK, J. & VAN TENDELOO, G. (2011). Exploring
Different Inelastic Projection Mechanisms For Electron Tomography. Ultramicroscopy 111(8),
1262-1267.
GORIS, B., DE BACKER, A., VAN AERT, S., GOMEZ-GRANA, S., LIZ-MARZAN, L.M., VAN TENDELOO, G.
& BALS, S. (2013a). Three-Dimensional Elemental Mapping At The Atomic Scale In Bimetallic
Nanocrystals. Nano Letters 13(9), 4236-4241.
GORIS, B., POLAVARAPU, L., BALS, S., VAN TENDELOO, G. & LIZ-MARZAN, L.M. (2014a). Monitoring
Galvanic Replacement Through Three-Dimensional Morphological And Chemical Mapping.
Nano Letters 14(6), 3220-3226.
GORIS, B., ROELANDTS, T., BATENBURG, K.J., HEIDARI MEZERJI, H. & BALS, S. (2013b). Advanced
Reconstruction Algorithms For Electron Tomography: From Comparison To Combination.
Ultramicroscopy 127, 40-47.
GORIS, B., TURNER, S., BALS, S. & VAN TENDELOO, G. (2014b). Three-Dimensional Valency
Mapping In Ceria Nanocrystals. Acs Nano 8(10), 10878-10884.
GROBELNY, J., DELRIO, F.W., PRADEEP, N., KIM, D.-I., HACKLEY, V.A. & COOK, R.F.
(2011). Size Measurement Of Nanoparticles Using Atomic Force Microscopy. Characterization
Of Nanoparticles Intended For Drug Delivery 697, 71-82.
HABERFEHLNER, G., BAYLE-GUILLEMAUD, P., AUDOIT, G., LAFOND, D., MOREL, P.H., JOUSSEAUME,
V., ERNST, T. & BLEUET, P. (2012). Four-Dimensional Spectral Low-Loss Energy-Filtered
Transmission Electron Tomography Of Silicon Nanowire-Based Capacitors. Applied Physics
Letters 101(6), 063108.
HABERFEHLNER, G., ORTHACKER, A., ALBU, M., LI, J. & KOTHLEITNER, G. (2014). Nanoscale Voxel
Spectroscopy By Simultaneous EELS And EDS Tomography. Nanoscale 6(23), 14563-14569.
HAIDER, M., ROSE, H., UHLEMANN, S., SCHWAN, E., KABIUS, B. & URBAN, K. (1998). A Spherical-
Aberration-Corrected 200 Kv Transmission Electron Microscope. Ultramicroscopy 75(1), 53-60.
HAISS, W., THANH, N.T.K., AVEYARD, J. & FERNIG, D.G. (2007). Determination Of Size And
Concentration Of Gold Nanoparticles From UV-Vis Spectra. Analytical Chemistry 79(11), 4215-
4221.
171
HEIDARI MEZERJI, H., VAN DEN BROEK, W. & BALS, S. (2013). Quantitative Electron Tomography:
The Effect Of The Three-Dimensional Point Spread Function. Ultramicroscopy 135, 1-5.
HEIDARI MEZERJI, H., VAN DEN BROEK, W. & BALS, S. (2011). A Practical Method To Determine
The Effective Resolution In Incoherent Experimental Electron Tomography. Ultramicroscopy
111(5), 330-336.
HERBIG, M., CHOI, P. & RAABE, D. (2015). Combining Structural And Chemical Information At
The Nanometer Scale By Correlative Transmission Electron Microscopy And Atom Probe
Tomography. Ultramicroscopy 153, 32-39.
HIGGINS, J.P., ARRIVO, S.M., THURAU, G., GREEN, R.L., BOWEN, W., LANGE, A.,
TEMPLETON, A.C., THOMAS, D.L. & REED, R.A. (2003). Spectroscopic Approach For On-Line
Monitoring Of Particle Size During The Processing Of Pharmaceutical Nanoparticles. Analytical
Chemistry 75(8), 1777-1785.
HILLHOUSE, H.W. & BEARD, M.C. (2009). Solar Cells From Colloidal Nanocrystals:
Fundamentals, Materials, Devices, And Economics. Current Opinion In Colloid & Interface
Science 14(4), 245-259.
HIRSCH, P., HOWIE, A., NICHOLSON, R., PASHLEY, D. & WHELAN, M. (1977). Electron Microscopy
Of Thin Crystals. Huntington, New York: Krieger.
HUTTER, E. & FENDLER, J.H. (2004). Exploitation Of Localized Surface Plasmon Resonance.
Advanced Materials 16(19), 1685-1706.
HWANG, B.J., SARMA, L.S., CHEN, J.M., CHEN, C.H., SHIH, S.C., WANG, G.R., LIU, D.G.,
LEE, J.F. & TANG, M.T. (2005). Structural Models And Atomic Distribution Of Bimetallic
Nanoparticles As Investigated By X-Ray Absorption Spectroscopy. Journal Of The American
Chemical Society 127(31), 11140-11145.
HSIEH, J. (2009). Computed Tomography: Principles, Design, Artifacts, And Recent Advances.
Wiley.
JAIN,P. K., HUANG, X., EL-SAYED, I.H., & EL-SAYED, M.A. (2008). Noble Metals on the
Nanoscale: Optical and Photothermal Properties and Some Applications in Imaging, Sensing,
Biology, and Medicine. Accounts of Chemical Research 41(12), 1578-1586.
172
JARAUSCH, K., THOMAS, P., LEONARD, D.N., TWESTEN, R. & BOOTH, C.R. (2009). Four-
Dimensional STEM-EELS: Enabling Nano-Scale Chemical Tomography. Ultramicroscopy
109(4), 326-337.
JESSON, D.E. & PENNYCOOK, S.J. (1995). Incoherent Imaging Of Crystals Using Thermally
Scattered Electrons. Proceedings Of The Royal Society-Mathematical And Physical Sciences
449(1936), 273-293.
JIANG, H.L., AKITA, T., ISHIDA, T., HARUTA, M. & XU, Q. (2011). Synergistic Catalysis Of
Au@Ag Core-Shell Nanoparticles Stabilized On Metal-Organic Framework. Journal Of The
American Chemical Society 133(5), 1304-1306.
JONES, S.D. & HAERTING, M. (2013). A New Correlation Based Alignment Technique For Use In
Electron Tomography. Ultramicroscopy 135, 56-63.
JOSHI, A. & STEIN, D.F. (1972). Chemistry Of Grain-Boundaries And Its Relation To
Intergranular Corrosion Of Austenitic Stainless-Steel. Corrosion 28(9), 321.
KANO, S., TADA, T. & MAJIMA, Y. (2015). Nanoparticle Characterization Based On STM And
STS. Chemical Society Reviews 44(4), 970-987.
KATO, M., KAWASE, N., KANEKO, T., TOH, S., MATSUMURA, S. & JINNAI, H. (2008). Maximum
Diameter Of The Rod-Shaped Specimen For Transmission Electron Microtomography Without
The "Missing Wedge". Ultramicroscopy 108(3), 221-229.
KELLY, T.F. & MILLER, M.K. (2007). Invited Review Article: Atom Probe Tomography. Review
Of Scientific Instruments 78(3), 031101.
KIM, J., CHOI, H., NAHM, C. & PARK, B. (2012). Surface-Plasmon Resonance For
Photoluminescence And Solar-Cell Applications. Electronic Materials Letters 8(4), 351-364.
KIMOTO, K., ASAKA, T., NAGAI, T., SAITO, M., MATSUI, Y. & ISHIZUKA, K. (2007). Element-Selective
Imaging Of Atomic Columns In A Crystal Using STEM And EELS. Nature 450(7170), 702-704.
KIRKLAND, E.J. (2010). Advanced Computing In Electron Microscopy, Second Ed. Advanced
Computing In Electron Microscopy, Second Ed.
KNOLL, M. & RUSKA, E. (1932). Das Elektronenmikroskop. Zeitschrift Für Physik 78(5-6), 318-
339.
KOGUCHI, M., KAKIBAYASHI, H., TSUNETA, R., YAMAOKA, M., NIINO, T., TANAKA, N., KASE, K. &
IWAKI, M. (2001). Three-Dimensional STEM For Observing Nanostructures. Journal Of Electron
Microscopy 50(3), 235-241.
KOSTER, A.J., GRIMM, R., TYPKE, D., HEGERL, R., STOSCHEK, A., WALZ, J. & BAUMEISTER, W.
(1997). Perspectives Of Molecular And Cellular Electron Tomography. J. Struct. Biol. 120(3),
276-308.
173
KOTHLEITNER, G., NEISH, M.J., LUGG, N.R., FINDLAY, S.D., GROGGER, W., HOFER, F. & ALLEN, L.J.
(2014). Quantitative Elemental Mapping At Atomic Resolution Using X-Ray Spectroscopy.
Physical Review Letters 112(8).
KOTULA, P., BREWER, L., MICHAEL, J. & GIANNUZZI, L. (2007). Computed Tomographic Spectral
Imaging: 3D STEM-EDS Spectral Imaging. Microscopy And Microanalysis 13(S2), 1324-1325.
KOTULA, P., MICHAEL, J. & ROHDE, M. (2008). Results From Two Four-Channel Si-Drift Detectors
On An SEM: Conventional And Annular Geometries. Microscopy And Microanalysis 14(S2),
116-117.
KRAKOW, W., JOSEYACAMAN, M. & ARAGON, J.L. (1994). Observation Of Quasi-Melting At The
Atomic-Level In Au Nanoclusters. Physical Review B 49(15), 10591-10596.
KRAMERS, H.A. (1923). On The Theory Of X-Ray Absorption And Of The Continuous X- Ray
Spectrum. Philosophical Magazine 46(275), 836-871.
KRAUSE, M.O. (1979). Atomic Radiative And Radiationless Yields For K And L Shells. Journal Of
Physical And Chemical Reference Data 8(2), 307-327.
KRIVANEK, O.L., DELLBY, N. & LUPINI, A.R. (1999). Towards Sub-Angstrom Electron Beams.
Ultramicroscopy 78(1-4), 1-11.
KRIVANEK, O.L., LOVEJOY, T.C., DELLBY, N., AOKI, T., CARPENTER, R.W., REZ, P., SOIGNARD, E.,
ZHU, J., BATSON, P.E., LAGOS, M.J., EGERTON, R.F. & CROZIER, P.A. (2014). Vibrational
Spectroscopy In The Electron Microscope. Nature 514(7521), 209.
KRIVANEK, O.L., LOVEJOY, T.C., DELLBY, N. & CARPENTER, R.W. (2013). Monochromated STEM
With A 30 Mev-Wide, Atom-Sized Electron Probe. Microscopy 62(1), 3-21.
KYOTANI, T., TSAI, L.F. & TOMITA, A. (1997). Formation Of Platinum Nanorods And Nanoparticles
In Uniform Carbon Nanotubes Prepared By A Template Carbonization Method. Chemical
Communications (7), 701-702.
LAOUFI, I., SAINT-LAGER, M.C., LAZZARI, R., JUPILLE, J., ROBACH, O., GARAUDEE, S.,
CABAILH, G., DOLLE, P., CRUGUEL, H. & BAILLY, A. (2011). Size And Catalytic Activity Of
Supported Gold Nanoparticles: An In Operando Study During CO Oxidation. Journal Of
Physical Chemistry C 115(11), 4673-4679.
LEARY, R., DE LA PENA, F., BARNARD, J.S., LUO, Y., ARMBRUESTER, M., THOMAS, J.M. & MIDGLEY,
P.A. (2013a). Revealing The Atomic Structure Of Intermetallic Gapd2 Nanocatalysts By Using
Aberration-Corrected Scanning Transmission Electron Microscopy. Chemcatchem 5(9), 2599-
2609.
LEARY, R., SAGHI, Z., MIDGLEY, P.A. & HOLLAND, D.J. (2013b). Compressed Sensing Electron
Tomography. Ultramicroscopy 131, 70-91.
174
LEE, J.-H., JANG, J.-T., CHOI, J.-S., MOON, S.H., NOH, S.-H., KIM, J.-W., KIM, J.-G., KIM, I.-S.,
PARK, K.I. & CHEON, J. (2011). Exchange-Coupled Magnetic Nanoparticles For Efficient Heat
Induction. Nature Nanotechnology 6(7), 418-422.
LEPINAY, K., LORUT, F., PANTEL, R. & EPICIER, T. (2013). Chemical 3D Tomography Of 28 Nm
High K Metal Gate Transistor: STEM XEDS Experimental Method And Results. Micron 47, 43-
49.
LEWIS, E., SLATER, T., PRESTAT, E., MACEDO, A., O'BRIEN, P., CAMARGO, P. & HAIGH, S. (2014a).
Real-Time Imaging And Elemental Mapping Of Agau Nanoparticle Transformations. Nanoscale
6, 13598-13605.
LEWIS, E.A., HAIGH, S.J., SLATER, T.J.A., HE, Z., KULZICK, M.A., BURKE, M.G. & ZALUZEC, N.J.
(2014b). Real-Time Imaging And Local Elemental Analysis Of Nanostructures In Liquids.
Chemical Communications 50, 10019-10022.
LI, R.F., WU, G.H., JIANG, L.T. & SUN, D.L. (2015). Characterization Of Multi-Scale Porous
Structure Of Fly Ash/Phosphate Geopolymer Hollow Sphere Structures: From Submillimeter To
Nano-Scale. Micron 68, 54-58.
LIAKAKOS, N., GATEL, C., BLON, T., ALTANTZIS, T., LENTIJO-MOZO, S., GARCIA-MARCELOT, C.,
LACROIX, L.-M., RESPAUD, M., BALS, S., VAN TENDELOO, G. & SOULANTICA, K. (2014). Co-Fe
Nanodumbbells: Synthesis, Structure, And Magnetic Properties. Nano Letters 14(5), 2747-2754.
LIAO, Y. & MARKS, L.D. (2013). Reduction Of Electron Channeling In EDS Using Precession.
Ultramicroscopy 126, 19-22.
LINK, S., WANG, Z.L. & EL-SAYED, M.A. (1999). Alloy Formation Of Gold-Silver Nanoparticles
And The Dependence Of The Plasmon Absorption On Their Composition. Journal Of Physical
Chemistry B 103(18), 3529-3533.
LIU, G.S., HOUSE, S.D., KACHER, J., TANAKA, M., HIGASHIDA, K. & ROBERTSON, I.M.
(2014). Electron Tomography Of Dislocation Structures. Materials Characterization 87, 1-11.
LIU, J.Y. (2005). Scanning Transmission Electron Microscopy and its application to the study of
nanoparticles and nanoparticle systems. Journal of Electron Microscopy 54(3), 251-278.
LIU, Z., MA, L., ZHANG, J., HONGSIRIKARN, K. & GOODWIN, J.G., JR. (2013). Pt Alloy
Electrocatalysts for Proton Exchange Membrane Fuel Cells: A Review. Catalysis Reviews-
Science and Engineering 55(3), 255-288.
LIZ-MARZAN, L.M. (2006). Tailoring surface plasmons through the morphology and assembly
of metal nanoparticles. Langmuir 22(1), 32-41.
175
LOPEZ-HARO, M., JIU, T., BAYLE-GUILLEMAUD, P., JOUNEAU, P.-H. & CHANDEZON, F. (2013).
Multiscale tomographic analysis of polymer-nanoparticle hybrid materials for solar cells.
Nanoscale 5(22), 10945-10955.
MAIRE, E. & WITHERS, P.J. (2014). Quantitative X-ray tomography. International Materials
Reviews 59(1), 1-43.
MARUYAMA, W., AKAO, Y., CARRILLO, M.C., KITANI, K., YOUDIUM, M.B.H. & NAOI, M. (2002a).
Neuroprotection by propargylamines in Parkinson's disease - Suppression of apoptosis and
induction of prosurvival genes. Neurotoxicology and Teratology 24(5), 675-682.
MARUYAMA, W., BOULTON, A.A., YOUDIM, M.B.H. & NAOI, M. (2002b). Anti-apoptotic function of
propargylamines. Catecholamine Research: from Molecular Insights to Clinical Medicine 53,
241-244.
MAYER, J., GIANNUZZI, L.A., KAMINO, T. & MICHAEL, J. (2007). TEM sample preparation
and FIB-induced damage. Mrs Bulletin 32(5), 400-407.
MAYORAL, A., DEEPAK, F.L., ESPARZA, R., CASILLAS, G., MAGEN, C., PEREZ-TIJERINA, E. & JOSE-
YACAMAN, M. (2012). On the structure of bimetallic noble metal nanoparticles as revealed by
aberration corrected scanning transmission electron microscopy (STEM). Micron 43(4), 557-564.
MERKLE, A., LECHNER, L., STEINBACH, A., GELB, J., KIENLE, M., PHANEUF, M., UNRAU, D., SINGH, S.
& CHAWLA, N. (2014). Automated correlative tomography using XRM and FIB-SEM to span
length scales and modalities in 3D materials. In Microscopy and Analysis, pp. S10-S13. Wiley.
MIDGLEY, P.A. & WEYLAND, M. (2003). 3D electron microscopy in the physical sciences: the
development of Z-contrast and EFTEM tomography. Ultramicroscopy 96(3-4), 413-431.
MIDGLEY, P.A. & DUNIN-BORKOWSKI, R.E. (2009). Electron tomography and holography in
materials science. Nature Materials 8(4), 271-280.
MISZTA, K., BRESCIA, R., PRATO, M., BERTONI, G., MARRAS, S., XIE, Y., GHOSH, S., KIM, M.R. &
MANNA, L. (2014). Hollow and Concave Nanoparticles via Preferential Oxidation of the Core in
Colloidal Core/Shell Nanocrystals. Journal of the American Chemical Society 136(25), 9061-
9069.
176
MOBUS, G., DOOLE, R.C. & INKSON, B.J. (2003). Spectroscopic electron tomography.
Ultramicroscopy 96(3-4), 433-451.
MOBUS, G. & INKSON, B.J. (2007). Nanoscale tomography in materials science. Materials Today
10(12), 18-25.
MOBUS, G., SAGHI, Z., GUAN, W., GNANAVEL, T., XU, X. & PENG, Y. (2010). Hybrid Tomography
for structural and chemical 3D imaging on the nanoscale. In Electron Microscopy and Analysis
Group Conference 2009, Baker, R. T. (Ed.), Bristol: Iop Publishing Ltd.
MOMOSE, A., TAKEDA, T., ITAI, Y. & HIRANO, K. (1996). Phase-contrast X-ray computed
tomography for observing biological soft tissues. Nature Medicine 2(4), 473-475.
MOURDIKOUDIS, S., CHIREA, M., ZANAGA, D., ALTANTZIS, T., MITRAKAS, M., BALS, S., LIZ-MARZAN,
L.M., PEREZ-JUSTE, J. & PASTORIZA-SANTOS, I. (2015). Governing the morphology of Pt-Au
heteronanocrystals with improved electrocatalytic performance. Nanoscale 7(19), 8739-8747.
MUGNAIOLI, E., GORELIK, T. & KOLB, U. (2009). "Ab initio" structure solution from electron
diffraction data obtained by a combination of automated diffraction tomography and precession
technique. Ultramicroscopy 109(6), 758-765.
MULLER, D.A. & SILCOX, J. (1995). Radiation damage of Ni3Al by 100 kev electrons.
Philosophical Magazine A 71(6), 1375-1387.
MYERS, B., MCILWRATH, K., WU, J., LI, S., INADA, H. & DRAVID, V. (2010). Performance
Characterization of Hitachi HD-2300A STEM with Dual-EDS Configuration. Microscopy and
Microanalysis 16(S2), 942-943.
NELLIST, P.D. & WANG, P. (2012). Optical Sectioning and Confocal Imaging and Analysis in
the Transmission Electron Microscope. Annual Review of Materials Research 42, 125-143.
NEWBURY, D.E. (2006). The New X-ray Mapping: X-ray Spectrum Imaging above 100 kHz
Output Count Rate with the Silicon Drift Detector. Microscopy and Microanalysis 12(1), 26-35.
NICOLETTI, O., DE LA PENA, F., LEARY, R.K., HOLLAND, D.J., DUCATI, C. & MIDGLEY, P.A. (2013).
Three-dimensional imaging of localized surface plasmon resonances of metal nanoparticles.
Nature 502(7469), 80.
177
PARK, J., ELMLUND, H., ERCIUS, P., YUK, J.M., LIMMER, D.T., CHEN, Q., KIM, K., HAN,
S.H., WEITZ, D.A., ZETTL, A. & ALIVISATOS, A.P. (2015). Nanoparticle imaging. 3D structure
of individual nanocrystals in solution by electron microscopy. Science 349(6245), 290-295.
PATTERSON, A.L. (1939). The Scherrer formula for x-ray particle size determination. Physical
Review 56(10), 978-982.
PENCZEK, P., MARKO, M., BUTTLE, K. & FRANK, J. (1995). Double-tilt electron tomography.
Ultramicroscopy 60(3), 393-410.
PENNYCOOK, S.J. & JESSON, D.E. (1990). High-resolution incoherent imaging of crystals.
Physical Review Letters 64(8), 938-941.
POTAPOV, P.L. & SCHRYVERS, D. (2004). Measuring the absolute position of EELS ionisation
edges in a TEM. Ultramicroscopy 99(1), 73-85.
POWELL, C.J. (1976). Cross sections for ionization of inner-shell electrons by electrons. Reviews
of Modern Physics 48(1), 33.
PRODAN, E., RADLOFF, C., HALAS, N.J. & NORDLANDER, P. (2003). A hybridization model
for the plasmon response of complex nanostructures. Science 302(5644), 419-422.
RAYLEIGH. (1879). XXXI. Investigations in optics, with special reference to the spectroscope.
Philosophical Magazine Series 5 8(49), 261-274.
REED, S.J.B. (1982). The single-scattering model and spatial-resolution in x-ray-analysis of thin
foils. Ultramicroscopy 7(4), 405-409.
178
RODRIGUEZ-FERNANDEZ, D., ALTANTZIS, T., HEIDARI, H., BALS, S. & LIZ-MARZAN, L.M. (2014). A
protecting group approach toward synthesis of Au-silica Janus nanostars. Chemical
Communications 50(1), 79-81.
RODRIGUEZ-GONZALEZ, B., BURROWS, A., WATANABE, M., KIELY, C.J. & LIZ-MARZAN, L.M. (2005).
Multishell bimetallic auag nanoparticles: synthesis, structure and optical properties. Journal of
Materials Chemistry 15(17), 1755-1759.
ROSSELL, M.D., RAMASSE, Q.M., FINDLAY, S.D., RECHBERGER, F., ERNI, R. & NIEDERBERGER, M.
(2012). Direct Imaging of Dopant Clustering in Metal-Oxide Nanoparticles. Acs Nano 6(8), 7077-
7083.
RUOFF, R.S., LORENTS, D.C., CHAN, B., MALHOTRA, R. & SUBRAMONEY, S. (1993). Single-crystal
metals encapsulated in carbon nanoparticles. Science 259(5093), 346-348.
SAGHI, Z., XU, X., PENG, Y., INKSON, B. & MOBUS, G. (2007). Three-dimensional chemical
analysis of tungsten probes by energy dispersive x-ray nanotomography. Applied Physics
Letters 91(25), 251906.
SAKAMOTO, T., CHENG, Z.H., TAKAHASHI, M., OWARI, M. & NIHEI, Y. (1998). Development of an
ion and electron dual focused beam apparatus for three-dimensional microanalysis. Japanese
Journal of Applied Physics Part 1-Regular Papers Short Notes & Review Papers 37(4A), 2051-
2056.
SASAKI, K., NAOHARA, H., CHOI, Y., CAI, Y., CHEN, W.-F., LIU, P. & ADZIC, R.R. (2012). Highly
stable Pt monolayer on pdau nanoparticle electrocatalysts for the oxygen reduction reaction.
Nature Communications 3, 1115.
SCOTT, M.C., CHEN, C.C., MECKLENBURG, M., ZHU, C., XU, R., ERCIUS, P., DAHMEN, U., REGAN,
B.C. & MIAO, J.W. (2012). Electron tomography at 2.4-angstrom resolution. Nature 483(7390),
444-491.
SHAO, M.H., HUANG, T., LIU, P., ZHANG, J., SASAKI, K., VUKMIROVIC, M.B. & ADZIC, R.R.
(2006). Palladium monolayer and palladium alloy electrocatalysts for oxygen reduction.
Langmuir 22(25), 10409-10415.
SHARMA, R. (2005). An environmental transmission electron microscope for in situ synthesis and
characterization of nanomaterials. Journal of Materials Research 20(7), 1695-1707.
SHERIDAN, P.J. (1989). Determination of experimental and theoretical kasi factors for a 200-kv
analytical electron-microscope. Journal of Electron Microscopy Technique 11(1), 41-61.
179
SILCOX, J. (1998). Core-loss EELS. Current Opinion in Solid State & Materials Science 3(4),
336-342.
SONDHEIMER, E.H. (2001). The mean free path of electrons in metals. Advances in Physics 50(6),
499-537.
SPENCE, J.C.H., ZUO, J.M. & LYNCH, J. (1989). On the holz contribution to stem lattice images
formed using high-angle dark-field detectors. Ultramicroscopy 31(2), 233-240.
SUN, Y.G. & XIA, Y.N. (2003). Alloying and dealloying processes involved in the preparation of
metal nanoshells through a galvanic replacement reaction. Nano Letters 3(11), 1569-1572.
SUN, Y.G. & XIA, Y.N. (2004). Mechanistic study on the replacement reaction between silver
nanostructures and chloroauric acid in aqueous medium. Journal of the American Chemical
Society 126(12), 3892-3901.
TARIQ, F., HASWELL, R., LEE, P.D. & mCcOMB, D.W. (2011). Characterization of hierarchical pore
structures in ceramics using multiscale tomography. Acta Materialia 59(5), 2109-2120.
TATTON, W.G., CHALMERS-REDMAN, R.M.E., JU, W.J.H., MAMMEN, M., CARLILE, G.W., PONG, A.W.
& TATTON, N.A. (2002). Propargylamines induce antiapoptotic new protein synthesis in serum-
and nerve growth factor (NGF)-withdrawn, NGF-differentiated PC-12 cells. Journal of
Pharmacology and Experimental Therapeutics 301(2), 753-764.
THIELE, E.W. (1939). Relation between catalytic activity and size of particle. Industrial and
Engineering Chemistry 31, 916-920.
180
THIELE, S., FURSTENHAUPT, T., BANHAM, D., HUTZENLAUB, T., BIRSS, V., ZIEGLER, C.
& ZENGERLE, R. (2013). Multiscale tomography of nanoporous carbon-supported noble metal
catalyst layers. Journal of Power Sources 228, 185-192.
THOMPSON, K., LAWRENCE, D., LARSON, D.J., OLSON, J.D., KELLY, T.F. & GORMAN, B.
(2007). In situ site-specific specimen preparation for atom probe tomography. Ultramicroscopy
107(2-3), 131-139.
TOSHIMA, N. & YONEZAWA, T. (1998). Bimetallic nanoparticles - novel materials for chemical
and physical applications. New Journal of Chemistry 22(11), 1179-1201.
TRAN, D.T., JONES, I.P., PREECE, J.A., JOHNSTON, R.L., DEPLANCHE, K. & MACASKIE, L.E. (2012).
Configuration of microbially synthesized Pd-Au nanoparticles studied by STEM-based
techniques. Nanotechnology 23(5), 055701.
TREACY, M.M.J. (2011). Z Dependence of Electron Scattering by Single Atoms into Annular
Dark-Field Detectors. Microscopy and Microanalysis 17(6), 847-858.
UCHIC, M.D., HOLZER, L., INKSON, B.J., PRINCIPE, E.L. & MUNROE, P. (2007). Three-
dimensional microstructural characterization using focused ion beam tomography. Mrs Bulletin
32(5), 408-416.
UNGER, V.M. (2001). Electron cryomicroscopy methods. Current Opinion in Structural Biology
11(5), 548-554.
VAN AERT, S., BATENBURG, K.J., ROSSELL, M.D., ERNI, R. & VAN TENDELOO, G. (2011). Three-
dimensional atomic imaging of crystalline nanoparticles. Nature 470(7334), 374-377.
VAN DEN BROEK, W., ROSENAUER, A., GORIS, B., MARTINEZ, G.T., BALS, S., VAN AERT, S. & VAN
DYCK, D. (2012). Correction of non-linear thickness effects in HAADF STEM electron
tomography. Ultramicroscopy 116, 8-12.
VAN HEEL, M., GOWEN, B., MATADEEN, R., ORLOVA, E.V., FINN, R., PAPE, T., COHEN, D.,
STARK, H., SCHMIDT, R., SCHATZ, M. & PATWARDHAN, A. (2000). Single-particle electron
cryo-microscopy: towards atomic resolution. Quarterly Reviews of Biophysics 33(4), 307-369.
VARELA, M., FINDLAY, S.D., LUPINI, A.R., CHRISTEN, H.M., BORISEVICH, A.Y., DELLBY, N., KRIVANEK,
O.L., NELLIST, P.D., OXLEY, M.P., ALLEN, L.J. & PENNYCOOK, S.J. (2004). Spectroscopic imaging
of single atoms within a bulk solid. Physical Review Letters 92(9), 095502 .
VILLINGER, C., GREGORIUS, H., KRANZ, C., HOEHN, K., MUENZBERG, C., VON WICHERT,
G., MIZAIKOFF, B., WANNER, G. & WALTHER, P. (2012). FIB/SEM tomography with TEM-like
resolution for 3D imaging of high-pressure frozen cells. Histochemistry and Cell Biology 138(4),
549-556.
181
VON HARRACH, H., KLENOV, D., FREITAG, B., SCHLOSSMACHER, P., COLLINS, P. & FRASER, H.
(2010). Comparison of the Detection Limits of EDS and EELS in S/TEM. Microscopy and
Microanalysis 16(S2), 1312-1313.
VON HARRACH, H.S., DONA, P., FREITAG, B., SOLTAU, H., NICULAE, A. & ROHDE, M. (2009). An
integrated Silicon Drift Detector System for FEI Schottky Field Emission Transmission Electron
Microscopes. Microscopy and Microanalysis 15, 208-209.
WANG, C.M., LIAO, H.-G. & ROSS, F.M. (2015). Observation of materials processes in liquids by
electron microscopy. MRS Bulletin 40(1), 46-52.
WANG, C.M., GENC, A., CHENG, H., PULLAN, L., BAER, D.R. & BRUEMMER, S.M. (2014). In-Situ
TEM visualization of vacancy injection and chemical partition during oxidation of Ni-Cr
nanoparticles. Scientific Reports 4, 3683.
WANG, D. & LI, Y. (2011). Bimetallic Nanocrystals: Liquid-Phase Synthesis and Catalytic
Applications. Advanced Materials 23(9), 1044-1060.
WANG, D., XIN, H.L., HOVDEN, R., WANG, H., YU, Y., MULLER, D.A., dIsALVO, F.J. & ABRUNA, H.D.
(2013). Structurally ordered intermetallic platinum-cobalt core-shell nanoparticles with enhanced
activity and stability as oxygen reduction electrocatalysts. Nature Materials 12(1), 81-87.
WANG, D., YU, Y., XIN, H.L., HOVDEN, R., ERCIUS, P., MUNDY, J.A., CHEN, H., RICHARD, J.H.,
MULLER, D.A., dIsALVO, F.J. & ABRUNA, H.D. (2012). Tuning Oxygen Reduction Reaction Activity
via Controllable Dealloying: A Model Study of Ordered Cu3Pt/C Intermetallic Nanocatalysts.
Nano Letters 12(10), 5230-5238.
WANG, P., D'ALFONSO, A.J., FINDLAY, S.D., ALLEN, L.J. & BLELOCH, A.L. (2008). Contrast
Reversal in Atomic-Resolution Chemical Mapping. Phys. Rev. Lett. 101(23), 236102.
WATANABE, M. & WILLIAMS, D.B. (2006). The quantitative analysis of thin specimens: a review of
progress from the Cliff-Lorimer to the new zeta-factor methods. Journal of Microscopy-Oxford
221(2), 89-109.
WILKINS, S.W., GUREYEV, T.E., GAO, D., POGANY, A. & STEVENSON, A.W. (1996). Phase-
contrast imaging using polychromatic hard X-rays. Nature 384(6607), 335-338.
WILLETS, K.A. & VAN DUYNE, R.P. (2007). Localized surface plasmon resonance
spectroscopy and sensing. In Annual Review of Physical Chemistry, pp. 267-297. Palo Alto:
Annual Reviews.
182
WILLIAMS, D.B. & GOLDSTEIN, J.I. (1991). Quantitative x-ray-microanalysis in the analytical
electron-microscope. Electron Probe Quantitation, 371-398.
WILLIAMS, D., GOLDSTEIN, J., NEWBURY, D. & LUND, M.W. (1995). Current Trends in Si(Li)
Detector Windows for Light Element Analysis. In X-Ray Spectrometry in Electron Beam
Instruments, pp. 21-31. Springer US.
WILLIAMS, D.B., PAPWORTH, A.J. & WATANABE, M. (2002). High resolution X‐ray mapping
in the STEM. Journal of Electron Microscopy 51(S1), S113-S126.
WILLIAMS, D.B. & CARTER, C.B. (2009a). Imaging. In Transmission Electron Microscopy, New
York: Springer US.
WILLIAMS, D.B. & CARTER, C.B. (2009b). Basics. In Transmission Electron Microscopy, New
York: Springer US.
WILLIAMS, D.B. & CARTER, C.B. (2009c). Spectrometry. In Transmission Electron Microscopy,
New York: Springer US.
WILSON, A.R., SUN, K., CHI, M., WHITE, R.M., lEbEAU, J.M., LAMB, H.H. & WILEY, B.J. (2013).
From Core-Shell to Alloys: The Preparation and Characterization of Solution-Synthesized aupd
Nanoparticle Catalysts. Journal of Physical Chemistry C 117(34), 17557-17566.
XIA, X., WANG, Y., RUDITSKIY, A. & XIA, Y. (2013). 25th Anniversary Article: Galvanic
Replacement: A Simple and Versatile Route to Hollow Nanostructures with Tunable and Well-
Controlled Properties. Advanced Materials 25(44), 6313-6333.
XIA, Y., YANG, H. & CAMPBELL, C.T. (2013). Nanoparticles for Catalysis. Accounts of
Chemical Research 46(8), 1671-1672.
XIN, H.L.L., DWYER, C., MULLER, D.A., ZHENG, H.M. & ERCIUS, P. (2013). Scanning
Confocal Electron Energy-Loss Microscopy Using Valence-Loss Signals. Microscopy and
Microanalysis 19(4), 1036-1049.
YAGUCHI, T., KONNO, M., KAMINO, T. & WATANABE, M. (2008). Observation of three-dimensional
elemental distributions of a Si device using a 360 degrees-tilt FIB and the cold field-emission
STEM system. Ultramicroscopy 108(12), 1603-1615.
YAMAMOTO, Y., BRADY, M.P., LU, Z.P., LIU, C.T., TAKEYAMA, M., MAZIASZ, P.J. & PINT,
B.A. (2007). Alumina-forming austenitic stainless steels strengthened by laves phase and MC
183
carbide precipitates. Metallurgical and Materials Transactions a-Physical Metallurgy and
Materials Science 38A(11), 2737-2746.
YAMASAKI, J., MUTOH, M., OHTA, S., YUASA, S., ARAI, S., SASAKI, K. & TANAKA, N. (2014). Analysis
of nonlinear intensity attenuation in bright-field TEM images for correct 3D reconstruction of the
density in micron-sized materials. Microscopy 63(5), 345-355.
YAZZIE, K.E., WILLIAMS, J.J., PHILLIPS, N.C., DE CARLO, F. & CHAWLA, N. (2012). Multiscale
microstructural characterization of Sn-rich alloys by three dimensional (3D) X-ray synchrotron
tomography and focused ion beam (FIB) tomography. Materials Characterization 70, 33-41.
YEDRA, L., ELJARRAT, A., ARENAL, R., PELLICER, E., CABO, M., LOPEZ-ORTEGA, A., ESTRADER, M.,
SORT, J., DOLORS BARO, M., ESTRADE, S. & PEIRO, F. (2012). EEL spectroscopic tomography:
Towards a new dimension in nanomaterials analysis. Ultramicroscopy 122, 12-18.
YEDRA, L., ELJARRAT, A., REBLED, J.M., LOPEZ-CONESA, L., DIX, N., SANCHEZ, F., ESTRADE, S. &
PEIRO, F. (2014). EELS tomography in multiferroic nanocomposites: from spectrum images to
the spectrum volume. Nanoscale 6(12), 6646-6650.
YEOH, C.S.M., ROSSOUW, D., SAGHI, Z., BURDET, P., LEARY, R.K. & MIDGLEY, P.A. (2015). The
Dark Side of EDX Tomography: Modeling Detector Shadowing to Aid 3D Elemental Signal
Analysis. Microscopy and Microanalysis 21(3), 759-764
YU, K. & CHEN, J. (2009). Enhancing Solar Cell Efficiencies through 1-D Nanostructures.
Nanoscale Research Letters 4(1), 1-10.
YU, Q., LEGROS, M. & MINOR, A.M. (2015). In situ TEM nanomechanics. Mrs Bulletin 40(1), 62-
68.
ZALUZEC, N.J. (2004). XEDS Systems for the Next Generation Analytical Electron Microscope.
Microscopy and Microanalysis 10(S2), 122-123.
ZALUZEC, N.J. (2009). Innovative Instrumentation for Analysis of Nanoparticles: The π Steradian
Detector. Microscopy Today 17(4), 56-59.
ZANKEL, A., KRAUS, B., POELT, P., SCHAFFER, M. & INGOLIC, E. (2009). Ultramicrotomy in
the ESEM, a versatile method for materials and life sciences. Journal of Microscopy-Oxford
233(1), 140-148.
ZANKEL, A., WAGNER, J. & POELT, P. (2014). Serial sectioning methods for 3D investigations
in materials science. Micron 62, 66-78.
184
ZHANG, D., OLEYNIKOV, P., HOVMOLLER, S. & ZOU, X. (2010). Collecting 3D electron
diffraction data by the rotation method. Zeitschrift Fur Kristallographie 225(2-3), 94-102.
ZHANG, Q., XIE, J., LEE, J.Y., ZHANG, J. & BOOTHROYD, C. (2008). Synthesis of Ag@agau metal
core/alloy shell bimetallic nanoparticles with tunable shell compositions by a galvanic
replacement reaction. Small 4(8), 1067-1071.
ZIESE, U., DE JONG, K.P. & KOSTER, A.J. (2004). Electron tomography: a tool for 3D structural
probing of heterogeneous catalysts at the nanometer scale. Applied Catalysis a-General 260(1),
71-74.
ZREIBA, N.A. & KELLY, T.F. (1988). Absorption and fluorescence corrections of characteristic x-
rays from thin spheres. X-Ray Spectrometry 17(6), 229-238.
185
Appendix A - Conditions for successful STEM-EDX tomography
The STEM-EDX tilt series’ used in both chapters 5 and 6 suffer from extremely low count
statistics, for the reasons described in section 2.3. An example single pixel spectrum is
presented in Figure A1 to demonstrate the low counts that can be expected in the case of small
nanoparticle samples.
Figure A1. EDX spectrum from the single pixel indicated in the inset image, taken from a
spectrum image at 0° tilt of the high-Au content AgAu nanoparticle described in chapter 6. A
For the small (less than 30 nm thickness) nanoparticles used in chapters 5 and 6, the conditions
used here resulted in single pixel spectra with between 0 – 5 counts for both the Au and Ag
peaks analysed. While the results in chapter 6 still demonstrate that this number of counts can
increase counts, as could be performed by binning pixels in this case. In order to achieve
maximum counts per pixel, it is advisable to attempt to make the pixel size as large as possible,
186
In the case of the stainless steel sample described in chapter 7, where the sample can tolerate
a larger dose, single pixel counts per X-ray peak ranged from 0 – 60. A single pixel spectrum
Figure A2. EDX spectrum from the single pixel indicated in the inset image, taken from a
spectrum image at 0° tilt of the stainless steel SS316 needle sample described in chapter 7. A
In chapters 5 and 6 a SIRT algorithm with 5 iterations was used to reconstruct the STEM-EDX
tilt series’. The choice of reconstruction algorithm and in the case of SIRT, the number of
iterations, is not straightforward and here has been carried out through qualitative inspection of
the reconstructions at a number of different conditions. Reconstructions of the tilt series’ used in
chapter 6 of an AgAu nanoparticle are shown in Figure A3 (for Au) and Figure A4 (for Ag). Here,
sharpness of the reconstruction and reduction in intensity of streaking artefacts and of the
missing wedge.
187
Figure A3. XY slices of reconstructions of the Au maps of a STEM-EDX tilt series of one of the
AgAu nanoparticles described in chapter 6, from a tilt series acquired with 5° angular
188
Figure A4. XY slices of reconstructions of the Ag maps of a STEM-EDX tilt series of one of the
AgAu nanoparticles described in chapter 6, from a tilt series acquired with 5° angular
The number of SIRT iterations was chosen based on the point at which high frequency
STEM-EDX almost always has very low signal levels and SIRT is known to enhance high
frequency information, including noise, as the number of iterations increase (Louis & Natterer,
1983; Defrise, 1988). Therefore, for the low signal-to-noise data obtained with STEM-EDX
tomography it may be the case that fewer iterations are used than for similarly acquired HAADF
tilt series'. Here, while the sharpness of the exterior of the nanoparticle improves up to 5
iterations in the Au reconstructions (Figure A3), subsequent iterations appear to act only to
resolve small features within the nanoparticle, which can be associated with noise in the initial
projections.
189