You are on page 1of 4

CASE STUDY

medical technologies have never had and expert knowledge needed to analyse
CARDIAC SERVICES & THE
CHANGING ENVIRONMENT OF their own appropriately reserved space. the installation environment and to map
MEDICAL TELEMETRY As more devices are being introduced out the correct position for access points
to the hospital environment, there is an to obtain the best performance. For
increased responsibility in managing the Cardiac Services, the whole process of
In the last 40 years Cardiac Services electromagnetic spectrum needed for deploying a telemetry system starts right
have witnessed the advances of Medical these wireless devices to communicate from the initial enquiry with mapping
Telemetry technology as its capabilities effectively. the hospital floor plans, understanding
move from beyond the CCU into general the structure of the building based on
medical-surgical units. Today, patients where the walls and obstructions are and
with potential cardiac concerns can be CHOOSING A SOLUTION - how many walls the radio waves must
monitored through a wireless centralised WIFI VS PROPRIETARY
travel through. Only at this point can the
system by staff who are not directly in the telemetry team begin to scope out the
patient’s vicinity. Monitoring from a remote Most manufacturers now offer solutions in installation at hand and ensure the correct
location within the hospital has given the 2.4/5 GHz frequency range. Essentially and most efficient number of access points
skilled nursing staff the ability to conduct there are two options using a proprietary for the system are installed.
24-hour visualisation across multiple units. solution or using the 802.11 wireless
This relieves local staff of responsibilities standard that most company/Hospital
for high acuity patient monitoring, freeing WIFI solutions use. It’s critical to note
WATERFORD HOSPITAL
up acute beds for critical cardiac patients that in choosing a WI-FI based system,
and overall it has a significant impact on the Hospital IT Department ultimately
improving the patient care. Today, there takes on the responsibility for delivering
are multiple players within the market an infrastructure that can maintain and Cardiac Services have a long standing
providing Radio Frequency Medical deliver a network that provides delivery partnership with Waterford University
Telemetry. Some providing solutions using of “life critical” data. Just having a WIFI Hospital, providing support and service
Wi-Fi and others through proprietary network does not mean it is suitable for for Medical Telemetry. In 2009, the
systems. However, both present challenges use for a critical patient monitoring system. hospital started to move into the 2.4 GHz
of their own, not least from the influx of For instance, most hospitals are not arena as it wished to extend the then UHF
mobile devices crowding an already busy equipped for people to send emails from telemetry system into the new AMAU.
radio frequency (RF) environment. the bathrooms, it’s simply not a design Cardiac Services began to roll out the new
consideration. But in a hospital/telemetry Philips ITS system by adding the AMAU
“With the increase in wireless setting, that is one place that definitely area of their hospital while retaining the
technologies available, needs a strong signal when mapping out existing UHF system. Over time the system
the environment in which the access points for installation. expanded through the hospital and
medical telemetry is
they slowly replaced the old system with
performing within has become Similarly, with the use of a proprietary the new.
crowded and polluted.” system, it is critical to have the support
-Colm Nolan, Services Manager,
Cardiac Services

UNDERSTANDING THE
WIRELESS SPECTRUM

Proper management and monitoring


of the radiofrequency environment
has never been more important. The
RF spectrum runs from VHF, UHF into
the present WIFI frequencies 2.4/5
GHz. Within these ranges there are
restricted bands where broadcast radio,
TV, military, Aircraft and mobile phones
are licensed to operate. Others such as
2.4/5 GHz, are less regulated, operators
have to comply with some technical
standards, but are then expected to
co-exist with each other. Unfortunately,
despite its obvious critical nature

2
OPERATING A WIRELESS
MONITORING SYSTEM IN A
CHALLENGING ENVIRONMENT

Towards the end of 2016, Cardiac Services


began to receive reports from the clinical
engineers of Waterford Hospital about
dropouts in signal in their telemetry
system. The problem manifested in the
intermittent loss of random telemetry
signals whereby those affected patients
were not monitored for certain durations.
Any situation where a telemetry signal is
interrupted is a matter of great concern
from a patient safety perspective, and this
further emphasised the need to ensure
certain progress and the earliest resolution
e of the problem. This fault was elusive
and intermittent which by its nature
necessitated a logical approach where
various assemblies were methodically
and incrementally eliminated as the
source of the problem. The intermittent
nature of this problem, and the inability

Designed
something interfering with the signal of the “that was the point that we stopped
to recreate the fault also meant that after
Telemetry System but they were struggling looking for the source of interference
every iteration of fault finding, a period
to identify what exactly it was. They and started dealing with the

responsibly
of observation/delay was necessary to
persevered and eliminated many things environment we were living in”
establish if the problem was improved or
that could potentially be the issue but still
resolved. Following standard procedures
to no avail. Cardiac Services began to measure the
and checks, the Cardiac team then
s signal strength for certain access points.
proceeded to explore possible changes The IntelliVue MX40 is designed with ease of cleaning in mind and
Followingto nearly
support8 your
months of trialling,
hospital’s environmentalBy moving these points to a more central
initiatives.
within the Hospitals environment, given
testing, and reporting, Cardiac Services and point in each area and adding a number
that it was in fact 7 years from the initial
Philips introduced an additional expert in of additional points signal/noise ratio
install of the equipment refresh.
RF troubleshooting from the USA. He came was significantly improved. They focused
The print quality of this copy is not an accurate representation of the original.
over to Ireland for a week in April with on the two operational wards and they
During the Christmas period the Hospital
highly sophisticated instrumentation and worked with the hospital to get the
reduced the telemetry service down to a
scanners that measured the RF spectrum in system performing to a capacity that
specific area to ensure CCU nursing staff
a more detailed way. Cardiac Services and they were happy with. They could then
could keep a close eye on patients that
Philips were determined to find a solution begin to talk about correcting the rest of
needed to be monitored and they would
that resolved the RF interferences and the system and after a period of time all
be able to assist quickly in the case of a
ensure that no other customer would have the remaining wards were restored to full
cardiac event. Cardiac Services continued
an issue like this going forward. working capacity.
to work closely with the hospital and
e
engaged with their contractors to run
During the fault finding process, a
some temporary cables in the hospital in DISCOVERY & RESOLUTION significant number of assemblies were
a bid to find a solution to the level of drop
replaced in an effort to resolve the
outs that were occurring. Although there
problem.
were some improvements, it just wasn’t
Following the ongoing collaboration
resolving the matter.
between Cardiac Services, Philips and “The perseverance, commitment and
Waterford University Hospital, it was strength of the local knowledge of
At this point, Cardiac Services engaged
discovered that the problem lay in Cardiac Services, backed up by the
the support of their partner Philips in the
the background/baseline noise within global expertise of Philips Healthcare,
UK to ensure they could find a resolution
the hospital. The levels of noise were proved to be an invaluable combination
as soon as possible. Philips sent over one
equivalent to the signal strength of the in the resolution of this problem”
of their own engineers with a spectrum
telemetry transmitters and the Philips ITS (Seamus Guiry, Chief Clinical Engineer/
analyser. At this point, both Cardiac
System was being overwhelmed when the Contracts Officer, University Hospital
and Philips suspected that there was
baseline noise rose slightly. Waterford)

3
Viewing information at the patient’s side g
caregiver access to the current patient stat
enhance care.

Having identified this critical systems over the last 50 years and
challenge, Cardiac Services now coupled with the global experience of
provide all Telemetry customers their partners, Cardiac Services and
with the option of having an annual Philips Healthcare have developed a
RF survey. The aim is to continue range of services that together ensure
to monitor and track any changes the correct and continued operation of
that may occur within the hospital installed wireless Patient monitoring
environments and to mitigate any systems for hospitals all over Ireland.
potential interferences before
they become a problem. These services start with pre-sale RF
and placement surveys. After
installation, a coverage assessment
survey verifies the performance of the
Key Take Away: system in the environment at that time.
Annual RF and coverage assessment
The RF environment in any hospital is surveys ensure that the system
never fixed. On a day to day basis the performance is monitored quantitatively
signals from the various broadcasters into the future. These latter surveys
flux in strength. Additional devices are will identify degradation in the RF
brought into the hospital continuously, environment prior to loss of performance
some for short periods (visitor devices) and allow for preventative corrective
others for longer (new systems). The print quality of this copy is not an accurate representation of the original.
action to be taken.
References:
No longer can we assume when we On the hospital side consideration John O’Rourke MSc in Clinical Engineering
install wireless monitoring systems that should be given to the establishment and Network Design and network design and
they will continue to function without of a group who would manage the RF implementation engineer, Oxygen Care LTD. (2017,
reassessing the environment. Historically, environment within the hospital confines. Spring). Fighting radio frequency interference in
it is only when the system starts to This group should have input from medical telemetry. BEAI Spectrum, pp. 35-38.
under preform or is reported as broken IT, Clinical Engineering and Technical
that corrective action would commence. Services as these are the groups most Patricia Brown Lazzara, M. R., Anjana
Traditionally, service contracts would likely to introduce systems that operate Redheendran Santos, M. R., Linda F. Hellstedt,
M. A., & and Rita Walter, M. R.-B. (2010). The
cover the repair of broken parts of the within the RF environment. Naturally this
evolution of a centralized telemetry program.
system, but who identifies and fixes the group should have inputs from both the
Nursing Management, 51-54.
environment when it is “faulty”? clinical and corporate side of the hospital
management, who are the internal Richard Swim. (2013, May/June). The Wireless Challenge
Having Built on their extensive consumers. - Understanding the Wireless Spectrum in a
experience with wireless monitoring Healthcare Facility. Biomedical Instrumentation
& Technology, pp.212 - 215.

CONTACT DETAILS:

Cardiac Services - Dublin


Unit 3, Northern Cross Business Park, Finglas,
Dublin 11, Ireland
T: +353 1 830 7499
F: +353 1 830 7662
E: sales@cardiac-services.com
www.cardiac-services.com

Cardiac Services - Belfast


6 Wildflower Way, Boucher Road,
Belfast, BT12 6TA
T: +44 28 9066 9000
F: +44 (0)28 90687100
E: orders-ni@cardiac-services.com
www.cardiac-services.com

Cardiac Services - Manchester


Europa House, Adlington Business Park,
Adlington, Macclesfield,
SK10 4NL, England
T: +44 1625 878 999
F: +44 1625 878 880
E: orders-uk@cardiac-services.com
www.cardiac-services.com

You might also like