Professional Documents
Culture Documents
20
0
−20
0 200 400 600 800
Frequency (MHz)
(a)
Fig. 2. Basic setup for microwave imaging in a two dimensional case when
one antenna (TX) transmits the signal and all other antennas (RX) are in
600 receiving mode. The imaging domain D contains the heterogeneous body to
be imaged. The measurement domain S contain the transmit and the receive
antennas. The whole system is in a homogeneous medium that acts as a
550
∆σ (%)
matching medium.
40
20
A. Quantitative Imaging
Quantitative imaging methods, also called microwave to-
mography, generates the image based on the values of electri-
cal properties, both the relative permittivity and the conductiv-
ity, of various tissues of the body. The generated image shows
how tissues with different electrical properties are distributed
inside the body. The body to be imaged is surrounded by
a number of transmit and receive antennas. Each transmitter
antenna illuminates the body with the microwave signal at a
time and the scattered signal by the body is collected by all
the receive antennas. Usually this complete setup is immersed
in a homogeneous medium with permittivity εb which acts
as a matching medium for coupling the microwave energy
transmitted to the body, reducing any reflection that may have
occurred at the air-body interface in its absence. The body,
that has to be imaged, lies within a bounded domain D and
the antennas are located on measurement domain S as shown
in Fig. 2. The microwave tomography problem is generally
formulated in terms of electric fields. Three electric fields are
The total electric field Etotal satisfies the domain equation in
D:
∫
Etotal(r) = Einc(r)+k2 G (r, r′)χ(r′)Etotal(r′)dv(r′). (3) b
D
t
where GS and GD are Green function’s operator in S and D , I(r) of a pixel in the image at r is the
Σsquares of the
2
respectively. The minimization of the cost-functional itera- coherently summed values [3]: I(r) = t ctBt(τt(r)) ,
tively constructs sequences of the contrast sources and the cially in the conductivity where the data obtained at one fre-
contrast. At each iteration step, each sequence is updated, quency are used as a starting point for the other frequency [59].
assuming the other one as constant. The cost-function is often
multiplied by a regularization term and the algorithm is called
multiplicative regularized contrast source inversion (MR-
CSI). The MGM is similar to CSI where the fields and the
contrast is updated in each step. Avoiding the forward
problem in each iteration in CSI and MGM leads to a large
number of unknowns. Moreover, they take a larger number of
iterations to converge [57]. However, in spite of these
difficulties, MR- CSI has been shown to be computationally
less demanding than the Newton-type algorithms [55].
The reconstruction algorithms can be applied at a single
frequency or at multiple frequencies in a frequency hopping
manner [58]. Ignoring the frequency dispersion of the tissues,
frequency hopping can be used to remove the artifacts, espe-
where τt(r) = 2|r − rt|/(V ∆T ) is the time delay from
the tth antenna at rt to the synthetic focal point at r in
the body. V is the propagation velocity of the signal
inside the
body, assuming a homogeneous medium and ct are
weights to compensate for the radial spreading of the
cylindrical waves as it propagates outward from the transmit
antenna. Similar to confocal approach, another algorithm
called tissue sensing adaptive radar (TSAR) [63] has also
been developed which assigns more weight to antennas
closest to the focal point. Further, TSAR algorithm includes
compensation term for the distance traveled within the body
and spreading of the waveform.
The time-shifting and the summation of the backscattered
signal done in the confocal imaging may not compensate for
frequency-dependent propagation effects that are associated
with an UWB signal. Moreover, it has limited ability to dis-
criminate against artifacts and noise [62]. Hence, microwave
imaging via space time beamforming has been proposed
[62]
which can spatially focus the backscattered signals to discrim-
inate against clutter and noise while compensating for fre- leading to comfort of patients. In [12], the Doppler radar was
quency dependent propagation effects. To achieve the spatial used at different frequencies, 2.4 GHz, 5.8 GHz, and 10 GHz,
focus, time shifting of the received signal is done to align and different power levels for heartbeat extraction. In [13], a
the backscattered signal from a hypothesized scatterer at a comparison between different preprocessing methods for the
probable location of the strong scatterer. This time aligned demodulation of the received signal using the data collected
signals are then passed through a bank of finite-impulse re- by the Doppler radar for heartbeat and respiration monitoring
sponse (FIR) filters, for each antenna location, whose weights is presented. Fig. 3 shows the phase of the heartbeat signal
are chosen using the least square method. This ensures that the obtained with the Doppler radar with different preprocessing
components of the backscattered signal originating from the technique presented in [13]. The plot of the ECG that corre-
probable strong scatterer location passes the beamformer with lates well with the extracted heartbeat is also shown.
unit gain while compensating for the frequency-dependent 2) Blood Flow and Pressure: Microwave sensing can also
propagation effect [62]. Further, the beamformer output is be used for detection in changes of blood flow and blood
time gated to the time interval which would contain the pressure. For example, in [66], it is shown that microwave
backscattered signal from the probable strong scatterer loca- sensing has potential for detection of 0.3 − 0.5% changes
tion. The beamformer is scanned to different locations inside in extremity blood flow and about 2.5% change in elevated
the body by appropriately changing the time shifts, gating, compartment pressure using the experiments done on a pig’s
and FIR filter coefficients and the energy at each location hind leg. Changes in the amplitude of the microwave signal
energy is calculated [62]. The reconstructed image shows the at 2.5 GHz transmitted through the pig’s thigh on reduction
backscattered signal energy at different locations. in the blood flow is shown in Fig. 4. In [67], an experimental
investigation is done on a phantom and a human subject using
C. Sensing an UWB radar to estimate the radius and changes in the radius
of the aorta that is related to the blood pressure.
One of the most commonly employed methods for mi- 3) Fluid or Water Accumulation: Another sensing appli-
crowave sensing applications, especially for heartbeat and cation is detection of fluid or water accumulation in various
respiration is based on the Doppler radar. In such a method, organs. The detection and monitoring of cerebral edema are
the radar transmits the microwave signal of wavelength λ that presented in [5]. A head phantom was used that consti-
is then reflected by the moving human body parts as the chest- tute of separate compartment of water and ethanol inside a
wall. The reflected signal is received by another antenna. The glass sphere to stimulate the head tissues (water for cerebral
phase of the reflected signal changes according to the Doppler
fluid and ethanol for brain tissues). The microwave signal at
theorem. The change in the phase ∆φ of the received signal
2.4 GHz from one side of the head phantom was transmitted
at time t is related to the displacement of the moving body and received from the other side. The received signal was
x(t) and is given by [13]: ∆φ = 4πx(t)/λ. x(t) contains compared with the reference signal. The amplitude and phase
displacement both due to heartbeat and respiration. Holding of the received signal are related to the variable amounts of
of breath may result in the displacement which is just due to excess water. It was found that 2◦ in phase change was related
the heartbeat. The heartbeat signal can also be extracted by to 10 ml of addition of water and 1◦ phase changes was
use of a low pass filter which can filter out high frequency
related to 10 ml change in the ethanol volume. Similar to this
respiration signal.
principle, water accumulation in the body using an UWB
Other sensing applications may involve sending microwave
radar is presented in [68] where it was found that the reflected
signal from an antenna towards the body and recording the
signal from the bladder with water had different received
reflected or the transmitted signal from another or the same signal than without water.
antenna. Due to changes in the electrical properties of the A microwave stethoscope for monitoring vital signs and
tissues during pathological conditions, the reflected signal will changes in lung water content is presented in [69]. A coplanar
be different from what it would be for a normal healthy case.
waveguide operating at 915 MHz and 920 MHz was used to
Based on this, pathological conditions can be detected. For
measure the reflection coefficient on a phantom as well as on
such a purpose, usually ultra-wideband (UWB) radar are used
the humans. From the measured reflection coefficient, various
as they can penetrate the body and also have a good temporal
vital signs like respiration rate, heart rate and changes in the
resolution [65].
lung water content was successfully extracted using signal
IV. APPLICATIONS
phase of
This section gives the overview of the different medical heart-beat signal
applications where MSI has been studied. (diff. est. methods)
ECG
A. Medical Sensing Applications 35 40 45
Time (s)
1) Heartbeat Detection: Heartbeat extraction using the
Doppler radar is a common sensing application. The
advantage of using the Doppler radar is that it is a non- Fig. 3. Phase of the received signal for heartbeat after filtering out the
contact method respiration signal [13].
TABLE I
SUMMARY OF SENSING AND IMAGING METHODS
Method Theoretical Basis Techniques/Algorithm [ref.] Positive Aspects Negative Aspects Frequency Remarks
Quantitative Iteratively solving Newton Methods: Less number of itera- Computationally Single frequency, 2-D or 3-D image with di-
(Section III-A) inverse EM scat- Newton-Kantorovich [44] tions required to con- demanding, multiple electric properties. Newton
tering problem. Gauss-Newton [52] verge, Limited 3-D ap- frequency methods are similar if
Levenberg-Marquadt [49] Lower number of un- plicability or wideband same regularization
Distorted Born [46] knowns. technique is applied.
Log-Magnitude and Phase Recon-
struction [51]
Contrast Source Inversion [55] Computationally less Larger number of
Modified Gradient Method [53] demanding unknowns, more
(than Newton Meth- iterations required
ods), 3-D applicabil- for convergence.
ity
Qualitative Strong scatterer Confocal Imaging [3] Avoids inverse prob- Detection UWB Image of backscattered en-
(Section III-B) detection using Tissue Sensing Adaptive Radar lem, computationally challenging ergy
radar techniques [63] Space-Time Beamforming very less demanding in low contrast
[62]
Sensing Change in ampli- Doppler Radar [13] Simple setup Motion artifacts, Single frequency No image but 2-D plot
(Section III-C) tude and/or phase Transmission/Reflection [6] reference needed or UWB
depth (cm)
span (cm)
(a)
depth (cm)
Fig. 6. A measurement setup for brain imaging using a physical head
phantom and the Vivaldi antenna [72].
(a) (b)
Fig. 10. Reconstructed images of the left leg and the right leg of a
patient affected by injury in the left leg through a Newton method with soft-
x (cm) x (cm) regularization (a) relative permittivity (b) conductivity. Left side is for the left
leg and right side images are for the right leg [16].
(a) (b)
D. Bone Imaging
Bone Imaging using microwave has been done for appli-
cations like detecting the leukemia in the bone marrow [15],
and to determine the bone density for detection of osteoporo-
sis [16]. Leukemia causes the cellular population in the bone
to increase that in turn increases the relative permittivity and
was done to acquire 2-D and 3-D images of the calcaneus
bone of two patients. Their one leg was immobilized for at
least six weeks during recovery from a lower leg injury. The
reconstructed images are shown in Fig. 10. A good
correlation was observed between the recovered electrical
properties of the injured calcaneus bone by microwave
imaging and the CT Hounsfield density measure. Hence,
microwave imaging of the bone can also help in determining
the bone density for detection of osteoporosis [16].
y (mm)
the change in temperature primarily due to changes in the
electrical properties of the water with temperature. The
change in electrical properties of the animal liver tissue is
reported with temperature in [98]. For example, it was
x (mm) observed that the properties change approximately 1% per ◦C
between 28◦C and 53◦C at 915 MHz. The idea of microwave
Fig. 11. Differential image when two tear are present in meniscus. The
differential image is obtained by subtracting images obtained by the TSAR imaging for remote thermal sensing was used in an early
imaging algorithm, with and without the tear. The meniscus on which work [99]. This has been reconsidered in recent works
measurements were done is shown in inset with the positions of the antennas [100]-[102]. In [101], tomographic images over the
marked with ’*’ [23].
frequency range of
300 − 1000 MHz of the pig abdomen were generated where
F. Heart Imaging the recovered conductivity was found to vary linearly with the
controlled temperature values. In [102], a real-time
Heart imaging is done to detect any pathological conditions
microwave imaging system at 915 MHz to image differential
as myocardial infarction [19] as the electrical properties of
temperature based on the change in the electrical properties of
myocardial tissue have a strong dependence on coronary
water with the goal of achieving noninvasive temperature
blood flow. It can also be used to obtain the temporal
monitoring of thermal therapy systems was developed, and
images of a beating heart for heartbeat extraction [22] due to
tested on a simple breast phantom. Hence, exploiting
millisecond temporal resolution. However, the heart is
temperature sensitivity of the tissues, MSI can also be used for
located inside an anatomically complex and highly
noninvasive thermal therapy monitoring.
electromagnetically scatter- ing structure of the thorax, that
make microwave imaging of the heart a challenging task
[2]. Moreover, there might be strong reflections from the H. Miscellaneous Biomedical Body Imaging
air-filled lungs that might make the detection of potentially There are many other studies which have been done for
weak field scattered by the heart challenging [20]. Hence, imaging the biomedical bodies that does not fit into any afore-
heart imaging requires minimum SNR of 30 dB and a mentioned applications. However, these shows the potential
dynamic range of 150 dB or more [20]. In [17], Semenov et fields where MSI can be utilized. Some of them are briefly
al., reconstructed the quantitative im- ages of an excised summarized here.
canine perfused heart in a static as well as in a beating The potential to image the human arm is shown using
case using a modified Rytov approximations at 2.45 GHz. the CSI method in [103]. In [104], the Newton-Kantorovich
In [18], authors were able to successfully reconstruct the algorithm is used to obtain a quantitative permittivity image of
internal structure of an excised canine heart including the a numerical human thorax and a real human arm at 434 MHz.
left and the right ventricles using the 3-D MGM and in [19] In [105], the author of this paper used the Levernerg-
for the detection of the myocardial infarction. In [20], Marquadt method presented in [49] for imaging the human
using the same 3-D gradient method, authors were able to torso at the level of small intestine in the 403.5 MHz
reconstruct the heart in an intact swine at 0.9 GHz. Further, MedRadio band. The reconstructed image was then used as a
measurements were done with and without the heart basis for the lo- calization of wireless capsule endoscopes.
in the thorax, and it was found that the presence of the heart Using this method, a localization accuracy within a centimeter
changes the amplitude of EM field by 10 − 20% and a change was obtained. The possibility to image the whole body using
of 7 − 10 degrees in the phase was observed. microwave imaging is shown in [106] where measurements
An UWB radar-based approach to image the human heart in were done on a dog and a 3-D gradient based iterative
real time is presented in [21], [22]. An UWB switched array algorithm was used to invert the data. In [45], comparisons
body-contact radar was used and a back-projection imaging between different quantitative image reconstruction methods
algorithm was used. The frequency utilized was 0.8 − 3 GHz. are done on the experimental data of a phantom and a pig’s
The purpose of imaging was to obtain the heartbeat rate. hind leg in the frequency range
Successive images were reconstructed of the beating heart 0.9−2.05 GHz. These methods are Newton based 2-D method
of a male subject with a frame rate of 25 Hz. Using the and 3-D gradient method, and MR-CSI both for 2-D and 3-D
reconstructed image, heartbeat waveforms were extracted. It cases. These research shows the feasibility of imaging large
was concluded that using this approach, different moving biological bodies using quantitative approaches and opens up
parts of the heart can be observed and may provide the basis opportunities in terms of several new application areas where
for an online diagnosis. MSI can be used.
that if the experimental data were collected on human/animal signal to the body. Due to a high difference in the electrical
and whether clinical trials have been performed. Clinical trials properties of the medium in which the antenna is placed
are initial testing of the MSI devices where prior approval
has to be taken from the health and ethical committee of the
country concerned. Moreover, many subjects either patients
or healthy volunteers participate after giving their written
consent.
B. Contrast Agents
A significant difference in the electrical properties of the
malignant tissue and the healthy tissues of the same kind is
used as a basis for detection of such malignant tissues. How-
ever, if any healthy tissue having a very small difference in
the electrical properties with this malignant tissue is present
in a close proximity, microwave imaging may fail to
distinguish between these two kinds of tissues. As for
example, there exist a large contrast between the breast
cancerous and the healthy fatty tissues. This difference is
small when the cancerous tissue is compared with the
healthy glandular or the fibro-connective breast tissue [36].
Hence, the detection of the cancerous tissue
C. Imaging Algorithms
As discussed before, the quantitative microwave imaging
algorithms are known to have inherit challenges of an in-
verse problem like the non-linearity and the ill-posedness.
Though several methods have been proposed to attack the
problem [45], they still suffer from several limitations. For
example, oscillations occur in the estimated electrical prop-
erties at the boundary of the high dielectric contrast tissues
(a) (b) with both Newton and MR-CSI methods though employing
different regularization methods [45]. It was suggested that
Fig. 12. Effect of varying concentration of contrast agent, single-walled an adaptive-density grid and regularization schemes for both
carbon nantotube (SWCNT), on tissue-mimicking material (a) relative per-
mittivity (b) conductivity, measured from 0.6 GHz to 20 GHz [110]. the EM field and the electrical properties might overcome
this problem [45]. Moreover, a 2-D based imaging algorithm
may not work for a 3-D biological object, and 3-D imag-
ing algorithms are computationally demanding. Moreover, as
listed in Table I, each of the imaging algorithms has its
in a close vicinity of the glandular tissue may be a challenging own limitations. Selecting a particular algorithm for a given
task with microwave imaging. One of the methods proposed application, depends upon available computational resources.
to tackle this challenge is to use contrast agents that can Hence, further in-depth studies are required for development
enhance the contrast of the cancerous tissue [108]-[111]. The of a computationally efficient and more accurate inversion
idea is to administer a contrast agent to the body by methods algorithm. Adoption of the algorithm for parallel computing
like intravenous injection. Some volume of the contrast agent can reduce the time complexity to some extent [45].
will then reach and bind the cancerous tissues enhancing
their electrical properties. In [108], a computational study
using 3-D realistic numerical breast phantoms is presented D. Signal Processing
using assumed effects of two contrast agents, microbubles and For sensing applications as the heartbeat detection, removal
carbon nanotubes. Differential imaging was done by taking of the motion and respiration artifacts of a patient, and the
the difference between the image with and without the background clutter remains an open challenge to effectively
contrast agent effect and was concluded that small tumors, extract the heartbeat from the reflected signal [13]. Hence, a
even below the resolution limit of the tomography system, can robust signal processing algorithm is required for such cases
be detected. In [109], a feasibility study is done using to remove such artifacts from the received signal. Moreover,
microbubbles and single-walled carbon nanotubes (SWCNT) qualitative imaging approaches use backscattered data for
as contrast agents in an UWB breast imaging to classify a breast tumor or brain stroke detection. In such cases, signal
lesion as malignant or benign. This is done by analyzing the processing is required to remove the skin backscatter and late
complex natural resonances of the differential backscatter time clutter response to effectively focus the tumor or the
responses before and after application of the contrast agents at stroke. Towards this end, some signal processing algorithms
the suspicious site. Similarly, [110] uses carbon nanotube for have been developed to tackle skin reflections [64] and clutter
contrast enhancing of the breast tumors. It was shown that response [80].
0.22% of SWCNTs concentration by weight can increase the
average relative per- mittivity by 37% and the conductivity by
E. Antennas and Measurement System
81% of tissue mim- icking material. The measured
enhancement in the contrast of the tissue mimicking material To relax the non-uniqueness and the ill-posedness of
is shown in Fig. 12 [110]. Use of magnetic nanoparticles as nonlin- ear inverse problems to some extent, large number of
the contrast agent is discussed in [111] for microwave antennas have to be used so that the scattered field data set
breast cancer imaging. The method is based on from which the electrical properties are retrieved is large [49].
reconstruction of the magnetic contrast induced by the However, the number of antennas that can be used is limited
magnetic nanoparticles using the differential scattered signal by their finite size. Moreover, placing the antennas too close
obtained at the receive antenna in the presence and the to each other may result in high mutual coupling between the
absence of the polarizing magnetic field. These research have antennas introducing error in the measured scattered field.
shown that the challenge of small contrast between the Other errors that might get introduced in a measurement setup
cancerous tissue and any healthy tissue can be addressed, are cable losses, phase shifts, or mismatch at the connectors
but there still remain several open challenges that need to [112]. Further, it is the electric field that is required in the
be addressed. For example, increasing the concentration level inversion, and using a vector network analyzer based
of the contrast agents within the pathological tissues [26] as measurement system would measure the path-gain between
only a low volume of contrast agent can reach the pathological the transmit and the receive antennas. Hence, calibration of
tissues [111]. The contrast agent may also distort the lesion the measured data is required. A calibration process to remove
border profile that may be used in classification of lesions as some of these errors is suggested in [112], [113]. Further,
benign and malignant [109]. Hence, development of effective [112] provides a method to avoid the frequencies where the
contrast agents can be a possible future direction. coupling is large enough to
prevent successful imaging in a wideband measurement setup.
Another method to avoid coupling is to use a virtual array to lightning) may also contribute to interference and noise in a
scan the body, but at a cost of increased scan-time. Therefore, medical facility [118]. Thus, development of MSI system that
the design of a measurement setup having an optimum scan- is robust to these EM interference is required. Furthermore,
time, low error, and also a better antenna design having a the MSI system must not be a source of interference to other
minimal mutual coupling can be a future research direction. wireless systems. It should follow the appropriate regulations
Moreover, further research on effective calibration methods of and standards set for medical devices.
the measured data is also needed.
The error may also occur in the EM system model for the H. Commercial Challenge
nonlinear optimization problem due to the presence of nonac-
Apart from aforementioned technical challenges, one of
tive antennas if a fixed antenna array is used for microwave
the non-technical challenge that would be faced by the MSI
imaging. Typically, the perturbations caused in the scattering
systems to be embraced by clinician is competition from
field due to the presence of these non-active antennas are
other well established imaging modalities like MRI, CT or
not taken into account in the EM system model. Moreover,
ultrasound. As mentioned earlier, though the MSI system
an error may also get introduced due to the assumption of
has advantages such as low risk, mobility, time resolution,
infinite matching medium. Hence, calibration methods as well
and cost-effectiveness, it lacks especially in spatial resolution
as a proper measurement system which introduces less error
compared to CT or MRI, that may be desired for a clear in-
are required. A compensation method for nonactive antennas
terpretation of the reconstructed images in some applications.
is discussed in [114]. A formulation that provides a way to
However, in spite of these limitations, there have been start-
compensate the perturbations resulting from the presence of
up companies based on MSI technology as EMTensor [8]
an array of antennas around the imaging body is proposed
and Medfield Diagnostics [119] for stroke diagnosis, and
in [115]. Mojabi et al. in [116] proposes a measurement setup
Micrima [120] for breast imaging, making a mark in the
using a rotatable conductive enclosure using a minimal
medical imaging industry that has been so far dominated by
antenna array. It is discussed that such a setup will reduce
MRI or CT modalities. The success of such companies based
errors due to the assumption of infinite matching medium as
on MSI technology would certainly depend upon how well
the boundary conditions that can be easily modeled.
MSI overcomes the technical as well as commercial
Moreover, lesser number of the antennas used in the setup will
challenges from other well established modalities.
further reduce any error because of the presence of many
nonactive antennas.
I. Other Future Directions
F. Frequency Band and Resolution Some of the other future research work will naturally
Another open issue in MSI is choosing a suitable imaging follow the ways and means to improve the signal excitation,
frequency. As discussed in Section IV, different applications detection, and the reconstruction techniques to obtain high
or even for the same application, different frequencies have quality, low noise signals and images. In addition, novel
been used. A choice of suitable frequency band is a tradeoff approaches like using multiple-input-multiple-output (MIMO)
between the penetration depth and the resolution. It is well techniques can be considered for reducing the complexity of
known that the penetration depth decreases with the imaging systems [121] or improving the image by increasing
frequency due to increased attenuation in the tissues, the signal to clutter ratio when compared with a bistatic or a
however, the resolution increases with the frequency. monostatic configuration [122]. Moreover, 3-D reconstruction
Moreover, a frequency used for imaging one body part may with temporal information in real time will be useful for new
not be optimal for another body part due to the difference in applications like obtaining temporal images of a beating heart.
the size and the tissue compositions. Hence, further Another direction that can be explored for MSI is com-
investigations are required for developing a standard with an pressive sensing. Compressive sensing is a method by which
optimum frequency and an acceptable resolution. signals can be reconstructed by sampling them at a rate
much lower than the Nyquist rate. However, for such a
reconstruction, following criteria have to be met: “sparse”
G. Electromagnetic Interference and Noise
representation of the unknown signal in some domain, and
A clinical environment is an environment where EM inter- “incoherence” of the signal used for measurement with respect
ference abounds. Hence, using an MSI system in a clinical to the unknown signal. Compressed sensing has been shown
environment is a challenging task as it will be susceptible to to substantially improve the performance of ranging with an
EM interference and noise resulting in error in the measured UWB medical radar in comparison with more conventional
data. The sources of such interference and noise in a clinical methods in scenarios with low signal-to-noise-ratio [123].
environment are RF emissions from wireless devices like
mobile phones, medical telemetry, wireless local area network
VI. CONCLUDING REMARKS
and other devices as radiology equipments, electrocautery
equipments, fluorescent lights and computers [117]. Further, Microwave imaging and sensing (MSI) opens new oppor-
noisy electrical power supplies and grounding (earth), mag- tunities in the field of medical imaging. MSI is shown to be
netic fields (static and alternating), and surges (static low-risk due to the non-ionizing EM signal used, low-cost,
discharge, portable, and have a good temporal resolution. This paper
presented an exhaustive summary of various quantitative and
qualitative imaging methods for medical MSI. It was [7] T. Henriksson et al., “Clinical trials of a multistatic UWB radar
for breast imaging,” in Proc. Loughborough Antennas Propag. Conf.,
discussed that the physical basis of these quantitative and Loughborough, Nov. 2011.
qualitative imaging methods is the difference in the electrical [8] EMTensor, TechGate Tower, Donau-City-Strasse 1, 1220 Vienna, Aus-
properties of different tissues. Quantitative methods solve the tria. Website: http://www.emtensor.com/
[9] L. E. Larsen, and J. H. Jacobi, “Medical Application of Microwave
ill-posed inverse EM problem through different iterative Imaging,” New York, IEEE Press, 1986.
schemes to minimize the error between the measured and the [10] C. Pichot et al., “Active microwave imaging of inhomogeneous bodies,”
modeled scattered field to obtain the electrical properties. IEEE Trans. Antennas Propag., vol. 33, no. 4, pp. 416-425, Apr. 1985.
[11] A. Broquetas et al., “Cylindrical geometry: a further step in active
Qualitative methods avoid solving ill-posed inverse problem microwave tomography,” IEEE Trans. Microw. Theory Techn., vol. 39,
by using radar-based techniques, and can be used in cases no. 5, pp. 836-844, May 1991.
where the objective is the detection of the strong scatterer like [12] D. Obeid et al., “Multitunable Microwave System for Touchless Heart-
beat Detection and Heart Rate Variability Extraction,” Microw. Optical
the tumor. Various medical imaging applications using Tech. Lett., vol. 52, no. 1, Jan. 2010.
microwaves for which proof-of-concept have been shown [13] M. Zakrzewski et al., “Comparison of Center Estimation Algorithms
were discussed. These applications include heartbeat, brain for Heart and Respiration Monitoring with Microwave Doppler Radar,”
imaging for stroke and edema detection, bone imaging for IEEE Sensors J., vol. 12, no. 3, pp. 627-634, Mar. 2012.
[14] N.K. Nikolova, “Microwave Imaging for Breast Cancer,” IEEE Microw.
density measures, breast cancer detection, heart imaging, and Mag., vol. 12, no. 7, pp. 78-94, Dec. 2011.
soft tissue imaging. Apart from these applications, MSI has [15] D. Colton, and P. Monk, “The detection and monitoring of leukemia
shown to work for anatomical imaging of different body parts using electromagnetic waves: numerical analysis,” Inverse Prob., vol.
11, no. 2, 1995.
like arms, torso, and also full-body. The feasibility studies of [16] P. M. Meaney et al., “Clinical Microwave Tomographic Imaging of the
MSI for these various applications shows the potential MSI Calcaneus: A First-in-Human Case Study of Two Subjects,” IEEE
has to be used as a low-risk and a low cost alternative to the Trans. Biomed. Eng., vol. 59, no. 12, pp. 3304-3313, Dec. 2012.
[17] S. Y. Semenov et al., “Microwave tomography: two-dimensional system
currently used imaging modalities like MRI and CT-scan. for biological imaging,” IEEE Trans. Biomed. Eng., vol. 43, no. 9, pp.
Successful initial clinical trials, especially for breast cancer 869-877, Sept. 1996.
imaging and stroke detection using MSI have been done. [18] S. Y. Semenov, “Three-dimensional microwave tomography:
experimen- tal imaging of phantoms and biological objects,” IEEE
Commercial start-up companies using MSI technology like Trans. Microw. Theory Tech., vol. 48, no. 6, pp. 1071-1074, Jun. 2000.
Micrima [120] (for breast cancer), EMtensor [8] and Medfield [19] S. Y. Semenov et al., “Microwave tomography for detection/imaging of
Diagnostics (for brain stroke) are involved in clinical studies. myocardial infarction. I. Excised canine hearts,” Ann. Biomed. Eng.,
vol. 31, no. 3, pp. 262-270, Mar. 2003.
However, for many other applications as listed in Table II, [20] S. Y. Semenov et al, “Microwave Tomographic Imaging of the Heart
MSI is still far from maturity as many open challenges remain in Intact Swine,” J. Electromag. Waves Applications, vol. 20, no. 7, pp.
that have to be addressed before MSI can be used in a 873-890, 2006.
[21] S. Brovoll et al., “Time-lapse imaging of human heartbeats using UWB
clinical setting. The paper discussed these open topics and radar,” IEEE Biomed. Circuits Sys. Conf. (BioCAS), pp. 142-145, Oct.
gave some insight on what could be the probable solutions and 2013.
future research directions for MSI. [22] S. Brovoll et al., “Time-Lapse Imaging of Human Heart Motion With
Switched Array UWB Radar,” IEEE Trans. Biomed. Circuits Sys., vol.
8, no. 5, pp. 704-715, Oct. 2014.
[23] S. M. Salvador et al., “Exploring Joint Tissues with Microwave Imag-
ing,” IEEE Trans. Microw. Theory Tech., vol. 58, no. 8, pp. 2307-2313,
REFERENCES Aug. 2010.
[24] P. M. Meaney et al. “Initial Clinical Experience with Microwave Breast
[1] L. Fass, “Imaging and cancer: a review,” Mol. Oncol., vol 2, no. 2, pp. Imaging in Women with Normal Mammography,” Acad. Radiol., vol.
115-152, Aug. 2008. 14, no. 2, pp. 207-218., Feb. 2007.
[2] S. Semenov, “Microwave tomography: review of the progress towards [25] E. C. Fear et al., “Microwave Breast Imaging With a Monostatic Radar-
clinical applications,” Philos. Trans. A Math. Phys. Eng. Sci., pp. 3021- Based System: A Study of Application to Patients,” IEEE Trans.
3042, Aug. 2009. Microw. Theory Tech., vol. 61, no. 5, pp. 2119-2128, May 2013.
[3] E.C. Fear et al., “Confocal microwave imaging for breast cancer detec- [26] S.C. Hagness et al., “Guest Editorial: Special Cluster on Microwave
tion: localization of tumors in three dimensions,” IEEE Trans. Biomed. Medical Imaging,” IEEE Antennas Wireless Propag. Lett. , vol. 11, pp.
Eng., vol. 49, no. 8, pp. 812-822, Aug. 2002. 1592-1597, 2012.
[4] S. Y. Semenov, and D. R. Corfield, “Microwave Tomography for Brain [27] H. Ohba et al., “Recent trends in medical microwave radiometry,” IEIC
Imaging: Feasibility Assessment for Stroke Detection,” Int. J. Antennas Trans. Commun. Eng., vol. E78-B, no. 6, pp. 789798, 1995.
Propag., vol. 2008, Article ID 254830, 2008. [28] K. R. Foster, and H. P. Schwan, “Dielectric properties of tissues and
[5] J. C. Lin, and M. J. Clarke, “Microwave imaging of cerebral edema,” biological materials: a critical review,” CRC Crit. Rev. Biomed. Eng.,
Proc. IEEE, vol. 70, no. 5, pp. 523-524, May 1982. vol. 17, no. 1, pp. 25104, 1989.
[6] M. Persson et al., “Microwave-Based Stroke Diagnosis Making Global [29] C. Gabriel et al., “The dielectric properties of biological tissues. I.
Prehospital Thrombolytic Treatment Possible,” IEEE Trans. Biomed. Literature survey,” Phys. Med. Biol., vol. 41, no. 11, pp. 2231-2249,
Eng., vol. 61, no. 11, pp. 2806-2817, Nov. 2014. 1996.
[30] W. T. Joines et al., “The measured electrical properties of normal and
malignant human tissues from 50 to 900 MHz,” Med. Phys., vol. 21,
no. 4, pp. 547-550, Apr. 1994.
[31] S. Gabriel et al., “The dielectric properties of biological tissues. III.
Parametric models for the dielectric spectrum of tissues,” Phys. Med.
Biol., vol. 41, no. 11, pp. 2271-2293, 1996.
[32] J. L. Schepps, and K. R. Foster, “The UHF and microwave dielectric
properties of normal and tumor tissues: variation in dielectric properties
with tissue water content,” Phys. Med. Biol., vol. 25, no. 6, pp. 1149-
1159, Nov. 1980.
[33] L. Sha et al., “A review of dielectric properties of normal and malignant
breast tissue,” in Proc. IEEE SoutheastCon., pp. 457-462, 2002.
[34] R. Pethig, “Dielectric Properties of Biological Materials: Biophysical
and Medical Applications,” IEEE Trans. Elec. Insul., vol. EI-19, no. 5, [57] A. E. Bulyshev, “Three-dimensional vector microwave tomography:
pp. 453-474, Oct. 1984. theory and computational experiments,” Inverse Problems, no. 20, pp.
1239-1259, 2004.
[35] M. Lazebnik et al., ”Highly Accurate Debye Models for Normal and
[58] W.C. Chew, and J.-H. Lin, “A frequency-hopping approach for mi-
Malignant Breast Tissue Dielectric Properties at Microwave Frequen-
crowave imaging of large inhomogeneous bodies,” IEEE Trans.
cies,” IEEE Microw. Wireless Comp. Lett., vol. 17, no. 12, pp. 822-824,
Microw. Guid. Wave Lett., vol. 5, no. 12, pp. 439-441, Dec. 1995.
Dec. 2007.
[59] M. Persson et al., “Microwave imaging for medical applications,” in
[36] M. Lazebnik et al., “A large-scale study of the ultrawideband Proc. 5th Eur. Conf. Antennas Propag. (EUCAP), pp. 3070-3072, April
microwave dielectric properties of normal, benign and malignant breast 2011.
tissues obtained from cancer surgeries,” Phys. Med. Biol., vol. 52, no. [60] A. Fhager et al., “Image Reconstruction in Microwave Tomography
20, pp. 6093-6115, 2007. Using a Dielectric Debye Model,” IEEE Trans. Biomed. Eng., vol. 59,
[37] G. Bindu, and K. T. Mathew, “Characterization of benign and malignant no. 1, pp. 156-166, Jan. 2012.
breast tissues using 2-D microwave tomographic imaging,” Microw. [61] A. Fhager et al., “Reconstruction Quality and Spectral Content of
Opt. Technol. Lett., vol. 49, pp. 2341-2345, Mar. 2007. an Electromagnetic Time-Domain Inversion Algorithm,” IEEE Trans.
[38] R. J. Halter et al., “The correlation of in vivo and ex vivo tissue Biomed. Eng., vol. 53, no. 8, pp. 1594-1604, Aug. 2006.
dielectric properties to validate electromagnetic breast imaging: initial [62] J. E. Bond et al., “Microwave imaging via space-time beamforming for
clinical experience,” Physiol. Meas., vol. 30, no. 6, pp. S121-S136, Jun. early detection of breast cancer,” IEEE Trans. Antennas Propag., vol.
2009. 51, no. 8, pp. 1690-1705, Aug. 2003.
[39] A.P. O’Rourke et al., “Dielectric properties of human normal, malignant [63] J. M. Sill, and E. C. Fear, “Tissue sensing adaptive radar for breast
and cirrhotic liver tissue: in vivo and ex vivo measurements from 0.5 to cancer detection - experimental investigation of simple tumor models,”
20 GHz using a precision open-ended coaxial probe,” Phys. Med. Biol., IEEE Trans. Microw. Theory Tech., vol. 53, no. 11, pp. 3312-3319,
vol. 52, no. 15, pp. 4704-4719, Aug. 2007. Nov. 2005.
[40] A. M. Deighton, “Differentiating between healthy and malignant lymph [64] S. Mustafa et al., “Novel Preprocessing Techniques for Accurate Mi-
nodes at microwave frequencies,” J. Undergrad. Reasearch Alberta, crowave Imaging of Human Brain,” IEEE Antennas Wireless Propag.
vol. 3, no. 1, 2013. Lett., vol. 12, pp. 460-463, 2013.
[41] T. Sunaga, “Measurement of the electrical properties of human skin and [65] R. M. Narayanan, “Technical considerations in medical radar,” in Proc.
the variation among subjects with certain skin conditions,” Phys. Med. 8th Int. Conf. Body Area Net. (BodyNets), pp. 526-535, 2013.
Biol., vol. 47, no. 1, pp. N11-N15, Jan. 2002. [66] S. Semenov, “Microwave tomography for functional imaging of extrem-
[42] P. M. Meaney et al., “Bone Dielectric Property Variation as a Function ity soft tissues: feasibility assessment,” Phys. Med. Biol., 52, pp. 5705-
of Mineralization at Microwave Frequencies,” Int. J. Biomed. Imag., 5719, 2007.
vol. 2012, Article ID 649612, 9 pages, 2012. [67] L .E. Solberg et al., “Experimental investigation into radar-based central
[43] S. Y. Semenov et al., “Microwave spectroscopy of myocardial ischemia blood pressure estimation,” IET Radar, Sonar and Navigation, vol. 9,
and infarction. 1. Experimental study,” Ann. Biomed. Eng., vol. 28, pp. no. 2, pp. 145-153, Feb. 2015.
48-54, Jan. 2000. [68] X. Li et al., “Ultra wideband radar for water detection in the human
[44] N. Joachimowicz et al., “Inverse scattering: an iterative numerical body,” in Proc. of German Microw. Conf., pp. 150-153, Mar. 2010.
method for electromagnetic imaging,” IEEE Trans. Antennas Propag., [69] N. Celik et al., “Microwave stethoscope: development and
vol. 39, no. 12, pp. 1742-1753, Dec. 1991. benchmarking of a vital signs sensor using computer-controlled
[45] S. Y. Semenov et al., “Microwave-tomographic imaging of the high phantoms and human studies,” IEEE Trans. Biomed. Eng., vol. 8, no. 8,
dielectric-contrast objects using different image-reconstruction ap- pp. 2341-2349, Aug. 2014.
proaches,” IEEE Trans. Microw. Theory Tech., vol. 53, no. 7, pp. 2284- [70] G. Marrocco, “Pervasive electromagnetics: sensing paradigms by
2294, July 2005. passive RFID technology,” IEEE Wireless Commun. Mag., vol. 17, no.
6, pp. 10-17, Dec. 2010.
[46] W. C. Chew, and Y. M. Wang, “Reconstruction of two-dimensional
[71] D. Ireland et al., “Microwave imaging for brain stroke detection using
permittivity distribution using the distorted Born iterative method,”
Born iterative method,” IET Microw. Antennas Propag., vol. 7, no. 11,
IEEE Trans. Med. Imag., vol. 9, no. 2, pp. 218-225, Jun. 1990.
pp. 909-915, Aug. 2013.
[47] A. Franchois, and A. G. Tijhuis, “A quasi-Newton reconstruction algo- [72] B. J. Mohammed et al., “Microwave System for Head Imaging,” IEEE
rithm for a complex microwave imaging scanner environment,” Radio Trans. Instrum. Meas., vol. 63, no. 1, pp. 117-123, Jan. 2014.
Sci., vol. 38, no. 2, pp. VIC 12-1- VIC 12-13, April 2003. [73] M. Jalilvand et al., “Hemorrhagic stroke detection via UWB medical
[48] A. E. Souvorov et al., “Microwave tomography: a two-dimensional imaging,” in Proc. of 5th Eur. Conf. Antennas Propag., pp. 2911-2914,
Newton iterative scheme,” IEEE Trans. Microw. Theory Tech., vol. 46, April 2011.
no. 11, pp. 1654-1659, Nov. 1998. [74] R. Scapaticci et al., “A feasibility study on microwave imaging for brain
[49] A. Franchois, and C. Pichot, “Microwave imaging-complex permittiv- stroke monitoring,” Progress Electromag. Res. B, vol. 40, pp. 305-324,
ity reconstruction with a Levenberg-Marquardt method,” IEEE Trans. 2012.
Antennas Propag., vol. 45, no. 2, pp. 203-215, Feb. 1997. [75] P. Mojabi, and J. LoVetri, “Microwave Biomedical Imaging Using the
[50] J. De Zaeytijd et al., “Full-Wave Three-Dimensional Microwave Multiplicative Regularized Gauss-Newton Inversion,” IEEE Antennas
Imaging with a Regularized Gauss-Newton Method-Theory and Propag. Lett., vol. 8, pp. 645-648, 2009.
Experiment,” IEEE Trans. Antennas Propag., vol. 55, no. 11, pp. 3279- [76] S. C. Hagness et al., “Two-dimensional FDTD analysis of a pulsed
3292, Nov. 2007. microwave confocal system for breast cancer detection: fixed-focus and
[51] P. M. Meaney et al., “Microwave image reconstruction utilizing log- antenna-array sensors,” IEEE Trans. Biomed. Eng., vol. 45, no. 12, pp.
magnitude and unwrapped phase to improve high-contrast object recov- 1470-1479, Dec. 1998.
ery,” IEEE Trans, Med. Imag., vol. 20, no. 2, pp. 104-116, Feb. 2001. [77] S. C. Hagness et al., “Three-dimensional FDTD analysis of a pulsed
[52] T. Rubæk et al., “Nonlinear Microwave Imaging for Breast-Cancer microwave confocal system for breast cancer detection: design of an
Screening Using GaussNewton’s Method and the CGLS Inversion Al- antenna-array element,” IEEE Trans. Antennas Propag., vol. 47, no. 5,
gorithm,” IEEE Trans. Antennas Propag., vol. 55, no. 8, pp. 2320-2331, pp. 783-791, May 1999.
Aug. 2007. [78] Xu Li, and S. C. Hagness, “A confocal microwave imaging algorithm
[53] R. E. Kleinman, and P. M. van den Berg, “A modified gradient method for breast cancer detection,” IEEE Microw. Wireless Compon. Lett., vol.
for two-dimensional problems in tomography, J. Comput. Appl. Math., 11, no.3, pp. 130-132, March 2001.
vol. 42, pp. 17-35, Jan. 1992. [79] Hooi Been Lim et al., “Confocal Microwave Imaging for Breast Cancer
[54] A. E. Bulyshev, “Three-dimensional microwave tomography: Theory Detection: Delay-Multiply-and-Sum Image Reconstruction Algorithm,”
and computer experiments in scalar approximation,” Inverse Problems, IEEE Trans. Biomed. Eng., vol. 55, no. 6, pp. 1697-1704, June 2008.
vol. 16, pp. 863-875, 2000. [80] D. J. Kurrant et al., “Tumor Response Estimation in Radar-Based
[55] A. Abubakar et al., “Imaging of biomedical data using a multiplicative Microwave Breast Cancer Detection,” IEEE Trans. Biomed. Eng., vol.
regularized contrast source inversion method,” IEEE Trans. Microw. 55, no. 12, pp. 2801-2811, Dec. 2008.
Theory Tech., vol. 50, no. 7, pp. 1761-1771, Jul. 2002. [81] Xu Li et al., “Microwave imaging via space-time beamforming: experi-
[56] P. Mojabi, “Investigation and Development of Algorithms and Tech- mental investigation of tumor detection in multilayer breast phantoms,”
niques for Microwave Tomography,” Ph.D. Dissertation, Dept. Elect. IEEE Trans. Microw. Theory Tech., vol. 52, no.8, pp. 1856-1865, Aug.
Comp. Eng., Univ. Manitoba, Winnipeg, Manitoba, Canada, 2010. 2004.
[82] Xie Yao et al., “Multistatic Adaptive Microwave Imaging for Early
Breast Cancer Detection,” IEEE Trans. Biomed. Eng., vol. 53, no. 8, [108] J. D. Shea et al., “Contrast-enhanced microwave imaging of breast
pp. 1647-1657, Aug. 2006. tumors: a computational study using 3D realistic numerical phantoms,”
[83] S. K. Davis et. al, “Ultrawideband microwave breast cancer detection: a Inverse Problems, vol. 26, No. 7, 2010.
detection-theoretic approach using the generalized likelihood ratio test,” [109] Chen Yifan et al., “Feasibility Study of Lesion Classification via
IEEE Trans. Biomed. Eng., vol. 52, no. 7, pp. 1237-1250, July 2005. Contrast-Agent-Aided UWB Breast Imaging,” IEEE Trans. Biomed.
[84] A. Santorelli et al., “Investigation of Classifiers for Tumor Detection Eng., vol. 57, no. 5, pp. 1003-1007, May 2010.
with an Experimental Time-Domain Breast Screening System,” [110] A. Mashal et al., “Toward Carbon-Nanotube-Based Theranostic
Progress Electromag. Research,, vol. 144, pp. 45-57, 2014. Agents for Microwave Detection and Treatment of Breast Cancer:
[85] Y. Chen et al., “Multiple-Input Multiple-Output Radar for Lesion Clas- Enhanced Dielectric and Heating Response of Tissue-Mimicking
sification in Ultrawideband Breast Imaging,” IEEE J. Sel. Topics Signal Materials,” IEEE Trans. Biomed. Engg., vol. 57, no. 8, pp. 1831-1834,
Process., vol. 4, no. 1, pp. 187-201, Feb. 2010. Aug. 2010.
[86] P. M. Meaney et al., “A clinical prototype for active microwave imaging [111] G. Bellizzi et al., “Microwave Cancer Imaging Exploiting Magnetic
of the breast,” IEEE Trans. Microw. Theory Tech., vol. 48, no. 11, pp. Nanoparticles as Contrast Agent,” IEEE Trans. Biomed. Eng., vol. 58,
1841-1853, Nov. 2000. no. 9, pp. 2528-2536, Sept. 2011.
[87] A. E. Souvorov et al., “Two-dimensional computer analysis of a mi- [112] C. Gilmore et al., “A Wideband Microwave Tomography System With
crowave flat antenna array for breast cancer tomography,” IEEE Trans. a Novel Frequency Selection Procedure,” IEEE Trans. Biomed. Eng.,
Microw. Theory Tech., vol. 48, no. 8, pp. 1413-1415, Aug. 2000. vol. 57, no. 4, pp. 894-904, April 2010.
[88] A. E. Bulyshev et al., “Computational modeling of three-dimensional [113] A. Fhager et al., “3D Image Reconstruction in Microwave
microwave tomography of breast cancer,” IEEE Trans. Biomed. Eng., Tomography Using an Efficient FDTD Model,” IEEE Antennas Wireless
vol. 48, no. 9, pp. 1053-1056, Sept. 2001. Propag. Lett., vol. 8, pp. 1353-1356, 2009.
[89] Dun Li et al., “Conformal microwave imaging for breast cancer detec- [114] K. D. Paulsen, and P. M. Meaney, “Nonactive antenna compensation
tion,” Microwave Theory and Techniques, IEEE Trans. Microw. Theory for fixed-array microwave imaging. I. Model development,” IEEE
Tech., vol. 51, no. 4, pp. 1179-1186, Apr. 2003. Trans. Med. Imag., vol. 18, no. 6, pp. 496-507, June 1999.
[90] M. J. Burfeindt et al., “Quantitative Microwave Imaging of Realistic [115] O. Franza, “SICS: a sensor interaction compensation scheme for
Numerical Breast Phantoms Using an Enclosed Array of Multiband, microwave imaging,” IEEE Trans. Antennas Propag., vol. 50, no. 2,
Miniaturized Patch Antennas,” IEEE Antennas Wireless Propag. Lett., pp. 211-216, Feb 2002.
vol. 11, pp. 1626-1629, 2012. [116] P. Mojabi, and J. LoVetri, “A Novel Microwave Tomography Sys-
[91] T. U. Gu¨rbu¨z et al., “A Nonlinear Microwave Breast Cancer Imaging tem Using a Rotatable Conductive Enclosure,” IEEE Trans. Antennas
Approach Through Realistic Body-Breast Modeling,” IEEE Trans. An- Propag., vol. 59, no. 5, pp. 1597-1605, May 2011.
tennas Propag., vol. 62, no. 5, pp. 2596-2605, May 2014. [117] S. E. Lapinsky MB, and A. C. Easty, “Electromagnetic interference in
[92] E. C. Fear et al., “Enhancing breast tumor detection with near-field critical care,” J. Critical Care, vol. 21, pp. 267-270, 2006.
imaging,” IEEE Microw. Mag., vol. 3, no. 1, pp. 48-56, Mar. 2002. [118] E. Hanada, and T. Kudou, “Electromagnetic noise in the clinical
[93] M. Klemm et al., “Radar-Based Breast Cancer Detection Using a Hemi- environment,” in Proc. 3rd Int. Symp. Med. Info. Comm. Tech.,
spherical Antenna Array-Experimental Results,” IEEE Trans. Antennas Montreal, Canada, Feb. 2009.
Propag., vol. 57, no. 6, pp. 1692-1704, June 2009. [119] Medfield Diagnostics AB, Medicinaregatan 8B, 430 90 Gothenburg,
[94] A. Diaz-Bolado et al., “Study of Microwave Tomography Measurement Sweden. Website:www.medfielddiagnostics.com
Setup Configurations for Breast Cancer Detection Based on Breast [120] Micrima, Bristol. Website:http://www.micrima.com/
Compression,” Int. J. Antennas Propag., vol. 2013, Article ID 701378, [121] Y. Qi et al., “Application of sparse array and MIMO in near-range mi-
12 pages, 2013. crowave imaging,” in Proc. SPIE 8179, SAR Image Analysis, Modeling,
[95] R. K. Amineh et al., “Near-Field Microwave Imaging Based on Techniq. XI, vol. 8179, Oct. 2011.
Aperture Raster Scanning with TEM Horn Antennas,” IEEE Trans. [122] O. T. Z. Daniel et al., “Design and experimental investigation of UWB
Antennas Propag., vol. 59, no. 3, pp. 928-940, Mar. 2011. microwave imaging via MIMO beamforming,” in Proc. IEEE Int. Conf.
[96] R. Bansal, “Battling cancer: the latest on microwave hyperthermia,” Ultra-Wideband (ICUWB), vol. 2, Nanjing, Sept. 2010.
IEEE Microw. Mag., vol. 6, no. 3, pp. 32-34, Sept. 2005. [123] T. Thiasiriphet et al., “Compressed Sensing for UWB medical radar
[97] C. J. Diederich, “Thermal ablation and high-temperature thermal ther- applications,” in Proc. IEEE Int. Conf. Ultra-Wideband (ICUWB), pp.
apy: Overview of technology and clinical implementation,” Int. J. 106-110, Sept. 2012.
Hyperthermia, vol. 21, no. 8, pp. 745-753, 2005.
[98] M. Lazebnik et al., “Ultrawideband temperature-dependent dielectric
properties of animal liver tissue in the microwave frequency range,”
Phys. Med. Biol., vol. 51, pp. 1941-1955, 2006.
[99] J. C. Bolomey et al., “On the possible use of microwave-active imaging
for remote thermal sensing,” IEEE Trans. Microw. Theory Tech., vol.
31, no. 9, pp. 777-781, Sep. 1983.
[100] M. Miyakawa, and J. C. Bolomey, “Non-Invasive Thermometry of the
Human Body,” London, CRC Press, Dec. 1995.
[101] P. M. Meaney et al., “Microwave thermal imaging: initial in vivo
experience with a single heating zone,” Int. J. Hyperthermia, vol. 19,
no. 6, pp. 617-641, Dec. 2003.
[102] M. Hayness et al., “Real-time Microwave Imaging of Differential
Temperature for Thermal Therapy Monitoring,” IEEE Trans. Biomed.
Eng., vol. 61, no. 6, pp. 1787-1797, June 2014.
[103] C. Gilmore et al., “Microwave Imaging of Human Forearms: Pilot
Study and Image Enhancement,” Int. J. Biomed. Imag., vol. 2013,
Article ID 673027, 17 pages, 2013.
[104] J. J. Mallorqui et al., “Quantitative images of large biological bodies in
microwave tomography by using numerical and real data,” Electronics
Lett. , vol. 32, no. 23, pp. 2138-2140, Nov. 1996.
[105] R. Chandra et al., “A microwave imaging based technique to localize
an in-body RF-source for biomedical applications,” IEEE Trans.
Biomed. Eng., to be published.
[106] S.Y. Semenov et al., “Three-dimensional microwave tomography: ini-
tial experimental imaging of animals,” IEEE Trans. Biomed. Eng., vol.
49, no. 1, pp. 55-63, Jan. 2002.
[107] V. Hamsakutty et al., “Coupling medium for microwave medical
imaging applications,” Electronics Lett., vol. 39, no. 21, pp. 1498-1499,
Oct. 2003.