fixation , dehydrates, clearing and infiltration of tissue specimen to help preserve the integrity of their cellular components for light or electron microscopy. 2nd DEFINITION:- An automated system used to process tissue specimen for examination through fixation , dehydration and infiltration. • It is an automatic tissue processor with 12 beakers which are thermostatically controlled.
• These beakers contain different grades of alcohol , clearing agents
and molten wax in wax bath. There is a timing disc.
• After fixation small piece of tissue are placed in tissue processing
baskets . These are perforated metal container. • This equipment automatically dehydrates , clears & impregnate tissues. • Hence it saves time , manpower and reduce human errors. • It also minimise the variation in time from one batch of tissue to the next. • The decrease in processing time is due to constant agitation ,which the tissue undergoes. PRINCIPLE • 1. The general working principle of all the models used are similar while the tissue is being processed on electric motor , agitate the baskets either by means of slowly rotating it or by up and down movement. • 2. This stops while the tissue is being transferred from one container to the next. The basket on being submersed in the fresh fluid , continue to agitate. PRINCIPLE- CONTD.... • 3. The length of time and basket in each reagent is determined by notch which are cut in a 24 hrs clock disc. • 4. These are pulled out at the required transferred time with the tab key which is provided in the machine. COMPONENTS • 1. A timer Clock • 2. Basket carrier in a circular structure. • 3. Receptacle and Receptacles (stainless steel or plastic capsule) • 4. Circular deck which holds the reagent beaker and paraffin bath. WORKING procedure • 1. Small blocks of tissue are enclosed in the perforated capsule and the capsule are placed in the baskets. • 2. The baskets is attached to one of the 12 yokes in the supra structure while it is in raised position. • 3. The entire superstructure descend immersing the basket in the first solution and sealing the other reagent beakers to prevent evaporation. • 4. To move the baskets from reagent to the next in the processing sequence the entire superstructure ascends and descends at schedule intervals controlled by the time clocks. WORKING OF HISTOKINETTE- CONTD.... • 5. During immersion in the fluid the basket oscillate up and down in a reciprocal motion to keep the tissue and reagent in a state of controlled in a state of controlled agitation which significantly increase the speed of penetration. • 6. After Prescribe amount of time in the last Paraffin wax bath, the tissue remain there until removed manually. • 7. It takes about 16 hours for routine Processing. Hence the machine is set in the afternoon/evening and the processed material is taken out on the following morning for embedding. REQUIREMENT • 1. Alcohol solution of various Concentration • 2. Xylene • 3. Paraffin Wax • 4. Tissue • 5. Automatic Tissue Processor PROCEDURE • 1. Place the solution and Paraffin in respective beaker of the equipment. • 2. After these steps follows:- PROCEDURE – CONTD..... After these steps follows:- STEPS Timing Solution Period
1 Starting 80% alcohol 0 hours
2 4:30 – 6:30 PM 80% alcohol 2 hours
3 6:30- 7:30 PM 95% alcohol 1 hours
4 7:30- 8:30 PM 95% alcohol 1 hours
5 8:30- 9:30 PM 100% alcohol 1 hours
6 9:30- 10:30 PM 100% alcohol 1 hours
7 10:30- 11:30 PM 100% alcohol 1 hours
8 11:30 – 12: 30 Xylene 1 hours
9 12:30 – 2: 30 Xylene 2 hours
10 2:30 – 4: 30 Paraffin 2 hours
11 4:30 – 6: 30 Paraffin 2 hours
12 6:30 – 8: 30 Paraffin 2 hours
. Place the solution & Paraffin in respective beakers of the equipment. The timing lever is set at zero & machine is started at 4:30 PM ( so that the tissue processing is complete during night hours ) The basket with the cassettes automatically changes position & takes a bath different reagent kept in the beaker in order to accomplish, a. dehydration b. clearing c. infiltration The final dip is in the warm paraffin. Cassettes are opened next day morning for embedding . CARE OF AUTOTECHNICON • 1. The care of autotechnicon is extremely important. • 2. Paraffin should be kept in Paraffin bath and removed from all other areas of the instruments with soft cloth soaked in Xylene.