Professional Documents
Culture Documents
: TDS-03-26
Release Top Drive Drilling Systems
October 20, 2003
Affected
Assemblies: Accumulator (P/N’s 82674, 84601, 107060,110562, 110563, 110564 & 115290)
Solution: The following sections relating to pre-charging and maintenance of accumulators will
be expanded in future service manuals.
1) Use an inert gas such as nitrogen for pre-charging accumulators. Do not use
oxygen.
a. The accumulator end caps are screwed flush into the accumulator body.
b. The gas valve is screwed in tight.
c. No oil remains trapped in the top of the accumulator.
d. The charging hose fitting on nitrogen charging kit is clean, as not to
introduce any contaminants to the accumulator when pre-charging.
Note: Depending on the size of the accumulator there are two possible types of gas
valves that may be found. Either the cored gas valve cartridge (ISO-4570-8V1;
Varco P/N 87979) or the heavy duty military poppet-type gas valve cartridge
(MS28889-2; Varco P/N 107060-GV1) may be supplied, as shown in Figure 1.
(ISO-4570-8V1) (MS28889-2)
4) Attach the charging hose to the nitrogen bottle and to the gas valve using the
following procedure:
If your accumulator has a cored gas valve cartridge follow Steps A through H.
Note: Use thread sealing and lubricating compound on the pipe threads of the
gauge.
a. Back gas chuck “T” stem completely out of the way out (CCW) before
attaching the charging assembly to accumulator gas valve. Caution: Make sure
not to loop or twist the hose, attach the swivel nut to the gas valve and tighten
(10-15 in-lbs).
b. Use a wrench to tighten the gas chuck swivel nut onto the gas valve. Close the
bleed valve.
d. Crack open nitrogen bottle valve and slowly fill accumulator. Shut off when the
gauge indicates the desired gas pressure.
e. Let the pre-charge set for 10-15 minutes, allowing the gas temperature to
stabilize. If gas pressure is exceeded, close nitrogen bottle valve, then slowly
open bleed valve (opposite the gas valve below the gauge). Caution: Do not
reduce pre-charge by depressing valve core with a foreign object. High
pressure will rupture rubber valve seat.
f. Before loosening the swivel nut, turn the gas chuck stem all the way out, then
open bleed valve.
g. To prevent the gas valve from turning, loosen swivel nut and remove the
assembly.
h. Replace the gas valve cap (torquing to 10-15 in-lbs.) and valve guard.
For heavy duty military poppet-type gas valve cartridges follow Steps I through P as
follows:
Note: Use thread sealing and lubricating compound on the pipe threads of the
gauge.
i. Back gas chuck “T” stem completely out of the way out (CCW) before
attaching the charging assembly to accumulator gas valve. Caution: Make sure
not to loop or twist the hose, attach the swivel nut to the gas valve and tighten
(10-15 in-lbs).
j. Use a wrench to tighten the gas chuck swivel nut onto the gas valve. Close the
bleed valve.
k. Hold the gas valve below the hex nut with one wrench while unscrewing hex
nut with a second wrench. This will open the poppet inside the gas valve.
Three (3) turns will fully open the valve.
l. Crack open nitrogen bottle valve and slowly fill accumulator. Shut off when the
gauge indicates the desired gas pressure.
m. Let the pre-charge set for 10-15 minutes, allowing the gas temperature to
stabilize. If gas pressure is exceeded, close nitrogen bottle valve, then slowly
open bleed valve (opposite the gas valve below the gauge). Caution: Do not
reduce pre-charge by depressing valve core with a foreign object. High
pressure will rupture rubber valve seat.
n. Tighten the hex nut with a wrench to internally close the poppet.
p. Replace the gas valve cap (torquing to 10-15 in-lbs.) and valve guard.
Maintenance-
Periodic testing of the pre-charge pressure will detect whether O-ring wear is sufficient
to degrade sealing performance. There is currently no maintenance schedule for
verification of accumulator pre-charge, however to be consistent with the
manufacturer, Varco is recommending that pre-charge be checked one week after
initial equipment installation, and thereafter on a monthly basis.
If an accumulator will not maintain pre-charge pressure, contact your local Varco
Service Center to order seal kits shown in table below.
Troubleshooting-
Refer to the Troubleshooting section of your service manual for symptoms, probable
causes and remedies related to loss of accumulator pre-charge. This includes but is
not limited to operation of the alignment cylinder, IBOP valve Actuator,
counterbalance system, dolly retract systems, on-board hydraulic power units and
stand jump on your top drive.
Below is a table outlining piston accumulators currently used on Varco Top Drive
models. Also included is the applicable seal kit part number and pre-charge
pressures.
Please contact your local Varco Service Center if you have any questions regarding this
product bulletin.