Professional Documents
Culture Documents
General Notices (1) apply to all monographs and other texts 139
2.4.20. Determination of metal catalyst or metal reagent residues EUROPEAN PHARMACOPOEIA 9.0
System suitability. A system suitability test must be carried C = concentration of metal in the analysed sample, in
out on the day of the analysis to ensure that the sample micrograms per gram ;
preparation and measurement system are appropriate. A = instrument reading of the concentration of the
Acceptance criterion for preparation of sample solution : a clear metal in the sample solution, in micrograms per
solution is obtained. millilitre ;
Acceptance criterion for measurement system : the measured m = mass of the sample in the initial sample solution,
concentration of a standard solution of the metal at a in grams ;
concentration within the range of the used calibration curve V1 = volume of the initial sample preparation, in
does not differ from the actual concentration by more than millilitres ;
20 per cent. V2 = total volume of any dilution performed, in
Calculation. The blank value of reagents must be taken into millilitres ;
account for the calculation of the content. Upon completion of V3 = volume of initial sample preparation used in any
the analysis, the concentration of a given metal in the sample dilution performed, in millilitres.
is calculated by the software of the instrument from the
concentration of the metal in the test solution. If no calculation
software is available or no indication for calculation is given in VALIDATION REQUIREMENTS
the corresponding general chapter in section 2.2. Physical and Some validation requirements provided below may differ from
physicochemical methods, the concentration of a given metal those provided in general chapters of the Ph. Eur. (e.g. 2.2.22
in the sample can be calculated from the concentration of the (AES), 2.2.23 (AAS), 2.2.57 (ICP-AES), 2.2.58 (ICP-MS)).
metal in the solution using the following expression : Before the initial use of the selected procedure, the analyst
must ensure that the sample preparation and measurement
method are appropriate for the metal(s), sample matrix and
General Notices (1) apply to all monographs and other texts 141