You are on page 1of 4

2.4.19. Alkaline impurities in fatty oils EUROPEAN PHARMACOPOEIA 9.

01/2008:20419 If no sample preparation and/or measurement method is


described in the specific monograph, a suitable sample
preparation and/or measurement method must be developed
and validated (see Figures 2.4.20.-1 and 2.4.20.-2).

2.4.19. ALKALINE IMPURITIES IN SAMPLE PREPARATION


FATTY OILS Sample preparation is critical to the success of elemental
In a test-tube mix 10 mL of recently distilled acetone R and analysis. Many techniques not using direct measurement are
0.3 mL of water R and add 0.05 mL of a 0.4 g/L solution of heavily dependent on sample transport.
bromophenol blue R in alcohol R. Neutralise the solution if If an atomisation system is used, the most conventional means
necessary with 0.01 M hydrochloric acid or 0.01 M sodium by which samples are introduced into the atomisation system
hydroxide. Add 10 mL of the oil to be examined, shake and is by solution nebulisation. In this case, solid samples must
allow to stand. Not more than 0.1 mL of 0.01 M hydrochloric be dissolved in order to be introduced into the atomisation
acid is required to change the colour of the upper layer to system. Samples may be dissolved in any appropriate solvent.
yellow. The use of aqueous or dilute nitric acid solutions is strongly
recommended, due to minimal interference with these
04/2013:20420 solvents compared to other solvents. Hydrochloric acid,
hydrofluoric acid, perchloric acid, sulfuric acid and hydrogen
peroxide, at various concentrations, can be used to dissolve
the samples. The viscosity of sulfuric acid is greater than
that of the other acids and is to be taken into account as it
2.4.20. DETERMINATION OF METAL can affect the overall fluidity of the solution. The choice
CATALYST OR METAL REAGENT of solvents includes, but is not limited to, the use of dilute
bases, straight or diluted organic solvents, combinations of
RESIDUES acids or bases, and combinations of organic solvents. Acids,
bases, and hydrogen peroxide of high purity must be used,
INTRODUCTION
especially when ICP-MS is employed. For aqueous solutions,
This chapter describes the general approach for the use deionised distilled water R. Diluents must be checked for
determination of metal catalyst or metal reagent residues interference if they are used in an analysis. Because it is not
in substances for pharmaceutical use. As the chemical always possible to obtain organic solvents that are free of
composition of the considered substances and the specification metals, organic solvents of the highest purity possible with
limits for the metal(s) of interest vary considerably, it is regard to metal contaminants must be used. Specifically
not possible to describe all suitable sample preparation and for ICP techniques, where samples are introduced into the
measurement methods. Therefore, any method that fulfils the plasma via solution nebulisation, it is important to consider
requirements described in this chapter may be used. the potential matrix effects and interferences that might arise
The results of the analysis are acceptable only if the system from the solvent. The use of an appropriate internal standard
suitability has been demonstrated by a suitable test. Before and/or matching the standard matrix with samples should
the initial use of a method, the analyst must ensure that the be applied for ICP-AES and ICP-MS analyses in cases where
method is appropriate for the samples and instruments used. accuracy and precision are not sufficient. In any case, the
This is accomplished by applying a validation procedure selection of an appropriate internal standard should take into
to methods not described in the specific monograph or account the metal(s) of interest, ionisation energy, wavelengths
by a system suitability test for methods described in the or masses, and the nature of the sample matrix.
monograph. Decision trees for the choice of the sample Where a sample is found not to be soluble in any acceptable
preparation and the measurement procedures are presented in solvent, a variety of digestion or incineration techniques can be
Figures 2.4.20.-1 and 2.4.20.-2. employed. These include hot-plate digestion, incineration and
microwave-assisted digestions, using open- and closed-vessel.
PROCEDURES The decision regarding the type of digestion technique to be
As a reference procedure is not provided for each metal, matrix used depends on the nature of the sample being digested,
and concentration, the choice of procedure according to as well as on the metal(s) of interest and the concentration
Figures 2.4.20.-1 and 2.4.20.-2, including sample preparation, range of the metals to be quantified. Open-vessel digestion
detection technique and instrument parameters, is the is not recommended for the analysis of volatile metals.
responsibility of the user. The suitability of a digestion technique, whether open-
or closed-vessel, should be supported by spike recovery
Use the flow chart in Figure 2.4.20.-1 to define the sample
experiments in order to verify that, within an acceptable
preparation method and the flow chart in Figure 2.4.20.-2 to
tolerance, volatile metals have not been lost during sample
define the measurement method. The sample preparation
preparation. The digestion cycle is suitable if a clear solution
method should yield a sufficient quantity of sample to allow
is obtained.
quantification of each metal at the specified limit stated in the
specific monograph or the general chapter. It is important to consider the selection of the type, the
All suitable sample preparation methods and measurement material of construction, the pretreatment, and the cleaning
techniques (e.g. 2.2.22. Atomic emission spectrometry (AES), of analytical labware used in elemental analyses. The material
2.2.23. Atomic absorption spectrometry (AAS), 2.2.37. X-ray must be inert and, depending on the specific application,
fluorescence spectrometry (XRFS), 2.2.57. Inductively resistant to caustics, acids, and/or organic solvents. For some
coupled plasma-atomic emission spectrometry (ICP-AES), analyses, care must be exercised to prevent the adsorption of
2.2.58. Inductively coupled plasma-mass spectrometry metals onto the surface of a vessel, particularly in ultra-trace
(ICP-MS), 2.4.2. Arsenic, 2.4.8. Heavy metals, 2.4.9. Iron, analyses. Contamination of sample solutions by metals and
2.4.10. Lead in sugars, 2.4.15. Nickel in polyols, 2.4.31. Nickel in ions present in the container can also lead to inaccurate results.
hydrogenated vegetable oils) can be used for the determination The use of volumetric glassware that does not comply
of metal residues, if the method has been verified before the with Class A requirements of the appropriate International
initial use by a system suitability test or a validation procedure Standard of the International Organization for Standardization
according to this chapter. (ISO) is acceptable if the validation or the system suitability

138 See the information section on general monographs (cover pages)


EUROPEAN PHARMACOPOEIA 9.0 2.4.20. Determination of metal catalyst or metal reagent residues

Figure 2.4.20.-1. – Metal residues decision tree : sample preparation

test of the method using such glassware have experimentally MEASUREMENT


demonstrated that the method is suitable for the intended Method. The choice of the techniques depends mainly on
purpose. the sample matrix and the characteristics and specification
CAUTION : when using high-pressure digestion vessels and limits of the metal(s) of interest. Analyse according to the
microwave laboratory equipment, the safety precautions and instructions of the manufacturer of the apparatus regarding
operating instructions given by the manufacturer must be programme and wavelength.
followed.

General Notices (1) apply to all monographs and other texts 139
2.4.20. Determination of metal catalyst or metal reagent residues EUROPEAN PHARMACOPOEIA 9.0

Figure 2.4.20.-2. – Metal residues decision tree : measurement

System suitability. A system suitability test must be carried C = concentration of metal in the analysed sample, in
out on the day of the analysis to ensure that the sample micrograms per gram ;
preparation and measurement system are appropriate. A = instrument reading of the concentration of the
Acceptance criterion for preparation of sample solution : a clear metal in the sample solution, in micrograms per
solution is obtained. millilitre ;
Acceptance criterion for measurement system : the measured m = mass of the sample in the initial sample solution,
concentration of a standard solution of the metal at a in grams ;
concentration within the range of the used calibration curve V1 = volume of the initial sample preparation, in
does not differ from the actual concentration by more than millilitres ;
20 per cent. V2 = total volume of any dilution performed, in
Calculation. The blank value of reagents must be taken into millilitres ;
account for the calculation of the content. Upon completion of V3 = volume of initial sample preparation used in any
the analysis, the concentration of a given metal in the sample dilution performed, in millilitres.
is calculated by the software of the instrument from the
concentration of the metal in the test solution. If no calculation
software is available or no indication for calculation is given in VALIDATION REQUIREMENTS
the corresponding general chapter in section 2.2. Physical and Some validation requirements provided below may differ from
physicochemical methods, the concentration of a given metal those provided in general chapters of the Ph. Eur. (e.g. 2.2.22
in the sample can be calculated from the concentration of the (AES), 2.2.23 (AAS), 2.2.57 (ICP-AES), 2.2.58 (ICP-MS)).
metal in the solution using the following expression : Before the initial use of the selected procedure, the analyst
must ensure that the sample preparation and measurement
method are appropriate for the metal(s), sample matrix and

140 See the information section on general monographs (cover pages)


EUROPEAN PHARMACOPOEIA 9.0 2.4.22. Composition of fatty acids by GC

instrument used. This is accomplished by following the 01/2008:20421


validation procedure before the initial use and the system
suitability test on the day of the analysis.
For metal residues, validation of a limit test must include
specificity and limit of detection.
The following section defines the characteristics for the 2.4.21. FOREIGN OILS IN FATTY OILS
acceptability of a quantitative procedure. It must be BY THIN-LAYER CHROMATOGRAPHY
demonstrated experimentally that such a procedure complies
with the validation requirements, with an appropriate system Examine by thin-layer chromatography (2.2.27) using
suitability test using material spiked with a suitable reference kieselguhr G R as the coating substance. Impregnate a plate by
material. The test materials must be spiked before any sample placing it in a chromatographic tank containing the necessary
preparation steps. For example, if a test material is to be quantity of a mixture of 10 volumes of liquid paraffin R
digested, the material must be spiked at the beginning of the and 90 volumes of light petroleum R so that the plate dips
digestion procedure. about 5 mm beneath the surface of the liquid. When the
SPECIFICITY impregnation mixture has risen by at least 12 cm from the
lower edge of the plate, remove the plate and allow the solvent
Specificity is the ability to ensure that the analytical procedures
to evaporate for 5 min. Carry out the chromatography in the
for sample preparation and measurement allow a reliable
same direction as the impregnation.
determination of the metal(s) in the presence of components
(e.g. carrier gas, impurities, matrix) that may be expected to Preparation of the mixture of fatty acids. Heat 2 g of the oil
be present. with 30 mL of 0.5 M alcoholic potassium hydroxide under a
Acceptance criteria : the procedure must be able to assess reflux condenser for 45 min. Add 50 mL of water R, allow
unequivocally each metal residue to be determined with to cool, transfer to a separating funnel and extract with
this procedure in the presence of components that may be three quantities, each of 50 mL, of ether R. Discard the ether
expected to be present, including other metal residues, matrix extracts, acidify the aqueous layer with hydrochloric acid R
components, and other sources of interference ; specificity is and extract with three quantities, each of 50 mL, of ether R.
demonstrated by complying with the accuracy requirement Combine the ether extracts and wash with three quantities,
for the metal(s) to be determined. each of 10 mL, of water R ; discard the washings, dry the ether
over anhydrous sodium sulfate R and filter. Evaporate the ether
RANGE on a water-bath. Use the residue to prepare the test solution.
Acceptance criterion : range is demonstrated by complying The fatty acids may also be obtained from the soap solution
with the recovery requirement. prepared during the determination of the unsaponifiable
ACCURACY matter.
Verify the accuracy using a certified reference material (CRM) Test solution. Dissolve 40 mg of the mixture of fatty acids
or by performing a test for recovery. obtained from the substance to be examined in 4 mL of
Recovery. Recovery may be determined on a sample of the chloroform R.
substance to be examined, spiked with a known quantity of a Reference solution. Dissolve 40 mg of the mixture of fatty acids
reference standard of the metal (3 concentration levels in the obtained from a mixture of 19 volumes of maize oil R and
range of 50-150 per cent of the intended specification limit, 1 volume of rapeseed oil R in 4 mL of chloroform R.
even if the original concentration of the reference standard is Apply to the plate 3 μL of each solution. Develop over a path of
at the specified value), in triplicate. 8 cm using a mixture of 10 volumes of water R and 90 volumes
Acceptance criterion : spike recovery is within 70 per cent and of glacial acetic acid R. Dry the plate at 110 °C for 10 min.
150 per cent for the mean of 3 replicates at each concentration. Allow to cool and, unless otherwise prescribed, place the plate
REPEATABILITY in a chromatographic chamber, with a tightly fitting lid, that
has previously been saturated with iodine vapour by placing
Test samples : either 6 independent samples of the substance
iodine R in an evaporating dish at the bottom of the chamber.
to be examined spiked with a suitable reference standard at
After some time brown or yellowish-brown spots become
the specified concentration level, or 3 concentration levels
visible. Remove the plate and allow to stand for a few minutes.
prepared in triplicate.
When the brown background colour has disappeared, spray
Acceptance criterion : the relative standard deviation is in both
with starch solution R. Blue spots appear which may become
cases not more than 20 per cent.
brown on drying and again become blue after spraying with
INTERMEDIATE PRECISION water R. The chromatogram obtained with the test solution
The effect of random events (intra-laboratory variations) on always shows a spot with an RF of about 0.5 (oleic acid) and
the analytical precision of the method must be established. a spot with an RF of about 0.65 (linoleic acid) corresponding
Acceptable experiments for establishing intermediate to the spots in the chromatogram obtained with the reference
precision include performing the repeatability analysis on solution. With some oils a spot with an RF of about 0.75 may
different days, or with different instrumentation, or by be present (linolenic acid). By comparison with the spot in the
different analysts. Only 1 of the 3 experiments is required to chromatogram obtained with the reference solution, verify the
demonstrate intermediate precision. absence in the chromatogram obtained with the test solution
Acceptance criterion : the relative standard deviation is not of a spot with an RF of about 0.25 (erucic acid).
more than 25 per cent.
LIMIT OF QUANTIFICATION 07/2016:20422
Determine the lowest concentration meeting the acceptance
criterion. Use the results from the accuracy study.
Acceptance criterion : the limit of quantification is below the
specification limit.
LIMIT OF DETECTION (ONLY APPLICABLE TO LIMIT 2.4.22. COMPOSITION OF FATTY
TESTS)
Determine the lowest concentration giving a signal clearly
ACIDS BY GAS CHROMATOGRAPHY
distinct from that obtained with a blank solution. The test for foreign oils is carried out on the methyl esters
Acceptance criterion : the limit of detection is not more than of the fatty acids contained in the oil to be examined by gas
0.5 times the concentration of the specification limit. chromatography (2.2.28).

General Notices (1) apply to all monographs and other texts 141

You might also like