You are on page 1of 8

ARTICLE

Retrospective Analysis of Pediatric and Adult


Populations Using an LC-MS/MS Method for
Oxcarbazepine/Eslicarbazepine Metabolite
Grace M. Kroner,a Ronald L. Thomas,b and Kamisha L. Johnson-Davisa,b,*

Downloaded from https://academic.oup.com/jalm/article/6/3/637/5961622 by guest on 13 October 2022


Background: Therapeutic drug monitoring of anti-epileptic drugs is important to manage seizure control in
patients with epilepsy. Oxcarbazepine is a second-generation anti-epileptic drug approved for use in pediatric
patients, and eslicarbazepine acetate is a newer generation drug used as adjunctive therapy and monotherapy for
partial-onset (focal) seizures. While several second and third generation anti-epileptic drugs have broader thera-
peutic efficacy in patients, these drugs can still have severe side effects and variable interpatient pharmacokinet-
ics. Consequently, there is a need for accurate and sensitive analytical methods to support therapeutic drug moni-
toring.
Methods: An assay improvement for a LC-MS/MS method was developed for the major metabolite of oxcarbaze-
13
pine and eslicarbazepine, licarbazepine (MHD), using a C-labeled form of the compound as the internal stan-
dard. Additionally, retrospective data analysis was used to compare the distribution of results observed in adult vs
pediatric patients.
Results: Accuracy and linearity across the analytical measuring range of 1 to 60 mg/mL was acceptable. Inter- and
intra-run precision was less than 6% at 3 concentrations tested. The limit of detection was determined to be
0.5 mg/mL. Significant interference from hemolysis, icterus, lipemia, or 187 other potential interferences was not
detected.
Conclusions: The improved assay for MHD was appropriate for clinical use. Retrospective data analysis showed
that pediatric and adult patients had a similar distribution of oxcarbazepine/eslicarbazepine metabolite concen-
trations in serum.

INTRODUCTION been approved for treating epilepsy since 1993


(1). Unlike some of the first generation anti-
Epilepsy can manifest through different types of epileptic drugs (AEDs), these second and third
seizures (either partial or generalized), and various generation drugs usually have fewer drug–drug
medications have been found to be more effective interactions and less severe adverse effects. While
with certain types of seizures. Several drugs have general therapeutic ranges in adult populations

a
Department of Pathology, University of Utah Health Sciences Center, Salt Lake City, UT; bARUP Institute for Clinical and Experimental Pathology,
Salt Lake City, UT.
* Address correspondence to this author at: 500 Chipeta Way, Mail Code 115, Salt Lake City, UT 84108. Fax (801) 584-5207; e-mail kamisha.davis@
hsc.utah.edu.
Previous presentations: poster presentation at MSACL 2019 meeting in Palm Springs, CA.
Received May 26, 2020; accepted September 10, 2020.
DOI: 10.1093/jalm/jfaa179
C American Association for Clinical Chemistry 2020. All rights reserved.
V
For permissions, please email: journals.permissions@oup.com.

...............................................................................................
May 2021 | 06:03 | 637–644 | JALM 637
ARTICLE MHD Method Validation and Retrospective Analysis

IMPACT STATEMENT
A mass spectrometry method was redeveloped and validated to improve analytical accuracy and specif-
icity for the quantification of the major metabolite of oxcarbazepine and eslicarbazepine. The improved LC-
MS/MS metabolite assay was appropriate for clinical use in both pediatric and adult populations and the
analytical gains in sensitivity and specificity will improve the accuracy of patient results.

Downloaded from https://academic.oup.com/jalm/article/6/3/637/5961622 by guest on 13 October 2022


are available for many drugs, ranges specifically limited numbers of subjects available for these
for pediatric patients are not well established, de- studies, wider scale investigation of concentra-
spite the fact that children comprise up to 25% of tions observed in children relative to adults would
all epilepsy cases (1). be helpful for clinical interpretation.
Oxcarbazepine is a second generation AED tar- The goal for this study was to perform an assay
geted for treatment of partial seizures (2) and esli- improvement for a liquid chromatography tandem
carbazepine acetate was approved to treat mass spectrometry (LC-MS/MS)-based assay for
partial-onset seizures. Both drugs have structures the oxcarbazepine/eslicarbazepine metabolite.
similar to carbamazepine, and both drugs are me- The assay was modified to optimize the mass
tabolized to licarbazepine, monohydroxy deriva- transitions for MHD to improve specificity, imple-
tive (MHD), which is the active form of the drugs, ment a MHD-13C6 internal standard to enhance
and glucuronide metabolites for renal elimination. assay precision, and change the analytical plat-
MHD reduces seizures by inhibiting sodium chan- form to gain a higher dynamic range to advance
nel function (2). Oxcarbazepine and MHD are 40% linearity, compared to the previous LC-MS/MS
bound by plasma proteins, while eslicarbazepine method. After implementation of the improved
is 30% bound (3). Side effects of these drugs in- method, we performed a retrospective analysis of
clude dizziness, fatigue, and vomiting (4, 5). patient results analyzed by the revised method to
Published clinical studies have demonstrated compare concentrations from pediatric and adult
that liver or kidney impairments alter drug metab- patient populations and to determine whether
olism, and higher doses may be necessary in chil- our method had an analytical measurement range
dren to reach an effective concentration in serum that was suitable for the patient populations
(6). For example, one pharmacokinetic model tested.
demonstrated that the weight-normalized clear-
ance was 0.083 L/hkg in a 15 kg, 96 cm child (ap-
proximately 3-4 years of age), and 0.060 L/hkg in a MATERIALS AND METHODS
35 kg, 135 cm child (approximately 11 years of
age), but only 0.046 L/hkg in a 70 kg, 176 cm adult A mass-spectrometry based method was devel-
(7). In agreement with that model, MHD half-life oped at our institution in order to monitor blood
and mean specific area under the plasma concentrations of the major metabolite for oxcar-
concentration-time curve (for 48 h) were both ap- bazepine and eslicarbazepine, MHD. A 13C-labeled
proximately 30% less in children 2-5 years of age derivative MHD was used as an internal standard
than in children 6-12 years of age (5). Given the in all of the samples (Cerilliant). Twenty-five

....................................................................................................
638 JALM | 637–644 | 06:03 | May 2021
MHD Method Validation and Retrospective Analysis ARTICLE

microliters of patient sample were mixed with 975 Accuracy was evaluated by comparison of the
lL of internal standard (MHD-13C6)/precipitation so- optimized method to the previous UPLC/mass-
lution (50% methanol/50% acetonitrile) to precipi- spectrometry method, using 108 previously tested
tate proteins in the sample, and the resulting samples (patient samples, CAP Proficiency Testing
supernatant was mixed with 0.1% formic acid in wa- samples, and spiked samples) over 5 days. Sample
ter (mobile phase A) before 5 mL was loaded onto concentration covered the analytical measurement
the instrument. Separation occurred on a range (AMR) of the assay, 1 to 60 mg/mL. Linearity
Phenomenex Kinetex C18 2.6 lm HPLC column (50 was assessed by analyzing 5 calibrator samples
x 2.1 mm) with a switch to mobile phase B of 0.1% that span the AMR. Three replicates at each con-

Downloaded from https://academic.oup.com/jalm/article/6/3/637/5961622 by guest on 13 October 2022


formic acid in acetonitrile and a flow rate of 800 mL/ centration were measured on 5 different runs.
min at a column temperature of 35  C. The HPLC Recovery was tested by splitting 5 positive patient
run time was 2.5 min; the first 30 s was set at 10% samples into 2 aliquots each and spiking one of
mobile phase B, followed by 90 s of 95% mobile them with 30 mg/dL of MHD. Both the fortified and
phase B, then finishing with 30 s of 10% mobile unfortified samples were tested in duplicate, and
phase B. The compounds were detected by multiple from these results, the percent recovery was calcu-
reaction monitoring on an AB Sciex 4000 mass lated. Intra-run imprecision was assessed by mea-
spectrometer, using electrospray ionization in posi- suring 20 replicates each of 3 levels of fortified
tive mode. The mass transitions used for the inter- samples (low: 5 mg/mL, medium: 20 mg/mL, and
nal standard were 261.1 and 198.2 Da for the high: 50 mg/mL concentrations) in 1 day. Inter-run
quantitative peak and 261.1 and 185.2 Da for the imprecision was assessed by measuring 4 repli-
qualitative peak. For MHD, the mass transitions cates each of the 3 levels of fortified samples over
used were 255.1 and 165.1 Da for the quantitative 5 days. Fortified samples were prepared by adding
peak and 255.1 and 179.1 Da for the qualitative standard obtained from Cerilliant into blank
peak. A 5-point calibration curve was used for quan- pooled plasma. The limit of quantification (LOQ)
tification (using concentrations of 1, 5, 10, 30, and was determined by testing at least 5 different con-
60 mg/mL MHD (Cerilliant)). Ascent software (Indigo centrations below and above the expected limit of
BioAutomation) was used for peak area integration. quantification in triplicate on 5 different runs and
The previous method was performed on a assigning the LOQ as the lowest concentration
Waters TQD instrument in positive electrospray measurable with CV less than 15%. The limit of de-
ionization mode. The mobile phase was identical tection was established at the lowest concentra-
to the optimized method; however, the column tion in which the analyte was present at the
utilized was a Waters Acquity UPLC HSS T3 1.8 correct retention time, with a signal to noise ratio
mm, 2.1 x 50 mm and 2 mL was injected onto the greater than 5, and imprecision within 20% CV.
mass spectrometer. The AMR was 1–40 mg/mL Specificity was evaluated by fortifying a blank (neg-
and the mass transitions used for the internal ative) serum sample with 187 potentially interfer-
standard, UCB 17025 (UCB Pharma), were 185.1 ing compounds and performing repeat testing to
and 69.0 Da for the quantitative peak and 185.1 assess for analytical interference. Additionally, 2
and 140.0 Da for the qualitative peak. For MHD, replicates each were analyzed after the addition of
the mass transitions used were 237.1 and hemolyzed, lipemic, or icteric material to evaluate
179.1 Da for the quantitative peak and 237.1 and possible interferences. Finally, 32 negative samples
194.0 Da for the qualitative peak. This method had underwent post-column infusion with MHD to as-
imprecision (þ/-20% CV) and limited specificity sess ion suppression. EP Evaluator was used for
and quantitative accuracy. method evaluation analysis.

..................................................................................................
May 2021 | 06:03 | 637–644 | JALM 639
ARTICLE MHD Method Validation and Retrospective Analysis

For the retrospective data analysis, we examined


2243 unique results for the oxcarbazepine or eslicar-
bazepine metabolite assay in serum/plasma at ARUP
Laboratories. We eliminated results from patients
with reported ages greater than 95 years (n ¼ 6).
Additionally, we eliminated results from patients with
more than one result; 175 results were eliminated.
Any results above or below the analytical measuring
range were eliminated from further analysis

Downloaded from https://academic.oup.com/jalm/article/6/3/637/5961622 by guest on 13 October 2022


(n ¼ 149). Therapeutic reference intervals and toxic
concentrations were from the literature (6, 8, 9). All
statistical analyses were performed using R Studio
and Microsoft Excel. The protocol for collection of the
limited dataset was approved by the University of
Utah Institutional Review Board.

RESULTS

Accuracy Fig. 1. Method comparison between the newly


developed LC-MS/MS method and the UPLC MS
The accuracy of the improved LC-MS/MS method. Dashed line is line of equivalence. Solid
line represents Deming regression.
method on an AB Sciex 4000 instrument was
compared to the previous assay on a Waters TQD
instrument. One hundred and eight patient sam-
ples, previously analyzed on the Waters platform, was 1.45. Linearity across the AMR was within our
were re-extracted and analyzed on the AB Sciex limits of 20% total allowable error and 10% allow-
4000 platform (Fig. 1). The results from Deming re- able systematic error.
gression statistics were: slope: 0.834 6 0.023, y-in-
Recovery
tercept: 2.2 6 0.7 (both shown as mean 6
standard deviation), R2 ¼ 0.96. The acceptance cri- The average recovery was 94%, meeting our
teria for accuracy was: slope: 0.8 6 15%, y-inter- expectations of recovery between 85% and 115%.
cept: 0-3, R2: 0.94 6 15%. The standard deviation
of residuals was 3.7. Of note, our slope criteria Precision
were centered at 0.8, due to limitations of accu- Average intra-run imprecision was 5.2% and av-
racy and specificity of the previous method, which erage inter-run imprecision was 4.3% (Table 1).
was demonstrated on CAP proficiency testing sur-
veys, as a positive bias trend. Sensitivity
The limit of quantification was 1 mg/mL and the
Linearity percent CV was less than 20%. The limit of detec-
We determined the AMR of the assay to be 1 to tion was determined to be 0.5 mg/mL, where the
60 mg/mL. The slope was 1.029 and the y-intercept signal of the analyte appeared at the expected re-
was -0.057 with an observed error of 5.3% and R2 tention time and exhibited a signal to noise ratio
of 0.996. The standard deviation of the residuals greater than 5 (Fig. 2).

....................................................................................................
640 JALM | 637–644 | 06:03 | May 2021
MHD Method Validation and Retrospective Analysis ARTICLE

Table 1. Intra- and inter-run precision data.

Intra-Run Precision (n ¼ 20) Inter-Run Precision (n ¼ 20)


Mean Standard Deviation CV Mean Standard Deviation CV
Low 5.20 0.25 4.8% 5.35 0.29 5.4%
Medium 19.55 1.18 6.0% 20.47 0.76 3.7%
High 52.13 2.44 4.7% 47.83 1.77 3.7%
The units for the mean and standard deviation is mg/mL.

Downloaded from https://academic.oup.com/jalm/article/6/3/637/5961622 by guest on 13 October 2022

Fig. 2. Chromatography of a sample at 0.5 mg/mL showed signal at the expected retention time (indi-
cated by arrow) and a signal to noise ratio greater than 5.

Specificity/Interferences compounds (Supplemental Fig. 1). The drug classes


No interference was detected after fortifying the for the compounds tested for interference were
negative sample with a cocktail containing 187 antidepressants, anticonvulsants, antipsychotics,

..................................................................................................
May 2021 | 06:03 | 637–644 | JALM 641
ARTICLE MHD Method Validation and Retrospective Analysis

Downloaded from https://academic.oup.com/jalm/article/6/3/637/5961622 by guest on 13 October 2022


Fig. 3. Frequency of results for A) pediatric patients and B) adult patients. Median concentration for pe-
diatric and adult patient populations is indicated by an orange or blue line, respectively. Therapeutic
intervals (3-35 mg/mL) are indicated by dashed black lines. Toxic concentration is indicated by a dashed
red line.

antidiabetics, benzodiazepines, barbiturates, significantly higher than the distribution of results


opioids, cardiac drugs, NSAIDS, stimulants, and il- from adult patients, with a median of 17 mg/mL
licit drugs. Additionally, accuracy for hemolyzed (Wilcoxon test P-value ¼ <0.001, Fig. 3). However,
(1800 mg/dL hemoglobin), lipemic (4700 mg/dL tri- both populations had a similar proportion of
glycerides), or icteric (7.25 mg/dL bilirubin) samples patients within the therapeutic range concentra-
containing approximately 10 mg/mL MHD was tions: 92.0% of adult patients and 91.5% of pediat-
greater than 87%. Finally, ion suppression was not ric patients. There was a slightly higher proportion
observed in any of the 32 tested negative samples. of pediatric patients who had concentrations
above the 40 mg/mL threshold for toxicity (4.33%
Retrospective Analysis of pediatric compared to 3.18% of adult patients).
There were 1913 results from unique patients Given the limited number of patients in the pedi-
used for retrospective data analysis. The pediatric atric group, we were unable to assess potential
population was defined as patients less than variation across age within the pediatric group.
18 years of age in our patient population. There
were 531 pediatric patients (55% male), with an
age range from 0 to 17 years, and 1382 adult DISCUSSION
patients (51% male), with an age range from 18 to
95 years. The distribution of results from pediatric An LC-MS/MS assay to quantify MHD was rede-
patients, with a median of 20 mg/mL, was veloped to provide accurate and precise

....................................................................................................
642 JALM | 637–644 | 06:03 | May 2021
MHD Method Validation and Retrospective Analysis ARTICLE

therapeutic drug monitoring results to support contribute to the need for higher dosage or
patient care. The previous method had limita- more frequent dosing in some cases to avoid
tions with accuracy and specificity because the dramatic fluctuations in drug concentrations (3).
method utilized a suboptimal product mass tran- During dosing optimization, these factors may
sition from the loss of a water molecule during cause slightly higher concentrations to be ob-
fragmentation. In addition, the method did not served in pediatric patients. In addition, approxi-
incorporate an isotopically-labeled internal stan- mately 50% of patients were estimated to
dard for MHD. These factors led to an unaccept- require more than one AED to control their
able performance on a CAP ZE proficiency seizures (1). Many AEDs have well-established

Downloaded from https://academic.oup.com/jalm/article/6/3/637/5961622 by guest on 13 October 2022


testing survey. The proficiency test specimens drug–drug interactions and are known to affect
contained multiple antiepileptic drugs and our the pharmacokinetics of oxcarbazepine (3, 6).
reported results had a 30% positive bias in com- Finally, adult patients may be more likely than
parison to the mean, due to poor specificity. The pediatric patients to have a stable dosing regi-
improved method employed mass transitions for men; recommended doses for pediatric patients
MHD that were optimal to minimize interfer- depend on the patient’s weight, and so doses
ences caused by specimen matrix effects or may require more frequent adjustments as the
commonly prescribed drugs. In addition, use of child grows.
a MHD 13C-labeled internal standard improved Our retrospective data analysis was limited by
accuracy of quantification by coeluting with MHD the lack of access to patient information such as
to compensate for ion suppression or enhance- diagnosis, dosing regimen, and medication history
ment. These modifications, along with changing to determine if patients were prescribed either
the analytical platform, improved the specificity, oxcarbazepine or eslicarbazepine. Therefore, we
accuracy, and linearity in comparison to the pre- were unable to correlate serum/plasma drug con-
vious assay. Performance of the new method on centration to drug dose. Another limitation of our
a CAP ZE proficiency testing survey was accept- study was the lack of outcome data for patients to
able, within < þ/- 2.0 SDI. determine whether seizures were controlled
Retrospective data analysis was used to evalu- when the drug concentrations were within the
ate over 1900 pediatric and adult results. Our therapeutic range or if patients displayed signs of
results demonstrated that the distribution of toxicity when serum MHD concentrations
results was similar between pediatric and adult exceeded the upper limit of the therapeutic range.
patients and the median concentrations for To mitigate these limitations, we removed results
both populations were within the range of 3– from the data set for patients who had serial
35 mg/mL. A slightly higher proportion of pediat- measurements performed to optimize therapy.
ric patients had results above 40 mg/mL. Our data suggest that MHD concentrations in se-
Elevated serum drug concentrations in pediatric rum were similar between adult and pediatric
patients may be due to the use of higher doses populations.
to provide equivalent MHD exposure to adult In summary, we have developed and validated a
patients. Previous research has documented LC-MS/MS method for the quantification of MHD
30% to 160% increased clearance in pediatric to support therapeutic drug monitoring for oxcar-
patients compared to adults, leading to higher bazepine and eslicarbazepine therapy. The assay
dose requirements than adults to achieve simi- was validated according to CLSI guidelines for
lar MHD concentrations (10, 11). Shorter elimi- mass spectrometry. The analytical method met cri-
nation half-lives in pediatric patients also teria for accuracy, imprecision, and specificity for

..................................................................................................
May 2021 | 06:03 | 637–644 | JALM 643
ARTICLE MHD Method Validation and Retrospective Analysis

interferences from matrix effects due to icterus, SUPPLEMENTAL MATERIAL


lipemia, and hemolysis and commonly prescribed
drugs, and thus will provide accurate results to Supplemental material is available at The Journal
support patient care. of Applied Laboratory Medicine online.

Nonstandard Abbreviations: AEDs, anti-epileptic drugs; MHD, monohydroxy derivative; LC-MS/MS, liquid chromatography tan-
dem mass spectrometry; AMR, analytical measurement range; LOQ, limit of quantification.

Author Contributions: All authors confirmed they have contributed to the intellectual content of this paper and have met the follow-

Downloaded from https://academic.oup.com/jalm/article/6/3/637/5961622 by guest on 13 October 2022


ing 4 requirements: (a) significant contributions to the conception and design, acquisition of data, or analysis and interpretation of
data; (b) drafting or revising the article for intellectual content; (c) final approval of the published article; and (d) agreement to be ac-
countable for all aspects of the article thus ensuring that questions related to the accuracy or integrity of any part of the article are ap-
propriately investigated and resolved.

G.M. Kroner, statistical analysis; K.L. Johnson-Davis, administrative support.

Authors’ Disclosures or Potential Conflicts of Interest: Upon manuscript submission, all authors completed the author disclo-
sure form. Disclosures and/or potential conflicts of interest: Employment or Leadership: R.L. Thomas, ARUP Laboratories.
Consultant or Advisory Role: None declared. Stock Ownership: None declared. Honoraria: None declared. Research
Funding: None declared. Expert Testimony: None declared. Patents: None declared.

Role of Sponsor: No sponsor was declared.

Acknowledgments: The authors wish to thank Dave Davis for his assistance with data collection.

REFERENCES
1. Knezevic CE, Marzinke MA. Clinical use and monitoring of drug monitoring, ILAE Commission on Therapeutic
antiepileptic drugs. J Appl Lab Med 2018;3:115–27. Strategies. Epilepsia 2008;49:1239–76.
2. Krasowski MD, McMillin GA. Advances in anti-epileptic 7. Sallas WM, Milosavljev S, D’Souza J, Hossain M.
drug testing. Clin Chim Acta 2014;436:224–36. Pharmacokinetic drug interactions in children taking
3. Italiano D, Perucca E. Clinical pharmacokinetics of new- oxcarbazepine. Clin Pharmacol Ther 2003;74:138–49.
generation antiepileptic drugs at the extremes of age: an 8. Krasowski MD. Therapeutic drug monitoring of the newer
update. Clin Pharmacokinet 2013;52:627–45. anti-epilepsy medications. Pharmaceuticals 2010;3:
4. Friis ML, Kristensen O, Boas J, Dalby M, Deth SH, Gram L, 1909–35.
et al. Therapeutic experiences with 947 epileptic out- 9. Borusiak P, Korn-Merker E, Holert N, Boenigk HE.
patients in oxcarbazepine treatment. Acta Neurol Scand Hyponatremia induced by oxcarbazepine in children.
2009;87:224–7. Epilespy Res 1998;30:241–6.
5. Rey E, Bulteau C, Motte J, Tran A, Sturm Y, D’Souza J, et al. 10. Battino D, Estienne M, Avanzini G. Clinical
Oxcarbazepine pharmacokinetics and tolerability in pharmacokinetics of antiepileptic drugs in paediatric
children with inadequately controlled epilepsy. J Clin patients. Part II. phenytoin, carbamazepine, sulthiame,
Pharmacol 2004;44:1290–300. lamotrigine, vigabatrin, oxcarbazepine and felbamate.
6. Patsalos PN, Berry DJ, Bourgeois BFD, Cloyd JC, Glauser Clin Pharmacokinet 1995;29:341–69.
TA, Johannessen SI, et al. Antiepileptic drugs- best 11. Perucca E. Clinical pharmacokinetics of new-generation
practice guidelines for therapeutic drug monitoring: a antiepileptic drugs at the extremes of age. Clin
position paper by the subcommission on therapeutic Pharmacokinet 2006;45:351–63.

....................................................................................................
644 JALM | 637–644 | 06:03 | May 2021

You might also like