Firefly

phototherapy

User Manual
MD Infant Phototherapy for Jaundice Treatment
# Firefly
UDI 111720641527 2265
Effective date: October 2023
Issue Number: 01 Version: 14 EN
Issue Date: October 2023 www.mtts-asia.com Firefly. User manual
Contents

Company Information 04

Introduction 05

Quick start guide

- Components 06

- Power supply 07

- Control panel 08

Warnings 10

Battery backup (optional) 12

Safety information 14

Cleaning and disinfection 17

Irradiance measurement 18

Irradiance measurement criteria 19

Specifications 20

Explanation of Symbols 22

Warranty Policy 23

V.1.14.EN Firefly. User manual


03
Company Information Introduction

Intended use
The Firefly Phototherapy is intended to provide phototherapy to treat infants jaundice
when a phototherapy is required and prescribed by a physician.

What does phototherapy do?


Reduces the bilirubin level in the blood through:
Photo-oxidation
MEDICAL TECHNOLOGY TRANSFER AND SERVICES Co., LTD
Configurational isomerisation
House No. 26, Alley 41, An Duong Vuong Street, Tay Ho District, Hanoi City, Vietnam
SRN: VN-MF-000036669
Structural isomerisation

Tel: +84 24 3766 6521 Phototherapy starts medical treatment immediately after it’s turned on.
Fax: +84 24 3718 8050 It emits very little heat because of the LED light source, so does not cause
Email: assistance@mtts-asia.com dehydration.
www.mtts-asia.com
Phototherapy with Firefly
Firefly double-sided LED phototherapy equips you to deliver high-intensity
phototherapy to treat this potentially hazardous condition.
Logic s.r.l.
Via Antonio Pigafetta 1
34147 Trieste, Italy
SRN: IT-AR-000012301

04 V.1.14.EN Firefly. User manual


05
Components Power Supply

3
4

5
12 V (DC)
1. Top Light
Blue light shines onto the top-side of the patient.

2. Bassinet
A removable plastic bassinet is used to carry the patient and hold the patient during
treatment.
Plug in and Power up
3. Control Panel
ON / OFF button and the display, which shows treatment time and total device hours. The connector plugs into the back of the Firefly
and then directly into the power outlet in the wall.
4. Lower Light 100 - 240 V (AC)
Blue light shines through a window onto the under-side of the patient.
Only use the cable and connector that is provided by MTTS.
5. Heat sink
Heat from the LED light source escapes from below the Firefly, ensuring the device The Firefly should only be used with the power outlets that have a
remains at a safe operating temperature. protective earth (ground) pin.

06 V.1.14.EN Firefly. User manual


07
Control Panel Control Panel

1. Plug in the Firefly.

The device recognises the connection


by illuminating every LED on the Control
Panel display.

This will be displayed for just a few


moments before returning to a blank
screen.

2. Turn on the Firefly.


C C

The display shows the “total device hours”


The “total device hours” is a multiple of 10
of what is actually displayed.

The control panel is located at the front of the Firefly device. This will be displayed for 3 seconds.
The image above shows the control panel, which consists of: This example shows;
17 500 hours

1. Power button
Turns the device ON and OFF. 3. The blue phototherapy lights will now
start to shine.
2. Display
The display will start counting the
Shows the ‘treatment time’, as well as ‘total device hours’ of Firefly usage. “treatment time”.

This example shows;


1 minute

C C
1
C
2
C
4. The phototherapy treatment will continue
until the doctor decides therapy is no
longer necessary and the device can
be switched OFF.

This example shows;


72 hours 30 minutes

08 V.1.14.EN Firefly. User manual


09
Warnings Warnings

Error messages: In the case that the Firefly has a minor error then the LED lifetime warning: The blue LEDs in the Firefly are rated to provide
error number will be displayed on the control panel. effective treatment for 60,000 hours.

If such an error occurs, this message will appear for 5 seconds after the device is
connected to the power supply but before the Firefly is switched on.

The Firefly is still safe to operate, however please contact MTTS and quote the error
number and device serial number. This error number will help the MTTS technical
team identify and solve the problem.
After turning ON the Firefly the display will show the “total device hours”
This particular example shows 60,000 hours.

C C

At 60,000 hours and anytime after 60,000 hours E04 code will be displayed.
This indicates that the LED’s have reached the recommended life span.
This message will appear for 5 seconds after the device is switched on following
total number of hours displayed for 3 seconds. After that E04 and total number of
hours will alternate 1 second each. C C

C The image above shows when exactly in the ‘start-up’ process the error message will be displayed.
C
General warning: In the rare occurrence that the Firefly experiences a more
serious error which results in the blue lights to “turn-off”, then immediately The blue phototherapy lights will turn ON and the treatment will start as normal.

C
remove the patient from the device, as they will not be receiving any The display will show the “treatment time” again.
phototherapy. An error code will be shown on the display panel, this error C will
Note: The estimated 60,000 hour life span is equal to a full 5 years of non-stop use. After this period of time the LED
most likely correct itself and the Firefly will automatically turn back on and
C
resume therapy, if not please contact MTTS and report the errorC code to MTTS.
phototherapy lights will continue to function, however treatment will be significantly less effective and the Firefly
should be replaced.

10 V.1.14.EN Firefly. User manual


11
Battery backup (optional) Battery backup (optional)
C C

Battery Backup E-MODE - extended battery operation


The battery backup allows operating the device for 2 hours without AC power supply. E-mode allows using the battery for extended period of time (additional 2 hours).
The backup activates automatically as soon as the AC power supply is off. During that time the battery is discharged further.

To check the battery charge level press (do not hold) ‘power button’. To activate the ‘E-mode’ press ‘power button’ during the 50 second low battery
‘C:XXX’ = charge (c) percentage (xxx) alarm.
The dot after the last digit indicates charging (connected to AC power source)

To protect the batteries and prolong their life time E-mode should be used only
in emergency situations.

This example shows battery charge This example shows battery charge
level 20% and charging Cnot chargingC
level 95% and

C C
When battery charge level goes down to zero (C: 0), visual and audio alarm goes off.
The message displayed is:

Treatment time: 3 sec ‘Low’: 1 sec ‘Battery’: 1 sec

The alarm continues for 50 second and there is no action, the devices powers off.
If during the duration of the alarm ‘power button’ is pressed, the alarm is muted and
the device enters the E-mode.

12 V.1.14.EN Firefly. User manual


13
Safety Information Safety Information

When using Firefly phototherapy the patient should ALWAYS be placed in the Always cover the patients eye’s with a shield or patch whenever the patients
bassinet. eye’s can be exposed to the Firefly’s light. Periodically, you may want to
ensure that the baby’s eyes are protected and infection free.
The Firefly has been designed so that the bassinet is removable.
The patient should be placed in the bassinet then the bassinet into the device Patient’s bilirubin values should be measured regularly while undergoing
to ensure patient comfort and safety. phototherapy treatment.

To accurately assess the baby’s skin colour turn off the light when checking
the baby’s condition.

Regularly monitor the patient’s temperature and fluid status.

DO NOT carry the Firefly while the patient is in the device.

DO NOT place the patient directly on the window of the lower light.
DO NOT disassemble any part of the Firefly phototherapy
DO NOT place the Firefly inside an incubator. (excluding the removal of the bassinet).

DO NOT use Firefly phototherapy with an MTTS infant warmer, or any other DO NOT place the Firefly in direct sunlight.
warmer as this will cause damage to the Firefly top light. Note: In the event of a power outage, eye coverings should be left on the patient whenever they are inside
the Firefly to protect the patient’s eyes in case the power returns.

14 V.1.14.EN Firefly. User manual


15
Safety information Cleaning and disinfection

The Firefly phototherapy should be used only by appropriately trained The Firefly bassinet should be disinfected, before and after the treatment of
personnel and under the direction of qualified medical personnel familiar with each patient. In the case that a patient undergoes treatment for more than one
currently known risks and benefits of infant phototherapy equipment use. week then be sure to disinfect the bassinet once a week.

Patients adjacent to the infant phototherapy equipment may need to be If visibly soiled, clean the affected parts with soap and water using a soft cloth.
protected with protective shields or eye protection. Then wipe with a wet cloth and dry with a soft cloth. For difficult areas use a
standard soft-bristled brush.
The operator may experience some effects during prolonged exposure to the
area irradiated by the infant phototherapy equipment.* To loosen heavy, dried-on soil, you may first need to saturate the spot with
water.
Blue light can hinder clinical observations by masking skin colour changes,
such as cyanosis. Only disinfect the bassinet with a solution containing at least 70% alcohol.
Apply the disinfectant with a clean cloth or soft sponge. Wait for the device to
An infant’s water balance may be influenced by phototherapy. Regular feeding dry after cleaning or disinfecting before use.
times may need to occur more often in order to prevent dehydration.
When cleaning the device disconnect the power cord.
Varying the ambient conditions (ambient temperature and/or different radiation
sources) may negatively affect the patient. Refer to your hospital phototherapy
policy and procedure regarding appropriate ambient conditions.

Bilirubin photoisomers may be toxic.

Do not use the Firefly if it appears that any part of the device is not functioning
properly or if any parts appear damaged. Contact MTTS.

In case of any malfunction DO NOT attempt to service this machine,


the Firefly should only be serviced by a qualified MTTS technician.

Due to photo effects, drugs and infusion liquids should not be stored in the
radiation area.

Do not use the in the presence of potentially combustible gases including


oxygen, nitrous oxide or other anaesthetics.

*Sensitive individuals may experience headache, nausea or mild vertigo if he/she stays too long in the
illuminated area. Using the Firefly in a well-lighted area or wearing glasses with yellow lenses can alleviate
potential effects. Guard Dog Bones glass es (p/n 413BB) are recommended and are available online at
www.safetyglasses.com DO NOT use alcohol to disinfect any other part of the device.

DO NOT immerse any part of the device in water or alcohol.

16 V.1.14.EN Firefly. User manual


17
Irradiance measurement Irradiance measurement criteria

Note: to ensure that the device is functioning properly, regularly monitor the Firefly in The lights will decay gradually over time, reaching approximately 70% of their
the following way. original irradiance at 60,000 hours.

Measurement instructions: Note that the irradiance should be measured when the device has reached 60,000
hours. After 60,000 hours the device should be replaced with a new one.
To test the irradiance of the Firefly measure using a lightmeter specifically designed
to measure wavelengths for phototherapy, preferably double sided such as MTTS LM-
800.
48cm
(18.89 in)

A stabilization time of 2 hours of continuous operation is required for accurate


measuring of the irradiance.
26 37 39 38 35 25
Measure each of the sections of the bassinet seen in the adjacent picture. Compare
the measurements to the expected values assuming some decay over time. 20cm
(7.87 in)
25 35 37 39 35 25

TOP LIGHT
IRRADIANCE
μW/cm2/nm

20.0 - 34.9
35.0 - 49.9
50.0 - 64.9
BOTTOM LIGHT

42 63 61 57 62 44

20cm
(7.87 in)

43 61 62 58 64 45

48cm
(18.89 in)

18 V.1.14.EN Firefly. User manual


19
Specifications Specifications
Performance Specifications Environmental Specifications

Illumination source High-power blue LEDs, 1-1.25W standard operating Operating Ambient temperature +10°C to +35°C
Peak wavelength 440 - 460 nm Humidity: 0% to 90% RH non condensing
Lamp duration 60,000 hours Atmospheric Pressure: 70-106kPa
Total spectral irradiance top light 40.0 µW/cm2/nm
bottom light 66.0 µW/cm2/nm Transport and storage Ambient temperature 0°C to +50°C
Effective surface area top light 48 cm x 20 cm Humidity: 0% to 90% RH non condensing
bottom light 48 cm x 20 cm Atmospheric Pressure: 70-106kPa
Relative local distribution top light 0.63 (IEC Compliant >0.4)
bottom light 0.64 (IEC Compliant >0.4) Exclusions Not for use inside an incubator
and or any radiant warmer
Time Totalizer Treatment time and total device hours
Measured with MTTS LM-800

Physical Specifications
Standards for Reference
Key features No moving parts
Enclosed designed toward NEMA Type 5 EN 60601-1:2006 + A1:2013: Medical electrical equipment – Part 1 General requirements for basic safety and
Dimensions (LxWxD) overall 66 cm x 38 cm x 49.5 cm 26in x 4.9in x 19.5in essential performance
top light 51.6 cm x 12.5 cm x 6.3 cm 0.3in x 4.9in x 2.5in
bottom light 64.7 cm x 38 cm x 10cm 25.5in x 15in x 3.9in EN 60601-1-2:2015: Medical electrical equipment – Part 1-2: General requirements for basic safety and es-
Total unit mass 13 kg 28.7lbs sential performance – Collateral standard: Electromagnetic compatibility – Requirements and tests

EN 60601-2-50: 2009: Medical electrical equipment – Part 2-50: particular requirements for the basic safety
and essential performance of infant phototherapy equipment
Electrical Specifications
EN 60601-1-6:2010: Medical electrical equipment – Part 1-6: General requirements for basic safety and
Power characteristics 60W, 100-240VAC, 47/63Hz essential performance – Collateral standard: Usability
Off-the-shelf external power supply International medical safety approvals
(ANSI/AAMI/EN 60601-1, UL/TUV)
Class I construction is standard (ground required)
Biocompatibility tests
100k hours MTBF
Skin sensitization as per ISO: 10993-Part 10
Energy Star Efficiency Level V compliant
RoHS 3 compliant
Cytotoxicity direct contact as per ISO 10993-Part 5
Over voltage / over current protection

Primary skin irritation as per ISO:10993-Part 10

20 V.1.14.EN Firefly. User manual


21
Explanation of Symbols Warranty Policy

This statement provides important General Terms Under this Limited Warranty, products purchased in one
Authorized Representative in the
information or highlights information that country or region may be transferred to another country
European Community This MTTS Limited Warranty gives you, the customer, or region where MTTS or its authorized service provid-
may be easily overlooked.
express limited warranty rights from MTTS, the manu- ers offer warranty service for the same product model
This statement is a warning. Not following
Device manufacturer. facturer for the duration specified on the Warranty Card. number. Warranty terms, service availability, and service
this statement could result in injury to the
Please refer to the MTTS Website for an extensive de- response times may vary from country or region to coun-
patient or operator, or damage to the device.
scription of your limited warranty entitlements. In addi- try or region. Standard warranty service response time is
Date of manufacture. tion, you may also have other legal rights under applica- subject to change due to local parts availability. If so, your
CE Marking with Notified Body Number. ble law or special written agreement with MTTS. MTTS authorized service provider can provide you with
2265
MTTS makes no other express warranty or details. MTTS will not alter form, fit, or function of this
Do not disassemble the device unless you condition whether written or oral and MTTS MTTS product to make it operate in a country for which
Serial number.
are an MTTS trained technician or have been expressly disclaims all warranties and con- it was never intended to function for legal or regulatory
instructed to by qualified personnel (exclud- ditions not stated in this limited warranty. reasons. MTTS is not responsible for any tariffs or duties
An eye patch should always be placed
ing the removal of the bassinet). to the extent allowed by local law of juris- that may be incurred in transferring the products.
on the patient before commencing
dictions outside vietnam, mtts disclaims all MTTS guarantees that the product that you have pur-
Consult instructions for use or consult the treatment.
implied warranties or conditions, includ- chased or leased from MTTS is free from defects in ma-
electronic insturctions for use ing any implied warranties of merchantabil- terials or workmanship under normal use during Lim-
Keep out of direct sunlight.
ity and fitness for a particular purpose. for ited Warranty Period. The Limited Warranty Period starts
This device contains electronic parts. Do not all transactions occurring in vietnam any on the date of purchase or lease from MTTS, or from
Liquid ingress protection - level 3 implied warranty or condition of merchant- the date MTTS completes installation. Your dated sales
dispose of it with normal waste. Dispose of
it according to local guidelines for disposal
IPX3 (spraying water). ability, satisfactory quality, or fitness for a or delivery receipt, showing the date of purchase of the
particular purpose is limited to the duration product, is your proof of the purchase or lease date. You
of eletronics. Dispose according to the WEEE DO NOT use Firefly phototherapy with
of the express warranty set forth above. some may be required to provide proof or purchase or lease as
directive in the European Union. an MTTS infant warmer, or any other countries do not allow a limitation on how a condition of receiving warranty service. You are entitled

MD
warmer as this will cause damage to long an implied warranty lasts or the exclu- to hardware warranty service according to the terms and
Medical device
the Firefly top light. sion or limitation of incidental or consequen- conditions of this document if a repair to your MTTS prod-
tial damages for consumer products. in such uct is required within the Limited Warranty Period.
Power supply polarity. countries, some exclusions or limitations of Unless otherwise stated, and to the extent permitted by

UDI Unique identification number


12V DC 5A
this limited warranty may not apply to you. for
consumer transaction, the limited warranty
local law, new MTTS product may be manufactured using
new materials and used materials equivalent to new in
12 Volts, Direct Current, 5 Ampere terms contained in this statement, except to performance and reliability. MTTS may repair or replace
the extent lawfully permitted, do not exclude, MTTS products (a) with new or previously used products
Website where a user can obtain additional
restrict, or modify but are in addition to the or parts equivalent to new in performance and reliability,
information on the medical product
Country of manufacture mandatory statutory rights applicable to the or (b) with equivalent products to an original product that
VN
sale of this product to you. has been discontinued. Replacement parts are warrant-

#
This Limited Warranty is applicable in all countries and ed to be free from defects in material or workmanship for
Model number
Temperature limit may be enforced in any country or region where MTTS or ninety (90) days or, for the reminder of Limited Warranty
its authorized service providers offer warranty service for Period of the MTTS product they are replacing or in which
the same product model number (subject to the terms they are installed, whichever is longer.
REF Catalogue number Humidity limitation and conditions set forth in this Limited Warranty)

Atmospheric pressure limitation

22 V.1.14.EN Firefly. User manual


23
Warranty Policy Warranty Policy

MTTS will, at its sole discretion, repair or replace any Limitation of Liability Limited Warranty Period Contacting MTTS
components or product that manifests a defect in materi-
als or workmanship during the Limited Warranty Period. If your mtts product fails to work as warrant- The Limited Warranty Period for this MTTS product is a If your MTTS product fails during the Limited Warranty
All component parts removed under this Limited War- ed above, the maximum liability of mtts under specified, fixed period commencing of the date of pur- Period and the suggestions in the product documentation
ranty become the property of MTTS. In the unlikely event this limited warranty is expressly limited to chase and specified on the Warranty Card. The date on do not solve the problem, you can receive support by do-
that your MTTS product has recurring failures, MTTS at the lesser of the price you have paid for the your sales receipt is the date of purchase unless MTTS or ing one of the following:
its sole discretion, may elect to provide you with (a) a product or the costs of repair or replace- your reseller informs you otherwise in writing. - Locate and contact your nearest MTTS service pro-
replacement unit selected by MTTS that is the same or ment of any hardware components that mal- vider via MTTS website:
equivalent to your MTTS product in performance or (b) function in conditions of normal use. Customer Responsibilities http://www.mtts-asia.com/support/
to give you a refund of your purchase price or lease pay- except as indicated above, in no event will mtts - Call the Technical Support Center:
ments (less interests) instead of a replacement. This is be liable for any damages caused by the prod- In order to avoid the risk of charges for issues not cov- +84 43 766 6521
your exclusive remedy for defective products. uct or the failure of the product to perform, ered by your limited warranty (issues that are not due to Before calling MTTS or an MTTS authorized service pro-
including any lost profits or savings or spe- defects in materials and workmanship on MTTS product), vider please have the following information available:
Exclusions cial, incidental, or consequential damages. you will be asked to assist MTTS as follows: - Product serial number and model name
MTTS is not liable for any claim made by a third Verify configurations, load most recent firmware, in- - Applicable error messages
MTTS does not warrant that the operation of party or made by you for a third party. stall software patches, run MTTS diagnostics and - Detailed questions
this product will be uninterrupted or error- this limitation of liability applies whether utilities;
free. mtts is not responsible for damage that damages are sought, or claim made, under this Implement temporary procedures or workarounds
occurs as result of your failure to follow the limited warranty or as a tort claim (includ- provided by MTTS while MTTS works on permanent
instructions intended for the product. ing negligence and strict product liability), a solution;
This Limited Warranty does not apply to expendable or contract claim, or any other claim. this limita- Cooperate with MTTS in attempting to resolve the
consumable parts and does not extend to any product tion liability cannot be waived or amended by problem using online chat, email, or telephone. This
from which the serial number has been removed or that any person. this limitation of liability will be may involve performing routine diagnostic procedures,
gas been damaged or rendered defective (a) as a result effective even if you have advised mtts of the installing additional software updates or patches;
of accident, misuse, abuse, contamination, improper or possibility of any such damages. this limitation Perform additional tasks as defined within each type
inadequate maintenance or calibration (if required) or of liability, however, will not apply to claims of warranty service provided by MTT and any other ac-
other external causes; (b) by operation outside the usage for personal injury. tions that MTTS may reasonably request in order to
parameters stated in the user documentation shipped this limited warranty gives you specific legal best perform the warranty support
with the product; (c) by software, interfacing, parts or rights. you may also have other rights that may
Customer is responsible for delivering the
supplies not supplier by MTTS; (d) improper site prepara- vary from country to country. you are advised
product (and all costs involved) from his lo-
tion or maintenance; (e) virus infection; (f) loss or dam- to consult applicable country laws for a full
cation to the MTTS authorized service point.
age in transit ; or (g) by modification or service by anyone determination of your rights.
other than (i) MTTS personnel, (ii) an MTTS authorized
service provider, or (iii) your own installation of end-user
replaceable MTTS or MTTS approved parts if available for
your MTTS product in the servicing country or region.

24 V.1.14.EN Firefly. User manual


25
MEDICAL TECHNOLOGY TRANSFER AND SERVICES Co., LTD
House No. 26, Alley 41, An Duong Vuong Street, Tay Ho District, Hanoi City, Vietnam
SRN: VN-MF-000036669

Tel: +84 24 3766 6521


Fax: +84 24 3718 8050
Email: assistance@mtts-asia.com
www.mtts-asia.com