TYPE PL3700
Maintenance Manual
SAS au Capital de 2 158 244 € - 444 871 933 R.C.S. Bourges - APE : 2651B
Headquarter : 9, rue Isaac Newton - 18000 Bourges - France
Level & Pressure Transmitter
Maintenance Manual
PL3700
1st Edition
Released February 2013
TABLE OF CONTENTS
2. DESCRIPTION ......................................................................................................................................4
3. MAINTENANCE .....................................................................................................................................4
3.1. Periodic control ............................................................................................................................4
3.2. Zero and Span calibration ............................................................................................................5
5. TROUBLESHOOTING ...........................................................................................................................9
2. DESCRIPTION
REFER TO TECHNICAL MANUAL NT5090
3. MAINTENANCE
RECOMMENDATION:
As calibration equipment, HONEYWELL MARINE recommends to use the
"HONEYWELL MARINE Test and Calibration kit", which includes a
vacuum/pressure generator and all necessary accessories.
SPAN
SPAN
Bw
Bw
Bu
RR
ZERO
RANGE
TAG
TAG
24v
S/N
24v
0v
0v + -
PL3700
- - Power supply
Ammeter 12 – 30 Vds
+ (28 Vdc for I.S.
+ Application)
Example:
PL3700 absolute range: 900 to 1100 mbar.
- 4.00 mA = 900 mbar
- 20.00 mA = 1100 mbar
- Span = 1100 – 900 = 200 mbar
- Actual value from atmospheric pressure = 996 mbar
- Current must be : {16 mA x (996 - 900) / 200 + 4 mA} ± 0.3%
- Calibration using test pressure equipment :
To test or calibrate absolute pressure sensor, using vacuum test pressure equipment is
recommended. Gauge test pressure equipment can be used (depending on its capability to
generated vacuum pressure).
Connect test pressure equipment.
Pressure test to apply = 900 mbar (abs. pressure) or -100 mbar (gauge).
If readings = 4.00 mA, correct.
If readings <> 4.00 mA ± 0.3%, adjust “Zero” potentiometer.
● “Span” calibration (gauge sensor application):
- Immersed installation :
Remove any liquid or pressure (inert gas) from tank.
Connect test pressure equipment
- Flange installation :
Close isolating valve.
If test flange, connect test pressure equipment into 1/8 NPT test connection.
If no test flange dismounts the transducer from isolating valve, connect test pressure
equipment with adapted test flange.
Apply full range of pressure indicated on transducer or remote box.
- Check mA readings :
If readings = 20,00 mA, correct.
If readings <> 20,00 mA ± 0.3%, adjust “Span” potentiometer.
Example:
To test or calibrate absolute pressure sensor, using vacuum test pressure equipment is
recommended. Gauge test pressure equipment can also be used.
PL3700 absolute range: 900 to 1100 mbar.
- 4,00 mA = 900 mbar
- 20.00 mA = 1100 mbar
- Span = 1100 – 900 = 200 mbar
Actual value from atmospheric pressure = 996 mbar.
- Pressure test to apply = 200 mbar -96 mbar = 104 mbar (gauge pressure) or 1100 mbar
(absolute pressure).
If readings = 20,00 mA, correct.
If readings <> 20,00 mA ± 0.3%, adjust “Span” potentiometer.
4. CALIBRATION CHANGE
4.1. General
In case of change the calibration on an existing tank (zero position or maximum tank
height or specific gravity have changed), you can use the same PL3700 only if new
calibration is still inside 50% of full range.
If it is not the case, you have to exchange to a new, adapted PL3700 (refer to table in
NT5090 § 3.1).
4.2. Calibration
● Connect test pressure equipment as explained § 3.2.
● Connect ammeter on PL3700 terminal block as explained § 3.2.
● Powering ON PL3700.
- Apply pressure requested for 4mA if any.
- Adjust “Zero” potentiometer to get 4,00 mA ± 0,3%.
- Apply pressure requested for 20 mA (example above).
- Adjust “Span” potentiometer to get 20,00 mA ± 0,3%.
- Apply again pressure of zero point, and check again if 4.00 mA still correct.
- If no, adjust again “Zero” potentiometer to get 4,00 mA ± 0,3% and repeat previews
operations until calibration become correct.
Contact:
Tel:
Fax:
Address:
L1
Hv
hz
Tank bottom
hz Zero line (4 mA)
1 2 3 4 5 6 7 8 9 10 11 12 13
PL3700
Pressure Transmitter PL3700
| | | | | | | | | | | |
2 | | | | | | | | | | | Type of measure
G | | | | | | | | | | | Gauge pressure
A | | | | | | | | | | | Absolute pressure
| | | | | | | | | | |
3 | | | | | | | | | | Range Type
P | | | | | | | | | | -200 to 1000 mbar forbidden if abs pressure (digit 2=A)
Q | | | | | | | | | | 900 to 1300 mbar forbidden if gage pressure (digit 2=G)
R | | | | | | | | | | 800 to 2000 mbar forbidden if gage pressure (digit 2=G)
S | | | | | | | | | | 0 to 20000 mbar forbidden if gage pressure (digit 2=G)
T | | | | | | | | | | 0 to 7000 mbar forbidden if gage pressure (digit 2=G)
U | | | | | | | | | | 0 to 25000 mbar forbidden if abs pressure (digit 2=A)
4 | | | | | | | | | Environment Comment
S | | | | | | | | | Non-Corrosive Product HNBR ( for PL3700 ATM )
C | | | | | | | | | Corrosive Product ( chemical ) Kalrez gasket, forbidden if digit 3=Q
| | | | | | | | |
5 | | | | | | | | Flange / process connection Comment
F | | | | | | | | Welded flange PN40 DN25 all except atm pressure (digit 3=Q)
P | | | | | | | | 1/2'' gas female connection only for manifold pressure ( digit 3=S )
0 | | | | | | | | Without flange only for atm pressure ( digit 3=Q )
Y | | | | | | | | Other welded flange On request
| | | | | | | |
6 7 8 | | | | | Cable Material and Lenght Comment
A | | | | | Polyethylen (PET), lenght in meter (min 01/max 50) all types, temp < 85°C
B | | | | | Teflon (FEP), lenght in meter (min 01/max 50) all types, high temp and corrosive atm
| | | | |
9 | | | | Electronics Housing Comment
A | | | | Polyester housing all types, for protected installation
C | | | | 316L stainless steel housing all types, for "on deck" installation
| | | |
10 | | | Flexible options and cable protection Comment
0 | | | No cable protection all types
A | | | Adaptators Kit (for flexible conduits) Only for SST box ( digit 9=C )
B | | | 1 flexible conduit (cell-electronic box) Only for polyester box (digit 9=A,B)
C | | | 2 flexible conduits Only for SST box ( digit 9=C )
D | | | 3 flexible conduits Only for SST box ( digit 9=C )
| | |
11 | | Calibration Comment
0 | | Standard calibration all types
| |
12 | Test flange option Comment
0 | Without test flange all types
|
13 Mechanical installation options Comment
0 Without any option all types
A 316L fixation kit only for atm pressure (digit 3=Q)
Handled by:
RMA #:
E-mail:
Tel: +33 (0) 248237918 RMA # only issued by Honeywell Marine factory.
Fax: +33(0) 248237902
Model # / Type #:
Serial #:
Part #: Description:
Problem description:
Required modification/
calibration/change/etc.:
By signing and returning this RMA Form I confirm that the equipment has been cleaned and decontaminated in accordance with good
industrial practices and in compliance with all regulations. This equipment poses no health or safety risk due to contamination.
If applicable, I attach corresponding International Chemical Safety Cards for the media the equipment was exposed to.
Also I agree that by returning above mentioned goods I will at least be charged with examination costs of EUR ,,,,, per item.
Name: Signature:
Department:
Date:
In order to avoid delays please ensure that all applicable fields are completed. Completed forms (one per instrument, spares can be itemized on a separate list) to be
mailed to service.marine@honeywell.com. Equipment to be sent to above mentioned "Ship To: " address with a copy of the completed form(s) on the packaging.
Do not send equipment before receiving a RMA number.