Product Information

Gas analysis
MKAS 710
Multi-component analyzer system
System solution MKAS 710

An economical standard design for all emission Regulations and standards


relevant applications
Approval regarding CEMS regulation
The multi-component analyzer system MKAS 710 is an
analyzing system designed to measure IR absorbing gases ■ Clean Air Guidelines
and oxygen. This system can be installed in many industrial
applications, e.g. for emission and ambient monitoring. ■ EMI regulations according to EN 50 081-1/EN 50 082-2

■ EU Guidelines EMI 89/336/EWC with respect to CE


Characteristics certification

■ Analyzer, sample handling and control unit are integrated ■ VDE Guideline according to EN 61 010
in one cabinet (IEC 1010-1, VD 411)

■ Protective cabinet with door window, cabinet lighting, fan ■ VDE 0100, VDE 0660 / part 500, VDE 0113
and filter or air conditioner*

■ Electrical supply and temperature regulation for Documentation


i.e. sample probe and sample line*
At time of delivery a documentation package is supplied:
■ Electrical supply and status signal processing for probe
purging ■ Operating manual for the S 710 analyzer and the
components
■ Sample gas delivery unit consisting of sample gas pump,
flow meter, regulating valve ■ Flow diagram

■ Sample gas conditioning consisting of sample gas cooler, ■ Wiring diagram


condensate pump, condensate collector, sample gas
filter ■ Terminal list

■ Test gas delivery unit consisting of solenoid valves or ■ Dimensional drawing


absorber unit respectively*
■ Parts list
■ Multi-component analyzer S 710, second multi-
component analyzer S 710 installed in series*,
NOx/NO converter*

■ Data acquisition unit DAU*

Depending on the configuration of the integrated S 710


multi-component gas analyzer, adjustments on the
MKAS 710 can be done with test gas or calibration
cuvettes.

* option
Technical data
S 710 Gas analyzer
Analyzer specifications ■ See separate product information S 710
I-10-signals ■ See separate product information S 710
Sample conditions
Sample volume ■ Max. 100 l/h
Sample temperature ■ Max. 200 °C (at cabinet inlet)
Dew point H20 ■ Approx. 65 °C
General data
Installation location requirement ■ Under a roof or with protection from a direct heating source, against high dust loads
and corrosive atmosphere
■ Not designed for installation in explosion-hazardous areas
Power supply ■ 230 V (+10/–15 %), 50 Hz, 25 A, max. 5 kVA,
external max. 3.5 VA
■ 230 V (+10/–15 %), 60 Hz, 25 A, max. 5 kVA,
external max. 3.5 VA *
■ 400 V (+10/–15 %), 50 Hz, max. 12 kVA,
external max. 7 kVA *
■ 400 V (+10/–15 %), 60 Hz, max. 12 kVA,
external max. 7 kVA
■ 115 V (+10/–15 %), 50 Hz, max. 2,5 kVA,
external max. 1.7 kVA *
■ 115 V (+10/–15 %), 60 Hz, max. 2,5 kVA,
external max. 1.7 kVA *
Ambient temperature during operation ■ +5 ... +35 °C without direct sunshine
■ +5 ... +50 °C * with built-in air conditioning unit IP 34
Transport and storage temperature ■ -20 ... +55 °C
Relative humidity ■ Class F (DIN 40 040)
75 % annual average
95 % temporary, not-condensing
Dimensions ■ Sheet steel cabinet
H x W x D 2150 x 800 x 600 mm3
■ FRP protective cabinet
H x W x D 2150 x 800 x 600 mm3
Color RAL 7032
Protection type IP 54
Weight Approx. 300 kg

* option
MEASURING GAS OUT

MEASURING GAS IN

test gas test gas test gas test gas


O rd e r d a t a

Basic version
630258- Basic version
Systems
1 Mounting of 1 analyzer
2 Mounting of 2 analyzers (in series)
0 Without converter or similar unit
1 Mounting of 1 converter or similar unit
0 Without DAU
1 Mounting 1 DAU / MEAC (max. 4 slide-in printboards)
Enclosure
2 Sheet steel, IP 54, RAL 7032, 2150 x 800 x 600 mm3
7 FRP, IP 54, 2150 x 900 x 600 mm3
Calibration
1 Automatic test gas delivery
2 Reference gas delivery calibration cuvette
Gas conditioning
1 Gas conditioning with 1 cooling path incl. pump, filter, moisture sensor
2 Gas conditioning with 2 cooling path incl. pump(s), filter, moisture sensor
External supply
0 Without controller for external heating circuits
1 Controller for 1 external heating circuit 230 VAC/3.5 kVA
(115 VAC/1.7 kVA)
2 Controller for 2 external heating circuits each 230 VAC/3.5 kVA
(115 VAC/1.7 kVA)
Electric connections
1 Electrical distribution 230 V / 50 Hz / 25 A, max. 5 kVA,
external max. 3.5 kVA
2 Electrical distribution 230 V / 60 Hz / 25 A, max. 5 kVA,
external max. 3.5 kVA
3 Electrical distribution 400 V / 50 Hz / 25 A, 3 L, N, PE,
max. 12 kVA, extern max. 7 kVA
4 Electrical distribution 400 V / 60 Hz / 25 A, 3 L, N, PE,
max. 12 kVA, external max. 7 kVA
5 Electrical distribution 115 V / 50 Hz / 25 A,
max. 2.5 kVA, external max. 1.7 kVA
6 Electrical distribution 115 V / 60 Hz / 25 A,
max. 2.5 kVA, external max. 1.7 kVA
Air conditioning
1 Fan and thermostat switch
2 Air conditioning 800 W
Option
A1 Bypass measuring gas
A2 Frost protection heating 500 W
Order number MKAS 710
630258-
Th e d i a l o g u e c o n t i n u e s .
Copy, complete and fax to +49 7641 469 1149

Company
R e p l y

Name

Job Title/Dept.

Street

ZIP, City
F a x

Phone/Fax

Industry/Field

PI 2.03 en
of Application

Yes, I would like to know more about I would like a detailed consultation with one of
the field of: your project advisors. Please arrange a meeting
for me.
Process gas analysis

Flue gas monitoring

Emission monitoring

Dust measurement

Volume flow measurement


w w w . s i c k - m a i h a k . d e

Data acquisition and evaluation

Water analysis

Liquid analysis

Level measurement

Tunnel sensors

Special measuring technology

Download more Product Informations at www.sick-maihak.de